Turner Designs Trilogy User Manual

Page 1
Trilogy Laboratory Fluorometer
User’s Manual
Version 1.2 September 15, 2010 P/N 998-7210
Turner Designs
845 W. Maude Ave. Sunnyvale, CA 94085 Phone: (408) 749-0994 FAX: (408) 749-0998
www.turnerdesigns.com
Page 2
This Page Intentionally Left Blank
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 2 of 38
Page 3
Highlights of the Trilogy Fluorometer
Multifunctional, Compact and Modern Design provides flexibility and precision
Spreadsheet Interface Software provides real time display and logging of data to Excel Spreadsheet
Color Touch-Panel Graphics User Interface provides intuitive ease of use
Sample Chamber accepts snap-in Application Modules for Fluorescence, Absorbance and Turbidity measurements
Trilogy saves time by storing up to 18 named calibrations
Automatically calculates and displays Extracted Chl a concentrations using filtered and solvent volumes
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 3 of 38
Page 4
Table of Contents
CHAPTER DESCRIPTION PAGE
Highlights of the Trilogy Fluorometer 3 I. Getting Started 5
Description 5 Unpacking and Inspection 5 Setup 6 Getting to the Home Screen 6 Precautions 7
II. Fluorometer Operation 8
Touch screen Basics (Fluorescence) 8 Measuring Samples 9
III. Absorbance Operation 11
Touch Screen Basics (Absorbance) 11 Measuring Samples 12
IV. Turbidity Operation 14
Touch Screen Basics (Turbidity ) 15 Measuring Samples 15
V. Calibration Overview 16 V. Calibration Overview 17
A. Why Calibrate 17 B. When to Calibrate 17 C. Trilogy Calibration Options, (Fluorescence and Turbidity) 17 D. Trilogy Calibration Procedures 17 E. “Direct Calibration” Procedure – a) Fluorescence and Turbidity Modules, Single Point and Multi Point Cal 18 F. Extracted Chlorophyll Measurements with the Chl a – Acidification Fluorescent Module 20 G. Calibrating and Displaying Corrected Readings for the Chl a Acidification and Non-Acidification Modules 20 H. “Direct Calibration” Procedure – b) Extraction, Acidification, Single Point and Multi Point 20 J. “Direct Calibration” Procedure – c) Extraction, Non-Acidification, Single Point and Multi Point 23 K. Using the Secondary Standard 23 L. Using the Solid Secondary Standard for in vivo Chlorophyll Applications 23 M. Use of the Solid Secondary Standard for Dye Applications: 24 N. Trilogy Calibration Procedure, (Absorbance Module) 24
VI. Touch Screen Basics 26
Touch Screen 26
VII. Systems Connections 27
Serial Output Functional Check 27
VIII. Spreadsheet Interface Software Installation 28
Excel Spreadsheet Example 28
Appendix A - Troubleshooting 29
Fluorescence Troubleshooting 29 Absorbance Troubleshooting 29 Turbidity Troubleshooting 30 Spreadsheet Interface Software Troubleshooting 30
Appendix B - Warranty and Obtaining Service 32
Warranty 32 Out-of-Warranty Service 32
Appendix C – Specifications 33 Appendix D – Principles of Fluorescence 34
Principles of Fluorescence 34
Appendix E – Recommended Measurement Practices 35
Linear Range and Quenching 35 Temperature Considerations 35
Appendix F - Chlorophyll a Acidification and Non-Acidification calculations 36 Appendix G - In vivo fluorescence measurements using Trilogy 37
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 4 of 38
Page 5
I. Getting Started
Description
The Trilogy Laboratory Fluorometer is a compact, multifunctional laboratory instrument that can be used for making fluorescence, absorbance, and turbidity measurements using the appropriate snap-in Optical Module. A color touch screen with simple menus makes for an intuitive user interface.
Fluorescence Modules are available for discrete sample measurement of various fluorescent materials including chlorophyll (in vivo and extracted), rhodamine, fluorescein, cyanobacteria pigments, ammonium, CDOM, optical brighteners and other fluorescent compounds.
The Absorbance Module accepts interchangeable filter paddles so measuremen ts can be made at different wavelengths in order to identify or place a sample in a particular class of compounds. The standard filter paddle wavelengths/bandwidths are: 560/10; 600/10 and 750/10 nm.
The Turbidity Module uses an Infrared (IR) LED with a wavelength of 850 nm as required for reference method: ISO 7027/DIN EN 27027, “Water Quality – Determination of Turbidity”. Using Infrared allows Turbidity to be measured at wavelengths that are not normally absorbed by organic matter thereby reducing susceptibility to interference by sample color.
When properly calibrated, the Trilogy Fluorometer will read out the actual concentration of the solution. Optical Modules contain the necessary light source and filters for the relevant application
Unpacking and Inspection
Upon receiving the Trilogy, please inspect it carefully and make certain all accessories are present. Refer to the checklist shipped with the instrument for order-specific items.
A typical Trilogy shipment includes:
- Checklist
- Trilogy Laboratory Fluorometer
- CD – Trilogy, includes files for: User’s Manual (pdf); Quick Start Guide (pdf); Spreadsheet Interface Software (pdf)
- Solid Secondary Standard, (not intended for use with UV or Turbidity optical modules)
- Power Supply Kit
- RS-232 Cable
- Quick Start Guide, (hardcopy)
- 10 x 10 mm Methacrylate Cuvettes, (quantity of 10)
- 12 x 35 mm Glass Round Vial, (quantity of 5)
- 12 mm Round Adaptor
- 50-Well Microtube Storage Box
- Optical Module(s) as ordered
Fluorescence Module Absorbance Module Turbidity Module
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 5 of 38
Page 6
Setup
Place the Trilogy Fluorometer on a flat, level surface. Allow at least 6 inches (16 cm) of clearance above the instrument to open and close the lid. Position the instrument so that the touch screen faces you .
Power Supply
Connect the power supply into the power connection of the instrument, see Figure 1, and plug it into a wall outlet. See Specifications for power requirements.
Figure 1.
Getting to the Home Screen
1. With the unit turned off, lift the lid and insert an optical module into position, (see Fig 2). Then press down on the module until you hear it snap into place. Be sure to close the fluorometer lid so that the Trilogy power on step will complete successfully.
