Tuttnauer Elara11 Service manual

(CAT Controller)
TECHNICIAN
MANUAL
Pre/Post-vacuum Class B
Table–top Autoclave
Cat. No. MAN205-0343001EN Rev. P 8-10-11
Manufactured by: Tuttnauer Co. Ltd., Har Tuv Industrial zone B P.O.Box 170, Beit Shemesh 99000, Israel Tel: 972 2 9904611, Fax: 972 2 9904730
Tuttnauer U.S.A. Co, Ltd. 25 Power Drive Hauppauge, NY, 11788, USA. Tel (631) 737 4850, (800) 624 5836, Fax: (631) 737 0720
TABLE OF CONTENTS
PARAGRAPH PAGE NO.
1 INTRODUCTION ................................................................................................... 4
2 SYMBOL DESCRIPTION ..................................................................................... 4
3 INSTALLATION .................................................................................................... 7
3.1 Lifting and carrying ...................................................................................... 7
3.2 Unpacking the autoclave ............................................................................... 7
3.3 Installation preparations ............................................................................... 7
3.4 Connections to Utility Supplies ..................................................................... 8
3.5 Final adjustments .......................................................................................... 8
3.6 Operating the autoclave ................................................................................ 9
4 PERIODICAL TESTS .......................................................................................... 10
5 WATER QUALITY ............................................................................................... 11
5.1 Water for Generating Steam ....................................................................... 11
6 CONTROL SYSTEM DESCRIPTION ................................................................ 12
6.1 The main board includes: ........................................................................... 12
6.2 The high voltage board (AC Board) includes: ........................................... 13
7 CHECKING AND CHANGING PARAMETERS AND OTHER DATA ........... 14
7.1 Menu ............................................................................................................ 15
7.2 Parameters ................................................................................................... 16
7.3 Digital Inputs ............................................................................................... 28
7.4 Digital Outputs............................................................................................. 29
7.5 Analog Inputs .............................................................................................. 30
7.6 Calibration ................................................................................................... 31
7.7 Setting the Clock .......................................................................................... 37
7.8 Printer Test .................................................................................................. 38
7.9 Language ..................................................................................................... 39
7.10 History .......................................................................................................... 40
7.11 More options ................................................................................................ 41
8 RESSETTING THE AUTOCLAVE ..................................................................... 43
9 MAINTENANCE AND REPLACEMENT PROCEDURES .............................. 44
9.1 Preliminary Operations for Each Technician Call .................................... 44
9.2 Safety tests after repair ................................................................................ 45
9.3 Dismantling the Outer Covers of the Autoclave ......................................... 46
9.4 Replacing the Safety Valve .......................................................................... 47
9.5 Replacing chamber heater .......................................................................... 48
9.6 Replacing steam generator heater .............................................................. 50
9.7 Replacing the steam generator cut-off thermostat ..................................... 51
9.8 Replacing and cleaning the generator’s water level electrode .................. 52
9.9 Replacing the Drain Valve .......................................................................... 54
9.10 Replacing the Door Cover ........................................................................... 55
9.11 Replacing the Front Panel .......................................................................... 56
Page 1 of 129 Pages
9.12 Replacing the Printer Door ......................................................................... 57
9.13 Replacing the Door Handle......................................................................... 58
9.14 Replacing the Plastic Handle Cover ........................................................... 59
9.15 Replacing the Printer .................................................................................. 60
9.16 Replacing the Door Switch .......................................................................... 61
9.17 Replacing the fuse on the Electronic Board ............................................... 63
9.18 Replacing the fuses of the transformer ....................................................... 64
9.19 Replacing the water pump ........................................................................... 65
9.20 Replacing the vacuum pump ....................................................................... 66
9.21 Replacing the Plunger or Coil of the 1/4" Solenoid Valve ........................ 67
9.22 Replacing the Electronic Boards ................................................................ 68
9.23 Replacing the switch .................................................................................... 73
9.24 Draining the generator ................................................................................ 74
9.25 Emergency Door Opening ........................................................................... 75
9.26 Replacing the water reservoirs .................................................................... 77
9.27 Cleaning the water reservoirs ..................................................................... 79
9.28 Replacing the cut-of thermostat for the Chamber Heater ......................... 80
9.29 Replacing the PT100 temperature sensor for the Chamber Heater .......... 82
9.30 Replacing the internal PT100 ..................................................................... 84
9.31 Replacing the pressure transducer ............................................................. 86
9.32 Cleaning the steam generator ..................................................................... 87
9.33 Replacing the cooler and the cooler's fan .................................................. 88
9.34 Testing procedure for the chamber PT100 ................................................. 90
9.35 Testing procedure for the heating element PT100 ..................................... 92
9.36 Testing procedure for the chamber pressure transducer ........................... 94
9.37 Testing procedure for the Generator pressure transducer” ...................... 95
9.38 Door Locking Solenoid................................................................................ 97
10 IN-OUT TEST ....................................................................................................... 99
11 TROUBLESHOOTING ...................................................................................... 101
11.1 Preliminary Check ..................................................................................... 101
11.2 Troubleshooting procedure ....................................................................... 102
12 PRESSURE VS TEMPERATURE FOR SATURATED STEAM .................... 114
Page 2 of 129 Pages
TABLE OF CONTENTS (Continue)
PARAGRAPH PAGE NO.
