TSI Instruments questemp 48n User Manual

Page 1
Page 2
Page 3
iii Contents

Contents

List of Figures
Contents
Contents ..................................................................... iii
Introduction ................................................................ 1
QUESTempº 48N Model ............................................. 1
Getting Started ........................................................... 2
Turning on, Logging/Capture an Event, and
Sensors ....................................................................... 6
Measurements ............................................................ 7
Remote measurements ............................................. 7
Operating QUESTempº 48N ...................................... 8
Displayed Items ........................................................ 12
Data Logging ............................................................ 14
Electronic Sensor Check ......................................... 15
Sensor Alignment .................................................... 16
Printing ...................................................................... 19
Powering ................................................................... 21
Appendix A: Specifications ................................... 22
QUESTemp° 48N Block Model ............................... 24
Appendix B: Heat Exposure Tables ....................... 25
Flag Conditions for U.S. Navy/Marine
Appendix C: Accessories (sold separately) .......... 29
Appendix D: Software ............................................ 30
Contact/Service information ................................... 34
Calibration ................................................................ 35
Warranty ................................................................... 36
Page 4
Contents
iv
List of Figures

List of Figures

1-1: QUESTempº 48N................................................. 1
1-2: Keypad keys identified ......................................... 3
1-3: Start-up screens (A – indicates model and
type B- indicates revision and battery power) .. 4
1-4: Index start-up screen ........................................... 4
1-5: Log rate start-up screen ....................................... 4
1-6: Main menu of the QT48N .................................... 5
1-7: Measurement screen (example) .......................... 5
1-8: Sensors identified ................................................ 6
1-9: Main menu example ............................................ 8
1-10: Viewing measured data ..................................... 8
1-11: Selecting temperature setting and setup ........... 9
1-12: Selecting temperature setting ............................ 9
1-13: Selecting log rate with event log ...................... 10
1-14: Run mode indicator .......................................... 11
1-15: Wet and Dry screen ......................................... 12
1-16: Globe and RH screen ...................................... 12
1-17: WBGT screen .................................................. 12
1-18: Time and Date screen ..................................... 12
1-19: Battery & memory screen ................................ 13
1-20: Navy PHEL stay times ..................................... 13
1-21: Index for ACGIH .............................................. 13
1-22: Navy PHELs screen ........................................ 13
1-23: Flag Conditions Navy/Marine Corp. Ashore .... 14
1-24: Main menu with view selected (A) &
measurement screen (B) ................................ 16
1-25: Alignment screen ............................................. 16
1-26: Sunshield & placement of finger cot prior
to RH alignment .............................................. 17
1-27: RH alignment with Low % example ................. 17
1-28: RH alignment with High% screen .................... 18
1-29: Sample DMS event data report ....................... 19
1-30: Sample printouts .............................................. 20
1-31: 9-volt battery .................................................... 21
1-32: Communicating with the QT 48N and DMS ... 30 1-33: Communicating with the QT 48N and DMS ... 31
1-34: Data Finder Page ............................................ 31
1-35: Quick Report .................................................... 32
1-36: Analyze button opens to the panel
layout page ..................................................... 33

List of Tables

1-1: Keypad explained ................................................ 3
1-2: Settings indentified .............................................. 9
1-3: Example of a Memory table .............................. 14
Page 5
1 Introduction

Introduction

List of Tables
Introduction
The QUESTempº 48N offers traditional heat stress monitoring without the aggravation of
maintaining a wet bulb. Through collaboration with Professor Dr. Thomas Bernard, from the College of Public Health at the University of South Florida, mathematical models were implemented to create a Waterless Wet Bulb calculation through a combination of dry bulb temperature, globe temperature, relative humidity, and air flow.
Figure 1-1: QUESTempº 48N

QUESTempº 48N Model

The QUESTempº 48N model was specifically engineered for the Department of Defense Ashore and Afloat Operations in which conditions require hand’s free monitoring, in-the-field user RH/WBGT sensor alignments, and snapshot measurement logging (called “event logging mode”).
The QT48N measures and calculates the dry bulb, wet bulb, globe, WBGT outdoors, and relative humidity with two different types of data logging modes – as a time history session or an event logging mode.
You also have the capability to measure stay times in order to manage work/rest regimens. Guidance is based on the screening criteria for heat stress as defined in the ACGIH TLV Handbook, U.S. Navy PHEL charts, and Flag Conditions for U.S. Navy/Marine Corp. Ashore.
Page 6
2 Getting Started
1. Place the QUESTempº 48N in the work area in a safe location approximately 3.5 feet off the ground.
2. Turn the unit On. If the battery voltage displayed during the power-on sequence is less than or equal to 6.4 volts, replace or recharge the batteries.
3. Be aware that the sensors require 10 minutes to stabilize to a new environment.
4. In the main menu, View will be selected (an indicator arrow denotes the selected menu). Press the I/O Enter key and the measurement screen will appear.
5. Press the Run/Stop key to begin datalogging. Use the Arrow keys to set the display to the desired items.

Getting Started

Up and Running Overview
Getting Started

Up and Running Overview

Placing the QUESTempº 48N on the Job Site

The QUESTempº 48N should be placed at a height of 3.5 feet (1.1 m) for standing individuals or 2 feet (.6 m) for seated individuals. Tripod mounting is recommended to get the unit away from anything that might block radiant heat or airflow. A 1/4" x 20 threaded bushing on the bottom of the instrument allows mounting to a standard photographic tripod. DO NOT stand close to the unit during sampling.
Before data logging, allow ten minutes for the sensors readings to stabilize.
Page 7
3 Getting Started
Keys
Explanation
I/O Enter key
The unit turns on with a single key press. The unit turns off by holding the key down while a countdown of 3-2-1 occurs in the lower right corner of the display. This key is also used to select a mode (such as Setup or View) or enter setup changes.
Pressing and releasing the key while viewing temperatures causes the display to view the next available sensor bar (indicated in the upper right corner of the display).
Up Arrow key
Changes items appearing in the display. Scrolls up.
Down Arrow key
Changes items appearing in the display. Scrolls down.
Run/Stop key
From the menu or view modes, pressing this key starts or stops the run mode. Pressing this key will exit the setup, print or reset modes.
Escaping or moving back one screen
If you are in the setup, print, reset, or sensor alignment screens, you can press Run/Stop key to escape or move back one screen.
I/O Enter key Run/Stop key
Up Arrow key Down Arrow key
Keypad Operation