2. Turn on the On/Off switch located on the back of the Trilogy, see Figure 1. Verify the display becomes active, and shows the module selection screen, see Figure 3 – Module Selection.
Figure 2 showing Optical Kit installed in preparation for getting to the Home Screen.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 6 of 38
Page 7
Figure 3 Module Selection Screen
3. Select the module inserted, and touch “OK” on the confirming screen.
4. When the Home Screen is displayed (Fig. 4), you are now ready to use Trilogy in its Raw Mode (measurements are relative) or calibrate the Trilogy and the snapped-in Optical Application Module to make quantitative measurements.
Figure 4: Trilogy Home Screen Display for Fluorescence Module Configuration.
Precautions
The Trilogy is intended for indoor use only. Wipe up spills immediately. The Trilogy contains sensitive optical components and precision-aligned mechanical assemblies. Avoid rough handling. Do not leave the lid open for extended periods of time. Turn OFF the Trilogy to change or install Optical Application Modules.
Note: After 20 minutes without activity or user stimulation, the Touch Screen hibernates. Lightly touch the screen to activate.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 7 of 38
Page 8
II. Fluorometer Operation
Fluorescence Optical Module
Installation
1. Power the Trilogy OFF
2. Grasp the handle of the Optical Application Module and align the kit with the sample
compartment.
3. Press do wn firmly to lock the Optical Application Module in place. You should hear or
feel a click indicating the module has snapped into place. (See Figure 5.)
4. Close the lid and power ON the Trilogy. Use the touch screen to identify the type of
Optical Application Module installed.
Figure 5.
Removal
1. Power OFF the Trilogy before removing the Optical Application Module
2. Gra sp the ha ndle and gently pull up to release the kit from the sample compartment.
Touch screen Basics (Fluorescence)
Home Screen
The "Home" screen appears after confirmation of the Optical Application Module. The "Home" screen provides orientation for the multiple functions of the Trilogy. From the "Home" screen, select "Calibrate," "Tools," "Mode," or "Help." The "Home" screen is also the measurement screen. (See over for Figure 6 showing Home Screen with callouts).
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 8 of 38
Page 9
Figure 6 – Trilogy Home Screen for Fluorescence Module Configuration.
Calibration
Refer to Section V.
Measuring Samples
There are two measurement modes available on the Trilogy when using the Fluorescence Module:
Touch “Mode” on the Home Screen to select the measurement mode.
1. Raw Fluorescence Mode: The Raw Fluorescence Mode should be used for qualitative
measurement, relative changes in fluorescence rather than absolute concentration values. Readings are displayed in Relative Fluorescence Units (RFU).
2. Direct Concentration Mode: The Direct Concentration mode makes absolute measurements
based on a calibration; see Section V for the Calibration Procedure. The Trilogy accommodates 10 x 10 mm methacrylate and polystyrene cuvettes (minimum 2 mL
volume). Use 12 x 35 mm or 12 x 75 mm glass test tubes for measuring solvent applications, such as extracted chlorophyll, and use methacrylate cuvettes for UV water applications.
Raw Fluorescence Mode – No calibration required Direct Concentration Mode – Calibration required (see Section V)
1. Open the lid of the Trilogy and insert the cuvette. Close the lid.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 9 of 38
Page 10
2. Touch "Samp le ID" to name your sample (optional).
3. Using the keypad, enter the sample name into the name field and touch "Save" to
save the sample ID. Touch "Measure Fluorescence" to make a measurement. The Trilogy will measure
4.
the sample for 6 seconds and report the average reading for the sample
The Trilogy reports data on the "Home" screen and displays the results for the most recent 20 measurements. Use the arrow keys to scroll through the most recent measurements. The data automatically exports to a printer or PC when properly connected, (see Section VII). Please note the Trilogy does not store more than 20 measurements at one time. If more than 20 readings are taken, the oldest reading will be overwritten. Measurements are not stored between power cycles.
Tools (Fluorescence) Touch the “Tools” key to access "Settings."
Tools - Settings
View Cal Details Touch "View Cal Details" to see information on the current calibration for Direct Concentratio n Mode. "View Cal Details" specifically provides information on the raw fluorescence for each standard and the blank as well as the unit of measure and the Optical Application Module.
Continuous Sampling
The Continuous Sampling feature enables repeat measurements at user-defined intervals.
.
1. Touch "Continuous Sampling" and turn the feature ON. Highlight the frequency of
measurement and the number of total measurements. The maximum number of total measurements is 9999.
2. Touch "OK" to return to the "Home" screen.
3. Connect the Trilogy to a printer or a PC to collect the data. Touching the screen repeatedly causes an early-abort of Continuous Sampling measure m ents.
Measurement Tip:
On the Home Screen, touch Settings then touch the Lid Start key to turn the feature ON. When the Lid
Start feature is ON, measurement begins as soon as the lid closes. The lid start feature allows for immediate measurement and eliminates the need to touch the "Measure" key. Also, the touch screen does not hibernate when Lid Start is ON.
Return to the Lid Start key under the Settings menu to turn the feature OFF.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 10 of 38
Page 11
III. Absorbance Operation
Absorbance Module
Installation
1. Power OFF the Trilogy.
2. Align the Absorbance Module with the sample compartment.
3. Press down to lock the Absorbance Module in place. You should hear or feel a click
indicating the module has snapped into place.
4. Close the lid and power ON the Trilogy. Select "Absorbance" from the list of options
on the touch screen.
5. Install the filter paddle that corresponds to the wavelength of absorbance for the
assay. (See Figure 7.)
Figure 7.
Removal
1. Power OFF the Trilogy before removing the Absorbance Module.
2. Grasp the handle and gently pull up to release it from the sample compartment.
3. Close the lid to protect the light detector from exposure to ambient light. Alternatively,
install an Optical Application Module.
Touch Screen Basics (Absorbance)
Home Screen The "Home" screen appears after confirming the photometer operation by installing the Absorbance Module. From the "Home" screen, select "Calibrate," "Tools," "Mode" or "Help." The "Home" screen is also the measurement screen. (see Figure 8 on next page)
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 11 of 38
Page 12
Mode Touch "Mode" to select the unit of measure for absorbance. The available options include Absorbance units (Ab) and Transmittance (%T). The following formulas describe the method of the Trilogy Absorbance Module for measuring % transmittance and absorbance:
%T = [(s-z) / (b-z)] * 100 Ab = 2 - Log
* (%T)
10
Where, z = zero b = baseline s = signal
Figure 8: Home Screen When Using Absorbance Module
Calibration
See Section V.
