This manual, together with the operator’s manual, forms the complete edition of the
Operation and Maintenance instructions. This manual is intended for the use of the
technician. It is forbidden for unqualified and unauthorized personnel to service the
autoclave in accordance with the instructions in this manual. Any unauthorized service
may result in the invalidation of the manufacturer’s guarantee.
The qualified technician shall be an authorized electrician with the right qualifications
in electronics and shall be familiar with the local technical/electrical regulations.
2.SYMBOL DESCRIPTION
Caution! Consult accompanying documents
Caution! Hot surface.
Caution! Hot steam.
Protective earth (Ground)
Stand by
Page 4 of 132 Pages
3.INSTALLATION INSTRUCTIONS
Recommended
For all the ELV models, the following utilities have to be connected (Refer to the
drawing below ‘Rear View’ of the autoclave).
Power outlet, as detailed in the table below:
ModelPower
2540
3150/70
3850/70
3850/70
3850/70 with switching box
3850/70, 5050/75
5050/75
5050/75
5050/75 with switching box
If the autoclave is a 3 phase 3870, 5050 or 5075 autoclave and has to be connected
to a one phase power network, 1 x 230V, 50/60Hz., connected the 1ph 230V
power source to the supplied switch box that has a 1 phase input (from the power
source) and a 3 phase output (to the autoclave). See details at the end of this
manual.
The power network must be protected by a current leakage relay.
Mineral-free water having a conductibility lower than 15µs (microsiemens),
through a 1/2” flexible hose.
1Ph, 230V/50/60Hz20A
1 or 2 Ph, 230V/50/60Hz25A
3 Ph, 208V/50/60Hz25A
3 Ph, 230V/50/60Hz25A
1 or 2 Ph, 230V/50/60Hz30A
3 Ph, 400V/50/60Hz20A
3 Ph, 208V/50/60Hz30A
3 Ph, 230V/50/60Hz30A
1 or 2 Ph, 230V/50/60Hz45A
Circuit Breaker
To obtain water quality meeting requirements a deionization column or reverse
osmosis apparatus can be installed. The water must be delivered at a pressure of 23bar. A pressure reducer shall be installed at the water source outlet as instructed
below.
For fast cooling (if this option is provided):
Feed water from the water network, pressure 2-3bars, connected through a 1/2”
pipe. A pressure reducer shall be installed at the water source outlet as instructed
below.
Compressed air, from a mobile compressor or compressed air network at pressure
3-4 bars, will be set at 1.6-2bars at the autoclave inlet is to be connected by a
flexible conduit of 3/16”.
The air must be of instrumental quality filtered at 5µ and free of humidity and oil
drops.
Drain connected by 1/2” pipe, located at the rear of the unit. The chamber exhaust
and coolant water is evacuated to an open waste funnel. The drainage piping must
be heat resistant, to 80ºC, non-continuous flow.
Attention:
Connection of water system to the autoclave must be performed through "BACK
FLOW PREVENTION SYSTEM" installation as per EN 1717:2001 regulations
and local regulations.
.
Page 5 of 132 Pages
54321
6
3.1Installation Site
Pressure Regulator, Water, In
-
Line,
1.Install the autoclave according to the following guidelines:
2.Place the autoclave on the floor. Verify that the surface is leveled.
3.All utility supplies must be prepared in accordance with requirements,
before autoclave installation e.g. mineral-free and tap water, compressed
air, one or three-phase power network, connection to the drain of the
building.
5.Leave the space free around the autoclave for maintenance and service
requirements.
3.2Connection of the pressure regulator
This paragraph refers to the pressure regulator of the feed water and to the
pressure regulator of the mineral free water.
1.Attach the pressure regulator to the facility's water outlet.
2.Connect the supplied flexible hose to the pressure regulator's assembly
outlet and to the autoclave's water inlet.
3.Verify that the pressure regulator is in right direction. The flow direction is
indicated by an arrow stamped on the pressure regulator.
4.Verify that water flows freely in the feed water and in the mineral free
water lines.
FLOW DIRECTION
No.DescriptionCat No.
