TSI Instruments AEROTRAK 9310, AEROTRAK 9510, AEROTRAK 9550, AEROTRAK 9350, AEROTRAK 9500 Operation Manual

AEROTRAK®
PORTABLE AIRBORNE
PARTICLE COUNTER
MODEL 9310/9350/9510/9550/9500
OPERATION MANUAL
P/N 6004217, REVISION S
MARCH 2018
SHIP TO/MAIL TO:
TSI Incorporated 500 Cardigan Road Shoreview, MN 55126-3996 USA
U.S. Technical Support:
(800) 874-2811/(651) 490-2811
Fax:
(651) 490-3824
E-mail address:
answers@tsi.com
Website:
http://www.tsi.com
INTERNATIONAL
Technical Support:
(001 651) 490-2811
Fax:
(001 651) 490-3824
AEROTRAK®
PORTABLE AIRBORNE
PARTICLE COUNTER
MODEL 9310/9350/9510/9550/9500
OPERATION MANUAL
P/N 6004217, Revision S
March 2018
Revision
Date
Revision
Date
A
July 2010
L January 2014
B
September 2010
M June 2014
C
November 2010
N January 2016
D
December 2010
P February 2016
E
February 2011
Q August 2016
F
May 2011
R January 2018
G
December 2011
S March 2018
H
November 2012
J October 2013
K October 2013
Manual History
The following is a manual history of the AeroTrak® Portable Airborne Particle Counter, Model 9310, 9510, 9350, 9550, 9500 Operation Manual (P/N 6004217).
ii
Part Number
6004217 / Revision S / March 2018
Copyright
©TSI Incorporated / 2010-2018 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
E-mail Address
answers@tsi.com
Limitation Of Warranty And Liability
(effective April 2014)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.) Seller warrants the goods, excluding software, sold hereunder, under normal use and service
as described in the operator's manual, to be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
Warranty
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of
shipment; b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals; c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment; d. Seller does not provide any warranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty
applies; e. This warranty does not cover calibration requirements, and seller warrants only that the
instrument or product is properly calibrated at the time of its manufacture. Instruments
returned for calibration are not covered by this warranty; f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the
manual allow an operator to replace consumables or perform recommended cleaning; g. This warranty is VOID if the product has been misused, neglected, subjected to accidental
or intentional damage, or is not properly installed, maintained, or cleaned according to the
requirements of the manual. Unless specifically authorized in a separate writing by Seller,
Seller makes no warranty with respect to, and shall have no liability in connection with,
goods which are incorporated into other products or equipment, or which are modified by
any person other than Seller. The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated
herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR
PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO SELLER’S BREACH
OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY
OR INDUCED INFRINGEMENTS. BUYER’S EXCLUSIVE REMEDY SHALL BE THE
RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON­INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR
iii
REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI’s Customer Service department at 1-800-874-2811 (USA) or +001 (651) 490-2811 (International).
Trademarks
TSI, the TSI logo, and AeroTrak are registered trademarks of TSI Incorporated. TrakPro is a trademark of TSI Incorporated. Microsoft, Windows, and Excel are registered trademarks of Microsoft Corporation. Modbus is a registered trademark of Modicon, Inc.
against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
iv AeroTrak® Portable Airborne Particle Counters
Contents
Manual History ............................................................................................ ii
Warranty ..................................................................................................... iii
Contents ...................................................................................................... v
Safety Information .................................................................................... vii
Laser Safety ............................................................................................ vii
Labels ...................................................................................................... viii
Description of Caution/Warning Symbols ................................................. ix
Caution .................................................................................................. ix
Warning ................................................................................................. ix
Caution or Warning Symbols ................................................................ ix
Getting Help ............................................................................................... x
CHAPTER 1 Introduction and Unpacking ............................................. 1-1
Unpacking the AeroTrak
Optional Accessories .......................................................................... 1-4
CHAPTER 2 Getting Started ................................................................... 2-1
Instrument Description ........................................................................... 2-2
Providing Power ..................................................................................... 2-4
To Install the Lithium-Ion Battery ........................................................ 2-5
To Use AC Power ............................................................................... 2-6
Using the Stylus...................................................................................... 2-6
Using the Integral Thermal Printer ......................................................... 2-7
Using Peripherals ................................................................................... 2-8
Connecting to the Alarm Closure ........................................................ 2-8
Filter Scanner ...................................................................................... 2-9
Probe ................................................................................................... 2-9
USB A ................................................................................................. 2-9
Alarm ................................................................................................. 2-10
Using Communications Ports ............................................................... 2-11
USB B ............................................................................................... 2-11
Ethernet Port ..................................................................................... 2-11
Using an Isokinetic Probe ..................................................................... 2-12
CHAPTER 3 Operation ............................................................................ 3-1
Screen Layout and Functionality ............................................................ 3-2
Software Input Panel (Keyboard or Keypad) ...................................... 3-2
Main Tab ............................................................................................. 3-2
Zoomed Data Screen ....................................................................... 3-6
Setup Tab ............................................................................................ 3-7
Zone Setup Screen .......................................................................... 3-8
Recipes Setup Screen ................................................................... 3-12
System Setup Screen .................................................................... 3-17
Change Power On Password Screen ............................................ 3-17
Change Setup Password Screen .................................................. 3-19
®
Airborne Particle Counter ............................. 1-2
v
Configuration Screen ..................................................................... 3-20
Print Setup Screen ......................................................................... 3-22
Print Schedule Screen ................................................................... 3-23
Clear Samples Screen ................................................................... 3-24
Purge Screen ................................................................................. 3-25
Information Screen ........................................................................ 3-26
Device Setup Screen ..................................................................... 3-26
Date and Time Screen ................................................................... 3-27
Display Screen ............................................................................... 3-28
Communications Screen ................................................................ 3-28
Regional Screen ............................................................................ 3-29
Environment Screen ...................................................................... 3-30
Data Tab ........................................................................................... 3-31
Export Data Screen ....................................................................... 3-33
Print Data ....................................................................................... 3-37
Reports Tab ...................................................................................... 3-40
CHAPTER 4 Data Handling ..................................................................... 4-1
USB Data Download .............................................................................. 4-1
USB Computer Communication ............................................................. 4-2
Installing Software ............................................................................... 4-2
Ethernet Communications ...................................................................... 4-2
CHAPTER 5 Maintenance ....................................................................... 5-1
Maintenance Schedule ........................................................................... 5-1
Zero Check ............................................................................................. 5-1
Cleaning the Instrument Enclosure ........................................................ 5-1
CHAPTER 6 Troubleshooting ................................................................. 6-1
CHAPTER 7 Contacting Customer Service ........................................... 7-1
Technical Contacts ................................................................................. 7-1
International Contacts ......................................................................... 7-1
Service ............................................................................................. 7-1
Technical Support ............................................................................ 7-2
Returning for Service .............................................................................. 7-3
APPENDIX A Specifications .................................................................. A-1
Compliance ............................................................................................ A-3
Dimensional Diagram ............................................................................ A-4
Index
vi AeroTrak® Portable Airborne Particle Counter
I M P O R T A N T
There are no user-serviceable parts inside the instrument. Refer all repair and maintenance to a qualified factory-authorized technician. All maintenance and repair information in this manual is included for use by a qualified factory-authorized technician.
Laser Safety
W A R N I N G
The use of controls, adjustments, or procedures other than those specified in this manual may result in exposure to hazardous optical radiation.
Safety Information
This section gives instructions to promote safe and proper handling of the AeroTrak® Portable Airborne Particle Counter.
This Portable Airborne Particle Counter is a Class I laser-based
instrument.
During normal operation, you will not be exposed to laser radiation.  Precaution should be taken to avoid exposure to hazardous radiation
in the form of intense, focused, visible light.
Exposure to this light may cause blindness. Take these precautions:
DO NOT remove any parts from the particle counter unless you are
specifically told to do so in this manual.
DO NOT remove the housing. There are no user-serviceable
components inside the housing.
vii
Labels
1. Serial number label (back panel)
2. Laser radiation label (internal)
3 Laser instrument
compliance label (back panel)
4. Calibration Label (back panel)
5. Laser radiation symbol label (back panel and internal)
6. European symbol for non­disposable item. Item must be recycled.
Advisory labels and identification labels are attached to the outside of the particle counter housing and to the optics housing on the inside of the instrument.
viii AeroTrak® Portable Airborne Particle Counter
Description of Caution/Warning Sym b o ls
C A U T I O N
Failure to follow the procedures prescribed in this manual might result in
irreparable equipment damage. Important information about the operation and maintenance of this instrument is included in this manual.
W A R N I N G
Warning means that unsafe use of the instrument could result in serious
injury to you or cause damage to the instrument. Follow the procedures prescribed.
Warns that uninsulated voltage within the instrument may have sufficient magnitude to cause electric shock. Therefore, it is dangerous to make contact with any part inside the instrument.
Warns that the instrument contains a laser and that important information about its safe operation and maintenance is included in the manual.
Warns that the instrument is susceptible to electro-static dissipation (ESD) and ESD protection procedures should be followed to avoid damage.
Indicates the connector is connected to earth ground and cabinet ground.
Appropriate caution/warning statements are used throughout the manual and on the instrument that require you to take cautionary measures when working with the instrument.
Caution
Warning
Caution or Warning Symbols
The following symbols may accompany cautions and warnings to indicate the nature and consequences of hazards:
Safety Information ix
Getting Help
To obtain assistance with this product or to submit suggestions, please contact Customer Service:
TSI Incorporated 500 Cardigan Road Shoreview, MN 55126 U.S.A. Fax: (651) 490-3824 (USA) Fax: 001 651 490 3824 (International) Telephone: 1-800-874-2811 (USA) or (651) 490-2811 International: 001 651 490 2811 E-mail Address: answers@tsi.com Web site: www.tsi.com
x AeroTrak® Portable Airborne Particle Counter
Model
Size Range
Flow Rate
9310-02
0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm
28.3 L/min (1 CFM )
9510-02
0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm
28.3 L/min (1 CFM )
9350-02
0.3, 0.5, 1.0, 2.0, 3.0, 5.0 μm
50 L/min (1.77 CFM)
9350-03
0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm
50 L/min (1.77 CFM)
9550-02
0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm
50 L/min (1.77 CFM)
9500-01
0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm
100 L/min (3.53 CFM)
C H A P T E R 1
Introduction and Unpacking
The AeroTrak® Portable Airborne Particle Counter (particle counter) has a touch-screen interface and operates on the included lithium-ion battery or AC power.
