TSI Instruments dusttrak II 8530, dusttrak II 8531, dusttrak II 8532, DUSTTRAK 8533, DUSTTRAK 8534 Operation And Service Manual

Page 1
Exposure Monitoring
Model 8530/8531/8532
DUSTTRAK™ II Aerosol Monitor
Operation and Service Manual
March 2010
Page 2
Copyright © TSI Incorporated / 2008–2010 / All rights reserved.
Limitation of Warranty and Liability (effective July 2000)
Seller warrants the goods sold hereunder, under normal use and service as described in the operator's manual, shall be free from defects in workmanship and material for twenty-four (24) months, or the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components
when indicated in specifications, are warranted for 90 days from the date of shipment.
b. Par ts repaired or replaced as a result of repair services are warranted to be free from defects
in workmanship and material, under normal use, for 90 days from the date of shipment.
c. Seller does not provide any w arranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies.
d. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with
respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems.
Page 3
i
CONTENTS
SAFETY INFORMATION ................................................................... III
Laser Safety ...................................................................................... iii
Labels ............................................................................................... iv
Description of Caution/Warning Symbols ......................................... iv
Caution .......................................................................................... iv
Warning ......................................................................................... iv
Caution and Warning Symbols ......................................................... v
Reusing and Recycling ..................................................................... v
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ........... 1
Unpacking the DUSTTRAK II Aerosol Monitor ................................ 1
Parts Identification for the DUSTTRAK™ II Desktop Aerosol
Monitor Models 8530/8531 ............................................................ 6
Parts Identification for the DUSTTRAK™ II Handheld Aerosol
Monitor Model 8532 ....................................................................... 7
CHAPTER 2 SETTING UP ................................................................ 9
Supplying Power to the DUSTTRAK™ II Aerosol Monitor .................. 9
Installing the Batteries in Model 8530/8531 Desktop .................... 9
Installing the Batteries in Model 8532 Handheld ........................... 9
Using the AC Adapter to Run Instrument .................................... 10
Battery Charging ......................................................................... 10
Inlet Cap ...................................................................................... 10
Size-Selective Impactors ............................................................. 11
Dorr-Oliver Cyclone ..................................................................... 12
Instrument Setup ............................................................................ 12
Connecting to the Computer ....................................................... 12
Installing TRAKPROTM Data Analysis Software ............................. 13
Connecting Analog/Alarm Output ................................................ 14
Wiring the Analog Output ............................................................... 14
Wiring the Alarm ............................................................................. 15
CHAPTER 3 OPERATION .............................................................. 17
Getting Started ................................................................................ 17
Setup Menu .................................................................................... 18
Zero Cal ....................................................................................... 19
Flow Cal ....................................................................................... 20
User Cal ....................................................................................... 21
Alarm ........................................................................................... 24
Analog ......................................................................................... 26
Settings ........................................................................................ 27
Run Mode .................................................................................... 29
Survey Mode ............................................................................... 30
Manual Mode ............................................................................... 31
Log Mode (1–5) ........................................................................... 32
Page 4
ii
Taking Mass Concentration Measurements ................................... 33
Screen Regions ........................................................................... 34
Stats ............................................................................................. 35
Graphing ...................................................................................... 35
Viewing Data ................................................................................... 37
Title Bar ........................................................................................... 38
CHAPTER 4 MAINTENANCE ......................................................... 39
Maintenance Schedule ................................................................... 39
Zeroing Instrument .......................................................................... 40
Cleaning the Inlet ............................................................................ 41
Cleaning and Oiling Impactors ........................................................ 42
Replacing the Internal Filters .......................................................... 42
Storage Precautions ....................................................................... 46
CHAPTER 5 TROUBLESHOOTING ............................................... 47
APPENDIX A SPECIFICATIONS ................................................... 51
APPENDIX B ZERO MODULE ...................................................... 53
INDEX ................................................................................................ 55
Page 5
iii
Safety Information
IMPORTANT
There are no user serviceable parts inside the instrument. Refer all repair and maintenance to a qualified factory-authorized technician. All maintenance and repair information in this manual is included for use by a qualified factory-authorized technician.
Laser Safety
The Model 8530/8531/8532 DUSTTRAK II is a Class I laser-based
instrument.
During normal operation, you will not be exposed to laser radiation.
Precaution should be taken to avoid exposure to hazardous radiation in
the form of intense, focused, visible light.
Exposure to this light may cause blindness.
Take these precautions:
DO NOT remove any parts from the D
USTTRAK II monitor unless you
are specifically told to do so in this manual
DO NOT remove the housing or covers. There are no serviceable
components inside the housing.
WARNING
The use of controls, adjustments, or procedures other than those specified in this manual may result in exposure to hazardous optical radiation.
WARNING
There are no user-serviceable parts insi de this instrument. The instrument should only be opened by TSI or a TSI approved service technician.
When operated according to the manufacturer’s instruction, this device is a Class I laser product as defined by U.S. Department of Health and Human Services standards under the Radiation Control for Health and Safety Act of
1968. A certification and identification label like the one shown below is affixed to each instrument.
Page 6
iv
Labels
Advisory labels and identification labels are attached to the instrument.
1. Serial Number Label (bottom)
2. Laser Radiation Label (internal)
DANGER!
VISIBLE LASER RADIATION WHEN OPEN. AVOID DIRECT EXPOSURE
TO BEAM
WARNING: NO USER SERVICABLE
PARTS INSIDE. REFER SERVICING
TO QUALIFIED PERSONNEL
3. European symbol for non­disposable item. Item must be recycled.
Description of Caution/Warning Symbols
Appropriate caution/warning statements are used throughout the manual and on the instrument that require you to take cautionary measures when working with the instrument.
Caution
Caution
Failure to follow the procedures prescribed in this manual might result in irreparable equipment damage. Important information about the operati on and maintenance of this instrument is included in this
manual.
Warning
WARNING
Warning means that unsafe use of the instr ument could result in serious injury to you or cause damage to the instrument. Follow the procedures prescribed.
Page 7
v
Caution and Warning Symbols
The following symbols may accompany cautions and warnings to indicate the nature and consequences of hazards:
Warns that the instrument contains a laser and that important information about its safe operation and maintenance is included in the manual.
Warns that the instrument is suscep tible to electro-static discharge (ESD) and ESD protection should be followed to avoid damage.
Indicates the connector is connected to earth ground and cabinet ground.
Reusing and Recycling
As part of TSI Incorporated’s effort to have a minimal negative impact on the communities in which its products are manufactured and used:
Do not dispose of used batteries in the trash.
Follow local environmental requirements for battery recycling.
If instrument becomes obsolete, return to TSI for
disassembly and recycling.
