This Technician's Manual, together with the operator’s manual, forms the
complete set of the Operation and Maintenance instructions for the Elara 11 pre
and post vacuum autoclave. This manual is intended for the use of the
technician. It is strongly recommended that only qualified and Tuttnauer
factory trained personnel service this autoclave and do so in accordance with
the instructions in this manual. Any unauthorized service may result in the
invalidation of the manufacturer’s guarantee.
2 SYMBOL DESCRIPTION
Caution! Consult accompanying documents
Caution! Hot surface
Caution! Hot steam.
Protective earth (Ground)
Page 4 of 129 Pages
Page 7
FRONT VIEW
1 23
17
16
5 46
718
8
9
1314
No.Description No. Description
1 Water reservoir funnel 10 Operating keypad
2 Autoclave cover 11 Printer cover
3 Water level sight gauge 12 Printer (standard)
4 Mineral-free water reservoir cover 13 Legs
Chamber and steam generator safety
5
valves
6 Waste water reservoir cover 15 Waste water reservoir drain valve
Display 16 Door cover
7
8 Ventilation grill 17 Door switch
9 Main switch/Circuit breaker 18 Door closing device
Mineral-free water reservoir drain
14
valve
1011151312
Page 5 of 129 Pages
Page 8
REAR VIEW
1
2
3
4
5
6
3
No.Description
1 Mineral-free water reservoir cover
2 Waste water reservoir cover
3 Ventilation grills
4 Air filter service cover
5 Opening for calibration
6 Main power electric cable socket
Page 6 of 129 Pages
Page 9
3 INSTALLATION
CAUTION:
Any operation described in section 3 shall be performed by a qualified
technician only.
3.1 Lifting and carrying
CAUTION:
Before moving the a utoclave, Make sure that the electric cord is
disconnected from the pow er, and there is no pressure in th e
chamber and in the generator.
Attention! The pressure of the generator does not decrease
immediately when the equipment is turned off. Wait approx. ½ an
hour to ve rify that the pressu re has decreased to atmospheric
pressure.
1. Disconnect the power supply cord.
2. Drain the water from both reservoirs.
To avoid injuries, lifting and carrying should be done with at least two
persons or by using a fork-lift or any other mechanical aid.
Do not drop the device!
3.2 Unpacking the autoclave
Unpack the autoclave and inspect for mechanical damage upon receipt.
Observe packing method and retain packing materials until the unit has
been inspected. Mechanical inspection involves checking for signs of
physical damage such as: scratched panel surfaces, broken knobs, etc.
To avoid injuries, lifting and carrying of the autoclave should be done
with at least two persons or by using a fork-lift or any other mechanical
aid.
3.3 Installation preparations
1. Check and verify that the counter carrying the autoclave isarigidand
leveledsurfaceandcancarryaloadof225 lbs (102kg).
Attention:
The Elara11 is not designed for use on any standard slide out shelf. If
it is necessary to use a slide out shelf, it must be tested and/or rated for
225 lb (102kg) or more.
Check and verify that the counter dimensions are, at least, 22” wide x
fromthewalltoallow ventilation and facilitate the device
disconnection.
3. If placed in a cabinet, verify that the rear of the cabinet is open to
allow ventilation.
4. Insufficient space for ventilation may result in an increase of
the
autoclave's temperature that may cause a malfunction or damage the
instrument.
5. It is recommended that enough space be left around the autoclave to
give a technician access for servicing the machine.
6. Check and verify that the room ventilation is 10 cycles per hour
minimum.
Page 7 of 129 Pages
Page 10
7. Check and verify that the ambient temperature range is 41˚F - 104˚F
(5˚C - 40˚C), it is preferable not to exceed 86˚F (30˚C).
8. Check and verify that the ambient relative humidity does not exceed
85%.
3.4 Connections to Utility Supplies
1. Check and verify that the power supply is a 1 phase, 230Vac ±5%,
50/60Hz (as appropriate), 15A supply.
2. Check and verify that the autoclave is connected to separate power
source to avoid flickers of light or sensitive devices.
3. Check and verify grounding of the autoclave.
4. Check and verify that the electrical net is protected with a current
leakage safety relay.
3.5 Final adjustments
At the time of installation, befo re the autoclave can be okayed for
daily operation the service technician needs to perform the follow ing
preliminary checks:
a. Integrity Check
Perform a visual check to verify that there are no dents, scratches or
broken components on the autoclave.
b. Support surface checkMake a visual check that the surface is level and strong enough to
support the autoclave.
c. The leakage current test
Test the precise operation of the earth leakage relay.
d. Ground Check
Test the continuity of the grounding connection.
e. Safety Valve Check
Test the safety valve as per the instructions in "Checking the Safety
Valve" in the Operator’s Manual.
f. Door check
Ensure that the door locking mechanism is functioning properly
g. Setting clock and date
See instruction in the user manual clause. 7.3.
h. Atmospheric Pressure check: the unit is set from the factory at 100
kpa, this is atmospheric pressure at sea level. If the unit is located
more that 500 ft above or below sea level then the Atmos. Press
parameter must be set, see sec 7.2.7
i. Cycle check: run a B&D Test to ensure that all systems in the unit
are functioning properly.
j. Reset the autoclave see sec 8.
After the above steps are performed, the autoclave is ready for daily
operation.
Page 8 of 129 Pages
Page 11
3.6 Operating the autoclave
1 Plug the power cord into the power socket.
2 Turn on the Main Switch / Circuit Breaker (see front view).
3 Select "Vacuum Test" cycle to keep the steam generator and heating
elements from heating up. See Selecting a Program in the Operator’s
Manual.
4 Open the door of the autoclave and remove the trays and the
packaging material.
5 Fill the Mineral Free Water Reservoir with water meeting the quality
specs in section 5.1 as follows:
5.1. Fill with 4 liters of mineral free water by pouring it into the
front reservoir fill opening at the top of the machine (see front
view).
5.2. Fill the remaining quantity by pouring water gently, into the
front funnel until it reaches the required level on the water level
sight gauge (See Front View).
6 Insert a paper roll in the printer (see printer handling in the
Operator’s Manual).
7 Close the door and perform a Vacuum Test. If the test fails perform
another test since the fail may be a result of moistureintheair.Ifthe
secondtestfailsitmaybenecessary to run Program 2 with the long
drying cycle to remove any excess moisture that is causing the test to
fail.
8 If the Vacuum Test is successful then select the B&D Test cycle.
At this stage the chamber and steam generator will be heating up. It
will take approximately 15 minutes (from selecting the B&D Test).
9 While waiting for the unit to heat up explain and instruct the operator
as follows (use the operation manual as reference):
9.1. Operation principals of the autoclave.
9.2. Preparation for sterilizing instruction including loading
instructions.
9.3. Intended use of each cycle.
9.4. Selecting a cycle.
9.5. Water filling method.
9.6. Displayed error and operational messages.
9.7. Monitoring and changing parameters.
9.8. Printer handling.
9.9. Maintenance instructions
10 Perform a B&D test with a chemical indicator.
11 The operator shall perform a cycle under supervision of the
technician.
Page 9 of 129 Pages
Page 12
4 PERIODICAL TESTS
PERIOD TEST
1 month Test the safety valve by operating it.
6 months
Removetheautoclave’scover,tightenthe heaters’ screws and
electricalconnections,valvesandconnectors in the control box.
Check the continuity of the grounding connections.
Check the temperature and pressure calibration.
Perform validation of the autoclave.
Check the precise operation of the earth leakage relay.
Check that the autoclave is leveled.
Check the safety elements; safety valve, safety and cut-off
thermostats door locking mechanisms.
Year
Run basic programs of the autoclave and check the operation
sequences, the sterilization parameters etc.
Checkthewaterreservoir,piping,plasticpartsand electric wires.
Check and tighten the piping joints to avoid leakage.
Check and tighten all screw connections in the control box, heaters
and valves and instrumentation.
Calibrate the temperature and pressure once a year or in reference to
local rules or regulations (refer to the section on Calibration).
5 years Observe the closing device for excessive wear
Safetytests(pressurevessel,efficiency,electrical) shall be performed in
accordance with local rules or regulations, by an authorized inspector.
Only an authorized technician shall perform the 6-months and yearly tests!
Page 10 of 129 Pages
Page 13
5 WATER QUALITY
5.1 Water for Generating Steam
The distilled or mineral – free water supplied to the sterilizer shall be
according to the table below:
Physical Characteristics and Maximum acceptable contaminants
Evaporate residue ≤15 miligrams/liter (mg/l)
Silica ≤2 mg/l
Iron ≤0.2 mg/l
Cadmium ≤0.005 mg/l
Lead ≤ 0.05 mg/l
Rest of heavy metals ≤0.1 mg/l
Chloride ≤3 mg/l
Phosphate ≤0.5 mg/l
Conductivity ≤50 μs/cm
levels in water for sterlizers
(According to ANSI/AAMI ST79:2006).
Element Condensate – allowable content
pH value 6.5 to 8
Appearance Colorless, clean, without sediment
Hardness ≤0.1 mmol/l
Compliance with the above data should be tested in accordance with
acknowledged analytical methods, by an authorized laboratory.
Attention: We recommend testing the water quality once a month. The use of
water that does not comply with the table above may have severe
impact on the working life of the sterilizer and can invalidate the
manufacturer’s guarantee.
Page 11 of 129 Pages
Page 14
6 CONTROL SYSTEM DESCRIPTION
The control system is based on 2 electronic boards, see the diagram below.
Control system block diagram
PC
Keypad
Main Board
Digital input
Press. Sensors
electrodes
Jack. temp
Chamb. Temp
Door Lock
Solenoid valves
Power Transformer
DISPLAY
AC Board
Heaters
Vacuum Pump
L N GND Power Inlet
The control system contains the parts listed below:
Main board
AC high voltage power board
Transformer (230VAC primary / 6, 12, 18, 24VAC secondary)
Electrical wires
6.1 The main board includes:
Stabilized power supply 24VAC & 5VAC
Analog inputs for reading the analog sensors (temperature sensors,
pressure sensor, water level electrodes)
Digital inputs for reading the digital indicators (door switch, float
reservoir if available)
Digital outputs to control the solenoid valves and the door lock.
Digital outputs to send signal to high voltage board or power
element drivers to control power elements.
Display
A memory logging the previous cycles
Drivers for printer
Real time clock to serve as clock to the system
Connection to the printer. It is connected directly to this board
through a flat cable 24 pins.
DPU30 Printer
Water Pump
Power Circuit
Breaker
Page 12 of 129 Pages
Page 15
6.2 The high voltage board (AC Board) includes:
High power outputs to control the heaters and the pumps.
Circuit breaker
TheboardprovidesfouroutputsforfourhighACdevices as
follows:
1. Steam generator heater (2200 Watt).
2. Chamber wall heater (500 Watt).
3. Vacuum pump.
4. Water pump.
Connection between the 2 cards
Page 13 of 129 Pages
Page 16
7 CHECKING AND CHANGING PARAMETERS AND OTHER DATA
For special application it may be required to change parameters.
Note!
The autoclave is sent from the manufacturer with validated parameters.
Any change of the parameters requires a validation of the autoclave.
The device is identified by its software and parameter versions.
Any change of the parameters will increase the parameter version number.
In order to change parameters and check various components, you can enter a
few sub-directories and check or change the required data.
