Tuttnauer EZ-Series User and maintenance manual (Rev.U)

Operation and
Maintenance Manual
Electronic Table-Top Autoclaves
Models EZ9Plus & EZ11Plus
Cat. No. MAN205-0443001EN Rev U Manufactured by:
Tuttnauer Co. Ltd., Har Tuv Industrial zone B P.O.Box 170, Beit Shemesh 99000, IsraelTel: 972 2 9904611, Fax: 972 2 9904730 Tuttnauer U.S.A. Co, Ltd. 25 Power Drive Hauppauge, NY, 11788, USA. Tel (631) 737 4850, (800) 624 5836, Fax: (631) 737 0720
http://www.tuttnauerusa.com/
Page 1
TABLE OF CONTENTS
PARAGRAPH PAGE NO.
1. GENERAL...................................................................................................................................5
1.1 INCOMING INSPECTION.......................................................................................................... 5
1.2 WARRANTY................................................................................................ ............................ 5
1.3 WARRANTY STATEMENT ................................................................................................ ....... 6
2. SAFETY INSTRUCTIONS..............................................................................................................7
3. GENERAL INFORMATION ..........................................................................................................8
3.1. INTRODUCTION .......................................................................................................................... 8
3.2. OPERATING CONDITIONS ................................................................ ........................................... 10
3.3. SPECIFICATIONS........................................................................................................................ 10
3.4. ELECTRICAL DATA ................................................................ ..................................................... 14
3.5. ENVIRONMENTAL EMISSION INFORMATION ................................................................................... 14
3.6. CONSTRUCTION................................................................ ........................................................ 14
3.7. SYMBOL DESCRIPTION ............................................................................................................... 15
3.8. WATER QUALITY ...................................................................................................................... 15
3.9. DIRECTIVES AND STANDARDS ................................................................ ...................................... 16
3.9.1. Technical Directives........................................................................................................ 16
3.9.2. Technical Standards ....................................................................................................... 16
4. CONTROL PANEL .....................................................................................................................21
4.1. DESCRIPTION AND FUNCTIONS OF THE FRONT PANEL....................................................................... 22
4.1.1. Visual Information Center.............................................................................................. 22
4.1.2. The Control Center ......................................................................................................... 22
4.1.3. Data Center ....................................................................................................................23
4.2. DISPLAYED ERROR MESSAGES / SYMBOLS ................................................................ ..................... 24
4.3. DISPLAYED OPERATIONAL MESSAGES / SYMBOLS ............................................................................. 25
5. EZVIEW SCREENS ....................................................................................................................26
5.1. SCREENS SHOWING A SUCCESSFULLY COMPLETED CYCLE.................................................................... 26
5.2. SCREENS SHOWING ABORTED CYCLES AFTER A COMPLETED STERILIZATION STAGE ................................. 27
5.2.1. Canceled by user AFTER complete sterilization stage ................................................... 27
5.2.2. Door is open AFTER the sterilization stage has finished ............................................... 28
5.2.3. Screens showing a High Pressure Failure AFTER a completed sterilization stage .......28
5.3. SCREENS SHOWING A FAILED CYCLE: ............................................................................................. 29
5.3.1. Screens showing a failure because of a Heat Time Error ..............................................29
5.3.2. Failure due to Cancellation by user BEFORE completing the sterilization stage.......... 30
6. STERILIZATION PROGRAMS.....................................................................................................31
6.1. PROGRAM 1: UNWRAPPED INSTRUMENTS..................................................................................... 33
6.2. PROGRAM 2: WRAPPED INSTRUMENTS, POUCHES.......................................................................... 34
6.3. PROGRAM 3: UNWRAPPED DELICATE INSTRUMENTS................................ ....................................... 35
6.4. PROGRAM 4: HANDPIECES ................................................................ ......................................... 36
6.5. PROGRAM 5: CHAMBER BRITE CLEANING...................................................................................... 37
6.6. PROGRAM 6: CALIBRATION CYCLE ............................................................................................... 38
6.7. PROGRAM 7: EXTRA DRYING TIME .............................................................................................. 39
6.8. PROGRAM 7: CUSTOM A (MAY BE ALTERED BY THE USER)................................................................ . 40
6.9. PROGRAM 8: CUSTOM B (MAY BE ALTERED BY THE USER)................................ ................................. 41
7. CHECKING AND CHANGING PARAMETERS AND OTHER DATA .................................................42
7.1. QUICK OPTIONS DIRECTORY ................................................................................................ ....... 43
7.1.1. Add extra dry time .........................................................................................................43
7.1.2. Export to USB ................................................................................................................. 44
Page 2
7.1.3. Print cycles...................................................................................................................... 46
7.1.4. Version information ....................................................................................................... 47
7.1.5. Start cycle by clock .........................................................................................................47