Keypad Operation

The unit operates using a keypad with four keys. The I/O Enter key responds when the key is released while all other keys respond when the key is pressed.
Table 1-1: Keypad explained
Figure 1-2: Keypad keys identified
Page 8
4 Getting Started
QUEST
QT-48N WWB
REVISION 1.25
BATTERY 8.3V
Index: - - - -
<ENTER>
Log Rate: 1 min
Nxt 01 <ENTER>
Turning on, Logging/Capture an Event, and Turning Off

Turning on, Logging/Capture an Event, and Turning Off

To quickly get you started with the QUESTempº 48N, the following section explains turning on the instrument, data logging/event logging, and stopping.
1. Press the I/O Enter key to turn on. The first two start-up screens detail the name, type of instrument, and then a
revision and battery power screen appears. (Ensure the battery level does not fall below 6.4V or you will want to replace the battery. See “Power options” for more details.)
NOTE: If you have data saved on the instrument, an additional screen stating “data
in memory” will appear prompting you to reset the instrument, if desired.
A B
Figure 1-3: Start-up screens (A – indicates model and type
B- indicates revision and battery power)
2. In the Index Start-Up screen, press I/O Enter key.
NOTE: If an index is selected from the Setup menu, this will appear on this screen.
Figure 1-4: Index start-up screen
3. A Log Rate Start-Up screen will appear, press I/O Enter key. (This screen details your
log rate time or event logging setup option and the last saved logged session or event.) (See Setup for more information on event logging.)
Figure 1-5: Log rate start-up screen
Page 9
5 Getting Started
VIEW PRINT
SETUP RESET
Indicator arrow
Indicates the selected menu option. Either press up/down arrows to select a new menu or press I/O Enter key to select.
Numeric counter
Event records or Data logging sessions displayed
Run indicator.
Will appear when you are in “Run mode”.
For an event, it will appear for about 1-2 seconds as an event is recorded in the instrument’s history.
WET 53.9º F 01
DRY 113.5º F
Turning on, Logging/Capture an Event, and Turning Off
4. The main menu (navigational screen) will appear after the start-up screens. Press I/O
Enter key. (This will open to the measurement screens.)
Figure 1-6: Main menu of the QT48N
5. Press Run/Stop key to either Data Log or log/record an Event and an asterisk will appear
in far right corner.
NOTE: Data logging or an Event is selected via the Setup menu.
Numeric counter
o Event logging/recording: As you record events, the numeric counter will
increase by one displaying the number of saved events.
o Data logging mode: This will display the number of saved data logged
sessions instead of Events. (The process is identical to event mode.)
6. To stop data logging, press and hold Run/Stop key.
NOTE: Skip this step for an Event log. An event log will capture the measurements
with the initial run/stop key press.)
7. To view different measurements, press the Up or Down Arrow key to toggle through the
views.
8. To return to the main menu, press and hold the I/O Enter key (3, 2, 1 countdown will
appear) and the main menu will display.
To select an option on the main menu, press the Up or Down Arrow until an arrow
appears directly in front of the appropriate menu selection and then press I/O enter key.
Figure 1-7: Measurement screen (example)
Page 10
6 Sensors
The dry bulb thermometer (right position) measures the ambient air temperature. This measurement is used in the outdoor WBGT calculation when a high solar radiant heat load may be present. The series of white plates surrounding the sensor shield it from radiant heat.
A B C
A. Globe thermometer
B. Relative humidity
sensor
C. Dry bulb thermometer

Sensors

Globe Thermometer
9. To power off, press and hold the I/O Enter key from the main menu.
Sensors

Globe Thermometer

The globe thermometer (left position) gives an indication of the radiant heat exposure on an individual due to either direct sunlight or hot objects in the environment. This is accomplished by placing a temperature sensor inside a blackened copper sphere and measuring the temperature rise. The WBGT index is based on the response of a 6 inch diameter globe. The QUESTemp° uses a 2-inch diameter globe for a faster response time. The temperature of the 2 inch globe is correlated to match that of a 6 inch globe.

Waterless Wetbulb & Relative Humidity Sensor

The relative humidity sensor (middle position) is used to calculate the Waterless Wetbulb from a combination of dry bulb temperature, humidity and wind speed measurements. The waterless wetbulb is used to calculate an estimated WBGT value.

Dry Bulb Thermometer

Figure 1-8: Sensors identified
Page 11
7 Measurements
**NOTE: Navy and Marine calculations are all based on the outdoor computation.
WBGT (outdoor) = 0.7WB + 0.2G + 0.1DB (denoted as “WBGT” on the display)

Measurements

WetBulb Globe Temperature (WBGT)
Measurements
The QUESTemp 48N data logging, with an event logging mode, heat stress monitor directly senses three parameters: dry bulb temperature (DB), globe temperature (G), and relative humidity (RH).
It computes the Wet Bulb (WB), the Wet Bulb Globe Temperature (WBGT), and the Stay Times Indices for ACGIH, U.S. Navy PHEL charts, and Flag Conditions for U.S. Navy/Marine Corp. Ashore.

