Failure to observe the following procedures may result in serious personal injury:
Not for use in Explosive or Hazardous locations. This product is NOT Intrinsically Safe.
Contains built-in lithium ion battery pack. DO NOT incinerate or dispose of in fire.
DO NOT disassemble, alter, or re-construct the li-ion battery.
Contains Class 3B visible laser radiation source (658 nm, nominal 5 mW). DO NOT look
directly into beam or view with telescopic devices.
This instrument will NOT detect natural gas.
WARNING!
Read the manual before operation.
DO NOT store in temperatures exceeding 60°C (140°F) and in direct sunlight.
DO NOT immerse in liquids.
The Sensors and batteries in this instrument have limited shelf-life, even if never used.
DO NOT disassemble, service, or clean the optical engine internal components while power is
applied. The laser source can hurt you.
Caution!
General
Condensation may damage your instrument and its sensors.
A non-condensing environment is required for proper measurements.
DO NOT charge batteries outside the range of 0°C to 40°C (32°F to 104°F).
Battery run-time may be somewhat reduced at lower than 20°C (68°F) temperatures.
Caution!
Temp/RH/Dew Point
The internal Temp/RH sensor in this instrument is not designed for use in bright sunlight or in the
presence of a strong radiant heat sources. It is intended primarily for indoor use. It is not equivalent
to a protected dry-bulb thermometer.
Dangers, Warnings, & Cautions
ENVIRONMENTAL MONITOR
EVM SERIES
EVM Series User Manual
Page 4
Dangers, Warnings, & Cautions
iv
Caution!
PM/dust
Clean the optical engine mirror with a soft cloth, if it is dirty. Scratches will affect performance.
Avoid measuring wet particles, fogs, and liquid mists that will tend to cling to the impactor,
optical engine, tubing, and pump surfaces.
Avoid oily mists as found near some machining operations.
This monitor is designed to measure dry particulates, not filaments.
DO NOT attempt to measure Asbestos with this instrument.
Keep a "Pump Protection Filter" or other gravimetric filter in place to help prolong the life of the
pump.
You must check and calibrate the flow rate, when using this device as a (gravimetric) sampling
pump.
Caution!
CO2
Avoid exposure to condensation, fogs, and liquid mists that can contaminate the optical path inside
the CO2 sensor.
Caution!
PID
This instrument cannot detect methane gas.
For best sensor life, avoid acidic condensing gases.
The presence of moisture can affect performance of the PID.
Caution!
Toxic
Beware of certain cross-sensitivities when measuring toxic gas levels. Readings can be
nonspecific in the presence of certain cross-contaminant gases.
Caution!
ir velocity
The tip of the Air velocity probe is very fragile.
EVM Series User Manual
Page 5
Dangers, Warnings, & Cautions
v
FCC Compliance Statement
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial or residential
environment. This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause harmful interference to
radio communications. If harmful interference occurs, the user will be required to correct the
interference at his or her expense.
EVM Series User Manual
Page 6
Page 7
Table of Contents
vii
Table of Contents
Table of Contents
Dangers, Warnings, & Cautions ....................................... iii
Table of Contents.............................................................. vii
Table 5-4: Past session files soft keys explained ............... 85
EVM Series User Manual
Page 11
Introduction
1
Caution
The EVM is intended to measure gases that are potentially dangerous to human health. To help
ensure worker safety through the proper use of your unit, it is important that you not only read but
also understand the contents of this manual. Familiarize yourself with the unit before using it in a
potentially hazardous situation.
CHAPTER
1
Table of Contents
Introduction
The EPA (Environmental Protection Agency) conducted a study over the last 25 years (1975-2000) and
determined that 107 measurable air pollutants, specifically carcinogens, exist in working environments.
Combined with the advent of closed windows, modern office buildings and various cleaning solvents,
the outcome has been energy efficient buildings and homes which contain levels of carcinogens. At high
levels, this is known as “Sick Building syndrome” and can cause acute effects such as headaches,
eye/nose/throat irritation, dry cough or itchy skin, fatigue, and trouble concentrating just to name a few.
Area monitoring or spot checking for high levels of pollutants, such as particulates, VOCs, and toxics,
should be a common practice to ensure good indoor air quality exists in your environment. If poor air
quality is determined, engineering controls such as proper air ventilation and the removal of toxic gases,
would be put into place in order to provide an adequate working/living environment.
The EVM is equipped to measure several different airborne containments or air pollutant applications
including:
Particulate matter sampling (dust monitoring)
Gas sampling
Temperature sampling
Relative Humidity and Dew Point sampling
To get you started quickly with the EVM, the first chapter is dedicated to introducing basic particulate
matter and toxic gas information. The subsequent chapters will guide you through setting up the EVM,
running studies, and assessing your results with easy to follow steps and illustrations.
.
EVM Series User Manual
Page 12
Introduction
2
(1) TURRET
Air travels though the impactor and the
larger, heavier particulates will stick to
the greased plates within the impactor.
(2) OPTICAL ENGINE
While the pump maintains the flow rate,
the smaller lighter particulates will pass
through the optical engine.
(4) PUMP
The remaining clean/filtered air passes
through the pump.
(3) GRAVIMETRIC CASSETTE
The particulates will collect/trap in the
gravimetric filter/cassettes.
(5) FLOW SENSOR
The clean/filtered air passes through
flow sensor (flow-control orifice hole).
(6) OUTLET
Lastly, it passes through to the Outlet on
the back of the instrument.
❶ ❷ ❸ ❹ ❺
❻
Particulate Sampling (Aerosols/Dust Vapors)
Particulate Sampling (Aerosols/Dust Vapors)
The EVM is a portable-area monitoring instrument with a laser-photometer that measures and stores
concentration levels of airborne-dust over time. It is designed to measure and help you identify the air
pollutants, existing as matter (gases and aerosols) in the environment. The methodology used entails a
handful of components including: an air inlet, the impactor (which is the particle size selector), the
collection of particulates, gravimetric sampling, and the pump which are addressed in the upcoming
sections.
Particulate Path
Figure 1-1: Particulate path diagram
EVM Series User Manual
Page 13
Introduction
3
Gases and
Aerosols
Defined
-----Liquid: Mist--------------►
-----Solid: Fume---►
◄--------Spray----------------
◄-------Particulates----------
Aerosols and
Particles
Medical
implications
◄---Oil smokes---►
◄-------------
◄---------------
Cement dust-----------►
---Fly ash----------------------
----Tobacco smoke--►
◄--------------
--- Coal dust---------------►
◄---Pollens-------------►
◄--Atmosphere/Airborne--►
inhalable particulates
◄--Wood.►Dust
◄--Textiles--► Dust
---------Respirable pm---------► (inhaled & penetrates the lungs)
---Thoracic pm------------------------►(inhaled & penetrates the larynx/throat)
Aerosols and gases are experienced in industrial processes, workplaces, and in everyday life. Keeping
in mind that inhalable dust is 100 microns and smaller, this dust is usually filtered through the hairs and
membranes in your sinuses. 10 microns and smaller is considered “thoracic”, 4-10 microns is “inhalable”
reaching into your windpipe and other regions, and 4 microns and smaller is “respirable”. These
particulate sizes will represent what you will be measuring.
The figure below identifies different types and sizes of particulates and how it affects the human body.
.01 0.1 1 10 100 1000
Particle matter (pm) diameter μm
Figure 1-2: Particulates and diameter size
EVM Series User Manual
Page 14
Gas Sensors Sampling
4
Gravimetric sampling is a type of filtering process in which the
particulates are collected and filtered into the gravimetric
cassette after the dust passes through the optical engine of the
EVM. A user supplied sealed gravimetric cassette can be used
in place of the cassette provided by TSI. The sealed cassette
can be sent to a laboratory for analysis. This testing is typically
used to measure personal exposure to inhalable dust
concentrations.
Particulates and Cut-Point
Particulates and Cut-Point
In consideration of how the size of particulates affect the human body, it is important to select a
particulate size cut-point. Essentially, the particle size selector on the EVM is a top-mounted rotary
impactor. Based on your setting, it will filter out all particulates at or above the selected size. The
instrument has four settings which include 2.5 μm, 4 μm, 10 μm, or none (100 μm). (This is explained in
more detail in Chapter 4.)
Collecting/Sensing the Particulates
Once the impactor setting is selected, the EVM collects (or measures) real-time aerosol/dust
concentration using a 90º optical light scattering photometer to determine the total mass concentration
(in units of mg / m3) of particulate matter. Additionally, gravimetric sampling, a more accurate method, is
also used to determine the mass of particulates.
Gravimetric Sampling
Pump
The pump is used in the particulate sampling process as a source of air movement so the particulates
are aspirated (or suctioned out) and collected for measurement. The flow rate is 1.67 Liter per minute.
The pump is typically calibrated before each gravimetric sampling.
Gas Sensors Sampling
The EVM uses a unique smart sensor technology includes automatic sensor recognition and measures
up to three gases simultaneously: a selection of optional nine toxic gases, Carbon Dioxide, combustible
gases, and VOCs.
EVM Series User Manual
Page 15
Gas Sensors Sampling
5
(1) AIR INLET COVER
The path begins with the air pulling
through the air inlet cover.
(2) GAS SENSOR CHAMBERS
The air continues through the gas
sensor bar passing the Toxic, PID, and
CO2 sensor chambers.
(3) Fan
It is continuously ventilated by the fan.
(NOTE: When the pump is off, the fan
can be heard near the louvers on the
back cover.)
(4) Temperature sensor
Measures the ambient air temperature.
❹
Temperature sensor
Gas sensor(s) path
❷
❶
Air inlet cover
❸
Fan
Very quiet!
Gas Sensors and Path
Gas Sensors and Path
TSI’s unique smart sensor technology includes automatic sensor recognition, calibration levels,
temperature compensation information and other valuable data that travels with the sensor from one
unit to another.
The following diagram illustrates the gas sensor path for Toxic, CO2, and PID sensors. It also identifies
the temperature sensor for additional analysis/measurements.
Figure 1-3: Gas sensor path illustration
EVM Series User Manual
Page 16
Introduction
6
Volatile Organic Compounds (VOCs) Sampling
Volatile Organic Compounds (VOCs) Sampling
According to the American Lung Association, Volatile Organic Compounds (also referred to as VOCs)
are one of the top Indoor Air Quality (IAQ) pollutants. These include:
Serious, long-term exposure could lead to health affects including leukemia, cancer, eye and respiratory
irritations and short term effects such as dizziness, memory loss, and visual impairments.
With the EVM you have the following parameters:
Range 0 to 2,000 ppm Isobutylene.
Calibrated with Sensitivity 1.0, using Isobutylene. Response varies with each VOC.
Units selection: ppm, ppb, mg/m
3
(for ppm PID sensor) or µg/m3 (for ppb PID sensor).
Temperature, Relative Humidity, Dew Point
It is important to monitor temperature, control moisture, relative humidity, and monitor dew point levels
in occupied spaces such as office buildings, schools, and hospitals, to name a few. With the presence of
too much moisture mixed with dirt/dust particles, this can contaminate the environment and enable mold
to thrive. High levels of relative humidity (RH) is a contributor of spreading and growing biological
pollutants such as molds, mildew, viruses, mites, building dust, and pollen. Failure to control water
leakage from humidifiers, refrigerators, and ventilation equipment can also lead to high levels of RH and
high Dew Point. For appropriate air quality control, RH levels should be maintained between 30 to
50 percent in a home/office environment. Some of the common side effects include nausea, headaches,
running nose, coughing, and difficulty breathing to name a few.
What the Standards Say
Many organizations have studied and analyzed the effects of toxic gas and particulate exposures. As a
result, standards were developed to regulate exposure. In the United States, the Occupational Safety
and Health Administration (OSHA), the Mine and Safety and Health Administration (MSHA), and the
American Conference of Governmental Industrial Hygienists (ACGIH), NFPA (National Fire and
Protection Agency), National Institute and Technology (NIST), and American Society of Heating,
Refrigeration, and Air-Conditioning (ASHRAE) have all created regulations and limitations on total
permissible exposure levels.
International standards are specified by the health and safety regulations such as the European Union
(EU) and the European Committee of Air Handling and Refrigeration Equipment Manufacturers
(EUROVENT-CEDOMAF). For Canadian standards, the Canadian Standards Association (CSA)
specifies air quality guidelines/regulations.
EVM Series User Manual
Page 17
What the Standards Say
7
OSHA
http://www.osha.gov
NFPA-654
http:///www.nfpa.org
MSHA
http://www.msha.gov
NIOSH
http://www.cdc.gov/niosh
ACGIH
http://www.cdc.gov/niosh
ASHRAE
http://www.ashrae.org
NIST
http://www.nist.gov
IEC
http://www.iec.ch
CSA
http://www.csa.ca
Resources
Resources
To access some of these copyrighted standards, refer to the following:
EVM Series User Manual
Page 18
EVM Series
8
Industrial setting - Used to monitor and ensure health
and safety in settings where toxic fumes are released
(such as testing levels at a paper mill or a foundry). This
information can be used to identify the source and
particle concentrations.
Sick Building syndrome – Used to evaluate air toxins
and ensure building standards maintain a comfortable
work setting.
Work site assessments.
Military applications.
Outdoor monitoring applications include construction
sites/dust outside of mines.
Regulatory compliance (i.e., OSHA, MSHA, IEC,
ACGIH).
As a mapping device- Layout a grid of high contaminant
areas, and quickly collect sample data at multiple
locations to help identify paths and sources, by relative
readings.
As a survey monitor
Specific area in a plant/facility – used to maintain a
quality work environment and conduct area
monitoring samples throughout an entire facility.
Specific areas in a plant/facility- used to identify
key areas that may be toxic or unsafe and implement
acceptable practices such as changes in the
ventilation.
Time history: manufacturing
Applications
EVM Series
Applications
A real-time, direct reading dust monitor is used in many different applications such as walk-through
surveys in manufacturing, wood making plants and/or bakery shops, background sampling, site dust
measurements, assessment of the effectiveness of dust control systems, and measurement of indoor air
quality (IAQ). It is also used to analyze work-place exposure and help identify dust control techniques
such as training or implementing engineering controls.
