Tuttnauer E-Series Repair manual (2004)

tuttnauer USA Co.

Sterilization & Infection Control Products

Technicians Service and
Repair Manual

for the

“E” Series Tuttnauer
This manual is intended for the qualified technician. The instructions and guidance go into great detail, but basic trouble shooting and diagnostic skills are still required.
I want to thank all the members of the technical staff at Tuttnauer USA with out whose help this manual could not have been completed
Written by Henry Dierschke
Illustrated by Edward A. Hendrickson
Revised Edition November 18, 2004
First Edition June 1, 2000
Section

Table of Contents

1 General Information

1.1 Tuttnauer Company
1.2 Warranty
1.3 Theory of Operation

2 Installation and Setup

2.1 Unpacking & Inspection
2.2 Location of Unit
2.3 Voltage Requirements
2.4 Setup

3 Front Panel Keypad

3.1 Front Panel Keypad

4 Display Messages

4.1 Display Messages

5 Operating Instructions

5.1 Preparation before Sterilizing
5.2 Filling the Reservoir
5.3 Sterilization Programs
5.4 Operating Instructions
5.5 Standard Sterilization Times

6 Maintenance

6.1 Cleaning
6.2 Door Assembly
6.3 Closing Device
6.4 Solenoid Valves
6.5 Air Jet
6.6 Water Sensing Electrode
6.7 Chamber
6.8 Dry Pump Filter
6.9 Safety Relief Valve
6.10 PC Board Maintenance
6.11 Filter Maintenance

7 Trouble Shooting

7.1 Power On Problem
7.2 Heat Up Problem
7.3 Pressure Problem
7.4 Temperature Problem
7.5 Keyboard Problem
7.6 Overheating Problem
7.7 Safety Thermostat Problem
7.8 Circuit Breaker or Fuse Problem
7.9 Dry Pump Problem
7.10 Drying Problem
7.11 Sterilizing Problem
7.12 Fan Problem
7.13 Unit Gives Off Odor
7.14 Door Problem
7.15 Water Filling Problem
7.16 Leaking Water Or Steam
7.17 Air Outlet Valve Problem
7.18 Display Problem
7.19 Power Supply Problem
7.20 Control Problem
7.21 Printer Problem
7.22 Add Water Indicator Problem
7.23 Door Closed Indicator Problem
7.24 Memory Problem
7.25 Exhaust Problem
7.26 Cut-Out Thermostat Problem
7.27 Water Pump Problem

8 Testing and Calibration

8.1 SSR (Solid State Relay)
8.2 Heating Elements
8.3 Solenoid Valves
8.4 Temperature Sensor
8.5 Pressure Sensor
8.6 Automatic Water Fill
8.7 Water Sensing Electrode
8.8 Safety Thermostat
8.9 Cut-Out Thermostat
8.10 Dip Switch Settings
8.11 Test Equipment and Use
8.12 Finding the Date Code
8.13 In – Out Test

9 Tables and Diagrams

9.1 Ohm and Amp Readings
9.2 Digital Predg board
9.3 Ajunc 2 board
9.4 Ajunc 3 board
9.5 Test Points for Ajunc 2
9.6 Test Points for Ajunc 3
9.7 Maximum Allowable Instrument Loads
9.8 Maximum Allowable Liquid Loads
9.9 Ajunc Board Schematics
9.10 Power Supply
9.11 LM34 Cross Reference Table
9.12 Solenoid Valve Layout
9.13 Autoclave Components Rear View
9.14 Autoclave Electronic Box Components
9.15 Chamber Brite Instructions

10 Replacement

10.1 Heating Elements
10.2 Safety Thermostat
10.3 Temperature Sensor
10.4 Pressure Sensor
10.5 Power Supply
10.6 Closing Device
10.7 Solenoid Valve
10.8 SSR (Solid State Relay)
10.9 Air Jet
10.10 Water Sensing Electrode
10.11 Door Bellows
10.12 Chamber
10.13 Printer
10.14 Dry Pump
10.15 Safety Relief Valve
10.16 Float Switch
10.17 Power Transistor
10.18 Ajunc Board
10.19 Digital Predg Board
10.20 Door Assembly
10.21 Fan Replacement
10.22 Water Pump Replacement
10.23 Fuse and Fuse Holder
10.24 Circuit Breaker

11 Component Function in the Autoclave

11.1 Safety Thermostat
11.2 Cut-Out Thermostat
11.3 Air Jet
11.4 Water Sensing Electrode
11.5 Door Bellows
11.6 Air Outlet Valve
11.7 Printer
11.8 Fuse and Circuit Breaker