These devices have either a 1.0 CFM (28.3 L/min) flow rate, 50 L/min (1.77 CFM) flow rate or a 100 L/min (3.53 CFM) flow rate and count bin sizes from 0.3 to 25 µm depending on the model ordered (see table below). Up to 10,000 data sets can be downloaded for analysis and reporting using the TrakProLite Secure Data Downloading Software
included with the device. Each model is also available with an “N” suffix
to indicate a “No Printer” option.
Typical applications for these particle counters include cleanroom monitoring, research, exposure assessment, indoor air quality, filter testing, clearance testing, quality assurance, and contaminant migration studies. All AeroTrak
®
particle counters meet JIS standards.
(continued on next page)
1-1
Unpacking the AeroTrak
Qty.
Item Description
Part/Model
Reference Picture
1
AeroTrak Airborne Particle Counter
9310-02 9510-02 9350-02 9350-03 9550-02 9500-01
1
Power Supply 24 VDC 3.0A
(Power cord included is country dependent)
700230
1
Country-Specific Power Cord
700057 (US) 700058 (UK) 700059 (Euro)
1 or 2
Battery Pack (can install up to 2 ea.)
700028 Qty = 1 (28.3
and 50 L/min models)
Qty = 2 (100 L/min models)
3 m (10 ft)
Sample Tubing
3/8 ID x 1/2 OD (28.3 or 50 L/min models)
1/2 ID x 5/8 OD (100 L/min model)
1
Isokinetic Probe (Aluminum)
700068 (28.3 L/min)
700089 (50 L/min)
700092 (100 L/min)
Particle Count e r
Carefully unpack the AeroTrak® Airborne Particle Counter from the shipping container and verify that all the items shown in the photos below and listed in the following tables are present. Contact TSI immediately if items are missing or broken (see Chapter 7, Contacting Customer
Service for more information).
AeroTrak® Portable Airborne Particle Counter Parts List
®
Airborne
1-2 AeroTrak® Portable Airborne Particle Counter
Qty.
Item Description
Part/Model
Reference Picture
1
Computer Cable (2 m), USB A to B
700033 2
Stylus
N/A
1
HEPA Zero Filter Assembly
700119 (for the
28.3 and 50
L/min model) 700098 (for the
100 L/min model)
1
TrakPro Lite Secure Software CD for 21 CFR Part 11 compliant data downloading (includes manuals)
7001901
1
Operation Manual
6004217
Included on TRAKPRO Lite
Secure software CD
1
Quick Start Guide
6004218 1
Alarm Accessories
One Clamp on Ferrite and One 2-pin connector
1
Calibration certificate
N/A
Introduction and Unpacking 1-3
Optional Accessories
Item Description
Part/Model
Reference Picture
Stainless Steel Isokinetic Probe (used with tubing)
700069 (28.3 L/min)700026 (50 L/min)
700091 (100 L/min)
Basic Filter Scanning Probe
700070 (28.3 L/min)700071 (50 L/min)
700088 (100 L/min)
Electronic Filter Scanning Probe (with start/stop, LED, sound)
700094 (28.3 L/min)700095 (50 L/min)
700096 (100 L/min)
Tubing, Superthane 3/8-inch ID x 1/2-inch OD, Clear 100 ft
700062 (28.3 or 50 L/min models) Printer Paper (10 rolls)
700027
Dual Battery Charger Supplied w/ US Power Cord, Order international cord separately
700029
Velocity/Temperature/ Relative Humidity Probe
964 (Straight) 966 (Articulated)
The following photos and table list optional accessories. If you ordered optional accessories, make certain they have been received and are in working order.
AeroTrak® Portable Airborne Particle Counter Optional Accessories
1-4 AeroTrak® Portable Airborne Particle Counter
Item Description
Part/Model
Reference Picture
Velocity/Temperature Probe
960 (Straight) 962 (Articulated)
Temperature/RH Probe with Cable
700097
Carry Case
700086
Heavy Duty Carry Case (rolling case)
700087
Introduction and Unpacking 1-5
(This page intentionally left blank)
1-6 AeroTrak® Portable Airborne Particle Counter
C H A P T E R 2
Getting Started
This chapter describes the features, connections, and installation of the AeroTrak® Portable Airborne Particle Counter (particle counter). It includes:
Instrument Description Providing Power Using the Stylus Using the Integral Thermal Printer Using Peripherals Using Communications Ports Using an Isokinetic Probe
(continued on next page)
2-1
Instrument Description
Description
Function
Battery access
This plate provides access to a battery bay. The instrument can be run on up to two rechargeable batteries to allow portable operation for up to two to seven hours (model dependent) or with a single battery to allow switching outlets between rooms without powering the unit down.
Pump exhaust diffuser
Provides an outlet for the filtered exhaust.
DC Power connection
This power connection is for use with an external supply. Only the TSI-supplied 24 VDC 3.0 A power supply model 6011436 should be used.
Power Switch
Touch Screen
Printer
Sample Inlet
Isokinetic Probe
The AeroTrak® Portable Airborne Particle Counter has many features to make measurements convenient. The power switch is located on the front panel in the lower-left. A power LED indicates when the instrument is powered up. The main interface for the user is the color touch-screen interface on the front (see the note below on using a stylus with the screen). The sample inlet is located on the top of the instrument. The barbed isokinetic probe can be used with a short section of tubing above the inlet or with longer tubing and an adjustable tripod mount (not included) to monitor particles in hard to reach places or that are flowing horizontally. A large handle is also located on top to carry the instrument. On the left side of the instrument is a built-in printer.
The back of the instrument has many features that are described in the table below.
2-2 AeroTrak® Portable Airborne Particle Counter
Description
Function
Alarm
This connector provides two pins for a contact closure to control an external alarm. The contact closure is normally open and rated for 0 to 60 V AC/DC at 1.5A peak, 0.5A continuous. The “alarm out” connection is rated for 60 V insulation. The contact is closed under alarm conditions determined by the programming of the device.
Filter scanner
This connector is used with the optional TSI Electronic Filter Scanner (TSI P/N 700103).
Probe
This connector is used with a variety of TSI velocity / temperature / humidity probes.
USB A
This standard USB connector is provided for use with “flash drive” devices to download data from the instrument and
transfer to a computer or other device. The data files are in ASCII format for easy use in spreadsheet programs.
This port can also be used to also connect a keyboard.
USB B
The standard USB connector is used to connect the instrument to a computer running TSI TrakPro Lite software for data downloading and recipe uploading. It is used with a standard USB cable (included).
Ethernet port
The particle counter is compatible with either 10 or 100 MBps systems. The green LED indicates that the network is connected. The yellow LED indicates activity on the network cable.
Alarm
Ethernet Port
Pump Diffuser
Power Connection
Battery Access
Filter Scanner
Probe
USB A (Flash Drive)
USB B (to computer)
Exhaust
Getting Started 2-3
Providing Power
N O T E S
When using AC power, the battery (if installed) charges when the
instrument is on, while actively sampling (trickle charging), and when the power is put in standbycharge battery mode.
Removing/changing the lithium-ion battery or disconnecting AC
power does not cause loss of data. The AeroTrak® Airborne Particle Counter has an internal, non-user accessible battery to maintain settings and save logged data.
Note that the battery provided has a built-in indicator of charge level.
Push on the “Check” button to see the charge level. If none of the
LEDs lights up, the battery is not charged.
These particle counters may be powered using rechargeable lithium-ion batteries (from one to four) or through an AC power cord.
2-4 AeroTrak® Portable Airborne Particle Counter
To Install the Lithium-Ion Battery
N O T E
It is also possible to “hot swap” batteries if you are careful. As long as one charged battery is engaged at all times, a fully charged battery can be inserted and then the depleted battery removed without powering down the unit.
W A R N I N G
The battery supplied by TSI (PN 1208057) has built-in protection against
explosion and fire hazard. Do not use a substitute.
Do not use any other type of battery in this instrument. Fire, explosions,
injuries, or other hazards may result.
1. Remove the battery door on the back of the instrument by turning the two thumbscrews counterclockwise.
2. One battery is provided for the 28.3 and 50 L/min models, but up to two batteries can be used to extend the run time. Two batteries are provided for the 100 L/min model (Model 9500) and both must be present for the unit to operate on batteries. This is due to the larger power draw at the 100 L/min setpoint. Slide the battery into the slot (it does not matter which one), pressing until it is flush with the back panel (note the orientation of the tabs).
3. Replace the battery door and secure with the two thumbscrews.
4. The batteries will be charged if the instrument is on and not actively taking a measurement. If you use the unit with batteries often, you might also consider using the TSI external battery charger (see
Optional Accessories in Chapter 1).
Getting Started 2-5
To Use AC Power
W A R N I N G
The instrument turns on automatically when the AC power supply is
plugged in.
Power Entry
Connect the country-appropriate power cord to the external power supply. Next connect the 24 VDC connector to the socket in the particle counter and then connect the other end to an AC outlet.
Using the Stylus
This particle counter is shipped with a plastic stylus for use with the touch screen interface. Use your fingertip or the stylus only. Do not use sharp objects, such as pens or pencils on the touch screen as they may damage it.
2-6 AeroTrak® Portable Airborne Particle Counter
Using the Integral T h e rmal Printer
The side-mounted integral thermal printer is standard on most models to print manually, automatically after each test is completed, or whenever the alarm function is activated (see Print Setup Screen and Print
Schedule Screen on the System Setup Screen).
Printer paper has a colored strip printed on the last few feet of each roll to indicate time to change the paper roll.
When installing a new roll of paper, the tag end should be from the bottom of the roll and pulled through the printer door.
The printer has an LED indicating that the printer is ready.
