Turku Bioscience Zeiss LSM780 User Manual

Zeiss LSM780
User manual
Cell Imaging & Cytometry
CONTACT:
microscopy@bioscience.fi
+358 (0)20 7032561(office)
050 369 7532 (mobile)
Turku Bioscience 5th floor, room 5077
Contents
1. Information ..................................................................................................... 1
2. Before Imaging ................................................................................................ 2
2.1 Check your sample with an ordinary fluorescence microscope ............................. 2
2.2 Clean your slides ................................................................................................. 2
2.3 Check the environment ....................................................................................... 2
2.4 Start the heating earlier if possible ...................................................................... 2
2.5 Cancellation ....................................................................................................... 2
2.6 Unsure? .............................................................................................................. 2
3. Working with immersion objectives ................................................................. 3
4. Starting up and shutting down the instrument ................................................. 4
4.1 Start up .............................................................................................................. 4
4.1.1 Definite focus ........................................................................................................ 4
4.1.2 Main switches ....................................................................................................... 4
4.1.3 Argon laser ............................................................................................................ 5
4.2 Shut down .......................................................................................................... 5
5. Starting the Zen software ................................................................................ 6
5.1 Start the ZEN software ........................................................................................ 6
6. Using the fluorescence microscope .................................................................. 7
6.1 Visualizing the sample in ocular mode ................................................................. 7
6.2 Using the confocal microscope settings ............................................................... 8
6.2.1 Reusing settings from previous experiments ....................................................... 8
6.2.2 Setting up new experiment .................................................................................. 8
6.2.3 Using Smart Setup ................................................................................................ 8
6.2.4 Manually create light paths .................................................................................. 9
6.2.5 Visualise sample on screen ................................................................................... 9
6.2.6 Adjust the Channels ............................................................................................ 10
6.2.7 Acquire images ................................................................................................... 10
6.3 Experiments ..................................................................................................... 11
6.4 Z-stacks ............................................................................................................ 11
7. Live Cell experiments ..................................................................................... 12
7.1 Live imaging equipment set up .......................................................................... 12
7.2 Setting up time-lapse ........................................................................................ 13
8. Troubleshooting ............................................................................................ 14
8.1 Do not see fluorescence in the Locate mode ...................................................... 14
8.2 Weak signal with the Argon laser ...................................................................... 14
8.3 Argon does not start ......................................................................................... 14
1
1. Information
The Cell Imaging and Cytometry Core is responsible for maintenance of this microscope. Each user must follow the CIC user policy. More information: https://bioscience.fi/services/cell-imaging
Only authorized users may use the CIC instruments.
Users must report any malfunction or problem to the CIC
personnel.
The user is responsible for removing their data from the hard
drive and should do so immediately.
Files older than 30 days are automatically wiped from the
system without prior notice
2
2. Before Imaging
2.1 Check your sample with an ordinary fluorescence microscope
It is a good practice to check the quality of your sample before making use of a more expensive instrument.
2.2 Clean your slides
Clean the remaining salt and mounting medium off the coverslip. Dirty coverslip compromises the image quality. You can use ethanol to aid the cleaning. Do not use the microscope lens tissue. Clean your slides beforehand in your own lab, as it is impractical to use microscopy time for cleaning.
2.3 Check the environment
Switch on the lights and check if the microscope environment is tidy. If there are oil spills or other issues, please inform CIC personnel. (microscopy@bioscience.fi)
2.4 Start the heating earlier if possible
When doing live cell imaging, it is good to switch on the heating at least two hours in advance. If someone is using the instrument, ask whether it is possible to switch it on. If no one is using the instrument, start the instrument and switch on the heating.
2.5 Cancellation
Cancellation must be done 24 hours before the reservation starts. However, if you suddenly cannot use your time, inform the next user and cancel your reservation. If you are the last user of the day and cannot come, you are responsible for the instrument shutdown.
2.6 Unsure?
If you feel that you need support, please contact CIC personnel.
3
3. Working with immersion objectives
The immersion medium should match the objective.