Front View ...................................................................................................................... 5
Rear View ....................................................................................................................... 6
Control System Block Diagram ................................................................................... 12
Assembled Nova4-D2 Board ........................................................................................ 69
Nova4-A2 Board ........................................................................................................... 70
Nova4-D2 Board .......................................................................................................... 71
Exploded Drawing – Rear.......................................................................................... 118
Table for Exploded Drawing – Rear ......................................................................... 119
Exploded Drawing – Door ......................................................................................... 120
Table for Exploded Drawing – Door ......................................................................... 121
Exploded Drawing – PT-100 and Cut-Off ................................................................ 122
Table for Exploded Drawing – PT-100 and Cut-Off ................................................ 123
Exploded Drawing – Generator ................................................................................. 124
Control Unit ............................................................................................................... 125
Partial Front View ..................................................................................................... 126
Electrical Diagram ..................................................................................................... 127
Piping Diagram .......................................................................................................... 128
Piping Layout ............................................................................................................. 129
Page 3 of 129 Pages
1 INTRODUCTION
This Technician's Manual, together with the operator’s manual, forms the
complete set of the Operation and Maintenance instructions for the Elara 11 pre and post vacuum autoclave. This manual is intended for the use of the technician. It is strongly recommended that only qualified and Tuttnauer factory trained personnel service this autoclave and do so in accordance with the instructions in this manual. Any unauthorized service may result in the invalidation of the manufacturer’s guarantee.
2 SYMBOL DESCRIPTION
Caution! Consult accompanying documents
Caution! Hot surface
Caution! Hot steam.
Protective earth (Ground)
Page 4 of 129 Pages
FRONT VIEW
1 2 3
17
16
5 4 6
718
8
9
13 14
No.Description No. Description
1 Water reservoir funnel 10 Operating keypad
2 Autoclave cover 11 Printer cover
3 Water level sight gauge 12 Printer (standard)
4 Mineral-free water reservoir cover 13 Legs
Chamber and steam generator safety
5
valves
6 Waste water reservoir cover 15 Waste water reservoir drain valve
Display 16 Door cover
7
8 Ventilation grill 17 Door switch
9 Main switch/Circuit breaker 18 Door closing device
Mineral-free water reservoir drain
14
valve
101115 13 12
Page 5 of 129 Pages
REAR VIEW
1
2
3
4
5
6
3
No.Description
1 Mineral-free water reservoir cover
2 Waste water reservoir cover
3 Ventilation grills
4 Air filter service cover
5 Opening for calibration
6 Main power electric cable socket
Page 6 of 129 Pages
3 INSTALLATION
CAUTION:
Any operation described in section 3 shall be performed by a qua lified
technician only.
3.1 Lifting and carrying
CAUTION:
Before moving the a utoclave, Make sure that the electric cord is
disconnected from the pow er, and there is no press ure in th e chamber and in the generator.
Attention! The pressure of th e generator does not decreas e
immediately when the equipment is turned off. Wait approx. ½ an hour to ve rify that the pressu re has decreased to a tmospheric pressure.
1. Disconnect the power supply cord.
2. Drain the water from both reservoirs. To avoid injuries, lifting and carrying should be done with at least two persons or by using a fork-lift or any other mechanical aid.
Do not drop the device!
3.2 Unpacking the autoclave
Unpack the autoclave and inspect for mechanical damage upon receipt.
Observe packing method and retain packing materials until the unit has been inspected. Mechanical inspection involves checking for signs of physical damage such as: scratched panel surfaces, broken knobs, etc.
To avoid injuries, lifting and carrying of the autoclave should be done
with at least two persons or by using a fork-lift or any other mechanical aid.
