Tuttnauer 3840 ELVG-D, 3850 ELVG-D, 3870 ELVG-D Technician Manual

Technician Manual
Electronic Laboratory Autoclaves
Models 3840, 3850, 3870 ELVG-D
ELVG-D: standard autoclave
C: optional system for fast cooling
PV: optional vacuum pump
BH: optional bio hazard filtration
The Netherlands Tel: 31 (0) 765423510, Fax: 31 (0) 765423540
Cat. No. MAN205-0491000EN Rev. F Tuttnauer Europe B.V., Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda,
http://www.tuttnauerusa.com/
Page 1
Table of Contents
Paragraph , page no.
1. Introduction 4
2. Installation Instructions 6
2.1 Installation Site 7
2.2 Connection of the pressure regulator 7
2.3 Front View 9
2.4 Front View for the Fan Configuration 10
2.5 Rear View 11
3. Tests 12
3.1 Installation Tests 12
3.2 Periodical Tests 12
4. Technical Data 14
4.1 Directives and Standards 14
4.2 Water Quality 14
4.3 Reverse Osmosis 15
4.4 Electrical data 16
4.5 Specifications 17
4.6 Overall Dimensions 18
4.7 Construction 18
4.8 Utilities 19
4.9 Symbol Description 20
5. Description of the Control System. 21
5.1 Hardware Control System Components: Block Diagram ................... 21
5.2 Application System Architecture 21
5.3 Interfaces to Users 22
5.4 Inputs and Outputs 22
5.5 Hardware Control System components 24
6. Checking and Changing Parameters and Other Data 28
6.1 Browsing through the menus 28
6.2 Changing a Parameter 29
6.3 Quick options screen 31
6.4 Logging in and entering the Main menu 37
6.5 Directories and subdirectories 38
6.6 System Parameters 41
6.7 Inputs/Outputs 43
6.8 Maintenance 49
6.9 Advanced Options 53
6.10 Version handling 59
Page 2
6.11 Cycle parameters 61
6.12 Temperature sensors 63
6.13 Displayed inputs 64
6.14 Purge (generator models only) 65
6.15 Create Pulse 65
6.16 Keep Heat 67
6.17 Heating 68
6.18 Sterilization 68
6.19 Cooling (“C” models only) 69
6.20 Exhaust 70
6.21 Drying 70
6.22 Ending 72
6.23 Global (for some models only) 72
7. Generator description 75
7.1 Generator Electrical Safety System 76
7.2 The Water Pump for the Steam Generator 77
7.3 GICAR Water Level Controller 77
8. Maintenance and Replacement Procedures: the Autoclave 79
8.1 Preliminary Operations for Each Technician Call 79
8.2 Replacing the Circuit Breaker 80
8.3 Replacing the Safety Valve 81
8.4 Replacing the door gasket 83
8.5 Replacing the door cover 84
8.6 Replacing the Printer 89
8.7 Replacing the Door Micro-Switch 1 91
8.8 Replacing the Door Micro-Switch 2 91
8.9 Replacing the air filter 92
8.10 Replacing the locking motor 93
8.11 Replacing the Plunger or Coil of the 1/4" Solenoid Valve 9
5
8.12 Replacing the Plunger or Coil of the BACCARA Solenoid Valve 96
8.13 Replacing the Mechanic Pressostat 96
8.14 Replacing the Pressostat 0.2 (NC) 97
8.15 Replacing the power supplies, water level board and relays
98
9. Maintenance and Replacement Procedures: the Generator 103
9.1 Access to Generator 103
9.2 Draining the generator 103
9.3 Cleaning and Replacing the Water Level Electrodes 104
9.4 Replacing the Heating Elements 106
Page 3
9.5 Replacing the Generator Side Gauge Glass Indicator Tube 106
9.6 Replacing the Generator Thermostat 108
9.7 The Water Pump Replacement 110
9.8 Releasing the Vacuum Pump Rotor 110
10. Troubleshooting 112
10.1 Autoclave Troubleshooting 112
10.2 Steam Generator Troubleshooting 117
11. List of Spare Parts 120
11.1 Water Outlet Strainer 128
12. Pressure vs Temperature for Saturated Steam 129
13. Valves Numbering 138
14. Drawings 140
14.1 Piping Drawing – Model 3840/3850/70 ELVCG-D 140
14.2 Piping Drawing – Model 3840/3850/70 ELVCPVG-D 141
14.3 Piping Drawing – Model 3840/3850/70 ELVCPVGBH-D
1
42
14.4 Electrical Drawing for The Autoclave 3840-50/3870
ELV(C)(PV)G 380V.......................................................................... 143
