TTI 1906 DATASHEETS

Page 1
THURLBY THANDAR INSTRUMENTS
1906
5½ digit high performance multimeter
Full autoranging, True RMS, RS-232 & GPIB interfaces
Page 2
5½ digitperformance at a 4½ digit price
higher resolution, accuracy & functionality
Greater resolution and accuracy
The 1906 is a true 51/2digit meter with a scale length of This gives it not just superior resolution
relative to 4 higher effective accuracy as well.
Thediagram showshow theeffective ac curacy varies with reading after taking into account the fixed errors. Over a range of voltages, the 1906 is around four times more accurate than a typical
1
/2digit meter.
4
± 210,000 counts.
1
/2digit meters, but much
Greater sensitivity and higher input impedance
With a maximum resolution of 1µV, 1m and 1nA, the1906 is ten times more sensi tive than a 4
This extra sensitivity enables accurate measurements to be made in areas previ ously impossible such as thermocouple junctions, switch contact resistance or ca pacitor leakages.
Industry standard 10Minput impedance appliestoall dcvoltageranges, butthelow est two can alternatively be selected at greater than 1000Mimpedance to elimi nate errors when measuring high imped ance circuitry.
1
/2digit meter.
True RMS ac functions
The 1906 provides True RMS ac response
-
whichgives accurate measurementsregard less of the waveform shape.
The wide bandwidth voltage attenuator provides high accuracy within the audio band and gives extended response to avoid distortion when measuring switching waveforms.
Auto or manual ranging
The 1906 offers fully automatic ranging on all functions including current. Alter natively any range can be selected and
-
held manually.
4 terminal Ohms for
-
improved stability
-
Resistance measurements can bemade in either 2 terminal or 4 terminal mode. High impedance sensing ensures that the meter switches automatically be
­tween modes.
-
For low resistance measurements, 4 ter
-
minal modeoffers much higher accuracy and repeatability.
Theresistance functionalso providesse lectable audible continuity as well as di ode test capabilities.
-
Closed case calibration
The 1906 stores its calibration constants in permanentmemory (EEPROM). Con­sequently it can be recalibrated without ever opening the case.
-
-
-
-
-
5½digit scale length ( 210,000 counts)
0.012% basic one year accuracy
1uV, 1m and 1nA resolution, 4 terminal Ohms
Full automatic or manual ranging
True RMS ac functions, wide ac bandwidth
Full set of current ranges from 200µA to 10A
Wide range of computing and data logger functions
Fully programmable by RS-232 or GPIB interfaces
Storage of front-panel set-ups
The 1906 can store up to six front panel set-ups in non-volatile memory.
In addition to the range and function each set-up defines the active comput ing or logging functions along with their parameters.
This facility is particularly valuable when similartests needto berepeated atinter vals.
-
-
Page 3
built-in computing & data logging
full bus programmability
Single key convenience functions for ease of use
Press the “null” key and the present dis played reading is stored and subtracted from all future readings.
This powerful feature has a host ofuses such as removing test lead resistance, observingdeviation levels, ormeasuring relative to alternative voltage points us ing a fixed ground lead connection.
Touch and Hold
Thetouch andholdfunction isinvaluable when taking measurements in awkward situations whereit may be difficult tosee the multimeter display.
Withtouchand holdselected,the display is automatically frozen every time a sta ble reading is achieved. A further press of the T/HOLD key un-freezes the dis play.
Digital filtering for better noise performance
The 1906 uses non-linear digital filtering which is highly effective in eliminating unwanted noise, but responds rapidly to a change in the reading.
When working with very noisy or unsta­blemeasurements, thecharacteristicsof thedigitalfilter canbeadjusted tosuitthe application.
Full protection for the instrument and the user
The 1906 incorporates extensive protec tion against damage from accidental overloads on all ranges including 10 Amps.
In additionthe 1906 meets the stringent safety requirements of IEC348 and IEC1010-1 for measurements up to 1kV.
Linear scaling with offset (Ax+b)
-
This function enables a reading to be multipliedby a scalefactor (A) andfor an offset (b) to be added or subtracted.
Thismeans, forexample, thatthe electri cal output of a transducer can be scaled to give a direct readout of the physical
-
parameter e.g. temperature or weight. It is particularly useful with 4-20mA cur
rent loops.
Limits comparison
This function enables the reading to be tested against high and low limits set by the user.
-
­The display then shows a code of H
(high), L (low) or P (pass) at the end of the display.
Percentage deviation ( %)
This function enables the meter to dis­playthepercentage amountbywhich the readingdiffers froma nominalvalue cho­sen by the user.
It is invaluable for measuring tolerance or stability.
Min-Max storage
The Min-Max function stores the highest and lowest values of a set of readings.
This facility has many uses such as checking for power supply glitches, re
-
cording peak temperature excursions and avoiding “missed” readings when using the Logger function.
Logarithmic measurements (dB)
Thisfunctionallows measurementstobe displayed in the form 20log zero reference scaling enables voltages to bedisplayed in dBV ordBm relative to any required impedance.
X. Built-in
10
Automatic data logging
The1906 canstore upto 100readingsat any requiredtime interval from1 reading per second up to 1 reading every 3 hours.
A simple recall sequence allows the
­readings to be scrolled onto the display
whenever required. Alternatively results can be downloaded using the RS232 or GPIB interfaces.
­As a result, tedious time related meas
urement sequences can be handled automatically without disrupting your work.
Manual storageof readings is also avail able eliminating the need for paper and pencilwhen makingaseries ofmeasure ments.
Storage can also be triggered from the RS-232 or GPIB interfaces.
Full bus control via RS-232 or GPIB
The 1906 is designed for complete bus control. Every function of the instrument canbecontrolled usingeither theRS-232 interfaceorthe GPIBinterface (optional).
