TSI Instruments SIZER EEPS 3090, SIZER EEPS 3090AK Operation And Service Manual

Page 1
ENGINE EXHAUST PARTICLE
SIZER™ (EEPS™) SPECTROMETER
MODEL 3090/3090AK
OPERATION AND SERVICE MANUAL
P/N 1980494, REVISION J
MARCH 2015
Page 2
Page 3
ENGINE EXHAUST PARTICLE
SIZER (EEPS) SPECTROMETER
MODEL 3090/3090AK
OPERATION AND SERVICE MANUAL
Product Overview
1
Unpacking and Setting Up
2
Quick Start
3
Instrument Operation
4
Software Operation
5
General Care, Troubleshooting and Service
6
Appendixes and
Index
Page 4
Manual History
The following is a history of the Model 3090/3090AK Engine Exhaust Particle Sizer™ (EEPS™) Spectrometer Operation and Service Manual, P/N 1980494.
Revision Date
A January 2004 B April 2004 C March 2005 D March 2006 E August 2006 F March 2009 G October 2010 H April 2011 J March 2015
ii
Page 5
Part Number
1980494 / Revision J / March 2015
Copyright
©TSI Incorporated / 2004–2015 / All rights reserved
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
E-mail Address
particle@tsi.com
Limitation Of Warranty And Liability
(effective February 2014)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.) Seller warrants the goods, excluding software, sold hereunder, under normal use and service as
described in the operator's manual, to be free from defects in workmanship and material for 12 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
Warranty
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components
when indicated in specifications, are warranted for 90 days from the date of shipment; b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals; c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in
workmanship and material, under normal use, for 90 days from the date of shipment; d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses,
batteries or other consumable materials. Only the original manufacturer's warranty applies; e. This warranty does not cover calibration requirements, and seller warrants only that the
instrument or product is properly calibrated at the time of its manufacture. Instruments returned
for calibration are not covered by this warranty; f. This warranty is VOID if the instrument is opened by anyone other than a factory authorized
service center with the one exception where requirements set forth in the manual allow an
operator to replace consumables or perform recommended cleaning; g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or
intentional damage, or is not properly installed, maintained, or cleaned according to the
requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller
makes no warranty with respect to, and shall have no liability in connection with, goods which
are incorporated into other products or equipment, or which are modified by any person other
than Seller. The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein.
NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR
MERCHANTABILITY IS MADE. WITH RESPECT TO SELLER’S BREACH OF THE IMPLIED
WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED
INFRINGEMENTS. BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S
OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS. TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER,
AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
iii
Page 6
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI Customer Service at 1-800-874-2811 (USA) or 651-490-2811.
Software License
(effective March 1999)
This is a legal agreement between you, the end user, and TSI Incorporated. BY INSTALLING THE SOFTWARE, YOU ARE AGREEING TO BE BOUND BY THE TERMS OF THIS AGREEMENT. IF YOU DO NOT AGREE TO THE TERMS OF THIS AGREEMENT, PROMPTLY RETURN THE UNOPENED PACKAGE AND THE ACCOMPANYING ITEMS (including written materials and binders or other containers) to TSI for a full refund.
1. GRANT OF LICENSE. TSI grants to you the right to use one copy of the enclosed TSI software program (the “SOFTWARE”), on a single computer. You may not network the SOFTWARE or otherwise use it on more than one computer or computer terminal at the same time.
2. COPYRIGHT. The SOFTWARE is owned by TSI and is protected by United States copyright laws and international treaty provisions. Therefore, you must treat the SOFTWARE like any other copyrighted material (e.g., a book or musical recording) except that you may either (a) make one copy of the SOFTWARE solely for backup or archival purposes, or (b) transfer the SOFTWARE to a single hard disk provided you keep the original solely for backup or archival purposes.
3. OTHER RESTRICTIONS. You may not rent or lease the SOFTWARE, but you may transfer the SOFTWARE and accompanying written material on a permanent basis, provided you retain no copies and the recipient agrees to the terms of this Agreement. You may not reverse-engineer, decompile, or disassemble the SOFTWARE.
4. DUAL MEDIA SOFTWARE. If the SOFTWARE package contains multiple types of media, then you may use only the media appropriate for your single-user computer. You may not use the other media on another computer or loan, rent, lease, or transfer them to another user except as part of the permanent transfer (as provided above) of all SOFTWARE and written material.
5. U.S. GOVERNMENT RESTRICTED RIGHTS. The SOFTWARE and documentation are provided with RESTRICTED RIGHTS. Use, duplication, or disclosure by the Government is
subject to the restrictions set forth in the “Rights in Technical Data and Computer Software”
Clause at 252.227-7013 and the “Commercial Computer Software - Restricted Rights” clause at
52.227-19.
6. LIMITED WARRANTY. TSI warrants that the SOFTWARE will perform substantially in accordance with the accompanying written materials for a period of ninety (90) days from the date of receipt.
7. CUSTOMER REMEDIES. TSI’s entire liability and your exclusive remedy shall be, at TSI’s option, either (a) return of the price paid or (b) repair or replacement of the SOFTWARE that does not meet this Limited Warranty and which is returned to TSI with proof of payment. This Limited Warranty is void if failure of the SOFTWARE has resulted from accident, abuse, or misapplication. Any replacement SOFTWARE will be warranted for the remainder of the original warranty period or thirty (30) days, whichever is longer.
8. NO OTHER WARRANTIES. TSI disclaims all other warranties, either express or implied, including, but not limited to implied warranties of merchantability and fitness for a particular purpose, with regard to the SOFTWARE and the accompanying written materials.
9. NO LIABILTY FOR CONSEQUENTIAL DAMAGES. In no event shall TSI be liable for any damages whatsoever (including, without limitation, special, incidental, consequential or indirect damages for personal injury, loss of business profits, business interruption, loss of information or any other pecuniary loss) arising out of the use of, or inability to use, this SOFTWARE.
Trademarks
Microsoft and Excel are registered trademarks of Microsoft Corporation. Windows and Access are trademarks of Microsoft Corporation. Swagelok is a registered trademark of Swagelok® Companies, Solon, Ohio. Engine Exhaust Particle Sizer, EEPS, Scanning Mobility Particle Sizer, and SMPS are trademarks of TSI Incorporated.
TSI Patent
US Patent Number 7,230,431.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
iv Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 7
C a u t i o n
Caution means be careful. It means if you do not follow the procedures
prescribed in this manual you may do something that might result in equipment damage, or you might have to take something apart and start over again. It also indicates that important information about the operation and maintenance of this instrument is included.
W A R N I N G
Warning means that unsafe use of the instrument could result in serious
injury to you or cause irrevocable damage to the instrument. Follow the procedures prescribed in this manual to use the instrument safely.
Safety
This section gives instructions to promote safe and proper operation of the Model 3090 Engine Exhaust Particle Sizer (EEPS) Spectrometer. Samples of warnings and information on labels attached to the instrument chassis are also presented.
Description of Caution S y mbol
The following symbol and an appropriate caution statement are used throughout the manual and on the EEPS spectrometer to draw attention to any steps that require you to take cautionary measures when working with the Model 3090:
Caution
Warning
v
Page 8
Warns you that uninsulated voltage within the instrument may have sufficient magnitude to cause electric shock. Therefore, it is dangerous to make any contact with any part inside the instrument.
Warns you that the instrument is susceptible to electro-static dissipation (ESD) and ESD protection procedures should be followed to avoid damage.
Indicates the connector is connected to earth ground and cabinet ground.
W A R N I N G
High voltage is accessible in several locations within these instruments. Make sure you unplug the power source before removing the cover or performing maintenance procedures.
1 Serial Number Label (back of cabinet)
Caution or Warning Symbols
The following symbols may accompany cautions and warnings to indicate the nature and consequences of hazards:
Electrical Safety
Labels
The EEPS spectrometer has high-voltage points within its cabinet. Only a qualified technician should perform service or maintenance.
The EEPS spectrometer has labels on the back of the instrument and on interior components. Labels are described below:
vi Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 9
2 Caution, No Serviceable Parts Label
(back of cabinet)
3 Customer Service Label (back of
cabinet)
4 High-Voltage Symbol Label (interior,
on power entry module)
5 Ground Symbol Label (interior,
bottom, next to ground stud)
Safety vii
Page 10
Lifting Caution
The EEPS spectrometer weighs 32 kg (70 lbs). To protect your back when lifting:
Get help from another person to move the instrument. The
Transport the instrument on a cart whenever possible. Lift with your legs while keeping your back straight. Keep the instrument close to your body as you lift.
instrument is equipped with handles to allow for two people to safely lift it.
viii Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 11
Contents
Manual History ............................................................................................ ii
Warranty ..................................................................................................... iii
Safety ........................................................................................................... v
Description of Caution Symbol .............................................................. v
Caution ................................................................................................ v
Warning ............................................................................................... v
Caution or Warning Symbols ............................................................. vi
Electrical Safety .................................................................................... vi
Labels .................................................................................................... vi
Lifting Caution ...................................................................................... viii
About This Manual ................................................................................. xvii
Purpose ............................................................................................... xvii
Related Product Literature .................................................................. xvii
Submitting Comments ......................................................................... xvii
CHAPTER 1 Product Overview .............................................................. 1-1
Product Description ............................................................................. 1-1
Applications ......................................................................................... 1-2
Instrument Description ........................................................................ 1-3
Front Panel ...................................................................................... 1-3
LCD Display ..................................................................................... 1-3
Control Knob .................................................................................... 1-3
Back Panel .......................................................................................... 1-4
Heat Exchanger Fan ........................................................................ 1-4
Aerosol Inlet/Cyclone ....................................................................... 1-5
Aerosol Outlet .................................................................................. 1-5
AC Connector and Switch ................................................................ 1-5
RS-232 Serial Connection ............................................................... 1-5
Analog and Trigger Input/Output ..................................................... 1-6
Analog Voltage Out Channels 1–4 .................................................. 1-7
System Diagram and Operational Overview ...................................... 1-7
CHAPTER 2 Unpacking and Setting Up ................................................ 2-1
Packing List ......................................................................................... 2-1
Unpacking ........................................................................................... 2-2
Setting Up the Engine Exhaust Particle Sizer Spectrometer .............. 2-4
Instrument Placement ...................................................................... 2-4
Installing the Cyclone ....................................................................... 2-4
Connecting Power ........................................................................... 2-5
Connecting the Computer ................................................................ 2-5
Installing the Engine Exhaust Particle Sizer Software ..................... 2-6
ix
Page 12
CHAPTER 3 Quick Start .......................................................................... 3-1
How To Perform a Sample Run .......................................................... 3-2
Step 1Open a New File ................................................................ 3-3
Step 2Start the Run ..................................................................... 3-4
Communications Errors ................................................................... 3-5
Run Properties ................................................................................. 3-5
Viewing Data ....................................................................................... 3-8
CHAPTER 4 LCD Display Operation ..................................................... 4-1
LCD Display Items .............................................................................. 4-1
Status Bar ........................................................................................ 4-2
Menu Options .................................................................................. 4-2
Concentration Display ..................................................................... 4-4
Particle Data Graph ......................................................................... 4-4
CHAPTER 5 Software Operation ........................................................... 5-1
Start the Engine Exhaust Particle Sizer Program ............................... 5-2
Open an Existing File .......................................................................... 5-3
Selecting a Particle Histogram to Display .......................................... 5-4
Maximum Concentration .................................................................. 5-5
Minimum Concentration ................................................................... 5-6
Change How Data is Viewed .............................................................. 5-6
Changing the Time Resolution ........................................................ 5-7
Changing the Concentration Scale .................................................. 5-7
Changing the Weighting in the 2D Histogram ................................. 5-9
Changing the Display Units ............................................................. 5-9
Using an Alternative Inversion Matrix ............................................... 5-10
Matrix File Format .......................................................................... 5-11
Playback (Review) a Sample Run .................................................... 5-11
Select a Data Hot Spot ..................................................................... 5-12
Delete Data ....................................................................................... 5-13
Zoom In and Out on Data in a Graph ............................................... 5-13
Print Information Displayed on the Desktop ..................................... 5-14
Export Data to a File ......................................................................... 5-15
To Manually Export Data ............................................................... 5-15
To Automatically Export Data ........................................................ 5-18
Arrange Open Windows/Change Layout .......................................... 5-19
Quit the Program .............................................................................. 5-19
Software Menus ................................................................................ 5-19
Desktop Menus ................................................................................. 5-20
File Menu ....................................................................................... 5-20
Run Menu ...................................................................................... 5-23
Playback Menu .............................................................................. 5-24
Format Menu ................................................................................. 5-24
View Menu ..................................................................................... 5-29
Window Menu ................................................................................ 5-40
Help Menu ..................................................................................... 5-41
Context-Sensitive Menus .................................................................. 5-41
Run View Popup Menu .................................................................. 5-41
Particle Table Popup Menu ........................................................... 5-43
x Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 13