Measuring Samples
The Absorbance Module accommodates 10 x 10 mm methacrylate and polystyrene cuvettes as well as glass cuvettes (minimum 2 mL volume).
1. Open the lid and insert the cuvette. Close the lid.
2. Touch "Sample ID" to name your sample (optional). Using the keypad, enter the sample name into the name field and touch "Save" to save the sample ID.
3. Touch "Measure Absorbance" to make a measurement. The Trilogy will measure the sample for 6 seconds and report the average reading for the sam ple.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 12 of 38
Page 13
The Trilogy reports data on the "Home" screen and displays the results for the most recent 20 measurements. Use the arrow keys to scroll through the most recent measurements. The data exports to a printer or PC if properly connected, (see Section VII). Please note the Trilogy does not store more than 20 measurements at one time. If more than 20 readings are taken, the oldest reading will be overwritten. Measurements are not stored betwe en power cycles.
Tools - Settings
Touch the "Tools" key to access "Settings." View Cal Details
Touch "View Cal Details" to see information on the current calibration for the baseline and the zero.
Continuous Sampling
The Continuous Sampling feature enables repeat measurements at user-defined intervals.
1. Touch "Continuous Sampling" and turn the feature ON.
2. Highlight the frequency of measurement and the number of total measurements. The maximum
number of total measurements is 9999.
3. Touch "OK" to return to the "Home" screen.
4. Connect the Trilogy to a printer or a PC to collect the data obtained during Continuous
Sampling. Touching the screen repeatedly causes an early-abort of Continuous Sampling measurements.
Measurement Tip:
On the Home Screen, touch Settings then touch the Lid Start key to turn the feature ON. When the Lid Start feature is ON, measurement begins as soon as the lid closes. The lid start feature allows for immediate measurement and eliminates the need to touch the "Measure" key. Also, the touch screen does not hibernate when Lid Start is ON.
Return to the Lid Start key under the Settings menu to turn the feature OFF.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 13 of 38
Page 14
IV. Turbidity Operation
Turbidity Optical Module
Installation
1. Power the Trilogy OFF
2. Grasp the handle of the Optical Application Module and align the kit with the sample
compartment.
3. Press down firmly to lock the Optical Application Module in place. You will hear or
feel a click indicating the module has snapped into place. (See Figure 9.)
4. Close the lid and power ON the Trilogy. Use the touch screen to identify the type of
Optical Application Module installed.
Figure 9.
Removal
1. Power OFF the Trilogy before removing the Optical Application Module
2. Gra sp the ha ndle and gently pull up to release the kit from the sample compartment.
Touch Screen Basics
See next page
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 14 of 38
Page 15
Touch Screen Basics (Turbidity )
Home Screen
The "Home" screen appears after confirmation of the Optical Application Module. The "Home" screen provides orientation for the multiple functions of the Trilogy. From the "Home" screen, select "Calibrate," "Tools," "Mode," or "Help." The "Home" screen is also the measurement screen.
Figure 10: Home Screen When Using Turbidity Module
Calibration
Refer to Section V.
Measuring Samples
There are two measurement modes available on the Trilogy when using the Turbidity Module:
Touch “Mode” on the Home Screen to select the measurement mode.
1. Raw Mode: The Raw Mode should be used for qualitative measurement, relative changes in
fluorescence rather than absolute concentration values. Readings are displayed in Relative Fluorescence Units (RFU).
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 15 of 38
Raw Mode – No calibration required Direct Concentration Mode – Calibration required – See Section V
Page 16
2. Direct Concentration Mode: The Direct Concentration mode makes absolute measurements
based on a calibration. Readings are displayed in Nephelometric Turbidity Units (NTU). (see Section V for the Calibration Procedure)
Use polystyrene cuvettes for measuring turbidity.
1. Open the lid of the Trilogy and insert the cuvette. Close the lid.
2. Touch "Samp le ID" to name your sample (optional).
Using the keypad, enter the sample name into the name field and touch "Save" to
3.
save the sample ID. Touch "Measure Turbidity" to make a measurement. The Trilogy will measure the
4.
sample for 6 seconds and report the average reading for the sam ple
The Trilogy reports data on the "Home" screen and displays the results for the most recent 20 measurements. Use the arrow keys to scroll through the most recent measurements. The data automatically exports to a printer or PC when properly connected, (see Section VII). Please note the Trilogy does not store more than 20 measurements at one time. If more than 20 readings are taken, the oldest reading will be overwritten. Measurements are not stored between power cycles.
Tools (Turbidity) Touch the “Tools” key to access "Settings."
Tools - Settings
View Cal Details Touch "View Cal Details" to see information on the current calibration for Direct Concentratio n Mode. "View Cal Details" specifically provides information on the raw fluorescence for each standard and the blank as well as the unit of measure and the Optical Application Module.
.
Continuous Sampling
The Continuous Sampling feature enables repeat measurements at user-defined intervals.
1. Touch "Continuous Sampling" and turn the feature ON. Highlight the frequency of
measurement and the number of total measurements. The maximum number of total measurements is 9999.
2. Touch "OK" to return to the "Home" screen.
3. Connect the Trilogy to a printer or a PC to collect the data. Touching the screen repeatedly causes an early-abort of Continuous Sampling measure m ents.
Measurement Tip:
On the Home Screen, touch Settings then touch the Lid Start key to turn the feature ON. When the Lid Start feature is ON, measurement begins as soon as the lid closes. The lid start feature allows for immediate measurement and eliminates the need to touch the "Measure" key. Also, the touch screen does not hibernate when Lid Start is ON.
Return to the Lid Start key under the Settings menu to turn the feature OFF.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 16 of 38
Page 17
V. Calibration Overview
A. Why Calibrate
When calibrated, the Trilogy converts the fluorescence signal to concentration units based o n the standard(s) used for the calibration. This saves you from having to perform any calculations.
B. When to Calibrate
- For greatest accuracy, calibrate or check the calibration using a secondary solid standard before running a new batch of samples.