Page 8
vi
(page intentionally left blank)
Page 9
1
Chapter 1
Unpacking and Parts Identification
Carefully unpack the Model 8530/8531/8532 DUSTTRAKTM II Aerosol Monitor from the shipping container. Use the tables and illustrations below to make certain that there are no missing components. Contact TSI immediately if anything is missing or damaged.
Note
If you purchased a DUSTTRAKTM II Model 8530-NA (no accessories) Aerosol Monitor, it only comes with the following items:
D
USTTRAK
TM
II Model 8530 Aerosol Monitor
Operations manual
TrakProTM Data Analysis Software CD
One-year calibration certificate
Service paperwork
2-year warranty
All accessories for the DUSTTRAKTM II Model 8530 Aerosol Monitor are sold separately Contact TSI at (800) 874-2811 for information on accessories and how to purchase them through a TSI sales representative.
Unpacking the DUSTTRAK II Aerosol Monitor
Compare all the components you received with those listed in the table below. If any parts are missing, contact TSI.
Item Qty
Part Number Description
or
1 8530
8531
8532
Desktop II Desktop II HC
Handheld II
Page 10
Chapter 1
2
Item Qty
Part Number Description
1 801670
801669
Desktop II Carrying Case
Handheld II Carrying Case
1 1090014 Data Analysis
Software CD­ROM
1 800663 Zero Filter
or
1 801680
801681
6600 mAH Lithium Ion Rechargeable Battery (Desktop)
Rechargeable lithium ion battery (Handheld)
1 1303740 USB cable
Page 11
Unpacking and Parts Identification
3
Item Qty
Part Number Description
1 801652 Analog/alarm
output cable (Desktop models only)
1 Operation and
Service Manual
1 Calibration
Certificate
1 801688 Conductive
Tubing
1 801668 Filter removal tool
(Spanner Driver)
4
2
1
801673
Spare Internal Filter Elements Desktop Model Only
37-mm filter includes: Filter body top Filter body bottom Mesh screen
Comes with
Page 12
Chapter 1
4
Item Qty
Part Number Description
37-mm cartridge opening tool
8 801666 Spare Internal
Filters Handheld Model
Only
1 801667 Impactor Kit
PM
2.5
assembled Top Bottom Impaction Plate PM
1.0
Top PM
4.0
Top PM
10
Top Extra Impaction Plate
1 801691 Dorr-Oliver
Cyclone
1 801692
801694
Power Supply – Desktop
Power Supply – Handheld
2 Stylus
When shipped, one stylus will be in the accessory bag, the second stylus attached to instrument.
Page 13
Unpacking and Parts Identification
5
Item Qty
Part Number Description
1 3012094 Screwdriver, dual
ended. (For Handheld Models only)
1 801674 Impactor Oil
2 801698 Inlet cap
When shipped, one inlet will be in the accessory bag, the second inlet attached to instrument.
Page 14
Chapter 1
6
Parts Identification for the DUSTTRAK II Desktop Aerosol Monitor Models 8530/8531
Figure 1-1: Features on Desktop Model
Filter Access
Zero Module
Connector
Battery Access
Touchscreen
On/Off
Inlet
Stylus
II
Power
Analog/Alarm Output
USB Host
USB Device
Ethernet
Page 15
Unpacking and Parts Identification
7
Parts Identification for the DUSTTRAK II Handheld Aerosol Monitor Model 8532
Figure 1-2: Features on Handheld Model
Filter Access
Battery Access
(Screw Lockdown)
Power
USB Device
USB Host
On/Off
Inlet
Stylus
Touchscreen
Port Cover
Page 16
Chapter 1
8
(page intentionally left blank)
Page 17
9
Chapter 2
Setting Up
Supplying Power to the DUSTTRAK II Aerosol Monitor
The DUSTTRAK II Aerosol Monitor must be powered by either batteries or use of the external AC adapter.
WARNING
The instrument has been design to be used with batteries supplied by TSI. Do not use a substitute.
WARNING
Do not use non-rechargeable batteries in this instrument. Fire, explosions, or other hazards may result.
Installing the Batteries in Model 8530/8531 Desktop
Remove the battery cover and slide one or two batteries into the battery slots. A single battery can be put into either slot. Orient the batteries with the label side facing up (see figure 2-1).
Figure 2-1: Batteries into Desktop Unit
Installing the Batteries in Model 8532 Handheld
Remove the battery cover by loosening captured screw on the bottom of the unit. Orient battery with brass connectors facing forward. Insert battery into cavity and slide forward to engage into pins. Replace the battery cover and secure by tightening screw (see figure 2-2).
Page 18
Chapter 2
10
Figure 2–2: Batteries into Handheld Unit
Using the AC Adapter to Run Instrument
The AC adapter allows you to power the D
USTTRAK monitor from an AC
wall outlet. When using th e AC adapter, the batteries (if installed) will be bypassed.
Battery Charging
This instrument will charge the Lithium Ion battery packs. Insert the batteries into the battery compartment, plug the instrument into AC power, and turn the instrument on. Batteries will charge only when the instrument is on and in stand-by mode. Batteries will not be charged if the instrument is turned off or is actively taken measurements. Charging will stop when the batteries have been fully charged.
Inlet Cap
When using the D
USTTRAK monitor to sample environmental air, the inlet
cap should be put over the instrument. This cap will keep large objects from dropping into and plugging the inlet. The cap will also keep direct light fro m shinning into the chamber and skewing the results.
The inlet cap can simply be pressed onto the instruments inlet.
Page 19
Setting Up
11
Figure 2-3: Putting on Inlet Cap
Size-Selective Impactors
Size-selective impactors can be attached to the inlet of the D
USTTRAK II
instruments. Size-selective impactors can be used to pre-condition the size range of the particles entering the instrument. PM
1
, PM
2.5
, PM4 (Respirable)
and PM
10
impactors are available. The instrument must run at the
factory default setting of 3.0 L/min for the impactors to achieve the correct cut points.
The size-selective impactor is composed of three parts; the cap, impaction plate and bottom. Selection of the cap will determine cut size of the impactor. Each cap is labeled with the particle cut size (1 µm, 2.5 µm,
4.0 µm or 10 µm). The same impaction plate and bottom are used on all impactor sizes.
Figure 2-4: Size-Selective Impactor
The impactor assembly is attached to the instrument in place of the inlet cap. The inlet cap does not need to be used if an impactor is being used. See
Chapter 4, "Maintenance,"
for instructions on how to add oil to the
impaction plate.
Impaction Plate
Cap (1 µm,
2.5 µm, 4 µm or 10 µm)
Bottom
Page 20
Chapter 2
12
Dorr-Oliver Cyclone
A Dorr-Oliver cyclone is shipped with the instrument. The Dorr-Oliver cyclone removes particles over 4.0 µm in size. The Dorr-Oliver cyclone is attached to the instrument by sliding the cyclone clip over the protruding catch. The tube from the Dorr-Oliver cyclone needs to be routed to the inlet of the instrument.