7.1.6. Set date and time ...........................................................................................................49
7.1.7. Login ............................................................................................................................... 50
7.1.8. Exit..................................................................................................................................52
7.2. MAIN MENU ................................................................ ........................................................... 52
7.2.1. Cycle Parameters for Custom A and B cycles ................................................................ 53
7.2.2. Modifying a parameter for the Custom A and B cycles ................................................ 54
7.2.3. System Parameters ........................................................................................................56
7.2.4. Maintenance .................................................................................................................. 59
7.2.4.1. Export Gain and Offset to USB .................................................................................. 60
7.2.4.2. Reset atmospheric pressure ...................................................................................... 61
7.2.4.3. Printer Test (for units with an optional printer) .......................................................61
7.2.4.4. Print All Gain and Offset (for units with an optional printer) ..................................63
7.3. TABLE OF PARAMETERS ................................................................ ............................................. 64
8. PRINTER.................................................................................................................................. 70
8.1. PRINTER OUTPUT (FOR UNITS WITH OPTIONAL PRINTER)................................................................... 70
8.2. PRINTER HANDLING (FOR UNITS WITH AN OPTIONAL PRINTER)........................................................... 73
8.2.1. Maintenance .................................................................................................................. 73
8.2.2. Installing printer paper .................................................................................................. 73
8.2.3. Notes on treatment of thermal papers: ........................................................................ 75
9. INSTALLATION INSTRUCTION .................................................................................................. 77
9.1. LIFTING AND CARRYING.............................................................................................................. 77
9.2. PLACING ................................................................................................................................. 77
9.3. ELECTRICAL ................................................................ ............................................................. 77
9.4. SETUP .................................................................................................................................... 78
9.5. FILLING THE MINERAL-FREE WATER RESERVOIR. ............................................................................ 78
10. PREPARATION BEFORE STERILIZATION ...................................................................................81
11. OPERATING INSTRUCTIONS..................................................................................................... 86
11.1. TURNING ON THE AUTOCLAVE ..................................................................................................... 86
11.2. PRE-HEATING (EZ11PLUS ONLY) ................................................................................................ 86
11.3. OPENING THE DOOR.................................................................................................................. 87
11.4. ADDING ADDITIONAL DRYING TIME................................................................ ............................... 88
11.5. LOADING ................................................................................................................................ 88
11.6. UNLOADING ................................................................ ............................................................ 92
11.7. STOPPING THE PROCESS MANUALLY.............................................................................................. 92
11.8. STOPPING THE PROCESS DUE TO CYCLE FAILURE............................................................................... 93
12. MAINTENANCE INSTRUCTIONS ...............................................................................................95
12.1. PREVENTIVE AND SCHEDULED MAINTENANCE ................................................................................ 95
12.1.1. Daily........................................................................................................................... 95
12.1.2. Weekly by the operator ............................................................................................ 95
12.1.3. Periodically ................................................................................................................ 99
12.2. DRAINING THE RESERVOIR........................................................................................................ 100
12.3. REPLACING THE DOOR GASKET.................................................................................................. 101
12.4. CHECKING THE SAFETY VALVE ................................................................................................ ... 104
12.5. CLEANING THE WATER OUTLET STRAINER................................ ..................................................... 105
12.6. WATER SENSOR CLEANING....................................................................................................... 106
12.7. REPLACING THE HEPA AIR FILTER ............................................................................................. 107
12.8. EMERGENCY DOOR OPENING..................................................................................................... 108
12.8.1. Relieve pressure from inside the chamber .............................................................109
12.8.2. Opening the door .................................................................................................... 110
Page 3
13. TROUBLESHOOTING..............................................................................................................111
14. SPARE PARTS LIST ................................................................................................................. 117
15. ACCESSORIES ........................................................................................................................117
Page 4
TABLE OF CONTENT (Cont.)
