This programmable true RMS multimeter has dual measurement capability and a dual display †
which can show either two independent measurements, a measurement together with its range or
a measurement with one of the many calculated functions available.
The key features are:
• Large, high contrast, liquid crystal dual display.
• 0.04% basic accuracy, 12000 counts.
• Manual or autoranging.
• DC and AC Volts, DC and AC current, Resistance, Capacitance, and Frequency measurement;
Continuity and Diode checks.
• True RMS AC and AC+DC measurement.
• Display nulling and Ohms null.
• Touch hold mode - holds onto a stable reading until updated.
• 100 point data logger.
• Measurement post-processing to give:
dB and power measurement with settable reference impedance
percentage deviation from a user-entered reference
linear scaling with offset
limits comparison for go/no go testing
automatic storage of minimum and maximum readings.
• Remote control via addressable RS232 (standard) or GPIB (factory-fit option) interfaces.
• Closed case software calibration.
• Fully compliant with EN61010-1 Safety and EN61326 EMC standards.
We Thurlby Thandar Instruments Ltd
Glebe Road
Huntingdon
Cambridgeshire PE29 7DR
England
declare that the
Model 1705 True RMS Programmable Multimeter and
Model 1705GP True RMS Programmable Multimeter with GPIB
meet the intent of the EMC Directive 89/336/EEC and the Low Voltage Directive 73/23/EEC.
Compliance was demonstrated by conformance to the following specifications which have been
listed in the Official Journal of the European Communities.
EMC
Emissions: a) EN61326 (1998) Radiated, Class B
b) EN61326 (1998) Conducted, Class B
c) EN61326 (1998) Harmonics, referring to EN61000-3-2 (2000)
Immunity: EN61326 (1998) Immunity Table 1, Performance B, referring to:
a) EN61000-4-2 (1995) Electrostatic Discharge
b) EN61000-4-3 (1997) Electromagnetic Field
c) EN61000-4-11 (1994) Voltage Interrupt
d) EN61000-4-4 (1995) Fast Transient
e) EN61000-4-5 (1995) Surge
f) EN61000-4-6 (1996) Conducted RF
Safety
Multimeter: EN61010-1
Installation Category I measurements to 1000V,
Installation Category II measurements to 600V,
Pollution Degree 1.
Probes: IEC1010-2-031
Rated to 1000V, Installation Category III.
This multimeter is a Safety Class I instrument according to IEC classification and has been
designed to meet the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use). It is an Installation Category II instrument intended for
operation from a normal single phase supply.
WARNING! THIS INSTRUMENT MUST BE EARTHED WHEN OPERATED FROM A MAINS SUPPLY
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
When operated from the internal batteries the multimeter meets the safety requirements for a Safety
Class II (reinforced insulation) product and does not require grounding for safety.
The test leads supplied with this instrument meet the requirements of IEC1010-2-031 and are rated
to 1000V Cat III; use only these test leads with the meter or a set of equivalent performance.
This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe
condition. This instruction manual contains some information and warnings which have to be
followed by the user to ensure safe operation and to retain the instrument in a safe condition.
This instrument has been designed for indoor use in a Pollution Degree 1 environment (no pollution,
or only dry non-conductive pollution) in the temperature range 5°C to 40°C, 20% - 80% RH (non-
condensing). It may occasionally be subjected to temperatures between +5° and −10°C without
degradation of its safety. It has been designed for Installation Category II measurement use to
600VDC/ACrms. and Installation Category I measurement use to 1000VDC/750VACrms. The full
definitions of Categories I and II can be found in IEC664, but the following can be taken as a guide:
Installation Category I is signal level e.g. telecommunication, electronic equipment, with smaller
transient overvoltages than Installation Category II.
Installation Category II is local domestic supply level, e.g. portable equipment and appliances. In
particular, Category II does not include distribution level supplies, e.g. three phase installations
which are classified as Installation Category III.
For this equipment 2500V is the maximum peak transient overvoltage that can be tolerated by any
terminal with respect to earth ground without impairing safety.
Use of this instrument in a manner not specified by these instructions may impair the safety
protection provided. Do not operate the instrument outside its rated supply voltages or
environmental range. In particular excessive moisture may impair safety.
When the instrument is connected to its supply or its inputs are connected to live voltages, terminals
may be live and opening the covers or removal of parts (except those to which access can be
gained by hand) is likely to expose live parts. The apparatus shall be disconnected from all voltage
sources before it is opened for any adjustment, replacement, maintenance or repair. Any
adjustment, maintenance and repair of the opened instrument under voltage shall be avoided as far
as possible and, if inevitable, shall be carried out only by a skilled person who is aware of the
hazard involved.
If the instrument is clearly defective, has been subject to mechanical damage, excessive moisture
or chemical corrosion the safety protection may be impaired and the apparatus should be withdrawn
from use and returned for checking and repair. Make sure that only fuses with the required rated
current and of the specified type are used for replacement. The use of makeshift fuses and the
short-circuiting of fuse holders is prohibited.
Do not wet the instrument when cleaning it and in particular use only a soft dry cloth to clean the
display window. The following symbols are used on the instrument and in this manual:
WARNING - risk of electric shock.
CAUTION - refer to accompanying documentation;
incorrect operation may damage the meter.
e) EN61000-4-5 (1995) Surge, 0.5kV (line to line), 1kV (line to ground), Performance A.
f) EN61000-4-6 (1996) Conducted RF, 3V, 80% AM at 1kHz (AC line only; DC Output
According to EN61326 the definitions of performance criteria are:
Performance criterion A: ‘During test normal performance within the specification limits.’