14.5 Electrical Wiring Diagram: The Generator 151
14.6 Generator Control Electrical Cascade Diagram 152
Page 4
1. Introduction
This manual, together with the operator’s manual, forms the complete edition of the Operation and Maintenance instructions. This manual is intended for the use of the technician. It is forbidden for unqualified and unauthorized personnel to service the autoclave in accordance with the instructions in this manual. Any unauthorized service may result in the invalidation of the manufacturer’s guarantee.
The qualified technician shall be an authorized electrician with the right qualifications in electronics and shall be familiar with the local technical/electrical regulations.
Available optional configurations
Models of 3870 ELVG-D model range are laboratory sterilizers designed especially for the sterilization of instruments, liquids, and other materials in hospital laboratories, medical laboratories, research institutes, food laboratories and pharmaceutical facilities.
C
A special feature of the ELVCG-D models is the fast cooling stage for liquids. In this stage pressure in the chamber is increased by means of compressed air to compensate the fast decreasing of pressure due to the fast cooling. The fast cooling shortens the time required for safe handling of bottles.
Note: there is no cooling option for WR configuration
PV
Special feature of ELVCPV-D models is the vacuum pump that enables air removal prior to sterilization and during the drying stage.
The advantages of the pre-vacuum sterilizer in comparison to the regular gravity displacement steam sterilizer are as follows:
Removal of air pockets from packs and porous loads and most kinds of
tubes (rubber, plastic etc.) by vacuum at the first stage of the cycle.
Better steam penetration into the load; resulting in effective sterilization.
Better temperature uniformity.
Better drying of materials with closed doors due to the vacuum
achieved in the chamber at the end of the sterilization cycle.
Page 5
A special feature of the autoclave is the fast cooling stage for liquids. In this stage pressure in the chamber is increased by means of compressed air to compensate the fast decreasing of pressure due to the fast cooling. The fast cooling shortens the time required for safe handling of bottles
There is a configuration equipped with fan that allows to shorten the cooling operation.
BH
In ELVCPVBH-D (bio-hazard) models, a bio-hazard filter installed in the chamber enables sterilizing bio-hazard load without contaminating the surrounding.
Page 6
2. Installation Instructions
The following utilities have to be connected (Refer to the drawing below ‘Rear View’ of the autoclave).
Power outlet, as detailed in the table below:
Power
3 Ph, 400V/50/60Hz
Recommended Circuit Breaker
20A
If the 3 phase autoclave has to be connected to a one phase power
network, 1 x 230V, 50/60Hz., connect the 1ph 230V power source to the supplied switch box that has a 1 phase input (from the power source) and a 3 phase output (to the autoclave). See details at the end of this manual.
The power network must be protected by a current leakage relay. Mineral-free water having a conductibility lower than 15µs
(microsiemens), through a 1/2” flexible hose.
To obtain water quality meeting requirements a deionization
column or reverse osmosis apparatus can be installed. The water must be delivered at a pressure of 2-3bar. A pressure reducer shall be installed at the water source outlet as instructed below.
For fast cooling (if this option is provided):
Feed water from the water network, pressure 2-3bars, connected
through a 1/2” pipe. A pressure reducer shall be installed at the water source outlet as instructed below.