WetBulb Globe Temperature (WBGT)

The WBGT is a type of temperature index which combines the effects of temperature, humidity, radiant heat, and air flow and provides a relatively simplistic tool to analyze thermal comfort. The resulting WBGT values can then be compared to indices of work-rest regimens (stay times) based upon workloads.
The WBGT calculation is a weighted average of the three temperature sensors using the following formulas:

Stay Times/Rest Times (Indicies)

Stay Times Indices represent how long a worker should be able to safely work under heat stress conditions. Select one of four indices for displaying and printing from the unit: ACGIH Stay Times, NAVY PHEL’s, or Flag Conditions for U.S. Navy/Marine Corp. Ashore. Refer to Appendix B for more information on the indices.
 NOTE: It will appear as “Index” on the measurement screen.

Remote measurements

The top sensor bar may be removed from the instrument and used through a remote cable. (Refer to Appendix C for remote cable options.) Shelter the instrument and remote the sensor bar if the measured environment is expecting heavy rain or if temperatures are above 60°C.
Page 12
8 Operating QUESTempº 48N
From a powered on mode, the main menu will appear (after the start-up screens are displayed.) Use the Up Arrow and Down Arrow keys to move the marker in the display in front of the desired mode. Pressing the I/O Enter key will select the mode.
VIEW PRINT
SETUP RESET
Figure 1-9: Main menu example
View mode
Arrow icon Indicates you are in view mode
Event #
Last logged event/session
WET 53.9º F 01
DRY 113.5º F
Operation
View

Operating QUESTempº 48N

View

Displays the measured data but does not log it. If more than one set of sensors is plugged into the unit, they can be displayed by pressing and releasing the I/O Enter key. The displayed sensor set is shown in the upper right corner.
NOTE: To return to the main menu, hold down the I/O Enter key while a 3, 2, 1
countdown is shown in the lower right corner of the display.
Figure 1-10: Viewing measured data
Page 13
9 Operating QUESTempº 48N
Settings
Explanation
Temperature units
Selectable: ºF/ºC
Language setting
Selectable: English, Spanish, French, Italian, and German
Time and Date
Time settings: 24-hour clock Day-month-year format
Log Rate
1, 2, 5, 10, 15, 30, 60 minutes log rate options Event Log: when run/stop is pressed, an event or
snapshot of the current conditions is logged into memory. The events range from 0-99.
Index/Stay Times
ACGIH, PHEL, FLAG, or none (denoted “----”)
VIEW PRINT
SETUP RESET
Press I/O Enter key to select Setup.
Fahrenheit English
Temperature setting – press I/O Enter key to switch between values.
Setup

Setup

The setup options are from the main menu when Setup is selected. At any time, to exit (or escape) a screen, press the Run/Stop key. Table 1-2 identifies the setup parameters.
Selecting Setup parameters
1. From the main menu, select Setup by pressing the I/O Enter key. Using the Up/Down
Arrow keys, select the settings in steps 2-6.
Table 1-2: Settings indentified
Figure 1-11: Selecting temperature setting and setup
2. For the Time and Date settings, once selected, press I/O Enter key to move to the
update field. (An underline will appear under the first field.)
Press Up/Down Arrows to change the number/month. Press I/O Enter key to
toggle through each field. Repeat as necessary until all values are selected.
3. For the Temperature setting, either Fahrenheit or Celsius will appear. Press the
Up/Down Arrow to select and I/O Enter key to change.
Figure 1-12: Selecting temperature setting
Page 14
10 Operating QUESTempº 48N
Event
Example of setting Log Rate to Event Log
Next event number
*To increase, press up arrow key (used to simulate skipping an event.)
Log Rate: event Nxt event: 11
Print
4. For the Language setting, once selected, press I/O Enter key. To select a language,
press Up/Down Arrow keys. Once selected, press Run/Stop key to return to setup parameters.
5. For the Log Rate, once selected, press I/O Enter key.
Repeatedly press I/O Enter keys to select either the Log Rate Event or a Log Rate
interval (for data logging; e.g. “5”).
Optional: changing Next Event field. Events are numbered starting at 1 and
incrementing. The number may be increased to simulate skipping events. This may be useful if the events represent locations and the starting location is somewhere other than 1.
Figure 1-13: Selecting log rate with event log
6. For the Index setting, once selected press I/O Enter key. Press Up/Down Arrow keys to
select: ---” (which is equivalent for no index), ACGIH, PHEL, or Flag.
7. Exit Setup by pressing the Run/Stop key.

Print

Allows printing to a parallel or serial printer or to a computer. The QUESTempº 48N Heat
Stress Monitor will recognize the cable plugged in and configure itself for serial or parallel. If no cable is plugged in, it will default to serial. Press I/O Enter key to begin printing. Press Run/Stop key to return to the menu.
NOTE: If you want to stop the printing, press I/O Enter key until you return to the main

Reset

Resetting enables you to clear the logged data from memory. Press the I/O Enter key to enter the Reset mode. Clear the memory by holding down the I/O Enter key while the display counts down from three.
menu. When the printer has stopped printing, remove the cable from the printer to the instrument.
Page 15
11 Operating QUESTempº 48N
GLOBE 90.3º F 05
RH 53.3º F
Run Indicator
Run
Run
For Data Logging setup, the run mode begins a session in memory and logs the data. For Event Logging, each run/stop is stored as an event and a numeric counter displays the number of saved events.
1. Begin a session by pressing the Run/Stop key from the view mode (or measurement
view). An asterisk in the lower right corner indicates the run mode.
To toggle through the views, press the Up or Down Arrow.
2. End the session by pressing the Run/Stop key again. (The session will stop recording
when the asterisk is no longer displayed.)
NOTE: If the logging memory is full or if there are no sensors plugged into the unit,
attempting to enter the Run mode will result in an error message. If the memory capacity is exceeded, the asterisk in the lower right corner of the display will turn into an “F” and the memory remaining screen will show “0.0”.
Figure 1-14: Run mode indicator
Page 16
12 Displayed Items
Screen 1: WET (Wet bulb)
DRY (Dry bulb)
WET 80.5º F 01 DRY 92.2º F
Screen 2: GLOBE and RH
GLOBE 40.4°C 03 RH 52.5%
Screen 3: WBGT
WBGT 30.7°C 3
Screen 4: Time (24 hour format)
Date (day, month, year)
TIME 11:04:13 04 DATE 14-MAY-18
Displayed
Items
Run