- Recalibrate if the ambient temperature changes by +/- 10°C.
- Recalibrate after changing Optical Application Modules or if you make measurements
on a new analyte.
- Verify the need to calibrate by reading a stable, known concentration standard immediately after calibration and again every few hours to see if readings have changed significantly. Recalibrate when the accuracy becomes u nacceptable for your study.
C. Trilogy Calibration Options, (Fluorescence and Turbidity)
There are two measurement modes available on the Trilogy when using either the Fluorescence or Turbidity Modules, (see section K for Absorbance Module Calibration Procedure):
1. Raw Fluorescence Mode: The Raw Fluorescence Mode should be used for qualitative measurement, relative changes in fluorescence rather than absolute concentration values. Readings are displayed in Relative Fluorescence Units (RFU).
2. Direct Concentration Calibration: Direct Concentrations can be calibrated by using single or multi-point calibrations. In multi-point calibrations, up to five standards and a blank can be read generating a calibration curve for increased accuracy. The software uses these points to calculate direct concentrations. The Trilogy will display the actual concentration of your samples in units that were chosen during calibration.
Raw Fluorescence Mode – No calibration required Direct Concentration Mode – Calibration required
D. Trilogy Calibration Procedures
1. Raw Fluorescence Mode: A calibration is not necessary to measure fluorescence with the
Trilogy. Simply use the Raw Fluorescence Mode to obtain the fluorescence value of a sample in Relative Fluorescence Units (RFU). Use a standard curve to determine the concentration of the analyte in the samples. The Trilogy does not manipulate data while operating in the Raw Fluorescence Mode. It is not necessary to zero the Trilogy for use in the Raw Fluorescence Mode; however a blank sample should be run to determine background fluorescence. A solid secondary standard may be used to verify instrument stability and function (see section J).
2. Direct Calibration Mode: The Direct Concentration Mode requires a calibration with one blank solution and at least one standard solution. The following procedure applies to all the fluorescence modules with the exception of the Chl Acidification and Non-Acidification modules. (There are separate procedures for these two exceptions). The procedure requires the use of at least one calibration standard, (Fluorescein, Rhodamine WT, etc.). Up to 5 standard solutions can be used for a Multi Point Calibration. Calibrations can be given a name and stored for future use.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 17 of 38
Page 18
E. “Direct Calibration” Procedure – a) Fluorescence and Turbidity
Modules, Single Point and Multi Point Cal
The following procedure applies to Trilogy and the Fluorescence or Turbidity Modules. See page 20 for procedures to calibrate the Chl Acidification and Non-Acidification Modules.
Instructions
a) Turn on the Trilogy
b) Select the module/application to be
calibrated and confirm by touching “OK”. Note: Refer to the module label to confirm
which module is in use.
c) On the home screen, touch "Calibrate" to
begin a calibration sequence
d) Select “Run New Calibration”
e) Select the unit of measurement
f) Insert the calibration “blank” and touch
“OK”
Sample Screen
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 18 of 38
Page 19
g) Enter the concentration for the first
Standard. If using the Turner Designs Chlorophyll a Standard, this will be the concentration data supplied with the Standard. If doing multi point calibrations, be sure to use Standards in order of increasing concentration.
h) Follow the screen prompt indicating that
the standard should be inserted, touch OK
i) After the calibration is complete, either
select “Proceed with Current
Calibration” or select “Enter More Standards”, in which case, return to “g”
above.
j) Name and Save the calibration for future
use (optional).
k) Subsequent readings in the Direct
Concentration mode should reflect the actual concentration of the analyte.
l) Confirm successful completion of the
calibration by measuring the same Standard. The displayed concentration should equal the value used in step “g” above. Optionally, the Solid Secondary Standard could now be adjusted to give the
This completes the “Direct Calibration” Procedure for the Fluorescence and Turbidity Modules.
same reading for future calibrations, see Section K.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 19 of 38
Page 20
F. Extracted Chlorophyll Measurements with the Chl a – Acidification
Fluorescent Module
EPA Method 445.0 is a standard method for measuring extracted chlorophyll a and pheophytin a in marine and fresh water algae by fluorescence. It requires extraction with 90% acetone, measurements before and after acidification and some fairly simple calculations to arrive at the chlorophyll a and pheophytin concentrations (see appendix F). Method 445.0 is detailed and straightforward. (If high concentrations of pure chlorophyll b are present, see the next paragraph).
A known concentration of pure chlorophyll a (as a standard) is required at least the first time you calibrate Trilogy. For greatest accuracy however, we recommend that you periodically (once every few months) use a known concentration of pure chlorophyll a in 90% acetone to recalibrate your instrument. (Liquid Primary and Solid Secondary Chlorophyll a standards are available from Turner Designs).
G. Calibrating and Displaying Corrected Readings for the Chl a
Acidification and Non-Acidification Modules
The Trilogy software can correct displayed chlorophyll readings for filtered and solvent volumes used during the calibration and measurement cycles. Be sure to touch either “Chl A” or “Chl NA” selections in the module selection screen so that the appropriate correction s are applied to measured data. This results in displayed readings that correspond to the actual sample concentration.
H. “Direct Calibration” Procedure – b) Extraction, Acidification, Single
Point and Multi Point
Instructions Sample Screen
a) Turn on the Trilogy
b) Touch “Chl – A” to select the Chlorophyll a
Acidification module and confirm by touching “OK”
c) On the home screen, touch "Calibrate" to
begin a calibration sequence
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 20 of 38
Page 21
d) Select “Run New Calibration”
e) Select the unit of measurement
f) Insert calibration “blank” and touch “OK”
g) Enter the concentration for the first
Standard. If using the Turner Designs Chlorophyll a Standard, this will be the concentration data supplied with the Standard. If doing multi point calibrations, be sure to use Standards in order of increasing concentration.
h) Follow the screen prompt indicating that
the Standard before Acidification (Fb) should be inserted. Insert sample and touch “OK”.
i) Now insert the Standard after Acidification
and touch “OK”. The (Fa) value will be measured and the ratio of the two readings will be displayed as seen in the next step.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 21 of 38
Page 22
j) If the ratio is within the required range,
touch “OK”. The ratio will be stored in the Trilogy for use in the EPA Method 445.0 for measurement of chlorophyll a and pheophytin a in water.
k) After the calibration is complete, either
select “Proceed with Current
Calibration” or select “Enter More Standards”, in which case, return to “g”
above.
l) Name and Save the calibration for future
use (optional).
m) It is recommended that the Solid Standard
is now measured and the displayed value noted to enable a quick calibration check prior to later use.