1Side of water supply sourceN/A
Reducer, Female 3/4" BSP to Male
2
1/2" BSP, Brass
Reducer, Female 1/2" BSP to Male
3
1/4" BSP, Brass
4
with Strainer, 1/4 x 1.5 bar
Fitting, Adaptor, M 1/2" BSP x M
5
1/4" BSP, Brass Ni plated
FIT100-0424
FIT100-0425
GAU029-0059
PNE100-0042
6Side of flexible hose (to autoclaveN/A
Page 6 of 132 Pages
REAR VIEW MODEL 2540
Page 7 of 132 Pages
REAR VIEW MODELS 3150, 3170
Page 8 of 132 Pages
REAR VIEW MODELS 3850, 3870
Page 9 of 132 Pages
REAR VIEW MODELS 5050, 5075
Page 10 of 132 Pages
4.TESTS
4.1Installation Tests
The service technician shall perform the following preliminary checks
before operating the autoclave:
a.Integrity Check
bLeveling Check
c.Leakage current test
d.Continuity Check
At this stage operate the autoclave and continue with the tests:
e.Safety Check
f.Programs Check
g.Validation
After the above steps are performed, the autoclave is ready for operation.
Perform a visual check to verify that there are no dents, scratches,
broken gauges, etc.
Check that the autoclave is leveled.
Check the precise operation of the earth leakage relay.
Check the continuity of the grounding connection.
Check the safety elements; safety valve and the door locking mechanisms.
Run basic programs of the autoclave and check the operation
sequences, the sterilization parameters etc.
Validate the sterilization cycles, taking in consideration the interface of
packaging/goods/autoclave.
4.2Periodical Tests
PERIODTEST
2
months
6
months
Test the safety valve by operating it
Remove the cover of the autoclave, tighten the screws of the heaters and
the electrical connections at the heaters, valves and connectors inthe
control box.
Check the continuity of the grounding connections.
Check the temperature and pressure calibration.
Perform validation of the autoclave.
Check the precise operation of the earth leakage relay.
Check that the autoclave is leveled.
Check the safety elements; safety valve, safety and cut-off thermostats
Year
door locking mechanisms.
Run basic programs of the autoclave and check the operation sequences,
the sterilization parameters etc.
Check the water reservoir, piping, plastic parts and electric wires.
Check and tighten the piping joints to avoid leakage.
Check and tighten all screw connections in the control box, heaters and
valves and instrumentation.
Observe the closing device for excessive wear
.
5 yearsObserve the closing device for excessive wear
Safety tests (pressure vessel, efficiency, electrical) shall be performed in accordance
with local rules or regulations, by an authorized inspector.
Only an authorized technician shall perform the 6-months and yearly tests!
Page 11 of 132 Pages
5.WATER QUALITY
5.1Water for Generating Steam
The distilled or mineral – free water supplied to the sterilizer shall be according
to the table below:
Physical Characteristics and Maximum acceptable contaminants levels
ElementCondensate – allowable content
in steam for sterilizers
(According to EN 13060:2004).
Silicium oxide. SiO
2
≤0.1 mg/kg
Iron≤0.1 mg/kg
Cadmium≤0.005 mg/kg
Lead≤ 0.05 mg/kg
Rest of metals except iron,
cadmium, lead
≤0.1 mg/kg
Chloride (Cl)≤0.1 mg/kg
Phosphate (P2O5)≤0.1 mg/kg
Conductivity (at 20°C)≤3 μs/cm
pH value (degree of acidity)5 to 7
AppearanceColourless clean without sediment
Hardness (Σ Ions of alkaline earth)≤0.02 mmol/l
Compliance with the above data should be tested in accordance with
acknowledged analytical methods, by an authorized laboratory.
Attention:
We recommend testing the water quality once a month. The use of water that
do not comply with the table above may have severe impact on the working
life of the sterilizer and can invalidate the manufacturer’s guarantee.
5.2Water for the Vacuum pump and drain cooling
The feed water supplied to the liquid ring vacuum pump must meet the
following requirements:
Hardness:0.7 - 0.2 mmol/l.
Water temperature:shall not exceed 15°C.
Note:
The use of heavy scaled water for the vacuum pump cooling, can cause
blocking of the rotor and put the pump out of operation. This invalidates the
guarantee for the vacuum pump.