Displayed Items

For the QUESTemp 48N Heat Stress Monitor, the number in the upper right corner indicates the saved session or the saved event log.
1indicates the sensor bar placed on (or attached to) the top of the instrument. Sensors
2 and 3 are labeled on the side of the unit as “Sensor 2”, and “Sensor 3”.
An *asterisk in the lower right corner indicates that the unit is in the run mode and is
logging data or saves an event.
The following measurements can be accessed on the display:
Figure 1-15: Wet and Dry screen
Figure 1-16: Globe and RH screen
Figure 1-17: WBGT screen
Figure 1-18: Time and Date screen
Page 17
13 Displayed Items
Screen 5: BAT (Battery voltage)
MEM (Logging memory available in days)
BAT 09:04:13 05 DATE 10.4 dy
Screen 6: Index
(Navy PHEL stay times)
PHEL_5 3:10  05 PHEL_6 8:03
L M H VH 04
60 45 30 15
PHEL_5 3:10 05
PHEL_6 8:03
PHELs stay time is greater than 8 hours.
Stay Time/Index
NOTE: A series of dashes appear in the display if one of the following occur:
The temperature is outside of its allowable range
A temperature sensor has failed
Stay times temperatures are outside of the their defined range

Stay Time/Index

The screen(s) displaying stay time data appear different for each of the possible indices.
If ACGIH is selected, the recommended working minutes per hour are shown for each of the workload categories Light (L), Moderate (M), Heavy (H), and Very Heavy (VH).
Figure 1-19: Battery & memory screen
Figure 1-20: Navy PHEL stay times
If the Navy PHELs are selected, the recommended working hours are shown based on a maximum of eight hours. Three screens are used to display the PHELs two at time.
NOTE: 8:01 following one of the PHELs indicates greater than eight hours.
Figure 1-21: Index for ACGIH
Figure 1-22: Navy PHELs screen
Page 18
Data Logging
14
HEAT CATEGORY 05
YELLOW FLAG
Navy/Marine Corp’s heat condition flag
warning example
Log Rate
1 min
2 min
5 min
10 min
15 min
30 min
60 min
1 sensor
11.2
22.5
56.2
112.4
168.6
337.3
674.5
2 sensors
5.6
11.2
28.1
56.2
84.3
168.6
337.3
3 sensors
3.7
7.5
18.7
37.5
56.2
112.4
224.8

Data Logging

Stay Time/Index
If Flag is selected, the Flag Conditions for Navy/Marines Corp. Ashore warning system screen will appear. There are five flag systems, no flag, yellow, green, red, and black, which provide heat exposure guidelines for acclimated individuals. (For more information, refer to the “Flag Conditions US Navy/Marine Corp. Ashore”.)
Data Logging
Data from each sensor is recorded at the interval set by the logging rate. Every time Run/Stop is pressed, a session is either started or ended in memory. Each session contains a header with time, date, and summary information.
Memory Table: Gives the number of logging DAYS.
Figure 1-23: Flag Conditions Navy/Marine Corp. Ashore
Table 1-3: Example of a Memory table
Page 19
15 Electronic Sensor Check
Example of Verification module
Electronic
Sensor Check
Performing an Electronic Sensor Check

Electronic Sensor Check

A verification module, TSI model 053-923, may be used to check the operation of the QUESTemp monitor’s wet bulb, dry bulb, and globe. The purpose is to verify that the electronic components are within a specific range with known values and a known source. The temperature tolerances should be within ±0.5oC.
NOTE: If the sensors are outside of the tolerances, this indicates the sensor alignment should be
serviced off-site for calibration.

Performing an Electronic Sensor Check

1. Ensure the instrument is reading in Celsius prior to your electronic sensor check.
To change the temperature setting, select Setup from the main menu. Then, either
Fahrenheit or Celsius will appear on the screen. When selected, press I/O enter key to switch between settings. For more information, refer to step 3.
2. Remove the top sensor bar, place to the side, and plug in the verification module into the
3. Verify the measurement readings on the screen are within ±0.5C tolerance to the readings
4. Once completed, remove verification module and place sensor bar back on the instrument.
center pins of the sensor housing.
printed on the verification module label. Example below:
a. Wet Bulb (WB): 11.1C b. Dry Bulb (DB): 45.3C c. Globe (G): 69.2C
NOTE: Relative humidity (RH) is not valid on this instrument.
(Tighten down the two bolts.) Change the Celsius reading back to Fahrenheit. (Refer to step 1 a-b.)
Page 20
Sensor Alignment
16
VIEW PRINT
SETUP RESET
WET 50.7° F
DRY 70.6° F
(RH Sensor) Alignment menu
Alignment RH
WBGT HIGH LOW
“High” is selected when the arrow appears in
front of the text.
NOTE: WBGT is not used in sensor alignment
Sensor
Alignment
RH Sensor Alignment

Sensor Alignment

For highly accurate measurement readings, you should align your QUESTempº 48N prior to data
logging or event logging/recording. For the RH sensor, it is recommended to align with a High concentration level using the 75% sensor salt, NaCl (sodium chloride), or a Low concentration level using the, the 33% sensor salt, MgCl (magnesium chloride). (To order sensor alignment salts, refer to an online vendor such as, colepalmer.com.)