The RS-232 interface, which is fitted as standard, iscompatible with theTTi ARC system.
The ARC system (Addressable RS232 Chain) enables up to 32 instruments to
-
be “daisy-chained” togetherand to be in dividually addressed and controlled us ing a single RS232 port of a PC.
Compatible with IEEE-488.2
The GPIB interface is an option. When installed, a rear panel switch allows the user to choose between RS232 or GPIB control.
The GPIB interface conforms fully with bothIEEE-488.1and IEEE-488.2. The.2 standard contains many enhancements which are particularly important when programming systems incorporating a number of instruments.
-
-
-
-
-
Using the null key allows gains and losses to be measured directly.
Page 4
Technical Specifications
INPUT CHARACTERISTICS
Input Current: < 100 pA. DC NMR: > 60dB at 50/60Hz. 1k Unbalanced
CMR:
ACCURA CY
Accuracies apply for 1 year 18oC to 26oC. Temperature coefficient outside these limits is <01 x quoted range accuracy per accuracy is 1 hour.
DC VOLTS
Range
200mV
2V
20V
200V
1000V
RESISTANCE (OHMS)
Range
200
2k
20k
200k
2M
20M
Max open circuit voltage 35V. Diode Test on 2k range. Audible continuity check available on all ranges.
AC VOLTS (True RMS)
Range
200mV
2V
20V 200V 750V
Additional error at crest factor=3 typically 07%. Input impedance 1MΩ / <100 pF any range. Max input 750V rms, 1kV peak, any range. Accuracies apply from 10% to 100% of full scale.
DC AMPS
Range
200µA0⋅08% + 12d
2mA
20mA
200mA
10A <2000mA
> 120dB at DC/50Hz/60Hz, DC ranges; > 60dB at DC/50Hz/60Hz, AC ranges.
o
C. Warm-up time to rated
Accuracy(±)
0017% + 3d 0012% + 3d 0019% + 3d 0019% + 3d 0019% + 3d
Accuracy(±)
0025% + 4d 1m 15mA 0019% + 3d 10m 15mA 0019% + 3d 100m 04mA 0019% + 3d 1 40µA
Res. Input Impedance Max Input
1µV
10µV
100µV
10Mor >1G
(Hi Z mode)
10M
1kV DC or
AC Peak
any range
1mV
10mV
Resol-
ution
Max. Measur
ing Current
Max. Input
-
300V DC or
RMS continu
ous any
Ohms input
any range
0022% + 3d 10 4µA
007% + 3d 100
Accuracy(±)
400nA
45Hz - 5kHz 5kHz - 10kHz 10kHz-20kHz 20kHz-50kHz 02% + 150d 02% + 250d 02% + 150d 02% + 250d 05% + 300d 02% + 150d 02% + 250d 05% + 300d 02% + 150d 02% + 250d 05% + 500d 02% + 150d
---
Accuracy(±)
008% + 12d 008% + 12d 008% + 12d 008% + 12d
>2000mA
012% + 12d
3% + 500d -
3% + 1500d 2% + 1000d
Resol
Voltage
ution
Burden
1nA 300 mV
10nA
max
100nA
1µA 100µA 650mV 100µA
max
-
Max. Input
1A, 300V
fuse pro
tected
10A, 300V
fused
AC AMPS (True RMS)
Range
Accuracy(±) 45 Hz - 1 kHz
200µA0⋅37% + 100d
2mA
20mA
200mA
10A <2000mA
>2000mA
Additional error at crest factor=3 typically 07%. Accuracies apply from 10% to 100% of full scale.
037% + 100d 037% + 100d 037% + 100d 037% + 100d
04% + 100d
Voltage Burden
300mV
max
650mV
max
DISPLA Y
Display Type: 13mm LED, 8 digit. Scale Length: Selectable 5½ digit or 4½ digit. Reading Rate: 3 readings/sec (5½ digit).
Overrange: Overflow: Annunciators: LEDs for range, function and program modes.
-
COMPUTING FUNCTIONS
Null: Operates over full range; values can be stored for Digital Filter: 10 options, including optimised default values for Touch & Hold: Reading is held when stable.
% DEV: Displays % deviation from entered reference value. dB: Displays measurement in dB relative to 1V, 1mA,
Ax+B: Linear scaling of results, with offset. Limits:
Min/Max: Minimum and maximum reading stored. Data Logger: Manual or automatic storage of 100 measurements,
10 readings/sec (4½ digit). Displays OL if input too great for range. Displays OFLO if calculated result overflows display.
every function. each range and function.
1kor user entered value, or in dBm. Reading displayed with H, L, or P (pass) with respect
to user-defined high and low limits.
Storage interval 1s to 9999s.
INTERF A CES
RS232: Baud rates 300, 1200 or 9600. Complies fully with
GPIB (Optional): Fully complies with IEEE-488.1 and IEEE-488.2
the ARC (Addressable RS232 Chain) interface stan dard. Address selectable from the front panel.
GENERAL
AC Input: 220 to 240 volts ±10% or 110 to 120 volts ±10% Consumption: 25VA max.
Operating Range: +5 Storage Range: -40
-
Environmental: Indoor use at <2000m, Pollution Degree 1. Safety: Complies with EN61010-1. EMC: Complies with EN61326. Size: 260(W) x 88(H) x 235(D)mm, excl. handle and feet Weight:
Thurlby Thandar Instruments Ltd. operates a policy of continuous devel opment and reserves the right to alter specifications without prior notice.
50/60Hz, Installation Category II.
o
C to +40oC, 20% to 80% RH
o
C to +60oC
22kg.
Max Input
1A, 300V
fuse pro
tected
10A, 300V
fuse pro
tected
-
-
-
-