Concentration Histogram and Total Concentration
Popup Menus ............................................................................. 5-44
3D Particle Concentration Popup Menu ........................................ 5-47
Status Bar Icons ................................................................................ 5-47
Shortcut Keys .................................................................................... 5-48
CHAPTER 6 General Care, Troubleshooting and Service ................... 6-1
Basic Care of the Engine Exhaust Particle Sizer Spectrometer ......... 6-1
Cleaning the Cyclone .......................................................................... 6-2
Cleaning and Replacing Charger Needles ......................................... 6-3
Cleaning the Negative Charger Needle ........................................... 6-3
Cleaning or Replacing the Positive Charger Needle ....................... 6-9
Cleaning the Charger Column .......................................................... 6-10
Basic Charger Column Cleaning Procedure ................................. 6-11
Detailed Charger Column Cleaning Procedure ............................. 6-14
Replacing the Filter Cartridges ......................................................... 6-21
Manifold Filters ............................................................................... 6-21
Pump Filters ................................................................................... 6-22
Replace the Extraction Pump ........................................................... 6-23
Troubleshooting ................................................................................ 6-24
Baseline Noise ............................................................................... 6-24
Noisy Channels .............................................................................. 6-26
Errors ............................................................................................. 6-29
Technical Contacts ........................................................................... 6-33
Returning the Engine Exhaust Particle Sizer Spectrometer
for Service ...................................................................................... 6-34
APPENDIX A Specifications .................................................................. A-1
APPENDIX B Theory of Operation ........................................................ B-1
Electrical Mobility History ................................................................... B-1
Size Distributions ............................................................................... B-2
Charger ........................................................................................... B-3
Measurement Column .................................................................... B-3
Flow Path ........................................................................................ B-4
Data Inversion ................................................................................. B-4
Concentration Limits ....................................................................... B-4
Selected References .......................................................................... B-6
APPENDIX C Particle Size Statistics .................................................... C-1
APPENDIX D Serial Commands ............................................................ D-1
Pin Connectors .................................................................................. D-1
Serial Protocol .................................................................................... D-2
Serial Commands .............................................................................. D-2
Command Summary .......................................................................... D-5
READ Commands .......................................................................... D-5
SET Commands .............................................................................. D-6
HELP Commands ........................................................................... D-6
Detailed Command Description ......................................................... D-7
READ Commands .......................................................................... D-7
SET Commands ............................................................................ D-12
Contents xi
Page 14
APPENDIX E Analog Output ................................................................. E-1
How Calculations are Made ............................................................... E-2
Specifications ..................................................................................... E-4
APPENDIX F AK-Protocol ....................................................................... F-1
Introduction ......................................................................................... F-1
Interface Specifications ................................................................... F-1
Protocol Specifications .................................................................... F-2
AK Command Overview ..................................................................... F-4
Control Commands – (S) ................................................................. F-4
Write Commands – (E) .................................................................... F-4
Read Commands – (A) .................................................................... F-4
Command Details ............................................................................... F-5
Control Commands .......................................................................... F-5
Write Commands ............................................................................. F-5
Read Commands ............................................................................. F-6
Particle Size Bin Numbers ................................................................ F-12
Index
Reader’s Comments
Figures
1-1 Model 3090 Engine Exhaust Particle Sizer Spectrometer .................. 1-1
1-2 View of the EEPS Spectrometer LCD Display .................................... 1-3
1-3 Back Panel of the EEPS Spectrometer .............................................. 1-4
1-4 Detail of EEPS Spectrometer Back Panel .......................................... 1-5
1-5 Schematic Diagram of the Model 3090 EEPS Spectrometer ............. 1-7
2-1 Removing the Shipping Strap ............................................................. 2-3
2-2 Cyclone Assembly .............................................................................. 2-5
3-1 EEPS Warm-up Display Screen ......................................................... 3-1
3-2 The Engine Exhaust Particle Sizer Desktop ....................................... 3-2
3-3 Open a New File ................................................................................. 3-3
3-4 Properties Dialog Box ......................................................................... 3-4
3-5 Communications Error Dialog Box ...................................................... 3-5
3-6 Sample Data ....................................................................................... 3-9
3-7 Popup Menu for the Histogram Window ........................................... 3-10
4-1 LCD Display Regions .......................................................................... 4-1
5-1 Engine Exhaust Particle Sizer Desktop .............................................. 5-2
5-2 Selecting a Time Window ................................................................... 5-4
5-3 Histogram Showing Clipping Due to Exceeding Maximum
Concentration .................................................................................. 5-5
5-4 Concentration Histogram Showing Minimum Concentration .............. 5-6
5-5 Concentration Dialog Box ................................................................... 5-7
xii Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 15
5-6 Popup Menu for the Concentration Histogram Window ...................... 5-9
5-7 Playback Setup Dialog Box ............................................................... 5-11
5-8 Use the Mouse to Directly View Data Values ................................... 5-13
5-9 Print Preview ..................................................................................... 5-14
5-10 Export Data Options Dialog Box ..................................................... 5-15
5-11 Sample Export File .......................................................................... 5-18
5-12 File Menu ......................................................................................... 5-20
5-13 Run Menu ........................................................................................ 5-23
5-14 Playback Menu ................................................................................ 5-24
5-15 Format Menu ................................................................................... 5-24
5-16 Concentration Scale Dialog Box ..................................................... 5-25
5-17 (A) The Graph Color Dialog Box, (B) The Table Color
Dialog Box ..................................................................................... 5-26
5-18 The Color Dialog Box for the 3D Graph .......................................... 5-27
5-19 Font Dialog Box ............................................................................... 5-28
5-20 View Menu ....................................................................................... 5-29
5-21 3D Graph View ................................................................................ 5-29
5-22 Particle Table Window .................................................................... 5-30
5-23 Particle Histogram Window ............................................................. 5-30
5-24 Total Concentration Window ........................................................... 5-31
5-25 Instrument Status ............................................................................ 5-32
5-26 Electrometer Offsets and RMS Screen ........................................... 5-34
5-27 Zero Electrometer Function Dialog Box .......................................... 5-35
5-28 Zero Electrometers Progress Dialog Box ........................................ 5-35
5-29 Zero Electrometers Zeroing Complete ............................................ 5-36
5-30 Instrument Status Expanded ........................................................... 5-36
5-31 Analog Out Setup ............................................................................ 5-38
5-32 Window Menu.................................................................................. 5-40
5-33 Help Menu ....................................................................................... 5-41
5-34 Run View Popup Menu ................................................................... 5-41
5-35 Set Y Scale for Analog Graph Dialog Box ...................................... 5-42
5-36 Enter Analog Input Labels Dialog Box ............................................ 5-43
5-37 Particle Table Popup Menu ............................................................. 5-43
5-38 Concentration Histogram (A) and Total Concentration
(B) Popup Menus ........................................................................... 5-44
5-39 Weight Menu ................................................................................... 5-45
5-40 3D Particle Concentration Popup Menu .......................................... 5-47
6-1 Remove Top Panel ............................................................................. 6-4
6-2 Disconnect Sample Inlet Tube ............................................................ 6-4
6-3 Remove the Screws of the Negative Charger Assembly .................... 6-5
6-4 Lift off the Negative Charger Head ..................................................... 6-5
6-5 Remove the Honeycomb for Reassembly .......................................... 6-6
6-6 The Charger Needle ............................................................................ 6-7
6-7 Hold the Flange with Tweezers ........................................................... 6-8
6-8 Disconnect the Charger Sheath Flow Tubing ..................................... 6-9
6-9 Remove the Screws of the Positive Charger Assembly ................... 6-10
6-10 Unscrew Charger Head Assembly Screws ..................................... 6-11
Contents xiii
Page 16
6-11 Unclip the Wiring Harness Retainer................................................ 6-12
6-12 Cleaning Tool .................................................................................. 6-12
6-13 Cleaning the Charger Column ........................................................ 6-13
6-14 Remove the Charger Sheath Flow Tube from the
Charger Column ............................................................................ 6-14
6-15 Remove the Column Retaining Screws .......................................... 6-15
6-16 Disconnect the Electrometer Ribbon Connector ............................ 6-15
6-17 Disconnect the Sheath Air Hose ..................................................... 6-16
6-18 Disconnect the Pressure Transducer Tube at the Column ............ 6-16
6-19 Column Assembly Base Screws ..................................................... 6-17
6-20 Move the Outer Column Assembly to a Clean Work Surface ........ 6-18
6-21 Clean the Center Column ............................................................... 6-19
6-22 Clean the Rings of the Electrometer Assembly .............................. 6-20
6-23 Replacing Filters ............................................................................. 6-22
6-24 Extraction Pump Location (inside back panel) ............................... 6-23
6-25 Instrument Status Dialog Box ......................................................... 6-24
6-26 Electrometer Offset Values ............................................................. 6-26
6-27 Advanced Run Settings .................................................................. 6-27
6-28 Changing Data Export Modes Dialog Box ...................................... 6-28
6-29 Software Showing Raw Electrometer Noisy Channel..................... 6-28
6-30 Raw Electrometer Data After Cleaning ........................................... 6-29
Tables
6-31 Instrument Errors Detected Dialog Box .......................................... 6-29
6-32 Installing the Shipping Strap ........................................................... 6-35
B-1 Schematic of a Differential Mobility Analyzer .................................... B-1
B-2 Schematic of EEPS Measurement Column ....................................... B-3
B-3 Flow Schematic of the EEPS Spectrometer ...................................... B-5
B-4 Graph of Concentration Limits ........................................................... B-6
E-1 Analog Out Screen ............................................................................ E-2
F-1 9-pin D-sub Connector........................................................................ F-1
1-1 Analog and Trigger Input/Output ...................................................... 1-6
2-1 Components of the EEPS System .................................................... 2-1
4-1 User Settings .................................................................................... 4-2
4-2 Diagnostics Information .................................................................... 4-3
5-1 Export Data Options ........................................................................ 5-16
5-2 Options/Parameters of the Concentration Scale
Dialog Box ..................................................................................... 5-25
5-3 Instrument Status Parameters ........................................................ 5-33
5-4 Weight Options ............................................................................... 5-45
xiv Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 17
6-1 Maintenance Schedule ...................................................................... 6-2
6-2 Non-Fatal Error Flags ...................................................................... 6-30
6-3 Fatal Error Flags.............................................................................. 6-32
D-1 Signal Connections for RS-232 Configurations ............................... D-1
Contents xv
Page 18
(This page intentionally left blank)
xvi Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 19
About This Manual
Purpose
This is an operation and service manual for the Model 3090/3090AK Engine Exhaust Particle Sizer (EEPS) Spectrometer.
Related Product Lite rature
The following TSI product manual may be of interest. Copies of these may be viewed on the TSI website, www.tsi.com.
Model 3936 SMPS
Manual (part number 1933796) TSI Incorporated
Submitting Comments
TSI values your comments and suggestions on this manual. Please use the comment sheet on the last page of this manual to send us your opinion
on the manual’s usability, to suggest specific improvements, or to report
any technical errors.
If the comment sheet has already been used, please mail, fax or email your comments on another sheet of paper to:
TSI Incorporated 500 Cardigan Road Shoreview, MN 55126 USA
Fax: (651) 490-3824 Email: particle@tsi.com
TM
Scanning Mobility Particle SizerTM Instruction
xvii
Page 20
(This page intentionally left blank)
xviii Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 21
The Model 3090 (Figure 1-1) is a high­performance instrument designed specifically for measuring particles emitted from internal combustion engines and vehicles. It measures particle size from 5.6 to 560 nanometers with a sizing resolution of 16 channels per decade (a total of 32 channels). Its unique design includes revolutionary features that allow fast distribution measurements.
Figure 1-1
Model 3090 Engine Exhaust Particle Sizer Spectrometer
C H A P T E R 1
Product Overview
This chapter contains an introduction to the Model 3090 Engine Exhaust Particle Sizer (EEPS) Spectrometer (U.S. patent No. 7,230,431) and provides a brief explanation of how the system operates.
Product Description
1-1
Page 22
Applications
The Model 3090 Engine Exhaust Particle Sizer spectrometer is ideal for analysis of a wide range of engine exhaust situations.
Engine tailpipe measurements
Engine Dynamometer test cells Chassis Dynamometer test cells Old diesel engines with high emissions New generation diesels with active controls Measurements upstream and downstream of Diesel Particulate Filters
or Particle Traps
Spark Emission engine emissionsGDI Engines
Mobile Measurements
(Limited to 90 minute testsnot continuous monitoring.)