RH Sensor Alignment

1. To open, navigate to the main menu and select View by pressing I/O Enter key (see A).
A measurement screen will appear (see B).
Figure 1-24: Main menu with view selected (A) & measurement screen (B)
2. Press and hold I/O Enter key and then press Down Arrow key from the View menu. The
Alignment screen will appear.
A B
3. Press Up/Down Arrow key to select either High or Low. Then press I/O Enter key.
4. Remove (or slide) the Sunshield (white globe) from the RH sensor and place it to the side.
(See Figure 1-26).
Figure 1-25: Alignment screen
Page 21
Sensor Alignment
17
Sunshield:
Slide off RH sensor to remove.
Place unrolled finger cot, with a small hole cut in the top, so it is completely stretched over the humidity sensor.
Alignment salt (remove rubber cap and place over RH sensor with the sunshield removed)
Once Alignment salt is placed, allow levels to stabilize (about 15 mins).
Low RH Alignment screen
Adjust % to 33% (using up/down arrow keys) and press Enter (after level has stabilized) to store sensor alignment level.
RH Sensor Alignment
5. Place a latex finger cot (user supplied) over the humidity sensor in order for the sensor to
stabilize quicker about 15 minutes (or up to 60 minutes without).
NOTE: Latex finger cots can be found at various hardware stores or online such as
colepalmer.com.
Figure 1-26: Sunshield & placement of finger cot prior to RH alignment
6. Place the Salt container, with the salt alignment cap removed, over the RH sensor (middle
sensor). Allow level to stabilize for at least 15 minutes. (Refer to Figure 1-25.)
About sensor alignment: if readings are within ±0.5%, a change in sensor alignment is
not required. Skip to step 7 to return to main menu.
NOTE: For a High alignment, use the 75% sensor alignment salt (NaCL) and for a Low
alignment, use the 33% alignment salt (MgCl).
Figure 1-27: RH alignment with Low % example
Page 22
Sensor Alignment
18
Press Enter to save the alignment.
Example of high RH percentage prior to sensor alignment. Press up/down arrows to adjust level after salt has stabilized.
RH 51.2%
<ENTER> HIGH
Sensor alignment with High % RH level
NOTE: WBGT is not used in sensor alignment
RH Sensor Alignment
7. In either the High/Low RH alignment screen, adjust the level to the percentage displayed on
the salt container by using the Up/Down Arrow keys. Press I/O Enter key to save. (The instrument will state either “Successful” or “Failure” if it passed/did not pass. For “Failure” repeat the steps above.)
Figure 1-28: RH alignment with High% screen
8. To return to the main menu, press Run/Stop key.
9. Verify the RH sensor is ±0.5% of the specific percentage. If not repeat the steps above.
10. Remove the alignment salt and finger cot. Replace the Sunshield over the RH sensor when
completed with the sensor alignment.
Page 23
19 Printing

Printing

Data/Communications
Printing
Printing options include to a RS-232 port or to a parallel printer. Serial transmission requires TSI cable #54-715. Parallel transmission requires TSI cable #56-875. With the applicable cable plugged in, select PRINT from the menu and press the I/O Enter key to enter the PRINT mode. Begin printing by pressing the I/O Enter key. Press the key again to abort the printing.

Data/Communications

TSI Detection Management Software DMS is recommended for downloading, storing, graphing, and printing reports (see below). Communications programs, such as Windows® HyperTerminal® may also be used to capture the printout into a file. The baud rate is fixed at 9600.
Figure 1-29: Sample DMS event data report
Page 24
Printing
20
EVENT: 1 Page 2 Sensor: 1
Degrees Fahrenheit Stay Times: ACGIH, Acclimated, WBGT, clo correction = 1.0 °C
TIME WBGT DRY GLOBE RH HI FLOW L M H VH 11:08 67.9 82.4 90.7 13 0 0.5 60 60 60 60 11:09 68.1 82.6 91.3 12 0 0.5 60 60 60 60
EVENT: 2 Page 3
Sensor: 1 Degrees Fahrenheit Stay Times: ACGIH, Acclimated, WBGT, clo correction = 1.0 °C
TIME WBGT DRY GLOBE RH HI L M H VH 11:08 79.3 98.9 104.5 15 0 60 45 30 15 11:09 80.2 99.2 105.6 15 0 60 45 30 15
EVENT: 3 Page 4
Sensor: 1 Degrees Fahrenheit Stay Times: ACGIH, Acclimated, WBGTi, clo correction = 1.0 °C
TIME WBGT DRY GLOBE RH HI L M H VH 11:08 68.1 88.0 92.7 11 0 60 60 60 60 11:09 68.4 88.3 92.9 11 0 60 60 60 60
EVENT: 4 Page 5
Sensor: WBGT(W-AVG) = .50*WBGT(1) + .25*WBGT(2) + .25*WBGT(3) Degrees Fahrenheit
Stay Times: ACGIH, Acclimated, WBGT, clo correction = 1.0 C WBGT WBGTo
TIME W-AVG W-AVG L M H VH
----- ----- ----- --- --- --- --­ 11:08 71.5 70.8 60 60 60 60 11:09 71.8 71.1 60 60 60 45
TSI QUEST TECHNOLOGIES Page 1 HEAT STRESS REPORT
File Name _________________________ QUESTempº48N Rev 1.00
Serial # TK09090909
Employee _________________________
Facility _________________________ Session (3) Start: 21-NOV-09 11:07:32 Department ________________________ Stop: 21-NOV-09 11:10:15
Job _________________________ Printed: 21-NOV-09 11:16:00
Comments/NOTEs_____________________________________________ _
____________________________________________________________
Logging Interval: 1 minutes Degrees Fahrenheit
MAXIMUM LEVELS, Sensor 1
WBGT 68.3 21-NOV-09 11:10:08 DRY BULB 82.7 21-NOV-09 11:09:56 GLOBE 91.4 21-NOV-09 11:10:12 REL HUMIDITY 14% 21-FEB-08 11:07:32 FLOW (m/s) 0.6 21-FEB-08 11:09:08
MAXIMUM LEVELS, Sensor 2
WBGT 80.5 21-FEB-08 11:10:11 DRY BULB 99.2 21-FEB-08 11:09:07 GLOBE 106.1 21-FEB-08 11:10:06 HEAT INDEX 0 00-XXX-00 00:00:00 REL HUMIDITY 15% 21-FEB-08 11:07:32
MAXIMUM LEVELS, Sensor 3
WBGT 68.6 21-FEB-08 11:09:56 DRY BULB 88.6 21-FEB-08 11:10:08 GLOBE 93.0 21-FEB-08 11:10:03 HEAT INDEX 0 00-XXX-00 00:00:00 REL HUMIDITY 11% 21-FEB-08 11:07:32
MAXIMUM LEVELS, Sensor(WEIGHTED AVERAGE) WBGT 71.4 21-FEB-08 11:10:14
Parallel

