Turck PS100R-501-2UPN8X-H1141 Data Sheet

Pressure transmitter (rotatable) 2 PNP/NPN transistor switching outputs PS100R-501-2UPN8X-H1141
Type designation PS100R-501-2UPN8X-H1141
Ident-No. 6832632
Pressure range
Power supply
Operating voltage 18…30 VDC Current consumption Voltage drop at I Protective measure SELV; PELV according to EN 50178 Short-circuit/reverse polarity protection yes / yes Protection type and class IP67 / IP69K / III
e
ð 420 bar ï 420 bar
ð 50 mA ð 2 V
Housing is rotatable after plugging the
process connection
Reading of adjusted values without tool
Recessed pushbutton and keylock for
secure programming
Permanent indication of pressure (bar,
psi, kPa, MPa, misc)
Peak pressure memory
Pressure range 0…100 bar rel.
Wiring Diagram
Functional principle
The pressure sensors of the PS series oper­ate with piezo-resistive ceramic measuring cells. The ceramic diaphragm is unbalanced in proportion to the pressure applied. Depend­ing on the sensor type, the voltage produced is made available either as switching or ana­log output signal. Non-rotatable and rotatable sensors, numerous thread types, front-flush or dead-zone free diaphragms and an accuracy of 0.5% of full scale guarantee highest flexibili­ty and safe process interfacing.
Outputs
Output 1 Switching output or IO-Link mode Output 2 switching output
Switching output
Output function NO/NC, PNP/NPN Accuracy switching output ± 0.5 % v. E. BSL Rated operational current 0.2 A Switching frequency Switching point distance Switch point: (min + 0.005 x range) up to 100% f.s. Release point(s) min up to (SP - 0.005 x range) Switching cycles
IO-Link
IO-Link Specification Specified acc. to version 1.0 Programming FDT / DTM Transmission physics corresponds to 3-wire physics (PHY2) Transmission rate COM 2 / 38.4 kbps Process data width 16 bit Measured value information 14 bit Switchpoint information 2 bit Frame type 2.2 Genauigkeit ± 0.5 % of final value BSL
Temperature behaviour
Medium temperature -40…+85 °C Temperature coefficient zero point Tk0 ± 0.15 % of full scale/10 K Temperature coefficient span T
kS
ð 180 Hz ï 0.5 %
ï 100 mil.
± 0.15 % of full scale/10 K
Edition • 2018-10-17T17:22:26+02:00
1 / 3 TURCK Inc.
Pressure transmitter (rotatable) 2 PNP/NPN transistor switching outputs PS100R-501-2UPN8X-H1141
Ambient conditions
Ambient temperature -40…+80 °C Storage temperature -40…+80 °C Vibration resistance 20 g (9…2000 Hz), according to IEC 68-2-6 Shock resistance 50 g (11 ms) , acc. to IEC 68-2-27 EMC EN 61000-4-2 ESD:4 kV CD / 8 kV AD
EN 61000-4-3 HF radiated: 15 V/m EN 61000-4-4 Burst: 2 kV EN 61000-4-5 Surge: 1000 V, 42 Ohm EN 61000-4-6 HF cable bound: 10 V
Housing
Housing material Stainless-steel/Plastic, V2A (1.4305) Pressure connection material Stainless steel A2 1.4305 (AISI 303) Pressure transducer material Ceramics AI2O Sealing material FPM spez. Process connection G¼” female thread Wrench size pressure connection / coupling nut 21/ 30 Electrical connection Connector, M12 × 1 Max. tightening torque housing nut 35 Nm
Reference conditions acc. to IEC 61298-1
Temperature 15…+25 °C Atmospheric pressure 860…1060 hPa abs. Humidity 45…75 % rel. Auxiliary power 24 VDC
3
Display 4-digit 7-segment display, rotatable by 180°, disen-
gageable Switching state 2 × LEDs,Yellow Programming options switch/release point, PNP/NPN; NO/NC; hystere-
sis/window mode, muting; pressure unit, peak pres-
sure memory Unit display 5 x LEDs green (bar, psi, kPa, MPa, misc)
MTTF 439 years acc. to SN 29500 (Ed. 99) 40 °C
Edition • 2018-10-17T17:22:26+02:00
2 / 3 TURCK Inc.
Loading...