This completes the “Direct Calibration” Procedure – b) Extraction, Acidification, Single Point and Multi Point
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 22 of 38
Page 23
I. J. “Direct Calibration” Procedure – c) Extraction, Non-Acidification, Single
Point and Multi Point
The Welschmeyer method is a simplified way to measure chlorophyll a without the need for acidification. It accurately measures chlorophyll a even in the presence of chlorophyll b and pheophytin. However, you cannot obtain a pheophytin a measurement with this procedure. Using this method, you extract your samples according to EPA Method 445.0, but skip the acidification step.
You still need to calibrate the instrument the first time using a known concentration of pure chlorophyll a in 90% acetone.
The calibration procedure for the Chlorophyll Non-Acidification follows the same steps as for the “Direct Calibration” mode, see Section E, however the measurement procedure will prompt for the filtered and solvent volumes.
K. Using the Secondary Standard
The following section will describe how to use the Solid Secondary Standard, P/No. 8000-952, with most of the Trilogy fluorescence modules, (Chlorophyll a, Rhodamine WT, etc., but not UV or Turbidity modules). The two main benefits of using the Solid Secondary Standard are:
1) It can be used in place of a primary liquid standard once a correlation between a primary standard and the solid standard has been established.
2) It can be used to check the fluorometer stability, and/or check for loss in sensitivity resulting from instrument/optical module problems.
The Solid Secondary Standard provides a very stable fluorescent signal. It has an adjustment screw so that you can tune the Solid Standard to provide a signal to match a specific sample. It should be noted that each Solid Standard/Fluorometer relationship is unique. This means that a given Solid Standard can not be used to calibrate multiple fluorometers or modules for identical readings of a given solution.
Figure 11. The Adjustable Secondary Standard includes a fluorescent rod that provides an extremely stable signal.
L. Using the Solid Secondary Standard for in vivo Chlorophyll
Applications
1. To establish a correlation between a known chlorophyll concentration and the fluorometer reading, measure a sample containing algae and note the fluorometer readin g.
2. Insert the Solid Standard in the Optical Module, and adjust the Solid Standard to produce the same reading on the fluorometer as in step 1, (turning the Secondary Standard adjustment screw clockwise produces a lower signal).
3. Next, perform a chlorophyll extraction using the Trilogy Laboratory Fluorometer, Spectrophotometer or HPLC to determine the actual chlorophyll a concentration in the sample chlorophyll a concentration.
1
Information on doing a chlorophyll extraction can be found on the Turner Designs web site at this URL:
http://www.turnerdesigns.com/t2/doc/appnotes/998_9000.html
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 23 of 38
1
. This will provide the correlation between the solid standard and the actual
Page 24
4. Now, at any time, the Solid Standard can be used to check/establish a new correlation between a known equivalent concentration and the current Trilogy reading.
M. Use of the Solid Secondary Standard for Dye Applications:
The Solid Secondary Standard accessory can also be used to check the fluorometer’s stability for dye tracing applications. If necessary, the Solid Standard can be used to establish a new correlation factor without the need to use a calibration solution each time.
1. To use the Solid Standard to establish a correlation between a known dye
concentration and the fluorometer reading, measure a dye solution of known concentration, say 50 ppb, and note the Trilogy reading.
2. Place the Solid Standard in the Optical Module, and turn the adjustment screw to
produce the same displayed concentration as in step 1, (turning the secondary standard adjustment screw clockwise reduces the displayed concentration).
Comprehensive information on dye trace measurements can be found at the following Turner Designs URL:
http://www.turnerdesigns.com/t2/doc/appnotes/main.html#fluorescent
N. Trilogy Calibration Procedure, (Absorbance Module)
Calibrate the Trilogy after powering up, and after changing filters. For best results, calibrate the Trilogy and Absorbance Optical Module immediately before reading a series of samples.
Instructions Sample Screen
a) Turn on the Trilogy
b) Touch “Absorbance” to select the
Absorbance Application Module and calibration firmware, and confirm by touching “OK”
c) On the home screen, touch "Calibrate" to
update the blank
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 24 of 38
Page 25
d) Insert a sample cuvette containing your
blank sample and touch “OK” to complete the calibration update.
e) When the blanking is complete, the display
will revert to the Absorbance Home/ Measurement Screen.
f) This completes the Absorbance calibration
procedure, and the Trilogy is now ready to make Absorbance measurements.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 25 of 38
Page 26
VI. Touch Screen Basics
Touch Screen
The touch screen provides a user-friendly method to operate the Trilogy. The touch screen is sensitive to the light pressure of a fingertip. It is not necessary to use a stylus. After 20 minutes without activity or user stimulation, the touch screen hibernates. Lightly touch the screen once to reactivate. To select a function, touch the button corresponding to the function once.
Tools Touch the "Tools" key to access "Settings" and "Diagnostics."
Settings
Contrast Touch the "Contrast" key to increase or decrease the brightness of the touch screen and
enhance visibility. The arrows increase or decrease contrast. Touch the "Home" key to save the adjustment and return to the "Home" screen.
Reset The "Reset" button restarts the Trilogy. Normal operation does not require this feature.
The Reset feature erases data displayed on the "Home" screen. Lid Start Touch the "Lid Start" key to turn the feature ON. While the Lid Start feature is ON,
measurement begins as soon as the lid closes and the touch screen does not hibernate. The Lid Start feature allows for immediate measurement and eliminates the need to touch the "Measure" key. Return to the Lid Start key under the "Settings" menu to turn the feature OFF.
Diagnostics
Touch Screen Calibration The "Diagnostics" menu contains a method for screen calibration. Although the touch
screen is calibrated at the factory, it may need re-calibration over time. Device Configuration The "Device Configuration" key contains useful information on firmware revisions and
instrument setup.
General
Please note the Trilogy does not store more than 20 measurements at one time. Measurements are not stored between power cycles.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 26 of 38
Page 27
VII. Systems Connections
Establish a connection between the Trilogy and a PC or a printer to export data. Connect the 9­pin RS-232 serial cable between the Trilogy and the PC or printer. The male 9-pin connector attaches to the Trilogy and the female connector attaches to the PC.