EVM Series User Manual
Page 19
EVM Series
9
Models and Sensors
Optional Sensor Accessories
EVM-7 includes:
Particulates
Carbon monoxide (CO) or (1) selectable toxic gas
Selectable toxic gas sensors include:
1. Carbon Monoxide (CO) 2. Chlorine (CL2)
3. Ethylene Oxide (ETO) 4. Hydrogen Cyanide (HCN)
5. Hydrogen Sulfide (H2S) 6. Nitric Oxide (NO)
7. Nitrogen Dioxide (NO2) 8. Oxygen (O2)
9. Sulphur Dioxide (SO2)
Volatile Organic Compounds with PID ppm or PID ppb sensor
ppm – parts per million (0 – 2,000 ppm)
ppb – parts per billion (0 – 50,0000 ppb)
Relative Humidity
Temperature
Carbon Dioxide (CO2)
Air Probe-10
EVM-4 includes:
Carbon monoxide (CO) or selectable toxic gas
Relative Humidity
Temperature
Carbon Dioxide (CO
2
)
Air Probe-10
Toxic sensor/CO sensor
CO
2
sensor
PID sensor
Particulates
EVM-3 includes:
Particulates
Relative Humidity
Temperature
Air Probe-10
Toxic sensor/CO sensor
CO
2
sensor
PID sensor
Models Explained
Models Explained
The EVM Series simultaneously measures and data logs specific sensors/parameters that were ordered
with your instrument. The models, sensors, and accessories are explained in detail below. Refer to the
part’s listing for further details (such as calibration accessories).
EVM Series User Manual
Page 20
EVM Series
10
Models Explained
(This page intentionally left blank)
EVM Series User Manual
Page 21
Getting Started
11
Environmental Monitor
(EVM)
USB cable- connects to the
external output of the EVM
to your PC
DMS software
**Universal Power adapter
(External Power supply for
EVM)
Sample senor (Your
sensors will be calibrated
and placed inside the
instrument, depending on
what you ordered. The
sensor type is labeled on
the side of the cylinder.)
Example of a cal gas and
regulator
HEPA filter/Zero filter for
particulate calibrations
This chapter is designed to guide you through basic information you will need to begin measuring with
your environment monitor.
Checking the Equipment
If your instrument was sent to you in a storage case, remove all the packaging and acquaint yourself
with the following equipment.
NOTE: Depending on the accessories you selected, the EVM may have the components displayed in
Figure 2-1. For additional sensors or other accessories, refer to “Replacement parts.”
Figure 2-1: Identifying your equipment
EVM Series User Manual
Page 22
Display and Keypad
12
1. Display
2. Backlight
3. Particulate pump
4. Left arrow
5. Run ►/Stop ■
6. Indicators
7. Soft Keys
8. Up arrow
9. Altf key
10. Enter key
11. Right arrow
12. Down arrow
13. On/Off
and Esc
Keypad
Explanation
1. Display
Used to view data, menus, and various indicators.
2. Backlight
Used to illuminate the background of the display screen. (i.e., nighttime study.)
3. Particulate pump
Used for particulate measuring. Toggles the internal pump on and off (used for
particulate measurements)
4. Left arrow
Used to toggle through the views and to move left through menu selections.
5. Run/Stop
Used to Run or Stop log session.
6. Indicators
Status information is provided by the instrument at the top of the display.
7. Soft keys
The soft keys are only applicable when a second menu appears above the keys. For
example, in the Dust measurement screen, a lookup soft key, a trend menu, and a
detail menu appear directly above the soft keys on the display.
8. Up arrow
Used to select menu options (in the setup screens) and select measurement values
(or data).
Diagram
Display and Keypad
The keypad is used to run and stop your studies, view your measurement values, select appropriate setup parameters, and power on and off.
Diagram
The following diagram explains the features on the display and keypad.
Figure 2-2: EVM keypad and display explained
Keypad description
EVM Series User Manual
Page 23
Turning on
13
Keypad
Explanation
9. Altf key
The Altf key (or Alternate Functions) is used to switch between soft key menu and
selected menu. This is an option in various screens (ie, calibration screens).
10. Enter key
Used to open a selected menu. Toggles through selected measurement parameters
when viewing measurement screens.
11. Right arrow
Used to toggle through the views and to move right through menu selections.
12. Down arrow
Used to select menus options (in the setup screens) and select measurement values
(or data).
13. On/Off/Escape
Used to turn on the EVM, turn off, and to Escape (or move back one screen).
Indicators
Battery power:
Stop: ■
Start Screen
Press the left or right arrow key to toggle through
measurement views.
Menu options
A selected menu is indicated by the shaded text.
Screen descriptions
A note explaining how to access the measurement screens
by pressing the right or left arrows.
Start Screen Explanation
Table 2-1: Keypad/user interface explained
Turning on
1. To turn on the EVM, press the On/Off /Esc .
A TSI splash screen will appear including the firmware version name.
Proceeding the splash screen, you will be viewing the Start screen.
Start Screen Explanation
The main screen you encounter and is referenced throughout the manual is the Start screen. From the
start screen, you can access the measurement screens, the setup screen, calibration screens, file
system screen (used to view past sessions and clear data), and unit information screen (includes serial
number information). (The measurement screens are explained in more detail in Chapter 5.)
1. To select a menu option, press ▲/▼Arrows and then press Enter
selected menu screen.)
2. To view the measurement screens, press the Right or Left Arrow ◄/►and this will toggle you
through the measurement views.
. (This will open the
Figure 2-3: Start screen defined
EVM Series User Manual
Page 24
Turning on
14
Indicator Icon
Explanation
■
Stop – indicates the instrument is currently in the stop status (opposed to run).
►
Run - indicates the instrument is currently logging a session/study.
Battery charge status - This icon shows the renaming capacity of the batteries. A
full battery indicates the instrument is fully charged.
Pause – The pause indicator is shown during the 5 second countdown to prime the
airflow prior to entering the run mode when a log session is started.
OL
Overload –will appear whenever the level measurement exceeds the range of the
instrument.
Altf key (Alternate Function key)- this icon appears below the Battery Charge
Indicator when alternate functions are being displayed. (For example this is used in
the Trend screen to view the time history on the chart.
UR
UnderRange – will appear whenever a level measurement is below the range of the
instrument.
00:00:00
Run Time – displays the current elapsed log session run time. When elapsed time
is less than 100 hours, time is displayed as HR:MIN:SEC. When the elapsed time
has exceeded 100 hours, time is displayed as HR:MIN where HR uses 4 digits and
MIN uses 2 digits.
Navigating
Navigating
Once the EVM is powered–on, press the Up/Down Arrow▲/▼ key to select a
menu or to edit a field.
Press the Left/RightArrow◄/► key to scroll through measurement screen, start
screen, or to move right or left in a menu (i.e., setup menu).
Press Enter key to open a selected screen (i.e. Setup screen). It is also used
to save Setup parameters.
The On/Off/Esc key also functions as a navigational tool and is used to move back one screen (or
escape). It can be used if you are viewing a Setup screen, File System screen Calibration screen or Unit
Information screen. At any time, you can cancel a set-up change, file system change (etc…) by pressing
the On/Off/Esc key.
Indicators/Icons on the display
The EVM is equipped to notify you of operational/status updates. These indicators (or alert notices) are
located at the top of the screen. Table 2-2 explains the function of each indicator.
When turning off the EVM, ensure you are in stop mode. A stop indicator will appear on the top left next
to the battery charge status icon. If it is in run mode, stop the session and then proceed below.
1. From any measurement view or the start screen, press the On/Off Esc .
2. “3, 2, and 1” countdown will appear as the EVM is powered off. Continue to press and hold until the
instrument is powered off.
NOTE: To cancel the shutdown process, release the On/Off/Esc key and it will return you to
the screen you were previously viewing (i.e., PM measurement screen).
Auto-Run Note: If an auto-run session has been set, the auto-run summary screen is
displayed prior to turning off. See section, Auto-Run in Chapter 3.
Instrument Information
The general information screen also called the “unit info” screen is used occasionally to check the type
of instrument you purchased and displays the serial number and firmware versions.
Unit Information Screen
Opening the unit info screen
1. Turn on the EVM by pressing the On/Off/Esc.
2. In the start screen, press the down arrow repeatedly until Unit Info is selected and press
Enter .
The unit info screen will appear.
Figure 2-4: Unit Information screen
3. To view the installed sensors, repeatedly press Enter and it will display the Unit Info screen
with the installed sensors. (See Figure 2-6).
EVM Series User Manual
Page 26
Instrument Information
16
Look-up key
Impactor and profile displayed.
Pop-up box
Appears when you press the
Look-up key.
Sensors
Select and then press Enter to
view the sensor’s parameter
A
B
EVM-7
Sensors 1 - 3
Sensors 4 - 7
Unit Information Screen
4. Or press the Look-up key. Then press the ▲/▼Arrow to select and press Enter to open the
screen. (Repeat this step to view the other sensor.)
5. Unit Info screens are displayed below.
Screen “A” displays the particulates (PM), temperature (temp), and humidity as “Installed” or
Screen “B” displays CO
Figure 2-5: Selecting Unit Info. Sensors
“Enabled” sensors.
, CO (or installed toxic sensor), and PID with the associated serial
2
numbers. Air Velocity is listed as “Not Found” which indicates that the sensor was not
installed.
Figure 2-6: Viewing installed sensors
EVM Series User Manual
Page 27
TSI Detection Management Software DMS
17
Powering and Outputs on the EVM
TSI Detection Management Software DMS
DMS is a software application designed by TSI that runs under the Windows operating system. You can
use DMS for a variety of functions such as: customizing your set-up parameters, displaying data,
manipulating data, and creating reports.
Communicating
A USB cable (included in the kit) is used for communicating between the EVM and the software. (First
remove the black rubber casing over the ports and then insert the USB cord in the bottom connector.)
(See Chapter 6 for more details on software and the EVM.)
Powering and Outputs on the EVM
Battery
The battery is an internal Lithium-ion Rechargeable battery pack. The minimum hours of operating time
is 8 hours of continuous operation including setup time, run time, and viewing your result time. This
battery is not user accessible or replaceable.
The universal power adapter cord is a 10 to 16 volt power inlet (nominal 12 VDC) with 1.5 amps and is
used as an external power source. (For more details on battery power, see “Checking battery power” in
Chapter 3).
Figure 2-7:Communicating with DMS via USB cable
EVM Series User Manual
Page 28
Communicating
18
Power jack
Analog & Digital Outputs
USB port
Powering and Outputs on the EVM
Charging
The rechargeable battery pack is charged by plugging in the universal power adapter cord into the
charging-port and into an electrical outlet. When using the EVM and the power adapter, it will charge as
you use the instrument. The estimated time to recharge a fully discharged battery pack is approximately
five hours. It is important to understand that the USB connection/cable will not charge the internal
battery. USB is used for communication purposes only.
NOTE: When the batteries are depleted and if you are charging the instrument, the temperature
sensor will read at a higher level. Thus, it is recommended to measure the temperature
when you are not charging the instrument.
About the Environment
If the environment you are in is above 40º C (104º F) when charging the battery then you will be unable
to charge the battery. (See Appendix D for screen warnings.)
Outputs
The digital output and/or analog output can be used to connect to external devices such as a chart
recorder or a light notification device. The monitor’s digital out and analog out channel are accessible
through the Input/Output connector. The air velocity wand also uses this connector. (The following
diagram depicts the backside of the instrument without the back cover and the black rubber casing
removed.) For more information on connecting and using a stereo plug, see Appendix D.
Figure 2-8: USB, power jack and digital output port
EVM Series User Manual
Page 29
Setting up the Environmental Monitor
19
CHAPTER
3
Setting up Parameters via the Keypad
Setting up the Environmental Monitor
The focus of this chapter is to address all of the setup options available on the Environmental Monitor
(EVM). Once your parameters are selected, refer to the Calibration chapter (see Chapter 4).
NOTE: If you are using DMS, you may want to program all of your settings via the software. The
program allows you to save setups and apply them for future measurements/tests. Refer to
Chapter 6, for DMS details.
Setting up Parameters via the Keypad
The following user-adjustable setup screens are explained will appear for all instrument models:
Time and date screen
Battery power check screen
Logging screen includes selecting the logging interval, averaging time, and logged measurements
Display settings include: key clicks, language, and display contrast features
Setting up Input/Output (I/O) options including digital-out, air velocity, or analog-out
If you ordered the EVM-7 model, a Particulate setup menu and a PID setup menu will also appear.
Time and Date Setting
The EVM uses a 24-hour clock (military time) and specifies date by day, month, and year. The date and
time setting are important elements used as you are analyzing your log data results.
Changing time and date settings
1. Turn on the EVM by pressing On/Off/Esc
The Start screen will appear (see Figure 3-2 for details).
2. Ensure Setup is highlighted and press Enter
3. In the Time/Date screen, press the ▲/▼ Arrows to select Time field or Date field. (Note the Day
field will update automatically based on the selected day from the date field.)
NOTE: It will be highlighted when selected.
4. Press Enter
5. To change the field values, press ▲/▼Arrows. To move to the next field, press the right/ left
arrow key. Repeat this step as necessary.
To return to the first column field (Time, Date, or Day), repeatedly press Enter
On/Off/Esc
.
. (The Setup menu will appear.)
to move to the update fields (2nd column).
to cancel session.
or press
EVM Series User Manual
Page 30
Setting up Parameters via the Keypad
20
Time
Date
Day – auto fills based on what you entered in the
Date field
Checking Battery Power
6. Continue to repeat steps 3-5 until all changes have been made.
Figure 3-1: Time-Date screen parameters
7.Press the On/Off/Esc
Remember, once you return to the start screen, you can press the right or left arrows to toggle
through the measurement screens.
Checking Battery Power
The battery power is supplied by an internal lithium-ion rechargeable battery pack with a minimum of
8 hours of continuous operation.
NOTE: The battery pack is not-user replaceable and would need to be changed at TSI or your
authorized dealer.
The battery screen is used to determine the battery capacity status. When operating from the battery,
you can check the remaining capacity of the battery pack at any time by accessing the Battery screen.
Additionally, a battery indicator will be present on all screens to notify you of the remaining battery
capacity. If the battery indicator is at 100%, it will look similar to the Figure 3-3 below.
If you are charging the battery, the battery screen will display “battery is charging” with an updating
indicator bar. The battery power percentage will update accordingly.
twice to return to the start screen.
EVM Series User Manual
Page 31
Setting up Parameters via the Keypad
21
Start screen
Highlighting indicates “Setup” menu is selected
Measurement screens
Press right or left arrow to view
Battery indicator has all bars representing fully charged
battery
Indicates remaining battery capacity
Checking Battery Power
Check battery power
1. From the start screen, highlight Setup (by pressing ▲/▼Arrows) and press Enter
.
NOTE: If viewing a measurement screen, press the right/left arrow until the start screen
appears.
Figure 3-2: Start screen
2. Repeatedly press ▲/▼ Arrows until Battery is selected.
3. In the Battery setup screen, it will display the remaining capacity on the battery pack.
4. To return to the Start screen, press On/Off/Esc
twice.