DRAWINGS PAGE NO.
FRONT VIEW – EZ9PLUS ........................................................................................ 17
REAR VIEW – EZ9PLUS ........................................................................................... 18
FRONT VIEW – EZ11PLUS ......................................................................................19
REAR VIEW – EZ11PLUS......................................................................................... 20
POUCH RACK AR910 ...........................................................................................117
TRAY HANDLE CMT240-0097............................................................................. 117
TRAY TRY254-0003............................................................................................... 118
TRAY HOLDER FOR EZ11PLUS TRH411-0021 ..............................................118
Page 5
1. GENERAL
Read the Operating Instructions carefully, before beginning any operation on the autoclave!
1.1 Incoming Inspection
Upon receiving your Tuttnauer Autoclave, carefully inspect the outside of the shipping carton for signs of damage. If any damage to the carton is found, note the location with respect to the autoclave and check that area of the autoclave carefully once it is fully unpacked. Observe packing method and retain packing materials until the unit has been inspected. Mechanical inspection involves checking for signs of physical damage such as: scratched panel surfaces, broken knobs, etc.
If any damage is found, contact your dealer as soon as possible so that they can file a claim with the shipping carrier and also notify Tuttnauer.
All Tuttnauer products are carefully inspected prior to shipment and all reasonable precautions are taken in preparing them for shipment to assure safe arrival at their destination.
Note: Lifting and carrying should always be done by two people.
1.2 Warranty
Tuttnauer warrantees, from the date of purchase, all new EZPlus autoclaves for a period of two full years, covering both parts and labor. This two year warranty covers defects in materials and workmanship on every part in the autoclave except door gaskets and HEPA filters (they are considered wear items). Tuttnauer warrantees the chamber for a period of ten (10) years against defects in materials and workmanship.
This warranty does not include installation or operator instruction which are covered in this manual for your convenience or which can be provided by your dealer.
This warranty does not apply to any instrument that has been subjected to improper use or accident, nor shall it extend to autoclaves that have been repaired or altered outside the factory without prior authorization from Tuttnauer.
The warranty also does not include routine cleaning or preventive maintenance, to be performed according to instructions in Sec. 12.1 (Preventive and Scheduled Maintenance).
Page 6
Tuttnauer’s obligation is limited to the repair or replacement of parts for the autoclave. No other warranties or obligations are expressed or implied.
The Autoclave should only be used in a manner described in this manual!
1.3 Warranty Statement
To activate the warranty, the registration card must be completed and mailed or faxed to Tuttnauer within fourteen (14) days of purchase or you may call our customer service department at the number listed below. Products will only be received and accepted for repair from an authorized dealer and only with prior return authorization from Tuttnauer. All transportation charges to and from Tuttnauer must be paid by the owner of the autoclave. Tuttnauer will not accept COD shipments. If repairs are needed during the first 90 days after purchase of this autoclave and a local authorized service dealer is not available, Tuttnauer will arrange pick up of the unit at Tuttnauer’s expense. This will be on an individually evaluated basis and ONLY with pre-approval. Note: If you have any questions or there are any difficulties with this instrument and the solution is not covered in this manual, please contact your dealer or Tuttnauer USA Co. Do not attempt to service this instrument yourself.
Tuttnauer USA Co., Ltd., 25 Power Drive Hauppauge, NY 11788, USA : (800) 624 5836, (631) 737 4850, Fax: (631) 737 0720 e-mail:info@tuttnauerUSA.com.
Page 7
2. SAFETY INSTRUCTIONS
The autoclave has unique characteristics. Please read and understand the operation instructions before first operation of the autoclave. This manual includes instructions and guidance provided by the manufacturer: how to operate the autoclave, the door safety mechanism, and the dangers involved in circumventing safety means, how to ensure that the door is closed, and how to select a correct sterilization program.
Make sure that you know where the main power switch is located.
Autoclave maintenance is crucial for the correct and efficient function of the device.