These sub-directories include the following:
Sub-directory Operation
Parameters Changing parameters
Digital Inputs Checking digital inputs
Digital Outputs Checking digital outputs
Analog Inputs Checking analog inputs
Calibration Calibration temperature and pressure
Set Clock Setting the clock (time and date)
Printer Test Testing the printer
History Printing the last cycles
Page 14 of 129 Pages
Page 17
7.1 Menu
Follow the instructions below for entering the sub-directories and
performing the menu operation.
1. Main screen
134°C
004’
001’
30.0°C
100kPaG
08:03:06
Turn on the autoclave with the main
power switch located on the right side of
the autoclave (see FRONT VIEW).
The main screen is displayed.
To enter the menu press simultaneously
the UPand DOWN pushbuttons.
2. Enter Code
ENTER CODE
0333
SET EXIT
ENTER CODE will be displayed.
Enter the technician code (0333) as
follows:
The cursor is under the right digit. To
change the right digit, press the UP or
DOWNpushbuttons. Each press on
the UP pushbutton will increase the
digit by 1 and each press on the DOWN
pushbutton will decrease it by 1.
Change this digit as required and press
the START/STOP pushbutton to
move the cursor to the second digit from
the right.
Repeat this until the curser is under SET
(SET is blinking), and then press the UP
pushbutton to enter the MENU.
3. Menu
The MENU contains 1 1 sub-directories
(see display on the left).
The 7 top sub-director ies are disp layed
when entering "Menu" and the
remaining 4 directories are displayed
when rolling dow n. Move to the
required sub-directory using the UP
andDOWN pushbutton. The pointed
sub-directory is blinking.
Pressing START/STOP pushbutton
will select the blinking directory and the
required directory will be displayed.
To return to the previous display press the
UP and DOWN pushbuttons until
reaching EXIT. When EXIT is blinking,
press the START/STOP pushbutton
and the previous screen will be displayed.
Page 15 of 129 Pages
Page 18
7.2 Parameters
This section describes the parameters, how they affect the process and
the way to change them.
Listed bellow are all the available parameters.
Each section describes the parameter, the minimum and maximum
allowed values and the changing resolution. Also included are the pre-
set values of the parameters for each cycle (default value).
NOTE:
If a parameter is modified, the only way to return to the original
value is to manually reenter it.
A global parameter is a parameter that by changing its value in
one program, it is changed in all the other programs to receive the
same value.
To change a non-global value, i.e. specific parameter for each
program, choose first the required sterilization program in the
main screen of the autoclave and then enter the MENU and the
sub-directory PARAMETERS.
7.2.1 Ster Temp - sterilization temperature -
This parameter will set the desired temperature for sterilization
Resolution: 0.1°C
Minimum value: Default value
Maximum value: 136°C
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
134 134 121 121 121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.2 Ster Time – sterilization time
This parameter will set the time desired for sterilization.
Resolution: 0.1 min
Minimum value: Default value
Maximum value: 99.9 min
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
4 4 20 20 20
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 16 of 129 Pages
Page 19
7.2.3 Dry Time – drying time
This parameter will set the time desired for drying.
Resolution: 0.1 min
Minimum value: Default value
Maximum value: 99.9 min
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
1 20 0 20 0
7.2.4 Vac Pulses – number of pulses
This parameter sets the number of vacuum pulses during the air
removal stage..
Resolution: 1
Minimum value: 0
Maximum value: 5
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
2 3 2 3 2
7.2.5 SterPressAdd - addition to the sterilization pressure
This defines the required addition to the sterilization pressure in
kPa in order to increase the sterilization temperature.
For example, for a sterilization temperature of 121°C the
required pressure is 205 kPa. Since the system controls the
sterilization process according to pressure, if SterPressAdd
equals “0”, the system will maintain the pressure at 205 kPa . If
the value is at 5 kPa, the system will be maintained at 210 kPa
and the temperature will be 121.8°C, and so on.
Resolution: 0.1 kPa
Minimum value: 0 kPa
Maximum value: 40 kPa
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
9 9 9 9 9
Delicate
121
Delicate
121
Delicate
121
B&D
Test
Fixed
value
B&D
Test
Fixed
value
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Vacuum
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 17 of 129 Pages
Page 20
7.2.6 End Temp - temperature at the end of the cycle
This parameter will set the temperature that the chamber must
reach before the cycle can end.
Resolution: 0.1°C
Minimum value: 50°C
Maximum value:136°C
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
120 120 110 110 110
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.7 ATMPressure – ambient pressure
This parameter tells the unit the atmospheric pressure of the
location in which the autoclave is installed. The pressure entered
must be with an accuracy of 5% of the actual atmospheric
pressure for that location. It is a global parameter that means that
you only need to enter this parameter once in anyone cycle and
all cycles will be updated.
This value can easily be calculated by knowing the altitude of
your location. The atmospheric pressure at Sea Level is 100 kPa
For every 100 meter above sea level, the atmospheric pressure
drops 1 kPa.
For every 100 meter below sea level, the atmospheric
pressure increases 1 kPa.
Changes in pressure do to weather will not affect the accuracy of
this unit.
Resolution: 0.1 kPa
Minimum value: 70 kPa
Maximum value: 110 kPa
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
100 kPa (global parameter)
Wdry
121
Delicate
121
B&D
Test
Vacuum
Test
7.2.8 Pulse vac 1 – Vacuum value in the first pulse
This parameter defines the vacuum value in pulse no.1 of the
prevacuum stage.
Resolution: 0.1°kPa
Minimum value: 5kPa
Maximum value:100kPa
Cycle
Default
Value
Flash
134
25.0 15.0 25.0 15.0 25.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 18 of 129 Pages
Page 21
7.2.9 Pulse vac T1 – vacuum time after Pulse vac 1
This parameter defines the period that vacuum is built after
reaching the value defined in Pulse vac 1.
Resolution: 1sec
Minimum value: 5sec
Maximum value:360sec
Cycle
Default
Value
Flash
134
10
Wdry
134
10 30 10 30 10
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.10 Pulse press 1 - Pressure value in the first pulse
This parameter defines the pressure value (above zero) in pulse
no.1 of the prevacuum stage.
Resolution: 0.1kPa
Minimum value: 70kPa
Maximum value:200kPa
Cycle
Default
Value
Flash
134
140.0 140.0 140.0 140.0 140.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.11 Pulse vac 2 – Vacuum value in middle pulses
This parameter defines the vacuum value in all pulses except
first and last pulse of the prevacuum stage.
Resolution: 0.1kPa
Minimum value: 5kPa
Maximum value:100kPa
Cycle
Default
Value
Flash
134
25.0 20.0 25.0 20.0 25.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.12 Pulse vac T2 - vacuum time after Pulse vac 2
This parameter defines the period that vacuum is built after
reaching the value defined in all pulses defined by Pulse vac 2.
Resolution: 1sec
Minimum value: 1sec
Maximum value:300sec
Cycle
Default
Value
Flash
134
10 30 10 30 10
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 19 of 129 Pages
Page 22
7.2.13 Pulse press 2 - Pressure value in the middle pulses
This parameter defines the pressure value (above zero) in the
middle pulses of the prevacuum stage.
Resolution: 0.1kPa
Minimum value: 70klPa
Maximum value:200kPa
Cycle
Default
Value
Flash
134
140.0 140.0 140.0 140.0 140.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.14 Pulse vac 3 - Pressure value in the last pulses
This parameter defines the vacuum value in the last pulse of the
prevacuum stage.
Resolution: 0.1kPa
Minimum value: 5kPa
Maximum value:200kPa
Cycle
Default
Value
Flash
134
30.0 20.0 30.0 20.0 30.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.15 Pulse vac T3 - vacuum time after Pulse vac 3
This parameter defines the period that vacuum is built after
reaching the value defined in all pulses defined by Pulse vac 3.
Resolution: 1sec
Minimum value: 1sec
Maximum value:360sec
Cycle
Default
Value
Flash
134
10 30 10 30 10
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.16 Pulse press 3 - Pressure value in the last pulse
This parameter defines the pressure value (above zero) in the last
pulse of the prevacuum stage.
Resolution: 0.1kPa
Minimum value: 7kPa
Maximum value:200kPa
Cycle
Default
Value
Flash
134
140.0 140.0 140.0 140.0 140.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 20 of 129 Pages
Page 23
7.2.17 Heat Exh on – Opening time of exhaust valve
This parameter defines the period that the exhaust valve is open,
during the heating stage, in order to remove the condensate.
Resolution: 0.1sec
Minimum value: 1sec
Maximum value: 50sec
Cycle
Default
Value
Flash
134
10 10 10 10 10
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.18 Heat Exh off - Closing time of exhaust valve
This parameter defines the period that the exhaust valve is
closed, during the heating stage.
Resolution: 0.1sec
Minimum value: 0sec
Maximum value: 1000sec
Cycle
Default
Value
Flash
134
200 200 200 200 200
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.19 Ster Exh on - Opening time of exhaust valve
This parameter defines the period that the exhaust valve is open,
during the sterilization stage.
Resolution: 0.1sec
Minimum value: 0sec
Maximum value: 50sec
Cycle
Default
Value
Flash
134
5 5 5 5 5
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.20 Ster Exh off - Closing time of exhaust valve
This parameter defines the period that the exhaust valve is
closed, during the sterilization stage.
Resolution: 0.1sec
Minimum value: 0sec
Maximum value: 1000sec
Cycle
Default
Value
Flash
134
200 200 200 200 200
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 21 of 129 Pages
Page 24
7.2.21 Exh Shoot on - Opening time of exhaust valve
This parameter defines the period that the exhaust valve is open,
during the exhaust stage.
Resolution: 0.1sec
Minimum value: 0sec
Maximum value: 1000sec
Cycle
Default
Value
Flash
134
1 1 1 1 10
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.22 Exh Shoot off - Closing time of exhaust valve
This parameter defines the period that the exhaust valve is
closed, during the exhaust stage.
Resolution: 0.1sec
Minimum value: 0sec
Maximum value: 1000sec
Cycle
Default
Value
Flash
134
0 0 0 0 100
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.23 Dry Air on - Opening time of air valve
This parameter defines the period that the air valve is open,
during the drying stage.
Resolution: 0.1sec
Minimum value: 0sec
Maximum value: 1000sec
Cycle
Default
Value
Flash
134
20 20 20 20 20
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
7.2.24 Dry Air off - Closing time of air valve
This parameter defines the period that the air valve is closed,
during the drying stage.
Resolution: 0.1sec
Minimum value: 0sec
Maximum value: 1000sec
Cycle
Default
Value
Flash
134
200 200 200 200 200
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 22 of 129 Pages
Page 25
7.2.25 SterGenPrsAd – Maximum pressure in the generator
The value of this parameter plus the sterilization pressure define
the maximum pressure in the generator in each program.
Resolution: 0.1kPa
Minimum value: 0kPa
Maximum value:50kPa
Cycle
Default
Value
Flash
134
20.0 20.0 20.0 20.0 20.0
Wdry
134
No Dry
121
Wdry
121
Delicate
121
B&D
Test
Fixed
value
Vacuum
7.2.26 Ster PrintT – printing rate during the sterilization stage
Test
Fixed
value
This parameter will set the interval between printings during the
sterilization stage.
7.2.30 SterPrsLimAd - maximum pressure in the chamber
This parameter plus the sterilization pressure define the
maximum pressure in the chamber.
Resolution: 0.1kPa
Minimum value: 0kPa
Maximum value: 40kPa
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
14.0 14.0 14.0 14.0 14.0
7.2.31 Heat Time Err - maximum time of the heating stage
This parameter defines the maximum time of the heating stage
until the beginning of the sterilization stage. If the heating
continues beyond this limit "ERROR 15" message will be
displayed and the cycle will fail.
