TSI Incorporated Alnor CF920, Airflow IAQ920 Operation And Service Manual

Indoor Air Quality

Indoor Air Quality Meter

Alnor® Model CF920
AIRFLOWTM Model IAQ920
Operation and Service Manual
Copyright©
TSI Incorporated / 2007-2008 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective July 2000)
Seller warrants the goods sold hereunder, under normal use and service as described in the operator's manual, shall be free from defects in workmanship and material for twenty-four (24) months, or the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions: a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of shipment.
b. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of shipment.
c. Seller does not provide any warranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies.
d. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty
with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call Customer Service department at +44 (0) 149 4 459200 (UK), (800) 874-2811 (USA), or (1) 651­490-2811 (International).
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............
CHAPTER 2 SETTING-UP................................................................. 3
Supplying Power to the Alnor CF920/AIRFLOW IAQ920 .................3
Installing the Batteries ............................................................... 3
Using the Optional AC Adapter ................................................. 3
Using The Probe.............................................................................3
Connecting to a Computer.............................................................. 3
CHAPTER 3 OPERATION .................................................................5
Keypad Functions ........................................................................... 5
Common Terms .............................................................................. 5
Menus ............................................................................................. 6
DISPLAY SETUP................................................................................. 6
SETTINGS .........................................................................................6
BAROMETRIC PRESSURE ................................................................... 6
DATA LOGGING .................................................................................7
Measurements...........................................................................7
Log Mode/Log Settings .............................................................7
Delete Data................................................................................ 7
% Memory .................................................................................7
APPLICATIONS ..................................................................................7
CHAPTER 4 MAINTENANCE............................................................9
Recalibration...................................................................................9
Cases..............................................................................................9
Storage ...........................................................................................9
1
CHAPTER 5 TROUBLESHOOTING................................................ 11
APPENDIX A SPECIFICATIONS..................................................... 13
iii
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below.. If anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
3. USB cable
4. CD-ROM with downloading software
5. Calibration collar
6. AC adapter
1
Chapter 2
Setting-up

Supplying Power to the Alnor CF920/AIRFLOW IAQ920

The Alnor CF920/A four size AA batteries or the optional AC adapter.

Installing the Batteries

Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The Alnor CF920/A is designed to operate with either alkaline or NiMH rechargeable batteries, although it will not recharge NiMH batteries. Battery life will be shorter if NiMH batteries are used. Carbon-zinc batteries are not recommended because of the danger of battery acid leakage.

Using the Optional AC Adapter

When using the AC adapter, the batteries (if installed) will be bypassed. Be sure to provide the correct voltage and frequency, which is marked on the back of the AC adapter. The AC adapter is not a battery charger.

Using The Probe

The sensing probe relies on the diffusion of air. For best results, try to keep the sensing probe surrounded by moving air. Do not breathe on the probe. Humans exhale CO take time for the probe to re-stabilize. Use the probe holder to support the probe when in continuous data logging mode.
IRFLOW IAQ920 can be powered in one of two ways:
IRFLOW IAQ920
levels exceeding 10,000 ppm and it may
2

Connecting to a Computer

Use the USB Cable provided with the Alnor CF920/A
IRFLOW IAQ920 to
connect the instrument to a computer. Connect the end labeled “COMPUTER” to the computer USB port and the other end to the data port of the Alnor CF920/A
IRFLOW IAQ920.
For more information on how to download stored data see Chapter 3 section
LogDat2 Downloading Software.
titled
Caution: This symbol is used to indicate that the data port of the Alnor CF920/A
IRFLOW IAQ920 is not intended for
connection to a public telecommunications network. Connect the USB data port only to another USB port.
3
Chapter 3
Operation