Never use the autoclave to sterilize corrosive products, such as: acids, bases and phenols, volatile compounds or solutions such as ethanol, methanol, chloroform or radioactive substances. Below are the operating instructions – safety instructions:
1. All autoclave users must receive training in proper usage from an experienced employee. Every new employee must undergo a training period under an experienced employee.
2. When sterilizing plastic materials, refer to the item manufacture information and make sure that the item can withstand the sterilization temperature. Plastic that melts in the chamber is liable to cause a great deal of damage.
3. On closing the device door, make sure it is properly locked before activating. Verify that the DOOR OPEN symbol is replaced by “System Ready”.
4. When withdrawing trays, use the enclosed tray handle or wear heat resistant gloves.
5. The door is electronically locked and will not open unless the pressure in the chamber equals the atmospheric pressure (chamber pressure is displayed on the screen).
6. The door also remains locked if there is a cycle failure or the unit has no power.
7. Open the door slowly to allow steam to escape.
8. Once a month, ensure that the safety valve is operating, and once every 5 years a certified inspector must perform a chamber pressure safety test.
9. Make sure there are no leaks, breaks, blockages, whistles or strange noises.
10. Perform maintenance operations as instructed. The owner of the autoclave is responsible to perform the maintenance operations.
11. Notify the person in charge immediately of any deviation from the proper function of the device.
12. Protective equipment and clothes and other safety instructions should be implemented in accordance with local and national regulations and/or rules!
Page 8
3. GENERAL INFORMATION
3.1. Introduction
The EZPlus tabletop autoclave is designed for sterilization of wrapped and unwrapped instruments, and related items found in dental, medical, and veterinary clinics, first aid rooms, hospitals, laboratories etc.
This autoclave is an electrically – heated sterilizer using steam
as the sterilizing agent. This unit uses steam flush pressure
pulse technology for removing air from the chamber. A
computerized control unit ensures a fully automatic sterilization
cycle, control and monitoring of physical parameters and a clear
documentation of the sterilization cycle.
The autoclave offers a choice of four automatic programs and
two custom programs designed to match the material to be
sterilized. In addition there is a dedicated cleaning program. The
autoclave is equipped with an Air Assisted Drying system. This
includes an air pump that during the drying stage draws air
through a HEPA filter (0.2µm) and circulates that air through the
heated chamber to remove moisture and facilitate the drying
operation. Drying is performed with the door closed.
On all models, a water pump is installed between the water
reservoir and the chamber. This pump guarantees fast and
accurate filling of the chamber every time. Entry of water may be
accompanied by a noise for approximately 30 seconds. This is
normal noise generated by the regular operation of the pump.
The EZPlus series features a digital absolute pressure display for monitoring and control purposes. The pressure display indicates if there is pressure in the chamber. The device is capable of displaying the pressure in psia, psig, or in kPa according to the operator’s requirements. When the pressure is displayed in psig, the atmospheric pressure is shown (at sea level) as 0 psig. If the pressure is defined in psia or kPa the absolute zero is displayed as “0” and the atmospheric pressure is shown (at sea level) as 14.7 psia or 100 kPa respectively.
Note: This unit comes from the factory with the pressure parameter set to display in psig.
The EZPlus can display temperature in ºF or ºC.
Note: The unit comes from the factory set to display temperature in ºF.
Page 9
The control system is designed to meet the most current sterilization standards to ensure efficacy, safety of personnel and reliable operation. See sec. 3.9
A printer is an optional addition to the autoclave. The printer prints the preset and actual parameters of the cycle (temperature, time and pressure).
The EZPlus features built in memory to record up to 100 sterilization cycles. These can be reprinted on the optional printer or exported to a USB device to be transferred to a PC.
The EZPlus has a built in Network Port for use with Tuttnauer’s R.PC.R software when connected to your local network.
The R.PC.R software has been developed especially for the Tuttnauer autoclaves.
This application will allow you to:
Monitor up to 8 autoclavesMonitor the real time activity of any autoclave connected via
the network port.
Manage the history files of the cycles run on your autoclave.
The history files can be downloaded either directly through a physical connection to the network port or transferred manually using a USB device.
Store all the history of the processes that have been run on
your autoclave
Track the parameter setting that have been used in each of
the cycles and stages run.
