Overall Dimensions Drawing for the 3150 / 3170 ELV..................................................14
Overall Dimensions Drawing for the 3850 / 3870 ELV..................................................15
Overall Dimensions Drawing for the 5050/ 5075 ELV...................................................16
Front Panel Keyboard.........................................................................................................26
Baskets and Containers.......................................................................................................56
Page 3 of 58 Pages
1GENERAL
1.1Incoming Inspection
The autoclave should be unpacked and inspected for mechanical damage
upon receipt. Observe packing method and retain packing materials until
the unit has been inspected. Mechanical inspection involves checking for
signs of physical damage such as: scratched panel surfaces, broken knobs,
etc.
If damage is apparent, contact your dealer or point of purchase, so that
they may notify the manufacturer and file a claim with the appropriate
carrier.
All
Tuttnauer products are carefully inspected prior to shipment and all
reasonable precautions are taken in preparing them for shipment to assure
safe arrival at their destination.
1.2Warranty
We certify that this instrument is guaranteed to be free from defects in
material and workmanship for one year against faulty components and
assembly with the exception of glassware, lamps and heaters.
The warranty does not include and does not replace routine treatment
and preventive maintenance to be performed according to instructions
in paragraph 10.1 (Preventive Maintenance).
Our obligation is limited to replacing the instrument or parts, after our
examination, if within one year after the date of shipment they prove to be
defective. This warranty does not apply to any instrument that has been
subjected to misuse, neglect, accident or improper installation or
application, nor shall it extend to products that have been repaired or
altered outside the factory without prior authorization from us.
The Autoclave should not be used in a manner not described in this
manual!
1.3Warranty Statement
The warranty registration must be completed and returned to our service
departments; within fourteen (14) days of purchase or the warranty will be
void.
Our Technical Service Depts can be reached at:
Tuttnauer Europe b.v., Hoeksteen 11, 4815 PR Breda, Netherlands.
Rudolf Gunz & Co. PTY LTD:
Service Department, 26-34 Dunning Avenue, Ros, 2018, Sydney,
Australia.
Service Department, Locked bag 690, Beaconsfield, NSW 2014,
Australia.
+61-2-99356600 Fax: +61-2-99356650
Page 4 of 58 Pages
Note:
If there is any difficulty with this instrument, and the solution is not
covered in this manual, contact our representative or us first. Do not
attempt to service this instrument yourself. Describe the difficulty as
clearly as possible so we may be able to diagnose the problem and provide
a prompt solution.
If the autoclave is equipped with a printer, send along a copy of the last
printout for our inspection. If replacement parts are needed, stipulate the
model and serial number of the machine.
No products will be accepted for repair without proper authorization from
us. All transportation charges must be paid both ways by the owner. This
warranty will be void if the unit is not purchased from an authorized full
service
Tuttnauer dealer.
1.4Ordering Information
Several items must be specified when ordering the unit from the dealer.
Exact model number (depending on desired chamber size).
Please specify the supply voltage available, 230/380V; 1PH or 3PH.
Temperature scale: Specify Centigrade or Fahrenheit.
Pressure scale: Specify kPa or psi.
1.4.1Options
Printer option
Hanging temperature control probe for use with liquids, to
be placed directly inside a sample solution.
Cooling option - specify cooling method: with or without
compressed air. Air is used to balance the chamber
pressure during the cooling stage.
1.4.2Accessories
Basket accessories: A set of two baskets is available for the unit.
The baskets are made of stainless steel wire and have a handle.
The basket allows the operator to load a large quantity of
materials into the chamber.
The pressure scale, printer option and cooling method can be set
up at any time by a technician.
Stainless steel containers: A stainless steel container, for waste
products, is available. This container has vent holes on its upper
part
Page 5 of 58 Pages
2SAFETY INSTRUCTIONS
The autoclave has unique characteristics. Please read and understand the operation
instructions before first operation of the autoclave. The following issues may
require instructions guidance provided by the manufacturer: how to operate the
autoclave, the door safety mechanism, the dangers involved in circumventing
safety means, how to ensure that the door is closed, and how to select a correct
sterilization program.