Parallel

Data can be sent directly to parallel printers that accept direct ASCII test input without special drivers. Make sure the printer is powered on and is online, ready to accept data, prior to printing.
Figure 1-30: Sample printouts
Page 25
Powering
21
The QUESTempº 48N Heat Stress Monitor uses a 9-volt alkaline battery. A door on the back of the instrument allows the user access to the 9-volt battery.
The 2-position switch located in the battery compartment is set by the user to the 9-volt battery.
Figure 1-31: 9-volt battery
9-Volt Alkaline Battery Replacement

Powering

9-Volt Alkaline Battery Replacement

WARNING: Replace batteries only in a non-hazardous environment.
The 9-volt battery should be replaced should be recharged when the voltage drops below
6.4 volts. The battery voltage is displayed when the instrument is turned on. While turned on,
the battery voltage can be displayed at any time by pressing the up or down arrow keys to move through the display until the battery voltage screen appears. If, while operating, the battery voltage drops below 6.4 volts, the display will automatically switch to the display showing the battery voltage along with a low battery message. After a low battery occurs, the unit will continue to operate for approximately 8 hours. When the battery voltage falls to
6.2 volts or below, the unit will automatically turn off.
Replace only with an approved 9-volt alkaline battery.

Approved 9-Volt Batteries

Eveready: Energizer 522, EN22, 6LR61 Duracell: MN1604 Panasonic: 6LR61, 6AM6X Rayovac: A1604 UltraLife: U9V
Page 26
Appendix A: Specifications
22
Appendix A:
Specifications
Measurements

Appendix A: Specifications

Measurements

Globe Temperature, Dry Bulb Temperature, Wet Bulb Temperature, % Relative Humidity,
and WBGT Index
Temperatures given in Celsius or Fahrenheit Index (displayed stay time with selected index): ACGIH TLV, U.S. Navy PHEL Charts,
Flag Conditions for U.S. Navy/Marine Corp. Ashore

Data Logging mode and Event logging mode

Data logging mode records and prints all measurements at user selected interval of 1, 2,
5, 10, 15, 30, or 60 minutes.
Event logging mode (1 second record) 128K bytes of data memory.

Languages

English, French, Spanish, Italian, German

Sensors

Dry Bulb Sensor- 1000 Ohm Platinum RTD
o Accuracy and Ranges: ±0.5 from 0°C to 120°C (±0.9°F from 32°F to 248°F)
Waterless Wet Bulb (Humidity) Sensor
o Accuracy and Ranges: Expanded measurement uncertainty of 1.1°C (k=2)
between 0°C and 80°C (32°F and 176°F)
Globe Sensor- 1000 Ohm Platinum RTD
o Accuracy and Ranges: ±0.5 from 0°C to 120°C (±0.9°F from 32°F to 248°F)
Relative Humidity Sensor - Integrated circuit with capacitive polymer sensor
o Accuracy and Ranges: ±5% from 20 to 95% (non-condensing)

Operating Temperature Range

Sensor Assembly: -5C to +100C Electronics: -5C to 60C

Housing

Designed water resistant to a light rain or mist. If rain is frequent, best practice would be
to remote the sensor bar and keep the instrument sheltered.
Page 27
Appendix A: Specifications
23
Size

Size

Height 9.2 in. (23.5 cm); Width 7.2 in. (18.3 mm); Depth 3.0 in. (7.5 mm) Dimensions include mounted sensor assembly

Weight

2.6 lbs. (1.2 kg) with mounted sensor assembly

Remote Sensor Bar

Using a cable, the top sensor bar can be setup for remote measurements, up to 200 feet
(61 m).

Power

9V alkaline

Battery Life

9V alkaline: 80 hours
Page 28
24 QUESTemp 48N Block Model
Keypad
Sensors 1, 2, & 3
Multiplexer
Gain
A/D
Microprocessor
LCD
RS232 Printer
Power
Controller
Battery
Data Memory
Program
Memory
Appendix A: Block Model
Battery Life
QUESTemp° 48N Block Model
Page 29
Appendix B
25
Work and recovery (TLV )
Light
Moderate
Heavy
Very Heavy
75% to 100%
31.0
28.0
26.0*
23.5*
50% to 75%
31.0
29.0
27.5
25.5*
25% to 50%
32.0
30.0
29.0
28.0
0% to 25%
32.5
31.5
30.5
30.0
Work and recovery (Action Limit )
Light
Moderate
Heavy
Very Heavy
75% to 100%
28.0
25.0
22.5*
20.0*
50% to 75%
28.5
26.0
24.0
22.5*
25% to 50%
29.5
27.0
25.5
24.5
0% to 25%
30.0
29.0
28.0
27.0
Appendix B:
Heat Exposure
Tables
ACGIH