Compressed air, from a mobile compressor or compressed air network
at pressure 3-4 bars, will be set at 1.6-2bars at the autoclave inlet is to be connected by a flexible conduit of 3/16”.
The air must be of instrumental quality filtered at 5µ and free of
humidity and oil drops.
Drain connected by 1/2” pipe, located at the rear of the unit. The
chamber exhaust and coolant water is evacuated to an open waste funnel. The drainage piping must be heat resistant, to 80ºC, non­continuous flow.
Attention: Connection of water system to the autoclave must be performed
through "BACK FLOW PREVENTION SYSTEM" installation as per EN 1717.
Page 7
2.1 Installation Site
1. Install the autoclave according to the following guidelines:
2. Place the autoclave on the floor. Verify that the surface is leveled.
3. All utility supplies must be prepared in accordance with requirements, before autoclave installation e.g. mineral-free and tap water, compressed air, one or three-phase power network, connection to the drain of the building.
4. Leave the space free around the autoclave for maintenance and service requirements.
2.2 Connection of the pressure regulator
This paragraph refers to the pressure regulator of the feed water and to the pressure regulator of the mineral free water.
1. Attach the pressure regulator to the facility's water outlet.
2. Connect the supplied flexible hose to the pressure regulator's assembly outlet and to the autoclave's water inlet.
3. Verify that the pressure regulator is in right direction. The flow direction is indicated by an arrow stamped on the pressure regulator.
4. Verify that water flows freely in the feed water and in the mineral free water lines.
No.
Description
Cat No.
1
Side of water supply source
N/A
2
Reducer, Female 3/4" BSP to Male 1/2" BSP, Brass
FIT100-0424
3
Reducer, Female 1/2" BSP to Male 1/4" BSP, Brass
FIT100-0425
4
Pressure Regulator, Water, In-Line, with Strainer, 1/4 x 1.5 bar
GAU029-0059
5
Fitting, Adaptor, M 1/2" BSP x M 1/4" BSP, Brass Ni plated
PNE100-0042 6
Side of flexible hose (to autoclave
N/A
Page 8
FLOW DIRECTION
54321
6
Page 9
2.3 Front View
No.
Description
No.
Description
1
Front leg
8
Display
2
Rear leg
9
Keyboard
3
Left service door
10
Printer cover
4
Left service door grip
11
Paper slot and paper cutter
5
Left service door lock
12
USB socket
6
Door cover
13
Main switch
7
Pressure gauge
14
Printer
9
10
5
6
7
8
11
14
432
1
12
13
Page 10
2.4 Front View for the Fan Configuration
There is a configuration equipped with a fan that allows shortening the cooling
operation. The fan is located on the door of the autoclave.
Fan
Page 11
2.5 Rear View
No.
Description
No.
Description
1
fan grill
9
air pressure regulator
2
RJ45 connector
10
generator drain outlet
3
circuit breaker
11
tap water inlet
4
electrical socket
12
mineral free water inlet
5
gravitation valve for vacuum pump
13
generator pressure gauge
6
compressed air inlet
14
side gage glass indicator tube
7
autoclave air pressure gauge
15
generator safety relief valve
8
autoclave drain outlet
16
chamber safety relief valve
6
7
9
1
2
3
5
14
131215
81110
4
16
Page 12
3. Tests
3.1 Installation Tests
The service technician shall perform the following preliminary
checks before operating the autoclave:
a. Integrity Check
Perform a visual check to verify that there are no dents, scratches, broken gauges, etc.
b Leveling Check
Check that the autoclave is leveled.
c. Leakage current test
Check the precise operation of the earth leakage relay.
d. Continuity Check
Check the continuity of the grounding connection.
At this stage operate the autoclave and continue with the tests:
e. Safety Check
Check the safety elements; safety valve and the door locking mechanisms.
f. Programs Check
Run basic programs of the autoclave and check the operation sequences, the sterilization parameters etc.
g. Validation
Validate the sterilization cycles, taking in consideration the interface of packaging/goods/autoclave.