3.3 Installation preparations
1. Check and verify that the counter carrying the autoclave is a rigid and
leveled surface and can carry a load of 225 lbs (102kg).
Attention:
The Elara11 is not designed for use on any standard slide out shelf. If
it is necessary to use a slide out shelf, it must be tested and/or rated for 225 lb (102kg) or more.
Check and verify that the counter dimensions are, at least, 22” wide x
24” deep (55cm x 63.5cm).
2. Keep
the back and the sides of the autoclave approximately 2” (5 cm)
away
from the wall to allow ventilation and facilitate the device
disconnection.
3. If placed in a cabinet, verify that the rear of the cabinet is open to
allow ventilation.
4. Insufficient space for ventilation may result in an increase of
the
autoclave's temperature that may cause a malfunction or damage the instrument.
5. It is recommended that enough space be left around the autoclave to
give a technician access for servicing the machine.
6. Check and verify that the room ventilation is 10 cycles per hour
minimum.
Page 7 of 129 Pages
7. Check and verify that the ambient temperature range is 41˚F - 104˚F
(5˚C - 40˚C), it is preferable not to exceed 86˚F (30˚C).
8. Check and verify that the ambient relative humidity does not exceed
85%.
3.4 Connections to Utility Supplies
1. Check and verify that the power supply is a 1 phase, 230Vac ±5%,
50/60Hz (as appropriate), 15A supply.
2. Check and verify that the autoclave is connected to separate power
source to avoid flickers of light or sensitive devices.
3. Check and verify grounding of the autoclave.
4. Check and verify that the electrical net is protected with a current
leakage safety relay.
3.5 Final adjustments
At the time of installation, befo re the autoclave can be okayed for
daily operation the service technician needs to perform the follow ing preliminary checks:
a. Integrity Check
Perform a visual check to verify that there are no dents, scratches or
broken components on the autoclave. b. Support surface check Make a visual check that the surface is level and strong enough to
support the autoclave.
c. The leakage current test
Test the precise operation of the earth leakage relay.
d. Ground Check
Test the continuity of the grounding connection.
e. Safety Valve Check
Test the safety valve as per the instructions in "Checking the Safety
Valve" in the Operator’s Manual.
f. Door check Ensure that the door locking mechanism is functioning properly g. Setting clock and date
See instruction in the user manual clause. 7.3. h. Atmospheric Pressure check: the unit is set from the factory at 100
kpa, this is atmospheric pressure at sea level. If the unit is located
more that 500 ft above or below sea level then the Atmos. Press
parameter must be set, see sec 7.2.7 i. Cycle check: run a B&D Test to ensure that all systems in the unit
are functioning properly.
j. Reset the autoclave see sec 8.
After the above steps are performed, the autoclave is ready for daily
operation.
Page 8 of 129 Pages
3.6 Operating the autoclave
1 Plug the power cord into the power socket. 2 Turn on the Main Switch / Circuit Breaker (see front view). 3 Select "Vacuum Test" cycle to keep the steam generator and heating
elements from heating up. See Selecting a Program in the Operator’s
Manual. 4 Open the door of the autoclave and remove the trays and the
packaging material. 5 Fill the Mineral Free Water Reservoir with water meeting the quality
specs in section 5.1 as follows:
5.1. Fill with 4 liters of mineral free water by pouring it into the
front reservoir fill opening at the top of the machine (see front view).
5.2. Fill the remaining quantity by pouring water gently, into the
front funnel until it reaches the required level on the water level sight gauge (See Front View).
6 Insert a paper roll in the printer (see printer handling in the
Operator’s Manual). 7 Close the door and perform a Vacuum Test. If the test fails perform
another test since the fail may be a result of moisture in the air. If the
second test fails it may be necessary to run Program 2 with the long
drying cycle to remove any excess moisture that is causing the test to
fail. 8 If the Vacuum Test is successful then select the B&D Test cycle.
At this stage the chamber and steam generator will be heating up. It
will take approximately 15 minutes (from selecting the B&D Test). 9 While waiting for the unit to heat up explain and instruct the operator
as follows (use the operation manual as reference):
9.1. Operation principals of the autoclave.
9.2. Preparation for sterilizing instruction including loading
instructions.
9.3. Intended use of each cycle.
9.4. Selecting a cycle.
9.5. Water filling method.
9.6. Displayed error and operational messages.
9.7. Monitoring and changing parameters.
9.8. Printer handling.
9.9. Maintenance instructions 10 Perform a B&D test with a chemical indicator. 11 The operator shall perform a cycle under supervision of the
technician.
