1-3: Example of a Memory table .............................. 14
Page 5
1 Introduction
Introduction
List of Tables
Introduction
The QUESTempº 48N offers traditional heat stress monitoring without the aggravation of
maintaining a wet bulb. Through collaboration with Professor Dr. Thomas Bernard, from the
College of Public Health at the University of South Florida, mathematical models were
implemented to create a Waterless Wet Bulb calculation through a combination of dry bulb
temperature, globe temperature, relative humidity, and air flow.
Figure 1-1: QUESTempº 48N
QUESTempº 48N Model
The QUESTempº 48N model was specifically engineered for the Department of Defense
Ashore and Afloat Operations in which conditions require hand’s free monitoring, in-the-field
user RH/WBGT sensor alignments, and snapshot measurement logging (called “event
logging mode”).
The QT48N measures and calculates the dry bulb, wet bulb, globe, WBGT outdoors, and
relative humidity with two different types of data logging modes – as a time history session or
an event logging mode.
You also have the capability to measure stay times in order to manage work/rest regimens.
Guidance is based on the screening criteria for heat stress as defined in the ACGIH TLV
Handbook, U.S. Navy PHEL charts, and Flag Conditions for U.S. Navy/Marine Corp. Ashore.
Page 6
2 Getting Started
1. Place the QUESTempº 48N in the work area in a safe location approximately 3.5 feet
off the ground.
2. Turn the unit On. If the battery voltage displayed during the power-on sequence is less
than or equal to 6.4 volts, replace or recharge the batteries.
3. Be aware that the sensors require 10 minutes to stabilize to a new environment.
4. In the main menu, View will be selected (an indicator arrow denotes the selected
menu). Press the I/O Enter key and the measurement screen will appear.
5. Press the Run/Stop key to begin datalogging. Use the Arrow keys to set the display to
the desired items.
Getting Started
Up and Running Overview
Getting Started
Up and Running Overview
Placing the QUESTempº 48N on the Job Site
The QUESTempº 48N should be placed at a height of 3.5 feet (1.1 m) for standing individuals
or 2 feet (.6 m) for seated individuals. Tripod mounting is recommended to get the unit away
from anything that might block radiant heat or airflow. A 1/4" x 20 threaded bushing on the
bottom of the instrument allows mounting to a standard photographic tripod. DO NOT stand
close to the unit during sampling.
Before data logging, allow ten minutes for the sensors readings to stabilize.
Page 7
3 Getting Started
Keys
Explanation
I/O Enter key
The unit turns on with a single key press. The unit turns off by
holding the key down while a countdown of 3-2-1 occurs in the
lower right corner of the display. This key is also used to select a
mode (such as Setup or View) or enter setup changes.
Pressing and releasing the key while viewing temperatures causes
the display to view the next available sensor bar (indicated in the
upper right corner of the display).
Up Arrow key
Changes items appearing in the display. Scrolls up.
Down Arrow key
Changes items appearing in the display. Scrolls down.
Run/Stop key
From the menu or view modes, pressing this key starts or stops the
run mode. Pressing this key will exit the setup, print or reset
modes.
Escaping or moving
back one screen
If you are in the setup, print, reset, or sensor alignment screens,
you can press Run/Stop key to escape or move back one screen.
I/O Enter key
Run/Stop key
Up Arrow key
Down Arrow key
Keypad Operation
Keypad Operation
The unit operates using a keypad with four keys. The I/O Enter key responds when the key is
released while all other keys respond when the key is pressed.
Table 1-1: Keypad explained
Figure 1-2: Keypad keys identified
Page 8
4 Getting Started
QUEST
QT-48N WWB
REVISION 1.25
BATTERY 8.3V
Index: - - - -
<ENTER>
Log Rate: 1 min
Nxt 01 <ENTER>
Turning on, Logging/Capture an Event, and Turning Off
Turning on, Logging/Capture an Event, and Turning Off
To quickly get you started with the QUESTempº 48N, the following section explains turning on
the instrument, data logging/event logging, and stopping.
1. Press the I/O Enter key to turn on. The first two start-up screens detail the name, type of instrument, and then a
revision and battery power screen appears. (Ensure the battery level does not fall
below 6.4V or you will want to replace the battery. See “Power options” for more
details.)
NOTE: If you have data saved on the instrument, an additional screen stating “data
in memory” will appear prompting you to reset the instrument, if desired.
A B
Figure 1-3: Start-up screens (A – indicates model and type
B- indicates revision and battery power)
2.In the Index Start-Up screen, press I/O Enter key.
NOTE: If an index is selected from the Setup menu, this will appear on this screen.
Figure 1-4: Index start-up screen
3. A Log Rate Start-Up screen will appear, press I/O Enter key. (This screen details your
log rate time or event logging setup option and the last saved logged session or event.)
(See Setup for more information on event logging.)