1 General Information

1.1 The Tuttnauer Company
The Tuttnauer Company founded in 1925 produces infection control equipment for the Dental, Medical,Veterinary and Laboratory markets. In addition Tuttnauer produces large walk-in units for industrial, commercial and hospital applications.
Tuttnauer equipment is distributed worldwide and the Tuttnauer Company is considered a
leader in the field of Infection Control Apparatus
Our main product line consists of manually operated and automatic sterilizers. The following list shows past as well as currently avaliable models of Tuttnauer sterilizers.
Manually operated models: M = Manual MK = Manual Kwiklave*
Chamber sizes: 7” x 12”; 9” x 18”; (diameter x depth) 10” x 18”; 15” x 20”
and 15” x 27”
Automatic models:
E = Electronic EA & EZ = Electronic with Air Assisted Drying EK = Electronic Kwiklave EKA & EZ10k = Electronic Kwiklave with Air Assisted Drying*
Chamber sizes: 7” x 12”; 9” x 18”; (diameter x depth) 10” x 18”; 15” x 20”
and 15” x 27”
Tuttnauer offers a wide variety of standard models of autoclave, as well as custom
designed units.
Additional Tuttnauer products,
Chamber Brite autoclave clean er
Clean & Simple ultrasonic enzymatic cleaning solution in tablet form
Ultrasonic Cleaners - 1 & 3 gallon
Water Distillers – 1gal, 3.5gal, 8gal and 12gal
* Kwiklave units have faster cycle times then a standard unit, while maintaining standard sterilization exposure times.
1.2 Warranty
Tuttnauer’s warranty covers defects in materials and workmanship on every part in the
autoclave. For exact details see a formal copy of the Warranty Policy or call Tuttnauer at 1 800 624-5836.
This warranty for new autoclaves covers both parts and labor.
Tuttnauer warrantees chambers (on select models) for a period of ten (10) years against any defects in materials and workmanship. This chamber warranty went into effect January 1997. (for more details call 1 800 624-5836)
These warranties do not apply to any improper installation or application; nor shall it extend to products which have been altered outside the factory without prior authorization from Tuttnauer; nor to products which have been improperly maintained.
No product will be received or accepted for repair without proper return authorization from Tuttnauer. All transportation charge to and from Tuttnauer are the responsibility of the owner of the autoclave. During the first 30 days after purchasing a new autoclave Tuttnauer will pay shipping costs on an individually evaluated basis and ONLY with pre approval.
This warranty will be void if the unit is not purchased from an authorized Tuttnauer dealer.
To activate the warranty, the registration card must be completed and returned to Tuttnauer within fourteen (14) days of purchase or you may call customer service at 1 800 624-5836.
Tuttnauer’s obligation is limited to repair or replacement of parts for the autoclave.
No other warranties or obligations are expressed or implied.
1.3 Theory of Operation
Theory of Operation – Electronic Steam Sterilizer Models E, EK, EA, EKA, EZ and EZ10k
The Tuttnauer Steam Sterilizer is designed as a gravity displacement system. This means that no other methods are used to move steam and air in or out of the Chamber other than the natural forces of gravity.
Water inside the autoclave Chamber is heated to produce steam. The rising steam forces any air inside the Chamber to the top of the Chamber where it is bleed off by the Air Jet. This event is due solely to the effect of gravity on the steam and air. As the pressure builds within the Chamber the air is continuously expelled through the unit’s Air Jet. The Air Jet is located in the water reservoir and connected by a copper tube to the top rear of the Chamber. The process of removing the air and leaving only steam in the Chamber is essential to the operation of the Sterilizer. Assisting the Air Jet in this function is the Air Outlet Valve. This valve participates in removing the air until a temperature of 195° is reached then the valve closes and any remaining air is removed by the Air Jet.
Steam temperature has a direct and important correlation to steam pressure. At every level of pressure, steam has a specific corresponding temperature, this is a universally accepted fact. There is one stipulation required to make this true, that is that there must be 0% air present. For this reason it is important that the air be removed as completely as
possible from the Chamber. Removing the air is what allows the temperature to rise properly inside the Chamber.
The importance of a clean working Air Jet can not be understated. The Air Jet has two important functions. First is to remove the air from in side the Chamber while the unit is heating up. If air were allowed to remain in the Chamber its presence would produce pockets of low and high temperature. These uneven temperatures within the Chamber would result in areas of no sterilization. Only by removing the air can more uniform temperatures be attained and as a result even and complete sterilization. In addition the Air Jet has a second function, that is to maintain circulation within the Chamber. It does this by remaining open after all the air has been bled off and continuing to purge the steam. This constant purging of steam causes motion within the Chamber. This constantly moving, constantly circulating steam is important in maintaining uniform temperature Uneven steam temperatures can be the result of the heating elements turning on and off during the sterile cycle. This can cause hot and cold pockets of steam within the Chamber. If an instrument is in one of these cold pockets it will not be sterilized even though the rest of the load was and the spore test confirmed a sterile load. The end result of keeping the steam in motion, because of the Air Jet, is that no pockets of uneven temperature will form and the load will be completely sterilized.
Why use steam in the first place? There are several reasons for preferring a steam Sterilizer.
The first is that steam is non-toxic. The second is that steam sterilization is fast. Steam has excellent heat transfer properties.
It allows for tremendous amounts of heat energy to be transferred to the instruments instantaneously. This flash of energy is what destroys the biological
contamination. Third, steam is readily available and easy to make from any water source. Forth, equipment designed for steam sterilization is simpler to manufacture and use
Basic operation of the autoclave
1. The operator closes the door and presses start
2. Water flows into the Chamber, the air outlet valve is open so that the water can flow in
smoothly. Water flowing into a hot sealed Chamber will build pressure
immediately exerting a force on the water that can slow or even stop it from
entering the Chamber. The open air outlet valve provides an escape for that
pressure, which allows the water to flow smoothly.
3. The autoclave heats to the proper temperature controlled by the temperature and
pressure sensor. The air outlet valve closes at 1950F. Air escapes the Chamber
through the Air Jet
4. Once temperature is reached the timer counts down the programmed amount of
sterilization time. Steam continues to purge through the Air Jet eliminating any
differences of temperature that can occur within the Chamber.
5. The autoclave exhausts and the sterilization is complete.