Keypad Functions

ON/OFF Key
Arrow (ST) Keys
(Enter) Key
Arrow (g or h) and Menu Soft Keys

Common Terms

In this manual there are several terms that are used in different places. The following is a brief explanation of the meanings of those terms.
Sample
Press to turn the Alnor CF920/A on and off. During the power up sequence the display will show the following: Model Number, Serial Number, Software Revision and Last Date Calibrated.
Press to scroll through choices while setting a parameter. Pressing the will lock the keypad to prevent unauthorized adjustments to the instruments. To unlock the keypad, press the
Press to accept a value or condition.
Press arrow keys to change choices while setting a parameter. Press the Menu soft key to select the Menu selections, which are Display Setup, Settings, Barometric Pressure, Data Logging, Applications, Calibration.
Consists of all of the measurement parameters stored at the same time.
ST keys simultaneously
ST keys simultaneously.
IRFLOW IAQ920
Test ID
A group of samples. The statistics (average, minimum, maximum, and count) are calculated for each test ID. The maximum number of test IDs is
100.
5
Time Constant
The time constant is an averaging period. It is used to dampen the display. If you are experiencing fluctuating flows, a longer time constant will slow down those fluctuations. The display will update every second, but the displayed reading will be the average over the last time constant period. For example, if the time constant is 10 seconds, the display will update every second, but the displayed reading will be the average from the last 10 seconds. This is also referred to as a “moving average”.
Logging Interval
The logging interval is a frequency period that the instrument will log readings. For example, if the logging interval is set to 30 minutes, each sample will be the average of the last 30 minutes.

Menus

DISPLAY SETUP

Display setup menu is where you will setup the desired parameters to be displayed on the running screen. With a parameter highlighted you can then use the ON soft key to have it show up on the running screen or select the OFF soft key to turn off the parameter. Use PRIMARY soft key to have a parameter show up on the running screen in a larger display. Only one parameter can be selected as a primary, and up to 2 secondary parameters can be selected at one time.

SETTINGS

Settings menu is where you can set the general settings. These include Language, Beeper, Select Units, Time Constant, Contrast, Set Time, Set Date, Time Format, Date Format, Number Format, Backlight and Auto Off. Use the g
or h soft keys to scroll through the settings for each option and
use the key to accept settings.

BAROMETRIC PRESSURE

Barometric Pressure menu is where you set the barometric pressure, which can affect the CO
measurement.
2
6 Chapter 3

DATA LOGGING

Measurements

Measurements to be logged are independent of measurements on the display, and must therefore be selected under DATA LOGGING Æ Measurements.

Log Mode/Log Settings

You can set Log Mode to Manual, Auto-save, Cont-key, or Cont-time.
Manual mode does not automatically save data, but instead prompts
the user to save a sample.
In Auto-save mode, the user manually takes samples that are
automatically logged.
In Cont-key mode, the user starts taking readings and logging by
pressing the key. The instrument will continue taking measurements until the key is pressed again.
In Cont-time mode, the user starts taking readings by pressing the
key. The instrument will continue taking samples until a set period of time has passed.
Auto-save, Cont-Key and Cont-time modes have the following
additional Log Settings:
Log Settings
Mode
Auto-save Log Interval Cont-key Log Interval Cont-time Log Interval
Test Length
Pressing the
ST keys simultaneously will lock the keypad to
prevent unauthorized adjustments to the instruments. To unlock the keypad, press the
ST keys simultaneously.

Delete Data

Use this to delete all data, delete test or delete sample.

% Memory

This option displays the memory available. Delete All, under Delete Data, will clear memory and reset the memory available.

APPLICATIONS

You can choose % Outside Air in the Applications menu. After choosing this application, take measurements or enter data for each line.
Operation 7
LogDat2
The Alnor CF920/A
Downloading Software
IRFLOW IAQ920 comes with special software called
LogDat2 Downloading Software, which is designed to provide you with maximum flexibility and power. To install this software on your computer, follow the instructions on the label of the LogDat2 CD-ROM.
To download data from the Alnor CF920/A supplied computer interface USB cable to the Alnor CF920/A
IRFLOW IAQ920, connect the
IRFLOW
IAQ920 and to a computer USB port. Then run the LogDat2 downloading software. Within the LogDat2 software, either select the tests to be downloaded or double-click on a test to open it.
8 Chapter 3
Chapter 4
Maintenance
The Alnor CF920/AIRFLOW IAQ920 requires very little maintenance to keep it performing well.