Choose the style of the report to view; either graph, table, or a
print out
All reports can be saved as a PDF.The graph style report offers the user an option to customize
the inputs and outputs used in the presentation.
For more information on the R.PC.R refer to the R.PC.R user guide.
This manual is intended for the user and gives the user a general understanding of the instrument and the best ways to operate and take care of it in order to obtain effective results.
Before operating this autoclave read carefully this operation manual. After reading this manual, operating the autoclave will be easy. However since this instrument is built with high technology sensitive components, no attempt should be made by the user or any other unauthorized person to repair or recalibrate it.
Page 10
Only technical personnel, of an authorized Tuttnauer dealer, having proper qualifications and holding technical documentation (including a technician manual) and adequate information are authorized to service the apparatus.
3.2. Operating Conditions
This device is for indoor use only! The minimum counter depth for the EZ9Plus and EZ11Plus is 22
inches. Counter tops or slide-outs need to have a minimum capacity of
175 pounds for the EZ9Plus and 190 pounds for the EZ11Plus. The sterilizer should be loaded only with autoclavable material! Minimum room ventilation shall be 10 cycles per hour. The environment shall not exceed an ambient temperature range
of from 41ºF (5ºC) to 104ºF (40ºC) and a relative humidity of 85%.
The operational altitude shall not be over 6562 feet (2000 meters) (ambient pressure shall not be lower than 11.6 psia (80 kPa)).
Operate the autoclave only in the manner specified in the manual. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
CAUTION!
Waste water should be brought into the public net in
accordance with the local rules or requirements
ONLY NON-HAZARDOUS LIQUIDS SHALL BE DISPOSED IN
PUBLIC SEWAGE!
3.3. Specifications
Property
Value
9"
11"
Chamber Dia.
9” (230 mm)
11” (280 mm)
Depth
19.8” (504 mm)
19.8” (504 mm) Chamber volume
5.2 gal (19.8 lit)
7.5 gal. (28.5 lit)
External dimensions
Width
19.1”(485 mm )
20.9” (530 mm)
Height
15.1” (384 mm)
17.3” (440 mm)
Depth
24.8” (630 mm)
24.8”(630 mm)
Maximum dimensions (door open)
36.2”(920 mm)
37.8”(960 mm)
Page 11
Property
Value
9"
11"
Distance between supporting legs
Between front and rear legs
16.3” (415 mm)
16.3” (415 mm) Front legs
10.5” (266 mm)
12.4” (315 mm)
Rear legs
13.8” (351 mm)
15.7” (400 mm)
Weight
94 lbs. (43 kg)
110 lbs. (50 kg)
Weight w/max solid load and full reservoir
114.6 lbs. (52 kg)
136 lbs. (62 kg)
Shipping weight
106 lbs. (48 kg)
126lbs. (57 kg)
Weight per support area (max. load)
Counter tops or slide-outs need to have a minimum capacity of 175 pounds for the EZ9Plus and 190 pounds for the EZ11Plus.
Minimum counter depth
22 inches
22 inches
Shipping dimensions
Width
25” (63.5cm)
25.5” (64.7 cm)
Height
19” (48cm)
22” (55cm)
Depth
29” (75 cm)
29” (75 cm)
Volume
7.97 ft3 (0.23 m3)
9.4 ft3 (0.267 m3)
Mineral-free water reservoir
Max. water volume
1.375 US gal (5.2 lit.)
Min. water volume
0.4 US gal (1.5 lit.)
Maximum solid load per item
1.1 lbs. (0.5 kg)
1.1 lbs. (0.5 kg)
Maximum solid load per tray
3.0 lbs (1.4 kg)
3.0 lbs (1.4 kg)
Maximum solid load
9.0 lbs (4.1 kg)
15 lbs (6.8 kg)
Maximum textile load
1.5lbs (0.7kg)
1.7lbs (0.8)
Tray dimensions
W
6.7” (169 mm)
6.7” (169 mm)
H
0.6” (16 mm)
0.6” (16 mm)
D
16.3” (413 mm)
16.3” (413mm)
Max. Allowable Working pressure (MAWP)
40 psi (2.8 bar)
No. of trays
3
5
Cassette capacity based on 8”x11” and 8”x 5.5” Miltex Thompson cassettes
2 full & 2 half Loaded horizontally
4 full & 4 half
Loaded vertically
Page 12
Overall Dimensions – EZ9Plus
Page 13
Overall Dimensions – EZ11Plus
Page 14
3.4. Electrical Data
Model
Specifications
9"
11"
EZ
EZ
Total Power
1400W
1400W
Voltage
120VAC ±5%, 60Hz / 1 ph*
120VAC ±5%, 60Hz / 1 ph*
Amperage
12A
12A
Recommended circuit breaker
15A
15A
Protection against electrical shock
Class I (IEC 60601-1) (GFCI)
Mains supply fluctuation
+/- 5%
Degree of protection by enclosure
IP31
Electrical Circuit
Dedicated electrical circuit
* According to the local network.