Figure 1-5: Log rate start-up screen
Page 9
5 Getting Started
VIEW PRINT
SETUP RESET
Indicator arrow
Indicates the selected menu option. Either press
up/down arrows to select a new menu or press I/O
Enter key to select.
Numeric counter
Event records or Data logging sessions
displayed
Run indicator.
Will appear when you are in “Run mode”.
For an event, it will appear for about 1-2
seconds as an event is recorded in the
instrument’s history.
WET 53.9º F 01
DRY 113.5º F
Turning on, Logging/Capture an Event, and Turning Off
4. The main menu (navigational screen) will appear after the start-up screens. Press I/O
Enter key. (This will open to the measurement screens.)
Figure 1-6: Main menu of the QT48N
5. Press Run/Stop key to either Data Log or log/record an Event and an asterisk will appear
in far right corner.
NOTE: Data logging or an Event is selected via the Setup menu.
Numeric counter
oEvent logging/recording: As you record events, the numeric counter will
increase by one displaying the number of saved events.
oData logging mode: This will display the number of saved data logged
sessions instead of Events. (The process is identical to event mode.)
6. To stop data logging, press and hold Run/Stop key.
NOTE: Skip this step for an Event log. An event log will capture the measurements
with the initial run/stop key press.)
7. To view different measurements, press the Up or Down Arrow key to toggle through the
views.
8. To return to the main menu, press and hold the I/O Enter key (3, 2, 1 countdown will
appear) and the main menu will display.
To select an option on the main menu, press the Up or Down Arrow until an arrow
appears directly in front of the appropriate menu selection and then press I/O enter
key.
Figure 1-7: Measurement screen (example)
Page 10
6 Sensors
The dry bulb thermometer (right
position) measures the ambient air
temperature. This measurement is
used in the outdoor WBGT
calculation when a high solar radiant
heat load may be present. The
series of white plates surrounding
the sensor shield it from radiant heat.
A B C
A. Globe thermometer
B. Relative humidity
sensor
C. Dry bulb thermometer
Sensors
Globe Thermometer
9. To power off, press and hold the I/O Enter key from the main menu.
Sensors
Globe Thermometer
The globe thermometer (left position) gives an indication of the radiant heat exposure on an
individual due to either direct sunlight or hot objects in the environment. This is accomplished
by placing a temperature sensor inside a blackened copper sphere and measuring the
temperature rise. The WBGT index is based on the response of a 6 inch diameter globe. The
QUESTemp° uses a 2-inch diameter globe for a faster response time. The temperature of the
2 inch globe is correlated to match that of a 6 inch globe.
Waterless Wetbulb & Relative Humidity Sensor
The relative humidity sensor (middle position) is used to calculate the Waterless Wetbulb from
a combination of dry bulb temperature, humidity and wind speed measurements. The
waterless wetbulb is used to calculate an estimated WBGT value.
Dry Bulb Thermometer
Figure 1-8: Sensors identified
Page 11
7 Measurements
**NOTE: Navy and Marine calculations are all based on the outdoor computation.
WBGT (outdoor) = 0.7WB + 0.2G + 0.1DB (denoted as “WBGT” on the display)
Measurements
WetBulb Globe Temperature (WBGT)
Measurements
The QUESTemp 48N data logging, with an event logging mode, heat stress monitor directly
senses three parameters: dry bulb temperature (DB), globe temperature (G), and relative
humidity (RH).
It computes the Wet Bulb (WB), the Wet Bulb Globe Temperature (WBGT), and the Stay
Times Indices for ACGIH, U.S. Navy PHEL charts, and Flag Conditions for U.S. Navy/Marine
Corp. Ashore.
WetBulb Globe Temperature (WBGT)
The WBGT is a type of temperature index which combines the effects of temperature,
humidity, radiant heat, and air flow and provides a relatively simplistic tool to analyze thermal
comfort. The resulting WBGT values can then be compared to indices of work-rest regimens
(stay times) based upon workloads.
The WBGT calculation is a weighted average of the three temperature sensors using the
following formulas:
Stay Times/Rest Times (Indicies)
Stay Times Indices represent how long a worker should be able to safely work under heat
stress conditions. Select one of four indices for displaying and printing from the unit: ACGIH
Stay Times, NAVY PHEL’s, or Flag Conditions for U.S. Navy/Marine Corp. Ashore. Refer to
Appendix B for more information on the indices.
NOTE: It will appear as “Index” on the measurement screen.
Remote measurements
The top sensor bar may be removed from the instrument and used through a remote cable.
(Refer to Appendix C for remote cable options.) Shelter the instrument and remote the sensor
bar if the measured environment is expecting heavy rain or if temperatures are above 60°C.
Page 12
8 Operating QUESTempº 48N
From a powered on mode, the main menu will
appear (after the start-up screens are
displayed.) Use the Up Arrow and DownArrow keys to move the marker in the display
in front of the desired mode. Pressing the
I/OEnter key will select the mode.