There is also a feed button before tearing it off. To tear off, pull steadily down on the paper from one side of the serrated edge to the other.
to allow a manual feed of the paper
Getting Started 2-7
The printer has a feed and stop button as well as an LED
W A R N I N G
The monitor Alarm Output function should not be used to detect
hazardous conditions or to provide an alarm for protecting human life, health, or safety.
C a u t i o n
The alarm switch must not be wired to AC power! Failure to properly
install the user alarm could damage the instrument and/or void the instrument warranty! Please read and follow all instructions before wiring or operating the user alarm.
Alarm
Filter Scanner
Probe
USB A (Flash Drive)
indicating that the printer is ready. The feed button can be held down at the end of a print to allow enough space to tear off the paper. If you unintentionally start a print (especially something very long), stop printing
with the stop
Using Peripherals
Connecting to the Alarm Closure
This connector provides two pins for a contact closure to control an external alarm. The contact closure is normally open and rated for 0 to 60 V AC/DC at 1.5A peak, 0.5A continuous. The “alarm out” connection is rated for 60 VDC insulation. The contact is closed under alarm conditions determined by the programming of the device. When used with an externally powered device, this can trigger a visual or audible local alarm (such as a light pole).
button.
2-8 AeroTrak® Portable Airborne Particle Counter
W A R N I N G
When connected to the alarm out connector, you must use safety
certified equipment and/or power sources.
Filter Scanner
Velocity/Temperature/Relative Humidity Probe
964 (Straight) 966 (Articulated)
Velocity/Temperature Probe
960 (Straight) 962 (Articulated)
This connector is used with the optional TSI Electronic Filter Scanner (TSI P/N 700103). Plugging in the connector from the Electronic Filter Scanner allows you to start and stop a sample from the probe head. In addition, the alarm for exceeding a given level can be heard at the probe as well as from the instrument.
Probe
This connector is used with a variety of TSI velocity/temperature/ humidity probes such as those listed below. The probe is plugged into this port with the power off. When a probe is connected to the instrument, additional velocity, temperature, and relative humidity information is displayed on the top, center of the main screen as well as stored in each data file. Data for these parameters is averaged over the sample time for each sample.
USB A
This standard USB connector is provided for use with “flash drive”
devices to download data from the instrument and transfer to a computer or other device. The data files are in XML format for easy use in spreadsheet programs. You can plug in a flash drive at any time. Make sure there is enough space free on the drive to download data files. See instructions on using the flash drive in Chapter 4.
A keyboard can also be used with this standard USB connector, to facilitate data entry at the portable, if desired, as an alternative to the touch screen keyboard.
Getting Started 2-9
Alarm
This connector is used to communicate the alarm status of the instrument. When the instrument is in flow or laser alarm, the dry relay will close. The mating connector and ferrite is provided. The user must supply the wiring. Install the clamp-on ferrite so that it is within 1 inch of the connector as shown.
2-10 AeroTrak® Portable Airborne Particle Counter
Using Communications Po r ts
Ethernet Port
USB B (to computer)
USB B
The standard USB B connector is used to connect the instrument to a computer running TSI TrakPro Lite Secure software for data downloading and analysis. It is used with a standard USB cable and TrakPro Lite Secure Software (both included).
Ethernet Port
The particle counter is compatible with either 10 or 100 MBps systems. The green LED indicates that the network is connected. The yellow LED indicates activity on the network cable. The instrument cannot be operated using power-over-Ethernet (POE).
The Ethernet LAN connector is a standard 10/100 Mbps 8-Position 8-Contact (8P8C, often called RJ45) modular plug connection.
Getting Started 2-11
Using an Isokinetic Probe
The isokinetic probe smoothly accelerates air into the inlet of the instrument. The barbed isokinetic probe can be used with tubing and an adjustable tripod mount (not included) to monitor particles in hard to reach places or that are flowing horizontally.
2-12 AeroTrak® Portable Airborne Particle Counter
C H A P T E R 3
Operation
The AeroTrak® Portable Airborne Particle Counter is controlled using a touch screen display. Use the plastic stylus or your finger tip. DO NOT use sharp objects (such as a pen point) that may damage the screen overlay.
To turn on the instrument, press the power switch . After a splash screen displays the TSI logo, a brief start-up sequence begins as the Windows® CE operating system boots up.
The instrument is ready for operation when the main tab (shown below) appears. If an optional temperature/humidity probe is attached, those values will be shown in the upper-left white area also.
The printer icon will come on when sampling is started.
3-1
Screen Layout a nd Functionality
There are four main screens (tabs): Main, Setup, Data, and Reports. The operation of each of these screens, the information displayed on them, and the operations you can perform from each are described in the remainder of this chapter.
Some screens require or allow you to enter information. To enter information, tap on the screen and an on-screen keyboard appears.
Software Input Panel (Keyboard or Keypad)
1. Throughout the setup screens, a keyboard or keypad appears on the screen when text or numbers may be entered.
2. When you enter information using the keyboard, press either the
 (Enter) or Esc keys when you are done. When you enter data
using the keypad, the data is entered when you press OK on the screen. The keyboard will then be hidden until another text entry box is selected.
3. When numeric input is needed, a numeric keypad will appear on the screen.
Main Tab
The Main Tab is the default screen. The left side of the screen summarizes the concentrations for the currently selected location. Tap on the size and count portion of the screen to enable Zoomed Data Screen (see Setup Tab).
The display shows:
Temperature*  Relative humidity*  Air Velocity*  Bin sizes  Particle count/concentration
*
Temperature, Humidity, and Air Velocity are displayed only if the optional T/H probe is installed.
3-2 AeroTrak® Portable Airborne Particle Counter
The status bar at the top of the screen shows the current time and date
Icon
Description
Instrument status error. If this icon is shown, it can be pressed and a more detailed description of the operational error will be shown. Refer to the troubleshooting section for appropriate actions. If an instrument error occurs, an audible alarm will also sound.
Sufficient flow through the instrument. NOTE: During Start Delay (Delay) and Hold Times (Hold), this is
only an indicator of flow On. During Sample Time (Time), this is an indicator of flow within specified tolerances.
Insufficient flow through the instrument. If this icon is shown, it can be pressed and a more detailed description of the flow error will be shown. Refer to the trouble shooting section for appropriate actions.
If a flow error occurs, an audible alarm will also sound. NOTE: During Start Delay (Delay) and Hold Times (Hold), this is
an indicator of flow Off. During Sample Time (Time), this is an indicator of flow not being within specified tolerances.
Operating on AC power, no battery installed.
Operating on AC power, battery is installed and charging. (The battery charges when the instrument is on but not actively taking a particle sample.)
Battery charged
Low battery and operating on battery.
Battery is very low!
Indicates that TrakPro Lite Secure software is interfacing with the AeroTrak particle counter. The front panel GUI is inoperable when the software is operational. Once the software is exited, normal front panel operation will resume.
settings (see the Setup Tab) and indicates:
Press and hold the (Zone) icon to display a summary of information for the current zone.
Tap the (Location) icon to step through the list of Locations for the Zone.
Operation 3-3
Field
Description
(Zone)
Displays the zone where the sample is being taken by the instrument. Press the icon to display a summary of information for the zone.
(Location)
This dropdown box allows selection of a preconfigured Location to associate the sampled data to.
Time
The time for each sample.
Delay/Hold
The Delay displays one of two times:
1. Before the Start button is pressed the Start Delay time is displayed and then immediately after the Start button is press the delay time begins a countdown.
2. During sampling and between cycles (after the Start Delay has been displayed), the Hold Delay is displayed and then begins a countdown.
Volume
The volume of air sampled for a given sample.
Sample
The specific sample being taken out of the total number of samples programmed to be taken.
Recs
The total number of records in the database/10000 (maximum number of records).
Manual/Automatic/Beep
Mode Indicator; refers to the “Data Count Mode”
(see section below).
Start/Stop button to begin and end sampling in the configured mode.
Press to print the current sample to the optional printer.
3-4 AeroTrak® Portable Airborne Particle Counter
1D barcode scanners that support USB 1.0 or 1.1 devices can be used (Zebra LS2208 works well) to select an existing Zone and Location pair in the list boxes on the Main page, if the barcodes use the format ZoneName:LocationName (i.e., the zone name followed by a colon followed by the location name). The barcode can then be scanned and the specified zone and location will be automatically selected in the list boxes.
There are free online barcode generators available (i.e.,
http://barcode.tec-it.com/en). Select linear Code 128 to generate the
barcodes.
Operation 3-5
Zoomed Data Screen
Field
Description
Location
Label that displays information about the currently selected location.
Manual/Automatic/Beep
Mode Indicator; refers to the “Data Count Mode”
(see section below).
Press the Start/Stop button the begin sampling in the configured mode.
The Zoomed Data screen is entered by touching in the size and count part of the main tab display. The bottom portion of the screen summarizes the concentrations for the currently selected location. Tap the size and count portion of the display to switch back to the Main Tab display.
The display shows:
Temperature*  Relative humidity*  Air Velocity*  Bin sizes  Particle count/concentration
*
Temperature, Humidity, and Air Velocity are displayed only if the optional T/H probe is installed.
3-6 AeroTrak® Portable Airborne Particle Counter
Setup Tab
Zones Setup
Identify and save the location information associated with collected samples.
Recipes Setup
Save a group of settings (a recipe) that you use over and over so you don’t have to reset individual settings.
Environment Setup
Sets which units are preferred for displaying environmental measurements taken using optional measurement probes.
System Setup
Change Power On Password, Setup Password, System Configuration, Print Settings, Print Schedule and Clear Samples.
Purge
Purge residual particles from flow path.
Info
Display instrument information such as model number, firmware version, serial number, etc.
Device Setup
Set Date and Time, Screen Alignment, Communications, Regional Settings, and get device information.
The setup tab provides access to the following:
Operation 3-7
Zone Setup Screen
Field
Description
Name
The name you assigned to the zone.
Standard
The classification standard to use for the samples taken in the Zone. Options include ISO 14644-1, EU-GMP, Fed Std 209E F, Fed Std 209E, and None. Use “None” for taking measurements that are not associated with standards classification.
Class
The Class selected for the classification of the Zone. Options vary by the Standard selected.
