TSI Incorporated Alnor LoFlo Balometer Owner's Manual

OWNER’S MANUAL
LoFlo
®
Balometer
Seller warrants the goods sold hereunder, under normal use and service as described in the operator's manual, shall be free from defects in workmanship and material for twenty-four (24) months, or the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions:
a. Hot-wire or hot-film sensors used with research
anemometers, and certain other components when indicated in specifications, are warranted for 90 days from the date of shipment.
b. Parts repaired or replaced as a result of repair
services are warranted to be free from defects in workmanship and material, under normal use, for 90 days from the date of shipment.
c. Seller does not provide any warranty on finished
goods manufactured by others or on any fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies.
d. Unless specifically authorized in a separate
writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO
OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call Customer Service at (800) 424-7427 (USA) and (1) 651-490-2811 (International).
.
1

TABLE OF CONTENTS

Section 1 General Description.........................................................2
Section 2 Theory of Operation........................................................2
Section 3 Safety...............................................................................3
Section 4 Instrument Setup and Storage..........................................3
Section 5 About the Instrument Display .........................................5
Section 6 Getting Started.................................................................6
Section 7 Detailed Operation ..........................................................6
Section 8 Troubleshooting ..............................................................9
Section 9 Maintenance..................................................................10
Section 10 Service Information.......................................................11
Appendix A Correction Factors for Standard
Flow Rate to Actual Flow Rate.....................................12
Appendix B Model 6200 LoFlo Balometer Field Setting Instruction for User Programmable
Correction Factors.........................................................13
1
SECTION 1
General Description
The Alnor LoFlo Balometer is a compact instrument that measures very low air flow rates in HVAC systems. It measures from 10–500 cfm (17–850 m
4.7–236 l/s). The LoFlo Balometer displays standard air volume rate when placed at supply or return diffusers, registers, or grilles located on the ceiling, wall, or floor.
The LoFlo Balometer has a vent mechanism which reduces the restriction of air flow caused by the instrument. This unique venting system (patent pending) allows the instrument to remain compact while still measuring a wide range of volume rates. The
3
/h,
LoFlo Balometer should be used with the vents closed at lower volume rates, but higher volume rates should be measured with the vents open to minimize any resistance effects. A 2-point measurement may be taken to compensate for the resistance effect of the instrument at higher volume rates.
The LoFlo Balometer is battery powered and may be used with or without optional hoods. It weighs just over 6 pounds with the 2' x 2' hood, helping to reduce worker fatigue.
SECTION 2
Theory of Operation: Capture Hoods
and Thermo-anemometry Sensing
As air exits a diffuser, it is captured in the hood or base and directed over a manifold. The manifold averages the velocity pressure and directs air over the sensors. The velocity and temperature of the air passing through the manifold are sensed, using a constant temperature thermo-anemometer principle. The voltage output from the sensor is then calibrated for the entire volume rate passing through the LoFlo Balometer.
All capture hoods are susceptible to “odd” air flow patterns. “Odd” could refer to any air flow pattern different from the pattern where it was calibrated. Substantial negative effects can be observed when
using a large hood on a small diffuser. For example, using a 2' x 2' hood on a 10" x 10" diffuser. This creates large recirculation regions on the sides of the fabric hood and causes an “odd” air flow pattern as it passes over the manifold. It is recommended to match the hood closely to the size of the diffuser.
3
! NOTE: Throughout this manual, the units cfm, m
and l/s are used for simplicity. Note, however, that the LoFlo Balometer actually reads in standard cubic feet per minute (scfm), standard cubic meters per hour (std. m (std. l/s).
3
/h), or standard liters per second
/h
2
3
SECTION 3
Safety
When using the LoFlo Balometer to check air flow at ceiling diffusers, make certain that you can safely raise and hold the unit while making measurements. Be especially careful when working on a ladder.
Observe all necessary precautions so that the unit does not become caught in moving machinery or touch any exposed electrical wiring.
SECTION 4
Instrument Setup and Storage

