APPENDIX A SPECIFICATIONS..................................................... 11
1
iii
www..com
information@itm.com1.800.561.8187
Page 3
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components below. If
anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
www..com
1
information@itm.com1.800.561.8187
Page 4
Chapter 2
Setting-up
Supplying Power to the Model 5815 DP-CALC Micromanometer
The Model 5815 is powered with four size AA batteries.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on the
inside of the battery compartment. The Model 5815 is designed to
operate with either alkaline or NiMH rechargeable batteries, although it
will not recharge NiMH batteries. Battery life will be shorter if NiMH
batteries are used. Carbon-zinc batteries are not recommended because
of the danger of battery acid leakage.
www..com
3
information@itm.com1.800.561.8187
Page 5
Chapter 3
Operation
Keypad Functions
ON/OFF Key
ft/min / m/s Key
in H2O / Pa / hPa /
kPa / mm HG Key
0 Key
Changing Units
Press to turn the Model 5815 on and off. During
the power up sequence the display will show the
following: Model Number, Serial Number,
Software Revision and Last Date Calibrated.
Pressing this key changes the display to read air
velocity, converted from.
Pressing this key changes the display to read
differential pressure.
Pressing the 0 key zeroes the pressure
measurement.
To change units, first put the desired measurement
(air velocity or temperature) on the display. Then
press and hold the left, unlabelled key for five
seconds. Finally, use the
select the units.
ST and ENTER key to
www..com
5
information@itm.com1.800.561.8187
Page 6
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions
for common problems encountered with the Model 5815. If your symptom is
not listed, or if none of the solutions solves your problem, please contact
TSI.
Table 5-1: Troubleshooting the Model 5815
Symptom Possible Causes Corrective Action
No Display Unit not turned on Switch unit on.
Low or dead batteries Replace batteries.
Dirty battery contacts Clean the battery
The pressure sensor is protected from damage up to 7 psi (48 kPa or
360 mmHg). At higher pressure it can burst!
Fluctuating flow Reposition probe in
less-turbulent flow.
Fault in instrument Factory service required
on instrument.
WARNING!
www..com
9
information@itm.com1.800.561.8187
Page 7
Appendix A
Specifications
Specifications are subject to change without notice.
Static/Differential Pressure:
1
Range
: -15 to +15 in. H2O (-28.0 to +28.0 mm Hg, -3735 to
+3735 Pa)
Accuracy: ±1% of reading ±0.005 in. H
Resolution: 0.001 in. H
Velocity From a Pitot Tube:
2
Range
: 250 to 15500 ft/min (1.27 to 78.7 m/s)
3
Accuracy
: ±1.5% at 2000 ft/min (10.16 m/s)
O (1 Pa, 0.01 mm Hg)
2
Resolution: 1 ft/min (0.1 m/s)
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C)
Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Pollution degree 1 in accordance with IEC 664
Transient over voltage category II
O (±0.01 mm Hg, ±1 Pa)
2
External Meter Dimensions:
3.3 in. × 7.0 in. × 1.8 in. (8.4 cm × 17.8 cm × 4.4 cm)
Meter Weight:
Weight with batteries: 0.6 lbs (0.27 kg)
Power Requirements:
Four AA-size batteries (included)
1
Overpressure range = 7 psi (190 in. H2O, 360 mmHg, 48 kPa).
2
Pressure velocity measurements are not recommended below 1,000 ft/min (5 m/s) and are
best suited to velocities over 2000 ft/min. Range can vary depending on barometric
pressure.
3
Accuracy is a function of converting pressure to velocity. Conversion accuracy improves
when actual pressure values increase.
11
www..com
information@itm.com1.800.561.8187
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.