Status
The occupancy status of the Zone. Options vary based on the selected standard, but include At Rest, Operational and As Built.
Air Flow
The direction of air flow through the zone: options are Unidirectional or Multidirectional.
Area
The area of the zone in ft2 or m2.
Largest Particle Size to Consider
The largest particle size to consider for classification measurements. Used by most standards to calculate minimum required sample volume.
Locations
The number of locations within the zone.
Recipe
The recipe assigned to the zone.
Zones” are a convenient way to group sample data for printing and export, and are required for creating standards-based classification reports. A Zone contains 1 or more “Locations;” this is modeled after cleanroom standards that prescribe the classification of a zone (or room) by taking samples at various locations within the zone.
Use the Zone Setup screen to add, delete or edit Zone configurations.
The Zone configuration screen provides the following information for each zone that is configured.
3-8 AeroTrak® Portable Airborne Particle Counter
To Delete A Zone
To delete a zone from the configuration screen, select (highlight) the zone name and press Delete. A verification message “Are you sure you want to delete this Zone?” appears. Press Yes to delete the zone.
A zone that has data associated with it cannot be deleted. The data associated with the zone must be deleted from the instrument before the zone can be deleted.
To Add a Zone
To add a zone, press Add. The Definition Screen is displayed.
1. Enter a name for the zone and select the Standard, Class, Status, Air Flow, and Largest Particle Size to Consider options from the dropdown boxes. Input the Area using the keypad and select either ft2 or m2 to describe the area of the zone.
2. Press the Locations tab. The Locations screen is displayed.
Operation 3-9
3. Enter names for each location in the zone and the press Add after
NOTES
You can also create recipes from the Setup Tab by selecting the
Recipe icon . But if you create a new recipe here, information
you have already entered for the zone is prepopulated into the required fields.
If you edit an existing recipe, your changes will affect all zones
using that recipe. Be certain that is what you want to do.
entering each. The name will be added to the box on the left side of the screen.
4. Press the Recipe tab. The Recipe screen is displayed with a default recipe in the “Selected Recipe” field.
5. Select the recipe you want to use from the “Selected Recipe” field or press Create Recipe to create a new recipe or Edit Recipe to edit the recipe shown in the “Selected Recipe” field.
See Recipes Setup Screen in the Setup Tab section for information about the fields and parameters of the recipe tabs.
6. When you are done selecting the recipe to use or adding a new recipe or editing an existing recipe, press Save or Save New Zone.
To Edit A Zone
1. To edit an existing zone configuration, press Edit. The following screen is displayed.
2. The display has three tabs: Definition, Locations, and Recipe. Select the tab for the information you want to edit.
3. The Definition Screen for each zone provides the same information as displayed on the main Zone Configuration screen with the addition of “Largest Particle Size to Consider” field.
3-10 AeroTrak® Portable Airborne Particle Counter
4. The Locations Screen displays the locations within the selected zone. You can add, rename, or remove a location from the zone. You can also move the location name up or down in the list. The Auto feature will generate the number of locations required by the chosen standard. This utilizes the room area entered in the zone definition. The locations can then be renamed or the default naming convention can be maintained.
5. You can take as many samples as possible (up to 10,000) under a specific location attached to a specific recipe and then be able to select the samples based on date to print a report or export data for a selected number of samples.
6. The Recipe Screen displays the recipe in use for the selected zone and information relevant to that recipe. You can select a different recipe for the zone or you can create a new recipe or edit an existing recipe. (For information about recipes see the Recipes Setup Screen in Setup Tab section.)
Operation 3-11
NOTE
If you edit a recipe, the changes will affect all zones using that recipe. Be certain that is what you want to do.
NOTE
The delete button will be grayed out and unavailable if the recipe has samples.
7. When all the changes have been made, press Save.
Recipes Setup Screen
Use the Recipes Setup screen to review recipes, add or delete recipes, and edit recipes. The “Default” recipe cannot be deleted. A recipe that has samples cannot be deleted.
3-12 AeroTrak® Portable Airborne Particle Counter
The steps for adding or editing a recipe are identical. Press either the
Field
Description
Automatic
If this mode is selected, the particle counter starts counting in automatic mode when the start button is pressed according to the settings on the Recipes Timing Screen.
Manual
If this mode is selected, the particle counter starts sampling immediately when the start button is pressed and stops at the end of the sample time, which is configured on the
Recipes Timing Screen.
Beep
The Beep mode enables the AeroTrak particle counter to
operate in a “Geiger Counter” mode. As particles are
detected, a beep is emitted. The frequency of beeps configured utilizing the alarm thresholds setting. It works on a single bin. If you wish to beep on total particulates, configure the unit in cumulative mode and set an alarm threshold for the .5 channel. The alarm threshold determines the beep frequency. The actual number of particulates measured in the preceding 1 second will be divided by the threshold and the corresponding number of beeps emitted. An alarm threshold of 0 will not emit beeps.
To alarm on viable particulates, be sure that all total particulate alarms are disabled and configure an alarm in the viable count channels. It is configured in the same manner described above.
Add button or the Edit button and proceed as follows:
1. Press Add or Edit button to display the Recipe Tab.
2. On the Recipe tab, enter a name or edit the name of the recipe. For a new recipe, a default name will appear, but you can type over it and name the recipe anything you want.
3. Select the Count Mode: options are Automatic, Manual, and Beep as described below.
Operation 3-13
Field
Description
Beep (continued)
If multiple alarms are configured, the AeroTrak particle counter will emit beeps only on a single channel. If multiple alarms are configured, the AeroTrak particle counter searches for which alarm to operating on starting with the smallest to largest Total Particulate Channel selected followed by the smallest to largest Viable Particulate Channel.
Settings in the sample timing screen are ignored in beep mode.
Example: Looking for a viable particle source that raises above 500 count background by 10000. Configure the unit in cumulative mode and enable the .5 viable particulate alarm to 500. The AeroTrak particle counter will now emit a single beep for every 500 particles. At steady state, a beep will be emitted once per second (500/500=1). When the source is encountered it will emit 20 beeps per second (10,000/500) = 20. If a higher frequency is desired lower the threshold to 200. This will result in 50 beeps per second being emitted.
The maximum number of beeps that can be emitted per second is 50.
4. Finally, enter the zones that this recipe is or will be assigned to.
NOTE
Entering the names of the zones in this box does not assign this recipe to the zone. This is for information only so when you change the recipe for a zone, update the information here.
5. Press the Timing tab to enter or edit start delay times, sample time, hold time, etc.
3-14 AeroTrak® Portable Airborne Particle Counter
6. To make changes to the timing settings, highlight the component to change (hours, minutes, seconds, etc.) and use the on-screen keypad to change the value.
W A R N I N G
Instrument status alarms are inactive if samples times are 10 seconds or
less. Flow error alarms may not occur if sample time is less than 15 seconds. To ensure proper instrument status and flow alarm operation, utilize a sample time of 15 seconds or larger.
NOTE
With firmware version 2.4 and higher, you will be able to print reports and/or export sample data selected by date. There is no need to create new zones to select data by date. There is no need to print all samples in a zone to view a specific sample or samples taken on specific dates. Samples can also be selected by date for a given recipe and zone.
Field
Description
Start Delay
Start Delay indicates how long it will be before the first sample is taken.
NOTE
It takes approximately 10 seconds for the pump to reach the flow set point; taking a measurement before the pump is functioning properly may result in a data and flow error.
Sample Time
Sample time indicates how long the instrument will run for each sample.
Hold Time
Hold Time indicates how long the instrument pauses between samples.
Cycles
Cycles is the total number of samples you want collected. In Automatic mode, a cycle value of 0 causes the instrument to count continuously using the settings for Delay, Time, and Hold until the Start/Stop button is pressed again.
Volume
Volume sets the volume of air that will pass through the instrument for each sample. After you enter a volume, you must select Cubic Feet, Cubic Meters or Cubic Liters for measurement.
Operation 3-15
7. Press the Channels tab.
NOTES
In Differential modes (Δ), disabling one or more channels will
disable all threshold alarms. Other alarms are not affected.
While in beep mode, a threshold of 0 will not trigger an audible
alarm even if Alarm is enabled.
Concentration display is unavailable in Beep mode.
8. This tab can be used to view or set the particle size for each channel (not supported in all models), enable/disable the channel, enable/disable alarm for each channel and set the alarm threshold for each channel. The threshold values are expressed in the units selected in the Count Units control. Select the appropriate Count Units from the list.
During sampling, when a channel value exceeds the threshold value set here, the channel data is highlighted in red on the Main screen, an audible alarm sounds, and the alarm icon.
To acknowledge the alarm and silence the buzzer, tap the alarm icon .
9. Press Save or Save New Recipe as appropriate when done.
3-16 AeroTrak® Portable Airborne Particle Counter
NOTE
The Channel Configuration screen has restrictions that must be noted when Differential mode () is selected.
When differential Δ particle count or concentration is selected, the total number of counts is the number of particles between enabled bin sizes. When particle concentration is cumulative Σ, the total number of counts includes all particles larger than the bin size.
In Differential Display/Alarm mode, there are two constraints:
If alarming is desired, all channels must be enabled. If channel selection is desired, then all alarms must be disabled.
The controls work in a mutually exclusive manner. When any of the
channel “Enable” boxes are unchecked, all “Alarm” enable boxes will be cleared. When any of the “Alarm” boxes are checked, all of the
channel “Enable” boxes will become checked.
For the Cumulative modes (), there are no such constraints. Any combination of “Enable” and “Alarm” selections can be made.
System Setup Screen
Use the System Setup screen to select (or change) the power on password, set up a password, select system configuration parameters, select print settings, schedule printing and clear samples.
Change Power On Password Screen
If a Power On password has been previously set, that password must be entered before being allowed to change the Power On password. If a Power On password is set, then on instrument startup a password screen will ask for the password before the instrument can be used. A blank
Operation 3-17
password is regarded as no password and if set as the new password,
NOTE
Keep the password in a safe place. It is difficult to reset the password and requires contacting the factory. If the password has been misplaced, please contact TSI technical support.
Field
Description
Old Password
Enter your existing password (if one has already been set) or leave blank.