5.3Reverse Osmosis
A Reverse Osmosis system may be used to improve the quality of the water
used to generate steam in the autoclave chamber. The use of mineral free will
contribute to better performance and longer life of the autoclave.
Page 12 of 132 Pages
6.DESCRIPTION OF THE CONTROL SYSTEM.
1
JP8
(See CDR Control diagram).
The control system is based on 3 electronic boards designed according to the autoclave
requirements, the digital board DIG-T2 containing the micro-controller memories,
buffers and digital ICs and the analog board ANL-T2 which performs the processing
of signals coming from the sensors and switches. The AC-T1 board consists of filters
and AC drivers. DC voltages 12V and 5V, supplied by a switching type power supply,
powers the 3 electronic boards.
CONTROL SYSTEM BLOCK-DIAGRAM
SKT
20
1
DIG-T2 BOARD
C2
+
U12
R3
C9
U8
R2
U16
R1
U11
U14
C5
U21
U17
DIG T3
1
U3
C3
+
RP1
U15
R9
U22
97-05
C24
C25
D1
C15
8
C23
JP3
C22
U1
2
DS1
1
PLUG
37
JP1
19
C7
U18
C4
R4
U4
C16
C17
Q1
Y1
BZ1
U19
C6
34
33
20
1
P1
1
JP7
JP2
JP3
1
JP12
1
37
19
JP1
12
1
JP6
JP5
JP4
1
1
1
JP9
JP10
1
The system is provided with communication interfaces RS 232 to PC and to parallel
printer.
6.1Digital Board DIG - T2
— The digital board is connected to the keypad panel, to the parallel printer
and to the analog board ANL-T2.
— The board contains the micro-controller (U15) type 80C32 that runs the
software program of the system.
— On the board, are three types of memories:
1)EPROM memory (U17), part no. 27512 storing the program codes.
2)RAM memory (U21) with a capacity of 32KB for the temporary data
during the running of the program.
3)EPROM memory (U22) that is a fixed serial memory with an
electrical writing and erasing.
— This EPROM serves as a non-volatile memory, enabling the system to
change follow-up tables during running of program codes, and ensuring
this data is not lost in case of power failure.
— The board contains an optimal Real Time Clock (U12), which serves as a
clock to the system, including a back-up battery, which ensures the clock,
runs continuously even when the autoclave is not powered.
Page 13 of 132 Pages
This component is optimal because it is related to the operation of the
BZ
1
1
RP
8
1
+
printer, which is also an optional item.
— The board contains a watchdog, which detects any faulty situation in the
running program code. It performs an automatic reset of the microcontroller and stops all the commands to avoid an uncontrolled activation
of any of the heating elements or the valves.
— The board functions as an MMI (Man-Machine Interface). It is connected
to a LCD display of two rows with 16 characters on each row and to the
following light indicators:
— START (autoclave in process),
— FAIL (the process failed),
— WATER (no water in the reservoir),
— DOOR (the door is not closed).
A keyboard connected to the digital board, serves as a control panel containing
the command and programming keys.
— The digital inputs and outputs are transferred to the system, as follows:
— Through the digital board to the analog board by means of buffers
74HC377.
— RS232 interface is performed on the board by the U13 component, the
signal is transmitted to the communication connector located on the ANLT2 board.
— The printer is connected directly to this board, connector DS1 enables to
connect the DPU-20 printer manufactured by SEIKO.
The printer receives the data and the supply voltage directly from this
connector.
The layout of the DIG-T2 board components is provided below
DIG- T2 BOARD
SK
20
R8
JP 3
DIG
U1
C1
U1
C9
U1
U1
C2
R7R6
C1
C2
C2
C2
DS
C1
+
C2
R3
R2
R1
C5
U2
U1
H2
97 -0 5
1
C1
U5
U1
C1
JP
C8
U 7
U8
U1
U3
C3
RP
U1
R9
U2
D
C1
2
1
U1
37
R5
R1
+
C1
U6
C4
U4
C1
Y1
JP 1
19
C1
C2
U2
R4
C1
C2
R1
R1
U2
C7
U1
Q 1
U1
C6
34
33
Page 14 of 132 Pages
6.2Analog Board ANL-T2
The analog board contains the drivers of the valves, heaters and pump. It
contains the sensors circuits connected to the control system, and serves as a
junction to the autoclave connections.