Note: In order to avoid any injury by electrical hazard, it is mandatory that a ground fault protection device (GFCI) be installed in the electrical panel feeding the autoclave (local codes may make this mandatory).
Attention:
The electrical net must be protected with a current leakage
safety relay (GFCI).
The electrical network must comply with local rules or
regulations.
Verify that there is an easy access to the main power switch
and to the current leakage safety relay (GFCI).
Surge protection is recommended in areas that experience
large voltage or ground fluctuations.
It is recommended that the autoclave be installed on a
dedicated line.
3.5. Environmental Emission Information
1. The peak sound level generated by the autoclave is 65dBa
with background noise of 48 dBa.
2. The total heat per hour transmitted by the autoclave is
<200Wh.
3.6. Construction
The main parts of the autoclave are made of materials as indicated below:
Chamber is built of stainless steel 316 L.
Door is made of stainless steel 304.
Trays are made of stainless steel 304.
Page 15
Water reservoir is made of polyethylene.
Door handle and door cover are made of hard plastic
material, which is safe to touch and thermo-insulated.
3.7. Symbol Description
3.8. Water Quality
Physical characteristics and contaminants levels
The distilled or mineral – free water supplied to the autoclave should have the physical characteristics and maximum acceptable level of contaminants indicated in the table below:
Physical Characteristics and Maximum acceptable
contaminants levels in steam for sterilizers
(According to EN 13060:2004).
Element
Condensate – allowable content
Silicon dioxide SiO2
≤0.1 mg/kg
Iron
≤0.1 mg/kg
Cadmium
≤0.005 mg/kg
Lead
≤ 0.05 mg/kg
All other metals except iron, cadmium, lead
≤0.1 mg/kg Chloride (Cl)
≤0.1 mg/kg
Phosphate (P2O5)
≤0.1 mg/kg
Conductivity (at 20°C)
≤3
μs/cm
pH value (degree of acidity)
5 to 7
Appearance
Colorless clean without sediment
Hardness (Σ Ions of alkaline earth)
≤0.02 mmol/l
Caution! Consult accompanying documents
Caution! Hot surface. Caution! Hot steam.
Protective earth (Ground)
Page 16
Compliance with the above data should be verified by testing in accordance with acknowledged analytical methods, by an authorized laboratory.
Attention: We recommend testing the water quality once a month. The use of water for autoclaves that does not comply with the table above may have severe impact on the working life of the sterilizer and can invalidate the manufacturer’s guarantee.
3.9. Directives and Standards
Every autoclave meets the provisions of the following Directives and is in compliance with the following Standards:
3.9.1. Technical Directives
Medical Device Directive 93/42/EEC
Pressure Equipment Directive 97/23/EEC
3.9.2. Technical Standards
USA Standards:
ANSI/AAMI ST55:2010
ASME Sec. VIII. Div 1
UL 61010-1
UL 61010-2-040
FDA 510(K) Cleared
ISO 17665-1:2006
The Quality System complies with ISO 13485:2003
and ISO 9001:2008
Canada Standards: CMDR
Page 17
FRONT VIEW – EZ9Plus
No.
Description
No.
Description
1
Reservoir drain
8
Safety valve
2
Door opening grip
9
EZView display
3
Door micro-switches
10
EZPad keypad
4
Door cover
11
Autoclave On/Off switch & circuit breaker
5
Water reservoir fill & level gauge
12
Printer cover 6
Autoclave cover
13
Printer (optional)
7
Mineral-free water reservoir cover
14
USB port
13
2
1
5
4
3
9
10
12
6
7
8
14
11
Page 18
REAR VIEW – EZ9Plus
No.