After the above steps are performed, the autoclave is ready for operation.
3.2 Periodical Tests
Period
Test
2 months
Test the safety valve by operating it.
6 months
Remove the cover of the autoclave, tighten the, valves and connectors in the control box.
Year Check the continuity of the grounding connections.
Check the temperature and pressure calibration.
Page 13
Perform validation of the autoclave.
Check the precise operation of the earth leakage relay.
Check that the autoclave is leveled.
Check the safety elements; safety valve, safety and cut-off thermostats door locking mechanisms.
Run basic programs of the autoclave and check the operation sequences, the sterilization parameters etc.
Check the water reservoir, piping, plastic parts and electric wires.
Check and tighten the piping joints to avoid leakage.
Check and tighten all screw connections in the control box, valves and instrumentation.
Observe the closing device for excessive wear
5 years
Observe the closing device for excessive wear
Safety tests (pressure vessel, efficiency, electrical) shall be performed in accordance with local rules or regulations, by an authorized inspector.
Only an authorized technician shall perform the 6-months and yearly tests!
Check connections to distilled or mineral free water. Open the
distilled or mineral free water valves. Manually test the distilled or mineral free water valves by over-riding the appropriate solenoid valves. If there are no leaks, –leave the water manual inlet taps open.
Observe the gauge and see that the generator reaches working
pressure.
Check that the jacket pressure gauge, reaches 2.4 bars (35 psi).
Page 14
4. Technical Data
4.1 Directives and Standards
Tuttnauer. Ltd. company meets the provisions of the following standards:
ISO 9001:2008 (Quality Systems)
Tuttnauer. Ltd. company also works in conjunction with and refers to:
ANSI/AAMI ST55 American Society of Mechanical Engineers
Section VIII, Division 1, for unfired pressure vessels.
EN 13060 Small Steam Sterilizers. UL UL 61010-1 PED 97/23EEC IEC IEC 61010-2-040 Safety ISO 17665-1:2006 (Validation and Routine Control)
4.2 Water Quality
The distilled or mineral-free water supply shall be according to the table below:
Physical Characteristics and Maximum acceptable contaminants levels in water or steam, for steam generator and sterilizers (According to EN 285: 2006)
Contaminants in water supplied to generator
Contaminants in condensate at steam inlet to sterilizer
Evaporate residue
≤ 10 mg/l
N/A
Silicate (SiO2)
≤ 1 mg/l
≤ 0.1 mg/l
Iron
≤ 0.2mg/l
≤ 0.1mg/l
Cadmium
≤ 0.005 mg/l
≤ 0.005 mg/l
Lead
≤ 0.05 mg/l
≤ 0.05 mg/l
Rest of heavy metals except iron, cadmium, lead
≤ 0.1 mg/l
≤ 0.1 mg/l
Chloride (Cl)
≤ 2 mg/l
≤ 0.1 mg/l
Page 15
Phosphate (P2O5)
≤ 0.5 mg/l
≤ 0.1 mg/l
Conductivity (at 25°C)
≤5 µs/cm
≤ 3 µs/cm
pH value (degree of acidity)
5 to 7.5
5 to 7
Hardness (Σ ions of alkaline earth)
≤ 0.02 mmol/l
≤ 0.02 mmol/l
Appearance
Colorless, clean, without sediments
[Minimum specific resistivity of 1 megohm per cm (1 M
Ω/cm)]
Compliance with the above data should be tested in accordance with acknowledged analytical methods, by an authorized laboratory.
Attention: The use of water for autoclaves that do not comply with the table
above may have severe impact on the working life of the sterilizer and can invalidate the manufacturer’s guarantee.
Use Only deionized water, Having a maximum conductance of 1 microSiemen per cm (1 μS/cm) Tap Water supply
Maximum hardness value 0.7-2.0 mmol/l The use of soft water is strictly forbidden! Please consult a water specialist!