Performance criterion B: ‘During test, temporary degradation, or loss of function or
performance which is self-recovering’.
Performance criterion C: ‘During test, temporary degradation, or loss of function or
performance which requires operator intervention or system reset occurs.’
Where Performance B is stated the multimeter will continue to function but accuracy may deviate
from Specification under the test conditions. However, the possible deviations are small and
infrequent and are unlikely to be a problem in practice.
Cautions
To ensure continued compliance with the EMC directive the following precautions should be
observed:
Performance B.
connections <3m not tested), Performance A.
a) only use screened cables and connectors to connect between the multimeter’s RS232
interface and other equipment.
b) after opening the case for any reason ensure that all signal and ground connections are
remade correctly before replacing the cover. Always ensure all case screws are correctly
refitted and tightened.
c) In the event of part replacement becoming necessary, only use components of an identical
Check that the instrument operating voltage marked on the rear panel is suitable for the local
supply. Should it be necessary to change the operating voltage, proceed as follows:
1) Disconnect the instrument from all voltage sources.
2) Unclip the front bezel by gently pulling the centre of each long edge up and forward.
The case halves are held together by 4 plastic push-rivets. Use the blade of a small
screwdriver in the slot beside each rivet to first ease out the rivet head and then fully remove
the rivet body. Separate the case halves. Visit www.tti-test.com
3) Change the transformer connections following the diagrams below:
Installation
for further details.
4) Reassemble in the reverse order.
5) To comply with safety standard requirements the operating voltage marked on the rear panel
must be changed to clearly show the new voltage setting.
Mains Lead
When a three core mains lead with bare ends is provided it should be connected as follows:-
WARNING! THIS INSTRUMENT MUST BE EARTHED WHEN OPERATED FROM AN AC LINE SUPPLY
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
Battery Operation
To fit or replace the six ‘C’ cells, slide off the battery cover at the rear of the instrument and
remove the battery tray. For safety reasons the battery cover can only be slid off after first
removing the security screw with a screwdriver. Always refit the security screw after changing the
batteries.
Either rechargeable or disposable cells may be used; fit into the battery tray observing the polarity
marked in the tray. Suitable cells are as follows:
Rechargeable: 2 Amp hour, e.g. NCC200, AN220, VR2C, RSH1.8, P-180C.
Disposable: Alkaline, e.g. MN1400.
Zinc Carbon, e.g. R14B, R14S.
If batteries are fitted it is essential that the disposable/rechargeable selector switch on the rear
panel is positioned correctly before applying external power via the AC socket.
WARNING! If disposable batteries are fitted and the selector is set in the rechargeable
position, there is a risk of explosion if the meter is connected to an AC supply.
The Rechargeable batteries are charged at a ‘trickle’ rate of C/20; the time to fully charge from
completely discharged is approximately 28 hours.
Low battery condition is indicated by showing the battery symbol in the top left-hand corner of the
display; when this symbol shows, approximately 5% of battery life remains.
Handle
Adjust the handle position by pulling out the ends and rotating the handle to the desired position.
There are sixteen possible positions including five facing forward which provide upward tilts for
the most convenient bench operation. The handle may be swung out of the way to the back or
front of the instrument if no tilt is required. The natural spring in the handle will lock it into position
when the handle is released.
This section is a general introduction to the features and organisation of the multimeter intended
to be read before using the instrument for the first time. Detailed operation is covered in later
sections, starting with Making Basic Measurements.
Connections
Input Sockets
The input sockets are 4mm safety sockets on a 19mm pitch designed to accept 4mm safety plugs
with fixed or retractable shrouds. The sockets are all rated to 1000V (Cat I)/600V (Cat II) with
respect to earth ground.
The input impedance between V/Ω and COM is nominally 10MΩ on dc ranges and 1MΩ on ac
ranges. The black COM socket is considered less positive than the red socket.
The mA/10A current sockets are low impedance; the voltage burden between mA/10A and COM
at full scale is <250mV for the mA ranges and <500mV for the 10A range. The black COM socket
is considered less positive than the white mA/10A sockets.
Multimeter Test Leads
General Operation
RS232
The test leads supplied meet the requirements of IEC1010-2-031 and are rated to 1000V Cat III.
Use only the test leads provided, or a set of similar performance, to ensure safe operation.
Alternative test leads should be rated to at least 1000V (Cat I), 600V (Cat II) and 10A current
capability.
The RS232 interface only operates when the instrument is run from an ac supply; RS232 signal
ground is then connected to the safety ground. The interface is fully isolated from the
measurement system.
9-pin D-connector compatible with the Thurlby Thandar ARC (Addressable RS232 Chain) system.
The pin connections are shows below:
PinNameDescription
1 - Power to optional PC-02
2 TXD Transmitted data from instrument
3 RXD Received data to instrument
4 - No internal connection
5 GND Signal ground
6 - No internal connection
7 RXD2 Secondary received data
8 TXD2 Secondary transmitted data
9 GND Signal ground
Pin 2, 3 and 5 may be used as a conventional RS232 interface with XON/XOFF handshaking.
Pins 7,8 and 9 are additionally used when the instrument is connected to the ARC interface.
Signal grounds are connected to safety ground.
To ensure compliance with EMC legislation use only screened cable assemblies with screened
connectors when connecting to other equipment.
GPIB (IEEE-488)
An IEEE-488 interface is available on the alternative mains-only version; IEEE-488 is not a
retrofittable option. The GPIB connector is located on the rear panel together with the switch
which selects between RS232 and GPIB.