New Password
Enter a new password. The password can be any length and use any characters.
Confirm New Password
Retype the new password then press OK. A confirmation message appears if the password is changed.
NOTE
Leave both New Password and Confirm New Password fields blank to turn off password protection.
Call TSI if you have forgotten the password.
will not prompt you for a password on system startup.
Tap on the screen to display the on-screen keyboard and enter the required information.
3-18 AeroTrak® Portable Airborne Particle Counter
Change Setup Password Screen
Field
Description
Old Password
Enter existing password (if one has already been set) or leave blank.
New Password
Enter a new password The password can be any length and use any characters.
Confirm New Password
Retype the new password then press OK. A confirmation message appears if the password is changed.
NOTE
Leave both New Password and Confirm New Password fields blank to turn off password protection.
Call TSI if you have forgotten the password.
If a Setup password has been previously set, that password must be entered before being allowed to change the Setup password. If a Setup password is set, clicking on the setup tab at the bottom of the main screen brings up a password screen. That password must be entered in order to change instrument settings.
Tap on the screen to display the on-screen keyboard end enter the required information.
Operation 3-19
Configuration Screen
Field
Description
and on Zoom
Select to zoom in on both cumulative (Σ) and differential (Δ) counts on the Main Tab. To zoom the Main Tab, select on the left side of the Main Tab (it takes a moment for the screen to update.) Click on the screen again to return to normal view.
Store Partial Samples
When selected, stores the partial record in the current database if the instrument is stopped during a sampling period.
Alarm Volume Level
Controls the alarm volume setting.
Use the Configuration screens to set configuration parameters. Press OK when finished.
3-20 AeroTrak® Portable Airborne Particle Counter
Field
Description
Relay Trigger
Check the box or boxes for the icon or icons upon which the relay trigger will close.
Relay will close on instrument status errors including laser error, algorithm corruption, or select control voltages.
Relay will close on any flow error including sample flow, sheath air flow, or concentrator flow.
Either of these errors signifies potentially invalid data. Relay closure can be selected to operate on any or all
alarm functions.
Trigger Delay
This is the number of sample cycles that must occur before the relay is activated.
(continued on next page)
Operation 3-21
Print Setup Screen
Field
Description
Serial Number
Indicates that the serial number of the particle counter used to collect the data will be printed.
Model Name
Indicates that the model number of the particle counter used to collect the data will be printed.
Separator
Indicates a line separator will be printed after the Model Name and Serial Number in the header of all printouts.
Differential
Indicates that the differential value of the data will be printed.
Cumulative
Indicates that the cumulative value of the data will be printed.
Last Calibration
The date and time the instrument was last calibrated by TSI.
Statistics
A statistical summary is printed after the sample data.
Statistics Only
A statistical summary is printed instead of sample data.
NOTE
Printer paper has a colored strip printed on the last few feet of each roll to indicate when it is time to change the paper roll.
A hard copy of a sample set or statistics can be printed from the instrument using an optional thermal printer. Use this screen to set print parameters. Press OK when finished. These selections will apply when printing automatically or when printing from the Data tab page.
3-22 AeroTrak® Portable Airborne Particle Counter
Print Schedule Screen
Field
Description
Automatic Printing
Enables automatic printing.
On Sample
Print data whenever a sample completes.
On Alarm
Print data when an alarm condition occurs.
English Only Printing
Always print data in English.
Use the Print Schedule screen to schedule automatic printing. Choose to either print when an alarm occurs or print whenever a sample is complete.
“English Only Printing can be selected to print reports in English even when the selected language is a language other than English. Prints default zones and locations in English. If zones or locations have been created with names in other languages, the names are not translated to English.
Operation 3-23
Clear Samples Screen
C A U T I O N
WHEN “YES” IS SELECTED ON THE CLEAR SAMPLES SCREEN,
ALL SAMPLE RECORDS WILL BE DELETED FROM THE INSTRUMENT! THERE IS NO WAY TO RECOVER THEM ONCE THEY HAVE BEEN DELETED.
Use the Clear Samples screen to clear all samples from the internal database. Select Yes to clear all samples. Select No to return to the System Setup screen.
3-24 AeroTrak® Portable Airborne Particle Counter
Purge Screen
Use the Purge feature to improve zero count test by removing any residual particles in the flow path and depositing them on the exhaust filter. TSI recommends that this feature be used when:
1. The instrument was exposed to high concentrations.
2. The instrument was not zeroed over long periods of time.
3. After a field calibration.
4. The instrument was moved around a lot. Attach filters to the inlet and click the Start Purge button. The pump will
continue to run for 15 minutes and then stop. To stop the pump before the 15 minutes have completed, click the Stop Purge button.
Operation 3-25
Information Screen
Use the Information screen to view the system’s model, serial number,
copyright, manufacture date, calibration date, next calibration date, firmware version, USB IP address and date and time format. Press Close when finished.
Device Setup Screen
Use the Device Setup screen to access screens to set or change the date and time, set visual parameters of the display, set up communications, or set regional features.
3-26 AeroTrak® Portable Airborne Particle Counter
Date and Time Screen
Field
Description
Date
Press the down arrow to display a calendar then select the date from the calendar.
Time
Select the time component you want to change (hours; minutes; seconds) and then use the left and right arrows to adjust to the current time.
24 Hour
Time display is in 24 hour format.
Use the Date and Time screen to set the current date and time and set the date format. Press OK when finished. Select options using the arrows or tapping on the screen which brings up the keypad.
Operation 3-27
Display Screen
Field
Description
Screen Alignment
Press this item to reset the screen alignment. Follow the directions on the alignment screen.
NOTE
The touchscreen display is aligned at the factory and typically will stay aligned for the life of the instrument. Only perform this alignment if tapping on the onscreen controls of the instrument seems to give poor results.
Use the Display screen to set or change visual parameters.
Communications Screen
Use the Communications screen to configure the IP address, subnet, and default gateway to which the instrument belongs. Addresses can be entered using the arrows or by selecting a field and using the on-screen keypad.
3-28 AeroTrak® Portable Airborne Particle Counter
Field
Description
IP Address
The numerical identification (logical address) that is assigned to this device when participating in a computer network utilizing the Internet Protocol for communication between its nodes.
Subnet Mask
A network of computers and devices that have a common, designated IP address routing prefix. All hosts within a subnet can be reached in one "hop" (time to live = 1), implying that all hosts in a subnet are connected to the same link.
Default Gateway
A node on the computer network that serves as an access point to another network and is chosen when the IP address does not belong to any other entities in the Routing Table.
Use DHCP (Dynamic Host Configuration Protocol)
When checked, this protocol is used to automatically obtain the information necessary for operation from a DHCP server running on your local network.
NOTE
TCP/IP is an industry standard networking protocol that allows computers and devices to communicate over Ethernet and other media access channels. Providing full details on how to configure an IP network is beyond the scope of this manual. Please contact your company IT department or a qualified networking professional if you are not qualified to configure such a network.
Regional Screen
Use the Regional screen to set the language in which the on-screen dialog is displayed and your regional format for numbers.
Operation 3-29
Field
Description
Language
Select the language in which you want on-screen dialog displayed; options are German, English, Spanish, French, Italian, Chinese (simplified), and Japanese.
Formats
Select the format that is commonly used to display real numbers and the date and time in your region.
Environment Screen
Field
Description
°F
Display temperature in degrees Fahrenheit.
°C
Display temperature in degrees Celsius.
ft/s
Display velocity in feet per second.
m/s
Display velocity in meters per second.
Use the Environment screen to set the units for temperature, which is displayed on the Main and Data Tabs, and the printouts when a humidity and temperature probe is hooked up to the instrument.
3-30 AeroTrak® Portable Airborne Particle Counter
Data Tab
NOTE
Counts displayed on the data tab concentration may have slight rounding errors when comparing all channels to values with selected channels enabled. The method for calculating concentration is to sum the raw counts for each location then calculate concentration from sample volume and then round. This may result in slight rounding errors when comparing counts with all channels enabled versus concentrations with selected channels enabled. The methodology utilized is covered in ISO 14644-1 Annex D.
Field
Description
#, ft3, m3
Button used to change between counts and concentration displays:
# = number ft
3
= particles per cubic foot
m3 = particles per cubic meter
Size
Channel size.
Differential concentration.
Cumulative concentration.
Export the data to a flash drive. See Export Data Screen below.
Print data to the optional printer. See Print Data below. Zone (Z)
Zone where the data was collected.
Location (L)
Location where the data was collected.
Use the Data tab to preview data that has been collected. To scroll though the records, use the elevator (slide) on the right. The record number is displayed at the bottom of the tab. As each record displays, its data and relevant parameters are displayed.
Operation 3-31
Field
Description
Sample
Duration of the sampling period.
Date
Date on which the data was collected.
Time
Time at which data was collected.
Temp
Temperature at the end of the time the data was collected (if probe connected during sampling).
RH
Humidity level at the end of the time the data was collected (if probe connected during sampling).
Alarm
Alarm threshold was triggered (Yes) or not (No).
Inst
Status of the instrument hardware. OK if no issues or an error abbreviation if instrument has a possible issue.
Vol
Volume of air that was sampled.
Vel
Velocity of the ambient air at the sampling location as measured by a connected TSI Velocity probe (optional).
Record
This record number.
Records
Total number of records.
(continued on next page)
3-32 AeroTrak® Portable Airborne Particle Counter
Export Data Screen
Field
Description
Secured file
This file is intended to be used with TrakPro Lite Secure software and maintains CFR 21 Part 11 compliance. The file has the extension file name_sec.xml.
Unsecured file
This file is intended for user input into Excel for graphing and data manipulation purposes and has the extension file name.xml.
Both
If using both file types, both file formats can be exported. Please note that the data export time is longer when both file formats are exported.
C A U T I O N
Do NOT modify the secure file. If the “_SEC” secure file is modified,
TrakPro Lite Secure software will not be able to open the file.
Use the Export Data screen to export sample data to a flash drive. Select the name of the file and range of data to export. Data is downloaded into an XML file that can be opened with commonly used spreadsheet programs.