Serial Output Functional Check
To perform a functional check on the Trilogy Serial Output, complete the above for the Trilogy, and then complete the following steps (note that the Serial Interface Software is not required):
1. Connect the Trilogy Serial Port to the PC Serial Port.
2. To receive ASCII data from the Trilogy, open a HyperTerminal program or some other appropriate data logging software on the PC.
3. Open a connection in HyperTerminal with the following COM Port settings:
Baud Rate: 9600 Data Bits: 8 Parity: None Stop Bits: 1 Flow Control: None
4. Make a measurement with the Trilogy
5. At the completion of the measurement, the concentration value will be displayed on the Trilogy display window, and also displayed in the HyperTerminal window.
This completes the Functional Check of the Trilogy Serial output. If you successfully completed these steps, the Trilogy Serial Output is functioning. If you were not able to complete these steps, refer to the troubleshooting section of this manual.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 27 of 38
Page 28
VIII. Spreadsheet Interface Software Installation
Install the Spreadsheet Interface Software (SIS) to send data to an Excel spreadsheet. The Spreadsheet Interface Software requires a PC loaded with Windows '98 or higher, an available serial port, and Excel. Insert the software CD-ROM into the CD-ROM drive of the PC to initiate the installation program. After the installation is complete, an icon for the SIS appears on the PC desktop and in the "Programs" menu.
(Make sure no other program on the PC is already using COM 1).
Figure12:
Spreadsheet Interface Software status window showing communications are set up on COM 1, and that the data will be displayed in an EXCEL spreadsheet
Excel Spreadsheet Example
Turner Designs SAMPLE-001 426.73 RFU SAMPLE-002 24.57 RFU SAMPLE-003 35.49 RFU SAMPLE-004 2.56 RFU
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 28 of 38
Page 29
Appendix A - Troubleshooting
Fluorescence Troubleshooting
Symptom Possible Solution
Bad calibration error message
Erratic reading
Negative values
Low readings
High background
Absorbance Troubleshooting
Symptom Possible Solution
Non-linear response
Low readings
Bad Calibration Error Message
A bad calibration error message may occur if the blank is brighter than the standard. Compare the reading of the standard and the blank in the Raw mode.
When direct fluorescence readings do not produce expected values, review the standard value entered during the calibration. The number of the standard value should correspond to the actual concentration of the standard.
After calibration, the blank value is automatically subtracted from subsequent readings. A negative reading can occur if a sample reading is less than the blank. This may also indicate that the module is not properly snapped into place. Press down on the module and listen for a click indicating the module has properly been installed.
Check the excitation and emission wavelengths of the analyte against the specifications of the Fluorescence Optical Application Module in use. Different analytes require different Optical Application Modules.
A wet cuvette or spill could contaminate the cuvette holder and increase the background signal. Carefully clean the cuvette holder with 70% ethanol
Many absorbance assays do not produce a linear response but instead produce a sigmoidal or pseudo-sigmoidal response. Refer to the Application Note for the assay for more information.
Check the filter installed in the Absorbance Module and make sure it is the correct filter for the assay. View the Calibration details from the Tools menu
Install the proper filter and use the ultrapure water in a clean cuvette to update the zero. Check the Calibration details from the Tools menu.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 29 of 38
Page 30
Turbidity Troubleshooting
Symptom Possible Solution
Trilogy readings do not agree with other Turbidity meters
Calibrate both meters with the same calibration standard solution. If meters still display significantly different readings, it may be that the second turbidity meter does not make an IR measurement, and the sample contains interference colors.
The turbidity readings change each time a reading
This is normal. Particles in a liquid sample do not remain in the same position, and these position changes affect the scattering of the light, and therefore the turbidity reading.
is taken My turbidity
readings seem to be different when I re­calibrated with a new primary
Formazine standards form the basis of all turbidity measurements and they are very susceptible to aging. ISO 7027 recommendation specifies that the 4,000 NTU Formazine solution can be kept for only 4 weeks. For consistent readings calibrate with current standards.
standard.
Spreadsheet Interface Software Troubleshooting
Symptom
Excel does not open
Excel does not open.
Both green lights are on, but data does not appear in Excel
New data does not report to Excel
Possible Cause Possible Solution
Excel is not installed on the PC.
The software cannot find Excel.
Make sure Excel is installed on your PC.
Open Excel from the Programs Menu on the PC then open the spreadsheet interface software.
Wrong COM port selected.
Click "STOP" then click on the "COM" button to change the COM port.
Fluorometer not connected to PC.
There is an editing process occurring within an Excel cell.
Check the RS-232 connection between the Trilogy and the PC.
Wait until all the data is collected before editing the Excel spreadsheet.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 30 of 38
Page 31
The software does not install
The software does not open
The PC allows only administrators to install new software.
The software was not installed properly.
Log in as Administrator for the PC, then install the software or contact your IT support desk.
Log in as Administrator and Remove the software and re-install.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 31 of 38
Page 32
Appendix B - Warranty and Obtaining Service
Warranty
Turner Designs warrants the Trilogy and accessories to be free from defects in material s and workmanship under normal use and service for a period of 12 months from the date of shipment from Turner Designs with the following restrictions:
• Turner Designs is not responsible for replacing parts damaged by accident or neglect. Your instrument must be installed according to instructions in the User’s Manual. Damage from corrosion is not covered. Damage caused by customer modification of the instrument is not covered.
• This warranty covers only Turner Designs products and is not extended to equipment used with our products. We are not responsible for accidental or consequential damages, except in those states where this limitation is not allowed. This warranty gives you specific legal rights and you may have other rights which vary from state to state.
• Damage incurred in shipping is not covered.
Warranty Service
To obtain service during the warranty period, the owner shall take the following steps:
1. Write, email or call the Turner Designs Technical Support depart ment and describe as precisely as possible the nature of the problem.
Phone: 1 (877) 316-8049 Email:
2. Carry out any adjustments or tests as suggested by the Technical Support Department.
3. If proper performance is not obtained you will be issued a Return Materials Authorization number (RMA) to reference. Package the unit, write the RMA number on the outside of the shipping carton, and ship the instrument, prepaid, to Turner Designs. If the failure is covered under the warranty terms, the instrument will be repaired and returned free of charge, for all customers in the contiguous continental United States.