VIEW PRINT
SETUP RESET
Figure 1-9: Main menu example
View mode
Arrow icon Indicates you are in view mode
Event #
Last logged event/session
WET 53.9º F 01
DRY 113.5º F
Operation
View
Operating QUESTempº 48N
View
Displays the measured data but does not log it. If more than one set of sensors is plugged into
the unit, they can be displayed by pressing and releasing the I/O Enter key. The displayed
sensor set is shown in the upper right corner.
NOTE: To return to the main menu, hold down the I/O Enter key while a 3, 2, 1
countdown is shown in the lower right corner of the display.
Figure 1-10: Viewing measured data
Page 13
9 Operating QUESTempº 48N
Settings
Explanation
Temperature units
Selectable: ºF/ºC
Language setting
Selectable: English, Spanish, French, Italian, and German
Time and Date
Time settings: 24-hour clock
Day-month-year format
Log Rate
1, 2, 5, 10, 15, 30, 60 minutes log rate options
Event Log: when run/stop is pressed, an event or
snapshot of the current conditions is logged into memory.
The events range from 0-99.
Index/Stay Times
ACGIH, PHEL, FLAG, or none (denoted “----”)
VIEW PRINT
SETUP RESET
Press I/O Enter key to select Setup.
Fahrenheit
English
Temperature setting – press I/O Enter
key to switch between values.
Setup
Setup
The setup options are from the main menu when Setup is selected. At any time, to exit (or
escape) a screen, press the Run/Stop key. Table 1-2 identifies the setup parameters.
Selecting Setup parameters
1. From the main menu, select Setup by pressing the I/O Enter key. Using the Up/Down
Arrow keys, select the settings in steps 2-6.
Table 1-2: Settings indentified
Figure 1-11: Selecting temperature setting and setup
2. For the Time and Date settings, once selected, press I/O Enter key to move to the
update field. (An underline will appear under the first field.)
Press Up/Down Arrows to change the number/month. Press I/O Enter key to
toggle through each field. Repeat as necessary until all values are selected.
3. For the Temperature setting, either Fahrenheit or Celsius will appear. Press the
Up/Down Arrow to select and I/O Enter key to change.
Figure 1-12: Selecting temperature setting
Page 14
10 Operating QUESTempº 48N
Event
Example of setting Log Rate to
Event Log
Next event number
*To increase, press up arrow key
(used to simulate skipping an event.)
Log Rate: event
Nxt event: 11
Print
4. For the Language setting, once selected, press I/O Enter key. To select a language,
press Up/Down Arrow keys. Once selected, press Run/Stop key to return to setup
parameters.
5. For the Log Rate, once selected, press I/O Enter key.
Repeatedly press I/O Enter keys to select either the Log Rate Event or a Log Rate
interval (for data logging; e.g. “5”).
Optional: changing Next Event field. Events are numbered starting at 1 and
incrementing. The number may be increased to simulate skipping events. This may
be useful if the events represent locations and the starting location is somewhere
other than 1.
Figure 1-13: Selecting log rate with event log
6. For the Index setting, once selected press I/O Enter key. Press Up/Down Arrow keys to
select: “---” (which is equivalent for no index), ACGIH, PHEL, or Flag.
7. Exit Setup by pressing the Run/Stop key.
Print
Allows printing to a parallel or serial printer or to a computer. The QUESTempº 48N Heat
Stress Monitor will recognize the cable plugged in and configure itself for serial or parallel. If
no cable is plugged in, it will default to serial. Press I/O Enter key to begin printing. Press
Run/Stop key to return to the menu.
NOTE: If you want to stop the printing, press I/O Enter key until you return to the main
Reset
Resetting enables you to clear the logged data from memory. Press the I/O Enter key to enter
the Reset mode. Clear the memory by holding down the I/O Enter key while the display
counts down from three.
menu. When the printer has stopped printing, remove the cable from the printer to
the instrument.
Page 15
11 Operating QUESTempº 48N
GLOBE 90.3º F 05
RH 53.3º F
Run Indicator
Run
Run
For Data Logging setup, the run mode begins a session in memory and logs the data. For
Event Logging, each run/stop is stored as an event and a numeric counter displays the
number of saved events.
1. Begin a session by pressing the Run/Stop key from the view mode (or measurement
view). An asterisk in the lower right corner indicates the run mode.
To toggle through the views, press the Up or Down Arrow.
2. End the session by pressing the Run/Stop key again. (The session will stop recording
when the asterisk is no longer displayed.)
NOTE: If the logging memory is full or if there are no sensors plugged into the unit,
attempting to enter the Run mode will result in an error message. If the memory
capacity is exceeded, the asterisk in the lower right corner of the display will turn
into an “F” and the memory remaining screen will show “0.0”.