2 Installation and Setup

2.1 Unpacking and Inspection
Upon receiving the autoclave carefully inspect the outside of the shipping carton for any signs of damage. If any damage to the shipping carton is found note the location with respect to the autoclave and check that area of the autoclave carefully once it is fully unpacked. In addition once the autoclave is fully unpacked carefully check for any signs of physical damage such as; scratched panels, broken knobs, broken door covers etc…
If any damage is found contact the dealer as soon as possible so that they can file a claim with the shipping carrier and also notify Tuttnauer.
All Tuttnauer products are carefully inspected prior to shipment and all reasonable precautions are taken in preparing them for shipment, to assure safe arrival at their destination
Note: Lifting and carrying should always be done by two people
2.2 Unit Location
The unit should be located on a stable, solid counter top. In the case of the 3850 and 3870 models a table is provided with the unit.
It is not recommended that units be stacked. Adequate clearance is required above the autoclave for the purpose of filling the reservoir with distilled water. In addition some steam escapes through the filling hole, if overhead cabinets are to close steam damage can occur to the underside of the cabinets.
A minimum one inch clearance is required on each side and at the back of the autoclave for access and ventilation.
Note: Lifting and carrying should always be done by two people
2.3 Voltage Requirements
All 110 volt units need to have a stable voltage between 110 and 125 volts AC.
All 220 volt units need to have a stable voltage between 220 and 235 volts AC.
For EK, EKA and EZ10k units check that the incoming voltage is between 220 volts and 235 volts AC. This is important because too high a voltage will damage the heating elements and too low a voltage will cause the sterilizer to run slower. In either case a Buck / Boost Transformer is recommended to correct the voltage. A Buck / Boost Transformer is relatively inexpensive and can be configured to either raise or lower the voltage
It is recommended that all autoclaves be installed on a direct line.
The use of a surge suppressor is recommended especially in areas where there is a
large fluctuation in voltage or frequent lightning strikes.
2.4 Setup
There are two procedures for setup depending on if the autoclave has
a Water Pump or not.
2.4.1 Setup and Automatic Filling for units without Water Pumps
These units will have Microprocessors with date codes that do not contain the letters WP. [see sec 8.12]
In these units the Chamber is filled from the Reservoir by gravity flow
Adjusting The Chamber Pitch
Proper adjustment of the Chamber pitch is one of the most important things you can do for the sterilizer. Proper Chamber pitch insures that among other things the sterilizer will have the proper amount of water in the Chamber at the beginning of each cycle. Insufficient water in the Chamber at the beginning of the cycle will generate a LOW WATER message at some point during the cycle when the water level becomes to low. If on the other hand there is too much water in the Chamber this will extend the heating portion of the cycle. In cases where the heating portion of the cycle is extended for more than 50 minutes (or 80 minutes for a 3850 / 3870) the sterilizer will abort that cycle.
Start with a sturdy, level counter.
Make sure all the feet are on the autoclave and none have been lost
Make sure the front feet are free to move in and out
Position the autoclave on the counter.
Fill the Reservoir with distilled water [see sec 5.2]
The Chamber should be empty of any instruments, trays or leftover water.
The autoclave should be turned off
The Chamber pitch now needs to be adjusted correctly
Measure out the proper amount of distilled water for the appropriate
model unit as listed below:
1730 = 10 oz – 12oz (300ml – 360ml) 2340 = 12 oz – 15oz (350ml – 440ml 2540 = 12 oz – 15oz (350ml – 440ml) 3850 = 20 oz – 23oz (600ml – 680ml) 3870 = 24 oz – 27oz (750ml – 800ml)
Pour the proper amount of water into the Chamber through the front door
of the unit.
This water should cover the bottom of the Chamber to within +/- ½ inch of
the grove in the front.
If necessary adjust the front Leveling Feet so that the water lays in the
Chamber correctly Once the Chamber pitch adjustment is completed, empty the water from the Chamber and check if the automatic filling is set correctly.
Checking the Automatic Fill
To check the automatic filling procedure, follow the next few steps: Remove any water that is in the Chamber Make sure the unit is turned on With the Door open, press and hold the Door Switch Then press the START Key. When water starts flowing into the Chamber release the Door Switch. Water should come up to the same spot as the measured amount had. If the water fill is not working correctly then try the adjustment procedure or check for a system problem [see sec 7.15]
Automatic Filling Adjustment Procedure
Make sure the power is off. The Door should be open
Press and hold the Water Inlet Key (this is the button on the
front Keypad with the two arrows). Turn the power on
When the normal display screen appears release the Water Inlet Key – wait one second and then press it in again. Water should begin flowing into the Chamber
Monitor the water flow into the Chamber.
Hold the Water Inlet Key until water reaches the groove at the
front Then release the button – wait ten seconds – the unit is now
reprogrammed.
2.4.2 Setup and Automatic Filling for units with Water Pumps Any unit with a Microprocessor date code ending in WP [see sec 8.12]
will have a Water Pump installed to insure proper filling.
Start with a sturdy, level counter. Make sure all the feet are on the autoclave and none have been lost Make sure the front feet are free to move in and out Position the autoclave on the counter. Fill the Reservoir with distilled water [see sec 5.2] The Chamber should be empty of any instruments, trays or leftover water.
To calibrate the automatic fill follow this procedure:
1. Press the STOP Key repeatedly until the message “Code: xxx”
appears.
2. Using the UP/DN arrow keys change the code to 105, then press the STOP Key.
3. A message will be displayed saying “Water in = xx sec”
4. Using the UP/DN arrow keys change the seconds according to
the following table:
2340 = 30 sec 2540 = 35 sec 3870 = 65 sec
now press the STOP Key
5. The message “Ea Type:” will appear, using the UP/DN arrow
keys select either “0” for an E or EK type unit or “1” for an EA or EKA type unit.
6. Press the STOP Key to finish