7.2.33 WaterTimeErr - maximum generator water filling time.
This parameter defines the maximum time for filling water in the
generator. If, before a cycle, the water pump has operated as
long as defined by this parameter and the water level electrode
did not sense water, "ERROR 20" will be displayed and a cycle
cannot be started. If this will be during a cycle – the cycle will
fail with "low temp" or "low pres".
7.2.34 Sleep Power - time until the autoclave turns into "sleep
mode"
This parameter defines the time until the autoclave turns into
"sleep mode". This time is counted from the last operation of the
autoclave or the last time a pushbutton has been pressed.
Resolution: 0.1hour
Minimum value: 0.1hour
Maximum value: 999.9hour
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
4.0 4.0 4.0 4.0 4.0
Delicate
121
Delicate
121
Delicate
121
B&D
Test
Fixed
value
B&D
Test
Fixed
value
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Vacuum
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 25 of 129 Pages
Page 28
7.2.35 TempInF – units of the temperature
This parameter enables the technician to set the displayed
temperature in ºC or in ºF.
Changing this parameter in one of the cycles will change this
parameter in all the other cycles to the value.
Access Code – available upon request
Resolution – 1
Value – 0 or 1
If TemInf = 1 the temperature is expressed in ºF.
If TemInf = 0 the temperature is expressed in ºC.
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
0 (global parameter)
Delicate
121
B&D
Test
7.2.36 PressInPSI – units of the pressure
Vacuum
Test
This parameter enables the technician to set the displayed
pressure in Kpa, psig or psia.
Changing this parameter in one of the cycles will change this
parameter in all the other cycles to the value.
Access Code – available upon request
Resolution – 1
Value – 0, 1 or 2
If PresInPSI = 0 the pressure is expressed in kPa.
If PresInPSI = 1 the pressure is expressed in psia.
If PresInPSI = 2 the pressure is expressed in psig.
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
0 (global parameter)
Delicate
121
B&D
Test
Vacuum
Test
7.2.37 Auto Add Water – automatic water filling
This parameter defines the possibility for automatic mineral free
water filling.
NOTES: 1. Automatic water filling requires connection to
mineral free water supply and to drain.
2. Not applicable on this unit
Automatic water filling 1
Manual water filling 0
Cycle
Default
Value
Flash
134
Wdry
134
No Dry
121
Wdry
121
Delicate
121
0 0 0 0 0
B&D
Test
Fixed
value
Vacuum
Test
Fixed
value
Page 26 of 129 Pages
Page 29
7.2.38 ElectFillWtr – water conductivity
This parameter defines the conductivity level of the mineral free
water in which the water level electrode senses water. .
2. To exit to the menu go to EXIT and when it is blinking press the
START/STOP pushbuttons.
Displayed
item
Air
Vac valve
Chamb
Steam
Fast exh
Displayed
symbol
Operation Remarks
Not operating
Operating
Not operating
Operating
Not operating
Operating
Air valve
(valve 43)
Vacuum valve
(valve 52)
Steam inlet valve to the chamber
(valve 93)
Not operating Fast exhaust valve to the
Operating
reservoir
(valve 73)
Dig Out #5
Not operating
Operating
Optional
Wtr to Res
Air
VacPump
Water Valve
Dig Out #9
Door lock
Heater 1
Heater 2
Vacuum
pump
Water pump
Buzzer
Backlight
Not operating
Operating
Not operating
Operating
Not operating
Not operating
Operating
Locked position
Unlocked position
Not operating
Operating
Not operating
Operating
Not operating
Operating
Not operating
Operating
Off
Buzzing
On
Off
N/A
Air inlet valve to the vacuum
pump (valve 44)
Mineral free water to generator
(valve 21)
Optional
Door locking pin
Steam generator heating element
Chamber heating element
Vacuum pump
Water pump
Buzzer
Screen
Page 29 of 129 Pages
Page 32
7.5 Analog Inputs
n
n
This directory enables checking the analog Inputs. It is possible to enter
this directory while the autoclave is performing a cycle in order to check
the devices sending these inputs.
2. To exit to the menu go to EXIT and when it is blinking press the
START/STOP key.
Displayed item Description
ChambPress Pressure inside the chamber
Chamb Temp Temperature in the chamber
Coil Temp Temperature in jacket of the chamber
Gen Press Pressure in the steam generator
Electr Gen Level of water in the steam generator
ElectrLow
ElectrLow2
This analog input measures the level of water i
the clean water reservoir.
This analog input measures the level of water i
the waste water reservoir.
Page 30 of 129 Pages
Page 33
7.6 Calibration
7.6.1 Calibration components
This directory describes the calibration of
the following 4 sensors:
Sensor name Sensor function
1.
ChambPress
2. ChambTemp
3. CoilTemp
4. GenPress
7.6.2 Required equipment for calibration
Reads the pressure in the chamber.
Reads the temperature in the chamber.
Reads the temperature in the jacket.
Reads the pressure in the steam generator.
Reference temperature tool: (a bead probe thermocouple)
Single element sealing gland (to insert the thermocouple
through the validation port).
¼” BSP adaptor (for the validation port)
Reference pressure tool: vacuum / pressure gauge in kPa.
7.6.3 Calibrating the sensors
Choose the sensor required to
calibrate among the 4 sensors
available with the UP and
DOWN keys.
When it is blinking select it by
pressing the START/STOP
key.
Move to Calc. GainOffset with
the UP and DOWN
keys, and then when it is
blinking select it with the
START/STOP key
This is a typical calibration screen.
There are two rows of data, each
row has two values.
RH, AH are for calibrating the
upper end of the range
RL, AL are for calibrating the
lower end of the range
AH, AL are the values from your
test equipment
RH, RL are the values from the
autoclaves display
7.6.4 Calibration port
In order to perform the calibration it is required to measure the
temperature or pressure inside the chamber. This is done as
follows:
1. Remove the rear cover of the autoclave (see para. 9.3).
2. Unscrew the plug sealing the validation port.
3. The validation port size is ¼”BSP
Page 31 of 129 Pages
Page 34
4. Assemble to this port the required measuring tool
(temperature sensor or vacuum/pressure gauge) as specified
in the calibration instruction.
Validation port plug
7.6.5 Calibration operation
Calibration can be preformed in psi and °F, HOWEVER, it is
recommended that the calibrations be performed in kPa and °C
(see changing parameters sec 7.2.35 & 7.2.36).
Make sure the insulation is securely wrapped around the chamber.
The side and top panels should also be attached.
When calibrating the autoclave you will:
First, acquire the appropriate data, as instructed below.
Second, access the calibration page of the software (see sec 7.1 &
7.6.3)
Third, use the START/STOP key to move from digit to digit
and the UP / DOWNkeys to increase and decrease the
values according to your data.
7.6.5.1 Calibrating
“ChambPress”
thepressureinthechamber
Calibrate the low pressure of the chamber
Attach your reference pressure gauge to the validation
port on the back of the chamber and operate a vacuum
test. When the unit has reached its lowest vacuum level
record the reading of your gauge and the pressure
reading for the chamber on the screen of the autoclave.
It is not necessary to abort the cycle, the calibration can
be done while the unit is running the cycle. access the
calibration page for ChambPress then enter the data as
follows (if necessary refer back to 7.6.3):
AL(actualpressurelowvalue):changetheAL to the
valueshownbyyourreferencepressure gauge.RL
(displayed pressure low value): change the RL to the
value read from the autoclave display.
Page 32 of 129 Pages
Page 35
TosavethedatamovetoSETbypressingthe
START/STOP key and when it is blinking press
the UP key.
If you did not abort the cycle then the chamber
pressure on the screen should match your reference
gauge.
Calibrate the high pressure of the chamber
Attach your reference pressure gauge to the validation
port on the back of the chamber and operate a program
of 134°C. When the unit has reached sterilization
record the reading of your gauge and the pressure
reading for the chamber on the screen of the autoclave.
It is not necessary to abort the cycle, the calibration can
be done while the unit is running the cycle. access the
calibration page for ChambPress then enter the data as
follows (if necessary refer back to 7.6.3):
AH (actual pressure high value): changetheAH to
thevalueshownbyyourreferencepressure gauge.RH (displayed pressure high value): change the RH
to the value read from the autoclave display.
To save the data move to SET by pressing the
START/STOP key and when it is blinking press
the UP key.
If you did not abort the cycle then the chamber
pressure on the screen should match your reference
gauge.
7.6.5.2 Calibrating the temperature in the chamber
“ChambTemp”
Calibrate the low temperature of the chamber
When the autoclave is cold, using the sealing gland
insert your reference thermocouple through the
validation port and into the chamber. Position it near
the Pt100 inside the chamber. Close the chamber door
to allow the temperature inside the chamber to stabilize,
then record the reading of your thermocouple and the
reading on the screen of the autoclave. Access the
calibration page for Chamb Temp then enter the data as
follows (if necessary refer back to 7.6.3):
AL
(actual temperature low value): changetheAL
to thevalueshownbyyourreferencethermocouple.
(displayedtemperaturelowvalue):change the
RL
RL to the value read from the autoclave display.
To save the data move to SET by pressing the
START/STOP key and when it is blinking press
the UP key.
Now chamber temperature on the screen should
match your reference thermocouple.
Page 33 of 129 Pages
Page 36
Calibrate the high temperature of the chamber
Using the sealing gland, insert your reference
thermocouple through the validation port and into the
chamber. Position it near the PT100 and operate a
program of 134°C. When the unit has reached
sterilization record the reading of your thermocouple
and the reading on the screen of the autoclave.
It is not necessary to abort the cycle, the calibration can
be done while the unit is running the cycle. access the
calibration page for Chamb Temp then enter the data as
follows (if necessary refer back to 7.6.3):
AH
(actual temperature high value): change the AH
to thevalueshownbyyourreferencethermocouple.
RH (displayed temperature high value): change the
RH to the value read from the autoclave display.
To save the data move to SET by pressing the
START/STOP key and when it is blinking press
the UP key.
If you did not abort the cycle then the chamber
temperature on the screen should match your
reference thermocouple.
7.6.5.3 Calibrating the temperature in the jacket “CoilTemp”
Calibrate the low temperature of the jacket
Whentheautoclaveiscoldand turned off, peel back the
insulation at the back top of the chamber. On your right
will be a PT100 inserted under the heating element (see
picture). Insert the reference thermocouple between the
heating element and the chamber next to the PT100.
Turn the unit on and immediately select the Vacuum
Test Cycle, this will keep the heaters from turning on.
Access the calibration page for Coil Temp calibration.
Record the reading of your thermocouple and the
reading on the screen of the autoclave then enter the
data as follows (if necessary refer back to 7.6.3):
AL (actual temperature low value): changethe AL
to thevalueshownbyyourreferencethermocouple.RL (displayed temperature low value): change the
RL to the value read from the autoclave display.
Page 34 of 129 Pages
Jacket PT100
Page 37
To save the data move to SET by pressing the
START/STOP key and when it is blinking press
the UP key.
Now the jacket temperature on the screen should
match your reference thermocouple.
Calibratethehightemperatureof the chamber wall
Peel back the insulation at the back top of the chamber.
On your right will be a PT100 inserted under the
heating element (see picture). Insert the reference
thermocouple between the heating element and the
chamber next to the PT100.
Select a cycle of 134°C and wait for the autoclave-
preheating icon to go off the screen. Access the
calibration page for Coil Temp calibration. Record the
reading of your thermocouple and the reading on the
screen of the autoclave then enter the data as follows (if
necessary refer back to 7.6.3):
AH (actualtemperaturehigh value): changetheAH
to thevalueshownbyyourreferencethermocouple.RH (displayed temperature high value): change the
RH to the value read from the autoclave display.