Figure 1-14: Run mode indicator
Page 16
12 Displayed Items
Screen 1: WET (Wet bulb)
DRY (Dry bulb)
WET 80.5º F 01
DRY 92.2º F
Screen 2: GLOBE and RH
GLOBE 40.4°C 03
RH 52.5%
Screen 3: WBGT
WBGT 30.7°C 3
Screen 4: Time (24 hour format)
Date (day, month, year)
TIME 11:04:13 04
DATE 14-MAY-18
Displayed
Items
Run
Displayed Items
For the QUESTemp 48N Heat Stress Monitor, the number in the upper right corner indicates
the saved session or the saved event log.
“1”indicates the sensor bar placed on (or attached to) the top of the instrument. Sensors
2 and 3 are labeled on the side of the unit as “Sensor 2”, and “Sensor 3”.
An “*” asterisk in the lower right corner indicates that the unit is in the run mode and is
logging data or saves an event.
The following measurements can be accessed on the display:
Figure 1-15: Wet and Dry screen
Figure 1-16: Globe and RH screen
Figure 1-17: WBGT screen
Figure 1-18: Time and Date screen
Page 17
13 Displayed Items
Screen 5: BAT (Battery voltage)
MEM (Logging memory
available in days)
BAT 09:04:13 05
DATE 10.4 dy
Screen 6: Index
(Navy PHEL stay times)
PHEL_5 3:10 05
PHEL_6 8:03
L M H VH 04
60 45 30 15
PHEL_5 3:10 05
PHEL_6 8:03
PHELs stay time is greater than 8 hours.
Stay Time/Index
NOTE: A series of dashes appear in the display if one of the following occur:
The temperature is outside of its allowable range
A temperature sensor has failed
Stay times temperatures are outside of the their defined range
Stay Time/Index
The screen(s) displaying stay time data appear different for each of the possible indices.
If ACGIH is selected, the recommended working minutes per hour are shown for each of the
workload categories Light (L), Moderate (M), Heavy (H), and Very Heavy (VH).
Figure 1-19: Battery & memory screen
Figure 1-20: Navy PHEL stay times
If the Navy PHELs are selected, the recommended working hours are shown based on a
maximum of eight hours. Three screens are used to display the PHELs two at time.
NOTE: “8:01” following one of the PHELs indicates greater than eight hours.
Figure 1-21: Index for ACGIH
Figure 1-22: Navy PHELs screen
Page 18
Data Logging
14
HEAT CATEGORY 05
YELLOW FLAG
Navy/Marine Corp’s heat condition flag
warning example
Log Rate
1 min
2 min
5 min
10 min
15 min
30 min
60 min
1 sensor
11.2
22.5
56.2
112.4
168.6
337.3
674.5
2 sensors
5.6
11.2
28.1
56.2
84.3
168.6
337.3
3 sensors
3.7
7.5
18.7
37.5
56.2
112.4
224.8
Data Logging
Stay Time/Index
If Flag is selected, the Flag Conditions for Navy/Marines Corp. Ashore warning
system screen will appear. There are five flag systems, no flag, yellow, green, red, and black,
which provide heat exposure guidelines for acclimated individuals. (For more information, refer
to the “Flag Conditions US Navy/Marine Corp. Ashore”.)
Data Logging
Data from each sensor is recorded at the interval set by the logging rate. Every time Run/Stop
is pressed, a session is either started or ended in memory. Each session contains a header
with time, date, and summary information.
Memory Table: Gives the number of logging DAYS.
Figure 1-23: Flag Conditions Navy/Marine Corp. Ashore
Table 1-3: Example of a Memory table
Page 19
15 Electronic Sensor Check
Example of Verification module
Electronic
Sensor Check
Performing an Electronic Sensor Check
Electronic Sensor Check
A verification module, TSI model 053-923, may be used to check the operation of the QUESTemp
monitor’s wet bulb, dry bulb, and globe. The purpose is to verify that the electronic components
are within a specific range with known values and a known source. The temperature tolerances
should be within ±0.5oC.
NOTE: If the sensors are outside of the tolerances, this indicates the sensor alignment should be
serviced off-site for calibration.
Performing an Electronic Sensor Check
1. Ensure the instrument is reading in Celsius prior to your electronic sensor check.
To change the temperature setting, select Setup from the main menu. Then, either
Fahrenheit or Celsius will appear on the screen. When selected, press I/O enter key to
switch between settings. For more information, refer to step 3.
2. Remove the top sensor bar, place to the side, and plug in the verification module into the
3. Verify the measurement readings on the screen are within ±0.5C tolerance to the readings
4. Once completed, remove verification module and place sensor bar back on the instrument.
center pins of the sensor housing.
printed on the verification module label. Example below:
a. Wet Bulb (WB): 11.1C b. Dry Bulb (DB): 45.3C
c. Globe (G): 69.2C
NOTE: Relative humidity (RH) is not valid on this instrument.
(Tighten down the two bolts.) Change the Celsius reading back to Fahrenheit. (Refer to
step 1 a-b.)
Page 20
Sensor Alignment
16
VIEW PRINT
SETUP RESET
WET 50.7° F
DRY 70.6° F
(RH Sensor) Alignment menu
Alignment RH
WBGT HIGH LOW
“High” is selected when the arrow appears in
front of the text.