3 Front Panel Keypad

3.1 Front Panel Keypad

The Front Keypad is divided into four sections, top, upper middle, lower middle and bottom
The Top Section has four lighted buttons, these represent the four available programs. Each program comes preset from the factory with default parameters. Each program can, however, be modified by the operator. If necessary the default or modified parameters can be lock-in [see sec 8.10]. When a program is selected the light in that button will illuminate and the program parameters will appear in the display.
Moving from left to right the programs are:
Unwrapped Instruments – symbolized by a pair of scissors
the default parameters are
273 0F for temperature 3 minutes of sterilization time rapid exhaust no drying time
Wrapped Instruments – symbolized by a gown
the default parameters are
273 0F for temperature 7 minutes of sterilization time rapid exhaust 30 minutes of drying time
Liquids – symbolized by a flask the default parameters are 250
0
F for temperature 30 minutes of sterilization time slow exhaust only no dry time, drying is not allowed
Extra Drying Cycle – symbolized by the sun the default parameter is 30 minutes of drying time
The purpose of the Extra Drying Cycle is to offer an alternative in situations where the dry time in the wrapped or unwrapped cycle is insufficient. Rather then wait for the items to air dry or run another complete cycle with a longer dry time just select the Extra Drying Cycle to continue the heat assisted drying process.
The Upper Middle Section consists of a:
Display –
The Display is comprised of a single row of 16 characters, this row is divided
into four sections.
When the system is running a program the screen will display the current
temperature and pressure within the Chamber and the remaining time for sterilization or drying
The first three sections from left to right are designed to show the parameters of
the selected program or any operating messages.
When the system is idle the display will show the parameters of the currently
selected program.
When the system is running a program the actual sterilization temperature is
If the program aborts as a result of a program check or manual stop a message
The last section of the screen on the right, will continually display the actual
TEMP Key –
The TEMP Key is used to change the temperature parameter of the Wrapped, Unwrapped or Liquid programs. This can only be done while the autoclave is
not running a cycle. Press the TEMP Key and a cursor will appear under the temperature parameter. Use the Up/Down Arrow Keys to change to the desired temperature. After a few seconds of inactivity the cursor will disappear and the parameter will be locked in. The acceptable range for proper sterilization of wrapped and unwrapped items is between 2500F and 2740F (1210C and 1340C). For liquids the maximum temperature is 2500F (1210C)
In addition the TEMP Key can be used to change the temperature display from Fahrenheit to Centigrade. This can be accomplished by simply turning the power off, press and hold the TEMP Key and turn the power on.
STE TIME –
The STE TIME Key is used to change the sterilization time parameter of the Wrapped, Unwrapped or Liquid programs. This can only be done while the
autoclave is not running a cycle. Press the STE TIME Key and a cursor will appear under the sterilization time parameter. Use the Up/Down Arrow Keys to change to the desired sterilization time. After a few seconds of inactivity the cursor will disappear and the parameter will be locked in.
DRY TIME –
The DRY TIME Key is used to change the dry time parameter of the Wrapped and Unwrapped programs. This can only be done while the autoclave is not running a cycle. Press the DRY TIME Key and a cursor will appear under the dry time parameter. Use the Up/Down Arrow Keys to change to the desired dry time. After a few seconds of inactivity the cursor will disappear and the parameter will be locked in. The acceptable range for drying time is 0 to 99 minutes
In addition the DRY TIME Key can be used to change the pressure display from psi to bar (on all machines up to an including Microprocessors dated T93N6) or from psi to kpa (on all machines with Microprocessors dated T96DN1 or T97DN6 or later)
displayed above the TEMP Key. The remaining sterilization time will be displayed above the STE TIME key and the remaining drying time will be displayed above the DRY TIME Key.
will be displayed on the screen. When a message is displayed, pressing any key will erase the message and redisplay the selected program.
current real pressure inside the Chamber, this occurs whether a program is running or not (provided the main power is on).