To save the data move to SET by pressing the
START/STOP key and when it is blinking press
the UP key.
Now the jacket temperature on the screen should
match your reference thermocouple.
7.6.5.4 Calibrating the pres sure in the steam generator
“GenPress”
Calibrate the low pressure of the steam generator
When the unit is off and the steam generator is cold and
there is no pressure in it, attach your reference pressure
gauge to the calibration port on the generator manifold
(see picture). Turn the unit on and immediately select
the Vacuum Test Cycle, this will keep the heaters from
turning on. Access the calibration page for GenPress
calibration. Record the reading of your reference
pressure gauge and the pressure reading for the
generator on the screen of the autoclave then enter the
data as follows (if necessary refer back to 7.6.3):
AL (actualpressurelowvalue):changetheAL to the
valueshownbyyourreferencepressure gauge.
RL (displayed pressure low value): change the RL
to the value read from the autoclave display.
To save the data move to SET by pressing the
START/STOP key and when it is blinking press
UP key.
the
The steam generator pressure on the screen should
match your reference gauge.
Page 35 of 129 Pages
Page 38
Calibrate the high pressure of the steam generator
When the unit is off and the steam generator is cold and
there is no pressure in it, attach your reference pressure
gauge to the calibration port on the generator manifold
(see picture). Select a program of 134°C the generator
will begin to heat. When the unit has reached the ready
state record the reading of your gauge and the reading
for the generator on the screen of the autoclave.
Access the calibration page for GenPress then enter the
data as follows (if necessary refer back to 7.6.3):
AH (actual pressure high value): changetheAH to
thevalueshownbyyourreferencepressure gauge.RH (displayed pressure high value): change the RH
to the value read from the autoclave display.
To save the data move to SET by pressing the
START/STOP key and when it is blinking press
UP key.
the
The steam generator pressure on the screen should
match your reference gauge.
Generator
Page 36 of 129 Pages
Calibration port
Page 39
7.7 Setting the Clock
This directory enables the operator to set the time and date.
The SET CLOCK screen is displayed when entering the SET CLOCK
directory:
When entering the directory the curser will be blinking on the "hour"
digit.
The time is displayed in the upper row in the form “hh:mm:ss”. The
hour range is 24 hour (i.e. from “0” to “24”).
The date is displayed in the lower row in the form “DD:MM:YYYY”.
1. To increase or decrease the time or the date use the UP and
DOWN keys.
2. To move the curser from one digit to another press the
START/STOP key.
3. After changing the time and the date move the curser to “SAVE”.
4. Confirm the new time and date by pressing the UP key. While the
new time and date are being saved the following screen is
displayed:
5. After saving is completed, the SET CLOCK screen is displayed
again, move the cursor to EXIT and press UP to return to the
previous MENU screen.
6. The printer will print the date in the format DD:MM:YYYY if the
chamber temperature is being displayed in ˚C. It will print the date
in the format MM:DD:YYYY if the chamber temperature is being
displayed in ˚F.”
Page 37 of 129 Pages
Page 40
7.8 Printer Test
This option checks the good functioning of the printer.
1. Select "Printer Test" as described in section 7.1 "Menu".
2. Press the START/STOP key to perform the Printer Test.
3. Check that the message “Printer Test” is written on the paper.
Page 38 of 129 Pages
Page 41
7.9 Language
THIS OPTION IS NOT AVAILABLE YET
Page 39 of 129 Pages
Page 42
This directory enables you to print the 40 previous cycles.
The HISTORY screen is displayed when entering the HISTORY
directory:
1. Choose the required cycle according to its number (Log Load Nu)
with the UP and DOWN key
2. The words "Log Load Nu" are replaced by "printing" and the
required cycle data is printed by the built in printer.
3. In order to leave this screen go to the EXIT option by using the
DOWN key and select it by pressing the START/STOP key.
7.10 History
Page 40 of 129 Pages
Page 43
7.11 More options
This sub-directory enable the technician to perform a few additional
operation.
1. To enter this sub-directory Select "more options" as described in
section 7.1 "Menu".
2. Use UP and DOWN keys to browse through the 4 options +
"EXIT".
7.11.1 Calib Default
This option enables the technician to return to the default
calibration data as defined at the factory.
Caution
This operation is a non-return operation. Activating this
option will delete all calibration data and it will be impossible
to retrieve it.
1. Press the START/STOP key to return to the default
calibration values..
2. To return to the main menu– move the cursor to "EXIT"
7.11.2 Set Serial Num
This option enables the technician to set the autoclave's serial
number. This is necessary after replacing the electronic unit.
1. Press the START/STOP key to enter the "SERIAL
NUMBER" display.
2. Press the START/STOP key to move the cursor from the
right digit leftward.
3. Use UP and DOWN keys to change the value of each
digit.
4. Complete the change by moving the cursor to "SAVE" and
save the number by pressing UP or DOWN keys.
Page 41 of 129 Pages
Page 44
7.11.3 Reset Load Nu.
This option enables the technician to reset the autoclave's load
number.
Caution
This operation is a non-return operation. Activating this
option will reset the load number to "0" and it will be
impossible to return to the previous setting it.
1. Press the START/STOP key to reset the load number.
2. The load number will reset to "0" and the following will be
displayed.
7.11.4 Enable Programs
This option enables the technician to choose the programs that
will be enabled or disabled.
When entering this option the following is displayed.
1. Use UP and DOWN keys to browse through the
programs.
2. The number on the right side of the program name defines the
status of the program.
1 – enabled
0 – disabled
Press the START/STOP key to switch between 1 and 0.
3. It is possible to enable or disable any program, but at least
one program must remain enabled.
4. To exit this option press both UP and DOWN keys
simultaneously.
Page 42 of 129 Pages
Page 45
8 RESETTING THE AUTOCLAVE
Whenever it becomes necessary to restore the system to normal operation, the
system must be reset. This will remove corrupted data from memory and
restore a healthy program. On occasion other situations require that a reset be
performed, they are as follows:
When the machine is operated for the first time.
If the machine has been sitting unused for a long period of time.
To reset the system proceed as follows:
Turn-off the main power switch on the right side of the autoclave.
Turn-on the main switch while pressing and holding the DOWN ▼key
until the message “RESET DONE!!!” is displayed.
This will not erase any settings or history.
Page 43 of 129 Pages
Page 46
9 MAINTENANCE AND REPLACEMENT PROCEDURES
9.1 Preliminary Operations for Each Technician Call
1. In order to mainta in efficient service, the technician must
perform the following:
1. Inspect and clean if needed, the following:
Chamber, trays and trays holder (see Maintenance
Instructions in the Operator’s Manual).
Filters.
Seats and plungers of the solenoid valves (see sec. 9.21).
Water level electrode in the generator (see sec. 9.8).
Water reservoir(see sec. 9.27).
Steam generator (including descaling) (see sec. 9.32).
2. Visual inspection for leaks or corrosion in the piping elements
3. Tightening loose screws and piping joints.
4. Visual inspection of the wiring.
5. Calibration and logging the calibration of the temperature and
pressure.
2. After completing the work, the technician must perfo rm the
following cycles:
1. A Wdry 134 cycle with out load
2. A Vacuum Test
3. A B&D Test.
Note:The warranty does not cover cleaning or maintenance. These procedures
are the responsibility of the equipment owner.”
Page 44 of 129 Pages
Page 47
9.2 Safety tests after repair
ATTENTION!
After every repair or dismantling of the enclosure, the technician
should perform the following two electrical safety tests.
1. Enclosure Leakage Current Test.
Every autoclave should pass this test as follows:
1. Connect the electrical cord to the autoclave. DO NOT plug
into the power outlet.
2. Turn on the main switch on the right side of the autoclave.
3. Short-circuit the L and L pins on the cord's plug.
4. Connect the Short-circuit pins to the L pole on the Megger.
5. Connect the GND pin to the GND pole on the Megger.
6. Impose an electrical potential of 1000V±10% on the tested
autoclave. The insulation resistance should be at least 2 MΩ.
The test is successful if there was no leakage.
2. Protective Earth Impedance Test
1. Connect the grounding pin of the power cord plug to one pole
of an Ohmmeter.
2. Connect any other metallic part (preferable – the metallic part
of the locking screw) to the second pole of the Ohmmeter.
3. The resistance should not exceed 0.3 Ω.
After performing these tests, the Service Engineer should complete
and sign the Work Order.
Page 45 of 129 Pages
Page 48
9.3 Dismantling the Outer Covers of the Autoclave
Caution!
Before starting, disconnect the instrument from the po wer source
and ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before removing outer covers.
1. Unscrew the screws (2) holding the upper cover (1).
2. Remove the upper cover.
3. Unscrew the screws (4) holding the side covers (3).
4. Remove the side covers (3).
5. Disconnect the air filter (6) from the air filter cover (7).
6. Unscrew the screws (5) holding the rear covers (8).
7. Remove the rear cover (8).
8. Remove the grounding wires from the rear cover.
1
2
3
4
8 7
6
5
Page 46 of 129 Pages
Page 49
9.4 Replacing the Safety Valve
Caution!
Before starting, be sure that the electric cord is disconnected and that
there is no pressure in the chamber or generator.
Allow the autoclave to cool before removing outer covers.
The safety valve is installed to protect the system from over pressurizing
should all the electrical controls fail.
1. Take off the autoclave cover (see para. 9.3 “Removing the
Autoclave’s Outer Covers”).
2. Remove the water reservoir cover.
3. Unscrew the safety valve (2) and remove it from the safety valve
base (1).
4. Replace the valve with a new safety valve (install only an original
equipment replacement!). Use liquid thread seal on the threads to
seal it. Tighten the safety valve to prevent leaking.
5. To check the new safety valve, perform the following:
12
43
No. Description
1 Safety valve base
2 Safety valve
3 Pulling device
4 Pressure relief ring
6. Operate the sterilization cycle according to the manual but with no
instruments.
7. Allow a pressure of approximately 300 kPa (29-psig) to build up in
the generator and the chamber.
8. Remove water reservoir cover.
9. Pull the ring (4) of each safety valve in turn using a tool, i.e.
screwdriver, hook etc. (3). Pull the safety valve ring for 2 seconds.
Be careful not to burn your hands.
10. Press the STOP key to stop operation, and exhaust steam from
chamber.
11. Wait until pressure decreases to zero, only then can the door be
opened.
Page 47 of 129 Pages
Page 50
9.5 Replacing chamber heater
Be careful not to bend the tubing or over-tighten with tie wraps.
Kinking the tubing will weaken that area of the tube and it will leak.
Replace any tubes that become damaged.
1. Perform this procedure only when the unit is cool and there is no
pressure in the chamber.
2. Unplug the unit
3. Drain the steam generator using a 6 mm Allen wrench to remove
the bottom plug.
4. Drain both reservoirs
5. Unscrew the screw holding the cooler to the right rear vertical
frame support.
6. Remove the left and right top frame supports.
7. Disconnect the chamber pressure transducer tubing from top rear
chamber manifold. Move the transducer and tubing out of your
work area.
8. Disconnect the air valve 52 tubing from the top rear chamber
manifold.
9. Disconnect tubing from the air check valve on the top rear
chamber manifold.
10. Remove the generator pressure transducer from generator
manifold. Move the transducer and tubing out of your work area.
11. Remove the safety valve tube from generator manifold and also
from safety valve at the reservoir.