Appendix B: Heat Exposure Tables

ACGIH

Screening Criteria for Heat Stress Exposure. WBGT values in C.
NOTE: According to the ACGIH’s guidelines, the temperature values represent a work and rest process which is explained in the standards. Refer to the ACGIH TLVs and BEIs for specific details.
*Values not specified by ACGIH have been estimated for continuity.
Page 30
Appendix B
26
Clothing type
Clothing correction (Addition to WBGT (ºC)
Work clothes (long sleeve shirt and pants)
Cloth (woven material) coveralls
Double-layer woven clothing
SMS polypropylene coveralls
0.5º
Polyolefin coveralls
Limited-use vapor-barrier coveralls
11º
WBGT(F)
I
II
III
IV V VI
80.0
>8:00
>8:00
>8:00
8:00
6:35
4:30
81.0
>8:00
>8:00
>8:00
8:00
6:35
4:30
82.0
>8:00
>8:00
8:00
7:05
5:25
3:40
83.0
>8:00
8:00
7:45
6:25
4:55
3:20
84.0
>8:00
8:00
7:05
5:55
4:30
3:05
85.0
8:00
7:45
6:30
5:20
4:05
2:50
86.0
8:00
7:05
5:55
4:55
3:45
2:35
87.0
7:25
6:30
5:25
4:30
3:25
2:20
88.0
6:45
5:55
4:55
4:05
3:10
2:10
89.0
6:10
5:25
4:30
3:45
2:50
2:00
ACGIH Clothing Corrections

ACGIH Clothing Corrections

The following clothing corrections are in degrees Celsius. When a clothing correction is entered into the setup portion of the QUESTemp 48N Heat Stress Monitor, the value is added to the WBGT only for looking up the stay times. The WBGT value displayed by the unit does not reflect corrections.
Cited from American Conference of Governmental Industrial Hygienists, Threshold Limit Values and Biological Exposure Indices for 2008. Reprinted with permission.

United States Navy

Physiological Heat Exposure Limits (PHEL) Time Table (Without the presence of fuel combustion gases/fuel vapors)
The recommended working hours are shown based on a maximum of eight hours. Naval personnel will follow a category, I - VI, based upon their function.
PHEL Curves (Total Exposure Time in Hours: Minutes)
Page 31
Appendix B
27
WBGT(F)
I
II
III
IV V VI
90.0
5:40
5:00
4:10
3:25
2:40
1:50
91.0
5:15
4:35
3:50
3:10
2:25
1:40
92.0
4:50
4:10
3:30
2:55
2:15
1:30
93.0
4:25
3:50
3:15
2:40
2:00
1:25
94.0
4:05
3:35
3:00
2:25
1:50
1:15
95.0
3:45
3:15
2:45
2:15
1:45
1:10
96.0
3:25
3:00
2:30
2:05
1:35
1:05
97.0
3:10
2:45
2:20
1:55
1:25
1:00
98.0
2:55
2:35
2:10
1:45
1:20
0:55
99.0
2:40
2:20
2:00
1:40
1:15
0:50
100.0
2:30
2:10
1:50
1:30
1:10
0:45
101.0
2:20
2:00
1:40
1:25
1:05
0:45
102.0
2:10
1:50
1:35
1:15
1:00
0:40
103.0
2:00
1:45
1:25
1:10
0:55
0:35
104.0
1:50
1:35
1:20
1:05
0:50
0:35
105.0
1:40
1:30
1:15
1:00
0:45
0:30
106.0
1:35
1:25
1:10
0:55
0:45
0:30
107.0
1:30
1:15
1:05
0:50
0:40
0:25
108.0
1:20
1:10
1:00
0:50
0:35
0:25
109.0
1:15
1:05
0:55
0:45
0:35
0:25
110.0
1:10
1:00
0:50
0:40
0:30
0:20
111.0
1:05
1:00
0:50
0:40
0:30
0:20
112.0
1:00
0:55
0:45
0:35
0:25
0:20
113.0
0:55
0:50
0:40
0:35
0:25
0:15
114.0
0:55
0:45
0:40
0:30
0:25
0:15
115.0
0:50
0:45
0:35
0:30
0:20
0:15
116.0
0:45
0:40
0:35
0:25
0:20
0:15
117.0
0:45
0:40
0:30
0:25
0:20
0:10
118.0
0:40
0:35
0:30
0:25
0:15
0:10
119.0
0:35
0:35
0:25
0:20
0:15
0:10
120.0
0:35
0:30
0:25
0:20
0:15
0:10
121.0
0:35
0:30
0:25
0:20
0:15
0:10
United States Navy
Page 32
Appendix B
28
WBGT(F)
I
II
III
IV V VI
122.0
0:30
0:25
0:20
0:15
0:15
0:10
123.0
0:30
0:25
0:20
0:15
0:10
0:10
124.0
0:25
0:25
0:20
0:15
0:10
0:05
No Flag
Green
Yellow
Red
Black
Flag Degrees F
<80.0
80.0–84.9
85.0–87.9
88.0–89.9
>90
Flag Degrees C
<26.7
26.7–29.4
29.4–31.1
31.1–32.2
>32.2
Flag Conditions for U.S. Navy/Marine Corp. Ashore

Flag Conditions for U.S. Navy/Marine Corp. Ashore

The following chart details the heat stress monitoring Flag Conditions for U.S. Navy/Marine Corp. Ashore in Degrees Fahrenheit and Celsius.
*Rest means minimal physical activity (sitting or standing) and should be accomplished in the shade if possible.
Page 33

Appendix C: Accessories (sold separately)

29
Appendix C: Accessories
Appendix C: Accessories (sold separately)
Replacement Relative humidity sensor (one included) ................................................. 057-317
Sensor array with 2-inch Globe and removable RH sensor (one included) ................. 057-909
Sensor alignment salt at 75% [ NaCl (sodium chloride)] (one included) ...................... 100-452
Sensor alignment salt at 33% [ MgCl (magnesium chloride)] (one included)............... 100-451
Verification module (one included) ................................................................................. 053-923
Neck Strap (one included) ............................................................................................... 0057-333
QT-3X series replacement battery cover ........................................................................ QT3X-BC
QT-3X series replacement dry bulb shield assembly-consists of
(5) shields and (1) cap/cover. ..................................................................................... QT3X-DBS
Remote sensor cable - 6 Foot ......................................................................................... 053-924
Remote sensor cable - 25 Foot ...................................................................................... 053-925
Remote sensor cable - 100 Foot .................................................................................... 053-926
Remote sensor cable - 200 Foot ................................................................................... 053-927
Serial/Computer interface cable (RS-232 to PC) 9 Pin ................................................. 054-715
RS-232 to USB converter (one included) ....................................................................... 053-810
Parallel printer interface cable ......................................................................................... 056-875
Storage case (one included) ........................................................................................... 053-922
QT48N User Manual (one included) ............................................................................... 057-330
Page 34

Appendix D: Software

30
QT 48 data port
Connected and powered on
Connected to serial port
Appendix D:
Software
Appendix D: Software
The QUESTemp 48N Heat Stress Monitor has the flexibility to be set up and controlled through computer software. The programmable start and stop time feature is only accessible through the computer. The instrument also has the capability of sending live data while measuring. These features are best utilized using TSI Detection Management Software DMS.