Page 9 of 129 Pages
4 PERIODICAL TESTS
PERIOD TEST
1 month Test the safety valve by operating it.
6 months
Remove the autoclave’s cover, tighten the heaters’ screws and electrical connections, valves and connectors in the control box.
Check the continuity of the grounding connections. Check the temperature and pressure calibration. Perform validation of the autoclave. Check the precise operation of the earth leakage relay. Check that the autoclave is leveled.
Check the safety elements; safety valve, safety and cut-off thermostats door locking mechanisms.
Year
Run basic programs of the autoclave and check the operation sequences, the sterilization parameters etc.
Check the water reservoir, piping, plastic parts and electric wires. Check and tighten the piping joints to avoid leakage.
Check and tighten all screw connections in the control box, heaters and valves and instrumentation. Calibrate the temperature and pressure once a year or in reference to local rules or regulations (refer to the section on Calibration).
5 years Observe the closing device for excessive wear
Safety tests (pressure vessel, efficiency, electrical) shall be performed in accordance with local rules or regulations, by an authorized inspector.
Only an authorized technician shall perform the 6-months and yearly tests!
Page 10 of 129 Pages
5 WATER QUALITY
5.1 Water for Generating Steam
The distilled or mineral – free water supplied to the sterilizer shall be
according to the table below:
Physical Characteristics and Maximum acceptable contaminants
Evaporate residue 15 miligrams/liter (mg/l)
Silica 2 mg/l
Iron 0.2 mg/l
Cadmium 0.005 mg/l
Lead 0.05 mg/l
Rest of heavy metals 0.1 mg/l
Chloride 3 mg/l
Phosphate 0.5 mg/l
Conductivity ≤50 μs/cm
levels in water for sterlizers
(According to ANSI/AAMI ST79:2006).
Element Condensate – allowable content
pH value 6.5 to 8
Appearance Colorless, clean, without sediment
Hardness 0.1 mmol/l
Compliance with the above data should be tested in accordance with acknowledged analytical methods, by an authorized laboratory.
Attention: We recommend testing the water quality once a month. The use of
water that does not comply with the table above may have severe impact on the working life of the sterilizer and can invalidate the manufacturer’s guarantee.
Page 11 of 129 Pages
6 CONTROL SYSTEM DESCRIPTION
The control system is based on 2 electronic boards, see the diagram below.
Control system block diagram
PC
Keypad
Main Board
Digital input
Press. Sensors
electrodes
Jack. temp
Chamb. Temp
Door Lock
Solenoid valves
Power Transformer
DISPLAY
AC Board
Heaters
Vacuum Pump
L N GND Power Inlet
The control system contains the parts listed below:
Main board AC high voltage power board Transformer (230VAC primary / 6, 12, 18, 24VAC secondary) Electrical wires
6.1 The main board includes:
Stabilized power supply 24VAC & 5VAC Analog inputs for reading the analog sensors (temperature sensors,
pressure sensor, water level electrodes)
Digital inputs for reading the digital indicators (door switch, float
reservoir if available)
Digital outputs to control the solenoid valves and the door lock. Digital outputs to send signal to high voltage board or power
element drivers to control power elements.
Display A memory logging the previous cycles Drivers for printer Real time clock to serve as clock to the system Connection to the printer. It is connected directly to this board
through a flat cable 24 pins.
DPU30 Printer
Water Pump
Power Circuit Breaker
Page 12 of 129 Pages
6.2 The high voltage board (AC Board) includes:
High power outputs to control the heaters and the pumps. Circuit breaker The board provides four outputs for four high AC devices as
follows:
1. Steam generator heater (2200 Watt).
2. Chamber wall heater (500 Watt).
3. Vacuum pump.
4. Water pump.
Connection between the 2 cards
Page 13 of 129 Pages
7 CHECKING AND CHANGING PARAMETERS AND OTHER DATA
For special application it may be required to change parameters.
Note! The autoclave is sent from the manufacturer with validated parameters. Any change of the parameters requires a validation of the autoclave. The device is identified by its software and parameter versions. Any change of the parameters will increase the parameter version number.
In order to change parameters and check various components, you can enter a
few sub-directories and check or change the required data.
These sub-directories include the following:
Sub-directory Operation
Parameters Changing parameters
Digital Inputs Checking digital inputs
Digital Outputs Checking digital outputs
Analog Inputs Checking analog inputs
Calibration Calibration temperature and pressure
Set Clock Setting the clock (time and date)
Printer Test Testing the printer
History Printing the last cycles
Page 14 of 129 Pages
7.1 Menu
Follow the instructions below for entering the sub-directories and
performing the menu operation.