Mobile test laboratories Electric vehicle measurement platforms Roadside measurements (in trailer or enclosure) to evaluate the effect
of individual vehicles passing monitoring location
1-2 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 23
Instrument Description
Use the following information to familiarize yourself with the location and function of controls, indicators, and connectors on your Model 3090 Engine Exhaust Particle Sizer Spectrometer.
Front Panel
The main components of the front panel are the color LCD display and the control knob shown in Figure 1-2.
Figure 1-2 View of the EEPS Spectrometer LCD Display
LCD Display
The color 640 480 pixel LCD display provides continuous real-time display of sample data and is used in conjunction with the control knob to display options and instrument parameters. Refer to Chapter 4 for details of how to make selections and change values on the menu.
Instrument status is displayed at the top of the LCD display to indicate that required instrument voltages and flows have stabilized and the instrument is ready to use. Error status is also provided.
Control Knob
Turning the control knob highlights an item on the LCD display. Pressing the knob inward selects the option. To spin the knob quickly, place your finger in the indent on the knob surface.
Product Overview 1-3
Page 24
Back Panel
Fan Guard/Filter
Inlet/Cyclone
Outlet
Location for SN
and service labels AC Connector and
Switch
RS-232 Serial Port
Analog & Trigger
Input/Output
As shown in Figure 1-3, the back panel of the Model 3090 has power and data connections, an analog and trigger input/output connection, an aerosol inlet, outlet, and a cooling fan.
Figure 1-3
Back Panel of the EEPS Spectrometer
Heat Exchanger Fan
This fan keeps the sheath air temperature stable, near the ambient air temperature. It is provided with a guard and removable filter that should be cleaned periodically.
1-4 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 25
Analog In CH1
and CH2
Trigger In and Out
Analog Voltage
Out Channels 1–4
Aerosol Inlet/Cyclone
The aerosol inlet is designed for use with the supplied Cyclone (Model 1031083). The cyclone removes large particles outside the instrument measurement range that can cause counting errors. For remote sampling, tubing may be attached to the cyclone inlet. The cyclone is attached to the aerosol inlet using a Swagelok®-type union connector, which is supplied.
Aerosol Outlet
Sampled aerosol flow is filtered and exhausted at this port.
AC Connector and Switch
Plug the supplied AC cable into this receptacle. The instrument power switch is integrated into this AC connector at the top of the receptacle.
RS-232 Serial Connection
The data connection for the EEPS spectrometer is a standard RS-232 serial port connection that allows communications between the system computer and the Model 3090 for control by the EEPS software (Figure 1-4). Serial commands are sent to and from the computer to collect instrument status information and provide control functions.
Appendix D provides information that may be useful for developing custom
software. The appendix provides a description of the basic serial data communication commands. (In some cases serial commands may be used for troubleshooting.) Appendix F provides information for communication via AK-Protocol (only available for the Model 3090 AK).
®
Swagelok is a registered trademark of Swagelok® Companies, Solon, Ohio.
Figure 1-4
Detail of EEPS Spectrometer Back Panel
Product Overview 1-5
Page 26
Connection
Description
Trigger In
A trigger input here will start a run if the run has already been initiated in software and Enable External Trigger is checked in the Properties dialog box.
A trigger input consists of bringing the + input to a logic low, which can be done two ways. A potential-free contact closure can connect the + input to the – input which is at GND. Otherwise, a logic circuit can drive the + input low, provided the circuit’s ground is connected to the – input. Logic thresholds are <0.5V for logic low and >2.5V for logic high.
Caution
The trigger + input should not exceed +12V or -8V with respect to the – input.
Trigger Out
A potential-free contact closes to connect the 2 outputs when a run starts, and opens when the run stops.
Analog Out CH1 to CH4
Four channels of 0–5 V or 0–10 V analog voltage output, 14-bit resolution. Updated each second (1 HZ). Channels are software configurable to represent concentration data. See paragraph below for details.
Analog In CH1 and CH2
0-10V analog voltage input, 12-bit resolution, sampled at 1 Hz (this input is displayed as a voltage at the bottom of the run view).
Caution
The analog + input should not exceed +17V or
-4V with respect to the – input. Exceeding these values may damage the electronics.
Analog and Trigger Input/Output
The Analog and Trigger Input/Output connection is provided to allow the use of various triggers to synchronize the EEPS spectrometer data with another instrument or analog signal. These connections are fully electrically isolated from the rest of the instrument up to 500 volts. Details of the functions of each connection are listed in the table below:
Table 1-1 Analog and Trigger Input/Output
1-6 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 27
A schematic of the EEPS spectrometer is shown in Figure 1-5.
Particles enter the instrument as part of the aerosol inlet flow through a cyclone with a 1 µm cut. This removes large particles that are above the instrument's size range. Next, the particles pass through an electrical diffusion charger in which ions are generated. These mix with the particles and electrically charge them to provide a predictable charge level based on particle size. The charger is mounted inline with the analyzer column and located at the top of the instrument.
Figure 1-5
Schematic Diagram of the Model 3090 EEPS Spectrometer
Analog Voltage Out Channels 1–4
In addition to RS-232 serial communications for data transfer, particle concentration can be output by the EEPS spectrometer as analog voltages, to be readily integrated with other instruments on common laboratory data acquisition platforms requiring analog inputs. The EEPS spectrometer provides four independently configurable analog channels for this purpose. The four channels give the user the flexibility to select different particle size ranges and different concentration weightings (e.g., number, surface, mass) for conversion to a proportional voltage value.
Full scale analog voltage settings are available as 0–5 or 0–10 volts. Entering parameters for the Analog Out feature are described in Chapter 5,
Software Operation.
A further discussion of the application of the analog feature is described in
Appendix E.
System Diagram and Oper a tional Overview
Product Overview 1-7
Page 28
Particles then enter the sizing region through an annular gap, where they meet a stream of particle free sheath air. The sizing region is formed by the space between two concentric cylinders. The outer cylinder is built from a stack of sensing electrode rings that are electrically insulated from each other. The electrodes are connected to a very sensitive charge amplifier, also called an electrometer, with an input near ground potential.
The inner cylinder is connected to a positive high voltage supply, which forms the high voltage electrode. This creates an electric field between the two cylinders. While the positive charged particles stream with the sheath air from the top to the bottom of the sizing region, they are also repelled from the high voltage electrode and travel towards the sensing electrodes.
Particles, which land on the sensing electrodes transfer their charge. The generated current is amplified by the electrometers, digitized, and read by a microcontroller. The data are processed in real time to obtain 10 particle size distributions per second. The entire system is automated and data analysis is performed using a computer system with customized EEPS software. The software collects and stores sample data. Data is displayed in graphs and tables and can be exported to other applications.
For information on the theory of instrument operation, refer to Appendix B.
1-8 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 29
Packing List
Qty
Part Number
Description
1
962002
Computer RS-232 Serial Cable
1
1031083
Cyclone, 10 L/min, 1µm cut
1
1031252
Inlet Filter
1
1102125
USB to Serial Cable Adapter
1
1601501
Union Elbow Fitting 3/8-inch
2
1602051
HEPA Capsule Filter (spare)
1
1602071
Filter Media for Fan Guard (spare)
2
1602230
Filter, Balston DFU-BX (spare)
1
1980494
Operation and Service Manual, Model 3090 Engine Exhaust Particle Sizer Spectrometer
1
2900009
Assembly Cleaning Tool
1
3305070
Hex driver Set .050-3/8-inch, 13-piece
1
3900135
Model 3090 EEPS Software
1
2900008
Charger Needle Assemblies (Pack of 2, spare)
20
12 15 in. (30 38 cm) cleaning cloths*
C H A P T E R 2
Unpacking and Setting Up
Use the information in this chapter to unpack your Model 3090 Engine Exhaust Particle Sizer (EEPS) Spectrometer and set it up.
Table 2-1 lists the components shipped with the EEPS spectrometer.
Table 2-1
Components of the EEPS System
*
Available directly from Kimberly-Clark as Kimtech Prep Scottpure® Wipers, Product Code: 06121, www.kcprofessional.com
2-1
Page 30
Unpacking
C a u t i o n
The EEPS spectrometer is a heavy instrument weighing 70 pounds (32 kg). Protect your back when lifting:
Get help from another person to move the instrument Transport the instrument on a cart whenever possible Lift with your legs while keeping your back straight Keep the instrument close to your body as you lift
The Engine Exhaust Particle Sizer spectrometer comes assembled with protective coverings on the inlets and electrical connections. To prevent contamination, do not remove the protective covers until you are ready to use the instrument. In addition, the measurement column is shipped with a ratcheting tie-down strap to hold the column in place on its shock­absorbing mounts. For best measurement results, the strap should be removed for measurements.
Refer to Figure 2-1 and follow the steps below to remove the shipping strap.
1. Make sure the instrument is unplugged from power. Loosen the two
screws on each side of the top panel (Note: They should not have to be removed).
2. Squeeze the release mechanism as shown to allow the ratchet lever to
be moved up.
3. While holding the release mechanism, swing the release lever through
180 degrees until it clicks and releases tension on the strap. While still holding the lever in release mode, pull the assembly away from the panel until all the strap has been removed.
4. Pull the strap up from the bottom of the cabinet to pull the strap out of
the cabinet. Retighten the four screws that you loosened in step 1.
When shipping the instrument, the strap should be placed back on the column again. Take care to pad the ratchet mechanism with cardboard that is taped to the side panel to prevent damage to the painted panel. Save the shipping strap and ratchet in the accessory kit for future shipping needs. Save the original packaging materials for future use should the instrument need to be returned to TSI for service. See Chapter 6, “Returning the
Engine Exhaust Particle Sizer Spectrometer for Service”.
2-2 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 31
1 2 3
4
Figure 2-1
Removing the Shipping Strap
If anything is missing or appears to be damaged, contact your TSI representative, or contact TSI Customer Service at 1-800-874-2811 (USA) or 001-(651) 490-2811. Chapter 6 includes instructions on returning the EEPS spectrometer to TSI Incorporated.
Unpacking and Setting Up 2-3
Page 32
C a u t i o n
Be certain that the Cyclone is installed on the Aerosol Inlet in the orientation shown in Figure 2-1 prior to making particle size measurements. Make sure that the cyclone is cleaned regularly and that it is removed and cleaned before shipping.
Setting Up the E ngine Exhaust Pa r ticle Sizer Spectrometer
The following provides information for setting up the Model 3090 Engine Exhaust Particle Sizer (EEPS) spectrometer.
Instrument Placement
The instrument cabinet is designed to be cooled by room air drawn in through a filter from the back of the cabinet and exhausted through the sides and front of the cabinet.
The cabinet should be installed with at least 3-inch (50-mm) clearance between the back panel and side panels and any other surface. The cabinet should be set on a flat, solid surface that isolates the instrument from outside vibration. Excessive vibration will produce artificial noise in the measurement.
When positioning the instrument, also allow adequate space at the back for access to the power switch and for the power cord.
Installing the Cyclone
The cyclone functions by removing large particles, preventing counting errors, and keeping the instrument clean. The cyclone is always used, and should only be removed when transporting or servicing the instrument. The cyclone will remove large particles and fibers that can cause excessive noise on one or more channels. Since cleaning the electrodes is a relatively involved procedure, it is much better to prevent contamination using the cyclone. The cyclone will be more effective in removing large fibers if the inside surfaces are lightly coated with a light oil.
Install the cyclone on the aerosol inlet using the union fitting shown in Figure 2-2. Make certain the components are tightly pushed together. Tighten the nuts on the union fitting finger tight and then ½-turn with a wrench (see Swagelok recommendations).
2-4 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 33
To Aerosol Inlet on
Instrument
Union Fitting
Aerosol In
Removable Base
Connect
Cyclone to
Union Fitting
Figure 2-2
Cyclone Assembly
Connecting Power
1. Connect the supplied power cord to the receptacle at the back of the instrument. The cord has been supplied with the appropriate connector for your country. Supply voltage can be 100 to 240 VAC and 50 or 60 Hz.
2. The instrument on-off switch is located above the power cable connection on the instrument.
Notes: Make certain the line cord is plugged into a grounded (earth
grounded) power outlet. Position the instrument so the power cord connector is not blocked and is easily accessible.
The internal power supply contains no user-serviceable parts.
If the power supply is not operating correctly, use the information in the troubleshooting section to contact TSI. This instrument should not be used in a manner not specified by the manufacturer.