Figure 2-5: Installing Door-Oliver Cyclone
Inlet attachments (impactors or inlet cap) should not be used when using the Dorr-Oliver Cyclone. The instrument flow rate must be changed to
1.7 L/min when using the Dorr-Oliver Cyclone in order to achieve a 4 µm (respirable) cut-point. See the Flow Cal
instructions in the Operations chapter for instructions on how to change the instruments flow rate.
Instrument Setup
The DUSTTRAK II monitor can be connected to a computer to download data and upload sampling programs.
Connecting to the Computer
Connect the USB host port of a Microsoft Windows®-based computer to the USB device port on the side of the D
USTTRAK monitor.
®
Windows is a registered trademark of Microsoft Corporation.
Page 21
Setting Up
13
Installing TRAKPROTM Data Analysis Software
T
RAKPRO software can preprogram the DUSTTRAK monitor, download
data, view and create raw data and statistical repo rts, create graphs, and combine graphs with data from other TSI instruments that use T
RAKPRO
software. The following sections describe how to install the software and set up the computer.
Note
To use TRAKPRO™ software with the DUSTTRAK™ Aerosol Monitor, the PC must be running Microsof t Windows
®
and the computer must have an available Universal Serial Bus (USB) port.
1. Insert the TRAKPRO Data Analysis Sof tw a re CD into the CD-ROM
drive. The install screen starts automatically.
Note
If the software does not start automatically after a few minutes, manually run the program listed on the label of the CD using the Run command on the Windows Start Menu.
2. Follow the directions to install TRAKPRO software. T
RAKPRO software contains a comprehensive installation guide. It is
recommended to print out this prior to starting the T
RAKPRO software
installation on your computer, so it may be consulted during the installation. The T
RAKPRO Software manual is located in the “Help” file in
T
RAKPRO software. There is no separately printed TRAKPRO Data
Analysis software manual.
®
Microsoft and Windows are registered trademarks of Microsoft Corporation.
Page 22
Chapter 2
14
Connecting Analog/Alarm Output
The Analog/Alarm Output Cable plugs into the alarm connection on the side of the instrument. This feature is on the desktop models (8530/8531) only.
The cable contains a 4-pin, mini-DIN connector. The pin-outs for the connector and the wiring for the cable are shown below.
4-pin miniDIN connector
Cable Wiring Diagram
Brown Wire Analog Ground Orange Wire Analog Out Red Wire Alarm (+) White Wire Alarm (-) Black Wire Shield
Figure 2-6: Cable Wiring Diagram
Wiring the Analog Output
System specifications:
Output voltage: 0 to 5 VDC.
Output impedance: 0.01 ohm.
Maximum output current: 15 mA.
Correct polarity must be observed (see pin-outs above).
The output cable supplied by TSI (part no. 801652) is labeled with the pin­out wiring diagram. Additional equipment may be needed for making connections to the system that TSI does not supply. It is the user’s responsibility to specify and supply all additional equipment.
Alarm Positive (+)
Alarm Ground (–)
Analog Output (+)
Analog Ground (–
)
Page 23
Setting Up
15
Wiring the Alarm
System specifications:
Maximum voltage: 15 VDC (DO NOT USE AC POWER)
Maximum current: 1 Amp
Correct polarity must be observed (see pin-outs above)
The alarm switch, located inside the D
USTTRAK monitor must be
located on the ground side of the alarm system.
!
WARNING
The DUSTTRAK monitor Alarm Output function should not be
used to detect hazardous conditions or to provi de an alarm for protecting human life, health or safety.
!
Caution
The alarm switch must not be wired to AC power! Failure to
properly install the user alarm could damage the DUSTTRAK
instrument and/or void the instrument warranty! Please read and
follow all instructions before wiring or operating the user alarm.
Page 24
Chapter 2
16
(page intentionally left blank)
Page 25
17
Chapter 3
Operation
Getting Started
The START UP screen is displayed initially when the instrument is turned on, following the initial TSI logo splash screen.
Using a stylus or finger tip, touch the “buttons” on the screen to activate different menus.
Page 26
Chapter 3
18
Setup Menu
Pressing Setup activates the Setup Menu touchscreen buttons along the left edge of the screen. Setup cannot be accessed when the instrument is sampling.
The main screen of the Setup screen displays the following information:
Serial Number
The instruments serial number.
Model Number
The instruments model number.
Firmware Versio n
Instruments current version of firmware.
Calibration Date
Date of the last factory calibration.
Pump Run Time
Pump running time in hours.
Cum Mass Conc
Amount of mass run through instrument over life.
Cum Filter Conc
Amount of mass run through instrument since last filter change.
Filter Time
Date of last filter change.
Page 27
Operation
19
Zero Cal
Zero Cal should be run the first time the instrument is used and should be repeated prior to every use. Zero Cal requires that the zero f ilter be attached prior to running.
Never perform a zero cal without attaching a zero filter.
1. Press Zero Cal Button
2. Attach Zero Filter
3. Press the Start button to start Zeroing process.
4. A count-down clock will appear indicating the time remaining. The
screen with indicate “Zero Cal Complete” when done.
Remove filter after zeroing has been completed. The instrument is now zero calibrated and ready for use.
Page 28
Chapter 3
20
Flow Cal
Flow Cal is run if the user wants to change the flow set point. The flow set point is factory set to 3 L/min total flow. 2 L/min of the total flow is measured aerosol flow. 1 L/min of total flow is split off, filtered and used for sheath flow. There is an internal P flowmeter in the D
USTTRAK II
instrument that controls flow rate to ±5% if factory setpoint. It is recommended to check the flows with an external flow reference meter, especially when collecting data. The pump will automatically start when entering the Flow Cal screen.
1. Attach a flow calibrator (reference flow meter) to inlet port. You may
use a bubble buret, mass flow meter, dry piston or rotameter as flow measurement devices.
2. Move the arrows up or down to achieve desired flow on the reference
flowmeter. Each up or down arrow will change the flow about 1%. Allow time between button presses to let pump change to the new flow rate.
3. Select save once the desired flow rate is achieved. Select Undo to return
to the factory set point.
Page 29
Operation
21
User Cal
User Cal allows the user to store and use 10 different calibration factors. The currently active user calibration is highlighted with an asterisk “*”.
Four variables can be set for each user calibration.
Name
User can rename calibration to a description name.
Photometric
Changes the factory calibration of particle signal, based on Arizona Road Dust, to actual aerosol being measured. See below for sets to set this calibration.