1. Main screen
134°C 004’ 001’
30.0°C 100kPaG
08:03:06
Turn on the autoclave with the main power switch located on the right side of the autoclave (see FRONT VIEW).
The main screen is displayed.
To enter the menu press simultaneously the UP and DOWN pushbuttons.
2. Enter Code
ENTER CODE
0333
SET EXIT
ENTER CODE will be displayed.
Enter the technician code (0333) as follows:
The cursor is under the right digit. To
change the right digit, press the UP or
DOWN pushbuttons. Each press on
the UP pushbutton will increase the
digit by 1 and each press on the DOWN
pushbutton will decrease it by 1.
Change this digit as required and press
the START/STOP pushbutton to
move the cursor to the second digit from
the right.
Repeat this until the curser is under SET
(SET is blinking), and then press the UP
pushbutton to enter the MENU.
3. Menu The MENU contains 1 1 sub-directories
(see display on the left).
The 7 top sub-director ies are disp layed
when entering "Menu" and the remaining 4 directories are displayed when rolling dow n. Move to the required sub-directory using the UP and DOWN pushbutton. The pointed sub-directory is blinking.
Pressing START/STOP pushbutton
will select the blinking directory and the required directory will be displayed.
To return to the previous display press the
UP and DOWN pushbuttons until
reaching EXIT. When EXIT is blinking, press the START/STOP pushbutton and the previous screen will be displayed.
Page 15 of 129 Pages
7.2 Parameters
This section describes the parameters, how they affect the process and
the way to change them. Listed bellow are all the available parameters. Each section describes the parameter, the minimum and maximum
allowed values and the changing resolution. Also included are the pre-
set values of the parameters for each cycle (default value).
NOTE:
If a parameter is modified, the only way to return to the original
value is to manually reenter it.
A global parameter is a parameter that by changing its value in
one program, it is changed in all the other programs to receive the same value.
To change a non-global value, i.e. specific parameter for each
program, choose first the required sterilization program in the main screen of the autoclave and then enter the MENU and the sub-directory PARAMETERS.
7.2.1 Ster Temp - sterilization temperature -
This parameter will set the desired temperature for sterilization
Resolution: 0.1°C
Minimum value: Default value
Maximum value: 136°C
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
134 134 121 121 121
B&D
Test
Fixed value
Vacuum
Test
Fixed value
7.2.2 Ster Time – sterilization time
This parameter will set the time desired for sterilization.
Resolution: 0.1 min
Minimum value: Default value
Maximum value: 99.9 min
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
4 4 20 20 20
B&D
Test
Fixed value
Vacuum
Test
Fixed value
Page 16 of 129 Pages
7.2.3 Dry Time – drying time
This parameter will set the time desired for drying.
Resolution: 0.1 min
Minimum value: Default value
Maximum value: 99.9 min
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
1 20 0 20 0
7.2.4 Vac Pulses – number of pulses
This parameter sets the number of vacuum pulses during the air
removal stage..
Resolution: 1
Minimum value: 0
Maximum value: 5
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
2 3 2 3 2
7.2.5 SterPressAdd - addition to the sterilization pressure
This defines the required addition to the sterilization pressure in
kPa in order to increase the sterilization temperature.
For example, for a sterilization temperature of 121°C the
required pressure is 205 kPa. Since the system controls the sterilization process according to pressure, if SterPressAdd equals “0”, the system will maintain the pressure at 205 kPa . If the value is at 5 kPa, the system will be maintained at 210 kPa and the temperature will be 121.8°C, and so on.
Resolution: 0.1 kPa
Minimum value: 0 kPa
Maximum value: 40 kPa
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
9 9 9 9 9
Delicate
121
Delicate
121
Delicate
121
B&D
Test
Fixed value
B&D
Test
Fixed value
B&D
Test
Fixed value
Vacuum
Test
Fixed value
Vacuum
Test
Fixed value
Vacuum
Test
Fixed value
Page 17 of 129 Pages
7.2.6 End Temp - temperature at the end of the cycle
This parameter will set the temperature that the chamber must
reach before the cycle can end.
Resolution: 0.1°C
Minimum value: 50°C
Maximum value:136°C
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
120 120 110 110 110
B&D
Test
Fixed value
Vacuum
Test
Fixed value
7.2.7 ATMPressure – ambient pressure
This parameter tells the unit the atmospheric pressure of the
location in which the autoclave is installed. The pressure entered must be with an accuracy of 5% of the actual atmospheric pressure for that location. It is a global parameter that means that you only need to enter this parameter once in anyone cycle and all cycles will be updated.