CLOCK Key –

Pressing the CLOCK Key first displays the current date with a cursor under the day parameter. Pressing the Up/Down Arrow Keys will change the day parameter. Pressing the CLOCK Key once again will move the cursor underneath the month and then the year parameters. Once the date has been updated pressing the CLOCK Key again will display the time with the cursor under the hour. Use the Up/Down Arrow Keys as before to change the hours then run through the minutes and seconds pressing the CLOCK Key each time to make the advance. After a few seconds of inactivity the cursor will disappear and the parameters will be locked in.
UP / DOWN Arrow Keys
Pressing these keys will raise or lower the values on any of the parameters that are user adjustable
The Lower Middle Section consists of a:

STOP Key –

This is the only key recognized by the system while a cycle is running. Pressing the STOP Key for over one second will cause the current program to abort and the MAN STOP message will be displayed.
In addition the STOP Key can be used to reset all the parameters back to their factory defaults. This includes the Automatic Fill, in which case it will be necessary to recalibrate the Automatic Fill.
Turn the power off Press and hold the STOP Key turn the power on
START Key –
Pressing this key will start which ever program cycle has been selected and cause the START Key light to turn on.
Water Inlet Key–
This key is symbolized by the two horizontal arrows pointing in through a channel. Pressing and holding this key allows for the manual filling of the Chamber with water. This is useful for calibrating the Automatic Fill, also during cleaning to flush out the Chamber and in case it becomes necessary to bypass the Automatic Fill before running a cycle. Water will flow into the Chamber only as long as the key is depressed.
The Bottom Section consists only of indicator lights. Looking from left to right they are the:
HEAT Light – a steady illumination is given when the autoclave is heating up at the
beginning of the cycle. Also this light will flash during the preheat / standby mode on units in which that option has been activated (all EK, EKA, EZ, EZ10k, 3850 and 3870 machines).
STE Light – a steady illumination is given while the autoclave is in the sterilization
portion of the cycle.
EXH Light – a steady illumination is given when the autoclave is exhausting the
Chamber
DRY Light - a steady illumination is given while the autoclave is in the Drying mode.
CYCLE FAIL Light – will illuminate anytime the autoclave detects a problem that
results in an aborted cycle.
ADD WATER Light – this indicator will light when the Reservoir is low on water. If the
indicator lights, after the Start Key has been pressed, the system will continue with the cycle. There is sufficient water in the reservoir to complete this cycle. The next cycle will not be allowed to begin until sufficient water is in the reservoir.
DOOR CLOSED Light – this indicator lights to signal that the Door of the autoclave
has been closed.