Make sure that you know where the main power switch is, where the water cut-off
valve is and where the compressed air disconnection valves (if applicable) is
located.
Autoclave maintenance is crucial for the correct and efficient function of the
device. We enclose a log booklet that includes maintenance recommendations,
with every device.
The weekly spore test is part of the preventive maintenance plan, along with the
annual validation of the sterilization processes that ensures appropriate
temperature dispersion within the chamber.
Never use the autoclave to sterilize corrosive products, such as: acids, bases and
phenols, volatile compounds or solutions such ethanol, methanol or chloroform
nor radioactive substances.
1.Never start using a new autoclave before the safety, licensing and
authorization department has approved it for use.
2.All autoclave users must receive training in proper usage from an
experienced employee. Every new employee must undergo a training
period under an experienced employee.
3.A written procedure must be established for autoclave operation,
including: daily safety tests, seal inspection and door hinge inspection,
smooth action of the closing mechanism, chamber cleaning, prevention of
clogging and preservation from corrosion, what is permitted and what is
prohibited for sterilization and choosing a sterilization program.
4.Liquids may be sterilized only with the “liquids” program. The container
must be covered but not sealed. Sealed bottles may only be sterilized using
a special program. The bottle must be either Pyrex or a Borosilicate glass
bottle.
5.When sterilizing plastic materials, make sure that the item can withstand
sterilization temperature. Plastic that melts in the chamber is liable to
cause a great deal of damage.
6.Individual glass bottles may be placed within an appropriate container that
will be placed in a basket. Never place glass bottles on the floor of the
autoclave. Never fill more than 2/3 of the bottle volume.
7.On closing the autoclave's door, make sure it is properly locked before
activating.
9.Before opening the door, verify that there is no pressure in the chamber
(chamber pressure gauge is located on the autoclave's front panel or door,
depends on model).
10.Open the door slowly to allow steam to escape and wait 5 minutes before
you remove the load. When sterilizing liquids, wait 10 minutes.
Page 6 of 58 Pages
11.Once a month, ensure that the safety valves are functioning, and once
annually a certified tester must conduct pressure chamber safety tests.
12.Once annually, or more frequently, effective tests must be performed, i.e.,
calibration and validation.
13.Examine the condition of assemblies on a regular basis. Make sure there
are no leaks, breaks, blockages, whistles or strange noises.
14.It is required to conduct maintenance operations as instructed.
15.Immediately notify the person in charge of any deviation or risk for the
proper function of the device.
Page 7 of 58 Pages
3TECHNICAL DATA
3.1Introduction
Models 2540, 3150, 3170, 3850, 3870, 5050 and 5075, 5090 ELV are
vertical sterilizers designed especially for sterilization of instruments,
liquids and other materials in hospital laboratories, medical laboratories,
research institutes, food laboratories and pharmaceutical facilities.
A special feature of this sterilizer is the fast cooling operation, provided as
an option for the liquids sterilization cycle.
The sterilizer is fully automatic with a choice of six programs, eliminating
any need for operator intervention during a cycle. The sixth program is an
optional program including a cooling stage (program 7 is for leakage test
only). A computerized control unit enables precise control and monitoring
of physical parameters and clear documentation of the sterilization cycles.
The autoclave is equipped with a safety valve, which blows off at an
overpressure of over 25 psi, which is located on the rear of the autoclave.
The control system provides adequate protection, to ensure the safety of
personnel and reliable operation with a minimum of down time.
On all models (except 1730), a printer is an optional addition to the
autoclave. The printer prints the preset and actual parameters of the cycle
(temperature, time and pressure/vacuum).
The autoclave is equipped with a pressure gauge designed for reference
only. In case that there is any problem with electricity supply during a
sterilization cycle the pressure gauge will be used to verify that there is
pressure in the chamber.
This manual provides information for the following models:
ELV – this is the basic model of the vertical laboratory autoclave.
ELVC – this model includes a cooling system that enables cycles with
fast cooling.