Figure 3-3: Battery screen
EVM Series User Manual
Page 32
Setting up Parameters via the Keypad
22
Display settings
Explanation
Language
With the language setting, there are six options including: English, Spanish,
German, French, Italian, and Portuguese.
Contrast
With the contrast setting, you may want to adjust the shading of the display
(or screen contrast) by darkening or lightening the screen to fit the
environment you will be operating the instrument in. The contrast setting
ranges from 1-12 in which 12 represents the darkest and 1 represents the
lightest.
Key Click
With the key clicksetting, you can turn the sound to “on” and each time you
press a key on the keypad you will hear a soft click sound. This can be
disabled by choosing the “off” setting.
Display fields
Select the field (via up/down arrows). Press
Enter. To change 2nd column, press up/down
arrows.
Display: Language, Contrast, and Key Click
Display: Language, Contrast, and Key Click
In the display screen, there are three selectable settings including language, contrast, and key click.
Table 3-1: Display settings explained
Changing language, contrast, and key click settings
1. From the start menu, press ▲/▼ Arrows to select Setup and press Enter
2. Repeatedly press the ▼Arrow to select the Display menu and press Enter .
3. To select a field in the display screen, press ▲/▼Arrows to select one of the fields. Press Enter
to move to the 2
4. Follow the fields that apply:
To change the Language, press ▲/▼Arrows to select a specific language. Press Enter
to move back to left column or escape to cancel selection.
To change the Contrast, press ▲Arrow repeatedly or press ▼Arrow repeatedly to
decrease the contrast setting. Press Enter
cancel selection.
To change the Key click sound setting, press ▲/▼ Arrows until On/Off is displayed. Press
Enter
nd
column (changeable field).
to move back to the left column or escape to cancel selection.
.
to move back to left column or escape to
Figure 3-4: Screen display settings
EVM Series User Manual
Page 33
Setting up Parameters via the Keypad
23
Log intervals
Explanation
Log interval
This sets the log interval between specific data points. For example, if logging
is set to 5 second intervals, then all parameter values are logged every 5
seconds. The logging parameters options are as follows: 1 second,5 seconds,
15 seconds, 30 seconds, 1 minute, 5 minutes, 10 minutes, 15 minutes, and 30
minutes, and 60 minutes.
Averaging
The Averaging parameter is used to average all measurements displayed on
the screen. If averaging is set to 15 then the last 15 measurements are
averaged. (New measurements are updated every second.) This averaged
value is displayed on the monitor. It’s basically a data filter. The averaged data
will change less rapidly than the 1 second data since it is averaged. The user
options are 1 – 30 with increments by 1 interval.
NOTE: 15 second averaging is recommended for the average setting. All
the sensors have settling times of 10 seconds or longer.
Measurements
Selectable logged measurements: Level, Lmin, Lmax, Lavg, and STEL.
If the measurement is set to “On”, this indicates the instrument will log data at
the logged interval (i.e., while in run mode). “Off” indicates the measurement
will not log when in run mode.
The measurements include the following: Level (instantaneous or averaged
measurement), STEL (short term exposure level calculated over last 15
minutes), Lmax (maximum measurement during log interval), Lmin (minimum
measurement during log interval), Lavg (average measurement during log
interval), and TWA (time weighted average measurement during interval).
NOTE: See “Table 5-3” for more information.
Run Time
Depending on the logged parameters that are turned “On” for data log and the
log interval, the Estimated Run Time screen will display the total remaining
run time. For example, if you want to log all parameters the overall run time will
decrease significantly. If you would like to extend the run time to a longer
duration, logging one or two parameters would be advisable.
Setting Logging and Viewing Run Time
5. To return to the Start screen, repeatedly press On/Off/Esc .
Setting Logging and Viewing Run Time
With the logging screen, you select the logging interval (i.e., 1 min. data logging), the averaging
parameter (i.e., a type of data filter) and which parameters will be logged and stored into memory. An
optional field, Run Time, allows you to view the amount of remaining log time when logged parameters
are selected.
For logging, there are two selectable fields- the logging interval rate and the measurement averaging.
Table 3-2: Logged intervals and selecting logged measurements
EVM Series User Manual
Page 34
Setting up the Environmental Monitor
24
Logging time/averaging time
Logging attributes/measurements
Run Time
When selected, the estimated run time screen
displays the total remaining time.
Setting up Parameters via the Keypad
Set logging intervals, logging attributes, and view run time
1. From the start menu, press ▲/▼ Arrows to select Setup and press Enter . Select Logging
and press Enter
2. To change Logging Interval or Averaging follow below:
Press ▲/▼Arrows to select Logging interval or Averaging.
.
Press Enter
Press Enter
to move to the changeable field, and then press ▲/▼ Arrows.
to return to the 1st column and save your settings.
3. To enable/disable logging, follow below:
Select a measurement by pressing ▲/▼Arrows.
Once selected, press Enter
(this will move your cursor to the “On/Off” field.)
Press ▲/▼Arrows to toggle and select On or Off.
Press Enter
again to return to the measurement.
Repeat step 3 until all logging parameters are enabled/disabled.
Figure 3-5: Logging screen
EVM Series User Manual
Page 35
Setting up Parameters via the Keypad
25
Estimated run time
**NOTE: To increase the total duration, log
only one or two measurements
and/or change the log interval.
PID Sensor Setup
4. Optional: To view the Estimated Run Time, select this field (by pressing ▲/▼ Arrows) and press
Enter . The estimated run time screen will appear.
NOTE: To extend the logging time, turn a logged measurement(s) to “Off” and/or change the
5. Press On/Off/Esc repeatedly to return to the Start screen.
PID Sensor Setup
For best PID testing results, it is recommended to add a correction factor to the relevant VOC you intend
to measure. Refer to the PID correction factors table in Appendix C for more details.
Additionally, the PID setup screen is used to set the ppb zero cal to either enabled or disabled
(if applicable).
log interval. Then, re-check the estimated run time.
Figure 3-6: Estimated run time for logging screen
EVM Series User Manual
Page 36
Setting up Parameters via the Keypad
26
PPB Zero Cal
Press up/down arrows to select either
enabled/disabled. Press Enter to save.
PID Factor
Press up/down arrows on keypad to change value.
Press Enter to save.
Default soft key
Press to restore PID factor to default (“1”).
PID and correction factor
The applied correction factor “3.5” is
displayed in the Gas measurement screen.
PID and correction factor
Displayed in the Composite
measurement screen.
PID Sensor Setup
Setting PID correction factor and/or ppb zero cal
1. From the start menu, press ▲/▼ Arrows to select Setup and press Enter
2. Select PID by pressing ▲/▼ Arrows and press Enter
.
.
3. In the PID screen, select either ppb zero cal or the PID factor by pressing ▲/▼ Arrows.
4. Press Enter
Arrows to change the value. Press Enter
5. Press On/Off/Esc
to move to either the disabled/enabled field or the value field. Press ▲/▼
repeatedly to return to the start screen.
Figure 3-7: PID correction factor/ppb setup
to save your changes.
NOTE: When changing PID factor to a value excluding “1”, the PID correction factor will
appear as a subscript on the measurement screens. Two samples are displayed
below:
Figure 3-8: PID correction factor displayed on measurement screens
EVM Series User Manual
Page 37
Setting up Parameters via the Keypad
27
Units
Temperature
Particulate Measurements, PID, or Air Velocity.
Screen message
Press the Enter
key to save the changed
setting.
Changing Units (Temperature, Particulates, PID, and Air Velocity)
Changing Units (Temperature, Particulates, PID, and Air Velocity)
In the units screen, the temperature, particulates, and the air velocity values can be displayed in
different measurement units.
Temperature’s setting is either Celsius (ºC) or Fahrenheit (ºF) with ºC as the default.
Particulates setting is either in mg/m
meter).
PID PPM/PPB sensor (VOCs) setting is either PPM or mg/m
Air velocity setting is in meters per second or feet per minute.
NOTE: If you do not have an air velocity sensor connected, it will not be displayed in this screen.
Changing temperature, particulates and/or air velocity units
1. From the Start screen (see Figure 3-2 for details), select Setupand press Enter
2. In the Units screen, press ▲/▼Arrows to select the Temperature, PM (Particulates), or
AirVelocity field.
3. Once selected, press Enter
4. Press ▲/▼Arrows to select the appropriate unit.
5. Press Enter
to save and move back to the first column.
6. To return to the start screen, press On/Off/Esc
3
(milligrams per cubic meter) or µg/m3 (micrograms per cubic
3
. PPB or μg/m3.
to select the units field.
twice to return to the screen.
.
Setting up Particulate Settings
There are four parameters in the particulate screen which include:
Particulate field- enables and disables when measuring particulates. (If disabled, it will not enable
the pump during run mode.)
Profile field – used to select the active particulate profile. The active profile can be edited by
pressing the edit quick key.
Gravimetric Mass and Gravimetric volume fields – displays the accumulated theoretical
particulate mass and air volume which has passed through the gravimetric filter (since the last
gravimetric reset). These can be reset by pressing the reset quick key.
Figure 3-9: Units screen
EVM Series User Manual
Page 38
Setting up Parameters via the Keypad
28
Setting Profiles (Particulate Correction Factors)
Setting Profiles (Particulate Correction Factors)
The EVM supports eight different particulate matter (PM) profiles with customized name settings and
user correction factor settings. When the profile field is selected from the particulate menu, you are able
to edit this field by pressing the Edit soft key. Once opened, you can change the profile name and
correction factor setting. (NOTE: Each profile stores up to four calibration verifies.)
1. Profile Name: You can change profile name and assign up to eight characters.
2. Factors: A user will have different correction factors based on the type of dust you are measuring.
How Particulate Factors are Adjusted and Calculated
All particulate profiles are factors are calibrated to Arizona Road Dust (ARD) with its particle size and
composition. Other dust compositions can be measured accurately by adjusting a profile’s correction
factor.
NOTE: For best measurement results, it is recommended to enter in the dust correction factor prior
to your particulate study. (This may require a gravimetric sampling.)
To calculate a correction factor, follow the guidelines below:
1. Run the instrument in the applicable environment for a set period of time/volume (i.e., 6–8 hours).
Accumulators:
The gravimetric mass accumulator will display the amount of dust accumulated based on
the active profile.
The gravimetric volume accumulator will display the amount of volume passed through the
gravimetric filter.
2. After the run, check the gravimetric mass accumulator, compare to what was actually measured
from the lab, and enter in a correction factor.
For example, before measuring ensure the gravimetric accumulators are reset and the
gravimetric filter is replaced. Next, you measure for four hours and the gravimetric
accumulator is displayed as 10 mg. Then, you measure the dust in the gravimetric filter (or
cassettes) and determine the actual weight. If the actual weight is 20 mg, you would change
the correction factor from 1 to 2.
Dust Correction Factors (Profiles)
When setting up the particulate parameters, it is good to keep in mind that the user defined profiles (or
correction factors) are a submenu of the particulate menu. Because of this layout, the following section
explains how to change the profiles/correction factors and apply a new profile.
Setting dust correction factors (profiles)
1. From the Start screen, select the Setup menu by pressing ▲/▼Arrows and press Enter
2. Press ▲/▼Arrows repeatedly until Particulate is selected and press Enter
setup screen will appear.)
3. Ensure Particulate is set to Enabled. (If not, press ▲/▼Arrows to enable and press Enter
to return to the first column.)
. (The particulate
EVM Series User Manual
.
Page 39
Setting up Parameters via the Keypad
29
Profile
Select this field and then press Edit soft key to
modify the Profile screen.
Edit soft key
Visible when profile menu item is highlighted
Profile name
Optional: customize the profile name
Factor
How Particulate Factors are Adjusted and Calculated
4.Select the Profile field by pressing ▲/▼ Arrows and press the Edit soft key.
Figure 3-10: Edit profiles for dust calibration
5. Press the Edit soft key and the Edit Profile screen will open.
6. (Optional, see step seven if not changing profile name.) To change the name of the profile, ensure
the profile field is selected and press the Enter . Press ▲/▼ Arrows to change the characters.
To move to the next position, press ◄/► Arrows then press Enter .
7. Press ▼ Arrow to select the Factor and press Enter
Arrows to change the numeric value. Press the left/right arrow key to move to the next position.
Continue this process until the appropriate number is selected and press Enter to confirm or
Esc key to cancel changes.
8. Press the On/Off/Esc
9. To apply a new profile, select the profile field and move to the second column (by pressing
Enter .) Then continuously press ▲/▼ Arrows until the appropriate profile/correction factor is
selected. Press Enter
10. Press On/Off/Esc twice to return to the start screen.
Figure 3-11: Entering a profile for particulate calibration
to move to the 2nd column. Press ▲/▼
to return to the particulate screen.
to return to the first column.
EVM Series User Manual
Page 40
Setting up Parameters via the Keypad
30
Reset soft key
Visible when grav. mass
menu item is highlighted
Displays accumulated
particulate mass & volume
Particulate Pump: Enable or Disable
Particulate Pump: Enable or Disable
You may want to disable the particulate pump for situations in which the environment is too dirty for the
particulate optical engine or if you want to keep the engine and pump clean and dry from oily mists,
liquids, or fogs elements. Additionally, you may want to log other parameters, such as CO2, Toxic,
and/or temperature and humidity parameters and exclude the particulate measurements from a log
session. This is changeable in the particulate screen.
Enable/Disable the pump
1. From the Start screen, select the Setup menu by pressing ▲/▼Arrows and press Enter
2. Press ▲/▼Arrows repeatedly until Particulate is selected and press Enter
setup screen will appear.)
3. To enable/disable the particulate setting/pump, ensure Particulate is selected and press
Enter .
. (The particulate
.
4. Press ▲/▼Arrows to select either Enabled or Disabled and press Enter
to view a sample screen.)
5. Press On/Off/Esc twice to return to the start screen.
Resetting the Gravimetric Filter
The Gravimetric accumulators display the calculated accumulated mass of particles and the
accumulated volume through the gravimetric filter (since the last time these accumulators were reset).
At the same time, the particles are actually collected in the gravimetric cassette in the back of the unit.
Typically the gravimetric accumulators are reset when the gravimetric cassette is replaced by the user.
Resetting gravimetric filter
1. From the Start screen (see Figure 3-2 for details), select Setup by using ▲/▼ Arrows. Press
Enter .
2. Press ▲/▼Arrows to select the Particulate screen and press Enter
will appear.)