This value can easily be calculated by knowing the altitude of
your location. The atmospheric pressure at Sea Level is 100 kPa For every 100 meter above sea level, the atmospheric pressure
drops 1 kPa.
For every 100 meter below sea level, the atmospheric
pressure increases 1 kPa.
Changes in pressure do to weather will not affect the accuracy of
this unit.
Resolution: 0.1 kPa
Minimum value: 70 kPa
Maximum value: 110 kPa
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
100 kPa (global parameter)
Wdry
121
Delicate
121
B&D
Test
Vacuum
Test
7.2.8 Pulse vac 1 – Vacuum value in the first pulse
This parameter defines the vacuum value in pulse no.1 of the
prevacuum stage.
Resolution: 0.1°kPa
Minimum value: 5kPa
Maximum value:100kPa
Cycle
Default
Value
Flash 134
25.0 15.0 25.0 15.0 25.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Page 18 of 129 Pages
7.2.9 Pulse vac T1 – vacuum time after Pulse vac 1
This parameter defines the period that vacuum is built after
reaching the value defined in Pulse vac 1.
Resolution: 1sec Minimum value: 5sec Maximum value:360sec
Cycle
Default
Value
Flash 134
10
Wdry
134
10 30 10 30 10
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed value
7.2.10 Pulse press 1 - Pressure value in the first pulse
This parameter defines the pressure value (above zero) in pulse
no.1 of the prevacuum stage.
Resolution: 0.1kPa Minimum value: 70kPa Maximum value:200kPa
Cycle
Default
Value
Flash 134
140.0 140.0 140.0 140.0 140.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.11 Pulse vac 2 – Vacuum value in middle pulses
This parameter defines the vacuum value in all pulses except
first and last pulse of the prevacuum stage.
Resolution: 0.1kPa Minimum value: 5kPa Maximum value:100kPa
Cycle
Default
Value
Flash 134
25.0 20.0 25.0 20.0 25.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.12 Pulse vac T2 - vacuum time after Pulse vac 2
This parameter defines the period that vacuum is built after
reaching the value defined in all pulses defined by Pulse vac 2.
Resolution: 1sec Minimum value: 1sec Maximum value:300sec
Cycle
Default
Value
Flash 134
10 30 10 30 10
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Page 19 of 129 Pages
7.2.13 Pulse press 2 - Pressure value in the middle pulses
This parameter defines the pressure value (above zero) in the
middle pulses of the prevacuum stage.
Resolution: 0.1kPa Minimum value: 70klPa Maximum value:200kPa
Cycle
Default
Value
Flash 134
140.0 140.0 140.0 140.0 140.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.14 Pulse vac 3 - Pressure value in the last pulses
This parameter defines the vacuum value in the last pulse of the
prevacuum stage.
Resolution: 0.1kPa Minimum value: 5kPa Maximum value:200kPa
Cycle
Default
Value
Flash 134
30.0 20.0 30.0 20.0 30.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.15 Pulse vac T3 - vacuum time after Pulse vac 3
This parameter defines the period that vacuum is built after
reaching the value defined in all pulses defined by Pulse vac 3.
Resolution: 1sec Minimum value: 1sec Maximum value:360sec
Cycle
Default
Value
Flash 134
10 30 10 30 10
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.16 Pulse press 3 - Pressure value in the last pulse
This parameter defines the pressure value (above zero) in the last
pulse of the prevacuum stage.
Resolution: 0.1kPa Minimum value: 7kPa Maximum value:200kPa
Cycle
Default
Value
Flash 134
140.0 140.0 140.0 140.0 140.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Page 20 of 129 Pages
7.2.17 Heat Exh on – Opening time of exhaust valve
This parameter defines the period that the exhaust valve is open,
during the heating stage, in order to remove the condensate.
Resolution: 0.1sec Minimum value: 1sec Maximum value: 50sec
Cycle
Default
Value
Flash 134
10 10 10 10 10
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.18 Heat Exh off - Closing time of exhaust valve
This parameter defines the period that the exhaust valve is
closed, during the heating stage.
Resolution: 0.1sec Minimum value: 0sec Maximum value: 1000sec
Cycle
Default
Value
Flash 134
200 200 200 200 200
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.19 Ster Exh on - Opening time of exhaust valve
This parameter defines the period that the exhaust valve is open,
during the sterilization stage.