4.2.1 Water for the Vacuum pump and drain cooling
The feed water supplied to the liquid ring vacuum pump must meet the following requirements:
Hardness: 0.7 - 0.2 mmol/l. Water temperature: shall not exceed 15°C.
Note: The use of heavy scaled water for the vacuum pump cooling, can
cause blocking of the rotor and put the pump out of operation. This invalidates the guarantee for the vacuum pump.
4.3 Reverse Osmosis
A Reverse Osmosis system may be used to improve the quality of the water used to generate steam in the autoclave chamber. The use of mineral free will contribute to better performance and longer life of the autoclave.
Page 16
4.4 Electrical data
230V configuration
Property
Value:
3840
Value:
3850
Value:
3870
Total Power
9000
9000
9000
Voltage
1Ph,
230VAC
1Ph,
230VAC
1Ph,
230VAC
Amperage
22A w/o PV, 25A
with PV
22A w/o PV, 25A
with PV
22A w/o PV, 25A
with PV
Protection against electrical shock
Class I (IEC 60601-1)
Mains supply fluctuation
+/- 10%
Degree of protection by enclosure
IP31
400V configuration
Property
Value:
3840
Value:
3850
Value:
3870
Total Power
9000
9000
9000
Voltage
3Ph,
400VAC
3Ph,
400VAC
3Ph,
400VAC
Amperage
13A w/o PV, 16A
with PV
13A w/o PV, 16A
with PV
13A w/o PV, 16A
with PV
Protection against electrical shock
Class I (IEC 60601-1)
Mains supply fluctuation
+/- 10%
Degree of protection by enclosure
IP31
Page 17
4.5 Specifications
Properties
Model
3840
3850
3870
Chamber diameter
380
380
380
Chamber depth
400
500
690
Chamber volume
52 liters.
65 liters.
85 liters.
Overall dimensions
Height
1000
1000
1000
Width
730
730
730
Length
700
700
700
Maximum dimensions (door open)
Width
800
800
800
Length
1500
1500
1500
Net weight (kg)
105kg
108kg
113kg
Shipping volume
0.9 m3
0.9 m3
0.9 m3
Shipping weight
133kg
136kg
141kg
Max. Allowable Working pressure (MAWP)
2.8 bar (40 psi)
Page 18
4.6 Overall Dimensions
4.7 Construction
Chamber and door material
Stainless steel
Outer Cabinet
Stainless Steel
Chamber insulation
Fiberglass with reinforced material
Page 19
4.8 Utilities
4.8.1 Electrical utility
Power
Switch box
Recommended Circuit Breaker
208-220V (3 Ph + Earthing)
No
20A
380-400V(3 Ph + Neutral + Earthing)
No
16A
230V 1ph
yes
32A
Frequency
50/60Hz
4.8.2 Other Utilities
Compressed Air (ELVCPVG-D only)
1/2" 3-4 Bar (44-58 psi)
Tap water
1/2", 2-3 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding
temp. of 80°C (176°F)
Attention:
A switch or circuit-breaker must be included in the building installation.
This switch or circuit-breaker shall be in close proximity to the equipment, within easy reach of the operator; and marked as the disconnecting device for the equipment.
The electrical net must be protected with a current leakage safety
relay.
The electrical network must comply with local rules or regulations. Verify that there is an easy access to the main power switch, to the
water cut-off valve and to the current leakage safety relay.
Make sure while placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
All water connections to autoclave must be performed through "BACK
FLOW PREVENTION SYSTEM" only, as per IEC 61770.
Page 20
4.9 Symbol Description
Caution! Consult accompanying documents
Caution! Hot surface. Caution! Hot steam.
Protective earth (Ground)
On-Off
Page 21
5. Description of the Control System.
5.1 Hardware Control System Components: Block Diagram
5.2 Application System Architecture
The system is divided into three main sections (dll)
1. GUI – holds all the Human Machine interface including the main application screen and all the configuration screens which enable the user to handle the machine.