NOTE: WBGT is not used in sensor alignment
Sensor
Alignment
RH Sensor Alignment
Sensor Alignment
For highly accurate measurement readings, you should align your QUESTempº 48N prior to data
logging or event logging/recording. For the RH sensor, it is recommended to align with a High
concentration level using the 75% sensor salt, NaCl (sodium chloride), or a Low concentration
level using the, the 33% sensor salt, MgCl (magnesium chloride). (To order sensor alignment
salts, refer to an online vendor such as, colepalmer.com.)
RH Sensor Alignment
1. To open, navigate to the main menu and select View by pressing I/O Enter key (see A).
A measurement screen will appear (see B).
Figure 1-24: Main menu with view selected (A) & measurement screen (B)
2.Press and hold I/O Enter key and then press Down Arrow key from the View menu. The
Alignment screen will appear.
A B
3. Press Up/DownArrow key to select either High or Low. Then press I/O Enter key.
4. Remove (or slide) the Sunshield (white globe) from the RH sensor and place it to the side.
(See Figure 1-26).
Figure 1-25: Alignment screen
Page 21
Sensor Alignment
17
Sunshield:
Slide off RH sensor
to remove.
Place unrolled finger
cot, with a small
hole cut in the top,
so it is completely
stretched over the
humidity sensor.
Alignment salt
(remove rubber cap
and place over RH
sensor with the
sunshield removed)
Once Alignment salt
is placed, allow levels
to stabilize (about
15 mins).
Low RH Alignment screen
Adjust % to 33% (using up/down arrow keys)
and press Enter (after level has stabilized) to
store sensor alignment level.
RH Sensor Alignment
5. Place a latex finger cot (user supplied) over the humidity sensor in order for the sensor to
stabilize quicker about 15 minutes (or up to 60 minutes without).
NOTE: Latex finger cots can be found at various hardware stores or online such as
colepalmer.com.
Figure 1-26: Sunshield & placement of finger cot prior to RH alignment
6. Place the Salt container, with the salt alignment cap removed, over the RH sensor (middle
sensor). Allow level to stabilize for at least 15 minutes. (Refer to Figure 1-25.)
About sensor alignment: if readings are within ±0.5%, a change in sensor alignment is
not required. Skip to step 7 to return to main menu.
NOTE: For a High alignment, use the 75% sensor alignment salt (NaCL) and for a Low
alignment, use the 33% alignment salt (MgCl).
Figure 1-27: RH alignment with Low % example
Page 22
Sensor Alignment
18
Press Enter to save the alignment.
Example of high RH percentage prior to
sensor alignment. Press up/down arrows to
adjust level after salt has stabilized.
RH 51.2%
<ENTER> HIGH
Sensor alignment with High % RH level
NOTE: WBGT is not used in sensor alignment
RH Sensor Alignment
7. In either the High/Low RH alignment screen, adjust the level to the percentage displayed on
the salt container by using the Up/Down Arrow keys. Press I/O Enter key to save. (The
instrument will state either “Successful” or “Failure” if it passed/did not pass. For “Failure”
repeat the steps above.)
Figure 1-28: RH alignment with High% screen
8. To return to the main menu, press Run/Stop key.
9. Verify the RH sensor is ±0.5% of the specific percentage. If not repeat the steps above.
10. Remove the alignment salt and finger cot. Replace the Sunshield over the RH sensor when
completed with the sensor alignment.
Page 23
19 Printing
Printing
Data/Communications
Printing
Printing options include to a RS-232 port or to a parallel printer. Serial transmission requires
TSI cable #54-715. Parallel transmission requires TSI cable #56-875. With the applicable
cable plugged in, select PRINT from the menu and press the I/O Enter key to enter the
PRINT mode. Begin printing by pressing the I/O Enter key. Press the key again to abort the
printing.
Data/Communications
TSI Detection Management Software DMS is recommended for downloading, storing,
graphing, and printing reports (see below). Communications programs, such as Windows®
HyperTerminal® may also be used to capture the printout into a file. The baud rate is fixed
at 9600.
MAXIMUM LEVELS, Sensor(WEIGHTED AVERAGE)
WBGT 71.4 21-FEB-08 11:10:14
Parallel
Parallel
Data can be sent directly to parallel printers that accept direct ASCII test input without special
drivers. Make sure the printer is powered on and is online, ready to accept data, prior to
printing.
Figure 1-30: Sample printouts
Page 25
Powering
21
The QUESTempº 48N Heat Stress Monitor uses a 9-volt
alkaline battery. A door on the back of the instrument allows
the user access to the 9-volt battery.
The 2-position switch located in the battery compartment is
set by the user to the 9-volt battery.
Figure 1-31: 9-volt battery
9-Volt Alkaline Battery Replacement
Powering
9-Volt Alkaline Battery Replacement
WARNING: Replace batteries only in a non-hazardous environment.