3. To reset the Gravimetric filter, select the Grav. Mass/Grav. Volume fields by pressing ▲/▼
Arrows. Once selected, the Reset soft key will appear. Press this soft key to reset.
. (Refer to Figure 3-7
. (The particulate screen
Figure 3-12: Particulate setup screen
EVM Series User Manual
Page 41
Setting up Parameters via the Keypad
31
Digital Out/Logic
Explanation
Active High Pulse
The output will assert an active high pulse for approximately 50 msec
when the threshold is exceeded.
Active Low Pulse
The output will assert an active low pulse for approximately 50 msec when
the threshold is exceeded.
Active High
The output will be at a logical high level when the threshold is exceeded.
Active Low
The output will be at a logical low level when the threshold is exceeded.
Threshold
Enter in a value to trigger the external device to turn on or off.
Auxiliary Input/Output
4.To return to the main menu, press On/Off/Esc twice.
Auxiliary Input/Output
Digital Output
With the digital output function, the user can connect an external device to the monitor. This can be
used to signal a specific level to turn on a notification device, an LED, or a fan motor at a specific
threshold (or set point). Once enabled, the following three are selectable parameters: a sensor, logic
setting, and threshold setting.
Sensors
When setting the Digital Output (“Digital-Out”) parameters, one of the seven selectable sensors is used
to trigger the output of the external device which includes:
RH (relative humidity), CO
(for measuring VOCs), Toxic gas (i.e, CO), and Dew Point.
NOTE: Air Velocity cannot be selected as a sensor.
Logic and Threshold
There are four selectable “Logic” types which is used in combination with the Threshold field to trigger
on and off the external device. These are explained in the table below.
NOTE: The digital output is an open collector output and as such requires the user to install a pull-
NOTE: The system maintains a hysteresis window around the threshold to prevent output chattering.
Setting or changing digital output
1. From the Start screen, select Setup menu and press Enter
2. Press ▼Arrow repeatedly until Aux I/O and press Enter
3. Select Digital Out and press Enter
.
.
.
EVM Series User Manual
Page 42
Setting up Parameters via the Keypad
32
Digital Output parameters
Auxiliary Input/Output
4. In the Digital-Out screen do one of the following:
Dig-Out field: select via ▲/▼ Arrows and press Enter
Enable/Disable.
Sensor field: select via ▲/▼ Arrows and press Enter
one of the sensors.
Logic field: select via ▲/▼ Arrows and press Enter
of the triggering modes.
Threshold field: select via ▲/▼ Arrows and press Enter
a value.
. Press ▲/▼Arrows to
. Press ▲/▼ Arrows to select
. Press ▲/▼ Arrows to select one
. Press ▲/▼ Arrows to select
5. To return to the Start screen, press the On/Off/Esc
Figure 3-13: Configuring digital output
twice.
EVM Series User Manual
Page 43
Setting up Parameters via the Keypad
33
Air Velocity – disable/enable
0 1 2 3 4 5 6 7 8 9 Hours
0 Volts
5
60
ºC 40
20
0
Air Velocity
Air Velocity
Setting the air velocity parameter is fairly simplistic; however, ensure no other outputs are enabled prior
to setup. (The EVM will display a pop-up message if more than one output is enabled. You will have to
disable all other auxiliary outputs before proceeding with air velocity setting.) When air velocity is
“enabled”, you can measure/log with the TSI Air Probe10 (optional accessory).
Setting air velocity
1. From the setup screen, select Aux I/O and press Enter .
2. Press ▲/▼Arrows to select Air Velocity and press Enter . To change the setting from either
enabled or disabled, press Enter .
3. Press ▲/▼Arrows to change the setting.
NOTE: If you have either the digital-out or analog-out setting enabled, a pop-up message will
4. Press On/Off/Esc twice to return to the start screen.
Analog-out
The analog-out channel provides a voltage reading that is proportional to the sensors output. The
analog-out channel can be configured to track any installed sensor. The example displayed in
Figure 3-12 shows the results of measuring with the thermometer from 0 to 60ºC reading from 0 to
5 volts.
Figure 3-14: Air Velocity setting
appear explaining to change the other settings before enabling air velocity.
Figure 3-15: Analog out results with temperature as output
EVM Series User Manual
Page 44
Setting up Parameters via the Keypad
34
Disable/Enable Analog-out field
Select a Sensor
Restoring Default settings
Select either (or both) to restore all
factory cals/settings.
Restoring Default Factory Settings
Setting Analog-Out
1. From the setup screen, select Aux I/O and press Enter .
2. Press ▲/▼Arrows to select Analog-Out and press Enter .
3. To enable or disable, press Enter
change the setting. Press Enter
4. To select a sensor, ensure Sensor is selected and press Enter . Then, press ▲/▼Arrows to
page through the selection list. Once the appropriate sensor is selected, press Enter .
5. To return to the start screen, press On/Off/Esc twice.
Figure 3-16: Analog-Out screen
Restoring Default Factory Settings
The EVM is equipped to restore all factory settings and factory calibration settings via the Defaults
screen.
NOTE: Dew Point and Air Velocity are not user calibrated and therefore, not listed in the restore
factory calibration screen.
1. In the Setup menu, press ▲/▼Arrows until Defaults is selected (or highlighted).
2. Press Enter
(to open the screen).
when Analog-out is selected and press ▲/▼Arrows to
Select a sensor, Choose Yes or No with
the screen prompt and press Enter key.
Restoring Default Factory Settings
3. To restore the Factory Calibration, ensure this is selected and press Enter
.
To restore defaults for a selected sensor(s), select a sensor using ▲/▼ Arrows and press
Enter . In the pop-up screen, select “Yes” via the Left Arrow (or No to cancel) and press
Enter .
Repeat as necessary to restore each selected sensor.
To restore all sensors, select All Sensors (via up/down arrows) and select Yes in the pop-up
screen (or No to cancel). Press Enter .
Figure 3-18: Restoring default calibrations
4. To restore the Factory Settings, press ▲/▼Arrows to select and press Enter
.
A Restore Factory Settings pop-up screen will appear. Press either Yes or No (with ◄/►
Arrows and Enter
to confirm).
EVM Series User Manual
Page 46
Setting up Parameters via the Keypad
36
Backlight key
Press to turn on. This feature has three levels
(off, dim, bright).
Backlight Setting
Backlight Setting
A key on the keypad is dedicated to turning on or off the backlight feature. When turned on, the
backlight will illuminate the screen from a dark gray to a light gray color. This feature has two levels of
brightness.
When turning on, click the backlight key once for a lighter backlight. Press a second time for a darker
contrast.
To turn off, repeatedly press the backlight key until it is turned off.
Resetting the EVM
At times, you may need to reset the instrument if it is unresponsive.
1. To manually reset, press and hold On/Off/Esc for about 8–10 seconds and the instrument will
power off.
Lock/Secure Run and/or Setup
The lock or security feature is designed to lock the EVM during a run to ensure the session will not be
stopped unless the lock feature is disabled. Likewise, this feature includes a locked (or secure) setup
mode which will stop you from editing a set-up. Once one or both of these options are programmed with
a code, it is set by pressing the Lock soft key in the Setup menu.
NOTE: When securing or locking the instrument, it will be in lock mode even when powered off/on.
Figure 3-19: Backlight key
EVM Series User Manual
Page 47
Setting up Parameters via the Keypad
37
Lock
Opens to the Lock screen.
Adding a numeric password
Press Up/Down arrows to enter a number.
Press Left/Right arrow keys to move to the next
field.
Repeat as necessary.
Lock/Secure Run and/or Setup
Setting lock setup or run
1. From the Start screen, select and open the Setup screen.
2. Press the Lock soft key.
Figure 3-20: Lock Soft key
3. In the Lock/Security Modes screen, Run and Setup are “disabled” as the default setting.
4. To enable either one, first select Run or Setup fields (by using Up/DownArrows).
5. Press Enter
and the first field will be shaded/selected, and then press ▲/▼Arrows to
change the numeric value.
6. Repeat step 5 until the password is entered. When you press Enter for the fourth digit, a
secure run mode or secure setup pop-up will appear as confirmation. Once set, the field will display
“Enabled”.
7. Once enabled, follow one of the scenarios below:
For Secure run mode to occur, navigate to one of the measurement screens and press the
Run key. A “Secure Run Mode” screen will appear.
For Securesetup mode to occur, navigate to the setup screen and the “Secure setup mode”
screen will appear.
Figure 3-21: Setting lock/secure
EVM Series User Manual
Page 48
Setting up Parameters via the Keypad
38
Auto-Run and Timed Run setting
Select Timed-run by pressing Enter and then
press Up/Down arrows to move through the menu.
Timed Run soft key
Select first soft key to access the Timed run screen
Unlock run and/or setup
Unlock run and/or setup
When in secure run mode or secure setup mode, a quick press of the lock/unlock soft key and entering
in the password will unlock the instrument. (See steps 5 and 6 above to for more information on entering
in the password.)
NOTE: If the entered password to unlock the instrument was forgotten, a default TSI password can
be used in lieu of the entered password, “9157”.
Auto-run
The EVM has four programmable auto-run settings including a timed-run, a date, a day of week (DOW),
and a disabled mode.
The EVM is shipped with the auto-run set to the disabled mode. Each mode has more than one setting
and is denoted by the “+” (plus) or “–”(minus) soft key symbols. When an alarm is activated (or
programmed), a “+” (plus) sign appears. The default alarm mode is set to an off mode and is denoted
via a minus symbol (i.e., “–”).
Auto-run and timed-run mode
With timed-run, you manually start the run (via the run key), and it will automatically stop based on the
programmed duration.
Setting auto-run with timed-run mode
1. From the Start screen, select and open the Setup screen.
2. Select Auto-Run by pressing ▲/▼Arrowsand press Enter
3. In the Auto-Run screen, press enter to select Timed-Run. If it is not displayed, press ▲/▼
Arrows repeatedly until selected.
4. Next, press the -ONE soft key or select View/Set Parameters.
.
Figure 3-22: Selecting auto-run with timed-run mode
EVM Series User Manual
Page 49
Setting up Parameters via the Keypad
39
Duration
Select the hours/minutes/seconds with up/down
arrows and press Enter to move to the next field.
Soft keys
Select a soft key to setup the Date Auto-Run
parameters
Once set, “-one” will appear as “+one” which is
turned on.
Auto-run
5. In the timed-run screen, press Enter
to select the duration field which is displayed in hours,
minutes, and seconds. To change the duration, press ▲/▼Arrows. To move between the fields,
press ◄/►Arrows. (Repeat as necessary).
Figure 3-23: Setting timed-run and duration
6. Press Enter
to save your settings.
NOTE: To begin the Timed-run, press the run/stop key and the instrument will automatically
stop at the programmed duration.
Auto-Run with Date Mode
With the date mode, it will wake up the instrument from a powered down or on condition, runs the study,
saves it, then powers down again. Up to four date settings are programmable with a specified date, time
(in hours, minutes, seconds), and duration via the soft keys.
Setting auto-run with date mode
1. From the Start screen, select and open the Setup screen.
2. Select Auto-Run by pressing ▲/▼Arrows and press Enter
3. Press Enter
to select the 2nd column (value field) and press ▲/▼Arrows repeatedly until
Date appears.
4. Next, press one of the soft keys. (This will open a Date/Auto-Run screen.)
.
Figure 3-24: Selecting auto-run with date mode
EVM Series User Manual
Page 50
Setting up Parameters via the Keypad
40
Start time
Select time of day
(24 hour clock) by
pressing up/down
arrows
Corresponds to soft key
Auto-Run with Date
Select day, month, year
Duration
Press up/down arrows to
select hours, minutes,
seconds
Auto-run
5. If off appears under Auto-Run, press Enter
6. To change the fields press ▲/▼Arrows. To move to the next field(s) press ◄/►Arrows.
(or skip and see step 6).
Repeat as necessary.
Figure 3-25: Setting auto-run date mode
7. To save, press Enter
key. Press key until you return to the start screen.
NOTE: To initiate the programmed run, either power down the instrument or leave on.
Auto-Run with Day of Week (DOW) Mode
The days of the week (DOW) setting is a pre-programmed weekly auto-run with a specified day, start
time, and duration with two auto on settings. The programmable settings include a day of the week, start
time, and duration.
Setting auto-run with day of week mode
1. From the Start screen, select and open the Setup screen.
2. Select Auto-Run by pressing ▲/▼Arrows and press Enter
3. In the Auto-Run screen, press Enter and press ▲/▼ Arrows repeatedly until DOW appears.
4. Press one of the soft keys (one or two).
5. In the DOW Auto-Run screen, press ◄/► Arrows to move through the Days, Time and Duration
fields. Press ▲/▼Arrows to change the value.
.
EVM Series User Manual
Page 51
Setting up Parameters via the Keypad
41
Days
Optional Programming is set from Sunday through
Saturday. Example displays Monday through Friday
S (hours:mins:secs)
Start time is programmed at 8 am
D (hours:mins:secs)
Duration is programmed for 4 hours
Auto-run
6.When programmed, press Enter
Figure 3-26: Setting auto-run DOW mode
. Press key repeatedly to return to the main screen.
NOTE: To initiate the programmed run, either power down the instrument or leave on.
EVM Series User Manual
Page 52
Setting up Parameters via the Keypad
42
Auto-run
(This page intentionally left blank)
EVM Series User Manual
Page 53
Calibrate Sensors
43
Manifold
Sensor housing
CHAPTER
4
Placement and Calibration of Sensors
Calibrate Sensors
Placement and Calibration of Sensors
Before you begin your measurement sampling, it is recommended to calibrate the sensors you will be
using. This chapter focuses on inserting and removing sensors, calibrating each sensor, viewing past
calibrations, and how to verify your calibration set points. (Reference the following sections which apply
to your installed sensor components.)
Calibrating sensors include:
Particulate sensor
Pump flow rate calibration
CO
Toxic sensors
Photo-Ionization Detector (PID) sensor (for VOCs)
Temperature (temp) and Relative humidity (RH) sensors
Inserting and Removing Sensors
1. The sensors you purchased with your EVM will be inserted, factory calibrated, and ready for usage.
Inserting sensors
2. Ensure the instrument is turned off and the external power cord is disconnected. Next, remove the
3. Lift the cover off and set aside. (The location of the sensors is indicated in Figure 4-2.)
4. Remove the black manifold casing.
and O2 sensor
2
To understand how to insert or replace a sensor this is discussed in the procedures below.
NOTE: To order new sensors when expired, see Appendix A, Replacement and optional parts.)
sensor bar by unscrewing two screws from the top compartment with a screw driver.