Connecting the Computer
1. Locate the serial cable supplied with the Model 3090. Connect one end of the cable to your computer’s serial interface port (COM 1, COM2, or other if available). If you have a USB port only, use the adapter cable (1102125) supplied in the accessory kit.
2. Connect the other end of the cable to the serial connection at the back of the EEPS spectrometer.
Unpacking and Setting Up 2-5
Page 34
Installing the Engine Exhaust Particle Sizer Software
Install the EEPS software as described below.
Computer Requirements
To use this software we recommend a personal computer with the following minimum features, components, and software:
An SVGA color monitor. Windows A hard drive large enough to accommodate Windows, the EEPS
software, and data files.
A CD-ROM drive. A mouse. An RS-232 serial interface port (in addition to the one that may be
required for the mouse) or USB connection (use with supplied USB to Serial Cable Adapter).
®
7 operating system or newer.
A Microsoft Windows-compatible printer is optional.
Program Installation
1. Shut down (exit) all programs/applications on the Windows desktop.
2. With the computer on and Windows running, insert the Model 3090 EEPS Software CD-ROM in your CD drive. To run the setup.exe from the CD:
a. If AutoPlay is enabled on your PC, the setup program begins
automatically and the introduction screen is displayed on the Windows desktop.
b. If AutoPlay is not enabled, select Run from the Start menu and
type: D:\SETUP in the Open box and press OK. (if D is the letter corresponding to your CD drive.)
3. Follow the instructions as the setup program runs. When the setup is complete, you may be asked if you want to view the readme.htm file. The readme.htm file contains important information that could not be included in this manual. If you decide not to view this file immediately, you may view the file later using a browser such as Internet Explorer.
4. When the installation program finishes, remove the CD-ROM and restart your computer. Store the CD-ROM in a safe place for later use.
2-6 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 35
The setup program creates a folder (directory) called “Engine Exhaust Particle Sizer” on your hard disk (assuming you accepted the default folder name). The folder contains the required program files and sample data files.
The setup program also creates a new item in the Start Menu called “TSI”
and an icon on your desktop for the EEPS software.
Note: Before creating a TSI menu item, the setup program checks for an
existing one. If one is present, it adds the icon only.
Unpacking and Setting Up 2-7
Page 36
(This page intentionally left blank)
2-8 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 37
C H A P T E R 3
Quick Start
Ease of operation is a key feature of the Model 3090 Engine Exhaust Particle Sizer (EEPS) spectrometer. After you are familiar with the components identified in Chapter 1 and have performed the setup steps in
Chapter 2, you are ready to begin taking measurements.
This chapter presents the very basics of operation. Detailed instrument and software operation instructions are in Chapters 4 and 5 and can be referenced as needed.
Turn the instrument on using the power switch at the back of the instrument, above the power cable connection. Once the instrument is turned on, the LCD display illuminates information as it appears in Figure 3-1. The instrument requires 10 minutes to warm up. When the warm up is finished the status will change to “Normal”.
Figure 3-1
EEPS Warm-up Display Screen
While the instrument is warming up, start the EEPS software program (if it is not already running). To start the program, proceed as follows:
1. From the Windows desktop, press the Start button and select Programs|TSI|Engine Exhaust Particle Sizer or select the EEPS icon from the desktop.
3-1
Page 38
The Engine Exhaust Particle Sizer desktop appears as shown in Figure 3-2.
Figure 3-2 The Engine Exhaust Particle Sizer Desktop
How To Perform a Sample R u n
There are two steps to collecting sample data.
Open a new file. Start data collection.
Before beginning, make certain the Model 3090 is connected to the computer, the instrument is warmed up, and the software is running.
3-2 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 39
Step 1Open a New File
Select File|New or or <Ctrl><N> on the toolbar. The dialog box shown in Figure 3-3 opens on your desktop.
Figure 3-3
Open a New File
Select a filename (and, optionally, a location other than the EEPS folder).
The default name is “EEPS1” for the first file you open, “EEPS2” for the
second file you open and so on. Accept the default name or enter any name you choose. You do not need to enter an extension in the filename box. It will be assigned automatically (.3090d).
After you enter a filename (or if you accept the default name), press the Open button.
After you press the Open button, the computer attempts to connect to the instrument. When the connection is complete, the Run View window opens. Other windows that were open the last time the application was closed are also opened. All windows are initially empty.
Quick Start 3-3
Page 40
Step 2Start the Run
Start the run one of three ways:
Select Run|Start Data Collection.  Click on the toolbar.
Press <F10>.
Each time you start a new run, you are shown the run properties you will use. The Run Properties dialog box shown in Figure 3-4 opens on the desktop.
3-4 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Figure 3-4 Properties Dialog Box
The parameters of the Properties dialog box are described below. For this “testrun, simply type a title and run length for the test and select OK.
Note: To open the Properties dialog box prior to pressing the Start button
on the toolbar, select Run|Properties. The values you enter in the properties dialog at this time will appear when the dialog box opens to start the run.
After you have completed the properties dialog box, sampling begins at the Start Time you selected or immediately after you press the OK button.
As the program begins collecting sample data, it is displayed in the windows open on the desktop.
Page 41
When the sample finishes, data for the sample is stored in the file, but the windows remain open.
To stop data collection before the selected run length time is complete, select Run|Stop Data Collection or on the toolbar.
Communications Errors
If you get the following error message (Figure 3-5) when you try to start a new sample, there is a communications problem between the computer and instrument (i.e., the cable is not connected) or a connection is already open to the instrument, such as another run, and are attempting to open another connection.
Figure 3-5 Communications Error Dialog Box
Check the cable connection and close all files and then try to reconnect.
Note: Most computers have only two active communications ports, COM1
and COM2. COM1 is most likely available for the EEPS connection.
Run Properties
The Run Properties dialog box contains three tabs: Run Settings, Analog Out, and Communications.
If this is the first time you have accessed the Properties dialog, you need to set up the parameters before continuing. If you have set up the properties before, verify they are correct before pressing OK. The following paragraphs describe the settings of the Properties dialog box.
The Run Settings Tab
Use the Run Settings tab to enter information about the run and control variables for the run. Once the run settings are set, you can store them for later use by selecting the Save Settings to File... and Load Settings from File... buttons.
Quick Start 3-5
Page 42
Title
Enter a descriptive title for the sample. This title will appear as a heading on various windows to identify this sample. The maximum number of characters allowed is 79.
Scheduling
Run Length (hh:mm).
Enter a run length by selecting the number of hours and minutes you want to collect data for this run. The maximum run length is 1 hour and 30 minutes. Remember, to stop the run at any time, use the button on the toolbar or Run|Stop Data Collection.
Start at Time
If you do not want the run to start immediately, you can schedule the run to start at a specific time.
Run Only Once or Run Continuous
Sampling can be set to run once or to run continuously (specifying a delay time between runs of 1 minute to 24 hours). When sampling is set up to run continuous, a new filename is opened for each run. The new filename is based on the current filename. For example, if the original filename is ModelCar.3090d, the new filename will be ModelCar_0.3090d the first time the sample is repeated and ModelCar_1.3090d the second time it is repeated and so on.
Enable External Trigger
The checkbox for Enable External Trigger, lets you tell the software that you want to start sampling using an external trigger to start a run remotely. This can be useful when different instruments need to be started at the same time.
A trigger input (on the back of the instrument) consists of bringing the + input to a logic low. This can be done using:
A potential-free contact closure (a toggle switch) that connects the +
input to the – input, which is at ground (GND).
A logic circuit to drive the + input low.
Data Collection and Display
Data Type
Select the type of data you would like to collect from the instrument. Select either Concentration data or Electrometer Current.
Zero-Check (Electronic Air Filter Mode)
Check this box if you would like to perform a run which has the same effect as a filter installed on the inlet. During a zero-check run the negative charger is set to its maximum current and the positive charger is turned off.
3-6 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 43
Note that this checkbox always defaults to “off”. If using it for consecutive runs, it must be checked each time.
Hide Invalid Values
Check this box to have the software hide any invalid data values in its graphs and tables. An invalid data value is one that is received by the instrument under a serious or fatal error condition. See Table 6-3 in
Chapter 6 for a detailed list of fatal errors.
Dilution Factor
To correct for a dilution factor to match measurement conditions, you can add a dilution factor here and turn it on/off with the checkbox. For example, if you used a 100:1 diluter before sampling into the instrument, the concentration can be corrected back to the original concentration using a dilution factor of 100. This will multiply the concentrations measured by the instrument by a factor of 100 when shown on graphs and tables. The original data is not changed and the dilution factor can always be turned back off.
Auto-Export
Select the Export Data at End of Run checkbox to automatically export data when the run is complete. Refer to Chapter 5 for a complete description of how to export data.
Instrument Matrix
The software displays the currently configured matrix on the instrument in the combo box. To select an alternate matrix, use the drop-down arrow. DEFAULT refers to the factory default matrix; “SOOT” and “COMPACT” are other pre-defined matrices which are installed with the software. When you select a matrix other than the one currently configured on the instrument, that matrix is uploaded to the instrument by selecting the Apply or OK button (to close the dialog).
Comment
The Comment box allows you to store comments about the sample run. The maximum number of characters is 255.
This information is stored with the sample data in the file and may be displayed when the Properties dialog box is opened for this file. It is also included in the exported data.
Quick Start 3-7
Page 44
Viewing Data
Save Settings to File.../Load Settings to File...
Use these two buttons to save the run settings you have set up and recall them for later use.
To save the current settings, click the Save These Settings... button and enter a filename (and, if desired, a different directory/folder/drive) and then select Save.
To recall settings you have saved, click the Load Settings from a File... button, highlight the filename you want to recall (after navigating to the appropriate folder if necessary) and selecting Open.
View data in a table or graphs while the data is being collected and change how data is viewed after sampling ends.
Note: To view the data on the LCD panel of the EEPS spectrometer, refer
to Chapter 4.
Change scales, time resolution, weighting, and other parameters, as well as play back the entire run.
Data can be viewed in the graphs either logarithmically or linearly and at different time resolutions. The particle data in the 2-dimensional graph represents an individual histogram at a selected time.
The following procedure provides an example of how to change how data is viewed for an existing run. Refer to Chapter 5 for a complete description.
1. Open an existing run file using File|Open or by pressing on the toolbar. Figure 3-6 shows sample data.
3-8 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 45
Figure 3-6 Sample Data
2. To toggle the graph scale between logarithmic and linear display modes, use the and toolbar buttons.
3. To toggle the display units between number concentration and dN/dlogDp, use the and toolbar buttons.
4. Using the mouse, point to a location in the Run View (contour graph) and click the left mouse button. This selects a time window during the run and this individual histogram of data is then displayed in the 2­dimensional Particle Histogram Graph window and in the Particle Table window. This subset of data is also highlighted in the 3D Graph window. Each time a new time window is selected, the histogram, table and 3D graph windows are updated. In the Total Concentration graph, the time window is identified with a crosshair annotation.
Quick Start 3-9
Page 46
5. To view the data at a different time resolution, select a different averaging interval using the drop-down style selector on the toolbar (see figure). Depending on the length of the run, you can select an averaging interval from .1 second to 60 minutes.
6. To select a maximum value for the concentration range or have the software auto­scale the data, use the second drop-down style selector on the toolbar (see figure).
7. To display data other than concentration in the 2D particle histogram graph, click the right mouse button in that window to display the popup menu (Figure 3-7). Select Weight|Mass. Notice the change in how data is displayed. Use Weight|Number to change the graph to display the number weighting.
Figure 3-7
Popup Menu for the Histogram Window
3-10 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 47
Status Bar
Concentration
Display
Total Mass
(3090AK only)
Particle Data
Graph
Menu
C H A P T E R 4
LCD Display Operation
This chapter provides information on the features and functions of the LCD display and the control knob. Refer to Chapter 5 for detailed operation information of the Engine Exhaust Particle Sizer (EEPS) software program
Use the control knob and LCD display to view instrument status, start and stop data display, read diagnostics, and view user settings.
To use the control knob, place your finger in the detent on the front of the knob and turn. Notice how items on the LCD display are highlighted (dark box around light text). To select a highlighted item, press the control knob in and release.
LCD Display Items
Figure 4-1 shows the LCD display while a sample is being taken. There are four unique regions of the display. Each is described below.
Figure 4-1
LCD Display Regions
4-1
Page 48
Menu Option
Description
Cursor
View individual particle channels.
Menu
View user settings and diagnostics.
Pause/Resume
Start and stop data display.
Setting
Options
Description
Y-Axis Setting
Autoscale; 1E4 to 1E9
Use this option to select the maximum displayed scale for the Y-Axis.
(1E4 can only be selected in Linear mode)
Y-Axis Scale
Linear/Log 10
Use this option to change the Y-Axis scale between Linear and Log 10 mode.