For customers outside of the contiguous continental United States who purchased equipment from one of our authorized distributors, contact the distributor. If you purchased directly, contact us. We will repair the instrument at no charge. Customer pays for shipping duties and documentation to Turner Designs. Turner Designs pays for return shipment (custom duties, taxes and fees are the responsibility of the customer).
Out-of-Warranty Service
Follow steps for Warranty Service as listed above. If our Technical Support department can assist you by phone or correspondence, we will be glad to, at no charge. Repair service will be billed on a fixed price basis, plus any applicable duties and/or taxes. Shipment to Turner Designs should be prepaid. Your bill will include return shipment freight charges.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 32 of 38
Address for Shipment:
Turner Designs, Inc.
845 W. Maude Ave.
Sunnyvale, CA 94085
Page 33
Appendix C – Specifications
Fluorescence Module Performance
Minimum Detection Limit:
Chlorophyll a, 0.025 ug/L
Rhodamine WT 0.01 ppb Linear Range: 0-300 ug/L Chl a
0-500 ppb RWT Linearity:
0.99R
2
Turbidity Module Performance
Minimum Detection Limit:
0.05 NTU
Linear Range:
0 – 1,000 NTU
Linearity:
0.99R
2
Absorbance Module Performance
Photometric Measuring Range:
0.0 – 4.0 Ab
Photometric Accuracy:
+/- 0.7% Photometric Precision:
<0.5% at 1 Ab
General Specifications
Sample Adaptors
Accommodates 10 x 10 mm square
plastic cuvettes, 12 x 75 mm round
tubes, and 12 x 35 mm round vials.
Readout:
Direct Concentration (ug/L, ppb,
etc.) or Raw Fluorescence (RFU)
Calibration:
One to Five point calibration with up
to 18 calibrations stored
Light Source & Detector:
LED and Photodiode
Blank:
Reads and subtracts blank
Data Output:
100% ASCII format through a 9 pin
RS-232 serial cable at 9600 baud
PC Operating System (optional if connected to PC):
Windows 98® or later
Power:
100 to 240VAC Universal Power
Supply included, Output 12VDC
0.84A Max
Dimensions:
12.92”D x 10.44”W x 8.42”H (32.82 cm D x 26.52 cm W x 21.39 cm H)
Operating Temperature:
60 – 105
o
F (15 - 40 oC)
Weight:
8.1 lbs (3.65 kg)
Humidity:
75% RH maximum
Warranty:
One year
Application Module Filter Wavelengths
Fluorescence Modules
Application
Chl a (Acid) 7200-040 485 nm 685/50 nm
Chl a (Non-Acid) 7200-046 485 nm 685/10 nm
Chl a (in vivo) 7200-043 485 nm 685/50 nm
Phycocyanin 7200-044 600 nm 640 nm
Rhodamine WT &
Phycoerythrin
Optical Brighteners 7200-047 365 nm 445/15 nm
Fluorescein 7200-048 485 nm 540 nm
Crude Oil 7200-06 3 365 nm 410-600 nm
Histamine 7200-049 365 nm 450/80 nm
CDOM & Ammonium 7200-041 365 nm 430 nm
Part
Number
7200-042 550 nm 610 nm
Excitation
Wavelength
Turbidity Module
Application Part Number Wavelength
Turbidity 7200-060 850 nm
Absorbance Module & Filter Kits
Description Part Number
Absorbance Module 7200-050 560/10 nm Filter Paddle 7200-051 600/10 nm Filter Paddle 7200-052 750/10 nm Filter Paddle 7200-053
Contact us for other applications or wavelengths
Specifications subject to change without notice
Included Accessories
10 each 10 x 10 mm Methacrylate Cuvettes 5 each 12 x 35 mm Screw Top Glass Vials 50-Well Storage Box Solid Secondary Standard Power Supply 12 mm Round Vial Adaptor CD with User’s Guide and Spreadsheet
Interface Software
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 33 of 38
Emission
Wavelength
Page 34
Appendix D – Principles of Fluorescence
Principles of Fluorescence
Fluorescence is a physical property of certain atoms and molecules. It is a molecule's ability to absorb light energy at one wavelength, then instantaneously re-emit light energy of another, usually longer, wavelength. Each compound that fluoresces has a characteristic excitation wavelength, (the wavelength of light that it absorbs) and a characteristic emission wavelength, (the wavelength of light that it emits when the molecules relax and return to their ground state). These excitation and emission wavelengths, (or spectra), are often referred to as the compou nd's fluorescence signature. Figure 13 shows the key components of a filter fluorometer.
ompound Wavelengths specific to c
Digital Readout
Light Detector
Lamp
Figure 13
Lamp: The lamp or light emitting diode provides the light energy that "excites" the compound of
interest. The lamp actually provides a broader range of light than that which excites the compound. This broad light range is illustrated by the "many wavelengths of light" shown in Figure 13.
Many wavelengths of light
Wavelengths created by compound, plus stray light
Excitation Filter
Specific wavelengths of light
Emission Filter
Cuvette or Sample Cell
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 34 of 38
Page 35
g
Appendix E – Recommended Measurement Practices
Linear Range and Quenching
The linear range is the concentration range in which the Trilogy output is directly proportional to the concentration of the fluorophore. The linear range begins with the smallest detectable concentration, and spans to an upper limit (concentration) that is dependent up on: the properties of the fluorescent material; the filters used; and the path length.
A non-linear relationship is seen at very high concentrations where the fluorescence signal does not increase at a constant rate in comparison to the change in concentration, see Figure 4 below. At even higher concentrations, the fluorescence signal will decrease even though the sample concentrations are continuing to increase. This effect is known as “signal quenching”.
Linearity may be checked by diluting a sample 1:1, or some other convenient ratio. If the sample is still in the linear range, the reading will decrease in direct proportion to the dilution. If the reading does not decrease in direct proportion to the dilution, or if the reading increases, the sample is beyond the linear range of the fluorophore.
Fluorometer Response Curve
Readin
Fluorometer
Sample
Linear Region
Sample
Quenching Region
Sample Concentration
Figure 14
Graph showing Linear and Quenching Regions of the sample’s response
Temperature Considerations
Fluorescence is temperature sensitive. As the temperature of the sample increases, the fluorescence decreases. For greatest accuracy, record the sample temperature and correct the sensor output for changes in temperature.