DMS quick overview

The focus of this section is to briefly introduce the following QSP-II topics: Downloading your data, setting up parameters, and viewing your data in charts, graphs, and reports. (For further details on DMS, refer to the online Help and select Contents.)

Communicating and Downloading in DMS

In order to download, review the data, and setup parameters in DMS software, this will require connecting the QT48N to a computer. The steps below explain connecting and communicating to DMS.
1. Using a TSI cable, part number 054-715, plug the cable into the computer and plug the
opposite end into the side jack of the QT 48 data port.
Figure 1-32: Communicating with the QT 48N and DMS
2. From the start page of DMS, select button and the instrument
communication panel will appear.
3. Select Heat Stress and then select the Model Type by clicking on QT48N. (See
Figure 1-34).
Page 35
Appendix D: Software
31
Data Finder checkbox. When checked, this will automatically open to the downloaded data page.
Download
Double-click to select/view QT44/46 data in charts and graphs
Data Finder and Quick Reports
4. Click on the button in the Instrument Communications page while Data
Finder checkbox is checked (see below).
NOTE: The Data Finder page will appear. See next section.
Figure 1-33: Communicating with the QT 48N and DMS

Data Finder and Quick Reports

In the Data finder page, the data you downloaded is stored by instrument family and then organized by models. NOTE: When working in this page, click on the Most Recent button if you are looking for your recently downloaded data.
1. In the Data Finder page, click on the downloaded data and select either Analyze or
Print Report.
Figure 1-34: Data Finder Page
Page 36
Appendix D: Software
32
Data Finder and Quick Reports
Quick Report: select the button to generate a report with basic
information and a summary table of your downloaded data.
Figure 1-35: Quick Report
Page 37
Appendix D: Software
33
Data Finder and Quick Reports
Analyze: allows you to add charts/graphs and customize how they will appear in a
report. Tip: Add the panels by using the add a panel and dragging them into the appropriate location.
Press button and it will generate a report based on the where the
charts/graphs (panels) are placed.
Figure 1-36: Analyze button opens to the panel layout page
Page 38
34 Contact/Service information
North America and Asia Pacific
Telephone: 1-800-680-1220 (USA);
+1 651-490-2860 (Outside USA)
Fax: +1 651-490-3824 E-mail: technical.services@tsi.com
Europe, Middle East, and Africa
Telephone: +49 241-52303-0
Fax: +49 241 52303-49 E-mail: tsigmbh@tsi.com
North America and Asia Pacific
TSI Incorporated
1060 Corporate Center Drive Oconomowoc, WI 53006-4828
Europe, Middle East, and Africa
TSI Instruments Ltd.
Stirling Road Cressex Business Park High Wycombe, Bucks HP12 3ST United Kingdom
Telephone: 1-800-680-1220 (USA);
+1 651-490-2860 (Outside USA)
E-mail: technical.services@tsi.com
Telephone: +44 (0) 149 4 459200
E-mail: tsiuk@tsi.com
Contact/Service
Information
Technical Support Contacts

Contact/Service information

This section gives directions for contacting TSI Incorporated for technical information and directions for returning the QUESTemp° 48N Heat Stress Monitor for service.

Technical Support Contacts

If you have any difficulty setting or operating the instrument, or if you have technical or application questions about this system, contact TSI’s Technical Support.

Service Contact Information

If your instrument does not operate properly, or if you are returning the instrument for service, visit our website at http://rma.tsi.com for a Return Material Authorization, or contact Customer Service.
Page 39

Calibration

35

Returning for Service

Visit our website at http://rma.tsi.com and complete the on-line “Return Merchandise Authorization” form or call TSI at 1-800-680-1220 (USA) or (651) 490-2860, or 001 651 490 2860 (International) for specific return instructions.
Customer Service will need the following information:
The instrument model number The instrument serial number A purchase order number (unless under warranty) A billing address A shipping address
Use the original packing material to return the instrument to TSI. If you no longer have the original packing material, seal off any ports to prevent debris from entering the instrument and ensure that the display and the connectors on the instrument front and back panels are protected. This instrument is very fragile and must be packed in a manner appropriate for a precision instrument.
Calibration
The QUESTemp 48N Heat Stress Monitor and TSI field calibrator devices should be examined regularly by the factory. An annual calibration is recommended. (See Service Information above.)
Page 40

Warranty

36
Warranty
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator's manual, to be free from defects in workmanship and material for 12 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when indicated in
specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material,
under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses, batteries or other
consumable materials. Only the original manufacturer's warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that the instrument or product is properly
calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory authorized service center with the one
exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or intentional damage, or is not
properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS
OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH
RESPECT TO SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS
LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED
INFRINGEMENTS. BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS
WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI’s Customer Service department at 1-800-874-2811 (USA) or +001 (651) 490-2811 (International).
Page 41
USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 1 41 19 21 99 Germany Tel: +49 241 523030
India Tel: +91 80 67877200 China Tel: +86 10 8219 7688 Singapore Tel: +65 6595 6388
TSI Incorporated – Visit our website www.tsi.com for more information.
P/N 057-330 Rev. F ©2018 TSI Incorporated Printed in U.S.A.
Loading...