The 9-volt battery should be replaced should be recharged when the voltage drops below
6.4 volts. The battery voltage is displayed when the instrument is turned on. While turned on,
the battery voltage can be displayed at any time by pressing the up or down arrow keys to
move through the display until the battery voltage screen appears. If, while operating, the
battery voltage drops below 6.4 volts, the display will automatically switch to the display
showing the battery voltage along with a low battery message. After a low battery occurs, the
unit will continue to operate for approximately 8 hours. When the battery voltage falls to
6.2 volts or below, the unit will automatically turn off.
Replace only with an approved 9-volt alkaline battery.
Temperatures given in Celsius or Fahrenheit
Index (displayed stay time with selected index): ACGIH TLV, U.S. Navy PHEL Charts,
Flag Conditions for U.S. Navy/Marine Corp. Ashore
Data Logging mode and Event logging mode
Data logging mode records and prints all measurements at user selected interval of 1, 2,
5, 10, 15, 30, or 60 minutes.
Event logging mode (1 second record) 128K bytes of data memory.
Languages
English, French, Spanish, Italian, German
Sensors
Dry Bulb Sensor- 1000 Ohm Platinum RTD
o Accuracy and Ranges: ±0.5 from 0°C to 120°C (±0.9°F from 32°F to 248°F)
Waterless Wet Bulb (Humidity) Sensor
oAccuracy and Ranges: Expanded measurement uncertainty of 1.1°C (k=2)
between 0°C and 80°C (32°F and 176°F)
Globe Sensor- 1000 Ohm Platinum RTD
oAccuracy and Ranges: ±0.5 from 0°C to 120°C (±0.9°F from 32°F to 248°F)
Relative Humidity Sensor - Integrated circuit with capacitive polymer sensor
o Accuracy and Ranges: ±5% from 20 to 95% (non-condensing)
Operating Temperature Range
Sensor Assembly: -5C to +100C
Electronics: -5C to 60C
Housing
Designed water resistant to a light rain or mist. If rain is frequent, best practice would be
to remote the sensor bar and keep the instrument sheltered.
Page 27
Appendix A: Specifications
23
Size
Size
Height 9.2 in. (23.5 cm); Width 7.2 in. (18.3 mm); Depth 3.0 in. (7.5 mm)
Dimensions include mounted sensor assembly
Weight
2.6 lbs. (1.2 kg) with mounted sensor assembly
Remote Sensor Bar
Using a cable, the top sensor bar can be setup for remote measurements, up to 200 feet
(61 m).
Power
9V alkaline
Battery Life
9V alkaline: 80 hours
Page 28
24 QUESTemp 48N Block Model
Keypad
Sensors
1, 2, & 3
Multiplexer
Gain
A/D
Microprocessor
LCD
RS232
Printer
Power
Controller
Battery
Data Memory
Program
Memory
Appendix A:
Block Model
Battery Life
QUESTemp° 48N Block Model
Page 29
Appendix B
25
Work and recovery
(TLV )
Light
Moderate
Heavy
Very Heavy
75% to 100%
31.0
28.0
26.0*
23.5*
50% to 75%
31.0
29.0
27.5
25.5*
25% to 50%
32.0
30.0
29.0
28.0
0% to 25%
32.5
31.5
30.5
30.0
Work and recovery
(Action Limit )
Light
Moderate
Heavy
Very Heavy
75% to 100%
28.0
25.0
22.5*
20.0*
50% to 75%
28.5
26.0
24.0
22.5*
25% to 50%
29.5
27.0
25.5
24.5
0% to 25%
30.0
29.0
28.0
27.0
Appendix B:
Heat Exposure
Tables
ACGIH
Appendix B: Heat Exposure Tables
ACGIH
Screening Criteria for Heat Stress Exposure. WBGT values in C.
NOTE: According to the ACGIH’s guidelines, the temperature values represent a work and
rest process which is explained in the standards. Refer to the ACGIH TLVs and BEIs for
specific details.
*Values not specified by ACGIH have been estimated for continuity.
Page 30
Appendix B
26
Clothing type
Clothing correction
(Addition to WBGT (ºC)
Work clothes (long sleeve shirt and pants)
0º
Cloth (woven material) coveralls
0º
Double-layer woven clothing
3º
SMS polypropylene coveralls
0.5º
Polyolefin coveralls
1º
Limited-use vapor-barrier coveralls
11º
WBGT(F)
I
II
III
IV V VI
80.0
>8:00
>8:00
>8:00
8:00
6:35
4:30
81.0
>8:00
>8:00
>8:00
8:00
6:35
4:30
82.0
>8:00
>8:00
8:00
7:05
5:25
3:40
83.0
>8:00
8:00
7:45
6:25
4:55
3:20
84.0
>8:00
8:00
7:05
5:55
4:30
3:05
85.0
8:00
7:45
6:30
5:20
4:05
2:50
86.0
8:00
7:05
5:55
4:55
3:45
2:35
87.0
7:25
6:30
5:25
4:30
3:25
2:20
88.0
6:45
5:55
4:55
4:05
3:10
2:10
89.0
6:10
5:25
4:30
3:45
2:50
2:00
ACGIH Clothing Corrections
ACGIH Clothing Corrections
The following clothing corrections are in degrees Celsius. When a clothing correction is
entered into the setup portion of the QUESTemp 48N Heat Stress Monitor, the value is
added to the WBGT only for looking up the stay times. The WBGT value displayed by the unit
does not reflect corrections.