Figure 4-1: Sensor housing and manifold
EVM Series User Manual
Page 54
Placement and Calibration of Sensors
44
Air inlet cover
Screw 1
Dial-in Impactors
Toxic sensor
Sensor bar- Remove screw 1 & screw 2
and lift off cover.
Temperature/
Relative Humidity sensor
(Dry bulb thermometer
measures ambient air
temperature)
Screw 2
PID sensor
CO2Sensor
Calibrating Particulates
5. To insert a sensor, gently push the sensor’s pins into the socket pin of the circuit board so it aligns
accordingly. It will fit securely.
6. To remove a sensor, pull the senor up horizontally and remove it from the sensor chamber.
7. When completed, replace the black manifold over the sensors and then place the cover over the
sensor housing. Tighten the screws back into place.
Calibrating Particulates
NOTE: The Temperature/ Relative Humidity sensor is not user removable.
NOTE: The user will receive a message on the display when a new sensor is detected or an
The EVM is factory calibrated to the “Arizona Road Dust” (also called “ISO Road dust” ISO 12103-1 A2
Fine) standard.
Before you measure particulates, it is recommended to calibrate to the environment you are in. If you
are in an ARD environment, you would adjust the impactor to correlate with a specific particulate setting,
run a calibration based on the particulates in that environment, and compare the results against the
ARD calibration. If you are in any other environment, you would calibrate based on the environment you
are in by determining a correction factor (also called “K-factor”) to the correct variation from the ARD
calibration standard.
Figure 4-2: Sensors’ locations
existing sensor is no longer detected.
EVM Series User Manual
Page 55
Placement and Calibration of Sensors
45
2. Pump – circulates the air
through the tubing
1. Filters through
impactor
3. Passes through
optical engine
4. Collects in Grav.
filter/cassettes
Impactor and use
The EVM supports eight different user profiles. Each profile stores a customizable profile name field,
and a correction factor. (Setting-up profiles/correction factors is discussed in the Set-up chapter,
“Setting profiles for particulates”).
The following sections will walk you through how the impactor operates, how to adjust the impactor, how
to run a zero calibration, and lastly running a gravimetric calibration.
Impactor and use
The impactor is a device used to remove larger sized particulates from the air-stream you are
measuring.
How does it operate? Air travels through the impactor and the heavy particulates will stick to the
greased plates within the impactor (1). The Pump will circulate the air as it passes through the tubing in
the back of the instrument (2). The smaller, lighter particulates will pass through the optical engine (3)
and collects in the gravimetric filter/cassettes (4).
Figure 4-3: How the impactor functions
NOTE: Depending on the environment (i.e., a dense, dusty manufacturing plant), you will want to
ensure the impactor is well greased and cleaned before using. Additionally, it is
recommended to monitor this process while measuring. You may need to stop your
measurements, clean and grease the impactors, and then resume measuring. To clean and
grease the impactors, see Chapter 7“Cleaning and maintenance”.
EVM Series User Manual
Page 56
Placement and Calibration of Sensors
46
PM 2.5 & profile displayed
Impactor setting updates as
the turret is rotated
(when in stop mode)
Adjusting the Impactor/Turret Setting
Adjusting the Impactor/Turret Setting
When measuring, the industrial processes (such as sanding, cutting, crushing, grinding, and
transportation) cause air pollutants. The size of the particles will enter the human body in different
areas, affecting the nasal, throat, and/or lung area.
The impactor setting is used to trap and filter out the larger size particles based on your cut-point (or
impactor setting). The range is programmed between 1 μm to over 100 μm. For example, if you are
measuring coal dust (aimed at the lung area), you would adjust your impactor setting to PM10. The
following are the user-selectable cut-points:
PM2.5, PM4, PM10, or PM (all dust ranging from 0 μm-100 μm)
NOTE: The particulate size is written on top of the turret. For PM there is no marking but is noted by
the (raised) air chamber.
Adjust impactor/turret for particulate logging
1. Using the keypad navigate to the particulate or composite measurement screen.
How to? From the Start screen, press the Right arrow key.
Figure 4-4: PM2.5 impactor setting example
EVM Series User Manual
Page 57
Placement and Calibration of Sensors
47
Air inlet hole for PM2.5
Air inlet hole for PM4
PM2.5 position
PM4 position
Air inlet hole for PM10
Air inlet hole for PM
PM10 position
PM position (no text marking)
Position used for zero calibration
Particulates Calibration (Zero Cal)
2. Twist the turret (red casing around the impactor) to the left or right until you see the appropriate
setting on the display. It will align to the impactor text and arrow on top of the housing.
NOTE: The actual diameter of the air inlet hole is directly to the left of the impactor text positioning.
Figure 4-5: Adjusting impactor/turret setting
3. If you see “< >PM ERR”, this indicates the impactor is not in the correct position and that it is
between two positions. Adjust the turret, until the correct reading is displayed on the screen.
Particulates Calibration (Zero Cal)
A zero calibration is recommended the first time you measure particulates
in the run mode. For particulate calibration, place a zero filter (also called a
HEPA filter) with attached coupler into the vent of the turret. The impactor
setting must be set at PM in order to conduct a zero cal.
NOTE: For the span on the particulate calibration, refer to the
“Gravimetric Filtering” instructions to determine the gain factor
from the correction factor.
Particulates zero cal
1. Insert the attached coupler and zero filter into the turret. Ensure the impactor is well-greased
before calibrating which requires a thin film of grease. (See “Impactor care” for details).
EVM Series User Manual
Page 58
Placement and Calibration of Sensors
48
Zero/HEPA filter
Used for particulate zero cal.
Zero filter with coupler and vent opening
vent opening.
Turret
the EVM. The raised section will align with the
“impactor” text.
Calibration – select to open to
the Calibration set screen
Particulates Calibration (Zero Cal)
Figure 4-6: Particulate zero cal with HEPA filter
2. Adjust the turret/impactor to PM (refer to “Turret” in Figure 4-6).
3. From the Start screen, press ▲/▼Arrows until Calibration is selected and press Enter
.
4. In the Calibration screen, press ▲/▼ Arrows and select PM (for particulate matter).
NOTE: If the impactor setting is incorrect, the instrument will flash a warning sign indicating
Figure 4-7: Selecting calibration screen
that you need to turn the impactor to “PM” before proceeding.
EVM Series User Manual
Page 59
Placement and Calibration of Sensors
49
PM
History soft key
Calibration soft key
Verify soft key
Start soft key
Particulates Calibration (Zero Cal)
5. Press the Cal soft key. (The pump will turn on.)
NOTE: The HISTR and VERIFY soft keys will not be present if a user calibration and a
calibration verify have not been performed.
Figure 4-8: Particulate cal screen
6.The ready screen with the PM, Profile will appear. Press the Start soft key.
Figure 4-9: Particulate calibration and ready screen
EVM Series User Manual
Page 60
Placement and Calibration of Sensors
50
A
B
Zero will appear
when it has
stabilized
Current Level
reading
Set soft key will
appear when
stabilized
Save soft key
Particulates Calibration (Zero Cal)
7. Allow the level to stabilize (approximately 1–3 minutes). It will state Waiting (A) and Set soft key
will appear when it has stabilized (B).
Figure 4-10: Particulate calibration and waiting screen
8. Press the Set soft key to set the Zero cal point.
9. The instrument will show a Pause state with the intended and actual values.
Figure 4-11: Particulate calibration pause screen
EVM Series User Manual
Page 61
Placement and Calibration of Sensors
51
Retry
Actual calibration value
Exit
Particulate Advanced Calibration (Gravimetric)
10. Press the Save soft key to save your calibration (or press Cancel and Retry to repeat this process.)
A screen pop-up will state “Zero cal set point” when completed (or “User canceled calibration”). A
pass (cancel) screen will be displayed.
Figure 4-12: Zero Cal set for Particulates
Press Retry soft key to repeat the zero cal procedures.
11. Press Exit soft key to close this screen. To return to the start screen, press On/Off/Esc or
press the Cal soft key to continue calibrating the sensors.
NOTE: A zero cal is performed for the instrument and will be used for all profiles.
Particulate Advanced Calibration (Gravimetric)
With the EVM, you can store up to eight different profiles (also called correction factors) to reflect the
environment or building zone you plan to measure.
The instrument is factory calibrated to Arizona Road dust. In some situations, you may want to perform
a gravimetric calibration in order to determine the correction factor for an alternative environment. This
is an optional procedure.
NOTE: If you do not plan on performing an advanced particulate calibration with gravimetric
sampling.
With a gravimetric calibration, you determine the mass of the particulate you want to measure. The
particulate mass is then used to calculate a correction factor to the environment you will be measuring.
The steps below outline a gravimetric calibration.
(For additional specifications on a gravimetric calibration, please refer to the NIOSH 0600 standard.)
EVM Series User Manual
Page 62
Placement and Calibration of Sensors
52
Reset soft key
Reset Gravimetric fields
to “0”
Gravimetric Calibration Overview
1. Reset the gravimetric and volume accumulators.
2. Insert a new (sealed) gravimetric cassette.
3. Measure in the specific location for approximately four to eight hours.
4. Proceeding the measurement, the gravimetric accumulator on the EVM will display
the amount of dust that has accumulated based on the active profile. (This value is
used in conjunction with the weight of the measured dust in the cassette.)
5. To determine the weight of particulates in the gravimetric filter, remove the
gravimetric cassette and weigh the contents. (Typically, sent to a lab for results.) Do
not include the weight of the cassette.
6. To determine the correction factor, calculate the ratio between the particulate mass
value and the weight from the gravimetric filter lab analysis.
For example, let’s say that you measured for 8 hours in a wood shop and the
PM mass accumulator displays 10 mg. Next, you measured the dust in the
gravimetric filter/cassette and determined the weight as 20 mg. To calculate the
correction factor, you divided 20 mg by 10 mg and determined the value as 2.
Lastly, the correction factor of “2” is multiplied by the existing active profile
correction factor, and is entered as the new correction factor (Setup\PM\EDIT
Performing a Gravimetric Calibration
Performing a Gravimetric Calibration
Ensure the gravimetric cassette is new before beginning.
1. Place the instrument in the appropriate environment.
2. Turn on by pressing On/Off/Esc.
3. In the Start screen, select the Setup menu and press Enter
pressing the Down arrow repeatedly and once selected, press Enter .
4. In the particulate screen, the reset gravimetric accumulator fields will display either a value of the
collected dust or it will display zero (if already reset). When beginning a gravimetric calibration,
ensure this is set to zero.
Figure 4-13: Reset gravimetric setting
. Next, select Particulates by
EVM Series User Manual
Page 63
Placement and Calibration of Sensors
53
Air Flow Outlet
Used to connect a primary standard calibrator with
tubing
Pump Flow Rate Calibration
To reset, press the down arrow repeatedly to select the gravimetric fields and press the Reset
soft key. (If it is not displayed, press the Altf key.)
5. Turn the pump on by pressing the Pump
6. Optional step:
a. It is optional to press the Run/Stop key to log the gravimetric filter sampling.
NOTE: If in “run mode”, a clock appears and displays the overall run time.
b. To view the particulate level, press the right arrow once from the start screen and the
particulate measurement screen will appear (noted as “PM”).
7. After the gravimetric collection, turn the pump off and reference the Gravimetric Mass field to view
the calculated particulates.
8. Lastly, calculate the ratio between the value in the Gravimetric Mass field and the weight of the
collected sample.
9. Enter in the correction factor via the Particulate Setup screen (see “Setting Profiles” section for
more details).
Pump Flow Rate Calibration
An accurate flow meter, such as a primary standard flowmeter calibrator, and inert tubing is required to
perform this calibration. The inert tubing is inserted on the dust air flow outlet (located on the back of the
housing).
NOTE: You may want to send the instrument in for a yearly calibration, if you do not own a
flowmeter calibrator for this calibration procedure.
key.
Flow rate calibration
1. Connect the flow meter (i.e, primary standard) and inert tubing to the EVM.
2. Turn the EVM on by pressing On/Off/Esc.
3. Select the Calibration menu by pressing the Down Arrow until it is selected and press Enter
4. Select Flow Rt (flow rate) by repeatedly pressing ▲/▼Arrows key until it appears.
Figure 4-14: Pump flow rate calibration
EVM Series User Manual
.
Page 64
Placement and Calibration of Sensors
54
1.67 L/min (lpm) is the
recommended
flow rate to mirror with
Percentage of air flowing
through the pump
Save
Toxic Gas Sensors
5.Press the CAL soft key and then press the Start soft key.
Figure 4-15: Flow Rate Calibration
6. Manually adjust the flow rate (as needed) by repeatedly pressing the Up/Down key to select the
desired level.
7. Press the Save soft key to store this as your flow rate calibration. The flow rate summary will
appear.
NOTE: You may want to repeat these steps if the desired level was not achieved. Press the
Cancel soft key to stop the calibration. Next, press the Retry soft key and repeat the
calibration steps to re-cal the flow rate.
8. Press the Exit soft key to exit this screen. (The pump will turn off.)
Toxic Gas Sensors
9. Press the On/Off/Esc
key to return to the start screen.
The gas sensor chamber will allow you to measure any one of the single toxic gases listed below. Refer
to Appendix A for detailed specifications.
1. Carbon monoxide (CO) with a range of 0–1000 ppm
2. Chlorine (CL
) with a range of 0–20 ppm
2
3. Ethylene oxide (EtO) with a range of 0–20 ppm
4. Hydrogen cyanide (HCN) with a range of 0–50 ppm
5. Hydrogen sulfide (HS
6. Nitrogen dioxide (NO
) with a range of 0–500 ppm
2
) with a range of 0–50 ppm
2
7. Nitric oxide (NO) with a range of 0–100 ppm
8. Oxygen (0
9. Sulphur dioxide (SO
10. Ozone (O
) with a range of 0–30%
2
with a range of 0–50 ppm
2)
) with a range of 0–1.0 ppm
3
The sensors are based on well-established electrochemical sensor technology. They are designed to be
maintenance free and stable for long periods of use.
EVM Series User Manual
Page 65
Placement and Calibration of Sensors
55
Gas
Interfering Gas
CO
H2S
SO2
NO
NO2
H2
CL2
HCN
C2H4
HCI
CL
O2
CO
100
~ 7
<10
<9
<20
<40
N/D
N/D
N/D
N/D
N/D
N/D
H2S
<2
100
~10
~1
~-20
~1
N/D
N/D
N/D
N/D
N/D
N/D
NO2
-5
<5
<-.5 0 100
N/D
N/D
N/D
N/D
N/D
N/D
N/D
NO 0 N/D
~ 5
100
<30
N/D
N/D
N/D
N/D
N/D
N/D
N/D
HCN
N/D
N/D
N/D
N/D
N/D
N/D
N/D
N/D
N/D
N/D
N/D
N/D
NH3 0 0
~ 60
~20
<10 0 ~-50
~-5
N/D
<10
N/D
N/D
*O3
-8
N/D
N/D
N/D
~60 0 ~120
N/D
N/D
~150
~150
~150
Cross Sensitivity
Cross Sensitivity
The following table illustrates the cross-sensitivity to a range of commonly encountered gases,
expressed as a reading of the sensor when exposed to 100 ppm of the interfering gas at 20ºC.