Page 30
Chapter 3
22
Size Corr
Changes the factory calibration of the particle distribution, based on Arizona Road Dust, to actual aerosol being measured. See below for sets to set this calibration.
User Cal [on,off]
Selecting On will activate current user calibration and deactivate the previously selected user calibration.
Photometric Calibration Factor
In most situations, the D
USTTRAK monitor with its built-in data logging
capability can provide very good information on how the concentration of an aerosol changes for different pro ces ses o ver time. Factory calibration to the respirable fraction of standard ISO 12103-1, A1 test dust is fairly representative of a wide variety of ambient aerosols. Because o ptical mass measurements are dependent upon particle size and material properties, there may be times in which a custom calibration would improve your accuracy for a specific aerosol.
Determining a aerosol specific photometric calibration requires that you determine a true mass concentration (e.g., gravimetric analysis) for the aerosol you want to measure. The true mass concentration is used to calculate the custom calibration factor for that aerosol. Once you have a custom calibration factor, you can reuse it each time you make measurements in the same aerosol environment.
Determining the Calibration Factor for a Specific Aerosol
The D
USTTRAK II monitor is factory calibrated to the resp irable fraction of
standard ISO 12103-1, A1 test dust. The D
USTTRAK monitor can be easily
calibrated to any arbitrary aerosol by adjusting the custom calibration factor. The D
USTTRAK monitor’s custom calibration factor is assigned the value
of 1.00 for the factory calibration to standard ISO test dust. This procedure describes how to determine the calibration factor fo r a specific aerosol. Using the value of 1.00 will always revert back to the factory calibration.
To determine a new calibration factor you need some way of accurately measuring the concentration of aerosol, hereafter ref er red to as the reference instrument. A gravimetric analysis is often the be s t choice, though it is limited to nonvolatile aerosols. The internal 37 mm filter cartridge, in the desktop units, can be used to collect the reference gr avimetric reference sample.
Page 31
Operation
23
To make an accurate calibration you must simultaneously measure the aerosol concentration with the D
USTTRAK monitor and your reference
instrument.
1. Zero the D
USTTRAK II monitor.
2. Put the instrument in Manual Log (Manual Logging is reviewed later in
this section).
3. Set the logging interval. One minute (i.e., “01:00”) is often a good
choice.
4. Co-locate the D
USTTRAK II monitor and the reference sampler
together so that they are measuring from the same area. The 37mm filter cartridge in the desktop unit can be used to collect the particles to be weighed for the gravimetric reference.
5. Start sampling aerosol with both instruments at the same time.
Note
Greater accuracy will be obtained with longer samples. The time you permit for sampling often depends on the reference instrument and characteristics of the measured aerosol. It may take some time to collect suffic ient aerosol onto a filter cassette for accurate gravimetric analysis. Refer to instructions of your reference instrument for sampling tim es.
6. Stop sampling w ith both instruments at the same time.
7. Record the D
USTTRAK monitor average concentration. This can be
done by viewing the sample average in the Data screen. (Data Screen is reviewed later in this chapter)
8. Determine the mass concentration in mg/m
3
from your reference instrument. For gravimetric sampling this means having the gravimetric sample weighed.
Note
If you used the internal gravimetric filter in the DustTrak™ Model 8530 or 8531, the flow rate used to compute the concentration should be 2 L/min, not 3 L/min since only 2 L/min of aerosol flow reaches the filter.
9. Compute the new calibration constant, NewCal, using the following
formula:
CurrentCal
ionConcentratDustTrak
ionConcentrat Reference
NewCal
 
 
=
10. Select Photometric from the User Cal drop down selection and enter
the NewCal factor using the onscreen controls.
Page 32
Chapter 3
24
Alarm
Alarm allows the user to set an alarm level that will be triggered if the instrument’s reading goes above the setpoint.
Alarm1 Setpoint [mg/m3]
The alarm1 setpoint is the mass concentration level upon which the alarm1 is triggered.
Alarm will be triggered if the mass concentration, taken at the logging interval, rises above the setpoint.
Note: Alarm 2 must be lower than Alarm 1 when both alarms are enabled.
Page 33
Operation
25
Alarm1 Relay [On, Off]
When the relay alarm is turned on, unit will close relay switch when Alarm1 level is surpassed.
Relay selection is available on the 8530 and 8531 desktop models only.
Alarm1 STEL [On, Off]
When the STEL alarm is turned on, STEL data will be collected when Alarm1 level is surpassed.
STEL selection is available on the 8530 and 8531 desktop models only.
See STEL Note below.
Alarm2 Setpoint [mg/m3]
The alarm2 setpoint is the mass concentration level upon which the alarm2 is triggered.
Alarm will be triggered if the mass concentration, taken at the logging interval, rises above the setpoint.
Note: Alarm 2 must be lower than Alarm 1 when both alarms are enabled
Alarm2 Enable [On, Off]
Enables Alarm2 to be logged and will activate the Audible or Visible alarms if they are enabled.
Alarm Audible [On, Off]
When the audible alarm is turned on, the instrument will activate internal beeper when Alarm1 or Alarm2 level is surpassed.
Alarm1 Visible [On, Off]
When the visible alarm is turned on, unit will show the alarm icon (Alarm1
,
Alarm 2
) in title bar when Alarm1 or
Alarm2 level is surpassed.
STEL Alarm
STEL stands for Short Term Exposure Limit. When a STEL alarm is selected, the instrument will inspect the data on a second by second basis, independent from the selected logging interval. If the mass exceeds the STEL limit, then a STEL even will be triggered and the following actions will be taken.
Page 34
Chapter 3
26
STEL indicator
The STEL indicator
will show Red on the main screen.
Data
Data will be taken a 1 minute logging interval for 15 minutes.
This data will be stored in a separate f ile named STEL_XXX, where XXX will be matched to the logged data file.
The instrument will also continue to log the mass concentration data at the logging interval selected.
STEL Alarm repeat
If the instrument remains over the STEL limit after the 15 minute interval, or if the instrument exceeds the STEL limit later during the sample period, additional STEL files will be generated.
Analog
Analog setup screen sets the parameters that will drive the analog out port. Applies to the 8530/8531 Desktop models only.
Analog out [On, Off]
Turns analog out port on.
Size Fraction
Selects the size channel that will drive the analog out.
Output Setting [V, mA]
Select between 0 to 5 V and 4 to 20 mA.
Page 35
Operation
27
Lower Limit [mg/m3]
Mass concentration reading of the selected channel that will correspond to 0 V or 4 mA.
Upper Limit [mg/m3]
Mass concentration reading of the selected channel that will correspond to 5 V or 20 mA.
Settings
Settings screen sets basic unit parameters.