4 Display Messages

4.1 Display Messages

Any time a cycle is aborted the Tuttnauer autoclave will give an error message. These messages are in the form of words that describe the problem the unit has encountered.
The following is a list of those Error Messages with descriptions of what they
mean and indications where the problem may be.
LOW WATER – This message will be displayed if during a normal Heat Up stage the
system determines that there is insufficient water in the Chamber to complete the cycle. This determination is made by the combined input of two sensors, the Water Electrode and the Safety Thermostat. Also if a power failure occurs during the Heat or Sterilization stage after the power returns the system will check the Water Electrode to see if there is sufficient water in the Chamber in order to resume the cycle. If not the cycle will be aborted, the message LOW WATER will be displayed, and the Cycle Fail indicator will light.
Possible causes for this message are:
a. Insufficient water entered the Chamber at the beginning of
the cycle. Check for proper leveling, a dirty or shorted Water Sensing Electrode, a clogged Water Pump, a partially clogged line or that the Air Outlet Valve is stuck closed
b. A leaky Solenoid Valve, Safety Valve, Air Jet, Door Gasket,
Door Bellows or a pipe fitting is allowing water or steam to escape at a higher than normal rate.
c. A power down has occurred and on power up if the water
Electrode tip is dry the Low Water message will be displayed
LOW HEAT – This message is displayed, the Cycle Fail indicator lights and the cycle is
aborted if the autoclave has not reached sterilization temperature after heating for 50 minutes in either Wrapped or Unwrapped programs (80 minutes in the Liquid program). Low Heat refers to the temperature in the Chamber before sterilization has begun.
Possible causes for this message are:
a. No power to the Heating Elements b. Bad Heating Elements c. Very low line voltage delaying heat up d. Safety Thermostat is opening prematurely, turning off the
Heating Elements -- this only applies to units with Microprocessors dated earlier than T93N5 or T93N6.
e. A clogged Air Jet
f. An Air Outlet Valve stuck closed.
LOW TEMP - This message is displayed, the Cycle Fail indicator lights and the cycle is
aborted if the temperature drops 2.5 0C (4.5 0F) below the required sterilization temperature.
Possible causes for this message are:
a. Insufficient water in the Chamber (see Low Water message)
b. The sterilization phase of the cycle has been set for too long
a period of time, allowing the Chamber water to boil away, and the Chamber to run dry.
c. The Safety Thermostat is opening prematurely, turning off the
Heating Elements -- this only applies to units with Microprocessors dated earlier than T93N5 or T93N6
d. A bad Temperature Sensor
LOW PRES – This message is displayed, the Cycle Fail indicator lights, and the cycle is
aborted if the pressure drops 4 PSI (0.27 BAR) below the required sterilization pressure.
Possible causes for this message are:
a. Insufficient water in the Chamber ( see Low Water message)
b. The Heating Elements not cycling on and off properly.
1. Problem is with the Solid State Relay
2. Problem with the control circuit c. Bad Heating Elements -- not producing enough wattage d. The Safety Thermostat is opening prematurely, turning off the
Heating Elements -- this only applies to units with Microprocessors dated earlier than T93N5 or T93N6
e. A bad Pressure Transducer
HIGH TEMP – This message is displayed, the Cycle Fail indicator lights and the cycle
is aborted if the temperature rises 9 0F (5 0C) above the required sterilization temperature during the Sterilization phase of the cycle. This message will also be displayed if the Temperature Sensor is damaged. In this case the message will appear just before the Heat phase starts
Possible causes for this message are:
a. The Heating Elements are remaining on instead of cycling on
and off. Check for a shorted Solid State Relay, shorted Heating Element or other short circuit
b. This message can ALSO indicate a bad Temperature Sensor
-- the message will display any time during the Heat Up phase.
HIGH PRES – This message is displayed, the Cycle Fail indicator lights, and the cycle
is aborted if the pressure rises 10 PSI (0.6 BAR) above the required sterilization pressure.
Possible causes for this message are:
a. The Heating Elements are remaining on instead of cycling on
and off. Check for a shorted Solid State Relay, shorted Heating Element or other short circuit
b. The Sterilization temperature has been set above 274 0F
MAN STOP - This message will be displayed and the Cycle Fail indicator will light
after the STOP Key is depressed for longer than 1 second.
RENEW WATER - This message is displayed only as information to the operator, that
the Water Reservoir should be drained and refilled with clean distilled water. This message will only appear on units with Microprocessors date coded earlier than and including T93N5.
POWER DN – This message is displayed once the power is restored, after a power
failure occurs during the running of a cycle. The POWER DN message will be displayed for several seconds, and if present the Printer will print POWER DN on the print out.
Once power has been restored the autoclave will make an attempt to
resume the current cycle from the point at which it was interrupted.
If a power failure occurs during the Heat Up phase, then heating will
resume (provided there is enough water in the Chamber. If not, the cycle will be aborted.)
Exhaust and Dry phases will automatically resume operation once power
is restored.
If the power down occurred during the Sterilization portion of the cycle,
then when power is restored, the autoclave will check if the temperature in the Chamber has fallen more than 4.5 0F (2.5 0C). If not then the Sterilization Cycle will resume automatically. If however, when the power returns, the system determines that the temperature has fallen more than 4.5 0F (2.5 0C), the Sterilization Cycle will abort and the Exhaust Cycle will start.
If a power failure occurs during the Liquids program, the system
will not allow a fast exhaust (as the exhaust valve is normally closed), nor will it fast exhaust when power comes back on.
ADD WATER - This message is displayed and the ADD WATER indicator lights to
show insufficient water in the Water Reservoir. If this message is displayed after the START Key has been pressed then the system is not allowed to proceed. After water is added to the Reservoir, the START Key must be depressed again in order for the selected cycle to begin.
DOOR UNLOCK – This message will be displayed and the DOOR CLOSED indicator
will remain unlit if the door is improperly closed when the START
Key is depressed. Once the door is properly closed, the DOOR CLOSED indicator will light and the START Key should be
depressed to start the desired cycle. If the door accidentally opens during any stage of the cycle, the same message and indicator will appear, and the system will abort, the Cycle Fail indicator will light and the DOOR UNLOCK message will be displayed.
WATER INLET – This message will be displayed as information to the operator while
water is entering the Chamber, during the Automatic Water Filling process.
CYC FINISHED – This message is displayed at the end of a successfully completed
cycle