12. Remove steam generator tube from steam valve 93.
13. Remove generator water inlet tube from water valve 21 at the
check valve.
14. Remove wires from generator pressostat.
15. Remove wire from generator water electrode.
16. Disconnect generator heater wires from both sides of terminal
block on back of valve assembly bracket.
17. Remove three screws holding the Toroidal transformer rear
mounting bracket and lay it back on the counter.
18. Remove the generator cut-off thermostat probe
19. Remove four screws from bottom of chassis holding steam
generator and remove steam generator.
Page 48 of 129 Pages
Page 51
20. Remove the water tube from the water pump outlet.
21. Remove air filter tube from valve 43.
22. Remove the fast exhaust tube from valve 73.
23. Remove the steam tube from valve 93 at the connector on the back
center of the chamber.
24. Remove the vacuum tube from the top connection of the cooler.
25. Remove two wires from the door switch.
26. Remove three screws holding the valve assembly bracket and
position bracket and valves over reservoirs.
27. Remove the fill tube from the fill cup and move out of the way.
28. Disconnect the two chamber heating element wires.
29. Cut tie wrap holding insulation blanket and discard.
30. Carefully remove insulation blanket. Pulling too hard will tear the
blanket. Be aware the two screws holding the waste reservoir
bracket
chamber
tothechassis my snag the blanket. The rear of the
canmovesidewaysslightly to add clearance for
removing the blanket.
31. Mark edge of heating element down length of chamber for
positioning the replacement.
32. Loosen the four bolts holding the chamber heating element.
33. Remove the chamber temperature sensor and cut-off thermostat
probe. Note their location for reinstallation.
34. Remove completely the four bolts holding the chamber heating
element.
35. Carefully remove the heating element. The rear of the chamber
can move sideways slightly to add clearance for removing the
heater.
36. Install the new heating element and follow these steps in reverse
to reassemble the unit.
37. When tightening the bolts on the heating element, tighten as tight
as possible.
38. When reinstalling the insulation make sure it is tucked in, in all
the proper places.
39. When the installation is complete run a cycle to verify proper
operation.
40. The steam generator may fail to heat initially due to air pressure in
the generator. If this occurs purge the air by opening the generator
safety valve.
Page 49 of 129 Pages
Page 52
9.6 Replacing steam generator heater
Caution:
Before starting, be sure that the electric cord is disconnected from the
power source and that there is no pressure in the autoclave chamber
or steam generator.
Allow the autoclave and the steam generator to cool before removing
outer covers.
1. Remove the autoclave cover (see para. 9.3 “Removing the
autoclave’s outer covers”).
2. Drain the water from the generator by opening the drain plug. Wait
until all the water has drained from the generator (see sec. 9.24).
3. Disconnect the heating element wires from the porcelain connector
(1).
4. Remove three screws holding rear mounting bracket and lay it back
on the counter.
5. Remove the cut-out thermostat probe from the generator.
6. Remove the heating element from the generator.
7. Install the new heating element using liquid Teflon sealant.
8. Insert the cut-out thermostat probe into the new heating element.
9. Reconnect the heating element wires.
10. Reattach the rear mounting bracket.
11. Run a test cycle.
1
Page 50 of 129 Pages
Page 53
9.7 Replacing the steam generator cut-off thermostat
The autoclave is equipped with a cut-off thermostat, which protects the
steam generator heating element from overheating.
Caution
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the chamber or steam generator.
Allow the autoclave to cool before removing outer covers.
1. Remove the rear cover (see para. 9.3 “Removing the autoclave’s
outer covers”).
2. Remove three screws holding rear mounting bracket and lay it back
on the counter.
3. Remove the probe of cut-out thermostat (2) from the generator.
4. Remove the cut-out thermostat probe from the generator.
5. Unscrew the nut holding the cut-off thermostat (2) to the rear
mounting bracket.
6. Insert the new cut-off thermostat probe into the generator as far as it
will go. Secure the cut-off thermostat with the nut removed earlier.
7. Reattach the rear mounting bracket.
8. Reassemble the rear cover.
1
2
No. Description
1 Chamber heater cut-off thermostat
2 Steam generator heater cut-off thermostat
Page 51 of 129 Pages
Page 54
9.8 Replacing and cleaning the generator’s water level electrode
The electrode that controls the water level of the generator is located in
the rear of the generator.
The Electrode performs the following:
a. It protects the heating element by switching it off when there is not
enough water in the generator.
b. It maintains the water level in the generator by switching the water
pump on and off as required.
To clean or replace the electrode proceed as follows (refer to the picture
and the drawings on the next page):
Caution!
Before starting, disconnect the instrument from the po wer source
and ensure that there is no pressure in the chamber or the
generator.
Allow the autoclave to cool before removing outer covers.
1. Take off the autoclave cover (see para. 9.3 “Removing the
Autoclave’s Outer Covers”).
2. Remove the electrical connection from the terminal of the electrode.
3. Unscrew the locking nut and remove the electrode from the top of
the housing.
4. Replace the electrode with a new one or reinstall the same electrode
after cleaning.
4.1 To clean the electrode, use a damp cloth or sponge. A mild
soapy solution may be used, rinse thoroughly. DO NOT use
any harsh chemicals
4.2 When installing a new electrode, make sure that the electrode
is positioned exactly as described in drawing "water level
electrode (see drawing ahead).
5. Tighten the locking nut to prevent any steam or water leakage.
6. Reconnect the electrical wire to the electrode terminal.
7. Turn the autoclave on. The generator will automatically fill with
water. Observe that the unit is heating. Wait for the unit to reach 30
psig (44.7 psi 308 kPa), and then check for leaks around the locking
nuts.
8. Reassemble the cover.
Page 52 of 129 Pages
Page 55
WATER LEVEL ELECTRODE
Water level Electrodes
0.16"-0.24"
Water level Electrodes
Page 53 of 129 Pages
Page 56
9.9 Replacing the Drain Valve
There are two drain valves on the front of the autoclave. One drains the
mineral free water from the clean reservoir and the other drains the
water from the waste water reservoir.
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before removing outer covers.
1. Drain the reservoir (See sec "Draining the Reservoirs" in the
Operator’s Manual). If the drain is not operational it will be
necessary to suction the water out of the reservoir from the top.
2. Remove the autoclave cover (see para. 9.3 “Removing the
Autoclave’s Outer Covers”).
3. Remove the right top frame support over the reservoirs..
4. Remove the metal shroud shielding the electronic components.
5. Remove the drain hose from the inside portion of the drain valve (1).
6. Unscrew the inside nut (6) using a 21mm socket wrench.
7. Remove the drain valve (1) from the panel (7).
8. Install the new valve and reattach the drain tube. Use a tie wrap to
secure the drain tube to the valve.
9. Verify there is no leakage.
10. Replace the metal shroud and right frame support.
VLV170-0013 VLV170-0015
1 6
7
Page 54 of 129 Pages
Page 57
9.10 Replacing the Door Cover
Caution!
Before starting, open the door first, then disconnect the instrument
from the power source and ensure that there is no pressure in the
chamber or generator.
Allow the autoclave to cool before removing outer covers.
1. Remove the two screws holding the Door Handle Cover and slide
it out from the door. It might be tight, but it will slide (see sec
9.14).
2. Unscrew the 6 screws (1) attaching the Door Cover to the upper
(2) and lower (3) cover holders.
3. Remove the Door Cover.
4. Install the new Door Cover and fasten with the 6 screws.
5. Reinstall the Door Handle Cover (see sec 9.14).
1
1
No.description
2
3
1 Fastening screws
2 Upper cover-holder
3 Lower cover-holder
Page 55 of 129 Pages
Page 58
9.11 Replacing the Front Panel
Caution!
Before starting, disconnect the instrument from the po wer source
and ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before replacing the Panel.
1. Open the printer's door (4).
2. Unscrew the three bottom fastening screws (5).
3. Unscrew the top fastening screw (1).
4. Pull the panel approx. 30 centimeters.
5. Remove the digital card (located on the rear side of the panel – as
per instructions in sec.9.22).
6. Remove the printer (3) (see sec. 9.15 "Replacing the Printer").
7. Assemble the digital card on the new panel (see sec. 9.22).
8. Assemble the printer on the new panel.
9. Assemble the panel and fasten it to the autoclave with the 4
fastening screws (1 & 5).
1
5
2
3
4
Page 56 of 129 Pages
Page 59
9.12 Replacing the Printer Door
Caution!
Before starting, disconnect the instrument from the po wer source
and ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before replacing the Printer Door.
1. Open the printer's door (3).
2. Unscrew and remove the 3 screws (2) holding the Printer Door to
the Integral Hinge (1).
3. Remove the Printer Door.
4. Assemble the new Printer Door to the Integral Hinge with the 3
screws (2).
3 1 2
Page 57 of 129 Pages
Page 60
9.13 Replacing the Door Handle
Caution!
Before starting, ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before replacing the Door Handle.
Do not disconnect the electric po wer, since disconnecting the po wer
will lock the handle and prevent replacement of the handle.
1. Open the autoclave's door.
2. Remove the two screws holding the Door Handle Cover and slide it
out from the door. It might be tight, but it will slide (see sec 9.14).
3. Remove the Door Handle by pulling it in the direction of the arrow
(1). It might be tight, but it will slide out of its holder.
4. Slide the new handle into position.
5. Reinstall the Door Handle Cover (see sec 9.14).
Page 58 of 129 Pages
Page 61
9.14 Replacing the Plastic Handle Cover
Caution!
Before starting, ensure that there is no pressure in the autoclave.
Allowthe autoclave to cool before replacing the Door Handle Cover.
Do not disconnect the electric po wer, since disconnecting the power
will lock the handle and prevent replacement of the handle cover.
1. Open the autoclave's door.
2. Open the handle and release the two screws (1).
3. Slide the handle cover out from the door. It might be tight, but it
will slide.
4. Transfer the two screws (2) from the rear of the old cover to the
new cover and tighten so as to provide a snug fit when the cover is
slid back into the slots on the metal bracket.
5. inserted screws (1) in the cover until two threads are above the
cover's surface.
6. Now insert the cover into the handle bracket.
7. Now with the cover in position tighten the two front screws (1).
1
1
2
Page 59 of 129 Pages
Page 62
9.15 Replacing the Printer
Caution!
Before starting, disconnect the in strument from the po wer source
and ensure that there is no pressure in the chamber or the
generator.
Allow the autoclave to cool before removing outer covers.
1. Remove the autoclave cover (see sec. 9.3 “Dismantling the Outer
Covers of the Autoclave”).
2. Disconnect the cable (1) from the back of the printer.
3. Unscrew the two screws (3) attaching the fastening bracket (4) to
the printer.
4. Remove the printer.
5. Set the dip switches on the new printer, located on the back side of
the printer, as follows:
1, 2, 3, 4,
up
5, 6, 7
down
8, 9
up
10
down
6. Insert the new printer into its frame.
7. Assemble the mounting bracket (4) to the printer (2) with the two
screws (3) and verify that it is securely in position.
8. Connect the cable (1) to the printer.
9. Verify connection of power by performing a printer test (see sec.
7.8).
10. Run a cycle and verify that the printer operates correctly.
1
2
3
No. Cat. No. Description
1 CTP201-0127 Cable, Printer, DPU-30, 30cm, 34p
2 THE002-0022 Printer, DPU-30, Seiko
Page 60 of 129 Pages
4
Page 63
9.16 Replacing the Door Switch
Caution!
Before starting, disconnect the instrument from the po wer source
and verify that there is no pressure in the chamber or the generator.
Allow the autoclave to cool before removing outer covers.
