TSI Instruments 854051, 854050 Quick Start Manual

C a u t i o n
Refer to the Manufacturer’s Manual and follow all instructions regarding; setup, installation, proper use, maintenance, safety, and troubleshooting.
Qty.
Item Description
Reference Picture
1
Lufft WS300/WS500 Metrology Station
1
Lufft Cable
1
Cable Gland
Qty.
Item Description
Reference Picture
1
Jam Nut
1
Communication Connector (3-pin)
www.phoenixcontact.com
connector - MSTB 2,5/ 3­ST-5,08
P/N 1757022
1
DIN Rail Power Connector
www.phoenixcontact.com
Plug - SPDB 2,5/ 2
P/N 3040410
Small Flat Blade Screw
Driver (not included)
14-mm Wrench
INSTALLING THE 854050/854051
INTO ENVIRONMENTAL ENCLOSURE
This guide will help you quickly attach the 10 m connection cable included with the Lufft WS300 (TSI P/N 854050)/ Lufft WS500 (TSI P/N 854051) to the DustTrak Environmental Monitor.
QUICK START GUIDE
METROLOGY STATION
Unpacking the (854050 /
854051) Metrology Station
1. Unpack and verify that all the items listed in the following table are present.
2. Contact TSI immediately if items are missing or broken.

Tools Needed for Installation

_______________ TSI and TSI logo are registered trademarks of TSI Incorporated. DustTrak is a trademark of TSI Incorporated. Thiamis is a trademark of Netronix Inc.
NO T E
Disconnect power from the instrument. The instrument SHOULD NOT have power during this installation procedure.
1. Remove the hole plug in the enclosure and install the cable gland and jam nut.
2. One end of the 10 m connection cable for the metrology station has been stripped back and ends added for connections inside the enclosure.
3. Insert the wire end of the cable through the gland. Tighten (hand tight) the gland around the cable to prevent leakage.
4. Insert the red and black wires into the connector as shown.
5. Attach RS-485/SDI-12 Communication Wires to green connector as shown. A small flat blade screw driver can be used to tighten the connector to the wires.
6. Insert the green connector into the Thiamis 1000 Node RS-485 socket.
7. Install the connector as shown to the DIN rail inside the enclosure.

Installation

-2-
8. Attach the other end of the cable to weather station. Tighten (hand tight) with a 14-mm open ended wrench (see figure to right).
9. See the weather station manual for mounting instructions and specifications.
10. Reapply power to the instrument.
11. Environet can be used to pair the weather station to the Thiamis 1000.
a. Click Add Sensor in the device configuration. b. Select Lufft from the pull-down menu and
click Save.