2. Logic – holds all the application logic for running the machine.
3. Utilities – Holds general functionality which is used by the logic section and the GUI section e.g: converting function for displaying different pressure or temperature units type, languages etc.
Main board
- CPU +memory- Backup
power input
-MCIMX27LVOP4A+MC13783VK5
- R.T.C (including Battery)
M41T81SM6E
- Memory card expansion
MT46H32M16LFBF-
6:B+EPM570F256C5+S71WS256P
Ethernet
Am79C874
I/O Board
CPU- STM32F103R6T6
-24 Digital Outputs
-2 Analog outputs
-9 Digital inputs
-4 Analog Inputs 4-20mA
-6 Analog inputs Temp pt100
-3 Inputs water level
Keypad
Graphic Display 3.5"
USB Memory
socket
ISP1504ABS
24/12
VDC
Input
COM1
(RS232)
MAX323
2CSE+
Printer
24/12 VDC
GUI
LOGIC
UTILITY
Page 22
Mapping of the software to the hardware – see Software Development Plan.
The Hardware architecture is based on Freescale i.MX27 PDK Evaluation Board.
5.3 Interfaces to Users
Keypad: The keypad has three push buttons:
o Down key
o Up key
o Start/stop key
Display: The control system has a graphical display
USB socket:
The USB socket is intended to load cycles' history from flash a memory (disk on key). The received file is in txt format that can be loaded onto a PC.
5.4 Inputs and Outputs
Analog inputs
Analog inputs
JP
Drain temperature
J5
Chamber temperature
J2
Bio hazard filter temperature
J6
Ref temperature
J3
Jacket temperature
J4
Chamber pressure
J7/1
Page 23
Digital inputs
Digital Inputs
JP
Door closed
J12/1
Door unlocked
J12/6
Door locked
J12/8
Digital outputs
Digital Outputs
JP
Steam to jacket
J13/8
Vacuum pump
J13/12
Exhaust to bio
J13/14
Main exhaust
J13/16
Atmospheric air valve
J14/2
Compressed air to chamber
J14/4
Slow exhaust
J14/6
Fast exhaust
J14/8
Vacuum valve
J14/10
Cool Drain
J14/12
Steam to chamber
J14/14
Cooling water
J14/16
Compressed air to jacket
J15/2
Close door
J15/4
Open door
J15/6
Air to door
J15/14
Actuators
The control system operates electrical valves (solenoids), Electric cylinder motors, pneumatic cylinders, electric pumps, heaters.
On / off switch
A Rocker Switch 250V AC, 16A
Page 24
5.5 Hardware Control System components
The hardware is consisted of cards: MAIN and IO
Main card
Operating system (MAIN card)
The Operating system is Microsoft Windows CE version 6. The code to the Operating system is supplied by Microsoft. The connection code between the Operating system and the hardware components (BST –
Power Supply
Connector
(24V DC)
Connector to
printer
+
Ethernet
Connector
SD
Not in use
USB
Connector
-
Connector
to IO Board
Connector to LCD display
Connector to Keypad
Lithium coin
battery T.H
orizontal 3V
48mAh
J5
BT1
J8
J4
P2
P4
T1
P3
J3
Page 25
Board Support Package) is supplied by FreeScale. Minimum suitability for this "Tuttnauer" project is performed in this code.
Development tools (MAIN card)
The software develop environment is on Microsoft Visual Studio 2005 that includes the Microsoft Platform Builder for Windows CE 6.0. The specific Tuttnauer system application is written in C Sharp.net on Microsoft Compact Framework.net
IO Card
Analog inputs
(PT-100)
Analog inputs
-
20mA)
Analog inputs
electrodes
J
J
J6
J3
J4
J1
J7
J11
J16
J17
J13
J15
J14
Connection to
MAIN Board
Analog
outputs
Digital
outputs
Digital Inputs
J12
Page 26
IO card is an independent card. It checks the card by a method of a continuous scan of all the ports, saving the data in the memory and transporting this data according to the request in the communication channel RS-232 to the personally adapted protocol.