1. Remove the autoclave top and left covers (see para. 9.3
“Dismantling the Outer Covers of the Autoclave”).
2. Open the door
3. Unscrew the 2 screws (1) holding the door switch adapter (5) on
the front panel of the autoclave.
4. Disconnect the 2 wires (2, 3) from the door switch (4).
5. Remove the door switch from the autoclave.
1 3
4
6
2 5
5
A
4
6. Release nut (6) and unscrew the door switch (4) from the door
switch adapter (5).
7. Assemble the new door switch to the adapter. Verify that the
wheel protrudes (A) approx. 0.37"-0.38" (9.5-10mm) from the
adapter (5) and tighten nut (6).
8. Assemble the door switch with its adapter to the front panel. Do
not tighten the two screws (1).
9. Perform final adjustment of the microswitch using a GO-NO GO
gauge (P/N JIG411-0102) as follows:
Page 61 of 129 Pages
Page 64
9.1 Insert the GO side (8) of the gauge (11) between the wheel
(7) of the microswitch and the door locking hooks (9) while
pressing the gauge to the hooks. Verify that you hear a
"click" indicating the closing of the switch.
9.2 Insert the NO GO side (10) of the gauge (11) between the
wheel (7) of the microswitch and the door locking hooks (9)
while pressing the gauge to the hooks. Verify that you DO NOT hear the "click" indicating the closing of the switch.
7 8 9 11 10
9.3 If these two conditions are fulfilled, dimension "A" is
correct, tighten screws (1).
9.4 If dimension "A" is not correct, remove the microswitch
from the panel, turn the adapter 1/2 a turn clockwise or
counterclockwise as required and reassemble it to the panel.
9.5 Repeat steps 9.1-9.4.
10. Reconnect the 2 wires (2, 3) to the door switch. Verify that one
wire is connected to screw no. 1 ("COM" - marked on the switch)
and the other wire is connected to screw no. 3 ("NO" marked on
the switch).
Makesurethatthedoorswitchisinstalledcorrectly (the w heel of
the microswitch is parallel to the floor)!
Check that the operational message “Door is Open" is displayed
when the door is opened and that the messag e disappears when the
door is closed.
Page 62 of 129 Pages
Page 65
9.17 Replacing the fuse on the Electronic Board
Caution!
Before starting, disconnect the instrument from the power source.
There are 3 fuses on the AC high voltage power board (Nova4-A2)
board. All fuses are 5x20mm.
1. fuse 3.15A FAST (3) to the chamber heater
2. fuse 2A FAST (4) to the vacuum pump
3. fuse 2A FAST (5) to the water pump
To replace the fuses proceed as following:
6. Separate the front panel from the autoclave (see sec. 9.11 "
Replacing the Panel").
7. Pull it outwards as far as possible.
8. With a small screwdriver disconnect the fuse holder (2) from the
board (1).
9. Remove the faulty fuse (6) from the holder and replace it with a
new one.
10. Return the fuse holder with the new fuse to the board.
11. Reassemble the panel.
1
2
453
2
Page 63 of 129 Pages
6
Page 66
9.18 Replacing the fuses of the transformer
Caution!
Before starting, disconnect the instrument from the power source.
Allow the autoclave to cool before removing outer covers.
There are 2 fuses, an input fuse 800ma slow blow located in the brown
wire and an output fuse 5A fast blow located in the black wire. Both
fuses are 5x20mm.
1. Remove the rear cover (see sec. 9.3 “Dismantling the Outer Covers
of the Autoclave”).
2. If necessary, cut one plastic cable tie to release the fuse capsule (4).
3. Twist and gently pull the two halves of the fuse capsule apart (1)
and (3).
4. Take out the burnt fuse and replace it with a new fuse (2).
5. Reassemble the fuse holder.
6. Replace the cut cable holder.
7. Re-assemble the covers.
4 4
3
1
2
Page 64 of 129 Pages
Page 67
9.19 Replacing the water pump
Caution!
Before starting, disconnect the instrument from the power source.
Allow the autoclave to cool before removing outer covers.
1. Empty the water reservoir (See Draining the Water Reservoir in
the Operator’s Manual).
2. Remove the autoclave rear cover (see sec. 9.3 “Dismantling the
Outer Covers of the Autoclave”).
3. Remove the three screws holding the electrical socket rear
mounting bracket and set it aside.
4. Disconnect the wires from the pump.
5. Disconnect the piping from the pump.
6. Remove the pump (7) from the rubber shock absorbers (1).
If the rubber shock absorbers are damaged, replace them.
7. Replace the damaged pump with a new pump.
8. Reconnect wiring and piping.
9. Reassemble the electrical socket bracket.
10. Reassemble the rear cover.
11. Refill the mineral-free water reservoir.
12. Turn on the autoclave and verify it operates correctly.
No. Description Cat. No.
1 Rubber shock absorber SKR203-0006
2 Fitting 5/16”x1/8” (straight) FIT100-0803
3 Screw
BOL191-0142
4 Nut NUT192-0155
5 washer NUT193-0283
6 Spring washer NUT193-0317
Fastening the
shockabsorber
7 ULKA water pump PUM055-0019
Page 65 of 129 Pages
Page 68
9.20 Replacing the vacuum pump
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the chamber or steam generator.
Allow the autoclave to cool before removing outer covers.
1. Take off the autoclave cover (see para. 9.3 “Dismantling the Outer
Covers of the Autoclave”).
2. Disconnect the electrical wires (Orange, Black & Green wires)
(3).
3. Remove fittings for inlet tube (1) and outlet tube (4).
4. Unscrew the 4 pump-leg screws (2) from the bottom of the
chassis.
5. Remove the vacuum pump.
6. Install the new pump and assemble the leg screws (2).
7. Install the inlet and outlet tube fittings (1) and (4) on to the new
pump.
8. Connect the electrical wires. Verify that the wires are connected as
they were originally.
9. Reassemble the covers.
10. Connect the autoclave to the power supply and turn on the
autoclave. Run program 2 (Wdry 134) to verify that the vacuum
pump operates O.K.
3
4
2
1
Page 66 of 129 Pages
Page 69
9.21 Replacing the Plunger or Coil of the 1/4" Solenoid Valve
Caution!
Before starting, disconnect the instrument from the po wer source
and ensure that there is no pressure in the chamber or steam
generator.
Allow the autoclave to cool before removing outer covers.
1. Take off the autoclave top cover (see para. 9.3 “Removing the
Autoclave’s Outer Covers”).
2. Remove nut (3) and washer (2).
3. Remove the coil (1).
4. Unscrew the plunger and replace it with a new kit. The plunger kit
consists of the plunger housing (6) and the plunger & spring (5).
5. Using a magnifying glass, inspect the seat in the valve base for
chips or cracks. If any are found the base will need to be replaced.”
6. Reassemble the coil (1) the nut (3) and the washer (2). If a new coil
is needed then unplug the electrical connector (4) from the old coil
and plug it into the new coil.
7. Reassemble the coil (1) and the nut (3).
8. Re-install the autoclave’s cover.
2 1 3 4
5 6
Page 67 of 129 Pages
Page 70
9.22 Replacing the Electronic Boards
Caution!
Before starting, disconnect the instrument from the po wer source
and ensure that there is no pressure in the chamber.
Allow the autoclave to cool before removing outer covers.
See drawings "Nova4-D2" and "Nova4-A2 on the following pages to
assist in locating connectors marked by JP.
9.22.1 Removing the Front Panel Assembly
1. Remove the control panel assembly (see sec 9.11
“Replacing the Front Panel;”).
2. Disconnect the flat cable (7) from the flat cable socket (1)
at the top of the big board (Nova4-D2).
3. Disconnect all the connectors along the outside edge of
the big board (Nova4-D2), JP11, JP2, JP3, JP1, JP12,
JP13, JP15, making note of their location.
4. Disconnect the GND wire at the top of the Nova4-D2
board
5. Disconnect the connectors at JP11, JP9, JP12, JP15, JP7,
JP8 and JP6 on the small board (Nova4-A2).
6. Disconnect the small green wire at the bottom right
7. Lay the front panel on a table with the keyboard down.
9.22.2 Replacing the "Nova4-A2" (small) board
1. Disconnect wires from JP2, JP3, JP4, JP5 (6), making
sure to note the location of each wire on the circuit board.
2. Disconnect connector JP10 (4).
3. Unscrew the 4 screws (3) attaching the small board (5) to
the big board (2).
4. Remove the small board.
5. Check the isolation sheet that is placed under the board.
Replace if damaged.
6. Place the new board and fasten it with the 4 screws (3).
7. Connect the connector JP10 and the wires JP2-JP5.
8. Return the control unit to the autoclave and reconnect the
wires and connectors following, in reverse order, the
instructions in sec 9.22.1.
9. Re-assemble the control panel to the autoclave
Page 68 of 129 Pages
Page 71
8
3
Assembled Nova4-D2 and Nova4-A2 Boards
1
2
3
4
3
5
6
3
7
Page 69 of 129 Pages
Page 72
Nova4-A2 Board
JP10
JP2
JP11 JP9 JP7 JP8
JP12
JP15
JP3
JP4
JP5
JP6
Page 70 of 129 Pages
Page 73
Nova4-D2 Board
JP8
JP6
JP11
JP2
JP4
JP9
JP3
JP1
JP12
JP13
JP15
JP5
Page 71 of 129 Pages
Page 74
1. Disconnect wires from JP2, JP3, JP4, JP5 (6 on
previous page) on the Nova4-A2 board. Making sure to
note the location of each wire on the circuit board.
2. Disconnect connector JP14 (8).
3. Unscrew the 4 screws (3) attaching the small board (5)
to the big board (2).
4. Remove the small board.
5. Check the isolation sheet that is placed under the board.
Replace if damaged.
6. Unscrew the 4 long spacers (9) connecting the small
board to the big board.
7. Unscrew the 2 screws (12) connecting the big board to
the plastic panel (11). Be careful not to lose the
grounding terminal (10) fastened by one of the screws.
8. Lift the board slightly and disconnect the keypad flat
cable (14) from its connector (13) on the rear side of the
board.
9. Replace the big board with a new board.
10. Reassemble by following steps 1 thru 8 in reverse order.
9.22.3 Replacing the "Nova4-D2" (big) board
9
1011
12
13
Page 72 of 129 Pages
14
Page 75
9.23 Replacing the switch
Caution!
Before starting, disconnect the instrument from the po wer source
and ensure that there is no pressure in the chamber.
Allow the autoclave to cool before removing outer covers.
1. Remove the control electronic boards (see para. 9.22 "Replacing the
Electronic Boards").
2. Disconnect the wires (1) from the switch (3). Log which wire is
connected to each leg (2) of the switch.
3. Press the switch's fasteners (5) and push the switch inwards to
release it from its seat on the panel (4).
4. Remove the switch and replace it with a new one. Pay attention to
the switch's position. The "0" on the switch must be opposite the
"0" in the plastic panel and the "|" on the switch must be opposite
the "|" on the panel.
5. Connect the wires to the switch verifying that the wires are
reconnected in the same positions as they were removed in
instruction #2.
6. Place the Nova4-D2 board on the panel and reconnect the keypads
flat cable to the board.
7. Reassemble the electronic boards according to para. 9.22.
21
3
4
5
Page 73 of 129 Pages
Page 76
9.24 Draining the generator
Caution!
Before starting, disconnect the instrument from the po wer source
and ensure that there is no pressure in the chamber or steam
generator.
Allow the autoclave to cool before draining the generator.