Flows
Off/On
Use this option to turn all flows on or off. When Off is selected, the status bar
displays “Flows and Chargers Turned
OFF.” There will be no draw at the sample inlet. The chargers are turned off to protect the electrodes from contamination.
Status Bar
The Status Bar indicates instrument status. Possible messages include: Warming Up, Normal, etc. Refer to Chapter 6 for a complete list of messages and actions you should take.
Menu Options
There are three menu options.
Cursor
Select CURSOR to view information about a particular particle size value in the graph. To view a specific value, use the control knob to highlight CURSOR then press the knob. A bar will appear on the graph and data for that size channel will be displayed in the concentration display region (see below). Rotate the control knob to view data to the left or right. To stop viewing data with the cursor, press the control knob again and turn the cursor to select another option from the display.
Menu
Select MENU to view user setting and diagnostics information. To view these items, highlight MENU with the control knob and press the knob. A menu appears for User Settings and Diagnostics. Use the control knob to highlight the item you want and press the control knob.
User Settings
The user settings (Table 4-1) let you change display settings and turn on/off the pumps or chargers and zero the electrometer. The pumps can be turned off to reduce noise and save power when the instrument is not being used, but still powered on.
Table 4-1 User Settings
4-2 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 49
Setting
Options
Description
Chargers
Off/On
Use this option to turn both chargers on or off. When Off is selected, the status bar displays “Chargers Turned OFF.”
Column Heater
Off/On
Turns the column heater on or off. Turn the heater on when the instrument is in a cool (i.e., air conditioned) environment and sampling from a warm, humid environment to prevent condensation that could cause the electrometers to become noisy. Normally, it should be turned off unless needed.
Display Brightness
0-100
Use this option to change the panel brightness and improve display visibility.
Note: On recently built units this
adjustment will have no effect.
Zero Electrometer
None
Use this option to calibrate the electrometer
offset (see “Troubleshooting, Baseline
Noise” in Chapter 6)
Parameter
Nominal Value(s)
Description
Sheath, Sample Flow
39.4 L/min, 8 L/min
Sheath flow through column and sample flow (note that sample flow plus extraction flow equals the flow into the instrument inlet).
Extraction, Charger Flow
2 L/min, 0.6 L/min
Extraction pump flow and charger sheath flow.
Temperature, Pressure
25.0 C,
700 to 1034 mBar
Temperature of the sheath air flow, in degree Celsius and absolute pressure inside the sizing column.
Negative Charger
35 nA -1000 to
-4200 V
Current of the negative charger, in nano ampere. Voltage level of the negative charger electrode, in volts.
Positive Charger
27 nA 1000 V to
4200 V
Current of the positive charger, in nano ampere. Voltage level of the positive charger electrode, in volts.
Column Voltages
85V, 470V, 1200V
Voltage on center electrode sections, top to bottom, in volts.
Analog Input Ch 1, Ch 2
0 to 10.0 V,
0 to 10.0 V
Voltage level on the analog input channel 1 and 2, in volts.
MCU Firmware Version
2.xx
Version of micro controller firmware.
DSP Firmware Version
2.xx
Version of digital signal processor firmware.
Error Message
None
Select with cursor to see a list with all error messages.
The diagnostics information (Table 4-2) is read-only, and is available to let you quickly see if any of the important values are out of range.
Table 4-2 Diagnostics Information
Instrument Operation 4-3
Page 50
Pause/Resume
Select Pause/Resume to start and stop real-time data display. The button changes from Pause to Resume each time it is selected. You must select Pause, before you can use the Cursor menu item, see below.
Concentration Display
While running a sample, you can view particle size data using the CURSOR (see above). As you move the CURSOR, the information in the Concentration Display shows the particle size and concentration values for the CURSOR location.
Particle Data Graph
A histogram of sampled particle size data is presented in the graph window as normalized concentration dN/dlogDp (#/cc) to allow easy comparison to other instruments. This particle concentration is normalized to the bin width. In this case it is calculated by concentration multiplied by 16, the number of channels per decade. Particle size is presented on a log scale with 32 equally sized bins in the size range of 5.6 to 560 nanometers. Other weightings, such as particle mass or surface area as well as concentration that are not normalized can be viewed using the EEPS software.
A yellow curve is shown superimposed on the particle size distribution to show the RMS lower noise limit of the instrument. For best results, the concentration should be above the noise curve. In addition, when the concentration is too high, a red box will appear on the end of each size channel that has been over-ranged. When this happens, the concentration should be reduced using additional dilution air into the EEPS spectrometer. Running the instrument near the upper concentration limit will require more frequent cleaning of the column. See Maximum Concentration and
Minimum Concentration in the Software chapter for more information.
4-4 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 51
C H A P T E R 5
Software Operation
This chapter provides specific information for the Model 3090 Engine Exhaust Particle Sizer (EEPS) software program including:
Starting the program Opening an existing file Displaying particle histograms at different times during a run Changing how data is viewed Using a different matrix Playing back data collection Selecting a hotspot Deleting data Zooming in and out on data Printing information displayed on the desktop Exporting data Arranging open windows and changing the layout Quitting the program Description of all program menus and menu items
If you have read Chapter 3, you will already be familiar with some of this information.
5-1
Page 52
Start the Engine Ex h aust Particle Sizer Program
To start the program, proceed as follows:
From the Windows desktop, press the Start menu and select Programs|TSI|Engine Exhaust Particle Sizer.
The Engine Exhaust Particle Sizer desktop appears as shown in Figure 5-1.
5-2 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Figure 5-1 Engine Exhaust Particle Sizer Desktop
Page 53
Open an Existing File
1. Select File|Open or on the toolbar.
2. When the Open file dialog box appears, it shows the files available in the default file location. The default file location is Program Files/TSI/Engine Exhaust Particle Sizer. Example files are included with the program and are installed in the Engine Exhaust Particle Sizer software folder when the program is installed.
Use the mouse to highlight the file you want to open. If you store data files in a location other than the default Engine Exhaust Particle Sizer folder, browse for the drive/directory where the data files are stored before you click Open.
3. Click Open. The program remembers the windows that were open on the desktop when you last closed the file and reopens them when you access the file again.
(continued on next page)
Software Operation 5-3
Page 54
Selecting a Particle Histogr a m to Display
To select a time at which to display a particle histogram during a run, click on the “contour graph” or Run View window. The Run View window is the controlling view for the application and is always open on the desktop. Closing the Run View window closes the run file.
Use the mouse to click on a point in the Run View window. An annotation (see the dashed line in Figure 5-2) shows the selected time as time since the start of the run. The cursor corresponds to the Particle Histogram shown in the lower left corner.
5-4 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Figure 5-2
Selecting a Time Window
When you select a time window using the mouse, the corresponding histogram is displayed in the Concentration Histogram window and in the Particle Table window. The Total Concentration graph window draws an annotation at the selected time. The 3D Graph window displays the same data as in the Run View window with the selected time highlighted in a different color.
Page 55
To continue, select an adjacent time window using the arrow keys on the keyboard or use the mouse to point and click at another location on the graph.
Maximum Concentration
The maximum concentration limits for each channel are plotted as a red line in the Concentration Histogram window. When the values for a given run reach the maximum concentration level defined for a given particle size or channel, an annotation is displayed on the graph. In the graph in Figure 5-3, the bars which are marked with the annotation indicate channels in which the concentration measured by the instrument (using the current averaging interval) reached or exceeded the defined maximum. The channels with a red box indicate channels where the concentration has been clipped. This affects the shape of the distribution considerably. If you see a graph with maximum concentration indicators, you should reduce the concentration sampled by the instrument. For a reference on the concentration limits see Figure B-4 in Appendix B.
Figure 5-3
Histogram Showing Clipping Due to Exceeding Maximum Concentration
Software Operation 5-5
Page 56
Minimum Concentration
Similarly, when the values shown in the Concentration Histogram window are small, it is useful to note the minimum concentration level defined for a given particle size or channel. The minimum concentration is shown for each channel in this graph as a yellow line. In the graph in Figure 5-4, this minimum concentration level is more apparent in a distribution where the concentration values measured by the instrument are relatively low. Note that this curve depends on the averaging time displayedthe longer the averaging time, the lower the concentration defined by the curve. This is because the electrometer noise is improved with longer averaging times. For a reference on the concentration limits see Figure B-4 in Appendix B.
Figure 5-4
Concentration Histogram Showing Minimum Concentration
Change How Data is Viewed
The data can be viewed in the following ways:
Changing the Time resolution Changing the Concentration scale Selecting a Weight (2D histogram) Selecting the Display units
5-6 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 57
To view data at different time resolutions, select an averaging interval in the average interval drop-down selector on the toolbar (see figure). Depending on the length of the run, you may select averaging intervals from .1 second to 60 minutes.
To change the concentration scale of the data displayed in the graphs, open the Concentration Scale dialog (Figure 5-5) by selecting
Format|Concentration Scaleor by selecting Concentration Scale
from the right-click popup menu on the Run View. This dialog lets you select linear or log scale and allows you to select the minimum and maximum concentration range for the data.
Figure 5-5 Concentration Dialog Box
Changing the Time Resolution
Changing the Concentration Scale
If the concentration scale is set to “Auto Scale,” the software determines
the minimum and maximum concentration values for the current average interval. To select a concentration range, select either the “Min”, “Max” or
the “Min/Max” radio button. The box below “Min Concentration”, “Max Concentration” or both will become enabled and you may enter a value,
where the Minimum Concentration is equal to or greater than zero and less than the Maximum Concentration.
To change the scale between log and linear mode, press the or buttons on the toolbar.
Software Operation 5-7
Page 58
Finally, to select the maximum concentration value, use the drop-down Concentration Scale selector on the toolbar (see below examples for linear and log).
5-8 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 59
Changing the Weighting in the 2D Histogram
To view data in the 2D histogram graph using different weighting, proceed as follows.
1. Open a Model 3090 Engine Exhaust Particle Sizer data file.
2. Make the concentration histogram window active.
3. Right-click the mouse to display the popup menu (Figure 5-6).
Figure 5-6
Popup Menu for the Concentration Histogram Window
4. Use the mouse to select Weight|Mass. Notice the change in how data is displayed. (A checkmark next to one of the menu items means that is the selected weight option.) The Y-axis title on the graph will display the current weighting also.
Note: Important additional information on how data can be viewed is found
later in this chapter “Software Menus.”
Changing the Display Units
To view data as number concentration or as dN/dlogDp, use the toolbar buttons or . Data in all graphs and tables will display in the
selected units. Particle concentration is normalized to the bin width by multiplying by 16, the number of channels per decade.
Software Operation 5-9
Page 60
Using an Alternat ive Inversion Matrix
Chapter 3 describes how to select an alternate matrix to upload to the
instrument before starting a new run. You may also select a different matrix to apply to the raw data in an existing file.
Note: If you have a data file which does not contain the raw data values
this option is not available. The data file must have been created using EEPS Firmware version 3.05 or later and EEPS Software version 3.1.0 or later.
Open the Run|Properties dialog for an existing data file. You may select a different matrix to apply to your raw data set using the combo box in the Instrument Matrix section. Select the Apply button to recalculate the number concentration values. The data in the graphs and tables will update. You may then save your file as a different file name or overwrite the file with these changes when you close the file.
The data inversion in the Model 3090 uses a matrix to convert the measured electrometer currents to size and concentration data. The DEFAULT matrix is based upon a combination of theoretical and experimental data. However, variation in particle shape can affect how the particles are charged within a unipolar charger which affects how they are sized. In general, agglomerates acquire a higher net charge than compact, near spherical particles.
To address differences in charging characteristics, two additional particle morphology-specific inversion matrices are included within EEPS software that result in better correlation between EEPS spectrometer and SMPS spectrometer data. Both matrices were developed by empirically calibrating the EEPS spectrometer to SMPS spectrometer measurements.
SOOT.matrix is designed to specifically change how soot particles, typically chain aggregates, in the accumulation mode (~>100nm) are sized. When measuring engine emissions (from constant volume sampling (CVS) tunnel or directly from the exhaust using partial flow dilution) it is recommended that the soot matrix be used.
COMPACT.matrix is designed to be used when measuring aerosols composed of compact, nearly spherical, particles. The compact matrix is useful for laboratory studies in which it is known (or assumed) that the sample aerosol is composed of compact particles such as atomized oil droplets, salt solution residue, or polystyrene latex spheres (PSL).