For further information on how temperature, light, water quality and the physiological state of the algal cells can all affect the measurement of chlorophyll a, please refer to the application section of Turner Designs’ web site at the following URL:
http://www.turnerdesigns.com/esupport/understanding.html
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 35 of 38
Page 36
Appendix F - Chlorophyll a Acidification and Non­Acidification calculations
When in direct concentration mode the following calculations will be use to calculate corrected chlorophyll a and pheophytin values for the acidification method and the corrected chlorophyll values for the non-acidification method.
Acidification Method
I. Variables stored during calibration phase of fluorometer
C F F F F
II. Variables required from the sample analysis phase
F F V V
III. Interpolation equation used in end calculation of chlorophyll a and pheophytin concentrations
Interp, Interp,
IV. End calculation for corrected chlorophyll and pheophytin a
Chlorophyll a concentration = [F Pheophytin a concentration = [F
Non Acidification Method
I. End calculation for corrected chlorophyll a
Chlorophyll a concentration = C
= Concentration of standard 1
stand[1]
= Fluorescence of Blank value
blank
= Fluorescence of standard 1 before acidification
stand[1],B
= Fluorescence of standard 1 after acidification
stand[1],A
= Acidification Ratio = (F
m
= Fluorescence of sample before acidification
samp,B samp,A solvent water
= C
B A
= Fluorescence of sample after acidification = Volume of solvent used to extract sample = Volume of water filtered
= C
stand[1] stand[1]
* (F
samp,B - Fblank
* (F
samp,A - Fblank
stand[1]
/(Fm-1)] * (Interp,B - Interp,A) * (V
m
/(Fm-1)] * [(Fm * Interp,A) - Interp,B] * (V
m
) / (F ) / (F
* [(F
stand[1],B
stand[1],B - Fblank stand[1],B - Fblank
- F
samp
– F
blank
blank
)/ (F
) / (F
stand[1],A
) )
solvent
stand[1] - Fblank
– F
)] * (V
blank
/ V
)
water
solvent
solvent
)
/ V
/ V
water
water
)
)
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 36 of 38
Page 37
Appendix G - In vivo fluorescence measurements using Trilogy
In vivo chlorophyll a analysis is the measurement of chlorophyll a fluorescence within a living cell. The advantage
of this type of analysis is that it is quick and simple and does not require special sample preparation or extraction. It allows the user to measure a large number of samples in the field. However, without comparisons to extractive analysis, in vivo readings are qualitative in nature. There are two modes for making in vivo measurements. The first utilizes the RAW fluorescence mode; the second requires calibrating to a secondary standard. No calibration is necessary for making in vivo measurements in the raw mode.
In-vivo measurements using the RAW mode:
1) Insert the in vivo chlorophyll a Trilogy module Turner Part Number 7200-043
2) Turn on the Trilogy fluorometer using the switch located on the back panel
3) Select the “Blue” module. This is indicated on the module label under GUI selection.
4) Press “OK” after verifying that the module loaded matches your selection. The default mode loaded is “Raw Fluorescence”.
5) Before measurements are made, it is advisable to filter a sample and measure the filtrate to obtain a bla nk or “Background” fluorescence reading for a given location
6) Thoroughly mix the sample by inverting or shaking to prevent settlement of algal cells and quickly insert sample
7) Close the lid and press the “Measure Fluorescence” button.
8) Subsequent readings will indicate relative changes in concentration levels. These readings will be presented in Relative Fluorescence Units of measure (RFU).
In-vivo Calibration Using Direct Concentration and a Secondary Standard:
1) Insert the in vivo chlorophyll a Trilogy module Turner Part Number 7200-043
2) Turn on the Trilogy fluorometer using the switch located on the back panel
3) Select the “Blue” module. This is indicated on the module label under GUI selection.
4) Press “OK” after verifying that the module loaded matches your selection. The default mode loaded is “Raw Fluorescence”.
5) Press “Calibrate”
6) Choose “Run New Calibration”
7) Choose the units to be display ed during measurement
8) Insert blank sample and press “OK”
9) When prompted enter a number, which will be correlated, to the sample’s response. Y ou may enter 1 to 5 standards of various concentration levels or calibration values.
• If concentration of sample is known, enter the known value. This will set the f luorescence r esponse
equal to the concentration entered and will give semi-quantitative results in subsequent measurements (example, 5(ug/L) = 137.4(RFU))
• If an arbitrary calibration value is entered, the fluorescence response of the sample will be correlated to
the new calibration value and subsequent measurements will be scaled proportionally. These
10) Thoroughly mix sample by inverting or shaking to prevent settlement of algal cells and quickly insert sample
11) After inserting sample immediately press “OK” to measure
12) If running a multi-point calibration repeat step 9 through 11 for each calibrant
Blank Subtracting:
Blanks provide background fluorescence values of samples excluding the fluorophore of interest. Subtracting a blank sample from subsequent samples increases accuracy of fluorophore estimates. An accurate “blank” is typically a water sample that has been filtered through a GF/F or membrane filter in order to remove the algal cells but to still retain additional dissolved components.
Using the Solid Secondary Standard:
1. To establish a correlation between a known chlorophyll a concentration and the fluorometer reading, measure
2. Insert the solid standard in the optical module and adjust the solid standard to produce the same reading on
3. Next perform a chlorophyll a extraction using the Trilogy Laboratory Fluorometer to determine the actual
4. The solid secondary standard can now be used to recalibrate and/or check the stability of the Trilogy
measurements will only give qualitative results (example, 500(RFU) = 137.4(RFU))
a sample containing algae and note the fluorometer reading the fluorometer as in step 1, (Turning the solid standard screw clockwise produces a lower signal) chlorophyll a concentration in the sample. This will provide the correlation between the solid standard and the
actual chlorophyll a concentration fluorometer.
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 37 of 38
Page 38
Turner Designs
845 W. Maude Ave.
Sunnyvale, CA 94085
Phone: (408) 749-0994
FAX: (408) 749-0998
www.turnerdesigns.com
Trilogy Fluorometer User’s Manual P/N 998-7210 Page 38 of 38
Loading...