c Setup
c Setup
Before the Opti-Check can be used, the supplied batteries must be installed.
Before the Opti-Check can be used, the supplied batteries must be installed.
1. On the back side of the instrument, remove the battery panel by loosening the center screw and then gently pulling on the screw.
1. On the back side of the instrument, remove the battery panel by loosening the center screw and then gently pulling on the screw.
2. Install the 4 - AAA batteries and verify the batteries are positioned correctly.
2. Install the 4 - AAA batteries and verify the batteries are positioned correctly.
3. Replace the battery panel and press it down into position.
3. Replace the battery panel and press it down into position.
4. Hold the batter y panel down in place and gently tighten the screw until it is snu g . Do not over tighten the screw.
4. Hold the batter y panel down in place and gently tighten the screw until it is snu g . Do not over tighten the screw.
Note: If the battery panel is very difficult to press down or remove, you can apply a small amount of silicon-based
Note: If the battery panel is very difficult to press down or remove, you can apply a small amount of silicon-based grease to lubricate the o-ring as needed.
grease to lubricate the o-ring as needed.
d
Calibration
d
Calibration
1. To begin turn the Opti-Check on by pressing <ON/OFF>. Wait 5 seconds for the Opti-Check to warm up.
1. To begin turn the Opti-Check on by pressing <ON/OFF>. Wait 5 seconds for the Opti-Check to warm up.
2. Press <A/B> to toggle between different channels and select the channel you want to use. To identify the configuration look at the
2. Press <A/B> to toggle between different channels and select the channel you want to use. To identify the configuration look at the label on the back side.
label on the back side.
3. Assign a Calibration Standard Value. This is the numeric value that you want the standard to read.
3. Assign a Calibration Standard Value. This is the numeric value that you want the standard to read.
4. Press the <STD VAL> button. Use the ↑ and ↓ arrow buttons to set the standard value.
4. Press the <STD VAL> button. Use the ↑ and ↓ arrow buttons to set the standard value.
5. Press the <ENT> or <ESC> button to accept the value and to return to the Home screen.
5. Press the <ENT> or <ESC> button to accept the value and to return to the Home screen.
6. To perform a calibration, press the <CAL> button.
6. To perform a calibration, press the <CAL> button.
7. Press <ENT> to start the calibration.
7. Press <ENT> to start the calibration.
8. Insert your blank sample and press <ENT>. The Opti-Check will average the readi ng for 10 seconds and set the blanking zero
8. Insert your blank sample and press <ENT>. The Opti-Check will average the readi ng for 10 seconds and set the blanking zero point.
point.
9. Insert the standard sample and press <ENT>. The readingis averaged for 10 seconds and the Standard Calibration value is set.
9. Insert the standard sample and press <ENT>. The readingis averaged for 10 seconds and the Standard Calibration value is set.
10. Press <ENT> when the calibration is complete to accept the calibration.
10. Press <ENT> when the calibration is complete to accept the calibration.
11. If you have more than one channel you can press <A/B> to toggle to the other channel and repeat steps 3-10.
11. If you have more than one channel you can press <A/B> to toggle to the other channel and repeat steps 3-10.
12. You are now ready to begin analyzing samples.
12. You are now ready to begin analyzing samples.
e
Sample Analysis
e
Sample Analysis
1. Insert your sample. The Opti-Check is designed with “ambient light rejection”. The black sample compartment cover does NOT
1. Insert your sample. The Opti-Check is designed with “ambient light rejection”. The black sample compartment cover does NOT need to be closed when reading sa mples.
need to be closed when reading sa mples.
2. The orientation and cleanliness of t he cuvettes can have an impact on the accuracy of your results. Refer to Section 5 in the
2. The orientation and cleanliness of t he cuvettes can have an impact on the accuracy of your results. Refer to Section 5 in the Users Manual for details.
Users Manual for details.
3. Press the <READ> button. The instrument will measure and average the fluorescence sig nal for 5 seconds.
3. Press the <READ> button. The instrument will measure and average the fluorescence sig nal for 5 seconds.
4. The reading result will be displayed on the top line of the Home screen.
4. The reading result will be displayed on the top line of the Home screen.
5. The top left corner will then display “ WAIT” for 5 seconds. Once “WAIT” disappears, another sample reading can be performed.
5. The top left corner will then display “ WAIT” for 5 seconds. Once “WAIT” disappears, another sample reading can be performed.
998-0863 Revision B Page 1 of 1