The 40x and 63x objectives need water or oil with a refraction index of 1.334. For
fixed samples and short time experiments, MilliQ water is appropriate
For long experiments water immersion oil is more suitable. The water immersion
bottle is marked with a W; the water-oil is very fluid and must be applied very carefully.
The 100x objective needs oil with a refraction index of 1.518. This oil is more
viscous than the water immersion oil.
The image will be suboptimal when incorrect oil is applied. Do not mix immersion
oils.
A small drop of oil is enough, as too much will make a mess, damaging the
instrument.
Start imaging with the objective in the lowest position. Then, focus the objective
upwards until the oil touches the coverslip. Next, focus the sample visually through the eyepieces.
After imaging, wipe the oil off from the objective softly with a dry lens tissue. Then,
finish the cleaning by wiping softly with a new lens tissue moistened with isopropanol. Only lens tissue may touch the objective lens.
When changing the sample with the immersion objective, the objective must be
lowered (inverted microscope) in between.
Press Load position in the remote unit, in the software (Focus window) or on the
top of the right focusing knob (closest, front most button). Objective moves down.
Remove the sample and add oil if needed Place the new sample. If you are changing the objective in between, wipe the
objective clean with a dry lens tissue. Otherwise, oil might spill into the microscope.
Press Work position in the remote unit, software or on the top of the right focusing
knob (second button). The objective will move up into the position before Load was pressed. It should be very close to the correct focus position.
4
4. Starting up and shutting down the instrument
4.1 Start up
Note: Switch lights on and check that the environment is clean. If you see some issues, report them to CIC personnel.
4.1.1 Definite focus
4.1.2 Main switches
3
2
1. Main on.
2. Systems/PC on. The computer starts.
3. Components on.
If needed, plug in and switch on definite Focus.
If definite focus remains off Zen will give an error on start up.
5
4.1.3 Argon laser
4.2 Shut down
If you are NOT the last user of the day
Clean the immersion objectives you used. Leave the Argon laser to standby mode. Remove your data.
If you ARE the last user of the day, additionally:
Switch off all lasers through the software (incl. Argon). Bring objective down. Close the ZEN software. Shut down the computer. Switch off Systems/PC and Components. Be sure that Argon cooling has finished (you will hear fan switch off). Switch off Main.
Turn the key on the Lasos control unit from 0 to 1.
Switch the Argon stand-by switch from Idle to Run.
The warm-up process will take several minutes. The green LED will lit when the laser is ready to use.
Note: the Argon laser will emit light also when in Idle but the light intensity is much weaker.
Do not touch the light control knob (the laser current selector). However, if the red LED is on, turn the knob counter­clockwise until the red light goes off.
6
5. Starting the Zen software
5.1 Start the ZEN software
Click the Start System button
Choose Acquisition
Switch on the lasers
Start-up will take a few minutes; a loading bar will be visible.
If there are any errors, contact CIC personnel
Switch on the lasers you need. The Diode 405 is always on.
7
6. Using the fluorescence microscope
6.1 Visualizing the sample in ocular mode
! NOTE: the first user must manually open the RL shutter on the Zeiss touchpad!
1 Choose the Locate tab to view sample using the oculars.
2 Transmitted light turn on bulb and open the shutter.
3 Objective lenses:
10x/0.45 dry 20x/0.80 dry 40x/1.2 Water 63x/1.2 Water 100x1.4 Oil
4 RL Reflected light shutter (fluorescence)
5 Fluorescence filters:
DAPI
GFP
• RFP
Analyser filter
(Polarization imaging)
NoneLSM (Confocal imaging)
3 4 5
8
6.2 Using the confocal microscope settings
6.2.1 Reusing settings from previous experiments
6.2.2 Setting up new experiment
6.2.3 Using Smart Setup
If you have earlier acquired images with desired light paths, open the image and use the Reuse button.
To create new light paths, you can Use the Smart Setup or create them manually.
You can save a new imaging configuration you created or open an existing configuration
The Smart Setup function allows you to insert your fluorophores and the software will provide light path suggestions.
Choose your fluorophore from the dropdown box
Fastest scans your labels with a single scan. This gives the most spectral cross talk/ bleed­through and should be avoided.
Best signal scans the light paths separately and is the most suitable option for everyday use.