Resolution: 0.1sec Minimum value: 0sec Maximum value: 50sec
Cycle
Default
Value
Flash 134
5 5 5 5 5
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.20 Ster Exh off - Closing time of exhaust valve
This parameter defines the period that the exhaust valve is
closed, during the sterilization stage.
Resolution: 0.1sec Minimum value: 0sec Maximum value: 1000sec
Cycle
Default
Value
Flash 134
200 200 200 200 200
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Page 21 of 129 Pages
7.2.21 Exh Shoot on - Opening time of exhaust valve
This parameter defines the period that the exhaust valve is open,
during the exhaust stage.
Resolution: 0.1sec Minimum value: 0sec Maximum value: 1000sec
Cycle
Default
Value
Flash 134
1 1 1 1 10
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.22 Exh Shoot off - Closing time of exhaust valve
This parameter defines the period that the exhaust valve is
closed, during the exhaust stage.
Resolution: 0.1sec Minimum value: 0sec Maximum value: 1000sec
Cycle
Default
Value
Flash 134
0 0 0 0 100
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.23 Dry Air on - Opening time of air valve
This parameter defines the period that the air valve is open,
during the drying stage.
Resolution: 0.1sec Minimum value: 0sec Maximum value: 1000sec
Cycle
Default
Value
Flash 134
20 20 20 20 20
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.24 Dry Air off - Closing time of air valve
This parameter defines the period that the air valve is closed,
during the drying stage.
Resolution: 0.1sec Minimum value: 0sec Maximum value: 1000sec
Cycle
Default
Value
Flash 134
200 200 200 200 200
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Page 22 of 129 Pages
7.2.25 SterGenPrsAd – Maximum pressure in the generator
The value of this parameter plus the sterilization pressure define the maximum pressure in the generator in each program.
Resolution: 0.1kPa Minimum value: 0kPa Maximum value:50kPa
Cycle
Default
Value
Flash 134
20.0 20.0 20.0 20.0 20.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
7.2.26 Ster PrintT – printing rate during the sterilization stage
Test Fixed value
This parameter will set the interval between printings during the sterilization stage.
Resolution: 1 sec Minimum value: 10 sec Maximum value: 360 sec
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
60 sec (global parameter)
B&D
Test
Vacuum
Test
7.2.27 Print Rate - printing rate except during the sterilization stage
This parameter will set the interval between printings during the other stages.
Resolution: 1 sec Minimum value: 10 sec Maximum value: 360 sec
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
180 sec (global parameter)
121
Wdry
121
Delicate
121
B&D
Test
Vacuum
Test
7.2.28 HeatSter Time – Pre-sterilization heating time
This parameter defines the heating time before sterilization. Increasing the value of this parameter will decrease the heating gradient define.
Resolution: 1 sec Minimum value: 0 sec Maximum value: 1000 sec
Cycle
Default
Value
Flash 134
1 1 1 1 1
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Page 23 of 129 Pages
7.2.29 WaterGen Time – Addition time for adding w ater into the
generator
This parameter defines the addition time that water enters the
generator after the generator's water level electrode senses water..
Resolution: 1 sec Minimum value: 0 sec Maximum value: 30 sec
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
10 10 10 10 10
7.2.30 SterPrsLimAd - maximum pressure in the chamber
This parameter plus the sterilization pressure define the
maximum pressure in the chamber.
Resolution: 0.1kPa Minimum value: 0kPa Maximum value: 40kPa
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
14.0 14.0 14.0 14.0 14.0
7.2.31 Heat Time Err - maximum time of the heating stage
This parameter defines the maximum time of the heating stage
until the beginning of the sterilization stage. If the heating continues beyond this limit "ERROR 15" message will be displayed and the cycle will fail.
Resolution: 1 sec Minimum value: 1000 sec Maximum value: 3600 sec
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
1500 1500 1500 1500 1500
Delicate
121
Delicate
121
Delicate
121
B&D
Test
Fixed
value
B&D
Test
Fixed
value
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Vacuum
Test Fixed value
Vacuum
Test Fixed value
Page 24 of 129 Pages
7.2.32 Vac Time Err - maximum time of the vacuum stage
This parameter defines the maximum time of the vacuum stage.
If the vacuum stage continues beyond this limit "ERROR 14" message will be displayed and the cycle will fail.