*Continuous exposures at ppm level over more than 30 minute might blind the sensor.
How Sensors Operate
Power is automatically applied to the gas sensors when they are connected and the instrument is turned
on. These are also referred to as “intelligent-sensors”. Additionally, the EVM will automatically identify
the type of gas sensor that is installed.
NOTE: It may take a few minutes for a newly inserted sensor to stabilize before using.
Table 4-1: Cross-sensitivity table of interfering gas
Quality Calibrations
The quality of the calibration process depends upon the accuracy of the calibration gas and allowing the
sensors to stabilize before saving the zero and span calibrations.
About Sensors and Cal Storage
Once a calibration is completed and a sensor is removed, the calibration records are stored in the
sensor.
Calibration History and Sensor Serial Numbers
After running, saving a calibration, and downloading the data to Detection Management Software
(DMS), the calibration type, levels, the serial numbers, and the date/time of the calibration are viewable
in the software. The serial number of the sensor is available in DMS in the calibration history chart.
Figure 4-16: Calibration history for EVM
EVM Series User Manual
Page 66
Placement and Calibration of Sensors
56
1. Connect the gas
source, regulator and
inert tubing.
2. Slide the inert tubing
over the black nozzle
of the cal cup.
3. Remove the air inlet
coverr on the EVM.
(See Figure 4-2 for an
example).
4. Place the Cal cup over
the air inlet.
5. When you are ready to
access the Cal screen,
turn on the regulator.
EVM
Tubing
Regulator
Calibration gas
source
Cal adapter
Important
The zero and span adjustment procedures for oxygen are different than other gas
sensors.
Applying a Gas Sensor
Applying a Gas Sensor
Gather your calibration gas source, the regulator (if not already connected), inert tubing, the cal adapter
and the EVM with the appropriate sensors installed. After your calibration setup is completed, follow the
proceeding applicable calibration sensor sections.
Applying a gas sensor
Figure 4-17: Applying a gas sensor
Calibrating Toxic Gas Sensors
It is recommended to conduct a zero cal and a span cal with toxic gas sensors. Essentially, a calibration
is conducted by checking the sensors response to a known concentration of a certified calibration gas. It
is important to note that with a zero cal and a span cal, you should be in a non-contaminated
environment when performing the calibration.
It is recommended to calibrate on a regular basis for best results. You will want to perform a zero cal
and a span cal (or adjustment) to properly calibrate a toxic sensor.
EVM Series User Manual
Page 67
Placement and Calibration of Sensors
57
Caution
Due to the presence of toxic gas during the calibration process, appropriate safety
procedures should be followed.
Additionally, you will want to ensure that your calibration is performed in an area
known not to contain hazardous or interfering gases if ambient air will be used as the
zero gas for operation. If this is not possible, pure bottled air (Nitrogen is
recommended) should be substituted. Instructions for both procedures are provided
below. You only need to perform one or the other.
Selecting a sensor
Press Up or Down Arrow on
keypad
Cal soft key
Calibrating Toxic Gas Sensors
Toxic Gas Zero Calibration
When conducting a zero cal for toxic gas, you will want to use a cylinder of Nitrogen or pure air with the
cal cup placed over the air inlet cover. (See “Applying a gas sensor” for details).
NOTE: If you are in clean air environment, you can use room air (for all toxics except for ozone).
Ozone Calibration for Zero and Span
For an Ozone calibration, a zero calibration is performed with Nitrogen and a span calibration is
performed using an ozone generator.
NOTE: It is recommended to factory calibrate the ozone sensor on a yearly basis.
Performing a Zero cal for toxic gas
1. While the instrument is powered off, connect the pure air cylinder, regulator, inert tubing, and cal
cup to the EVM.
2. Power on the EVM by pressing the On/Off/Esc.
3. In the start screen, press the Down ▼ arrow to select Calibration and press the Enter
4. Press ▲/▼Arrows to select the toxic you want to calibrate.
5. Press the Cal soft key and a specific toxic gas (i.e., CO) Calibrate Set Point screen will appear.
Figure 4-18: Selecting sensor for calibration process
.
EVM Series User Manual
Page 68
Placement and Calibration of Sensors
58
* Flashing asterisk indicates
cal. mode.
Level
Allow to stabilize before
setting calibration point
Press Set soft key to save
the level set point
Zero Cal
+ (plus sign) indicates
Zero cal has been set.
Span cal
- (minus sign) indicates User
Span cal has not been set.
Soft keys
Explanation
Cal
When pressed, this will allow a calibration to be performed.
Zero When pressed, this will enable you to a zero calibration.
Span
When pressed, this will enable you to perform a span calibration.
Exit
Enables you to exit at any time.
Calibrating Toxic Gas Sensors
6. Turn regulator On. Wait for the current level to stabilize (allow at least 90 seconds to stabilize.)
(The screen will show +Ready.)
7. When the level has stabilized, press Zero soft key.
Figure 4-19: Calibration zero screen
8. Press the Set soft key. A screen displaying the zero cal intended and actual set point values will
appear. Press the Save soft key.
Figure 4-20: Calibration zero pause screen
9. To set the span cal, proceed to “Toxic gas –span calibration”.
NOTE: To re-do the zero calibration, press the +Zero soft key and repeat steps 7 and 8.
Table 4-2: Gas calibration soft keys explained
EVM Series User Manual
Page 69
Placement and Calibration of Sensors
59
Level
Allow to stabilize
Set
Calibrating Toxic Gas Sensors
Toxic Gas Span Calibration
When conducting a span calibration, the calibrated gas used should match the sensor you will be
calibrating.
NOTE: For Ozone span calibration, an ozone generator should be used.) If you are continuing from
the Toxic Zero cal section, the screen you will be viewing is the Calibrate/Pause screen. (If
you are performing a Span cal only, ensure you are at the Start screen and select the
calibrate menu. Then select the specific sensor by pressing up/down arrows. Press the Cal
soft key and then press the Span soft key. Proceed to Step 2.
Performing a span cal for Toxic gas
1. Continuing from the zero cal procedures, you should be viewing the Cal/Pause screen (as
displayed in Figure 4-17). Press the Span soft key.
2. Apply the gas at full flow rate while observing the “Level” field on the display. Wait for the current
level to stabilize (allow a couple of minutes).
Figure 4-21: Span Calibration
3. Adjust the span value by pressing ▲/▼Arrowson the keypad.
4. Press the Set soft key to store the span cal. point. (A screen will appear displaying that you set the
span cal. point.)
EVM Series User Manual
Page 70
Placement and Calibration of Sensors
60
* New value
Previous calibration result
SAVE soft key
Calibrating CO2 AND O2
5. Once your calibration is set, intended and actual cal values will appear on the display screen.
Press Save to store these as your calibration set points.
NOTE: To redo the span calibration, press the Span soft key and repeat steps 2–5.
6. Turn the regulator off and disconnect the tubing from the instrument.
7. Repeatedly press On/Off/Esc until you return to the start screen.
8. Press ◄/►Arrows to view measurement screens.
Calibrating CO2 AND O2
With the Carbon Dioxide (CO2) or Oxygen (O2) calibration, you may perform a
zero calibration and a span calibration (when the cal gas is connected) in the
field. Each of these screens will be discussed below.
The CO2 sensor and O2 sensor requires a calibration gas, a regulator with at
least 1 L/min flow rate and inert tubing.
A zero calibration using a Nitrogen (N2) gas is recommended. For a span
calibration, the range (or span) should be a sizeable fraction of the selected
full scale range of the gas.
Figure 4-22: Saved Calibration screen
O2 Accuracy
The range for O2 sensor is 0–30% with a typical span cal of 20.9%
Calibrating a zero cal and span cal for CO
1. Connect one of the following gases with the regulator and inert tubing: N
gas (used for zero cal.)
2
CO
Room air for 0
2. Attach the regulator and tubing to the gas cylinder.
3. Place the Cal. cup over the Air inlet cover and slide the tubing over the black air inlet casing.
4. From the start screen, press ▼Arrow to select Calibration and press Enter
calibration gas
2
calibration gas (i.e., 20.9%)
2
or for 02 sensor
2
.
EVM Series User Manual
Page 71
Placement and Calibration of Sensors
61
Selecting a sensor
Press Up or Down Arrow
CAL soft key
Soft key Menu
Zero, Span or Exit
Calibrating CO2 AND O2
5.In the Calibration screen, press ▲/▼ Arrows to select CO
Figure 4-23: Selecting a calibration sensor
6. Press the CAL soft key to open the CO
“Calibration/Calibrate” screen. (Allow at least 2.5 minutes
2
or O2.
2
for CO2 to stabilize.)
For the Set Points, -zero and –span soft keys will appear. Press either Zero Cal or Span Cal
soft key. (See Table 4-2 for more information.)
Figure 4-24: Calibration CO2 ready screen
NOTE: If “–” (a minus sign) appears in front of Zero or Span, this has not been calibrated.
Alternatively, the “+” (plus sign) signifies the cal point has been calibrated.
7. Turn the regulator on. The regulator should have a flow rate of 1.0 L/min.
8. Allow the level to stabilize. For CO
For Zero Cal, the screen will display Zero with a Set soft key.
For span cal, press ▲/▼Arrows on the keypad to adjust the appropriate level.
, wait at least two minutes to stabilize.
2
EVM Series User Manual
Page 72
Placement and Calibration of Sensors
62
Level
Allow to stabilize before setting
calibration point
Set Soft key
Intended cal. values
❸ Cal before Zero cal
❹ Cal before Span cal
Actual cal. values
❶ After Zero Cal
❷ After Span Cal
Save
❶ ❷ ❸
❹
Calibrating CO2 AND O2
9. Press the Set soft key. Turn off regulator and unhook the N
bottle (for zero cal) or CO2 bottle (for
2
span cal). (To perform a Zero or Span cal, repeat steps 1–8.)
NOTE: To redo a zero or span calibration, press the zero or span soft key and repeat steps
above to program new set points.
Figure 4-25: Setting CO2 Span Cal screen
10. Once the zero and span cal values are set, the calibration save screen will appear. Press Save to
save your calibration.
NOTE: To repeat the cal, press the Cancel soft key and you will return to the calibration
screen. Repeat the steps above to program new set points.
Figure 4-26: Saved Calibration
11. Once saved, the display will indicate if it passed. You have another option to Retry the calibration
procedure by pressing the Retry soft key.
12. To exit CO
calibration, press the Exit soft key.
2
13. Then, press On/Off/Esc twice to return to the Start screen.
It is preferred to flush the PID ppm sensor with Zero Grade Air for a zero calibration. Alternatively, you
may flush the PID ppm sensor with Nitrogen. In clean environments (with no presence of toxic
chemicals, such as in a school setting with no toxic cleaners present) applying room air can be used to
set your zero calibration point.
For the flow rate, it is recommended to set it at 1 L/min for best response time.
PID parts per billion (ppb) zero calibration
NOTE for PID ppb sensor: The PID ppb sensor may be susceptible to baseline settlement after
more than 2 weeks of storage. In these cases, the PID PPB sensor should be operated in
clean air for several hours prior to performing a calibration and prior to performing
measurements of low concentration of gases (< 1000 PPB).
The PID ppb sensor uses a static base line for the zero offset. If the user adjusts zero base line, the PID
ppb needs to be enabled in the setup menu.
NOTE: To enable, select Setup from the main screen and then choose PID. In the PID screen,
enable the ppb zero feature. Once set, return to the calibration screen to resume the
calibration process. (For more details on setting, refer to “PID sensor setup” in Chapter 3.)
PID ppm/ppb zero cal
1. Connect the Zero Grade Air and cal cup to the EVM. (Skip this step if you are not applying a gas.)
2. From the start screen, press the Down▼ arrow to select Calibration and press Enter
3. Press the ▲/▼Arrows to select the PID sensor.
.
Figure 4-27: Selecting PID sensor for calibration process
EVM Series User Manual
Page 74
Placement and Calibration of Sensors
64
Zero
With PID ppb sensor, if
“Zero” is not displayed,
this indicates the PID
setup PPB sensor
should be set. See PID
ppb zero cal. section
above.
Units: ppm/ppb
The EVM will automatically
display the attached sensor
which is either PID ppm or
PID ppb for VOC testing.
Flashing asterisk
indicates cal. mode.
Level
Allow to stabilize before
setting calibration point
4.Press the Cal soft key and the Calibrate PID/Ready screenwill appear. Press Zero soft key.
NOTE: Allow at least 60 seconds to stabilize.
5. When the level has stabilized, press the Set soft key. (Or press Cancel soft key and then Retry
soft key to repeat this process.)
Figure 4-28: Calibration mode screen
6.The PID/Pause screen will appear displaying the intended and actual calibration values. Press the
Save soft key. (Or press the Cancel soft key to repeat this process.)
To set the span cal, proceed to “PID – span calibration”.
7. To restore the factory calibration for the ppb sensor, select the HISTR soft key from the PID
calibration screen. Press the FACTRY soft key. This restores the ppb zero calibration value.
PID Span Calibration
Typically for PID 100 ppm, isobutylene is used for a span calibration when the VOCs are unknown or
apply the gas of interest if detecting a specific VOC for best results.
To measure a specific VOC with PID PPM sensor, you will want to use the relevant VOC for your
calibrated gas source. However, for PID PPB span calibration, it is recommended to use 10 ppm (or
10,000 ppb) isobutylene whether the gas of interest is known/unknown.
NOTE: A correction factor is a number that relates to the PID sensor of a specific VOC and
correlates to the PID response to the calibration gas (sometimes referred to as a response
factor). Please reference Appendix C for specific correction factors provided by Ion
Science.
Performing a PID Span Calibration
Allow up to 15 minutes of warm-up time for PPB PID sensor and 5 minutes for PPM PID sensor. For
PPB span calibration, refer to PID ppb note if the sensor has not been used for over two weeks.
If you are proceeding from the Zero cal section, the screen you will be viewing is the Calibrate/Pause
screen. (If you closed the screens, refer to the Note in step 1.)