To Export Data
1. Click the USB drive icon on the Data tab. The Export Data screen appears.
Operation 3-33
2. Select a file from the list and click: a. “Export” to overwrite an existing file. b. “Export As…” to enter a file name. Then select OK.
3. Select Sample data by Zone or by Sample index range with option of limiting samples to a date range.
a. Check “Use Start Date” checkbox and select a start date to
exclude data that was collected prior to that start date.
b. Check “Use End Date” checkbox and select an end date to
exclude data that was collected after that end date. Date is displayed as month/day/year.
3-34 AeroTrak® Portable Airborne Particle Counter
4. This form allows you to select data for export by zone. Move a desired zone to the box on the right to select it. To select data by sample index range instead, tap the Select by Record button at the bottom.
5. This form allows you to select data for export by range of sample Index. Tap Select Range radio button and then select the lower and upper sample index numbers. To select data for export by zone instead, tap the Select by Zone button at the bottom.
6. Once the records or the zone have been selected, press Export… to begin exporting. Status screens allow viewing the progress of the export.
or
Operation 3-35
C A U T I O N
Do NOT remove the external drive during the export process. If the thumb
drive is removed, re-insert and restart the download process. Data stored on the instrument is not lost during the transfer.
3-36 AeroTrak® Portable Airborne Particle Counter
Print Data
The print button allows a range of sample data to be printed using the internal printer.
The “Print by Record” form or the “Print by Zone” form can be used.
Both forms provide the option to limit data selection to a date range.
Check the “Use Start Date” checkbox and select a start date to
exclude samples collected before that date.
Check the “Use End Date” checkbox and select an end date to
exclude samples collected after that date.
Check the “Recipe” checkbox to include recipe details for each recipe
used in the report.
To Print Data
1. Click the Printer icon on the Data tab. The “Print by Zone” screen appears.
2. Move a zone to the box on the right to select it for printing.
3. Optionally filter samples by date by checking start date and/or end date checkboxes and selecting the dates.
Operation 3-37
4. To select data by range or sample index instead, tap the Select by Record button at the bottom.
5. The “Select by Record” screen above allows you to select data by range of sample index number. Tap the Select Range radio button and select the lower and upper sample index numbers. To select data by zone instead, tap the Select by Zone button at the bottom.
6. To filter the data by start and/or end date, check the appropriate checkboxes and select the date.
7. Check “Recipe” to include the details of each recipe used by samples that are included in the report.
8. Once you have identified the records or zone to print, press the Print… button.
3-38 AeroTrak® Portable Airborne Particle Counter
9. The print data screen shows progress on the current selected range of sample data to be printed. Press the Cancel Printing button to cancel the rest of the print job.
Printouts that include sample data will identify the samples as “Sample X of Y” where Y is the total number of samples in a set being printed (not
the total number of samples on the instrument) and X is the current sample in the set (not the sample ID for the sample).
Operation 3-39
Reports Tab
Use the Reports Tab screen to select various standard reports for viewing and printing.
1. Select a zone to be included in the report and you have the option to restrict data to a date range.
a. Check the Use Start Date checkbox and enter a start date to
exclude data collected before that date.
b. Check the Use End Date checkbox and select an end date to
exclude data collected after that date.
2. Check the “Recipe” checkbox to include recipe details on the report.
3. When you press the Generate button, you are given the option of entering a title for the report.
4. Enter a title or leave the textbox blank.
5. Press the Enter key on the visual keyboard or press the Close button of the form to close the form and generate the report.
3-40 AeroTrak® Portable Airborne Particle Counter
6. Report will show the title you provided.
Field
Description
Zone
Select the zone from the dropdown list.
Standard
Select the standard from the dropdown list.
Class
Select the class from the dropdown list.
Use Start Date
Exclude samples collected before a specified date.
Use End Date
Exclude samples collected after a specified date.
Recipe
Check to include the recipe details on the report.
Exclude Samples button.
Tap Exclude Samples button to display form that allows you to enter IDs of samples to exclude and IDs of replacement samples (if required by standard).
When you close the Exclude Samples form, the excluded samples will be listed in the listbox on the Reports page.
Generate
Press to begin generating a report that you can view on­screen or print.
The “Exclude Samples” form shown below is displayed when you click the Exclude Samples button. Enter the ID of a sample to exclude in the left column and enter the ID of a sample to replace it (if required by the selected standard) in the right column. The software allows you to enter up to five excluded and five replacement samples for ISO 14644-1:2015. If the standard (e.g., ISO 14644-1:1999) allows only one excluded sample and no replacements, only the first edit box in the left column will be enabled and the replacement boxes will all be disabled.
Operation 3-41
To save the selections for the report, click Save button. To clear the textboxes (i.e., cancel use of excluded samples and replacements) click the Clear button.
3-42 AeroTrak® Portable Airborne Particle Counter
Ethernet Port
USB A (Flash Drive)
USB B (to computer)
C H A P T E R 4
Data Handling
There are three basic ways to get data from the AeroTrak® Portable Airborne Particle Counter:
1. Data download to a USB Flash Drive.
2. USB port connection to a computer using TrakPro Lite Secure software.
3. Ethernet connection to a TSI FMS Server.
USB Data Download
The AeroTrak® Portable Airborne Particle Counter is equipped with a USB A host drive that will allow for the downloading of stored data to a USB flash drive (also commonly called a USB thumb drive). To download data, attach a flash drive to the USB A host port and follow the instructions in the operation section of this manual. The data is downloaded in XML format that can be opened in Microsoft Excel® version 2003 or greater. The data files can also be opened in the latest versions of OpenOfficeTM.
4-1
USB Computer Commun i cation
The AeroTrak® Portable Airborne Particle Counter is equipped with a USB compatible to USB B cable for connection to a personal computer. The cable plugs into the right side of the instrument. Data from the particle counter can be downloaded to a PC, through TrakPro Lite Secure software and recipes can be uploaded to the particle counters using the TrakPro Lite Secure software.
Installing Software
See the TrakPro Lite Secure Software (version 3.0 or later) User’s Guide (P/N 6004404) on CD (P/N 7001901) for installation instructions.
Ethernet Communicat i o ns
An Ethernet port is provided for use with TSI Facility Monitoring Software (FMS). Refer to the FMS Software documentation and the TSI service and installation manual for detailed configuration and operation information on Modbus TCP over Ethernet.
4-2 AeroTrak® Portable Airborne Particle Counter
NOTE
There are no user-serviceable parts inside this instrument. Opening the instrument case may void the warranty. TSI recommends that the AeroTrak
®
Airborne Particle Counter be returned to the factory for any
required maintenance or service not described in this manual. Cleaning can be done without removing the instrument case.
Item
Frequency
Zero check
Daily or according to application
Factory cleaning and calibration
Annually.
Cleaning the instrument enclosure
As needed
C H A P T E R 5
Maintenance
The chapter contains maintenance and troubleshooting solutions for the Model 9310/9510/9350/9550/9500 AeroTrak® Portable Airborne Particle Counters.
Maintenance Schedule
TSI recommends annual factory cleaning and calibration for the AeroTrak® Airborne Particle Counter. See Chapter 7, "Contacting
Customer Service" for service/calibration.
Recommended Field Maintenance Schedule
Zero Check
The zero check ensures that the instrument is properly assembled and free from leaks, residual particles, and electronic noise.
Cleaning the Instrum e n t Enclosure
To clean the enclosure, dampen a lint-free cloth and gently wipe the surface until surface contamination is removed.
5-1
(This page intentionally left blank)
5-2 AeroTrak® Portable Airborne Particle Counter
Symptom
Possible Cause
Corrective Action
Counts are too low
Instrument is being operated outside temperature or relative humidity specifications.
Operate instrument within specifications.
Internal parts have been damaged because instrument was stored at a temperature greater than 122°F (50°C).
Return to factory or factory authorized service centers for service.
Instrument has contamination on the optics due to condensation or excessive loading.
Return to factory or factory authorized service centers for service.
Laser or pump control is damaged.
Return to factory or factory authorized service centers for service.
Unit is due for calibration.
Return to factory or factory authorized service centers for service.
Instrument does not turn on
The on/off button is not being
pressed properly.
Press and hold the on/off button for one second.
Battery is not charged.
Recharge battery or connect to AC power.
Power cord is not plugged into unit.
Connect power cord.
Instrument does not meet zero count specification (<1 particle/5 mins)
Instrument inlet is contaminated with particles
HEPA zero filter is not connected properly and room air is leaking into the HEPA zero filter assembly.
Follow inlet cleaning instructions in manual.
Check that the HEPA zero filter has been tightly connected to the inlet. Check that rubber O-ring (black) on the inlet is in place.
Residual particles from previous samples are shedding off internal parts and into the optics.
Purge instrument by running the instrument for 15 to 30 minutes before attempting zero count test. Longer purge times may be required depending on amount of particle contamination in the instrument.
An internal component has been damaged due to operation outside of temperature specifications or one or more excessive bumps or jolts, and electronic noise is inducing false counts.
Return to factory or factory authorized service centers for service.
A leak has developed in the aerosol flow path.
Return to factory or factory authorized service centers for service.
Internal optics have become dirty.
Return to factory or factory authorized service centers for service.
C H A P T E R 6
Troubleshooting
6-1
Symptom
Possible Cause
Corrective Action
Battery does not charge
The unit must be turned on but not in sampling mode for the battery to charge.
Turn on unit. Green LED by on/off button should be lit.
Unit is actively taking data.
The unit will not recharge the battery while taking a measurement. Once the measurement is complete, the unit will begin recharging the battery.
Unit not put in standby mode when turned off.
Select Standby/Charge when shutting off the instrument if you want the battery to be charged.
BATTERY ERROR
Low battery (<10%).
Recharge battery or connect AC cord.
Single battery in 100 L/min Model 9500, or two batteries where one has been only partially charged
The 100 L/min Model 9500 requires two batteries (within 25% charge of each other) to take a sample. If only 1 battery is installed, then the unit will not start a measurement and will display the battery warning. To correct, place a second charged battery in the instrument.
FLOW ERROR
Instrument was unable to control flow rate [i.e., if sample tubing used is too long (greater than 12 m on 28.3 and 50 L/min instruments, 8 m on 100 L/min instrument)].
