Copyright
TSI Incorporated / 2007–2013 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective June 2011)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods sold hereunder, under normal use and service as described in
the operator's manual, shall be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator's manual, from the date of
shipment to the customer. This warranty period is inclusive of any statutory warranty. This
limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date
of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s
manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's
warranty applies;
e. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty
with respect to, and shall have no liability in connection with, goods which are
incorporated into other products or equipment, or which are modified by any person
other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated
herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID
WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES
CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS. BUYER’S EXCLUSIVE
REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR
REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF
THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR
BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED
ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL
BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE
PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE
GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE
DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE
PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST
PROFITS OR ANY SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER
SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR
REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be
brought against Seller more than 12 months after a cause of action has accrued. The
goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be
returned, if at all, at Seller's risk of loss.
Page 4
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This
LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms
waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are
to our customers, our service policy is designed to give prompt attention to any
problems. If any malfunction is discovered, please contact your nearest sales office or
representative, or call TSI's Customer Service department at (800) 874-2811 (USA) or
(001 651) 490-2811 (International) or visit www.tsi.com.
Page 5
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............ 1
APPENDIX A SPECIFICATIONS ..................................................... 11
iii
Page 6
(This page intentionally left blank)
Page 7
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components below. If
anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
1
Page 8
2
(This page intentionally left blank)
Chapter 1
Page 9
N o t e
For temperature measurements, make sure that at least 3 inches
(7.5 cm) of the probe is in the flow to allow the temperature sensor to
be in the air stream.
Chapter 2
Setting-up
Supplying Power to the Model 9515 VELOCICALC Air Velocity
Meter
The Model 9515 is powered with four size AA batteries.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on the
inside of the battery compartment. The Model 9515 is designed to
operate with either alkaline or NiMH rechargeable batteries, although it
will not recharge NiMH batteries. Battery life will be shorter if NiMH
batteries are used. Carbon-zinc batteries are not recommended because
of the danger of battery acid leakage.
Using the Telescoping Probe
The telescoping probe contains the velocity and temperature sensors. When
using the probe, make sure the sensor window is fully exposed and the
orientation dimple is facing upstream.
Extending the Probe
To extend the probe, hold the handle in one hand while pulling on the
probe tip with the other hand. Do not hold the cable while extending the
probe as this prevents the probe from extending.
Retracting the Probe
To retract the probe, hold the handle in one hand while pulling gently
on the probe cable with the other hand.
3
Page 10
(This page intentionally left blank)
4Chapter 2
Page 11
ON/OFF Key
Press to turn the Model 9515 on and off. During
the power up sequence the display will show the
following: Model Number, Serial Number,
Software Revision, and Last Date Calibrated.
ft/min / m/s Key
Pressing this key changes the display to read air
velocity.
°C / °F Key
Pressing this key changes the display to read
temperature.
Changing Units
To change units, first put the desired measurement
(air velocity or temperature) on the display. Then
press and hold the left, unlabelled key for five
seconds. Finally, use the and ENTER key to
select the units.
Chapter 3
Operation
Keypad Functions
5
Page 12
(This page intentionally left blank)
6 Chapter 3
Page 13
U.S. & International
TSI Incorporated
500 Cardigan Road
Shoreview MN 55126-3996
Tel: (800) 874-2811
(651) 490-2811
Fax: (651) 490-3824
Chapter 4
Maintenance
The Model 9515 requires very little maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model 9515 to TSI for annual recalibration.
Please contact one of TSI’s offices or your local distributor to make service
arrangements and to receive a Return Material Authorization (RMA)
number. To fill out an online RMA form, visit TSI’s website at
http://service.tsi.com.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft
cloth and isopropyl alcohol or a mild detergent. Never immerse the
Model 9515. If the enclosure of the Model 9515 becomes broken, it must be
replaced immediately to prevent access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month to
prevent damage due to battery leakage.
7
Page 14
(This page intentionally left blank)
8 Chapter 4
Page 15
Symptom
Possible Causes
Corrective Action
No Display
Unit not turned on
Switch unit on.
Low or dead batteries
Replace batteries.
Dirty battery contacts
Clean the battery
contacts.
Velocity reading
fluctuates unstable
Fluctuating flow
Reposition probe in
less-turbulent flow
Instrument Error
message appears
Fault in instrument
Factory service required
on instrument.
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions
for common problems encountered with the Model 9515. If your symptom is
not listed, or if none of the solutions solves your problem, please contact
TSI.
Table 5-1: Troubleshooting the VELOCICALC Model 9515
W A R N I N G !
Remove the probe from excessive temperature immediately:
excessive heat can damage the sensor. Operating temperature limits
can be found in Appendix A, Specifications.
9
Page 16
(This page intentionally left blank)
10 Chapter 5
Page 17
Appendix A
Specifications
Specifications are subject to change without notice.
Velocity:
Range: 0 to 4000 ft/min (0 to 20 m/s)
Accuracy
Resolution: 1 ft/min (0.01 m/s)
Temperature:
Range: 0 to 200°F (-18 to 93°C)
Accuracy3: ±0.5°F (±0.3°C)
Resolution: 0.1°F (0.1°C)
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C)
Operating (Probe): 0 to 200°F (-18 to 93°C)
Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Pollution degree 1 in accordance with IEC 664
Transient over voltage category II
1&2
: ±5% of reading or ±5 ft/min (±0.025 m/s),
whichever is greater
External Meter Dimensions:
3.3 in. 7.0 in. 1.8 in. (8.4 cm 17.8 cm 4.4 cm)
Meter Weight:
Weight with batteries: 0.6 lbs (0.27 kg)
Power Requirements:
Four AA-size batteries (included)
1
Temperature compensated over an air temperature range of 40 to 150°F (5 to 65°C).
2
The accuracy statement of ±5.0% of reading or ±5 ft/min (±0.025 m/s), whichever is greater,
begins at 30 ft/min through 4000 ft/min (0.15 m/s through 20 m/s).
3
Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F (0.03°C/°C) for
change in instrument temperature.
11
Page 18
(This page intentionally left blank)
12 Appendix A
Page 19
USA Tel: +1 800 874 2811
UK Tel: +44 149 4 459200
France Tel: +33 4 91 11 87 64
Germany Tel: +49 241 523030
India Tel: +91 80 67877200
China Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
TSI Incorporated – Visit our website www.tsi.com for more information.