Hood Setup

1. Locate the four (4) aluminum channels for the hood and position them into a square with the foam facing up. Slide the “U” opening on the channel end (Fig. 1) and corner bracket toward the brass eyelets until you feel it latch.
2. Continue with the o ther two channels until you have a square frame that is locked tightly in all four corners.
The LoFlo Balometer is not designed for gas mixtures other than air. Use with corrosive or other dangerous or explosive gas mixtures is not recommended and is at the user’s risk.
8. Locate the hood support rods. Extend the hood so it is taut and insert one support rod onto the spring­loaded pin closest to you.
9. Push the support rod down and insert the top end of the rod into the metal cup on the hood frame. Fig 1.
10. Install the rod on the opposite side, then install the last two rods.
Foam
3. Locate the hood for the frame size you have assembled.
4. Position a corner seam of the hood onto the corner of the frame. Stretch the nylon hood to an adjacent corner until the hood corner meets the frame corner.
5. Push the elastic cord into the channel along the side of the frame.
6. Repeat steps 4 and 5 until the hood assembly is complete and ready to mount onto the base.
7. Place hood on a table or clean floor with the base inside the opening. Pull the hood over the base. Rotate the hood assembly until the metal cups are above the spring-loaded pins.
“U” Opening
Metal Cup
Figure 1
Hood Frame Channel
3

Intrument Storage

The hood and frame assemblies can be detached from the instrument base and placed inside their own carrying case without disassembly.
The LoFlo Balometer should always be turned off before re-packing the instrument. If storing the LoFlo Balometer for an extended period of time, remove the batteries to prevent damage due to leakage.

Battery Installation

To install the 4 C-size batteries that ship with the LoFlo Balometer:
1. Make sure the LoFlo switch is off.
2. Remove the battery cover located on top of the
meter.
3. Slide the battery holder up until the snap connector
is exposed.
4. Disconnect the battery holder from the LoFlo
Balometer.
3. Reconnect the fully charged battery holder to the LoFlo Balometer and slide battery holder into the battery compartment.
4. Replace battery cover.
Do not try to charge alkaline batteries.
5. Install the 4 C size batteries in the holder.
6. Reconnect the snap connector and slide the battery holder into the battery compartment.
7. Replace battery cover.
To use and install rechargeable batteries:
! NOTE: As an alternative to alkaline batteries, four
Nickel Cadmium (NiCd) rechargeable batteries are available from TSI for use in the LoFlo Balometer.
1. Connect the battery holder loaded with rechargeable NiCd batteries to the charger and charge for 12–14 hours.
2. Turn the LoFlo off.
Figure 2
Battery Holder
Do not use the charger as a battery substitute. Damage to the instrument will result.
4
SECTION 5
About the Instrument Display
This display is an analog display, simulated with 26 LCD segments to give the user the response of a mechanical Balometer. It also displays an exact digital number, using 3½ digit numeric section. The LCD has indicators for supply, return, manual range, vent modes, and low battery.
Figure 3 shows the front of the LoFlo Balometer.
Index No. Name
1 LCD 2 On/Off Switch 3 Supply/Return Button 4 Manual Range Button 5 Handle Button 6 Battery Holder 7 Vents 8 Vent Open/Close Slide
5
Figure 4 shows the LCD.
Index No. Name
1 Digital Readout 2 Simulated Analog Display 3 Manual Range Indicator 4 Supply/Return Indicator 5 2-Point Mode Indicator 6 Vent Indicator 7 Units Indicator 8 Low Battery Indicator
Figure 3
Front View of LoFlo Balometer
Figure 4
LoFlo Balometer LCD
5
SECTION 6
Getting Started
! NOTE: The instrument should be allowed to
stabilize to room conditions in order to achieve specified accuracy.
1. If you are using a hood with the unit, assemble
the LoFlo Balometer as described in Section 4, “Instrument Setup and Storage.”
2. Turn on the unit using the slide switch located on
the right side of the meter.
3. When turned on, the LoFlo Balometer displays
the vent closed indicator. Close the vent on the base by turning the top baffle counter clockwise.
4. Select the proper mode for air flow direction,
supply or return.
5. Place the LoFlo over the diffuser or grille to be
measured.
SECTION 7
Detailed Operation
6. The auto range feature of the LoFlo Balometer will choose the correct scale and display the volume rate in the digital readout section. A quick press on the handle switch will freeze the display.
7. Another quick press will unfreeze the display.
3
8. For volume rates above 150 cfm (255 m l/s), it is recommended to use the vent open mode to minimize air flow resistance effects. See Section 7, “Detailed Operation.”
! NOTE: Make sure that there are no gaps between
the LoFlo and the surface around the diffuser or grille. Also, make sure that your body or other objects are not blocking or diverting the air flow where the air is exiting (supply) or entering (return). This will affect the reading.
/h, 71