Restart measurement. Pressure drop across inlet may be too large.
Lower pressure drop across inlet by using larger diameter tubing, less tubing, and/or adding a bleed valve.
Inlet tubing kinked.
Straighten tubing.
Inlet not at ambient pressure.
Do not subject the unit to other than ambient pressure conditions.
LASER POWER / DETECTOR WARNING
Excessive direct light is entering the aerosol inlet.
Remove instrument from direct light. Optical path blocked.
Return to factory for service.
Nozzle is misaligned. Fiber attached on the nozzle tip.
Contact TSI and return to factory.
Detector board damaged. Laser power is normal.
Return to factory or factory authorized service centers for service.
6-2 AeroTrak® Portable Airborne Particle Counter
C H A P T E R 7
Contacting Customer Service
This chapter gives directions for contacting people at TSI Incorporated for technical information and directions for returning the AeroTrak® Portable Airborne Particle Counter for service.
Technical Contacts
If you have any difficulty setting up or operating the AeroTrak
Portable Airborne Particle Counter, or if you have technical or application questions about this system, contact an applications engineer at TSI Incorporated, 1-800-874-2811 (USA) or (651) 490­2811 or e-mail technical.service@tsi.com.
If the AeroTrak
properly, or if you are returning the instrument for service, visit our website at http://rma.tsi.com, or contact TSI Customer Service at 1-800-874-2811 (USA) or (651) 490-2811.
®
Portable Airborne Particle Counter, does not operate
International Contacts
Service
TSI Instruments Singapore Pte Ltd
150 Kampong Ampat #05-05 KA Centre Singapore 368324
Telephone: +65 6595-6388 Fax: +65 6595-6399 E-mail: tsi-singapore@tsi.com
TSI Instrument (Beijing) Co., Ltd.
Unit 1201, Pan-Pacific Plaza No. 12 A, Zhongguancun South Avenue Haidian District, Beijing, 100181 CHINA
®
Telephone: +86-10-8219 7688 Fax: +86-10-8219 7699 E-mail: tsibeijing@tsi.com
7-1
TSI Instruments Ltd.
Stirling Road Cressex Business Park High Wycombe, Bucks HP12 3ST UNITED KINGDOM
Telephone: +44 (0) 149 4 459200 Fax: +44 (0) 149 4 459700 E-mail: tsiuk@tsi.com
Technical Support
TSI Instruments Singapore Pte Ltd
150 Kampong Ampat #05-05 KA Centre Singapore 368324
Telephone: +65 6595-6388 Fax: +65 6595-6399 E-mail: tsi-singapore@tsi.com
TSI Instrument (Beijing) Co., Ltd.
Unit 1201, Pan-Pacific Plaza No. 12 A, Zhongguancun South Avenue Haidian District, Beijing, 100181 CHINA
Telephone: +86-10-8219 7688 Fax: +86-10-8219 7699 E-mail: tsibeijing@tsi.com
TSI GmbH
Neuköllner Strasse 4 52068 Aachen GERMANY
Telephone: +49 241-52303-0 Fax: +49 241-52303-49 E-mail: tsigmbh@tsi.com
TSI Instruments Ltd.
Stirling Road Cressex Business Park High Wycombe, Bucks HP12 3ST UNITED KINGDOM
Telephone: +44 (0) 149 4 459200 Fax: +44 (0) 149 4 459700 E-mail: tsiuk@tsi.com
7-2 AeroTrak® Portable Airborne Particle Counter
TSI France Inc.
Hotel technologique BP 100 Technopôle de Château-Gombert 13382 Marseille cedex 13 FRANCE
Telephone: +33 (0)1 41 19 21 99 Fax: +33 (0)1 41 19 21 96 E-mail: tsifrance@tsi.com
Returning for Serv i c e
Visit our website at http://rma.tsi.com and complete the on-line “Return Merchandise Authorization” form or call TSI at 1-800-874-2811 (USA), (651) 490-2811, or 001 651 490-2811 (International) for specific return instructions.
Customer Service will need the following information:
The instrument model number The instrument serial number A purchase order number (unless under warranty) A billing address A shipping address
Use the original packing material to return the instrument to TSI. If you no longer have the original packing material, seal off any ports to prevent debris from entering the instrument and ensure that the display and the connectors on the instrument front and back panels are protected. This instrument is very fragile and must be packed in a manner appropriate for a precision instrument.
Contacting Customer Service 7-3
(This page intentionally left blank)
7-4 AeroTrak® Portable Airborne Particle Counter
Specification
Description
Size Range
9310/9350: 0.3 to 25 µm 9510/9550/9500: 0.5 to 25 µm
Channel Sizes (additional channel sizes available)
9310-02: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm 9350-02: 0.3, 0.5, 1.0, 2.0, 3.0, 5.0 µm 9350-03: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm 9510-02: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm 9550-02: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm 9500-01: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm
Counting Efficiency
9310/9350: 50% @ 0.3 μm; 100% for particles
>0.45 μm (per JIS and ISO 21501-4)1 9510/9550/9500: 50% @ 0.5 μm; 100% for particles
>0.75 μm (per JIS and ISO 21501-4)1
Concentration Limits
9310/9510: 820,000 particle/ft3 at 10% coincidence loss 9350/9550: 710,000 particle/ft3 at 10% coincidence loss 9500: 425,000 particle/ft3 at 10% coincidence loss
Light Source
Long life laser diode
Zero Count Level
<1 count/5 minutes (meets JIS B9921 and ISO 21501-4)
Flow Rate
9310/9510: 28.3 L/min (1.0 CFM) with 5% accuracy
(meets JIS and ISO 21501-4) 9350/9550: 50 L/min (1.77 CFM) with 5% accuracy 9500: 100 L/min (3.53 CFM) with 5% accuracy
Flow Control
Automatic
Calibration
NIST traceable
Sample Probe/Tubing
Isokinetic sampling probe
Sampling Modes
Manual, automatic, beep, cumulative/differential count, or concentration
Sampling Time
1 second to 99 hours
Sampling Frequency
1 to 9999 cycles or continuous
Sample Output
Internal HEPA filter
Vacuum Source
Internal pump with patented* flow control technology
1
A P P E N D I X A
Specifications
All specifications meet or exceed ISO 21501-4 and JIS B9921 and are subject to change without notice.
Meets all ISO 21501-4 specifications for ISO 14644-1 Class 3 and higher cleanrooms
*
Patent Number 6,167,107
A-1
Specification
Description
Communication Mode
Modbus® TCP over Ethernet (TCP/IP) or USB output
Data Storage
10,000 sample records
Data Security
Password protected
Alarm/Status
Audible alarm sounds for counts over threshold values set in recipe and for instrument or flow errors also
Alarm
This connector provides two pins for a contact closure to control an external alarm. The contact closure is normally open and rated for 0 to 60 V AC/DC at 1.5 A peak, 0.5 A continuous. The “alarm out” is rated for 60 V insulation. The contact is closed under alarm conditions determined by the programming of the device.
Display
VGA 5.7-inch touch screen with Windows® CE
Languages
English, German, French, Spanish, Japanese, Chinese (simplified), Italian
Reports
On screen viewable and printable ISO 14644-1, FS-209E & EU GMP
Printer
Built-in thermal printer (also available without printer)
External Surface
Stainless Steel
AC Power (power to AC adapter)
110 to 240 VAC universal power supply
DC Power (power to instrument)
24 VDC @ 3.0 A Battery
Removable/rechargeable Li-Ion (up to 2)
Battery Life
9310/9510: Up to 7 hours of continuous use 9350/9550: Up to 4 hours of continuous use 9500: Up to 2 hours of continuous use
Dimensions (L x W x H)
24.1 x 23.0 x 23.2 (9.5 x 9.0 x 9.2)
Standards
CE, JIS B 9921, ISO 21501-4 as listed above
Weight
5.8 kg (12.8 lbs) with battery, 5.4 kg (11.8 lbs) without battery
Warranty
2 years. Extended warranties available
Operating Conditions
2 to 35°C (36° to 95°F); 20% to 95% non-condensing relative humidity
Storage Conditions
0 to 50°C (32 to 122°F); Up to 98% non-condensing relative humidity
Included Accessories
Operating manual on CD, power cord, battery, isokinetic probe, 3 m (10 ft) tubing, purge filter, printer paper, USB cable, and TrakPro Lite Secure data download software
Optional Accessories
Electronic filter scanning probe, basic filter scanning probe, TSI velocity probes, Temp/RH probe, additional battery, dual port external battery charger, isokinetic probes, sample tubing, carrying case
A-2 AeroTrak® Portable Airborne Particle Counter
Compliance
CE Marking
EN61326 / EN 55011, Class BA: Radiated Emissions EN61326 / EN 55011, Class BA: Conducted Emissions EN61000-3-2: Harmonics EN61000-3-3: Voltage Fluctuations EN61000-4-2: Electrostatic Discharge Immunity EN61000-4-3: Electromagnetic Field Immunity EN61000-4-4: Burst Immunity EN61000-4-6: Conducted PS Immunity EN61000-4-5: Surge Immunity EN61000-4-8: Rated Power-Frequency Field Immunity EN61000-4-11: Voltage Dips\Short Interruptions Immunity
RoHS Marking
Yes
Laser Safety
Complies with 21 CFR 1040.10 and 1040.11
Specifications A-3
Dimensional Diagram
Dimensions are given in millimeters with inch equivalents in parenthesis.