The input is 12VDC & 5VDC and the output is for all the signals to the
autoclave.
— The power supply provides the DC voltages 12V and 5V through the
connector JP3 to the board and further to the DIG-T2 board.
— The computer is connected to the board through the connector P1, by
RS232 interface.
— The analog and digital inputs, from the temperature sensor, electrode door
switch, etc. are connected through connector JP6, the conversion circuits
of the sensors are located on this board.
— The pressure sensor MPX2200 that measures the chamber pressure is
connected to the board through connector JP8, to the conversion circuit of
the sensor that is located on this board.
— The analog to digital conversion circuit A/D (U5) and the analog
multiplexer (U6) for 8 analog inputs (temperature, pressure, and electrode)
are located on this board and transfer digital signal to the DIG-T2 board.
— All the drivers and power circuits to the solenoids (valves, door locking)
are included on this board.
Layout diagram of the ANL-T2 is provided below.
ANL - T2 BOARD
PLUG
P1
JP2
JP12JP8
1
ANL T1 BOARD
20
1
JP1
1211
1
JP7
JP3
1
1
JP5
JP4
1
JP9
1
37
19
JP6
JP10
1
JP11
1
Page 15 of 132 Pages
6.3AC - Board - AC-T1
This board receives command signals from ANL-T2 board designated to
activate AC devices. It converts them to AC mode.
The AC-T1 board provides AC filtering for the power supply to the control
system and protection from sharp current fluctuations.
The AC-T1 board includes:
— AC drivers for the AC devices (pump, heaters, etc.) up to 6A.
— AC - input connector JP3.
— Filter circuit based on L1C5, the power transfer from the filter to power
supply via JP2.
— Spike protection - RV1÷3.
— Three - 230/115 VAC output channels, triac control circuit.
Q1 - 16Amp - JP4/1,2
Q2 - 8 Amp - JP7/1
Q3 - 8 Amp - JP7/2
AC-T1 BOARD
(JP3,JP4,JP7 ON THE BOTTOM SIDE OF THE BOARD)
U3
R12
JP5
R9
C3
R11
Q3
1
R10
R8C2
JP7
U2
JP1
R5
R6
R7C1
JP4
1
Q2
R4
1
R1
U1
R3
Q1
RV3RV2
JP2
R2
L1
JP6
JP3
C5
ETYPEAC-97
CAT TECH.
RV1
Page 16 of 132 Pages
7.CALIBRATION OF TEMPERATURE AND PRESSURE
a X
Y
310
100
7.1Method of Calibration:
The calibration of temperature and pressure is performed digitally.
The temperature and pressure measuring circuits are designed with components
having 1% accuracy.
The temperature circuit is linear and has an output of 100mV÷2400mV for a
temperature range 20°C÷150°C.
The pressure circuit is also linear and has an output of 100mV÷2400mV for a
pressure range 0÷400 kPa.
The measuring at the A/D is limited for values higher than 2400mV or below
100mV.
It is obvious that although the accuracy of the circuit components is 1%, the
sum deviation may reach 5%. Therefore calibration is necessary.
The system has a non-volatile memory in which the offset and gain data of the
sensors are stored.
Calibration is performed by use of the keyboard (or the communication port for
computer use).
Following is a description of the calibration of temperature and pressure using
the keyboard. The calibration procedure is identical to all autoclave models.
Definition of Gain and Offset:
Every straight line can be defined by the equation ax+b=y. “a” is the deviation
from y=0 and “b” is the slope of the line. In our case a=Gain and b=Offset.
The system is based on calculation of offset and gain of two points as in the
following example
If the actual measured pressures are 100 kPa & 300 kPa and the displayed
pressure are 90 kPa & 310 kPa respectively, the ‘input Vs pressure” graph will
be as follows.
READ
300
90
PRESURE
V
1
INPUT FROM SENSOR
V
2
ACTUAL
The calibration method enables to input this data into the system in order to
perform automatic correction of OFFSET and GAIN.
It is necessary to define the two points, which have to be input to the system,
prior to performing calibration.