5 Operating Instructions

5.1 Preparation Before Sterilizing

Instruments to be sterilized must be free from all residual matter, such as blood or organic tissue. Instruments must also be dry and free from mineral deposits. Such substances may cause damage to the instruments themselves or the Sterilizer.
1. Clean instruments immediately after use to remove any residue. It is
recommended that all instruments be ultrasonically cleaned using Tuttnauer's CLEAN AND SIMPLE enzymatic cleaning tablets or other suitable solution.
2. After cleaning, rinse instruments for 30 seconds & pat or air dry.
3. Follow the instrument manufacturer’s instructions on the use of products for
cleaning and lubricating instruments that have been ultrasonically cleaned.
4. Be sure that instruments of dissimilar metals (stainless steel, carbon steel, etc.) are
separated. Carbon steel instruments should be bagged or placed on autoclavable towels and not directly on stainless steel trays.
5. When using a paper / plastic bag the plastic side should always be down.
6. Check the instructions of the item manufacturer as to the proper procedure for
sterilizing each item.
7. Items must be sterilized in an open position. Surfaces that are hidden because the
item is in a closed position will not be exposed to the steam and will not be sterilized
8. Place a sterilization indicator in each tray or inside each wrapped pack.
9. At least once a week use a biological spore test (Bacillus Stearothermophilus) in
any load to insure proper sterilization. (Be aware testing standards may vary) Always follow the spore test manufacturer’s instructions
10. Make sure that all instruments remain apart during the sterilization cycle. Surfaces
that are hidden because items are covering other items will not be exposed to the steam and will not be sterilized
11. Empty canisters should be placed upside-down in order to prevent the
accumulation of water.
12. Do not overload the Sterilizer trays. Overloading will cause inadequate
sterilization & drying. (see table 9.7 for loading limits for each model)
13. Allow a distance of approximately 1" between trays to permit steam circulation.
14. Wrapped instruments should be placed in material which will allow steam
penetration and promote drying, such as autoclave bag, autoclave paper, or muslin towels.
15. Do not stack pouches. It is recommended that a pouch rack such as the Tuttnauer
Pouch Rack be used to insure proper steam penetration and adequate drying. Surfaces that are hidden because the items are being stacked will not be exposed to the steam and will not be sterilized
16. Tubing should be rinsed after cleaning. When placed in the tray make sure that
both ends of the tubing are open and there are no sharp bends or twists
17. Packs should be placed upright on the tray. They should not be touching each
other or the Chamber walls. There should be about 1” between packs for proper steam circulation.
18. Liquids should only be sterilized in heat proof glass. The beaker should only be
filled 2/3 full and the lid should be on loosely to allow for expansion (see the table
9.8 for the maximum liquid capacity for each model)
19. If spotting is detected on the instruments the first step would be to use an ordinary
eraser to remove the spot. If there is no pitting under the spot then the spot was only dirt. Dirt spots on an instrument may be an indication that the autoclave needs to be cleaned or that the instruments were not adequately cleaned or dried. If removal of the spot reveals pitting then the spot was most likely rust. Rust spots on an instrument are not uncommon on inexpensive instruments. It may also be an indication that the instruments were rinsed in tap water with a high content of minerals. These minerals when exposed to high temperature and steam will accelerate the oxidation of the metal. One suggestion would be to final rinse the instruments in a distilled water or alcohol bath.
20. If the instruments exhibit a discoloration this can be due to the mixing of carbon
steel and stainless steel. When these two metals come into contact with each other an electrolysis occurs that breaks down the metal. The best solution is to separately wrap the carbon steel to insulate it from other instruments or the trays.
21. Items should not be allowed to touch the walls of the Chamber as the hot metal
can damage the item.
5.2 Filling The Reservoir
Always use DISTILLED WATER in the autoclave for sterilizing. Using water of a poorer quality will cause increased maintenance due to the mineral residue that accumulates in the various parts of the autoclave.