Cited from American Conference of Governmental Industrial Hygienists, Threshold Limit Values and Biological Exposure Indices for 2008. Reprinted with permission.
United States Navy
Physiological Heat Exposure Limits (PHEL) Time Table
(Without the presence of fuel combustion gases/fuel vapors)
The recommended working hours are shown based on a maximum of eight hours. Naval
personnel will follow a category, I - VI, based upon their function.
PHEL Curves (Total Exposure Time in Hours: Minutes)
Page 31
Appendix B
27
WBGT(F)
I
II
III
IV V VI
90.0
5:40
5:00
4:10
3:25
2:40
1:50
91.0
5:15
4:35
3:50
3:10
2:25
1:40
92.0
4:50
4:10
3:30
2:55
2:15
1:30
93.0
4:25
3:50
3:15
2:40
2:00
1:25
94.0
4:05
3:35
3:00
2:25
1:50
1:15
95.0
3:45
3:15
2:45
2:15
1:45
1:10
96.0
3:25
3:00
2:30
2:05
1:35
1:05
97.0
3:10
2:45
2:20
1:55
1:25
1:00
98.0
2:55
2:35
2:10
1:45
1:20
0:55
99.0
2:40
2:20
2:00
1:40
1:15
0:50
100.0
2:30
2:10
1:50
1:30
1:10
0:45
101.0
2:20
2:00
1:40
1:25
1:05
0:45
102.0
2:10
1:50
1:35
1:15
1:00
0:40
103.0
2:00
1:45
1:25
1:10
0:55
0:35
104.0
1:50
1:35
1:20
1:05
0:50
0:35
105.0
1:40
1:30
1:15
1:00
0:45
0:30
106.0
1:35
1:25
1:10
0:55
0:45
0:30
107.0
1:30
1:15
1:05
0:50
0:40
0:25
108.0
1:20
1:10
1:00
0:50
0:35
0:25
109.0
1:15
1:05
0:55
0:45
0:35
0:25
110.0
1:10
1:00
0:50
0:40
0:30
0:20
111.0
1:05
1:00
0:50
0:40
0:30
0:20
112.0
1:00
0:55
0:45
0:35
0:25
0:20
113.0
0:55
0:50
0:40
0:35
0:25
0:15
114.0
0:55
0:45
0:40
0:30
0:25
0:15
115.0
0:50
0:45
0:35
0:30
0:20
0:15
116.0
0:45
0:40
0:35
0:25
0:20
0:15
117.0
0:45
0:40
0:30
0:25
0:20
0:10
118.0
0:40
0:35
0:30
0:25
0:15
0:10
119.0
0:35
0:35
0:25
0:20
0:15
0:10
120.0
0:35
0:30
0:25
0:20
0:15
0:10
121.0
0:35
0:30
0:25
0:20
0:15
0:10
United States Navy
Page 32
Appendix B
28
WBGT(F)
I
II
III
IV V VI
122.0
0:30
0:25
0:20
0:15
0:15
0:10
123.0
0:30
0:25
0:20
0:15
0:10
0:10
124.0
0:25
0:25
0:20
0:15
0:10
0:05
No Flag
Green
Yellow
Red
Black
Flag Degrees F
<80.0
80.0–84.9
85.0–87.9
88.0–89.9
>90
Flag Degrees C
<26.7
26.7–29.4
29.4–31.1
31.1–32.2
>32.2
Flag Conditions for U.S. Navy/Marine Corp. Ashore
Flag Conditions for U.S. Navy/Marine Corp. Ashore
The following chart details the heat stress monitoring Flag Conditions for U.S. Navy/Marine
Corp. Ashore in Degrees Fahrenheit and Celsius.
*Rest means minimal physical activity (sitting or standing) and should be accomplished in the
shade if possible.
Storage case (one included) ........................................................................................... 053-922
QT48N User Manual (one included) ............................................................................... 057-330
Page 34
Appendix D: Software
30
QT⁰ 48 data port
Connected and powered on
Connected to serial port
Appendix D:
Software
Appendix D: Software
The QUESTemp 48N Heat Stress Monitor has the flexibility to be set up and controlled
through computer software. The programmable start and stop time feature is only accessible
through the computer. The instrument also has the capability of sending live data while
measuring. These features are best utilized using TSI Detection Management Software DMS.
DMS quick overview
The focus of this section is to briefly introduce the following QSP-II topics: Downloading your
data, setting up parameters, and viewing your data in charts, graphs, and reports. (For further
details on DMS, refer to the online Help and select Contents.)