A-4 AeroTrak® Portable Airborne Particle Counter
Index
2
24 hour, 3-27 2-pin connector, 1-3
A
AC power, A-2 accessories
included, A-2
optional, 1-4, A-2 add zone, 3-9 air flow, 3-8, 3-41 air velocity, 3-6 alarm, 2-3, 2-8, 2-10, 3-32, A-2 alarm accessories, 1-3 alarm closure, 2-8 alarm out, 2-3, 2-8, A-2 alarm out connector
warning, 2-9 alarm output function
warning, 2-8 alarm switch
caution, 2-8 area, 3-8 automatic, 3-13 automatic printing, 3-23
B
barcode scanner, 3-5 basic filter scanning probe, 1-4 battery
does not charge, 6-2
life, A-2
low error, 6-2
specifications, A-2 battery access, 2-2 battery low, 3-3 battery pack, 1-2 battery, warning, 2-5 beep, 3-13 bin size, 3-2, 3-6 both, 3-33
C
calibration, A-1 calibration certificate, 1-3 Calibration label, viii cancel printing button, 3-39 carry case
optional, 1-5 carry case, heavy duty
optional, 1-5
caution
alarm switch, 2-8 description, ix description of symbol, ix
symbol, ix CE marking, A-3 change setup, 3-19 change setup password
confirm new password, 3-19
new password, 3-19
old password, 3-19 channel sizes, A-1 class, 3-8 class 1 laser product label, viii class level, 3-41 cleaning instrument enclosure, 5-1 clear button, 3-42 clear samples screen, 3-24 communication mode, A-2 communications, 4-1 communications screen, 3-28
default gateway, 3-29
IP address, 3-29
subnet mask, 3-29
use dhcp, 3-29 compliance, A-3 computer cable, 1-3 concentration limits, A-1 configuration screen, 3-20
relay trigger, 3-21
trigger delay, 3-21
volume, 3-20
zoom, 3-20 confirm new password, 3-18, 3-19 contacting TSI, 7-1
email address, iii counting efficiency, A-1 counts are too low, 6-1 cumulative, 3-22 cumulative concentration, 3-31 customer service, 7-1
D
data handling, 4-1, 4-2 data security, A-2 data storage, A-2 data tab, 3-31
alarm, 3-32
cumulative concentration, 3-31
date, 3-32
differential concentration, 3-31
export data, 3-31
humidity, 3-32
data tab (continued)
laser, 3-32 location, 3-31 print data, 3-31 record, 3-32 records, 3-32 sample, 3-32 size, 3-31 temperature, 3-32 time, 3-32 vol, 3-32
zone, 3-31 date, 3-27, 3-32 date and time screen, 3-27
24 hour, 3-27
date, 3-27
time, 3-27 DC power, A-2 default gateway, 3-29 delay, 3-4, 3-15 delete zone, 3-9 detector warning, 6-2 device setup, 3-7 device setup screen, 3-26 differential, 3-22 differential concentration, 3-31 dimensional diagram, A-4 dimensions, A-2 display, A-2
screen alignment, 3-28 display screen, 3-28 dual battery charger
optional, 1-4
E
edit zone, 3-10 edit zone screen, 3-10 electronic filter scanning probe, 1-4 English only printing, 3-23 environment screen, 3-30 environment setup, 3-7 ESD protection, ix Ethernet communications, 4-2 Ethernet port, 2-3, 2-11, 4-1 European symbol for non-disposable
item, viii exclude samples screen, 3-42 exclude samples button, 3-41 exclude samples form, 3-41 export data, 3-31, 3-33 export data screen, 3-33
both, 3-33
completed successfully, 3-36
Index-1
export data screen (continued)
exporting recipes, 3-35 secured file, 3-33 select range of data, 3-35 select zone data for export, 3-34
unsecured file, 3-33 external alarm, A-2 external surface, A-2
F
factory cleaning and calibration, 5-1 ferrite, 1-3 filter samples, 3-37 filter scanner, 2-3, 2-8, 2-9 flow control, A-1 flow error, 6-2 flow rate, 1-1, A-1 formats, 3-29
G
generate, 3-41 getting help, x getting started, 2-1
H
help, x, 7-1 HEPA zero filter assembly, 1-3 hold, 3-15 humidity, 3-32
I–J
info, 3-7 information screen, 3-26 installing lithium-ion battery, 2-5 instrument description, 2-2 instrument does not meet zero count
specification, 6-1 instrument does not turn on, 6-1 instrument enclosure
cleaning, 5-1 instrument inlet contaminated, 6-1 integral thermal printer, 2-7 international contacts, 7-1 introduction, 1-1 IP address, 3-29 ISO 14644-1
1999, 3-41
2015, 3-41 isokinetic probe, 1-2
optional, 1-4
using, 2-12
K
keyboard, 3-2 keypad, 3-2
L
labels, viii language, 3-29 languages, A-2 largest particle size to consider, 3-8
laser, 3-32 laser power warning, 6-2 laser radiation label, viii laser radiation symbol label, viii laser safety, vii, A-3 last calibration, 3-22 light source, A-1 lithium-ion battery, 2-5 location, 3-4, 3-6, 3-31 location icon, 3-3 locations, 3-8 locations screen, 3-9, 3-11 locations tab, 3-9
M
main tab, 3-2 maintenance, 5-1
cleaning instrument enclosure, 5-1 factory cleaning and calibration, 5-1 schedule, 5-1
zero check, 5-1 manual, 3-13 manual history, ii manual/automatic beep, 3-6 manual/automatic/beep, 3-4 model name, 3-22
N
name, 3-8 new password, 3-18, 3-19
O
old password, 3-18, 3-19 on alarm, 3-23 on sample, 3-23 operating conditions, A-2 operation, 3-1 optional accessories, 1-4
P–Q
packing instructions, 7-3 particle count/concentration, 3-2, 3-6 parts list, 1-2 password
changing, 3-19 peripherals, 2-8 power, 2-4
specifications, A-2
using AC, 2-6 power connection, 2-2 power cord, 1-2 power entry, 2-6 power on password
confirm new password, 3-18
new password, 3-18
old password, 3-18 power on password screen, 3-17 power supply, 1-2 power switch, 3-1
print
cancel button, 3-39 icon, 3-4
schedule screen, 3-23 print by record, 3-37 print by zone
form, 3-37 print data, 3-31, 3-37
button, 3-37 print schedule screen, 3-23
automatic printing, 3-23
on alarm, 3-23
on sample, 3-23 print setup screen, 3-22
cumulative, 3-22
differential, 3-22
last calibration, 3-22
model name, 3-22
separator, 3-22
serial number, 3-22
statistics, 3-22
statistics only, 3-22 print summary, 3-38 printer
specifications, A-2 printer icon, 3-1 printer paper
rolls, 1-4 probe, 2-3, 2-8, 2-9 product registration, ii pump exhaust diffuser, 2-2 purge, 3-7 purge screen, 3-25
R
recipe, 3-8, 3-41 recipe screen, 3-11 recipes screen
channels T tab, 3-15
recipe tab, 3-12
timing tab, 3-14
cycles, 3-15 delay, 3-15 hold, 3-15 sample time, 3-15
volume, 3-15 recipes setup, 3-7 recipes setup screen, 3-12
automatic, 3-13 beep, 3-13
manual, 3-13 records, 3-32 recs, 3-4 regional screen, 3-29
formats, 3-29
language, 3-29 relative humidity, 3-2, 3-6 relay trigger, 3-21 reports, A-2 reports tab
air flow, 3-41
class level, 3-41
Index-2 AeroTrak® Portable Airborne Particle Counter
reports tab (continued)
exclude samples button, 3-41 generate, 3-41 recipe, 3-41 room area, 3-41 use end date, 3-41
use start date, 3-41 returning for service, 7-3 RoHS marking, A-3 room area, 3-41
S
safety, vii sample, 3-4, 3-15, 3-32 sample output, A-1 sample probe/tubing, A-1 sample time, 3-15 sample tubing, 1-2 sampling frequency, A-1 sampling modes, A-1 sampling time, A-1 save, 3-10 save button, 3-42 save new zone, 3-10 sconfiguration screen
store partial samples, 3-20 screen alignment, 3-28 screen layout, 3-2 secured file, 3-33 select by record, 3-34
button, 3-38 select by zone, 3-35, 3-38 select range, 3-35, 3-38 select zone data to print screen, 3-37 separator, 3-22 serial number, 3-22 serial number label, viii service
international contacts, 7-1 setup tab, 3-7
device setup, 3-7
environment setup, 3-7
info, 3-7
purge, 3-7
recipes setup, 3-7
system setup, 3-7
zones setup, 3-7 size, 3-31 size range, 1-1, A-1 software
installation, 4-2 software input panel, 3-2 specifications, A-1 stainless steel isokinetic probe, 1-4 standard, 3-8 standards, A-2 start purge, 3-25 start/stop buttons, 3-4, 3-6 statistics, 3-22 statistics only, 3-22 status, 3-8 stop purge, 3-25
storage conditions, A-2 store partial samples, 3-20 stylus, 1-3, 2-6 subnet mask, 3-29 superthane tubing, 1-4 system setup, 3-7 system setup screen, 3-17
T
TCP/IP, 3-29 technical contacts, 7-1 technical support
international, 7-2 temperature, 3-2, 3-6, 3-32 thermal printer, 2-7
feed button, 2-7 time, 3-4, 3-27, 3-32 TRAKPRO Lite CD, 1-3 trigger delay, 3-21 tripod mount, 2-2 troubleshooting, 6-1
U
unpacking, 1-2 unsecured file, 3-33 USB A, 2-3, 2-8, 2-9, 4-1 USB B, 2-3, 2-11, 4-1 USB communication, 4-1 USB computer communication, 4-2 USB data download, 4-1 use dhcp, 3-29 use end date, 3-34, 3-37, 3-40, 3-41 use start date, 3-34, 3-37, 3-40, 3-41
V
vacuum source, A-1 velocity/temperature probe, 1-5 velocity/temperature/relative humidity
probe, 1-4 vol, 3-32 volume, 3-4, 3-15, 3-20
W–X–Y
warning
alarm out connector, 2-9
alarm output function, 2-8
battery, 2-5
description, ix
description of symbol, ix
hazardous optical radiation, vii
symbol, ix warranty, iii, A-2 weight, A-2
Z
zero check, 5-1 zero count
level, A-1 zone, 3-4, 3-31 zone icon, 3-3
zone setup
add zone, 3-9 air flow, 3-8 area, 3-8 class, 3-8 delete zone, 3-9 edit zone, 3-10 locations, 3-8 name, 3-8 recipe, 3-8 standard, 3-8
status, 3-8 zone setup screen, 3-8 zones screen, 3-9 zones setup, 3-7 zoom, 3-20 zoomed data screen, 3-6
Index Index-3
Loading...