Handle Button

In 1-point mode, the handle button will freeze the digital reading and flash the value when it is pressed for less than three seconds. When the handle button is pressed for more than three seconds, it switches the LoFlo Balometer between vent closed mode and vent open mode.
In 2-point mode, the handle button cycles the instrument through the 2-point measurement process. See 2-point mode section for detail.

Auto/Manual Ranging Button

When turned on, the instrument is in auto-range mode. The user can switch to manual range mode by pressing the range button. The manual range indicator will show. By pressing the range button, the meter display switches to the next volume range. To return to auto­range mode, press and hold the range switch until the manual range indicator disappears.

Changing Measurement Units

To change measurement units:
1. Turn the instrument off.
2. Press and hold the supply/return button and switch
the instrument on at the same time.
3. The display will flash one of the three available
measurement units: cfm, m
4. Press and hold the supply/return button or the
range button to scroll through the measurement units.
5. When the units you desire are displayed, switch the
instrument off.
6. When you turn the instrument on again, your
chosen units will be displayed.
3
/h, l/s.
6

Supply/Return Button

The LoFlo Balometer is in supply mode when turned on. This is indicated on the display with the supply indicator
. To take return air measurements, press the
supply/return button on the front of the meter. Return air mode will be indicated on the display with the return indicator
.
Significant errors may result if:
Instrument is not in proper mode.
Fabric hood is not used with vent open
mode.

Vent Closed/Vent Open Modes

When the instrument is turned on, it is in the vent closed mode the vents must be closed to take the measurement. Close the vents by loosening the thumbscrew and sliding the vent baffle to the fully closed position and locking the vent in place again.
The vent open mode is indicated with the vent open indicator on the LCD the thumb screw and sliding the vent baffle to the fully open position and locking it into place again. Then press and hold the handle switch for more than 3 seconds to switch the LoFlo Balometer to the open vent mode.
. When the vent closed indicator is on,
. Open the vents by loosening
7
3
10–150 cfm (17–255 m
! NOTE: In Vent Closed Mode, volume rates below
8 cfm (13 m
3
/h, 3.6 l/s) are displayed as 0; volume
rates above 517 cfm (879 m
/h, 4.7–71 l/s).
3
/hr, 244 l/s) are
displayed as Or.
Use Vent Open Mode for measurements between:
3
150–500 cfm (255–850 m
! NOTE: In Vent Open Mode, volume rates below
150 cfm (255 m
3
volume rates above 517 cfm (879 m
/h, 71–236 l/s)
/h, 71 l/s) are displayed as Ur;
3
/h, 244 l/s) are
displayed as Or.

2-Point Measurement Mode—Supply Only

To activate the 2-point measurement mode, turn the instrument off. While pressing the handle button, turn the instrument back on. The 2-point mode indicator will show.
The 2-point measurement mode takes both vent open and vent closed measurements to calculate a resistance­compensated volume flow rate.
The process starts with the vents closed and the display showing the vent closed indicator.
1. Place the LoFlo Balometer over the diffuser to be measured and press the handle switch once to take the vent closed measurement.
2. The instrument stores that reading.
3. The display will now show the open vent indicator and you should open the vents.
The vent open mode was developed for higher volume rates on larger diffusers; the 2' x 2' or 650 mm x 650 mm hood should always be used with vent open mode.