1. The generator’s drain plug is located under the autoclave on the left
side. In order to drain the generator it will be necessary to position
the autoclave so it is partially hanging over the edge of the counter.
CAUTION when attempting to drain the generator make sure the
autoclave is well supported and will not fall.
2. Place a vessel under the autoclave to collect the drain water.
3. Unscrew the drain plug, using a 6mm allen wrench.
1
4. If the water does not drain out easily then pull on the ring of the
generator safety valve to break the vacuum in the generator.”
5. After draining the water, reassemble the drain plug.
6. Move the autoclave back to its leveled position, reconnect it to the
power source and turn the autoclave on.
7. Check and verify that the generator is filled with water and that the
plug does not leak.
Page 74 of 129 Pages
Page 77
9.25 Emergency Door Opening
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the autoclave.
Caution!
Keep hands and face away from the safety valve to avoid being burned
by hot steam.”
9.25.1 Decreasing the pressure
Since the autoclave is not equipped with an analogue pressure
gauge, and the display is off (no electrical power), it is
recommendedtowaitapproximatelyonehour to verify that the
pressureandtemperaturehavedecreasedtoambientconditions.
If it is required to open the door urgently, decrease the pressure
by opening both, the chamber's and generator's safety valves as
follows. The safety valves are located in the mineral-free water
reservoir.
1
A. ASME-approved type safety valve
1. Remove water reservoir cover (1).
2. Pull the ring (4) of the safety valve using a tool, i.e.
screwdriver, hook etc. (3). Pull the safety valve ring until
steam ceases to escape from the safety valve. Be careful
not to burn your hands.
243
No. Description
1 water reservoir cover
2 Safety valve
3 Pulling device
4 Pressure relief ring
Page 75 of 129 Pages
Page 78
B. PED-approved type safety valve
1. Remove the water reservoir cover.
2. Turn the pressure relief nut clockwise until steam ceases
to escape from the safety valve.
Attention:
Use protective gloves in order not to burn your hands with
the hot steam.
1. Insert a 2 mm (0.08") metal pin (1) into the hole (2)
located above the door opening handle (3).
2. Press the pin downwards and, simultaneously open the
door by pulling the door handle (3) in the direction of the
arrow (4).
1
2
3
Page 76 of 129 Pages
4
Page 79
9.26 Replacing the water reservoirs
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the autoclave.
Note:
The following instructions refer to both, the mineral-free water
reservoir and to the waste-water reservoir.
1. Drain the water reservoir (See sec "Draining the Reservoirs" in the
Operator’s Manual).
2. Remove the right, rear and top covers of the autoclave (see para.
9.3 "Dismantling the Outer Covers of the Autoclave").
3. Unscrew the screws (4) holding the cooler to the right rear vertical
frame support and bottom frame.
4. Unscrew screws (1 & 2) and remove the right top frame support
(10).
5. Disconnect the relevant cable terminals (3 or 9) from the water
level electrodes of the reservoir to be removed.
6. Disconnect the tubing and fittings (5, 6, 7, 8 etc.) from the bottom
and top of the reservoir to be replaced.
7. Pull the reservoir straight up and out of the machine.
8. Insert the new reservoir and follow the steps in this section in the
reverse order to reassemble the unit.
Page 77 of 129 Pages
Page 80
5
6
10
9
8
1
2
3
7
4
Page 78 of 129 Pages
Page 81
9.27 Cleaning the water reservoirs
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the autoclave.
Note:
The following instructions refer to both, the mineral-free water
reservoir and to the waste-water reservoir.
1. Drain the water reservoir(See sec "Draining the Reservoirs" in the
Operator’s Manual).
2. Remove the top, right and rear covers of the autoclave (see para.
9.3 "Dismantling the Outer Covers of the Autoclave").
3. Unscrew the screw (3) holding the cooler to the right rear vertical
frame support.
4. Unscrew screws (1 & 2) and remove the right top frame support
(6).
5. Remove any clips holding the reservoir cover.
6. Remove the top cover of the reservoir.
7. Clean the inside of the reservoir by wiping with a clean cloth.
8. Clean the water level electrodes (connected to 4, 5) inside the
reservoir with a soft cloth.
9. Verify the electrode terminals (4, 5) are connected firmly to the
electrodes.
10. Follow the steps in this section in the reverse order to reassemble
the unit. Be sure to reattach clips to the waste-water reservoir
cover.
1
2
6
5
4 3
Page 79 of 129 Pages
Page 82
9.28 Replacing the cut-of thermostat for the Chamber Heater
These instructions refer to pictures on the following page.
1. Perform this procedure only when the unit is cool and there is no
pressure in the chamber.
2. Unplug the unit
3. Drain both reservoirs
4. Unscrew the screw holding the cooler to the right rear vertical
frame support.
5. Remove the left (15) and right (16) top frame supports.
6. Disconnect the chamber pressure transducer tubing (10) from top
rear chamber manifold (8). Move the transducer and tubing out of
your work area.
7. Disconnect the air valve 52 tubing (11) from the top rear chamber
manifold (8).
8. Disconnect tubing (9) from the air check valve on the top rear
chamber manifold
9. Remove three screws (7, 13, 14) holding the valve assembly
bracket (12) and move the assembly down to give clearance for
removing the insulation blanket.
10. Cut tie wrap holding insulation blanket and discard.
11. Carefully move the insulation blanket (5) to expose the cut-off
thermostat sensor (6).
12. Loosen the two bolts (18) holding the section of the chamber
heating element (17) that secures the temperature sensor and cutoff thermostat.
13. Remove the chamber cut-off thermostat probe.
14. Trace the cut-off thermostat tubing back to the cut-off thermostat
control box (2).
15. Remove the two wires on the control box
16. Unscrew the nut (1) holding the control box to the mounting plate
(3).
17. Install the new cut-off thermostat and follow these steps in reverse
to reassemble the unit. Position the probe as shown in the picture
(6).
18. When tightening the bolts on the heating element, tighten as tight
as possible.
19. When reinstalling the insulation make sure it is tucked in, in all the
proper places.
20. When the installation is complete run a cycle to verify proper
operation
21. The steam generator may fail to heat initially due to air pressure in
the generator. If this occurs purge the air by opening the generator
safety valve.
Page 80 of 129 Pages
Page 83
1
2
3
18 5
16
15
14
13
12
17
6
7
Page 81 of 129 Pages
8
9
10
11
Page 84
9.29 Replacing the PT100 temperature sensor for the Chamber Heater
These instructions refer to pictures on the following page.
1. Perform this procedure only when the unit is cool and there is no
pressure in the chamber.
2. Unplug the unit
3. Drain both reservoirs
4. Unscrew the screw holding the cooler to the right rear vertical
frame support.
5. Disconnect the chamber pressure transducer tubing (6) from top
rear chamber manifold (4). Move the transducer and tubing out of
your work area.
6. Disconnect the air valve 52 tubing (7) from the top rear chamber
manifold (4).
7. Disconnect tubing (5) from the air check valve on the top rear
chamber manifold
8. Remove three screws (3, 10, 11) holding the valve assembly
bracket (8) and move the assembly down to give clearance for
removing the insulation blanket.
9. Cut tie wrap holding insulation blanket and discard.
10. Carefully move the insulation blanket (9) and (1) to expose the
PT100 (14).
11. Loosen the two bolts (16) holding the section of the chamber
heating element (15) that secures the temperature sensor and cutoff thermostat.
12. Remove the PT100 temperature sensor.
13. Trace the sensor cable back to the electronic box and main control
board.
14. Remove the connector from the main control board (2) (JP3).
15. Install the new PT100 and follow these steps in reverse to
reassemble the unit. Position the probe as shown in the picture (14).
16. When tightening the bolts on the heating element, tighten as tight
as possible.
17. When reinstalling the insulation make sure it is tucked in, in all the
proper places.
18. When the installation is complete run a cycle to verify proper
operation
19. The steam generator may fail to heat initially due to air pressure in
the generator. If this occurs purge the air by opening the generator
safety valve.
Page 82 of 129 Pages
Page 85
16
15
14
1
2
13
12
11
10
9
8
3
4
5
6
7
Page 83 of 129 Pages
Page 86
9.30 Replacing the internal PT100
Caution!
Before starting, disconnect the in strument from the po wer source
and ensure that there is no pressure in the chamber.
Allow the autoclave to cool before removing outer covers.
1. Remove the autoclave's rear and right covers (see para. 9.3
“Removing the Autoclave’s Outer Covers”).
3. Cut the tie wrap (3) around the PT100 isolation (2). Be careful not
to damage the isolation.
4. Reveal the PT100 (5) and unscrew its nut (4) with a 1/2" wrench
(6).
5. Cut all the tie wraps holding the cable of the PT100.
6. Disconnect the plug of the PT100 (7) from the Nova4-D2 board
connector JP1.
7. Replace the PT100 with a new one. use only an original PT100
supplied by Tuttnauer.
8. Place the isolation sleeve on the new PT100.
9. Insert the PT100 into its seat and tighten its nut (4) loosely. Verify
that dimension "A" is 0.3"-0.5" (8-13 mm) and that the electrode's
tip does not touch the bottom tray.Continue tightening nut (4) to
lock the PT100 in position.
10. Cover the PT100 with the isolation sleeve and tighten it with tie
wraps.
11. Attach the cable of the internal PT100 to the cable of Chamber
Heater PT100 with tie wraps.
12. Connect the plug of the internal PT100 to the Nova4-D2 board
connector JP1.
13. Reassemble the electrical bracket to the autoclave's frame.
14. Reassemble the covers.
Page 84 of 129 Pages
Page 87
1 2 3
4 5 6 7
Page 85 of 129 Pages
Page 88
9.31 Replacing the pressure transducer
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the chamber and in the generator.
Note:
The following instructions refer to both, the chamber and the generator
pressure transducer.
1. Remove the top cover (see para. 9.3 “Removing the Autoclave’s
Outer Covers”).
2 1
2. Lift slightly the pressure transducer (1) as in picture 1.
3. Unscrew screw (2) as in picture 1.
4. Remove the connector (3) as in picture 2.
5. Unscrew the pressure transducer. Use two open wrenches (5) as
shown in picture 3.
6. Remove the pressure transducer and replace it with a new one.
8. Assemble the connector (3) and fasten it with the screw (2).push
down the pressure transducer to its place, below the front panel top
(4).
43
55
Page 86 of 129 Pages
Page 89
9.32 Cleaning the steam generator
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the chamber and in the generator.
1. Drain the generator (see sec. 9.24).
2. Remove the generator's heating element (see sec. 9.6).
3. Place a vessel to collect water under the drain opening.
4. Insert a tap-water hose in the heating element opening.
5. Clean the inner part of the generator with water under low
pressure.
6. Verify that no water is left in the generator
7. Reinstall the drain plug.
8. Reinstall the heating element (see sec. 9.6).
9. If the generator has an excessive build up of mineral deposits then
proceed as follows”
10. Drain the generator see sec 9.24”
11. Disconnect all the piping connected to the generator manifold”
12. Disconnect piping at the front of the generator”
13. Remove the water sensing electrode”
14. Remove the generator heating element see sec 9.6”
15. Remove the four screws holding the generator to the chassis and
remove the generator
16. Remover the foam insulation material”
17. Soak the generator in a mineral dissolving solution.”
18. Flush thoroughly with clean water and reinstall following these
instructions in reverse order”
Page 87 of 129 Pages
Page 90
9.33 Replacing the cooler and the cooler's fan
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the chamber and in the generator.