The Reservoir is filled from the top of the autoclave. Remove the Reservoir Cover and pour water through the opening. Continue filling until the water reaches the base of the Safety Valve Holder. Under no circumstances should the Reservoir be filled above the Safety Valve Holder. The Reservoir should never be filled while the autoclave is running a cycle. If the Reservoir is filled while the Autoclave is running then at the end of the cycle water exhausted from the Chamber can cause the Reservoir to overflow.
Overfilling or the failure to use Distilled Water will lead to clogging of the hole in the Air Jet. This will be evidenced by the lack of both a hissing sound and a stream of steam coming from the Air Jet during sterilization. When this situation occurs follow the instructions in sec 6.5 for cleaning the Air Jet.
5.3 Sterilization Programs
Program 1 – Unwrapped Instruments
This program is for sterilizing unwrapped instruments and materials that the manufacturer of these items has recommended autoclaving at a temperature between 2500F and 2740F (1210C and 1340C).
This program comes set with these default parameters: Sterilization temperature 2730F Sterilization time 3 minutes Dry time none
These values can be altered to fit the needs of a particular office
The parameters can only be changed while the autoclave is not running a
cycle.
Press the TEMP Key and a cursor will appear under the temperature
parameter. Use the Up/Down Arrow Keys to change to the desired
temperature. The acceptable range for proper sterilization of unwrapped
items is between 2500F and 2740F (1210C and 1340C).
*** Caution – in no case should the temperature be set higher than
Any change of temperature must be coordinated with a corresponding
change is sterilization time.
Press the STE TIME Key and a cursor will appear under the sterilization
time parameter. Use the Up/Down Arrow Keys to change to the desired
sterilization time.
If drying is desired then press the DRY TIME Key and a cursor will
appear under the dry time parameter. Use the Up/Down Arrow Keys to
change to the desired dry time. The acceptable range for drying time is 0
to 99 minutes
After a few seconds of inactivity the cursor will disappear and the
parameter will be locked in.
0
F (1340C) ***
274
Program 2 – Wrapped Instruments
This program is for sterilizing wrapped instruments and materials that the
manufacturer of these items has recommended autoclaving at a
temperature between 2500F and 2740F (1210C and 1340C)
This program comes set with these default parameters:
Sterilization temperature 2730F
Sterilization time 7 minutes
Dry time 30 minutes
These values can be altered to fit the needs of a particular office
The parameters can only be changed while the autoclave is not running a
cycle.
Press the TEMP Key and a cursor will appear under the temperature
parameter. Use the Up/Down Arrow Keys to change to the desired
temperature. The acceptable range for proper sterilization of unwrapped
items is between 2500F and 2740F (1210C and 1340C).
*** Caution – in no case should the temperature be set higher
0
than 274
Any change of temperature must be coordinated with a corresponding
change is sterilization time.
Press the STE TIME Key and a cursor will appear under the sterilization
time parameter. Use the Up/Down Arrow Keys to change to the desired
sterilization time.
If a longer or shorter drying is desired then press the DRY TIME Key and
a cursor will appear under the dry time parameter. Use the Up/Down
Arrow Keys to change to the desired dry time. The acceptable range for
drying time is 0 to 99 minutes
After a few seconds of inactivity the cursor will disappear and the
parameter will be locked in.
F (1340C) ***
Program 3 – Liquids
This program is for sterilizing liquids solutions, distilled water, medicines
and other liquid preparations. For proper liquid sterilization it is
recommended that only the default parameters be used.
This program comes set with these default parameters:
Sterilization temperature 2500F
Sterilization time 30 minutes
Dry time drying is not allowed

Program 4 – Extra Drying Cycle

The purpose of the Extra Drying Cycle is to offer an alternative in
situations where the dry time in the wrapped or unwrapped cycle is
insufficient. Rather then wait for the items to air dry or run another
complete cycle with a longer dry time, just select the Extra Drying Cycle
to continue the heat assisted drying process.
This program comes set with these default parameters: Dry time 30 minutes
These values can be altered to fit the needs of a particular office
If a longer or shorter drying is desired then press the DRY TIME Key and
a cursor will appear under the dry time parameter. Use the Up/Down
Arrow Keys to change to the desired dry time. The acceptable range for
drying time is 0 to 99 minutes
After a few seconds of inactivity the cursor will disappear and the
parameter will be locked in.
Loading...
+ 176 hidden pages