Date Time
Sets current date, current time and date/time format. Time can be set in 12 or 24 hour format. Date can be set in yyyy/dd/mm, yyyy/mm/dd or yyyy/dd/mm.
Page 36
Chapter 3
28
IP
USB PORT IP Address: USB IP is the address assigned to the instrument
by the NDIS driver. It is shown but cannot be changed.
Ethernet Port IP parameters: (Model 8530, 8531 Desktop only.) IP method can be set to static or dynamic. For static IP, IP address, default gateway, and
subnet mask can be set. For Dynamic, The IP assigned by the network is
shown. This cannot be changed. See Note below.
Display
Switches between blue and while backgrounds.
Touch Cal
Calibrates the touch cal screen.
IP Notes
After changing the instrument to Dynamic or Static, the instrument must be rebooted.
In Dynamic Mode, the unit will show the IP to which is assigned (after being rebooted).
Page 37
Operation
29
Run Mode
The RunMode tab brings up sampling mode options.
Sampling mode options include Survey Mode, Manual Log, and Log Mode 1-5.
Survey
Survey Mode runs a real time, continuous active sample, but does not log data.
Manual
Manual Log sets the instrument to log data for a specified run time.
Log Modes
Log Mode starts and stops the instrument at specified times, run for a specified test length, and perform multiple tests of the same length with a specified time period between tests.
Page 38
Chapter 3
30
Survey Mode
Time Constant
Time Constant can be set from 1 to 60 seconds. This will control the update rate of the main screen. It is the rolling average of data displayed on the main screen and is not linked to logged data in either Manual or Program Log modes.
Auto Start on Power Up
When set to “Yes”, unit will start a measurement upon being powered on, if the unit was set to “Survey” when it was turned off.
When set to “No”, the un it w ill be in idle when it is powered on.
Page 39
Operation
31
Manual Mode
Log Interval
The log interval can be set from 1 second to 60 minutes. It is the amount of time between logged data points.
Test Length
Test length can be set from 1 minute to the limit of the data storage.
Time Constant
Time Constant can be set from 1 to 60 seconds. This will control the update rate of the main screen. It is the rolling average of data displayed on the main screen and is not linked to logged data in either Manual or Program Log modes.
In Manual mode, data will be stored to a file named “Manual_XYZ” where XYZ is an incrementing integer.
Page 40
Chapter 3
32
Log Mode (1–5)
Log Name
Log Name, brings up a virtual keypad to name the Logged Data file.
Start Date
Start Date, select the date the test will start.
Start Time
Start Time, select the time the test will start.
Log Interval
The log interval can be set from 1 second to 60 minutes. It is the amount of time between logged data points.
Auto Zero Interval
Interval between re-zeroing the instrument using the Auto-Zero accessory. Models 8530 and 8531 desktop only.
Test Length
From 1 minute to the limit of the data storage.
Number of Tests
Number of tests, 1 to 999.
Time between Tests
Time between tests, 1 minute to 30 days.
Time Constant
Time Constant can be set from 1 to 60 seconds. This will control the update rate of the main screen. It is the rolling average of data displayed on the main screen and is not linked to logged data in either Manual or Program Log modes.
Page 41
Operation
33
Use Start Date
Use Start Date, option to use programmed start date or by pass programmed start date.
Use Start Time
Use Start Time, option to use programmed start time or bypass programmed start time.
In Log mode, data will be stored to a file named “LogName_XYZ” where LogName is the user entered log name and XYZ is an incrementing integer.
Taking Mass Concentration Measurements
Measurements are started and controlled from the main screen.
Prior to starting a measurement the instrument should be zeroed from the
Setup screen and the run mode should be configured and selected from the RunMode screen.
When the instrument is on, but not taking any mass measurements the start button will be green and instruments pump will not be running. To start taking a measurement, press the green start button.
While taking a measurement the screen will display the current measured mass concentration. The various regions of the screen are shown below.
Page 42
Chapter 3
34
Screen Regions
Mass Reading
Shows the instruments mass measurements.
Run Mode Region
Shows the run mode selected from the RunMode screen.
File Name Region
Displays the file name to which the data is currently being saved.
Test Progress Region
Shows the time-based progress of the test.
Error Indicator Region
Shows the current stats of the instrument STEL: Shows if STEL is in progress
(desktop instruments only) Flow: Status of the flow control Laser: Status of the Laser Filter: Status of the Filter See Chapter 5, "Troubleshooting,"
to
resolve any of these error conditions.
Run Mode
File Name
Test Progress
Mass Reading
Error Indicators
Page 43
Operation
35
Stats
The Stats button will show the statistics of the mass measurement. When the Stats button is pressed, the main mass reading will reduce in font size, and the measurement statistics will show on the right side of the screen.
Graphing
During sampling, pressing the Graph button displays current readings in graphical form.
During Survey Mode, five (5) minutes of running real-time data is
displayed graphically.
During Logging Mode, the entire log test time is displayed on the graph.
Time Display
(live area)
Scale Display
(live area)
Data Region
(live area)
Page 44
Chapter 3
36
Time Display
Pressing the Time x-axis label on the graph screen switches between Time (s),
Time (abs), and Time (rel). Time (s): Elapsed time from first logged
point (log interval) to the last logged point (test length).
Time (rel): Relative time from zero to last logged point (test length – log interval).
Time (abs): Absolute time from first logged point (test start + log in terval) to last logged point (test stop).
Scale Display
Pressing in the Scale Display area will bring up a dialog that will allow changing between auto scaling and user scaling of the Y-axis.
Data Region
Pressing the data region will bring up a dialog to show TWA or Average lines.
TWA: Will show a secondary line on the graph showing the time weighted average of the data. This line will not show if test time is less then 15 minutes.
Average: Show a secondary line on the graph of the running average of the data.
In Graphing Mode, pressing Main returns the instrument to the Main Screen display.
Page 45
Operation
37
Viewing Data
The Data button opens a list of data files for viewing.
Select File
Press the arrows on the right side of the screen to scroll up or down to the data file to be viewed.
Data Statistics
Statistics on the selected file
o File Name o Sample Average o Sample TWA o Sample Maximum Reading o Sample Minimum Reading
o
Number of Data Points in the F ile
Save All Button
Downloads data to a USB thumb drive. USB thumb drive must be attached to the USB host port. Data is saved as a .csv file that can be viewed in Microsoft
®
Excel®
spreadsheet software.
Delete Button
Deletes the currently highlighted file.
Delete All Button
Deletes all the files stored on the instrument.
Graph Button
Data can also be viewed in graphical form by pressing the Graph button while the data file is highlighted.
Page 46
Chapter 3
38
Title Bar
The Title Bar shows common instrument information.
Current Screen
Title of the current screen that is being displayed.
Instrument Lock
Icon shows if the instrument touchscreen is in an unlocked or locked condition.