1-Point Measurement Mode

The default mode of the LoFlo Balometer is the 1-point measurement mode. If you are in 2-point measurement mode, simply turn the instrument off and turn back on.
The 1-point measurement mode is described in the Getting Started section earlier in this manual. This is the faster, simpler way to take measurements.
Use Vent Closed Mode for measurements between:
! NOTE: Make certain you are using a fabric hood
when in 2-point mode.
4. Again place the LoFlo Balometer over the diffuser to be measured and press the handle switch to take the vent open measurement.
5. The instrument calculates and displays the compensated reading.
6. The numeric display will flash; press the handle switch again and the instrument will display the vent closed indicator. Close the vents.
7. Return to Step 1 to take another resistance compensated reading.
! NOTE: Return to 1- point measurement mode at
any time by turning the power off and on.
7
When “–––” is displayed at the last step of the 2-point mode, the calculation for the resistance effect cannot be made. It can be caused by any of the conditions listed here.
Condition Reason
The vent closed or vent open measurement is less than 150 cfm (70 l/s, 255 m
The vent closed or vent open measurement is greater than 500 cfm (235 l/s, 850 m
The vent open measurement minus the vent closed measurement is too small (< 0).
3
/h).
3
/h).
Calculations are not made at the low end since the
effect is small and can generally be neglected. Use 1­point mode.
Calculations are not made beyond the calibrated range.
Use the Standard or Electronic Balometer (Range: 50– 2000 cfm), or perform a traverse of the system instead.
A mistake has occured in the process. Repeat the two-
point measurement.
The resistance effect is small and was overwhelmed
by natural fluctuations. Use single-point mode.
The vent open measurement minus the vent closed measurement is too large.
A mistake has occurred in the process. The vent may
not have been open for the open vent measurement. Repeat the two-point measurement.
The effect is beyond the range of the compensation
equation. Perform a traverse of the system instead.
! NOTE: The display will also show “–––” when the return mode is selected. The 2-point measurement mode is for
supply only.
8
9
SECTION 8
Troubleshooting
Symptom Possible Cause and Corrective Action
Meter does not turn on.
Meter reading lower than expected.
Battery holder snap connector not connected. Connect
battery holder to instrument.
Batteries may be discharged. Charge or replace them.
Switch failure, wiring failure, or circuit failure. Call
TSI.
Wrong vent mode chosen. Choose correct vent mode,
open or closed.
Wrong air direction mode chosen. Choose correct
direction, supply or return.
Back pressure effects could be significant. Use 2-
point measurement mode.
Hood frame not sealing properly around diffuser or
grill. Press hood evenly against diffuser.
Meter reading higher than expected.
Meter not at zero at zero flow.
Meter displays Er. Handle button not working.
Odd air flow pattern present. Perform traverse and use
proportional balancing.
Hood torn. Replace, or repair tear with duct tape or
other non-porous material.
Meter out of calibration. Call TSI.
Damage to manifold. Call TSI.
Wrong vent mode chosen. Choose correct vent mode,
open or closed.
Wrong air direction mode chosen. Choose correct
direction, supply or return.
Odd air flow pattern present. Perform traverse and use
proportional balancing.
Meter out of calibration. Call TSI.
Instrument not at room temperature. Allow instrument
to acclimate before turning on the power.
The LoFlo is reading real room air currents. Place on
floor with a piece of cardboard over the top to determine if it is measuring real air currents.
Meter out of calibration. Call TSI.
Self test failed. Call TSI.
Call TSI.
9
SECTION 9
Maintenance

Hoods

Hoods may be cleaned in cool water with a mild detergent.

Sensing Manifold

The manifold should be checked visually before using to be certain that the sensing holes have not become clogged with dirt or dust particles. Do not immerse the manifold in water. It is recommended that cleaning be done with extreme care with the manifold in place. Do not use highly compressed air to clean the manifold tubes.

Calibration

It is recommended that your LoFlo Balometer be returned to the factory once a year for a calibration check. When shipping the LoFlo Balometer for factory calibration, pack it carefully, and follow the Instructions for Return in this manual.

Vent Operation

If the vent becomes difficult to operate, remove both thumbscrews from the vent mechanism and spray the contacting surfaces with silicone spray lubricant. Wipe off excess and reassemble vent mechanism.