Resolution: 1 sec Minimum value: 600 sec Maximum value: 3600 sec
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
1000 1000 1000 1000 1000
7.2.33 WaterTimeErr - maximum generator water filling time.
This parameter defines the maximum time for filling water in the
generator. If, before a cycle, the water pump has operated as long as defined by this parameter and the water level electrode did not sense water, "ERROR 20" will be displayed and a cycle cannot be started. If this will be during a cycle – the cycle will fail with "low temp" or "low pres".
Resolution: 1 sec Minimum value: 10 sec Maximum value: 200 sec
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
90 90 90 90 90
7.2.34 Sleep Power - time until the autoclave turns into "sleep
mode"
This parameter defines the time until the autoclave turns into
"sleep mode". This time is counted from the last operation of the autoclave or the last time a pushbutton has been pressed.
Resolution: 0.1hour Minimum value: 0.1hour Maximum value: 999.9hour
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
4.0 4.0 4.0 4.0 4.0
Delicate
121
Delicate
121
Delicate
121
B&D
Test
Fixed
value
B&D
Test
Fixed
value
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Vacuum
Test Fixed value
Vacuum
Test Fixed value
Page 25 of 129 Pages
7.2.35 TempInF – units of the temperature
This parameter enables the technician to set the displayed
temperature in ºC or in ºF.
Changing this parameter in one of the cycles will change this
parameter in all the other cycles to the value. Access Code available upon request Resolution – 1 Value 0 or 1 If TemInf = 1 the temperature is expressed in ºF. If TemInf = 0 the temperature is expressed in ºC.
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
0 (global parameter)
Delicate
121
B&D
Test
7.2.36 PressInPSI – units of the pressure
Vacuum
Test
This parameter enables the technician to set the displayed
pressure in Kpa, psig or psia. Changing this parameter in one of the cycles will change this
parameter in all the other cycles to the value. Access Code available upon request Resolution – 1 Value 0, 1 or 2 If PresInPSI = 0 the pressure is expressed in kPa. If PresInPSI = 1 the pressure is expressed in psia. If PresInPSI = 2 the pressure is expressed in psig.
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
0 (global parameter)
Delicate
121
B&D
Test
Vacuum
Test
7.2.37 Auto Add Water – automatic water filling
This parameter defines the possibility for automatic mineral free
water filling. NOTES: 1. Automatic water filling requires connection to
mineral free water supply and to drain.
2. Not applicable on this unit
Automatic water filling 1 Manual water filling 0
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
0 0 0 0 0
B&D
Test
Fixed
value
Vacuum
Test Fixed value
Page 26 of 129 Pages
7.2.38 ElectFillWtr – water conductivity
This parameter defines the conductivity level of the mineral free
water in which the water level electrode senses water. .
Resolution: 1 Minimum value: 0 – infinite conductivity Maximum value:9999 – infinite resistance (zero conductivity)
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
6000 6000 6000 6000 6000
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.39 JackTempStBy - temperature during the Stand-By mode
This parameter defines the temperature of the chamber's heating
element during the Stand-By mode.
Resolution: 0.1°F Minimum value: 68°F Maximum value:284°F
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
140.0 140.0 140.0 140.0 140.0
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.40 JackTempProc – temperature of the chamber's heating element
This parameter defines the temperature of the chamber's heating
element during the sterilization stage.
Resolution: 0.1°F Minimum value: 32.1°F Maximum value:284°F
Cycle
Default
Value
Flash 134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
248.8 248.0 239.0 239.0 239.0
B&D
Test
Fixed
value
Vacuum
Test Fixed value
7.2.41 FixGenPress
N/A
Page 27 of 129 Pages
e
This directory enables monitoring the status of the digital inputs. It is
possible to view the digital inputs during cycle.
7.3 Digital Inputs
1. To move from one item to another item use the UP and DOWN
pushbuttons.
2. To exit to the menu go to EXIT and when it is blinking press the
START/STOP pushbuttons.
Displayed
item
Thermostat (not
applicabl
on this model)
Float RESERV (not applicable on this model)
Door Switch
Float Res (not applicable on this model)
Displayed
symbol
Operation Status
Thermostat disconnects the
The temperature is above the
acceptable temperature. heaters. The temperature is below the
acceptable temperature. Senses water in the clean water
The float doesn’t sense water in
the clean water reservoir. reservoir. The float senses water in the
clean water reservoir. Senses if the door is open or closed.
The sensor senses the door is
closed.
The sensor senses the door is
open. Senses water in
the waste water
The float doesn’t sense water in
the waste water reservoir. reservoir.
The float senses water in the
waste water reservoir.
Page 28 of 129 Pages
Loading...
+ 101 hidden pages