Description
1
Main power electric cable socket
2
Cut-off thermostat
3
Network port
4
Ventilation grills
5
Mineral-free water reservoir cover
6
Autoclave cover
7
HEPA filter cover
8
Ventilation grills
9
Water outlet strainer
5
4
2
1
6
7
8
3
9
Page 19
FRONT VIEW – EZ11Plus
No.
Description
No.
Description
1
Reservoir drain
8
Safety valve
2
Door opening grip
9
EZView display
3
Door microswitches
10
EZPad keypad
4
Door cover
11
Autoclave On/Off Switch & circuit breaker
5
Water reservoir fill & level gauge
12
Printer cover 6
Autoclave cover
13
Printer (optional)
7
Mineral-free water reservoir cover
14
USB port
13
2
1
5
4
3
9
10
12
6
7
8
14
11
Page 20
REAR VIEW – EZ11Plus
No.
Description
1
Main power electric cable socket
2
Cut-off thermostat
3
Network port
4
Ventilation grills
5
Mineral-free water reservoir cover
6
Autoclave cover
7
HEPA filter cover
8
Ventilation grills
9
Water outlet strainer
6
5
4
7
2
1
9
3
8
Page 21
4. CONTROL PANEL
No.
Description
1
EZView display
2
Main switch and circuit breaker
3
EZPad keypad
4
USB port
5
Printer (optional)
1
2
3
4
5
Visual Information
Center
Data Center
Control Center
Page 22
4.1. Description and Functions of the Front Panel
The front panel is composed of 3 sections (see picture on previous page):
1. Visual Information Center
2. Control Center.
3. Data Center
4.1.1. Visual Information Center
The information center contains the EZView display which is an LCD panel used to display the current status of the autoclave and any Operational Messages or Error Messages.
4.1.2. The Control Center
The Control Center contains the ON/OFF switch/Circuit Breaker and the EZPad keypad.
The keypad consists of 3 keys as described below:
UP key
This key has the following functions: In the main screen:
o This key enables the operator to browse through the
cycles.
In the menu directories:
o When the cursor is blinking on a number, the UP ▲ key
increases its value.
o When the cursor is blinking on a menu selection, the UP
Program description: Program icon and name
Chamber temperature
Chamber pressure
System statusName of stageError and operational messages
Program Parameters:
Sterilization temperature
Sterilization time Drying Time
Page 23
▲ key allows browsing backward through the menu.
o When adjusting a parameter and the cursor is blinking on
SET or EXIT the UP ▲ key activates that procedure.
DOWN key
This key has the following functions: In the main screen:
o This key enables the operator to browse through the
cycles.
In the menu directories:
o When the cursor is blinking on a number, the DOWN
key decreases its value.
o When the cursor is blinking on a menu selection, the
DOWN ▼ key allows browsing forward through the menu.
o When adjusting a parameter and the cursor is blinking on
SET or EXIT the DOWN ▼ key activates that procedure.
START/STOP key
This key has the following functions: In the main screen:
o Starts the process when the required program was
chosen.
o Stops the current process. o Cancels the ERROR message displayed on the screen
and opens the electric door lock.
In the menu directories:
o When the cursor is blinking on a number, the
START/STOP key enables moving to the next position.
o When the cursor is blinking on a menu selection, the
START/STOP key activates that selection.
4.1.3. Data Center
The Data Center contains a USB port, a network port and an optional printer.
The USB port can be used to upload or download
software and settings and download cycle history for transferring to a PC for storage or printing.
The network port (located on the rear of the unit) can be
used to connect to a local network and download information to Tuttnauer’s R.PC.R software.
The RPCR software has been developed especially for the Tuttnauer autoclaves and is an excellent report generating tool.
This software will allow you to:
Monitor up to 8 autoclavesMonitor the real time activity of any autoclave connected via
the network port.
Page 24
Manage the history files of the cycles run on your autoclave.
The history files can be downloaded either directly through a physical connection to the network port or transferred manually using a USB device.