Performance Check

If the calibration is to be checked, the best way is to use a reference flow standard more accurate than the LoFlo Balometer. Flow standards that may be used for this purpose include orifice plates, venturis, nozzles, and laminar flow elements. These devices require accurate differential pressure, barometric pressure, and temperature reading instruments to measure either actual or standard volume rate.
If the LoFlo Balometer is checked against a velocity instrument such as a pitot probe/manometer or thermo­anemometer, there may be errors. If a velocity standard is used, the average velocity must be obtained by taking a traverse.* The accuracy of the traverse is dependent on the flow uniformity, the number of readings taken, the accuracy of the velocity instrument, and operator expertise. This average velocity reading must then be multiplied by the area over which the traverse was taken.
Finally, air flow instrumentation is dependent on environmental conditions such as temperature, atmospheric pressure, humidity, and even turbulence. These conditions can have very different effects on various instrument types. Caution must be exercised when making comparisons.
*A traverse is a set of velocity readings taken in a prescribed pattern which will provide an overall velocity value when averaged together. Any of the Alnor velocity measuring instruments can be used for this purpose.
10
11
SECTION 10
Service Information

Service and Repair

Please return your Product Registration Card immediately. This allows us to send you service reminders, special offers, and important information about your product.
Before sending your instrument for calibration or repair, you should call TSI Customer Service. The Service Department will provide you with the cost of service or calibration, Return Material Authorization (RMA) number, and shipping instructions.
Please have the following information available when you call:
! Owner’s name, address, and phone number ! Billing address, if different and applicable ! Instrument name or model ! Serial number ! Date of purchase ! Where purchased
TSI recommends that you keep a “calibration log” and keep all records of service on your instrument.

Instructions for Return

Include the Purchase Order showing instrument model number, cost of service and/or calibration, and the RMA number. Mark the outside of the shipping container with the RMA number. This will expedite processing of your instrument when we receive it.

Damaged in Transit

All orders are carefully packed for shipment. On receipt, if the shipping container appears to have been damaged during shipment, the instrument should be thoroughly inspected. The delivering carrier’s papers should be signed noting the apparent damage. DO NOT DISCARD THE BOX.
If the instrument itself has been damaged, a claim should be promptly filed against the carrier by the customer. The selling agent will assist the customer by supplying all pertinent shipping information; however, the claim must be filed by the insured. If the instrument is damaged beyond use, a new order should be placed with TSI while awaiting reimbursement from the carrier for the damaged instrument.
Call TSI directly for assistance if necessary.
Send the instrument prepaid. Securely package your instrument in a strong container surrounded by at least two inches (5 cm) of suitable shock-absorbing material.
11
APPENDIX A
Correction Factors for Standard Flow Rate to Actual Flow Rate
The LoFlo Balometer measures standard volumetric flow rate. Standard flow is defined as the flow rate at standard conditions, 70°F (21.1°C) and 14.7 psia (29.92 inHg, 760 mmHg). Actual flow rate is the true volumetric rate of air at the local temperature and barometric pressure. If you desire standard flow rate, use the displayed reading. However, if the actual flow rate is desired, use the information below.
To correct the standard flow rate to actual flow rate use the equation below.
V
= V
act
V
= Actual flow rate
act
V
= Standard flow rate (this is displayed by
std
std
x CF
the instrument)
T
= Actual Temperature
act
ρ
= Actual barometric pressure
act
3
ρ
= Standard density 0.075 lb/ft
std
(at 70°F and 14.7 psia)
ρa = Actual density in lb/ft
3
CF = ρ
(Correction Factor)
std/ρa
Using the equations or chart below, you can bypass the full density calculation.
Calculations in imperial units
CF = (14.7/ ρ Where ρ
is in psia and T
act
)*(460+ T
act
act
is in °F
act
)/530
Calculations in imperial units
CF = (29.92/ ρ Where ρ
is in inHg and T
act
)*(460+ T
act
act is in °F
act
)/530
Calculations in metric units
CF = (760/ ρ Where ρ
act
)*(273.15+ T
act
is in mmHg and T
act
is in °C
act
)/294.25
For your convenience, we have made a chart for determining the correction factor given temperature and pressure.
! NOTE: The atmospheric pressure as reported by
the National Weather Service is corrected to sea level and cannot be used if measurements are not at sea level.
12
Loading...
+ 32 hidden pages