TSI Instruments 3302A, 3340 User Manual

3302A Aerosol Diluter
3340 Laser Aerosol Spectrometer
ISOKINETIC COUPLER FOR THE
MODEL 3302A AEROSOL DILUTER
AND MODEL 3340
LASER AEROSOL SPECTROMETER
USERS GUIDE

Description

Operation

1. Install the 1:100 capillary into the 3302A Aerosol Diluter.
2. Connect the 1140001 Isokinetic Coupler to the bottom outlet port of the 3302A.
3. Place the assembly on top of the 3340 using a spacer or stand (not included). The tube from the 3340 inlet to the Isokinetic Coupler should be straight.
4. Connect and switch on the external pump (e.g. TSI model 3032).
5. Set the LAS sampling flow to 0.10 L/min.
6. Adjust the pressure across the capillary (p supplied with the instrument.
) in the 3302A according to the calibration sheet
aerosol
USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 4 91 11 87 64 Germany Tel: +49 241 523030
India Tel: +91 80 67877200 China Tel: +86 10 8219 7688 Singapore Tel: +65 6595 6388
3302A Aerosol Diluter
Coupler (114001), connectors, filter, orifices, and connection to pump
Aerosol Diluter
Stand (Not Included)
3340 Laser Aerosol Spectrometer

Schematic

Parts List

Isokinetic Coupler (P/N 1140001) includes:
Connector fitting with isokinetic sampling probe  Critical orifice to set auxiliary flow  Absolute filter to protect orifices  Tubing

Accessories Needed

External pump to maintain main flow through coupler (e.g., TSI model 3032)  Stand to place 3302A on top of LAS (if not mounted in test rig)