At the end of the check process of the digital and analog ports, the communication channel is checked, in case of receiving a request, the request is checked and if the request is legitimate it will be taken care of.
The IO card is controlled by the MAIN card. Only when requests are received from the MAIN, the requests will be checked and performed.
The IO card will perform an electrical restart of the systems (IO card) if it does not receive requests within 5 seconds. In the next request received from the MAIN, the IO card will indicate to the requester of IO card systems restart.
The "brain" in the IO card runs on microcontroller type: STM32F103R6T6.
The Microcontroller is an electrical Integrated circuit with an actualized complete system (SoC) including a processing unit, external interfaces, memory, etc.
The Microcontroller is constructed of a number of main components:
Central processing unit – the main component in the microcontroller is a central processing unit (CPU) This unit is in charge of receiving the request (command) code from the memory, deciphering, receiving variable data from the memories and performing the command.
Memory – divided into 2 sections.
software memory – in this memory the software that samples the IO card and request (command) from the user are saved. As this memory is a flash memory it is possible to burn the code many times (efficient to the development stage).
information memory – in this memory the data and other variables values are saved. This memory is not flash memory, but is RAM memory.
Digital / Analog ports (in/out) – through them the microcontroller can receive data of digital/analog signs from the system, save them and pass them on respectively to the user request.
Ports (in / out) to the communication – in the IO card there is a UART component. This component enables communication with the MAIN card through the RS-232 in suitable protocol.
MAIN and IO cards :communication protocol between IO board to Main Board
Page 27
Communication protocol between IO to Main boards is described below.
The communication manged by the Main board (master), the IO functions as slave.
There are six communication functions. The communication functions (to read or write) identified by number. The communication functions listed below:
01 Read – command to IO send all the information of digital analog
inputs.
02 Write – Command to IO to activate analog and digital outputs. 03 Read version – IO software version to verify the version. 04 N.A. 05 Read number that identifies the ID of IO board and the software
version.
06 Write (only at the first time) to IO his ID and software number.
First Byte (Byte 1).
The first byte identifies ID functionality request. It can be one of the numbers in a paragraph.
The second byte will present the data size if the main board asks to write information to the IO. If the request is to read this byte will be 00.
The number of the request. This number is increased from 00-ff. Check sum of the package – to prevent mail functionality of the board if
there is interference in communication. The information package transfer. The information contains number of
the input or output and data for/from each I/O. Bytes identify number of input or output followed by two bytes
containing the data of the input or outputs. Last byte contains a check sum of the entire package that is
transferred. This is to verify that the information is not corrupted.
Example: The main board sends:
01;00;0003;0004; 01 Ask to Read inputs. 00 not transfer information. 0003 Request number 3 0004 Checksum to verify the request
Page 28
6. Checking and Changing Parameters and Other Data
Bacsoft control panel allows changing parameters of the cycle and of the system, exporting various data to, and importing from, a USB device or to the printer, and some other options.
Cycle parameters are changeable for Custom programs only (see Duplicate cycles), with the exception of the Temperature sensors, Displayed inputs, and Dry Time.
6.1 Browsing through the menus
Now you will learn how to browse through the folders. When you read the Directories and subdirectories chapter with links to specific menus, you will need to know how to browse through the folders using the autoclave control panel. Below is the instruction.
Login as Technician (see 6.4). The Main menu screen appears. To browse through the menus:
1. Press the Up and Down keys to scroll through the menus.
2. Press the Start/Stop key to enter the next screen (i.e. to get one level down).
3. Repeat steps 1 and 2 to enter the next screen until you get to required screen.
Below are the example screens for the following menu: Cycle Parameters\ Drying\ Dry Time:
Loading...
+ 124 hidden pages