Unlocked: Locked: To lock the touchscreen controls, touch the
“lock” icon, immediately followed by three (3) quick touches on the current screen (Main) word along the top tool bar.
Repeat the process to unlock the screen.
Battery Status
Show the current % life of the battery and show if the battery is currently being charged:
Charging:
(unfilled portion of the
icon filled yellow) Not Charging:
(unfilled portion of
the icon transparent)
Date and Time
Indicates the instruments current date and time.
Alarm
If the instrument is in an alarm status, an alarm icon
will appear in the title bar.
Current Screen
Date
,
Time
Battery Status
Instrument Lock
Page 47
39
Chapter 4
Maintenance
The D
USTTRAK II aerosol monitor can be maintained in the field using the
instructions below. Additionally, TSI recommends that you return your D
USTTRAK II to the factory for annual calibration. For a reasonable fee,
we will quickly clean and calibrate the unit and return it to you in “as new” working condition, along with a Certificate of Calibration. This “annual checkup” helps ensure that the D
USTTRAK II is always in good operating
condition.
WARNING
There are no user-serviceable parts insi de this instrument. The instrument s houl d only be opened by TSI or a TSI approved service technician.
Maintenance Schedule
Your DUSTTRAK II Aerosol Monitor requires maintenance on a regular basis. Table 4–1 lists the factory recommended maintenance schedule.
Some maintenance items are required each time the D
USTTRAK monitor is
used or on an annual basis. Other items are scheduled according to how much aerosol is drawn through the instrument. For example, cleaning the inlet sample tube is recommended after 350 hours of sampling a 1 mg/m
3
concentration of aerosol. This recommendation should be pro-rated according to how the instrument is used. 350 hours at 1 mg/m
3
is the same
amount of aerosol as 700 hours at 0.5 mg/m
3
or 175 hours at 2 mg/m3, etc.
Table 4–1. Recommended Maintenance Schedule
Item Frequency
Perform zero check Before each use. Clean inlet 350 hr. at 1 mg/m3* Clean 2.5 µm calibration impactor Before every use. Replace internal filters 350 hr. at 1 mg/m3* or
when indicated by the main screen filter error indicator.
Return to factory for cleaning and calibration Annually
*Pro-rated, see discussion above.
Page 48
Chapter 4
40
The DUSTTRAK monitor keeps track of the accumulated amount of aerosol drawn through it since its last cleaning. When the internal filter replacement is due, the filter error indicator will turn from green to red.
TSI recommends that you perform a zero check prior to each use for the D
USTTRAK monitor and certainly before running any extended tests, and
after the instrument experiences a significant environmental change. Examples of significant environmental changes would be ambient temperature changes that exceed 15
°F (8 °C) or moving from locations with
high aerosol concentrations to low concentrations.
Zeroing Instrument
1. Attach the zero filter to the inlet of the instrument.
Figure 4-1: Attach Zero Filter to Inlet
2. Follow zero calibration instructions detailed in the operations section of
this manual.
Page 49
Maintenance
41
Cleaning the Inlet
The inlet should be cleaned based on the schedule in Table 4–1.
1. Turn the D
USTTRAK monitor off.
2. Unscrew the inlet nozzle from the instrument (Figure 4-2).
Figure 4-2: Unscrew Inlet Nozzle
3. Clean the inlet port. A cotton swab can be used to clean the outside of
the inlet port. The swabs can be dampened with water or a light solvent (e.g., isopropanol). The inside of the sample tube can be cleaned using a small brush, along with a light solvent. Dry the tube by blowing it out with compressed air, or let it air-dry thoroughly.
Note
Be careful not to blow particles into the DUSTTRAK monitor inlet port.
Figure 4-3: Do NOT Blow into Instrument
4. Screw (hand-tighten) inlet back into instrument.
Page 50
Chapter 4
42
Cleaning and Oiling Impactors
The calibration impactor should be cleaned prior to every use, using it to perform a Standard Calibration (size correction) on the instrument, as described in the Operations
section.
1. Unscrew Impactor. Check O-ring on the impactor base.
2. Clean outside and inside of Impactor and the impactor plate using a
clean brush and a light solvent. Dry impactor parts by blowing it out with compressed air, or let it air-dry thoroughly.
3. Apply 2 drops of oil (included) to the impactor plate. Do not over-fill
impaction plate.
Figure 4-4: Apply 2 Drops of Oil to Impactor Plate
4. Screw (hand-tighten) impactor back together.
Replacing the Internal Filters
The internal filters should be replaced based on the schedule in Table 4–1 or when the filter indicator on the main screen changes to red.
1. Turn the instrument off.
2. Remove old filters from the instrument.
Handheld Model
a. Use the enclosed filter removal tool (PN 801668) tool to unscrew
the two filter caps located on the bottom of the instrument.
b. Pull the old filters out of the two filter wells. If filter wells are
visibly dirty, blow out with compressed air.
Apply
2 Drops of Oil
Page 51
Maintenance
43
Figure 4-5: Pull Filters Out of Two Filter Wells (Handheld Model)
c. Put two (2) new filters into the filter wells and screw filter caps
back into place.
Note
Replacement filters were shipped with the new instrument. Additional filters can be ordered from TSI under PN 801666.
Desktop Model
Put two (2) new filters into the filter wells and screw filter caps
a. Open filter access door on the back of the instrument. b. Use the enclosed filter removal tool (PN 801668) to unscrew
filter cap.
c. Pull out single cylindrical filter from filter well. If filter well is
visibly dirty, blow out with compressed air.
Figure 4-6: Pull out Single Cylindrical Filter from Filter Well
(Desktop Model)
Page 52
Chapter 4
44
d. Put new filer back into filter we ll and screw filter cap back
into place.
e. Open blue retention clip by pinching ends inward and
pushing down.
Figure 4-7: Open Blue Retention Clip
f. Remove 37-mm filter cassette by pulling downward and outward.
Figure 4-8: Remove 37-mm Filter Cassette
Page 53
Maintenance
45
g. Open filter cassette using enclosed tool PN 7001303.
Figure 4-9: Open Filter using Enclosed Tool
h. Remove screen mesh from filter cassette and blow out using
compressed air. Blow in reverse direction to remove captured particulate.
i. Replace mesh in filter cassette and press halves together. Make sure
filter has been fully closed. T he filter tool PN 7001303 can be used to ensure the filter is fully closed.
Figure 4-10: Replace Mesh in Filter Holder
j. Place filter cassette back into position and close blue retaining clip.
Make sure retaining clip snaps back into place.
Notes
Replacement filters (HEPA and 3-mm Filter Cassette with mesh filter) were shipped with the new instrument. Additional filters can be ordered from TSI under PN 801673.