Store all the history of the processes that have been run on
your autoclave
Track the parameter settings that have been used in each of
the cycles and stages run.
Choose the style of the report to view; either graph, table, or a
print out
All reports can be saved as a PDF.The graph style report offers the user an option to customize
the inputs and outputs used in the presentation.
For more information on the R.PC.R refer to the R.PC.R user guide.
The printer is an optional device. It prints the detailed history of
each cycle performed by the autoclave. The printing is on thermal paper with 24 characters per line and records the sterilization cycle information for subsequent consideration.
4.2. Displayed Error Messages / Symbols
An error message is displayed when a failure occurs. The failures are divided into two categories.
1. Failures that occur before completing the sterilization stage,
which in this case will leave the load unsterilized
2. Failures that occur after completing the sterilization stage,
which in this case will leave the load sterilized
For the list of Displayed Error Messages / Symbols see sec. 13 Troubleshooting
Page 25
4.3. Displayed operational messages / symbols
Operational messages tell you the status of the machine before or after a cycle.
Message / Symbol Name
Message / Symbol Description
Required Action
This symbol is displayed when the door is open.
Close the door.
"Door is open"
This message is displayed in stand-by when the door is opened and the START/STOP key is pressed.
Close the door to perform a new cycle. If the problem persists, call the technician.
This message is displayed if the electrode in the chamber senses water.
DO NOT open the door, water will spill out. Run a new cycle to drain the chamber.
This symbol is displayed when there is no water in the mineral-free water reservoir.
Pour water in the front funnel until it reaches the full level.
"Cycle Ended"
This message is displayed when the cycle has ended successfully.
Open the door. The Instruments are ready to be removed.
This symbol is displayed when Cycle by Clock mode is active.
Enter the Quick Options menu as described in this manual to change the time or to cancel this option.
"Start cycle by clock is active”
This message is displayed if the user presses START/STOP key while the "start cycle by clock" mode is active. Starting another cycle is not allowed.
Enter the Quick Options menu as described in this manual to change the time or to cancel this option.
"Atmospheric pressure not set"
This message is displayed when the ATM needs to be set.
Opening the door for 2 minutes will allow the Atmospheric pressure to be set automatically.
"Please restart machine in order for changes to be updated"
Changes to the system software require that the autoclave be restarted.
Restart the autoclave in order for changes to be updated.
Page 26
5. EZView SCREENS
During the cycle the EZView screen will change to let you know how the cycle is progressing. The following screens are a representation of what will be seen during a cycle.
5.1. Screens showing a successfully completed cycle
1. System Ready
2. Pre-heating
3. Starting
4. Water Filling
5. Steam Flush
6. Pulse L
7. Pulse H
8. Heating
9. Sterilization
10. Exhaust
11. Drying
12. Ending
Page 27
13. Cycle Ended (successful cycle)
5.2. Screens showing aborted cycles AFTER a completed sterilization stage
The sterilization phase ended successfully – the cycle was aborted and the reason for the failure is displayed. When the sterilization portion of the cycle is successful the EZView display remains white even though the cycle was aborted. Note: There is a mandatory 1 minute of drying at the end of any aborted cycle. The next three scenarios show examples of possible error messages:
5.2.1. Canceled by user AFTER complete sterilization stage
The sterilization stage ended successfully, however the operator manually aborted the remainder of the cycle, by pressing the START/STOP key. This resulted in the following sequence of screens showing the reason for the aborted cycle.
Page 28
Note: The user will have to press the START/STOP key, after the mandatory 1 minute drying, to clear the message and unlock the door.
5.2.2. Door is open AFTER the sterilization stage has finished
The sterilization stage ended successfully, however the door switch indicated that the door was opened. This resulted in the following sequence of screens showing the reason for the aborted cycle.
Note: The user will have to press the START/STOP key, after the mandatory 1 minute drying, to clear the message and unlock the door.
5.2.3. Screens showing a High Pressure Failure AFTER a completed sterilization stage
The sterilization stage ended successfully, however the chamber indicated that there was high pressure during the exhaust phase. This resulted in the following sequence of screens showing the reason for the aborted cycle.
Note: The user will have to press the START/STOP key, after the mandatory 1 minute drying, to clear the message and unlock the door.
Loading...
+ 90 hidden pages