PID Span Cal
1. From the PID/Calibrate Pause screen, press the Span soft key on the keypad.
NOTE: To navigate to the PID Span cal screen, select the Calibration screen. Select PID
using the up/down arrows. Then, press the Cal soft key. Zero and Span soft keys will
appear. It will look similar to the screen below but it will not have the intended/actual
zero cal set points displayed. See step 2 to proceed.
Figure 4-29: PID Calibration Span soft key
2. Adjust the span value by pressing ▲/▼Arrows once the sensor has warmed-up.
3. Apply the gas at full flow rate while observing the “Level” field on the display. Wait for the current
level to stabilize (allow a couple of minutes).
4. Press the Set soft key (and it will state Cal set point stored).
5. Once your calibration is completed, a Save screen will appear with the details of your parameters.
NOTE: If you want to repeat the span cal, press cancel and a message will appear stating it is
canceling. Press the Retry soft key to repeat the span cal steps.
EVM Series User Manual
Page 76
Placement and Calibration of Sensors
66
Calibration results after
completed span cal.
Previous calibration
SAVE soft key
Calibrating Relative Humidity and Temperature
6.Press Save soft key to store this into the memory.
Figure 4-30: Saved Calibration screen
7. Turn the regulator off and disconnect the tubing from the instrument.
8. Repeatedly press On/Off/Esc until you return to the start screen. (Press right or left to view
measurement screens.)
Calibrating Relative Humidity and Temperature
The relative humidity and temperature sensors are factory calibrated. To perform a calibration, you
would place the instrument into the location you will be testing with the appropriate temperature sensing
instrument (such as a thermometer for a temperature calibration). Then, access the calibration screen
and allow the Relative Humidity and Temperature readings to stabilize. Preceding the calibration level,
you then store these values.
NOTE: When performing a user calibration on Relative Humidity and Temperature, do not
externally power/charge the instrument. If you are currently charging or using an external
power adapter, disconnect before proceeding.
Relative Humidity or Temperature calibration
1. Open the Calibration screen.
2. Press ▲/▼Arrows and select either RH or Temperature calibration.
3. Select the Cal soft key and then press the Start soft key.
4. Adjust the level to match the current value of a calibrated temperature sensing instrument.
To adjust the level, press ▲/▼Arrows until the appropriate value is selected.
5. Press the Set soft key and a Save screen will appear.
EVM Series User Manual
Page 77
Placement and Calibration of Sensors
67
Temperature cal
Intended & actual readings
Saved Cal
Before Set point
Saved Cal
After Set point
Calibration History
6. Press the Save soft key to save the calibration.
7. To calibrate the other sensor (RH or Temp), repeat these steps.
Or press the Exit soft key and then press On/Off/Esc.
To open the calibration screen: from the start screen, press the Down▼arrow to select
Calibration and press Enter . The Calibration screen will appear.
8. To exit and return to the main screen either press the On/Off/Esc twice.
Calibration History
The calibration history screen will show your previous calibration of a specific sensor. When the sensor
is selected, the last calibration performed will be shown. To view the calibrated values, an optional
History menu is selectable via a History soft key. The figure below is an example of a CO2 cal. history.
Figure 4-31: Temperature calibration
Figure 4-32: Calibration History screen
Accessing the Calibration history screen
1.Select the Calibrate menu from the start screen and then choose a sensor.
NOTE: Press up/down arrow in the calibrate menu to select a sensor.
2. If the sensor selected was recently field calibrated, the History softtkey menu will appear along
with the date and time of the last calibration.
3. Press the HISTR soft key to view the saved calibration results.
EVM Series User Manual
Page 78
Placement and Calibration of Sensors
68
Calibrate offset value
Verify Calibration and Drifting Analysis
4.To exit and return to the start screen press On/Off/Esc twice.
Verify Calibration and Drifting Analysis
The verify screen is typically used to analyze drifting. This process verifies if the zero calibration
remained stable or unchanged during the run-time/logging time. If it did not remain stable, you can set a
new calibration offset and save this as your new set point.
Verify calibration
1. Select the Calibration menu from the start screen and select a sensor you want to verify.
NOTE: Press up/down arrow menu to select a sensor in the calibration menu.
2. Press the Verify soft key. (The Calibrate offset will display on the screen.)
NOTE: For this example, PID sensor was selected in step 1. If you selected a different sensor,
the screens will look very similar to the figures below.
Figure 4-33: Calibrate offset verify screen for PID sensor
3. Press the Start soft key and allow the level to stabilize. (Apply any gas if applicable as described in
the Calibration sections above.)
4. Press the Set soft key to set a new calibration value.
Figure 4-34: Calibrating PID
EVM Series User Manual
Page 79
Placement and Calibration of Sensors
69
Verify Calibration and Drifting Analysis
5. The Intended and Actual values will appear. Press the Save soft key to save these new values. Or press the Cancel soft key and press Retry to repeat this process.
6. To return to the start screen, press On/Off/Esc twice. To retry the offset cal, press the Retry soft key and repeat the steps above.
EVM Series User Manual
Page 80
Placement and Calibration of Sensors
70
Verify Calibration and Drifting Analysis
(This page intentionally left blank)
EVM Series User Manual
Page 81
Measuring and Viewing
71
Quick Help List: typical gas/particulate/air quality overview
1. Turn the EVM on and check your battery power (see Chapter 2).
2. Reset/Clear memory (if necessary). o This will clear out all of the data history before your study begins to ensure you have sufficient
memory space.
3. Set-up your study parameters (see Chapter 3).
4. Calibrate each sensor/parameter (see Chapter 4).
5. Place the EVM in a specific location for area monitoring.
6. Press the Run/Stop key to start the study (see Chapter 5)
7. When you are ready to end your study, press the Run/Stop key to stop the study.
8. Review your results either on the instrument or download to DMS for further analysis (graphs/charts,
saving and printing).
CHAPTER
5
Overview of Running a Session
Measuring and Viewing
Once you have setup your logging parameters and calibrated the instrument, you are ready to begin
measuring. The focus of this chapter is to give you a brief overview of what is measured, how to
navigate, what measurements screens will appear, and how to run and stop your measurements. The
chapter concludes with an explanation of the file system management functionality which is comprised
of the following: reviewing past measurements, deleting files, viewing file storage and memory, and
viewing the estimated run-time.
NOTE: The subsequent sections show examples of views which you may have disabled on your
instrument (refer to Chapter 3, “Setting-up the EVM” for more details).
Overview of Running a Session
The list below is a brief overview of the steps for running a study.
Quick Help
Figure 5-1: Quick help list for running a study
Measuring
Once you have placed the Environmental Monitor (EVM) in a specific location, you are ready to
take measurements.
With a couple pushes on the keypad buttons, you will be running and viewing your sessions. Following
your sessions you can load the data/files into DMS to manipulate, analyze, save, and create reports
(See Chapter 6 for viewing your logged data in DMS).
EVM Series User Manual
Page 82
Measuring
72
Understanding What Data is Stored
Understanding What Data is Stored
When you are performing air quality monitoring, there are two types of data stored in the instrument
which includes summary data and time history data (or logged data).
Summary Data
Summary data represents values that occurred over the entire run of the study. An average
measurement is the average of all arithmetic data points for a sensor (i.e., dust, humidity, CO, etc..),
Max and Min values are the highest and lowest measurements that occurred throughout the run time.
Figure 5-2: Sample summary data from a DMS table
Logged Data/Time History
Logged data, also referred to as “time history,” plots measurements with a specified logging interval
ranging from 1 second intervals to 60 minute intervals (as set up in the EVM monitors via the logging
screen).
Time History data represents measurements that occurred over the most recent time period, which is
the logging interval. For example, if the log interval is 1 minute, then the maximum value is the highest
value for that 1 minute, and the Lavg value is the average value for that 1 minute.
NOTE: DMS has two default logged data charts that appear: temperature and humidity on the first
logged chart, and Dust and CO2 data on the second logged data chart. (If you do not have
those sensors installed, they will not appear.) In DMS, to view different sensor
measurement parameters on the logged data chart, click on the and click in
the specific sensors’ checkboxes.
EVM Series User Manual
Page 83
Measuring
73
Particulates
Will display max.
and low peaks in
logged data
chart.
CO2 data
Run mode
Level values for each
sensor
Shortcut: Press Enter
key to toggle through
different measurement
units.
Running and Stopping
Figure 5-3: Sample time history data (or logged data) chart
Running and Stopping
When running a session, the pump will automatically turn on during your logging/run if particulate is
enabled. (This is explained in the Particulate setup section.) At any time, you can turn the pump off
(disable) if you are not measuring particulates or the environment is extremely dirty and you want to
keep the pump clean. Simply press the particulate key (as indicated in Figure 2-2) to turn this feature
off. To page through the measurement screens, press ◄/► Arrows.
NOTE: For particulates ensure the impactor is well-greased before measuring (see “Impactor Care”
for details).
Running and Stopping measurements
1. From the start screen or any of the measurement screens, press the Run/Stop
key to begin
measuring.
A 5 second countdown will appear with a run message (in pause mode) when the particulate
setting is enabled.
2.Press and hold the Run/Stop
Figure 5-4: Composite measurement screen in run mode
key to stop measuring.
EVM Series User Manual
Page 84
Measurement Screens and Navigation
74
Abbreviations
on the EVM
Explanation
OL
Over Load – This indicates a measurement (on the current screen) is above the maximum
value (for that sensor) which the instrument can measure.
UR
Under Range – A measurement (on the current screen) is below the minimum value (for
that sensor) which the instrument can measure.
ºF
Degrees Fahrenheit – This is displayed with readings for Temperature and Dew Point.
ºC
Degrees Celsius – This is displayed with readings for Temperature and Dew Point.
%
Percent- This is displayed for the percentage measurement for the pump flow rate, the
relative humidity sensor, and the Oxygen sensor.
ppm
Parts per Million – This is displayed with measurement readings of CO2 , toxic gas, and
PID sensor(for VOCs).
ppb
Parts per Billion – This is displayed with reading of PID sensor (for VOCs).
mg/m3
Milligrams per cubic meter – This is a unit of measurement (of mass per unit volume) and
is used for particulate measurements and PID PPM (PID sensor for VOCs).
μg/m3
Micrograms per cubic meter – This is a unit of measurement (of mass per unit volume)
and is used for particulate measurements and PID PPB (PID sensor for VOCs).
m3
Volume – Gravimetric volume for the particulate measurements.
mg
Mass in Milligrams – Gravimetric accumulator mass for the collected particulates.
m/s
Meters per second – This is the unit of measurement for metric air velocity.
ft/m
Feet per minute – This is the unit of measurement for air velocity.
+
Unit defined
Measurement Screens and Navigation
To navigate through the measurement screens repeatedly press ◄/►Arrows key from the Start
screen. (The start screen is the second screen you encounter after the initial splash screen.
Looking at the measurement screens clockwise (or pressing ►Arrow key), you will see the particulate
measurement screen. The subsequent screens that appear include the relative humidity and
temperature screen, the gas screen, the composite screen, and the air velocity screen. Depending upon
the sensors you purchased, the installed sensors, and the measurement setup you may see some
variations in the upcoming measurement screens in comparison to your instrument. These will be noted
(or explained in full detail).
Unit defined
While viewing your measurement data, you will want to have a good understanding of the abbreviated
units used on the instrument. These are listed in Table 5-1.
Table 5-1: Unit abbreviations
EVM Series User Manual
Page 85
Measurement Screens and Navigation
75
Soft key menus
Used to select additional
menus or different
measurements/units
Measurement
Soft keys
Explanation
Lookup soft
key
(eyeglasses)
Lookup icon (eyeglasses) soft key – used to select and view different measurement values
(on the particulate and air velocity screens).
Level, Maximum, Minimum, Average, Short Term Exposure Level (STEL) and Time Weighted
Average (TWA).
Select soft key
Select soft key - if more than one sensor/parameter is selectable on a measurement screen,
this is used to select a sensor such as: temperature, toxic gas sensor, PID sensor etc. If you
are on the composite screen and the screen is magnified, you can select each sensor and the
measurement units will be displayed on the bottom of the screen.
+/–
Magnifying glass symbol – Used to Zoom in (+) or out (–). (The plus denotes zoomed in
and the minus sign denotes zoomed out.)
Detail soft key
Detail soft key – used to view a sensor’s measurement values in a summary screen. For
example, if you are on the composite measurement screen and you want to view the
summary data for CO2, you would first select the CO2 field and then select the Detail soft key.
You may need to press the up/down arrow to scroll through the data.
Trend soft key
Trend soft key – is graphical chart and is used to give you a quick synopsis of trends over
time such as a high volume of particulates at a specific time. The following settings are
always logged and are viewable by pressing the Altf key when you are viewing the trend
screen: 90 sec, 180 sec, 900 sec, 90 min, 3 hrs, 8 hrs, 12 hrs, 24 hrs. The timescale (x-axis)
can be expanded by pressing the right or left arrow keys. The measurement level (y-axis) can
be expanded and contracted (some sensors) by pressing the up or down arrow keys.
About the Measurement Soft Keys
About the Measurement Soft Keys
Each measurement screen has soft keys (extended menus) which are used to view different
measurements.
NOTE: Press the Enter key to toggle through the various measurements.)
Figure 5-5: Example of measurement soft key menus
Table 5-2: Measurement soft keys explained
EVM Series User Manual
Page 86
Particulate Measurement Screen
76
Measurement displayed
is changeable via the
lookup/glasses icon or
by pressing Enter key
Impactor and profile
displayed
Particulate Level is
0.445 mg/m3
Press Detail soft key to view
all measurement values
Impactors and Particulate Measuring
Particulate Measurement Screen
The particulate measurement screen displays the user-adjustable impactor setting (i.e., PM 2.5), the
applied profile (or correction factor) and various measurements (level, max, min, etc.)
Figure 5-6: Particulate Measurement screen
Impactors and Particulate Measuring
As discussed in the calibration chapter, there are four selectable particulate settings. To move the
impactor, ensure you are in stop mode. Then, twist to the right or left until you feel a tactile detent. The
impactor setting will align to the gray marking on the housing (see Figure 4-4 for an example).
NOTE: The particulate screen will display “ERR” if it is not clicked into one of the appropriate
settings. The selections include: PM2.5, PM4, PM10, and PM.
NOTE: The impactors may require cleaning during or after your measurement study. Please see
Chapter 4, “Impactor and Use” for more information.
Extending Run-Time for PM Measurements
If you will be measuring in a dusty environment (such as an environment that is oily) or for a long period
of time (10+ hours), you may need to clean the impactors frequently in order to remove the debris buildup on the impactor. (See Chapter 7 for Maintenance and cleaning for more detailed instructions.)