ELVPRC – this model includes only the cooling coil as a preparation
for a cooling system but performs as ELV. A Tuttnauer technician can
upgrade this model to ELVC.
This manual is intended to give the user a general understanding of how
the autoclave works and indicates best ways to operate and maintain it in
order to obtain optimum results and a trouble free operation.
After reading this manual, operating the autoclave should be straight
forward. However, since the autoclave is built using high technology
sensitive components, no attempt should be made by the user or any other
unauthorized person to repair or re-calibrate it.
Only technical personnel having proper qualifications, holding technical
documentation and adequate test instrumentation, are authorized to
undertake repair or service.
Page 8 of 58 Pages
3.2Operating Conditions
The autoclave is intended to work in ‘indoor’ conditions only.
Only autoclavable materials shall be used.
The ambient temperature shall not exceed 40ºC and a relative
humidity up to 85%.
The autoclave shall not be used in a manner not specified in this
manual!
Do not use the autoclave in the presence of dangerous gases.
The packed or unpacked device shall be stored in ‘indoor’ conditions.
Caution
Waste water should be brought into the public net in
accordance with the local rules or requirements i.e. only nonhazardous liquids shall be disposed in public sewage!
3.3Directives and Standards
Every autoclave meets the provisions of the following Directives and is
constructed in compliance with the following Standards:
The electrical net must be protected with a current leakage safety relay.
The electrical network must comply with local rules or regulations.
Page 11 of 58 Pages
3.9Symbol Description
On-Off
Caution! Consult accompanying documents
Caution! Hot surface.
Caution! Hot steam.
Protective earth (Ground)
Page 12 of 58 Pages
OVERALL DIMENSIONS 2540 ELV
Page 13 of 58 Pages
OVERALL DIMENSIONS DRAWING FOR THE 3150 / 3170 ELV
Note:
The dimensions A, B and C are different for the models 3150 and 3170, as
indicated below
TYPEABC
3150
3170
1165780530
1345960710
Page 14 of 58 Pages
OVERALL DIMENSIONS DRAWING FOR THE 3850 / 3870 ELV
Note:
The dimensions A, B and C are different for the models 3850 and 3870, as
indicated below
TYPEABC
3850
3870
1220745500
1400925680
Page 15 of 58 Pages
OVERALL DIMENSIONS DRAWING FOR THE 5050/ 5075 ELV
Note:
The dimensions A, B and C are different for the models 5050 and 5075, as
indicated below
TYPEABC
5050
5075
1440780503
16701010733
Page 16 of 58 Pages
3.10Water Quality
Physical Characteristics and Maximum acceptable
Condensate
–
allowable
Rest of metals except iron,
Colourless clean without
Hardness (Σ Ions of alkaline
3.10.1Water for Steam Generation
The distilled or mineral – free water supplied to the autoclave
should have the physical characteristics and maximum
acceptable level of contaminants indicated in the table below:
contaminants levels in steam for sterlizers
(According to EN 13060:2004).
Element
Silicium oxide. SiO
2
≤0.1 mg/kg
content
Iron≤0.1 mg/kg
Cadmium≤0.005 mg/kg
Lead≤ 0.05 mg/kg
cadmium, lead
≤0.1 mg/kg
Chloride (Cl)≤0.1 mg/kg
Phosphate (P2O5)≤0.1 mg/kg
Conductivity (at 20°C)≤3 μs/cm
pH value (degree of acidity)5 to 7
Appearance
earth)
sediment
≤0.02 mmol/l
Compliance with the above data should be tested in accordance
with acknowledged analytical methods, by an authorized
laboratory.
Attention:
month. The use of water for autoclaves that does not comply
with the table above may have severe impact on the working
life of the sterilizer and can invalidate the manufacturer’s
guarantee.
3.10.2Drain Cooling
The feed water supplied to the drain cooling must meet the
following requirements:
Hardness:0.7 - 0.2 mmol/l.
Water temperature shall not exceed 15°C.
We recommend testing the water quality once a
Page 17 of 58 Pages
Loading...
+ 41 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.