Pressure Transmitter
PT004R-13-LI3-H1131
Type designation PT004R-13-LI3-H1131
Ident-No. 6831612
Protection type and class IP67
■ Compact and robust design
■ Pressure connection with integrated
peak pressure aperture
■ Accuracy hardly affected by tempera-
ture within entire range -40…+125°C
■ Excellent EMC properties
Functional principle
The pressure transmitters of the PT…-1 series operate with piezo-resistive ceramic measuring cells. The ceramic diaphragm is unbalanced in proportion to the pressure applied.
Depending on the sensor type used, the processed signal is provided as an analog output
signal 4…20mA (2-wire) or 0…10V (3-wire).
Vibration resistance 20 x g (9…2000 Hz, 2…9 Hz with an amplitude
+/-15 mm), acc. to IEC 68-2-6
Shock resistance 100 g (11 ms) , acc. to IEC 68-2-27
Housing
Housing material Stainless-steel/Plastic, V2A (1.4305)
Pressure connection material Stainless steel A2 1.4305 (AISI 303)
Pressure transducer material Ceramics AI2O
Sealing material FPM
Process connection NPT ¼”-18 male thread
Wrench size pressure connection / coupling nut SW 21
Electrical connection Connector, M12 × 1
Max. tightening torque housing nut 20 Nm
MTTF 114 years acc. to SN 29500 (Ed. 99) 20 °C
3
Edition • 2018-10-10T15:54:28+02:00
1 / 1 TURCK Inc.