TSI does not supply any filter media for the filter cassette. Any commercially available 37-mm filter media may be used with the D
USTTRAK™ II or DRX desktop instruments to collect gravimetric
reference samples.
3. It is important to reset the instruments filter count er a fter replacing
filters. Resetting the counter will clear the filter error condition shown on the main screen. Reset the counters by the following:
a. Turn on the instrument. b. Press the Setup button to go into the setup screen.
Page 54
Chapter 4
46
c. Touch the Cum Filter Conc: (live key) to reset the aerosol mass.
d. Replace user serviceable filters? Dialog will appear. Press OK. e. Reset filter concentration? Dialog will appear. Press Yes to reset
the cumulative filter concentration to zero.
f. The Setup screen will not show zero for the Cum Filter
Concentration and the current date for the Filter Time.
Storage Precautions
When storing the DUSTTRAK monitor for more than 30 days, you should charge and remove the batteries. This prevents damage due to battery leakage.
This instrument must be stored in a location where the temperature remains between –20 and 60°C (–4 and 140°F).
Touch
Page 55
47
Chapter 5
Troubleshooting
The table below lists the symptoms, possible c auses, and recommended solutions for common problems encountered with the D
USTTRAK II
monitor.
Symptom Possible Cause Corrective Action
Erratic zero
reading.
Leak.
Dirty inlet port and/or sample tube.
Internal filter(s) not installed properly (leaking).
Check connections for leaks.
Replace zero filter.
Clean inlet port. Clean or replace tubing.
Inspect internal filter wells to make certain the filters and o­rings are seated properly. Replace internal filters if
necessary.
Run Mode
Error: The start time has passed
Run Mode Error: The selected log mode will exceed the allowed number of samples
The selected Run Mode
program has “Use Start Date” selected, but the start date is prior to the current date.
The selected Run Mode program is programmed to save more samples then is room in memory.
Correct or change the run
mode program
Reduce the number of samples by reducing the test length or increasing the logging interval.
Instrument runs slow
Large amount of data in memory
Large data files or many small data files will cause instrument to slow, due to need to read and display large amounts of
data.
Page 56
Chapter 5
48
Symptom Possible Cause Corrective Action
No display.
Unit not switched on.
Low or dead batteries.
Switch unit on.
Recharge the batteries or plug
in the AC adapter.
No touch ­screen response.
Instrument currently busy
Instrument Touchscreen is locked
The instrument will take time to open large data files and save configuration information. During this time, the instrument will not respond to additional touchscreen touches.
If the lock in the title bar is red, unlock the instrument following the instructions in the Chapter 3, Operation : Title
Bar section of this manual.
Analog
output does not work
Cable/connector not
correctly installed.
Output wired with reverse polarity.
Make sure cable connector is
fully seated.
Make sure analog out (+) and analog ground (-) are wired
correctly to data-logger.
Analog
output is not in proportion to display
Analog output range in D
USTTRAK monitor
may be set incorrectly.
Data logger scaling factor
may be set incorrectly.
Check analog output setting in the Setup->Analog
screen. Make sure the channel of interest is selected. Make sure that the correct output (0 to 5V, 4 to 20 mA) is selected.
Review the scaling factor set in
the Setup-Analog screen.
Alarm
output does not work.
Alarm does not turn on correctly.
Alarm function not turned
on.
Alarm setting incorrect.
Alarm output wired with reverse polarity.
Turn the alarm function on in
the Settings->Alarm screen.
Check the alarm settings in the
Settings->Alarm
screen.
Alarm wires are polarized. Voltage input must be wired to
alarm input (+).
Page 57
Maintenance
49
Symptom Possible Cause Corrective Action
Instrument
does not store new data
Memory is full.
Instrument is in Survey mode.
Delete or transfer historic data.
The instrument does not store data in survey mode. Can to
manual or program log mode.
Flow Error
is indicated on front screen
If sampling from a duct,
instrument may have problems overcoming pressure differences.
Flow obstruction.
Internal pump failing, indicated by inability to adjust flow rate to full range.
Filter Cassette clogged or has mass loading.
Attach both the input and the
exhaust port into the duct.
Remove obstruction if still present. Press any key to bypass.
Factory service may be required.
Replace the filter cassette. See the maintenance section of the
manual.
Laser Error
indicated on front screen
Laser background is too
high.
Laser is failing
Remove and clean inlet nozzle.
Pay close attention to the tip of the nozzle that is inserted into the instrument to ensure it is clear of any contamination.
Factory service may be
required.
Filter Error
indicated on front screen.
Filters need to be replaced
Replaced the filters per
instructions in the maintenance section of this manual. Make sure to reset the filter mass and date once the filters have been
changed.
Page 58
Chapter 5
50
(page intentionally left blank)
Page 59
51
Appendix A
Specifications
Specifications are subject to change without notice.
Sensor Type 90° light scattering Range 8530 Desktop 0.001 to 150 mg/m3
8531 Desktop HC 0.001 to 400 mg/m
3
8532 Handheld 0.001 to 150 mg/m
3
Resolution ±0.1% of reading of 0.001 mg/m3, whichever is
greater Zero Stability ±0.002 mg/m3 24 hours at 10 sec time constant Particle Size Range Approximately 0.1 to 10 µm Flow Rate 3.0 L/min set at factory
1.4 to 3.0 L/min adjustable
Flow Accuracy ±5% factory setpoint
Internal flow controlled Temperature Coefficient +0.001 mg/m3 per °C Operational Temp 0 to 50°C Storage Temp -20 to 60°C Operational Humidity 0-95% RH, non-condensing Time Constant Adjustable 1 to 60 sec onds Data Logging 45 days at 1 minute samples Log Interval 1 second to 1 hour Physical Size (HWD) Handheld: 4.9 x 4.75 x 12.45 in.
Desktop: 5.3 x 8.5 x 8.8 in. Weight Handheld: 2.9 lb, 3.3 lb with battery
Desktop: 3.45 lb, 4.45 lb – 1 battery,
5.45 lb – 2 batteries
Communications 8530/31: USB (Host and Device) and Ethernet.
Stored data accessible using thumb drive
8532: USB (Host and Device). Stored dada
accessible using thumb drive.
Power—AC AC power adapter included.