1. Drain the waste reservoir.
2. Remove the autoclave's covers (see para. 9.3 “Removing the
Autoclave’s Outer Covers”).
3. Disassemble and remove the right top frame (1).
4. Disconnect the wires (2) from the water level electrodes of the
waste reservoir.
5. Remove the two rear screws from the left top frame.
6. Disconnect the waste water inlet hoses on the left side of the
reservoir and the drain hose (4) below the reservoir.
7. Pull the rear top frame back and remove the waste water reservoir
from the autoclave by pulling it upward. The reservoir bracket
supports are slotted so the reservoir will slide straight up.
8. Disassemble the inlet and outlet hoses from the cooler (7).
9. Unscrew 3 screws (8, 9, 10).
10. Remove the cooler (5) with the fan (6) from the autoclave.
11. Disconnect the fan's electrical connector (3).
12. Detach the fan from the cooler.
13. Replace the fan or the cooler as required.
14. Reconnect the fan to the cooler.
15. Connect the fan's electrical connector.
16. Insert the assembled cooler into the autoclave.
17. Assemble the cooler to the autoclave with the 3 screws (8, 9, 10).
18. Reattach the hoses to the inlet and outlet of the cooler (7). Use tie
wraps to secure the hoses
.
19. Insert the waste water reservoir into its place.
20. Connect the hoses to the reservoir. Use tie wraps to secure the
hoses.
21. Connect the electrical wires to the water level electrodes.
22. Reassemble the covers.
Page 88 of 129 Pages
Page 91
1
2
3
4
5
6
10
9
7
8
3
Page 89 of 129 Pages
Page 92
9.34 Testing procedure for the chamber PT100
Caution!
Before starting ensure that the autoclave is off and that there is no
pressure in chamber.
For items numbering – see pictures on the next page.
1. Prepare a flexible temperature probe with a reading device, a
1/4"BSP sealing nut and a PT100 simulator.
2. Verify using a DVM that the PT100 simulator reads 100Ω at 0˚C
and 151.4 Ω at 134˚C
3. Disconnect the connector of the internal PT100 from the
electronic board (2) JP1.
4. Connect a PT100 simulator to the socket of the PT100 JP1.
5. Turn on the autoclave and select the Vacuum Test Cycle to keep
the unit from heating up.
6. Change the temperature display to ˚C by changing parameter
TempInF (see sec 7.2.35) to 0.
7. Set the simulator to 70˚C and verify that the reading on the display
is 70˚C Using the PT100 simulator recalibrate if necessary (See
sec 7.6).
8. Setthesimulatorto134˚Candverifythatthereadingonthe
display is 134˚C Using the PT100 simulator recalibrate if
necessary (see sec. 7.6).
9. If the readings on the display are correct the electronic system is
functional and you can proceed. If the calibration does not align
the simulator with the display readout, then replace the Nova4-D2
board.”
10. Disconnect the simulator and reconnect the PT100 (2).
11. Using the sealing nut insert the temperature probe into the
chamber through the Validation port (1) and place it close to the
PT100.
12. With the door closed and the Vacuum Test cycle still selected
allow the chamber temperature to stabilize.
13. Verify that the Elara display and the independent temperature
reading device show the same temperature. Recalibrate if
necessary (see sec 7.6).
14. Select and run a 134°C cycle
15. When the pressure has stabilized, verify that the deviation
between the reading of the PT100 and the probe does not exceed
±1/2°C.
16. If the deviation is greater than ±1/2˚C then calibrate the Chamber
Temperature Sensor. (see sec 7.6).
17. If the calibration is not successful then change the Chamber
PT100.
Page 90 of 129 Pages
Page 93
2 1
Page 91 of 129 Pages
Page 94
9.35 Testing procedure for the heating element PT100
Caution!
Before starting ensure that the autoclave is off and that there is no
pressure in chamber.
For items numbering – see pictures on the next page.
1. Prepare a flexible temperature probe with a reading device and a
PT100 simulator.
2. Verify using a DVM that the PT100 simulator reads 100Ω at 0˚C and
151.4 Ω at 134˚C.
3. Disconnect the connector of the heating element PT100 (1) from the
electronic board JP3.
4. Connect a PT100 simulator to the socket of the PT100 JP3.
5. Turn on the autoclave and select the Vacuum Test Cycle to keep the
unit from heating up.
6. Change the temperature display to ˚C by changing parameter
TempInF (see sec 7.2.35) to 0.
7. Set the simulator to 70˚C and verify that the reading on the display is
70˚C. Using the PT100 simulator recalibrate the Coil Temp if
necessary (See sec 7.6).
8. Setthesimulatorto134˚Candverifythatthereadingonthedisplay
is 134˚C. Using the PT100 simulator recalibrate the Coil Temp if
necessary (See sec 7.6).
9. If the readings on the display are correct the electronic system is
functional and you can proceed. If the calibration does not align the
simulatorwiththedisplayreadout,thenreplacetheNova4-D2 board.
10. Turn off the autoclave.
11. Disconnect the simulator and reconnect the PT100 (1).
12. Release the bolts of the heating element. Follow instructions #4 thru
#12 sec 9.29.
13. If the probe has a small diameter insert it into the groove of the
PT100 (3) (together with the PT100). If the diameter of the probe is
too big for this, remove the cut-off sensor and place the probe in the
cut-off groove(2).
14. Turn on the autoclave.
15. With the door closed and the Vacuum Test cycle still selected allow
the chamber temperature to stabilize.
16. Enter the calibration display for the heating element PT100 (Coil
Temp).
17. Verify that the Coil Temp display and the independent temperature
reading device show the same temperature. Recalibrate if necessary
(see sec 7.6).
18. Select any cycle except vacuum test cycle.
19. Enter the calibration display for the heating element PT100 (Coil
Temp).
20. Wait approx 1/2 hour or until the temperature readings have
stabilized and verify that the deviation between the reading of the
PT100 and the probe does not exceed ±1/2°C.
Page 92 of 129 Pages
Page 95
21. If the deviation is greater than ±1/2˚C then calibrate the Heating
element PT100 (see sec 7.6).
22. If the calibration is not successful then change the Heating element
PT100.
4 3 2 1
Page 93 of 129 Pages
Page 96
9.36 Testing procedure for the chamber pressure transducer
Caution!
Before starting ensure that there is no pressure in the autoclave
chamber.
1. Prepare a current loop calibrator-simulator 4-20mA .
2. Turn the autoclave on and select the Vacuum Test Cycle to keep
the unit from heating up.
3. Change the pressure display to kPa by changing parameter
PressInPSI (see sec. 7.2.36) to 0.
4. Unscrew screw (1) and disconnect the chamber transducer's plug
(2).
5. Disassemble the plug by pulling it (5) out from its housing (3) and
attach the "+" lead of the loop calibrator to the black wire (4).
1
3
4
2
5
6. Attach the "–" lead of the loop calibrator to the GND (ground)
point on the frame of the autoclave.
7. Apply a 5mA signal. The pressure displayed on the autoclave's
display must be 25±3 kPa. With the Current Loop Calibrator
attached calibrate the low pressure as necessary (See sec 7.6).
8. Apply a 16mA signal. The pressure displayed on the autoclave's
display must be 300±3 kPa. With the Current Loop Calibrator
attached calibrate the high pressure as necessary (See sec 7.6).
9. If the displayed pressure is within the required limits the electronic
circuit is functioning correctly.
10. If the calibration does not align the Current Loop Calibrator with
the display readout, then replace the Nova4-D2 board.
11. Reconnect the pressure transducer.
12. Connect a test pressure gauge to the validation port.
13. Select and run the Vacuum Test.
14. When the vacuum pump stops running verify that the chamber
pressure on the autoclave display matches the test gauge. The
allowable deviation is +/-3kPa. Recalibrate if necessary. See sec
7.6.
15. Select and run a 273˚F (134˚C) cycle.
16. When the pressure has stabilized verify that the pressure deviation
is no greater than +/-3kPa
17. Calibrate low and high pressure as needed (see sec 7.6).
18. If calibration is not successful then change pressure transducer
Page 94 of 129 Pages
Page 97
9.37 Testing procedure for the Generator pressure transducer”
Caution!
Before starting ensure that there is no pressure in the generator.
For items numbering – see also pictures on the next page.
1. Prepare a current loop calibrator-simulator 4-20mA .
2. Turn the autoclave on and select the Vacuum Test Cycle to keep
the unit from heating up.”
3. Change the pressure display to kPa by changing parameter
PressInPSI (see sec. 7.2.36) to 0.
4. Unscrew screw (1) and disconnect the steam generator transducer's
plug (2).
5. Disassemble the plug by pulling it (5) out from its housing (3) and
attach the "+" lead of the loop calibrator to the black wire (4).
1
3
4
2
5
6. Attach the "–" lead of the loop calibrator to the GND (ground)
point on the frame of the autoclave.
7. Apply a 5mA signal. The pressure displayed on the autoclave's
display must be 25±3 kPa. With the Current Loop Calibrator
attached calibrate the low pressure as necessary (see sec 7.6).
8. Apply a 16mA signal. The pressure displayed on the autoclave's
display must be 300±3 kPa. With the Current Loop Calibrator
attached calibrate the high pressure as necessary (see sec 7.6).
9. If the displayed pressure is within the required limits the electronic
circuit is functioning correctly.
10. If the calibration does not align the Current Loop Calibrator with
the display readout, then replace the Nova4-D2 board.
11. Reconnect the pressure transducer.
12. Ensure that the generator is cold and did not heat or build pressure.
13. Remove the plug (8) from the generator manifold (7) port and
connect a test pressure gauge (6) to the manifold
14. Withthevacuumteststillselected,verifythatthegenerator
pressure on the autoclave display matches the test gauge. The
allowable deviation is +/-3kPa. Recalibrate if necessary. See sec
7.6.
15. Select a 273˚F (134˚C) cycle. This will cause the generator to heat.
16. When the pressure has stabilized verify that the pressure deviation
is no greater than +/-3kPa.
17. Calibrate low and high pressure as needed (see sec 7.6)
18. If calibration is not successful then change pressure transducer
Page 95 of 129 Pages
Page 98
6 78
Page 96 of 129 Pages
Page 99
9.38 Door Locking Solenoid
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the chamber.
1. Insert a pin (1) into the hole (2) in the handle, pressing down the
locking pin.
2. Open the door.
1
3
2
3. Remove the door cover (3) (see sec. 9.10).
4. Release the locking mechanism spring (5) from the front spring
holder (6) or brace the locking mechanism so it will not move
while you are working on it..
5. Remove the solenoid pin stop bracket (4).
4
5 6
6. Disconnect the door solenoid wires (7) at the terminal block (8).
8
7
Page 97 of 129 Pages
Page 100
7. Using the correct size screwdriver (9) insert it from the top and
unscrew the door solenoid mounting cylinder (11) from the door
locking mechanism brass holder (10).
9
10
Upper door bridge
Door handle
8. Disassemble the solenoid (15) from the mounting cylinder (11).
12 11 13 14 15
9. Insert a new spring (14) and reassemble the solenoid and mounting
cylinder screw the assembly into the door locking mechanism.
10. Tighten with the screwdriver.
11. Follow instructions in the reverse order to complete the installation.
LIST OF DOOR LOCKING SOLENOID
Item Cat. No. Description
11 LOK411-0019 Solenoid extension adaptor T4x12
12 LOK411-0018 Solenoid pin -Elara
13 LOK254-0001 Washer, locking solenoid pin
14 SPR177-0004 Spring, door locking solenoid
15 SOL027-0003 Solenoid, door lock
Page 98 of 129 Pages
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.