5-10 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 61
Matrix File Format
A matrix file is a file with the extension “.matrix” and consists of a header string followed by 22 rows x 33 columns of floating point values. The header string “TSI 3090/3091 Matrix” identifies it as a matrix file. Values are separated by either a comma or a tab character. In addition to the default matrix, TSI currently provides the matrix file “SOOT.matrix” and “COMPACT.matrix”. You may use your own custom matrix file provided it conforms to this format.
Be aware that only the name of the matrix used is stored in the data file. The EEPS software will display all *.matrix files found in its installation directory in the instrument matrix combo box of the Run|Properties dialog. It is up to you to manage these files, and ensure that the file names are unique. The maximum length of a matrix name (filename only, excluding the .matrix extension) is 20 characters. The matrix name “Default” is reserved and refers to the factory default matrix already saved in the instrument memory.
Playback (Review) a Sample R un
Review data collection for the run (or portions of the run) as follows:
1. Select Playback|Play or on the toolbar. The Playback Options box shown in Figure 5-7 appears.
Figure 5-7
Playback Setup Dialog Box
2. Select the playback time, update rate, and whether you want the run playback displayed once or continuously. Playback begins immediately when you select Play Once or Play Continuous. For better viewing, it may be necessary to maximize the window of particular interest to you.
Software Operation 5-11
Page 62
3. If you select Full Speed for Update Rate, the data is played back at full speed in the 3D graph view only. All other views are minimized until playback is complete or you stop playback.
4. If you select an update rate of 2x/second or 1x/second, all views are updated at that rate. An indicator arrow in the Run View window shows the current time of the playback.
5. Pause the playback at any time by selecting Playback|Pause. Select Playback|Play to continue playback where you left off. Stop playback at any time by selecting Playback|Stop. When you stop playback, you must restart it from the Playback dialog again. Use Playback|Pause and Playback|Play if you want to stop and start playback.
Control playback by using icons on the toolbar. The icons perform the following operations:
Stop. Stop playback. Pause. Pause playback.
Start. Start playback or restart playback.
Select a Data Hot Spot
When a concentration histogram window or the total concentration graph is active on the desktop, find the values of any data point as follows:
1. Position the pointer on the bar, line, or peak (depending on the type of graph that is active). The pointer becomes a pointing hand.
2. Press the left mouse button to display the corresponding values. Figure 5-8, shows how the data values would be displayed with the crosshair annotation at the selected data point.
Tip: After you have a data hot spot selected, use the “>“ and “<“ or arrow keys to move the crosshair cursor right or left (from one value to the next), or grab the vertical crosshair annotation and drag it to another location. Refer to the end of this chapter for a list of other keys or key combinations for navigating and performing operations without using the mouse.
5-12 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 63
Figure 5-8
Use the Mouse to Directly View Data Values
Delete Data
To redo a run or delete data for a run just taken, on the toolbar select Run|Delete or . A verification message is displayed.
You can only delete data for the run you just collected. Once you close this file, you cannot delete the data from this file using this method the next time this file is opened using File|Open or . If you need to delete a saved file, use a tool such as Explorer to delete the file.
Zoom In and Out on Data in a Graph
Zoom in on data displayed in a graph as follows:
1. Use the mouse to position the cursor (pointer) on the graph surface, at one corner of the graph area you want enlarged.
2. Press down on the left mouse button and drag the mouse to the opposite corner of the data you want enlarged.
3. Release the left mouse button. The area you selected is enlarged.
4. Continue to zoom further by repeating steps 1 through 3.
To unzoom, select <Z> or Format|Undo Zoom or on the toolbar. The enlarged area is returned to normal view.
Software Operation 5-13
Page 64
Print Information Displaye d o n the Desktop
Print the information displayed in the active window on the desktop as follows (a printer must be properly installed):
1. With a graph or table window active on the desktop, select File|Print Preview or on the toolbar.
2. Review that what you see in the print preview window is what you want to print (an example is shown in Figure 5-9) and select Print from the Preview box. The contents of the window are sent to your printer.
3. You can also select Print or Print Preview from the popup menu of each window.
To immediately print a window without previewing it, select File|Print or
from the toolbar or press <Ctrl><P>.
5-14 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Figure 5-9
Print Preview
Page 65
Export Data to a File
Data from a Model 3090 EEPS file can be exported for use in another program such as Microsoft Excel® or Microsoft Access® programs. Data can be exported either manually or automatically.
To Manually Export Data
To manually export data:
1. Open the data file by selecting File|Open or or <Ctrl><O>.
2. Select File|Export. An Export Data Options dialog box appears as shown in Figure 5-10.
®
Excel and Access are trademarks of Microsoft Corporation.
Figure 5-10
Export Data Options Dialog Box
Software Operation 5-15
Page 66
Parameter
Description
Data Types
Concentration
The total number of particles per unit volume of air sampled (i.e., number concentration expressed as #/cm3). Concentration is the primary measurement of the Model 3090 EEPS system.
Volume
Volume represents the total volume of the particles per unit volume of air sampled (i.e., volume concentration expressed as nm3/cm3). The volume concentration calculation assumes that all the particles are perfect spheres.
Surface
Surface represents the total surface area of the particles per unit volume of air sampled (i.e., surface area concentration expressed as nm2/cm3). The surface area concentration calculation assumes that all the particles are perfect spheres.
Mass
Mass represents the total mass of the particles per unit volume of air sampled (i.e., mass concentration expressed as µg/cm3). The mass concentration calculation assumes that all the particles are perfect spheres with the density defined in the particle table for each channel.
Raw Data (Instrument Record)
Note: Valid only with 3090 instruments with firmware newer than 3.05.
Each data set collected has attached to it the raw electrometer data. This data can be exported and reviewed.
Units
Normalized (dW/dlogDp)
Concentration, surface, volume, mass values exported are normalized to 16 channels per decade.
Number
Actual (non-normalized) values are exported.
Total Concentration
Total concentration on all channels.
Electrometer Current
If the run file contains data which is electrometer current (see run settings), this option may be selected.
Analog Input
Input voltage for Analog Input Channel 1 and 2 in Volt.
Sample Temp
Sample temperature recorded during the run.
Pressure
Sample pressure recorded during the run.
Min-Max Concentration Limits
Export the minimum and maximum concentration limits for each channel. (Number or dN/dlogDp units apply.)
Analog Output
Exports the Model 3090 EEPS Analog Output settings that were in effect when the EEPS data was collected. Only valid for Model 3090 with Analog Out Option.
3. Select the type of data you want to export, the data range, the type, averaging interval, and format of the file to which you want to export the data. Refer to Table 5-1 for a description of the parameters.
Table 5-1 Export Data Options
5-16 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 67
Parameter
Description
Time Range and Resolution
From/To
Starting time in the run to begin exporting data.
Use Averaging Interval
Just as the averaging interval can be when viewing the data, it can be set for the exported data independently.
Display Time as:
Format of the time corresponding to each data point in the exported output.
File Type
Excel
File extension will be either .xls or .csv (comma­separated-values) to be read by Excel®.
Text
Output will be written to a text (.txt) file.
Delimiter
Delimiter used in output.
Output Filename
Name of output file for exported data.
4. Verify the default file name (and path) or press the Save As button to change the filename and/or location if you want to save the file to another location or provide a new filename. The extension is automatically selected when you chose a file type for exporting.
5. Press OK.
Figure 5-11 illustrates how an exported *.xls file is formatted for a Model 3090 Engine Exhaust Particle Sizer (EEPS) spectrometer when Concentration is selected as Data Types for exporting.
Software Operation 5-17
Page 68
Figure 5-11
Sample Export File
To Automatically Export Data
To export data automatically at the end of the run to a file for use in another program (such as Microsoft Excel® spreadsheet software):
1. When setting up the properties for the run, select Auto-Export on the Run Settings tab. This activates the Export Options button on the dialog. When you press the Export Options button, the Export Data Options dialog opens (see Figure 5-10 above).
2. Select the type of data you want to export, the data range, and the file format in which you want to export the data.
3. Verify the default file name (and path) or press the Save As button and change the filename and/or location if you want to save the file to another location or provide a new filename. The extension is automatically selected when you chose a file type for exporting.
4. Press OK. Data will be exported to the file at the end of the run.
5-18 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 69
Arrange Open Windows/Change L a yout
When several windows are open on the desktop, it is helpful to arrange them for easier viewing.
There are two ways to arrange windows:
The traditional Windows method.
To arrange windows, select an item from the Window menu. Windows on your desktop can be arranged horizontally, vertically, or cascaded. Refer to your Windows documentation for examples and more information.
Using the EEPS spectrometer layout function.
Select one of the layout icons (5-pane layout), (4-pane layout),
(3-pane layout), or (2-pane layout).
Each icon changes the number and way in which windows are displayed on the desktop. (These layout functions can also be selected from the Window menu.)
All of the program windows that are currently open on the desktop are listed at the bottom of the Window menu under the Close All item. To make a window active (and bring it to the front of all windows), select the name of the window from the list of open windows.
Quit the Program
To end the program, select File|Exit. All windows and files open on the desktop are closed. If you have made changes to a file and have not yet saved it, you will be prompted to do so.
Software Menus
This section describes the software functions available for the EEPS spectrometer software. There are two types of menus: desktop menus and context-sensitive or popup menus.
Desktop menus are those menus listed at the top of the Engine
Exhaust Particle Sizer desktop. They include: File, Run, Playback, Format, View, Window, and Help.
Popup menus are those menus that appear when you click the right
mouse button when the cursor is positioned in an active window.
Software Operation 5-19
Page 70
Nearly all menu items are available from both a desktop menu and a popup menu, though a few items are only available in one or the other. This section presents the items of the desktop menus first and then describes the menu items that are available only through the popup menus.
In addition to these menus, a toolbar, located just beneath the desktop menus, provides shortcuts to the most commonly used menu functions.
Shortcut keys (keyboard keystrokes that do not require the use of the mouse) are also available. They are described at the end of this section.
Desktop Menus
Note: All menus and the menu items are described below. Depending on
the operation you are currently performing and the window that is active on the desktop, the menu may appear different than shown; that is, some menu items may not be available.
File Menu
The items of the File menu are used to open, save, and recall files and perform other program operations.
Figure 5-12
File Menu
5-20 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 71
New
Select File|New or collect sample data.
After you select New, you are prompted to enter a filename. Accept the default filename or enter any filename you chose then select OK.
A Run View window opens on the desktop with the filename you entered. If other windows were open at the time you last closed a file, these windows also open on the desktop. All graphs and tables are initially empty. You are ready to start collecting data. Refer to “How to Perform a Sample Run” for a complete description.
or <Ctrl><N> to open a new file and prepare to
Open
Select File|Open or or <Ctrl><O> to open an existing file. By default, the program will look for files in the same directory location as the last file you opened or in the same directory as the EEPS program if this is the first time running the program. All files are automatically given the appropriate filename extension (.3090d). If you store data files in another directory or on another drive, you must first display the file pathname in the Open dialog box before you can select and open it.
When the file opens, a window (or windows) will open on the desktop displaying the data in this file.
Multiple files can be open on the desktop at the same time. Only one window, however, is the active window.
Close
Select File|Close or or <Ctrl><W> to close a file (and all the windows associated with it). If there are windows open on the desktop from more than one file, Close will close only those windows associated with the file whose window is currently active. If you attempt to close a file that has been changed but not saved, you will be prompted to save the changes before closing the file.
Since the Run View is the controlling view for a file, closing this window using the button in the upper right corner also closes the file.
Software Operation 5-21
Page 72
Save As
Select File|Save As or to save data to a new filename. (The file contents are duplicated to the new filename. To delete the original filename later, use Windows® Explorer® to do so.)
After you select Save As, select a drive/directory in the file dialog box. The same filename can be used if you save the file to another drive/directory, but to save the file to the same directory, you must give it a new name.
Do not add the filename extension. It is added automatically when you select Save As.
Export
Select File|Export to export data for use in another program. Data is
exported in a delimited text file. Refer to “Export Data to a File” earlier in
this chapter for information.
Print
Select File|Print or desktop in a report format. To preview the output before printing it, select File|Print Preview.
Refer to your Windows® documentation for information about the Print dialog box.
or <Ctrl><P> to print the active window on the
Print Preview
To avoid printing something you do not want, select File|Print Preview or
to see what your printed output will look like before selecting Print.
Print Setup…
Select File|Print Setup to set up the printer for printing. Refer to your Windows® documentation and your printer’s documentation for information about setting the printer parameters.
Properties
Select File|Properties to see the run properties for the current run. The parameters that can be set in the Properties dialog for a new run are described earlier in this chapter. When viewing the properties for an existing run, only the title and comment fields may be modified and you may also select or de-select “Hide Invalid Data Values”.
5-22 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
List of Recently Accessed Files
Between the Print Setup and Exit menu items is a list of the most recently accessed data files (a maximum of four files is displayed). These are accessed from the menu by selecting File|1, 2, 3, or 4.