115 to 240 VAC
Page 60
Appendix A
52
Battery 8530/31: Up to 2 Removable Li-Ion
External and Internal charging Life, 1 battery: >6.5 hours (9 hours
typical for a new battery) Life, 2 battery: >13 hours 8532: 1 Removable Li-Ion External and Internal charging Life: 6 hours typical
Analog out 8530/31:
User selectable output 0 to 5 V or 4 to 20 mA User selectable scaling
Alarm Out 8530/31:
Relay or sound buzzer Relay No latching MOSFET User selectable set point 5% deadband Connector 4-pin, Mini-DIN connectors
8532: Sound buzzer
Screen 8530/31: 5.7” color touchscreen
8532: 3.5” color touchscreen
Gravimetric Sampling 8530/31: Removable 37-mm Cartridge EMI/RF Immunity: Complies with Emissions Directive Standard:
EN50081-1:1992 Complies with Immunity Directive Standard:
EN50082-1:1992*
*ESD Shock may require instrument reboot
Page 61
53
Appendix B
Zero Module
The Zero Module (PN 801690) allows for automatic re-zeroing of the D
USTTRAK Instrument during long sampling runs. The Zero Module
works only with the 8530 and 8531 desktop models.
The AutoZero module is attached to the main instrument in two steps. The first step is to place the Zero module over the instrument’s inlet and press down. The Zero module has an O-ring seal that will engage with the instrument’s inlet.
Figure B-1: Place Zero Module Over Inlet and Press Down
The second step is attaching the cable from the Zero module to the Zero module connector located on the back of the instrument.
Figure B-2: Zero Module Connector
Page 62
Appendix B
54
The Zero Module can only be used in a program log mode. The Zero module function is controlled through these two program mode options:
Auto Zero Interval
Interval between re-zeroing the instrument using the Auto-Zero accessory.
Use Auto Zero
Select Yes to use the Zero Module. Select No to not use the Zero Module.
Important points on Zero Module operation:
The Zero module will take one (1) minute to take a zero reading. The
first 45 seconds of that period is used to clear the chamber of particles. Readings from last 15 second of the period, when the chamber is cleared of particles, will be averaged to determine the Zero offset.
The log interval, when the Zero module is activated, must be two (2)
minutes or greater. Data will not be recorded to the log file when the Zero module is activated.
Page 63
55
Index
4
4-pin miniDIN connector, 14
A
AC adapter, 10 accessories
8530-NA, 1 advisory labels, iv aerosol monitor
maintenance, 39
specifications, 51
troubleshooting, 47 alarm, 24, 38
audible, 25
wiring, 15 alarm1
relay, 25
STEL, 25
visible, 25 alarm1 setpoint, 24 alarm2 enable, 25 alarm2 setpoint, 25 analog, 26
lower limit, 27
out, 26
output setting, 26
size fraction, 26
upper limit, 27 analog cable, 3 analog output
wiring, 14 analog/alarm output
connecting, 14 audible, 25 auto start on power up, 30 auto zero interval, 54
B
battery
charging, 10
battery installation, 9
desktop unit, 9 handheld unit, 10
battery status, 38
C
calibration certificate, 3 calibration date, 18 calibration factor
for specific aerosol, 22
calibration impactor
cleaning and oiling, 42 charging battery, 10 Class I, iii cleaning
inlet conditioner, 41 conductive tubing, 3 connecting
analog/alarm output, 14 connecting computer, 12 current screen, 38
D
data, 37
delete, 37
delete all, 37
save all button, 37
select file, 37
statistics, 37 delete, 37 delete all, 37 determining calibration factor for
specific aerosol, 22 Dorr-Oliver cyclone, 4, 12
E
error indicator region, 34 Ethernet port IP parameters, 28
Page 64
Appendix B
56
F
file name region, 34 filter cassette, 44, 45
troubleshooting, 49 filter opening tool, 3 filter removal tool, 3 filter time, 18 filter tool, 45 firmware version, 18 flow cal, 20
G–H
graph, 35
button, 37
data region, 36
scale display, 36
time display, 36
I–J–K
impactor oil, 5 inlet cap, 5, 10 inlet conditioner
cleaning, 41 installing batteries, 9 instrument lock, 38 instrument setup, 12 internal filter, 4
replacing, 42 internal filter element, 3 internal gravimetric filter, 23
L
laser radiation label, iv log interval, 31, 32 log mode, 29, 32
auto zero interval, 32
log interval, 32
log name, 32
number of tests, 32
start date, 32
start time, 32
test length, 32
time between tests, 32
log mode (continued)
time constant, 32 use start date, 33
use start time, 33 log modes, 29 log name, 32
M
maintenance, 39
schedule, 39 manual log, 29 manual mode, 31
log interval, 31
test length, 31
time constant, 31 mass concentration
measurements, 33 mass reading, 34 model number, 18
N–O
name, 21 newcal, 23 number of tests, 32
P–Q
packing list, 1 parts identification, 6, 7 photometric, 21, 23 photometric calibration factor, 22 power supply, 4 pump run time, 18
R
relay, 25 replacing internal filter, 42 reusing and recycling, v run mode region, 34 runmode, 29
S
safety information, iii
Page 65
Maintenance
57
save all button, 37 screen regions, 34 screwdriver, 5 select file, 37 serial number, 18 serial number label, iv service policy, 2 setting up, 9
instrument, 12
settings, 27
date time, 27 display, 28 IP, 28 touch cal, 28
setup menu, 18
calibration date, 18 cum filter conc, 18 cum mass conc, 18 filter time, 18 firmware version, 18 model number, 18 pump run time, 18
serial number, 18 size corr, 22 size fraction, 26 size-selective impactors, 11 software
installation, 13 spanner driver, 3 specifications, 51 start date, 32 start time, 32 start up, 17 START UP screen, 17 stats, 35 STEL, 25
alarm repeat, 26
data, 26
indicator, 26 storage precautions, 46 stylus, 4 supplying power, 9 survey mode, 29, 30
auto start on power up, 30
time constant, 30
T
test length, 31 test progress region, 34 time between tests, 32 time constant, 30, 31, 32 title bar, 38
alarm, 38 battery status, 38 current screen, 38 date and time, 38 instrument lock, 38
T
RAKPRO™ software
installation, 13
troubleshooting, 47
U
unpacking, 1 USB cable, 2 USB port
connector, 12 USB port IP address, 28 use auto zero, 54 use start date, 33 use start time, 33 user cal, 21, 22
name, 21
photometric, 21
size corr, 22
V
visible, 25
W–X–Y
warranty, 2 wiring alarm, 15 wiring analog output, 14
Z
zero cal, 19 zero filter, 2, 40 zero module, 53 zeroing instrument, 40
Page 66
58
(page intentionally left blank)
Page 67
Thank you for reading this data sheet.
For pricing or for further information, please contact us at our UK Office, using the details
below.
UK Office
Keison Products,
P.O. Box 2124, Chelmsford, Essex, CM1 3UP, England.
Tel: +44 (0)330 088 0560
Fax: +44 (0)1245 808399
Email: sales@keison.co.uk
Please note - Product designs and specifications are subject to change without notice. The user is responsible for determining the
suitability of this product.
Loading...