Page 17 of 132 Pages
7.2Temperature Calibration Procedure
— Press the PROGRAMkey.
— The system enables to set the system’s date the time.
— This operation is skipped by pressing PROGRAMkey six times.
—
CODE: will be displayed.
— Select Technician Code (011) by the UP/DN keys and press the
PROGRAM key.
— Select CALIB CODE: 107 to calibrate the system by means of the UP/DN
keys and press the PROGRAM key again until TEMP CAL:
will be displayed.
— Set calibration code to TEMP CAL: 1
— Press the PROGRAMkey again and do not change the code.
— The following data will be displayed:
UPPER ROW:
LOWER ROW: A 130.0 R130.0
Enter the actual data (A) and the displayed data (R) for two known points as in
the following example:
The actual temperatures are 70°C & 120°C and the displayed temperatures are
68°C & 120°C respectively. the
Proceed as follows:
— On the upper row, change A60.0 to A 70.0 by means of UP/DN keys.
— Press the PROGRAMkey.
— The cursor on the upper row will move to R60.0, change to
R68.0 by means of UP/DN keys.
— Press the PROGRAMkey.
— The cursor moves to the lower row to A 130.0, change it to A 120.0 by
means of UP/DN keys.
— Press the
— The cursor on the lower row will move to R130.0
means of keys UP/DN, then press the PROGRAM key
PROGRAMkey.
A 60.0 R 60.0
, change to R123.0 by
This completes calibration procedure for temperature.
Page 18 of 132 Pages
Remarks:
— When cursor is set under any figure, pressing the START/STOP key stops
calibration.
— When the cursor is set under any figure, it enables to feed this value to the
reading of temperature by pressing the
Example:
If the cursor is set under R68.0 on the upper row and the actual temperature of
the autoclave is now 66.0ºC, by pressing the PARAMETERS key, the
temperature reading will be
corrected from R68.0 to R 66.0.
This calibration method enables the technician to change a certain point
without affecting a second point.
Example:
When the working temperatures are 134ºC and 121ºC but the actual
temperature are 132ºC and 121ºC
The measured temperature for 134ºC must be corrected without changing the
temperature at 121ºC,
Considering the above example, the setting of data can be done as follows:
A 121.0R 121.0
A 132.0 R 134.0
The system will calibrate the new offset and gain and retain them into the nonerasable memory.
As a result, while running a 134ºC program the correction is automatically
made, while at 121ºC no change is felt.
PARAMETERS key.
Page 19 of 132 Pages
7.3Pressure Calibration Procedure
To perform the pressure calibration two points have to be defined allowing the
calculation of new OFFSET and GAIN values.
— Press the PROGRAM key.
— Skip date and the time setting by pressing the PROGRAMkey six
times.
— CODE: will be displayed.
— Select the Technician Code (011) by the UP/DN keys and press the
PROGRAMkey.
— Select CALIB CODE: 107 to calibrate the system by means of the
UP/DN keys and press the PROGRAMkey again until the system
displays: PRES CAL:
— Press the PROGRAM key.
The following is displayed:
Point 1 – Upper Row A 100 R 100
Point 2 - Lower Row A 300R 300
— Open the door, do not alter A 100.
— Press the PROGRAM key, the cursor will move to R 100.
— Press the PARAMETERS key. The system performs reading
pressure and displays it instead of R100. E.g. R97 (i.e. the
system reads out 97kPa, when the door is open).
— Press the PROGRAMkey.
— The cursor moves to the lower row to A300.
If the system reads out 280kPa, do not change A300. Press the PROGRAM
key to move to R300and change it to the actual
pressure value 280kPa by means of the keys UP/DN.
— Press the PROGRAMkey again.
The system calculates the new values for offset and gain and stores these
values in the non-volatile memory.
This completes the calibration procedure for the pressure.
Page 20 of 132 Pages
8.TEST POINTS
The test point list provides testing points on the junctions on board to assist in locating
the malfunction.
The software for the control system of laboratory autoclaves is constructed as
one software for all types of laboratory autoclaves (table and vertical), version
LAB00VN1.
The software contains a table of parameters of which part of them defines the
autoclave, and part defines the processes in the autoclave. This document
describes parameters and their task in the software.
