1 GENERAL
1.1 Incoming Inspection
Upon receiving your Tuttnauer Autoclave carefully inspect the outside of
the shipping carton for signs of damage. If any damage to the carton is
found note the location with respect to the autoclave and check that area of
the autoclave carefully once it is fully unpacked. Observe packing method
and retain packing materials until the unit has been inspected. Mechanical
inspection involves checking for signs of physical damage such as:
scratched panel surfaces, broken knobs, etc.
If any damage is found contact your dealer as soon as possible so that
they can file a claim with the shipping carrier and also notify
Tuttnauer.
All Tuttnauer products are carefully inspected prior to shipment and all
reasonable precautions are taken in preparing them for shipment to assure
safe arrival at their destination.
Note: Lifting and carrying should always be done by two people.
1.2 Warranty
To activate your warranty or for warranty information on this unit please
contact your dealer or Heidolph Brinkmann at one of the #'s listed below:
Heidolph Brinkmann, LLC, 1241 Jarvis Avenue, Elk Grove Village, IL
If you have any questions or there are any difficulties with this instrument
Do not attempt to service this instrument yourself.
1.3 Ordering Information
Several items must be specified when ordering the unit from your dealer.
e-mail:info@tuttnauerUSA.com.
and the solution is not covered in this manual, please contact your dealer
or Heidolph Brinkmann.
• The chamber diameter and chamber depth required
• Please specify the supply voltage available (i.e. 115v/208v; 1Ph/3Ph)
• The temperature scale needed (Celsius or Fahrenheit).
• The pressure scale needed (kPa or psi)
Page 4 of 51
1.3.1 Options
• Printer - An optional, built in, thermal printer is available
• Rapid Cooling – these units are available with or without
Rapid Cooling for liquid cycles.
οThe cooling option requires an additional water and
compressed air connection.
οWhile cooling water is circulated around the outside of
the chamber compressed air is used inside the chamber to
maintain the liquid at a high boiling point.
1.3.2 Accessories
• Baskets – Different size baskets are available for these units
(see pages 49 & 50).
The baskets are made of stainless steel wire and have a
handle. The basket allows the operator to load a large
quantity of materials into the chamber.
• Stainless steel containers - Different size containers are
available for these units (see pages 49 & 50).
The containers are designed for sterilizing waste material.
The containers have vent holes along the upper rim.
Page 5 of 51
2 TECHNICAL DATA
2.1 Introduction
Models 2540, 3850, 3870, ELV/ELVC are vertical sterilizers designed
especially for sterilization of instruments, liquids and other materials in
laboratories, research institutes, food laboratories and pharmaceutical
A special feature of the ELVC sterilizer is the ability is to rapidly cool
The sterilizer provides four fully automatic programs eliminating any need
A safety valve, mounted at the back top of the autoclave, is set to 40psig
The autoclave is equipped with a pressure gauge (for reference only) that
According to the standard for laboratory autoclaves these units are
Thismanualisintendedtogive the user a general understanding of how
After reading this manual, operating the autoclave should be straight
Only technical personnel having proper qualifications, holding technical
facilities.
down a liquid load at the end of the sterilizing portion of the cycle.
for operator intervention during a cycle. A computerized control unit
enables precise control and monitoring of physical parameters and clear
documentation of the sterilization cycles.
and protects the system from excessive pressure.
can be used in case there is any problem with the electricity supply during
the sterilization cycle. The pressure gauge can verify if there is pressure in
the chamber before an attempt is made to open the door.
equipped with two temperature sensors that control the cycle and prevent
the opening of the door if the temperature is too high.
the autoclave works and indicates best ways to operate and maintain it in
order to obtain optimum results and a trouble free operation.
forward. However, since the autoclave is built using high technology
sensitive components, no attempt should be made by the user or any other
unauthorized person to repair or re-calibrate it.
documentation and having adequate test instrumentation, are authorized to
install or undertake repair or service.
Page 6 of 51
2.2 Operating Conditions
• The autoclave is for “Indoor use” only.
• Only autoclavable materials shall be used.
• The ambient temperature shall be in the range of 50°F - 104°F (10°C -
40°C) and a relative humidity up to 85%.
•The operation altitude of this autoclave is restricted to 6600ft
(2000m).The ambient pressure shall not be lower than 11.6psi (80kPa)
•The autoclave shall not be used in a manner not specified in this
manual!
Caution
Waste water should be brought into the public net in accordance
with the local rules or requirements i.e. only non-hazardous liquids
shall be disposed in public sewage!
2.3 Directives and Standards
Every autoclave meets the following Standards:
2.3.1 Technical Standards
1. The pressure vessel is an integral part of the equipment
and complies with the ASME code, Section VIII div.1 for
unfired pressure vessels
2. UL-61010–2-041 & UL-61010–1: Safety requirements for
medical device.
3.
– 1998 Software in Programmable Components
UL
2.3.2 Quality standards
The manufacturing plant meets the following quality standards:
1. EN ISO 9001:2000– Quality System
2. ISO 13485-2003 – Quality systems – Medical devices.
2.4 Environment Emission Information
A. Peak sound level generated by the sterilizer is
a back sound level of 60 dBA.
B. Total heat transmitted by the sterilizer is less than 300 W/h
Air supply pressure (ELVC only)
after pressure reducer - psi (Bar)
Feed water pressure (ELVC only)
after pressure reducer - psi (Bar)
Mineral free water pressure after
pressure reducer - psi (Bar)