Page 73
The list provides a shortcut to these files so you can bypass the Open command. To open one of the files listed, use the mouse to highlight it and click the left mouse button. The file opens on the desktop.
Exit
Select File|Exit to end the program.
Run Menu
The items available under the Run menu are used to start and stop data collection.
Figure 5-13
Run Menu
Start Data Collection
With a new file open on the desktop, select Run|Start Data Collection or
or <F10> to begin collecting data. Data is collected according to the
parameters set in the Properties dialog (see earlier in this chapter).
Once you select Start Data Collection, data collection begins immediately or at the time selected in the Parameters dialog box. As data is collected, it is displayed in the open windows.
Stop Data Collection
Select Run|Stop Data Collection or to stop collecting sample data immediately. When you select this item, sampling stops and data already collected is saved in the file.
Delete
Select Run|Delete or to delete data. This is useful if you need to stop and restart a run. You can only delete data for a new run. It cannot be used to delete data once the file is closed. To delete data for an existing run, use Explorer® or a similar tool to delete the file.
Properties
This menu item provides a shortcut to the same Properties dialog box listed under the File menu. Refer to that description or see the description of the Properties dialog box earlier in this chapter.
Software Operation 5-23
Page 74
Playback Menu
The Playback menu lets you display the sample data series, so that many timed samples can be viewed quickly, visually identifying trends or special events from a large list of samples.
The playback feature was described earlier in this chapter under “Playback
(Review) a Sample Run.” Refer to that description.
Figure 5-14 Playback Menu
Format Menu
The menu items of the Format menu let you control how information is presented in tables and graphs.
Figure 5-15
Format Menu
5-24 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 75
Option/Parameter
Description
Linear
Data plotted on a linear scale.
Log
Data plotted using a logarithmic scale.
Min
Select this radio button to specify a minimum value for the concentration range. When selected, enter a minimum value in the edit box below. This value must be greater than or equal to 0.
Max
Select this radio button to set a maximum value for the concentration range. When selected, enter a maximum value in the edit box below. Enter a value, m, such that 100 <= m <= 2e8.
Min/Max
Select this radio button to set both the minimum and maximum values for the concentration range.
Auto Scale
Select to let the software automatically determine the minimum and maximum concentration values. Auto­scale is the default.
Min Concentration
Enter the minimum concentration value here. When the graphs are auto-scaled, the value shown in this box (grayed) is the minimum value as calculated by the software.
Max Concentration
Enter the maximum concentration value here. When the graphs are auto-scaled, the value shown in this box (grayed) is the maximum value as calculated by the software.
Concentration Scale
Select Concentration Scale to open the Scale dialog box (Figure 5-16).
Figure 5-16
Concentration Scale Dialog Box
Use this dialog box to set options and parameters as described in Table 5-2. Make your changes and select OK.
Table 5-2 Options/Parameters of the Concentration Scale Dialog Box
Software Operation 5-25
Page 76
Color
Select Format|Color or to change the colors used to display items in the active window.
When Format|Color is selected, the Graph Color or Table Color window opens depending on the active window on the desktop (Figure 5-17 shows both). These windows include a preview screen, a drop down list and a color palette. From the drop down list, select the name of the item you want to modify. The current color for that item is indicated in the color palette. Use the mouse to point to the new color for that item and click the left mouse button to select it. The preview screen then displays the item in the new color.
(A) (B)
Figure 5-17
(A) The Graph Color Dialog Box, (B) The Table Color Dialog Box
Select items from the drop down list and colors as desired. When finished press OK. The colors you selected will be used from that point on in all graphs and/or tables.
5-26 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 77
The items selectable for windows containing graphs are:
Window Background Plot Area Background Labels Axis Data View Boundaries Data Hotspot Lines
The items selectable for windows containing tables are:
Window Background Cell Text Grid Color Cell Background Fixed Area Text
The Color dialog box for the 3D graph is slightly different as shown in Figure 5-18.
Figure 5-18
The Color Dialog Box for the 3D Graph
Software Operation 5-27
Page 78
Use this dialog to change the color of the following items:
Data Highlighted Data Window Background Graph Background Labels Axis
Font Style
Select the Format|Font|Style… menu to change the style of the text in all graphs or tables. See Figure 5-19.
Select any font and font style available on your computer. The preview box displays what the text will look like before you implement it by selecting
OK.
Figure 5-19
Font Dialog Box
Font Size
To change the font size, select Format|Font|Small, Medium, or Large (the default is Medium). To change the text size without using the menus,
select the icon from the toolbar. Each time you select the icon the text size changes to the next text size. (Rotation is from small to medium to large.) The font size for text can be changed in all views except the Run View where only the font style can be changed.
Undo Zoom
Select the Format|Undo Zoom, select the Undo Zoom icon or type Z in the graph that is currently zoomed to return the graph to its normal viewing size. Refer to the section “Zoom In and Out on Data in a Graph earlier in this chapter.
5-28 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 79
View Menu
The View menu lets you select the view windows you may open on the desktop. A checkmark next to an item indicates the item is already open. If it is not visible, look under the Windows menu to find the item and make it the active window.
Figure 5-20
View Menu
3D Graph
Select the 3D Graph menu item when you want the 3D graph window open on the desktop (Figure 5-21).
Figure 5-21
3D Graph View
Software Operation 5-29
Page 80
Particle Table
Select the Particle Table menu item when you want the Particle Table window open on the desktop (Figure 5-22).
Figure 5-22 Particle Table Window
Particle Histogram
Select the Particle Histogram menu item when you want the Particle Histogram window open on the desktop (Figure 5-23).
Figure 5-23
Particle Histogram Window
5-30 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 81
Total Concentration
Select the Total Concentration menu item when you want the Total Concentration window open on the desktop (Figure 5-24).
Figure 5-24 Total Concentration Window
Software Operation 5-31
Page 82
Instrument Status <Ctrl><I>
This menu item is available at all times. You do not need to have a data file open in order to view instrument status. When you select the Instrument Status menu item, a dialog window opens detailing the status of the instrument as in Figure 5-25. This dialog will attempt to connect to the instrument if necessary. The Instrument Status dialog reads instrument status values once per second and updates the values for each item shown. Table 5-3 describes each of the parameters and buttons on this dialog.
5-32 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Figure 5-25 Instrument Status
Page 83
Parameter
Description
Sheath Flow
Volumetric flow rate of the particle free sheath air flow in liter per minutes.
Sample Flow
Volumetric flow rate of the sample flow, in liter per minutes. Sample flow plus extraction flow equals the aerosol inlet flow.
Extraction Flow
Volumetric flow rate of the extraction flow in liter per minutes.
Charger Flow
Volumetric flow rate of the particle free charger sheath air in liter per minutes.
Internal Pressure
Absolute pressure in the sizing column in millibar. Used for the volumetric flow calculation.
Column Voltage 1-3.
Voltage on the center electrode sections in Volt. Column Voltage 1 = Top Section Column Voltage 2 = Middle Section Column Voltage 3 = Bottom Section
Sheath Temp.
Temperature of the particle free sheath air flow in degree Celsius before it enters the sizing column.
Neg. Charger Current and Voltage
Ion current of the negative charger in nano Ampere. Voltage at the charger needle in Volts.
Pos. Charger Current and Voltage
Ion current of the positive charger in nano Ampere. Voltage at the charger needle in Volts.
External Inputs
Analog 1
Voltage of the external analog input Analog In CH1 in Volts.
Analog 2
Voltage of the external analog input Analog In CH2 in Volts.
Trigger channel 1
Status of the external Trigger IN (On or OFF)
External Outputs
Analog Out
Activates the Analog Output Settings Dialog. The Analog Out button is only displayed if the Model 3090 EEPS spectrometer has the analog out option.
Note: The Instrument Status Dialog is inactive while the Analog Out dialog is displayed.
Controls
Neg. Charger
Use this push button to turn on/off the negative charger.
Pos. Charger
Use this push button to turn on/off the positive charger.
Column Heater
Use this push button to turn on/off the column heater. Turn the heater on when the instrument is in a cool (i.e., air conditioned) environment and sampling from a warm, humid environment to prevent condensation that could cause the electrometers to become noisy. Normally, the heater should be turned off unless needed.
Flow Control
Use this push button to turn on/off the flow control.
Analyzer Voltages
Use this push button to turn on/off the analyzer voltages.
Table 5-3 Instrument Status Parameters
Software Operation 5-33
Page 84
Parameter
Description
Electrometers
Electrometer Offsets and RMS
Press this button to open a window which will display the offsets and RMS values for each of the electrometer channels (see Figure 5-26).
Zero the Electrometers
Pressing this button will start the process of zeroing the electrometers on the instrument (see Figure 5-27 to Figure 5-29).
more >> or <<less
Press this button to expand or contract the dialog. See Figure 5-30.
Figure 5-26
Electrometer Offsets and RMS Screen
5-34 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 85
When the Zero the Electrometers... button is pressed, the following message box is displayed (Figure 5-27). To continue with the zeroing procedure, select the Yes button. Refer to the "Troubleshooting" section in Chapter 6 which discusses this procedure and some possible reasons for doing a zero on the instrument.
Figure 5-27 Zero Electrometer Function Dialog Box
If you answer “Yes” to continue, the zeroing procedure will start. A
progress dialog will display while the zeroing is taking place. The zeroing takes approximately 45 seconds.
Figure 5-28
Zero Electrometers Progress Dialog Box
Software Operation 5-35
Page 86
When the zeroing is complete, you will see a message window like this:
Figure 5-29
Zero Electrometers Zeroing Complete
5-36 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Figure 5-30
Instrument Status Expanded
Page 87
When the instrument status dialog is expanded (>>more button is pressed), a graph showing current particle concentration similar to the instrument front panel is displayed. Control the data displayed in the graph in the following ways:
Press the Start Sampling button to begin sampling. The graph will
begin updates of current particle concentration once per second.
Linear or Log button: Click on each of the buttons to toggle between
the two y-axis scales.
dN/dlogDp or Number button: Click on each of the buttons to toggle
between the two display units.
Concentration or Electrometer Current button: Click on each of the
buttons to toggle between these data types sampled from the instrument.
Press the Stop Sampling button to stop sampling and the instrument
status display will resume updates to the status values on the left side of the dialog.
Use the <<less button to contract the dialog back to its original size.
Analog Out
The Analog Output Setup dialog displays the output voltage and the EEPS Model 3090 Analog Out settings. If the dialog is active, it shows the current 3090 settings. If the current document is an old data file with Analog Data included, the Properties | Analog Out Tab will display the Model 3090 EEPS Analog Out settings state at the time the EEPS data file was created. If the Analog Out dialog is accessed from the Instrument Status dialog, it will always be active and display/adjust the current Model 3090 settings. If the Model 3090 does not have the Analog Out option or the data file was taken with a Model 3090 without the Analog Out feature, the Analog Out dialog will not be available. The information entered is immediately sent to the Model 3090 when data entry is completed. For direct entry, this is after the Enter key, Esc key, or Tab key is pressed. Invalid entries will leave the previous value for that field unchanged.
Note: Cancel does not restore the Model 3090 EEPS spectrometer to any
previous state. The Model 3090 Instrument Analog Out settings have already been updated by the time Cancel is pressed.
Save these Settings… button. Brings up a file dialog box where all of
the values currently set will be saved in a Settings (*.set) file that can be used later to restore the Model 3090 settings back to the current settings.
Load Settings from a File… button. Brings up a file dialog box where
previous Model 3090 settings can be used to restore the Model 3090 settings back to a previous saved state. File type is *.set.
Software Operation 5-37
Page 88
Voltage Range
Select the voltage range for analog outputs. Either 0–5 volts or 0-10 volts.
lin/log
Select whether the channel will display the analog voltage response linearly proportional to concentration (lin) or with a logarithmic (Log) response characteristic.
ngeSelectioVoltageRan
MinConcMaxConc
MinConcncparticleCo
V(lin)
geSelecionVoltageRan
c)Log(MinConc)Log(MaxCon
)LN(MinConceConc)LN(particl
V(Log)
dN dN/dlogDp
Select whether the channel will display the analog voltage response “normalized” dN/dLog(Dp) or non-normalized (dN) over the selected particle size range. dLog(Dp) = Log(MaxConc) – Log(MinConc).
Min/Max Dp
Spin Control but values can be entered manually. The Maximum Value is 562.300, minimum is 5.624.
Power Dn
The Maximum Value is 10, minimum is 0.0000.
Weighting symbols. These are used to represent three predefined
values: Number(1), Surface(3.1416), and Volume(.5236).