Communicating and Downloading in DMS
In order to download, review the data, and setup parameters in DMS software, this will require
connecting the QT⁰48N to a computer. The steps below explain connecting and
communicating to DMS.
1. Using a TSI cable, part number 054-715, plug the cable into the computer and plug the
opposite end into the side jack of the QT⁰ 48 data port.
Figure 1-32: Communicating with the QT⁰ 48N and DMS
2. From the start page of DMS, select button and the instrument
communication panel will appear.
3. Select Heat Stressand then select the Model Type by clicking on QT⁰48N. (See
Figure 1-34).
Page 35
Appendix D: Software
31
Data Finder checkbox. When
checked, this will automatically
open to the downloaded data
page.
Download
❶
Double-click
to select/view
QT⁰44/46
data in charts
and graphs
❶
Data Finder and Quick Reports
4. Click on the button in the Instrument Communications page while Data
Finder checkbox is checked (see ❶ below).
NOTE: The Data Finder page will appear. See next section.
Figure 1-33: Communicating with the QT⁰ 48N and DMS
Data Finder and Quick Reports
In the Data finder page, the data you downloaded is stored by instrument family and then
organized by models. NOTE: When working in this page, click on the Most Recent button if
you are looking for your recently downloaded data.
1. In the Data Finder page, click on the downloaded data and select either Analyze or
Print Report.
Figure 1-34: Data Finder Page
Page 36
Appendix D: Software
32
Data Finder and Quick Reports
Quick Report: select the button to generate a report with basic
information and a summary table of your downloaded data.
Figure 1-35: Quick Report
Page 37
Appendix D: Software
33
Data Finder and Quick Reports
Analyze: allows you to add charts/graphs and customize how they will appear in a
report. Tip: Add the panels by using the add a panel and dragging them into the
appropriate location.
Press button and it will generate a report based on the where the
charts/graphs (panels) are placed.
Figure 1-36: Analyze button opens to the panel layout page
1060 Corporate Center Drive
Oconomowoc, WI 53006-4828
Europe, Middle East, and Africa
TSI Instruments Ltd.
Stirling Road
Cressex Business Park
High Wycombe, Bucks
HP12 3ST
United Kingdom
Telephone: 1-800-680-1220 (USA);
+1 651-490-2860 (Outside
USA)
E-mail: technical.services@tsi.com
Telephone: +44 (0) 149 4 459200
E-mail: tsiuk@tsi.com
Contact/Service
Information
Technical Support Contacts
Contact/Service information
This section gives directions for contacting TSI Incorporated for technical information and directions
for returning the QUESTemp° 48N Heat Stress Monitor for service.
Technical Support Contacts
If you have any difficulty setting or operating the instrument, or if you have technical or application
questions about this system, contact TSI’s Technical Support.
Service Contact Information
If your instrument does not operate properly, or if you are returning the instrument for service, visit our
website at http://rma.tsi.com for a Return Material Authorization, or contact Customer Service.
Page 39
Calibration
35
Returning for Service
Visit our website at http://rma.tsi.com and complete the on-line “Return Merchandise Authorization”
form or call TSI at 1-800-680-1220 (USA) or (651) 490-2860, or 001 651 490 2860 (International) for
specific return instructions.
Customer Service will need the following information:
The instrument model number
The instrument serial number
A purchase order number (unless under warranty)
A billing address
A shipping address
Use the original packing material to return the instrument to TSI. If you no longer have the original
packing material, seal off any ports to prevent debris from entering the instrument and ensure that the
display and the connectors on the instrument front and back panels are protected. This instrument is
very fragile and must be packed in a manner appropriate for a precision instrument.
Calibration
The QUESTemp⁰ 48N Heat Stress Monitor and TSI field calibrator devices should be examined
regularly by the factory. An annual calibration is recommended. (See Service Information above.)
Page 40
Warranty
36
Warranty
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator's
manual, to be free from defects in workmanship and material for 12 months, or if less, the length of time specified in the
operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This
limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when indicated in
specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material,
under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses, batteries or other
consumable materials. Only the original manufacturer's warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that the instrument or product is properly
calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory authorized service center with the one
exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended
cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or intentional damage, or is not
properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a
separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods
which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS
OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH
RESPECT TO SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS
LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED
INFRINGEMENTS. BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE
DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS
WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF
SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING
CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN
OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE
REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE
DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN
NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL,
INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR
INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be
brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to
Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the
complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended,
modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy
is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales
office or representative, or call TSI’s Customer Service department at 1-800-874-2811 (USA) or +001 (651) 490-2811
(International).
Page 41
USA Tel: +1 800 874 2811
UK Tel: +44 149 4 459200
France Tel: +33 1 41 19 21 99
Germany Tel: +49 241 523030
India Tel: +91 80 67877200
China Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
TSI Incorporated – Visit our website www.tsi.com for more information.