Smartest (Line) reduces the number of scans and groups together light paths, which would produce least cross talk.
9
6.2.4 Manually create light paths
6.2.5 Visualise sample on screen
1
2
3
5
1 Click + or trash button to add or delete track. 2 Ch1 is a standard photomultiplier tube
(PMT). ChS1 is a very sensitive detector, for very weak/sensitive samples. Ch2 is a cooled PMT, which is more (red-) sensitive than Ch1.
Once you select your fluorophore, an emission spectrum appears. You can adjust the slider, which will change the range of the emission
filter. 3 Visible light includes Argon and both HeNe
lasers.
Invisible light contains the 405nm diode. 4 You must select the correct mirror to reflect
the laser line toward the sample (arrow) 5 T-PMT is used for collected transmitted light
from lasers to create a wide field image.
The buttons mentioned here are used to visualise your sample on screen. One you have selected 1 track (like in the adjust channels image) you can start optimizing the visualization of your sample.
AF Find Focus and Set Exposure are automatic methods. They take extra time and are not necessarily optimal compared to manually doing so.
Live continuously scans the sample in a fast (maximum) speed.
Continuous scans continuously like Live, but with a final (Snap) image quality.
Snap takes a snapshot.
10
6.2.6 Adjust the Channels
6.2.7 Acquire images
1 Expand all shows all available tracks. 2 Laser power and activated lasers can be
adjusted. Adjust pinhole to equalize the optical
Section thicknesses in each channel. Do not equalize the Airy units.
Gain (Master) defines the PMT gain (brightness). Too much gain produces noise in the image.
Digital Offset adjusts the background intensity level. When using the range indicator, zero pixels are blue. Increasing the Digital Offset brings up the background and blue pixels disappear.
Digital Gain increases the PMT gain but can also create more noise.
1 Objective can be chosen both here and in the Locate window.
2 Click Optimal for the maximum resolution for current objective and zoom level.
Increasing the Line Step value increases scanning speed as fewer lines are scanned.
Decreasing the Speed and increasing the Averaging increases signal-to-noise ratio with the cost of increased photo bleaching and acquisition time.
Unidirectional scanning has more accuracy but a slower scan speed than bidirectional.
Scan Area defines the zoom level and angle.
3
11
6.3 Experiments
6.4 Z-stacks
Experiments are for a protocol more advanced than a single z slice.
They are activated by ticking desired boxes through the corresponding window.
Once parameters are selected they will open and new window, such as for z-stack.
Execute experiments through Start Experiment button
During Live scan, mark the lowest and highest desired focus levels with Set First and Set Last.
Selecting Optimal will calculate the best interval between slices based on objective and lasers selected.
Alternatively, interval between Z-slices or number of Z-slices can be manually adjusted.
NOTE! Keeping interval spacing will alter the Z-range. Keeping slices will keep Z-range and alter intervals between Z-slices.
Execute the procedure with the Start Experiment button
12
7. Live Cell experiments
7.1 Live imaging equipment set up
Check that there is distilled water in the humidifier bottle.
Place the heating insert onto the stage. Place your sample into the insert. Place the CO2 incubator lid onto the heating insert. Make sure the air holes on the lid
are facing the sample.
Switch on/off the incubator and CO2 through the software. Typically, 37°C and 5% CO2 are used for normal cell imaging.
13
7.2 Setting up time-lapse
Setting up a time-lapse mean you can allow the software to run the experiment without the need to be at the microscope. You optimized your imaging set up and then you tell the software how often to capture and how long to wait between captures.
Time Series is activated in the Experiments (page 11)
Tick Show all to reveal more options Cycles specifies the number of times you
want to run your image capture set up Interval specifies the gap of time
between each captured image The combination of cycles and interval
will determine how long the experiment runs.
14
8. Troubleshooting
8.1 Do not see fluorescence in the Locate mode
Check that that the HBO 100 lamp unit is on and manually open the FL shutter in the touch-screen control unit.
8.2 Weak signal with the Argon laser
Check that the switch is in Run mode, not Idle.
8.3 Argon does not start
The LASOS cooling unit is perhaps switched off.
Loading...