9.2Changing Cycle Parameters
Access to the program is achieved by inserting the access code.
Access code 1 –This access code enables change of the sterilization
time, sterilization temperature and the drying time.
Access code 11 –changes that may be done by the supervisor.
Enables to change a few parameters
Access code 33 –changes that may be done by the technician.
Enables to change all the parameters
To change the parameters listed below, proceed as follows:
a.Select the cycle.
b.Enable the operation by entering the ACCESS CODE (11).
c.Set the parameter that has to be modified, by pressing successively the
PROGRAM key.
d.Set the desired data by means of the UP/DN keys.
e.Pressing the PROGRAM key enters the modified data into memory, and
moves to the next parameter.
The pre-set parameters values will replace the last default values and become
the updated default values.
9.2.1STER Temp – Required sterilization temperature
Entry Code–1
Resolution–1ºC
Minimum value for the change–60ºC
Maximum value for the change–137ºC
Set
value
9.2.2STER Time – Required sterilization time for the process
Entry code–1
Resolution–1 minute
Minimum value–3 minutes
Cycle
Value
1234567
1341211211211051210
Maximum value–99 minutes
Set
value
Cycle
Value
1234567
315202018300
Page 22 of 132 Pages
9.2.3Dry Time – Required drying time for the process
Entry Code–1
Resolution–1 minute
Minimum Value–0 minutes
Maximum Value–99 minutes
Set
value
9.2.4Water Time – Time for entering water to the autoclave
This value defines the entry time of the water to the autoclave to
locate the electrode touching the water.
This time will change from process to process and even for different
autoclaves.
When the autoclave is vertical and the defined time is Ø seconds, the
water valve will not open at all.
Entry Code–11
Resolution–1 second
Minimum Value–1 second
Maximum Value–200 seconds
Set
value
Cycle
Value
Cycle
Value
1234567
0000000
1234567
20202020202020
9.2.5Heat T.O. - The maximum time allowed for the heating stage
(testing)
After this time, if the system does not enter the sterilization stage, it
will receive a ‘Low Heat’ message and the process will abort.
Entry Code–11
Resolution–1min
Minimum Value–20min
Maximum Value–300 min
Set
value
Cycle
Value
1234567
6060609090900
Page 23 of 132 Pages
9.2.6End Temp - The ending temperature of the process
If at the end of the process the temperature is higher than this
temperature, the process will not end and the door will remain closed.
Entry Code–11
Resolution–1ºC
Minimum Value–40ºC
Maximum value–137ºC
Set
value
9.2.7PulsNum. – No. of pulses in the prevacuum stage
Entry Code–11
Resolution–1
Minimum Value–1
Maximum Value–5
Set
value
9.2.8VacDip1 – Vacuum value in the first pulse
Defines the vacuum value in pulse no.1 of the prevacuum stage.
Entry Code–11
Resolution–1kpa
Minimum Value–1kpa
Cycle
Value
Cycle
Value
1234567
120120100858585134
1234567
2121111
Maximum Value–90kpa
Set
value
9.2.9VacDip2 – Vacuum value in the remaining pulses
This value defines the vacuum in the remaining pulses in the
prevacuum stage (not just pulse no. 2)
Entry Code–11
Resolution–1kpa
Minimum Value–1kpa
Maximum Value–90kpa
Set
value
Cycle
Value
Cycle
Value
1234567
12202020202020
1234567
20202020202020
Page 24 of 132 Pages
9.2.10Vac Time 1 – Vacuum Time in the First Pulse
This value defines the time the system will continue to maintain
vacuum after reaching Vac Dip1 for the first pulse.
Entry Code–11
Resolution–30sec
Minimum Value–1sec
Maximum Value–1800sec
Set
value
9.2.11VacTime2 – Waiting time for the remaining pulses
In the remaining pulses starting from pulse no. 2 in the prevacuum
stage, after arriving at the VacDip2, stage there is a time delay,
defined as VacTime2 (in seconds), that the vacuum system continues
working.