To extend the run-time for PM measurements, attach a cyclone to the turret. The cyclone is used to
prevent the larger dust particles from building-up on the impactor which eventually clogs the impactor. It
forces the larger particulates, using centrifugal force, down and around the filter and collects in the
bottom holding cap. (See Figure 4-6 for more details.)
NOTE: You use a cyclone that is designed to work at 1.67 L/min pump flow rate.
EVM Series User Manual
Page 87
Particulate Measurement Screen
77
Look-up/
Glasses icon.
Press Up/Down Arrow to
select a measurement
value.
Trend soft key.
Detail soft key.
LEVEL
MIN
MAX
G
STEL
TWA
AVG
Changing Particulate Measurements
Changing Particulate Measurements
To change the displayed measurements, there are three options in the particulate screen. You can
press the Enter key to page through each measurement, select a specific measurement from the
lookup/glasses soft key, and/or press the Detail soft key. (The measurements are defined in Table 5-3).
NOTE: If “---” characters appear on the screen, this denotes that the measurement is invalid.
To change your measurement view
1. Click on the “Look-up” icon.
2. In the pop-up measurement menu press the Up/Down arrow to select a different type.
3. Press Enter
4. To view a graphical view of the data, press the Trend soft key. In the Trend screen, press the Line
soft key. This switches the screen from a dotted line to a shaded bar chart view. To exit this
screen, press On/Off/Esc.
NOTE: Press the Altfkey to toggle between the trend time scale and the soft keys.
to select. On the particulate screen, the selected measurement type will appear.
Figure 5-7: Selecting a measurement calculation
EVM Series User Manual
Page 88
Relative Humidity and Temperature Measurements
78
Measurement
values
Explanation
LEVEL
Level - Notation used to represent current live reading of the sensor. (This is updated once a
second.)
MIN
Minimum - Minimum level value which occurred during a log session.
MAX
Maximum - Maximum level value which occurred during a log session.
AVG
Average - Arithmetic average of all level values which occurred during a log session.
STEL
Short term exposure level - is a calculation used to indicate an exposure level (to a toxin)
over a short-term period. This value is calculated by averaging the last 15 one minute
averages. This value is invalid for the first 15 minutes and is updated every minute thereafter.
Essentially, it is used when the building occupant/worker is exposed to high concentrations of
chemical gases. The measurement will average after 15 minutes of data logging and continue
to compute after each minute of logging, using measurements from the most recent fifteen
minutes. If less than 15 minutes has elapsed, the STEL will be invalid by displaying dashes.
TWA
Time weighted average - is an average based on an 8-hour interval. The summing all of the
level values taken and dividing by the number of samples in 8 hours (at one sample per
second the total would be 28,800). The important thing here is that the average is calculated
the same way regardless of whether you've run the instrument for 1 hour, 8 hours, 24 hours
or any other amount of time (i.e. the number of samples used to divide by is the same). This
value is updated once a second.
To select a different
measurement type, press
the Enter key repeatedly.
Select, Trend, Detail soft
keys (used to change the
measurement view).
Relative Humidity and Temperature Screen
5. Or select the Detail soft key and a summary of all the measurements will appear. (Below is an
explanation of measurements.)
Table 5-3: Measurement values explained
Relative Humidity and Temperature Measurements
The relative humidity and temperature sensor will display on the left-hand side of the screen with two
measurement values. The measurement displayed in the second column is user-selectable via
Enter .
Relative Humidity and Temperature Screen
Figure 5-8: RH and temperature measurement screen
EVM Series User Manual
Page 89
Gas Measurements
79
Gas Screen
To change the RH and Temperature measurement view
1. Press the Select soft key, Trend soft key, or the Detail soft key. (See Table 5-1 for details).
NOTE: The ▲/▼ Arrows can be used instead of the Select soft key.
Gas Measurements
The gas measurement screen displays the measurements for CO2, toxic gas(es), and PID sensor.
NOTE: Depending on what sensors you ordered and which sensors are installed, the appropriate
gas sensors will display on this screen.
Gas Screen
Figure 5-9: Gas measurement screen
To change the gas measurements view
1.Press the Select soft key, Trend soft key, or the Detail soft key. (See “Viewing summary/details”
or “Viewing trend details.)
NOTE: The ▲/▼ Arrows can be used instead of the Select soft key.
EVM Series User Manual
Page 90
Composite Measurement
80
+ Zoomed in: The composite measurement screen (when zoomed- in) combines all
sensors/parameters into one screen with two columns displaying two different
measurement calculations. The second column is changeable by pressing Enter .
(See “B” below.)
– Zoomed out: The composite measurement screen (when zoomed-out), appears with
the measurements for all installed sensors. Press Enter
to change the
measurement values (as explained in Table 5-3). Press ▲/▼Arrows or Enter
to
page through sensors. (See “A” below.)
A- Zoomed -out (Magnify soft key)
B – Zoomed-in (Magnify soft key)
Composite Screens
Composite Measurement
When you first access the composite screen, all of the sensors appear on the screen. There are two
main views which is dependent on if you pressed the zoomed-in or zoomed-out soft key. (The “zoomedout” is the default composite view.)
Composite Screens
Figure 5-10: Composite measurement screen
To change the composite measurements/view
1.Press the Select soft key, Trend soft key, or the Detail soft key (see “Viewing summary/details” or
“Viewing trend data” for more details).
2.To zoom in on the results, press the Magnify soft key (see “B”).
EVM Series User Manual
Page 91
Air Velocity Meaursurement Screen
81
Air Probe
bracket
attachment
Output
*Pull back rubber
covering in order
to plug in cable.
Air velocity measurement
screen
Connect
Air Probe to the
back housing
(two inlets located
near the air flow
outlet. See air
flow diagram on
previous page. )
Composite Screens
Air Velocity Meaursurement Screen
There are a couple quick steps to ensure air velocity is setup and that the Air Probe is attached before
measuring.
1. The air velocity setting is enabled in Setup /AUX/Air Velocity screen (see “Air Velocity” Setup in
Chapter 3 for more details).
2. Connect the Air Probe bracket attachment to the backside of the EVM. Attach the Air Probe to the
bracket. Next, plug in Air Probe plug from the Air Probe to the digital output on the instrument.
(See Air Probe output below.)
NOTE: It is optional to attach the Air Probe to the bracket on the instrument.
3. Turn Air Probe switch to On.
4. Navigate to the Air Velocity measurement screen using the ◄/►Arrows.
5. To log air velocity data, press the Run/Stop key.
Figure 5-11: Attaching Air Probe and measuring Air velocity
EVM Series User Manual
Page 92
Summary Data
82
PM (Composite) details
screen
Exit soft key
Viewing Details in the Measurement Screen
Summary Data
In each of the measurement screens, a Detail soft key will appear. Once selected, you can view a quick
summary of all measurement calculations in one screen. This is explained below.
Viewing Details in the Measurement Screen
Viewing the composite measurements in individual summary screens
1. To view details of a sensor, select a measurement screen and then select/highlight a specific
sensor (if more than one is displayed).
For example, to select PM from the Composite screen, press ▲/▼Arrows or the Select
soft key.
2. Press the Detail soft key and a summary screen of the selected measurement screen (i.e., PM 2.5
– particulates screen) will appear.
Figure 5-12: Details of the composite measurement screen
3.Press On/Off/Esc to return to the previous screen.
Repeat these steps to view other sensor measurements.
EVM Series User Manual
Page 93
Trend Data
83
Trend soft key
Press third soft key to view
trend data
Viewing trend data in the measurement screen
Trend Data
The trend data screen is a graphical logged data chart and is viewable in all the measurement screens.
Depending on the parameter you are viewing (i.e., PID sensor, Particulates, etc...), it will display the
measurement level values in a graphical format.
Viewing trend data in the measurement screen
The example in Figure 5-13 displays particulate trend data with one second level values.
Trend data screen
1. To select a trend screen, first select a measurement screen you want to view.
Figure 5-13: Selecting Trend soft key
2.In the measurement screen, press the Trend soft key and a graphical data chart will appear.
You can either view the data as a “Line” format or a “Bar” chart format by pressing these soft
keys on the bottom of the display.
3. (Optional) Press the Altf key to toggle between the trend time scale units and the soft key menu.
4. (Optional) Press ▲/▼Arrows to change the vertical range.
EVM Series User Manual
Page 94
Past Session
84
Scale
Press up/down to
change range
Trend Screen (Dust
example)
Plots data logged points over
90 second intervals
Line (or Bar) soft keys
View trend data as a bar
chart or as a line chart by
pressing this soft key.
FILES
Allows user to choose any
past session. (This appears
when File field is selcted.)
DETAIL
Used to view the summary
data of the selected session.
Summary Data Screen
5. (Optional) Press ◄/►Arrows to change the horizontal range.
6. Press On/Off/Esc to return to previous screen.
Past Session
The past session summary screen allows you to view your session summary data. You can view
sessions by name and file size, name and date, or name and run time via the lookup soft key. Once a
session is selected, the measurement summary information is viewable and selectable via
◄/►Arrows.
NOTE: The session run time will appear at the top of display.
Figure 5-14: Trend Screen for PM (Dust)
Summary Data Screen
Selecting and viewing past session summary data
1. From the Start screen, select past session summary menu and press Enter
.
2. In the past session screen, to view the last session, ensure this field is selected then press Detail
soft key.
Figure 5-15: Past session screen
EVM Series User Manual
Page 95
Past Session
85
Look-up
Re-arrange how the
sessions are organized.
Past sessions listed by
Name and date order
Files soft keys
Explanation
Lookup
Look-up soft key: used to organize the session files by the following: Name &
file size, Name &date, or Name& run time
PG-UP
Displays the previous page of session files.
PG-DN
Displays the next page of session files.
LOAD
Once you select a session file to view, press the Load soft key in order to
LOAD and then view the detailed summary data screens.
Summary Data Screen
3. To select a specific file, first select the File field and then press the Files soft keyand the
LoadFile screen will appear.
Optional: press the Look-up soft key to view the file by name and file size, name and date, or
name and run time. To select, press up/down arrow and then press the Load soft key to
select.
NOTE: Pressing the Enter key will sequence the files size, date, and run time.
To select a session, press ▲/▼ Arrows and press the Load soft key.
Figure 5-16: Past Session with Files selected
Table 5-4: Past session files soft keys explained
EVM Series User Manual
Page 96
File System Menu
86
Session run time
Summary data
Measurements viewable
for particulates
PM4
Indicates the summary
data you are viewing.
Deleting Sessions Individually
4. Once a file is selected and loaded, press the Detail soft key to view the summary data.
NOTE: The particulates, toxic gas, CO2 gas, and PID sensors will detail the following
measurement parameters: Min, Max, Avg, Max STEL, TWA. For RH, Temp, and Air
Velocity, the measurement parameters include: Max, Min, and Avg.
Figure 5-17: Summary screen of a past session
5. When viewing summary data, press ◄/►Arrows to view summary data for each sensor. (This is
similar to viewing measurements during run or stop mode.)
In the summary data screen, the session run time will appear in the upper hand corner of the
display.
NOTE: For particulates summary data, it will display the impactor setting applied during the
measurement (i.e., PM4).
File System Menu
Deleting Sessions Individually
6. To return to the main screen, press On/Off/Esc twice to return to the Start screen.
With the File system menu, you can delete or clear sessions one-by-one or delete all sessions. You can
also verify the used and free memory of the instrument via the Properties menu. The instructions below
explain the Session Directory menu, Delete All Session menu, and the File System Proprieties menu.
Deleting sessions individually
1. Open the File System screen from the Start menu and select the Session Directory menu. Then,
press Enter to open.
2. In the File System/Sessions screen, you have the same soft key selections as the Past Session
Summary menu. (See Table 5-4 for soft key explanation.)
3. To delete a selected session, first select a specific session to delete by pressing the
▲/▼ Arrows.
EVM Series User Manual
Page 97
File System Menu
87
Delete message
Press left/right arrow to
select “Yes” or “No”. Then
press Enter key.
Delete all sessions
message
Press left/right arrow to
select “Yes” or “No”.
Then press Enter key.
Delete All Sessions
4. Press the Delete soft key. A Deletepop-up screen will appear. To confirm the delete, press the
Left arrow and then press the Enter . (This will delete the selected session.)
5. Repeat these steps until all of the appropriate sessions are deleted (or see the next section,
“Delete All” to delete all of the sessions.)
Delete All Sessions
Once you have reviewed your data and possibly downloaded to DMS for future storage, you can delete
all sessions.
Deleting all sessions
1. Open the File System screen from the Start menu and select the Delete All Sessions menu.
Then, press Enter
2. A delete message will appear. Press the left arrow to select Yes and press Enter
all sessions.
Figure 5-18: Delete session message
to open.
to delete
Figure 5-19: Delete all sessions message
3. A message stating “deleted all sessions will appear”.
4. To return to the start screen, press On/Off/Esc.
EVM Series User Manual
Page 98
File System Menu
88
File system properties
Summarizes the memory
remaining on the
instrument.
Memory Remaining and File Properties
Memory Remaining and File Properties
The properties screen displays the number of stored files, the total amount of space used on the
instrument, and the amount of free memory space.
Viewing remaining memory and file properties
1. Open the File System screen from the Start menu and select the Properties menu by pressing
the Down arrow key.
2. Then, press Enter
Num Files field – indicates the number of files saved on the instrument.
Used field – indicates the amount of memory used in the instrument’s file storage.
Free field – indicates the amount of memory available (or free) on the instrument.
key to open the Properties screen.
Figure 5-20: File Systems properties screen
EVM Series User Manual
Page 99
File System Menu
89
Estimated Run Time
Estimated Run Time
The Estimated Run Time screen displays the amount of days, hours, and minutes’ remaining on the
instrument. You can extend the run time if you log less parameters (such as, only logging MAX or
increasing the log interval.)
You may want to download and save your files in DMS and then clear the memory on the EVM. This will
also extend the logged time. For more information on estimated run time charts, see Appendix A, “Estimated logged run times.”
NOTE: The estimated run time screen is also viewable via the logging setup screen (for more
details, see Chapter 3, “setting logging and viewing runtime”).
View estimated run time
1. Open the File System screen from the Start menu and select the Estimated Run Time menu by
repeatedly pressing the Down arrow key.
2. Then, press Enter
key and the Estimated Run Time screen will appear.
Figure 5-21: Viewing estimated run time
EVM Series User Manual
Page 100
File System Menu
90
Estimated Run Time
(This page intentionally left blank)
EVM Series User Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.