Figure 5-31
Analog Out Setup
5-38 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 89
Weighting
The Maximum Value is 1e38, minimum is 0.0000. If 0.000, output will be 0.000.
Density
The Maximum Value is 1000, minimum is 0.0000. If 0.000, output will be 0.000.
Min/Max Conc.
Spin Control but values can be entered manually. The Maximum Value is 1e38, minimum is 0.000 lin or 0.001 log.
Avrg Time
Value is in integer seconds. The Maximum Value is 10, minimum is
1.
Analog Out (Volts)
Display of Scaled analog voltage out of the EEPS Model 3090 based on the selected channel settings.
Toolbars
Select View|Toolbars to display or hide the toolbars that appear on the desktop. Each toolbar is illustrated below:
Main Toolbar
Color and Font Toolbar
The Color and Font toolbar will be displayed if it was displayed the last time you ran the EEPS software. Select View|Toolbars|Color and Font Toolbar to display the color and font toolbar.
To view what each icon button does, position the cursor on the icon. A balloon appears to describe the function of the icon.
To move a toolbar, position the cursor on a gap between two icons and press and hold the left mouse button. As you move the mouse, the toolbar moves with it. When you release the mouse button, the toolbar remains where it is.
Software Operation 5-39
Page 90
Window Menu
Use the Window menu items (Figure 5-32) to select your window layout, close all windows, and open, close and arrange the windows on your desktop. Refer to your Windows® documentation, if necessary, for an example of what the cascade, tile horizontal, and tile vertical commands do.
Figure 5-32 Window Menu
5, 4, 3, and 2 Pane Layout
These menu items and toolbar buttons are intended to provide a shortcut to viewing and sizing each of the windows within the bounds of the desktop in a convenient layout. Each of these window layout selections correspond to toolbar buttons:
5 Pane Layout displays the Run View, Particle Table, Particle Histogram, Total Concentration Graph and the 3D graph. Each window is sized so that all five panes fit on the desktop.
4 Pane Layout displays the Run View, Particle Table, Particle Histogram, and the 3D graph. Each window is sized so that all four panes fit on the desktop.
3 Pane Layout displays the Run View, Particle Histogram, and 3D graph. Each window is sized so that all three panes fit on the desktop.
2 Pane Layout displays the Run View and Particle Histogram. Each window is sized so that both panes fit on the desktop.
5-40 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Close All
Closes all the windows on the desktop.
Page 91
Help Menu
The Help menu (Figure 5-33) provides access to information about the program.
Figure 5-33 Help Menu
Help Manual
Select Help Manual to open a PDF file of this manual. Adobe® Acrobat® Reader must be installed on your computer to view the manual. The PDF file of the manual is searchable and printable.
About EEPS Software
Select About EEPS Software to see the copyright statement for the program and view the version number of the software.
Context-Sensitive M enus
To access context-sensitive menus, also called “popup menus,” click the right mouse button when the cursor is in an active window. Each window has a popup menu as shown below. The table following each popup menu describes where to find a description of that menu item.
Run View Popup Menu
Figure 5-34
Run View Popup Menu
Software Operation 5-41
Page 92
Menu Item
For a description, see ...
Concentration Scale ...
Format Menu
Show Max Concentration Value
Below
Set Analog Scale
Below
Analog Input Labels...
Below
Font...
Format Menu
Copy
View Menu
Undo Zoom
Format Menu
Print
File Menu
Print Preview
File Menu
Properties
File Menu
Show Max Concentration Value
Select this option or press <Ctrl><M> in the Run View and the software will select and display the time window corresponding to the maximum concentration value recorded for the current averaging interval.
Set Analog Scale
Select Set Analog Scale to open a dialog box (Figure 5-35) that lets you enter new values for the minimum and maximum Y values for the analog input data. You may enter a new value for the Min Y and or the Max Y.
5-42 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Figure 5-35
Set Y Scale for Analog Graph Dialog Box
Page 93
Menu Item
For a description, see ...
Units
Below
Load Density Values...
Below
Save Density Values...
Below
Color...
Format Menu
Font...
Format Menu
Copy
View Menu
Print
File Menu
Print Preview
File Menu
Analog Input Labels
Select Analog Input Labels to open a dialog box (Figure 5-36) that lets you enter new labels for the analog inputs that are displayed in the analog graph in the lower section of the Run View window. Enter any text up to 47 characters.
Figure 5-36 Enter Analog Input Labels Dialog Box
Particle Table Popup Menu
Figure 5-37
Particle Table Popup Menu
Units
dNdlogDp or Concentration: select either of these two display units.
Software Operation 5-43
Page 94
Load Density Values...
Select Load Density Values to load previously saved density values from a text file. When you select this menu item, a Load Density File dialog box opens. Navigate to the file you want to use and select Open. The density values are read and displayed in the density column of the table. Density values are saved with the data file.
Save Density Values...
Select Save Density Values to save the current density values to a file for later use. When you select this menu item, a Save Density Values to a File dialog box opens. Enter a file name and select Save. The current values in the density column are written to this file.
Concentration Histogram and Total Concentration Popup Menus
The menu items for the popup menus for the Concentration Histogram and Total Concentration windows (Figure 5-38) are the same except Concentration Scale or View Boundaries cannot be selected from the Total Concentration Graph popup menu. This is because the data in the Total Concentration Graph window is always auto-scaled and represents totals for all channels. You can only change the scale type in the Total Concentration Graph window between logarithmic and linear using the toolbar button.
5-44 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
A B
Figure 5-38
Concentration Histogram (A) and Total Concentration (B) Popup Menus
Page 95
Menu Item
For a description, see ...
Concentration Scale...
Format Menu
Weight
Below
View Boundaries
Below
Graph Type
Below
Grid Lines
Below
Color...
Format Menu
Font...
Format Menu
Copy
View Menu
Print
File Menu
Print Preview
File Menu
View Units
Description
Number
Number represents the total number of particles per unit volume of air sampled (i.e., number concentration expressed as #/cm3). Number concentration is the primary measurement of the Model 3090 EEPS system. The sensor is sensitive to the number of particles in the aerosol sample as opposed to the particle mass, color, shape, composition or other characteristic. The distributions of diameter, surface area, volume or mass concentrations of the particles are calculated based on the particle number distribution.
The number concentration, dN, measured by the Model 3090 is the concentration of particles in a given channel. The normalized concentration dN/dlogDp is the number concentration of each channel multiplied by the number of channels per decade (16).
Surface
Surface represents the total surface area of the particles per unit volume of air sampled (i.e., surface area concentration expressed as nm2/cm3). The surface area concentration calculation assumes that all the particles are perfect spheres.
Surface area concentration is calculated by: dS = dN D
p
2
,
where Dp is the geometric midpoint of the particle size channel.
Weight
Use the Weight menu to choose how to “weight” the units that are displayed.
The options are number, surface, volume, and mass.
Figure 5-39
Weight Menu
The currently selected weighting is marked with a checkmark. Table 5-4 gives a description of each unit.
Table 5-4
Weight Options
Software Operation 5-45
Page 96
View Units
Description
Volume
Volume represents the total volume of the particles per unit volume of air sampled (i.e., volume concentration expressed as nm3/cm3). The volume concentration calculation assumes that all the particles are perfect spheres.
Volume concentration is calculated by: dV = dN (/6)D
p
3
,
where Dp is the geometric midpoint of the particle size channel.
Mass
Mass represents the total mass of the particles per unit volume of air sampled (i.e., mass concentration expressed as µg/cm3). The mass concentration calculation assumes that all the particles are perfect spheres with the density defined in the density column of the Particle table for each channel.
Mass concentration is calculated by: dM = dN (/6)D
p
3

where Dp is the geometric midpoint of the particle size channel and  is the density. This quantity is related to Volume concentration by the simple factor 
In many of the tables, the values for all four weightings are supplied in tabular format.
Note: Surface, Mass and Volume weighting calculations assume that all
the particles are spherical. Mass calculations use the density defined in the particle table for each channel. However, comparisons to methods that measure particle surface, volume or mass directly may give results different from those calculated by the software.
View Boundaries
Select View Boundaries to display the channel view boundaries in the graph window. After you select View Boundaries, drag view boundaries by positioning the cursor on the boundary, clicking the left mouse button, and dragging the boundary to a new location.
Graph Type
Select Graph Type to select the type of graph to display in the active window. The options are bar, line, or area.
Grid Lines
Select Grid Lines to select the lines for the graph in the active window. The options are horizontal, vertical, both or none.
5-46 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 97
Menu Item
For a description, see ...
Concentration Scale...
Format Menu
Graph Type
Below
Color...
Format Menu
Font...
Format Menu
Copy
View Menu
Print
File Menu
Print Preview
File Menu
3D Particle Concentration Popup Menu
Figure 5-40 3D Particle Concentration Popup Menu
Graph Type
Select Graph Type to select the type of graph to display in the active window. The options are 3D Bar or Waterfall.
Status Bar Icons
A status bar at the bottom of the desktop displays information about the current run as described below.
Software Operation 5-47
Page 98
Icon/Text
Description
Matrix Name
Name of the matrix currently configured on the instrument for new runs or, the name of the matrix used to calculate the concentration data for the current file. Refer to the description for the Run|Properties dialog.
Dilution Factor
The current dilution factor. This is set in the Run|Properties dialog.
Attention. Errors conditions were detected during the run. Move the mouse over the icon to view the tool tip which displays the error text. See Chapter 6 for a complete description of instrument errors that may be reported here.
Instrument status information. Move the mouse over the icon to view instrument status information saved for this run.
The run completed without errors.
Shortcut Keys
Key(s)
Action for EEPS Sensor
F1
Help; opens the help manual (PDF file)
F10
Start Data Collection
Ctrl C
Copy current view to clipboard
Ctrl I
Opens Instrument Status window
Ctrl N
Open a New document
Ctrl O
Open a document
Ctrl P
Print current view
P
Print current view
Ctrl S
Save As
Ctrl W
Close current file
Z
Undo Zoom in graph
Shortcut keys and key combinations can be used to perform operations using only the keyboard (no mouse required).
5-48 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Page 99
C a u t i o n
In general, if the instrument is functioning properly (sizing particles correctly compared to a standard, and without excessive noise), cleaning is not recommendedit can lead to leaks or damage if not done correctly.
C H A P T E R 6
General Care, Troubleshooting and Service
This chapter is written for a service technician with skill in both electronics and mechanics. Static preventative measures should be observed when handling any printed circuit board connectors.
The Model 3090 Engine Exhaust Particle Sizer (EEPS) spectrometer is designed for many hours of maintenance-free operation. This chapter describes basic care of the instrument, some troubleshooting suggestions, and provides information on where to go for service.
The Model 3090 is a complex instrument. Feel free to contact TSI prior to any maintenance procedure or for general questions regarding engine emission sampling. Read through the entire section and make sure all the tools necessary are available before starting to disassemble the instrument.
Basic Care of the E ngine Exhaus t Particle Sizer Spectrometer
Table 6-1 provides a general indication of the maintenance requirements. In applications where high concentrations may be generated, maintenance is required more often.
6-1
Page 100
Maintenance
Frequency
Remove the cyclone cover and clean interior surfaces (especially the wall opposite the inlet nozzle).
100 hours of operation or as needed Check/clean the charger electrodes.
100 hours of operation
Clean the column and outer electrode. Clean the high voltage column and sensing electrodes.
500 hours of operation
Replace the manifold filter cartridges.
>6000 hours of operation or as needed
Replace the pump filter cartridges.
>6000 hours of operation or as needed
Table 6-1 Maintenance Schedule
Cleaning the Cyclon e
To help maintain the instrument at peak performance, use the cyclone whenever you take measurements. The cyclone removes particles greater than those measured by the EEPS spectrometer and improves measurement accuracy by eliminating multiple charged particles behaving as smaller particles. The cyclone also reduces the particle load on filters and the accumulation of particles on interior surfaces of the charger column.
Clean the cyclone prior to the start of testing. When cleaning, refer to the illustration of the cyclone in Figure 2-2.
To clean the cyclone:
1. Separate the cyclone from the SS union fitting by loosening the hex nuts.
2. Remove the cyclone’s removable base by turning it counterclockwise.
3. Use compressed air to blow out the interior surfaces and rinse it in isopropyl alcohol, acetone, or water.
4. Examine the interior inlet orifice to make sure it is clear.
5. Dry the parts thoroughly, and add a very light coating of oil or grease to the bottom sides of the cyclone. This will increase its collection efficiency especially for fibers.
6. Reassemble and reinstall the cyclone.
6-2 Model 3090/3090AK Engine Exhaust Particle SizerTM Spectrometer
Loading...