Entry Code–11
Resolution–30 seconds
Minimum Value–1 second
Maximum Value–1800 seconds (30 minutes)
Default Value–240 seconds
Set
value
9.2.12Ex Mode – The method for exhausting the steam at the end of the
process
Entry Code–11
Cycle
Value
Cycle
Value
1234567
30303030303030
1234567
30303030303030
Resolution–1
Ex Mode
Fast Ex1
N.A.2
Slow Ex (Waste)3
Slow Ex (Liquids)4
Cooling with compressed air 5
Cooling with water circulation 6
Set
value
Cycle
Value
1234567
1134461
Page 25 of 132 Pages
9.2.13LimitP add – The allowable addition to the maximum pressure in
every process
If the maximum pressure is 225kpa, the system will not heat beyond
this pressure. If LimitP add will be Ø , the maximum pressure will
remain 225kpa. If LimitP add equals 10, the maximum pressure will
be 235kpa.
Entry Code–11
Resolutio–1kpa
Minimum Value–Ø kpa
Maximum Value–30kpa
Set
value
9.2.14Calib Code – Calibration Code
This allows entry to the calibration mode.
It must be set on 107 in order to allow the code to be entered.
Entry Code–11
Resolution–1
Minimum Value–Ø
Maximum Value–255
Set
value
9.2.15Ex Time
This defines the time the exhaust valve should remain open after
reaching a pressure of 127kpa for removing the water in the chamber.
In some autoclaves 15 seconds is enough, and in the large autoclaves
it is possible that there will be a need to extend this to 40 seconds.
Cycle
Value
Cycle
Value
1234567
0000500
1234567
100100100100100100100
Entry Code–11
Resolution–10 sec
Minimum Value–Ø sec
Maximum Value–240 sec
Set
value
Cycle
Model
254045453030303030
3150/7045453030303030
Value
3850/7045453030303030
5050/7560606060304530
Page 26 of 132 Pages
1234567
9.2.16VacPump – Does the autoclave have a vacuum pump?
VacPump
There is no vacuum pump in the autoclave Ø
There is a vacuum pump in the autoclave1
Entry Code–33
Resolution –1
Set
value
9.2.17AutoSize – Size of the Autoclave:
This parameter defines the size of the autoclave by the following
details:
2540Ø
3850,38701
3150,31701
5050,50702
Entry Code–33
Resolution–1
Set
value
Cycle
Value
Value
1234567
0000000
AutoSize
Cycle
Model
25400000001
3150/70
3850/70
1234567
1111111
9.2.18Req Prs+
This defines the required addition to the sterilization pressure in kpa.
For example, for a sterilization temperature of 121ºC the required
pressure is 204kpa. Since the system controls the sterilization process
according to pressure and temperature, if Req Prs+ equals Ø , the
system will maintain the pressure at 204kpa. If the value is at 5kpa,
the system will be maintained at 209kpa, and so on.
Entry Code–33
Resolution–1
Minimum Value–0kpa
Maximum Value –15kpa
Set
value
Cycle
Value
5050752222222
1234567
5311111
Page 27 of 132 Pages
9.2.19NO Drain
This is intended for vertical autoclaves only, and defines if the water
will drain at the end of the process or not.
Water drains at end of process–Ø
Water remains in the chamber–1
Entry Code–11
Resolution–1
Minimum Value–0
Maximum Value–1
Set
value
9.2.20Water Max
The maximum allowable time in seconds for water entery (generally it
will be 180 seconds).
Entry Code–11
Resolution–1 sec.
Minimum Value–1 sec.
Maximum Value–1200 sec.
Set
value
9.2.21Ex Rate
It sets the rate of the exhaust in the slow exhaust. It is possible in
autoclaves that have a top exhaust valve. A combination of natural
exhaust (through the condense valve) and the operation of the top
exhaust valve in Shoot Mode when:
Cycle
Value
Cycle
Value
1234567
0000000
1234567
400400400400400400180
Entry Code–11
Resolution–1 (step)
Minimum Value–0
Maximum Value–5
Ex RateTop Exhaust
0=Closed
1=5% of the time Open
2=20% of the time Open
3=35% of the time Open
4=50% of the time Open
5=70% of the time Open
Set
value
Cycle
Value
1234567
0000000
Page 28 of 132 Pages
Loading...
+ 104 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.