Recalibration

To maintain a high degree of accuracy in your measurements, we recommend that you return your Alnor CF920/A annual recalibration. Please contact one of TSI’s offices or your local distributor to make service arrangements and to receive a Return Material Authorization (RMA) number. To fill out an online RMA form, visit TSI’s website at
U.S. & International UK
TSI Incorporated Alnor Products Stirling Road 500 Cardigan Road Cressex Business Park Shoreview MN 55126-3996 High Wycombe Tel: (800) 874-2811 Bucks (651) 490-2811 HP12 3RT United Kingdom Fax: (651) 490-3824 Tel: +44 (0) 149 4 459200 Fax: +44 (0) 149 4 459700
The Alnor CF920/A using the CALIBRATION menu. These field adjustments are intended to make minor changes in calibration to match a user's calibration standards. The field adjustment is NOT intended as a complete calibration capability. For complete, multiple-point calibration and certification, the instrument must be returned to the factory.
http://service.tsi.com.
TSI Instruments Ltd.
IRFLOW IAQ920 can also be recalibrated in the field
IRFLOW IAQ920 to TSI for

Cases

If the instrument case or storage case needs cleaning, wipe it off with a soft cloth and isopropyl alcohol or a mild detergent. Never immerse the Alnor CF920/A IAQ920 or the AC adapter becomes broken, it must be replaced immediately to prevent access to hazardous voltage.

Storage

Remove the batteries when storing the unit for more than one month to prevent damage due to battery leakage.
IRFLOW IAQ920. If the enclosure of the Alnor CF920/AIRFLOW
9
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the Alnor CF920/A IAQ920. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI.
Table 5-1: Troubleshooting the Alnor CF920/A
Symptom Possible Causes Corrective Action
No Display Unit not turned on Switch unit on. Low or dead batteries Replace batteries or plug
Dirty battery contacts Clean the battery
Humidity reading near zero or not believable High CO2 levels Breathing on probe Shield probe from breath
No response to keypad
Instrument Error message appears
Fault in instrument Factory service required
Probe exposed to intense light
Keypad locked out Unlock keypad by
Memory is full Download data if desired,
IRFLOW IAQ920
in AC adapter.
contacts. Shade the probe while taking samples.
and allow instrument to re-stabilize.
pressing simultaneously.
then DELETE ALL memory.
on instrument.
WARNING!
Remove the probe from excessive temperature immediately: excessive heat can damage the sensor. Operating temperature limits can be found in Appendix A, Specifications.
IRFLOW
ST keys
11
Appendix A
Specifications
Specifications are subject to change without notice.
CO2:
Range 0 to 5000 ppm Accuracy ±3% of reading or ±50 ppm, whichever is greater Resolution 1 ppm Sensor type Non-Dispersive Infrared (NDIR)
Temperature:
Range: 32 to 122°F (0 to 50°C) Accuracy Resolution: 0.1°F (0.1°C) Response time: 30 seconds (90% of final value, air velocity at 2 m/s) Display units: °F or °C (user selectable) Type: Thermistor
Humidity:
Range: 5 to 95% RH Accuracy Resolution: 0.1% RH Response time: 20 seconds (for 63% of final value) Sensor type: Thin-film capacitive
1
: 1.0°F (±0.6°C)
2
: ±3% RH (includes ±1% hysteresis)
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C) Operating (Probe): 14 to 140°F (-10 to 60°C) Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters Relative humidity up to 80% RH, non-condensing Pollution degree 1 in accordance with IEC 664 Transient over voltage category II
Data Storage Capabilities:
Range: More than 12,700 samples and 100 test IDs (one sample
can contain fourteen measurement types)
Logging Interval:
Intervals: 1 second to 1 hour
Time Constant:
Intervals: User selectable
13
External Meter Dimensions:
3.3 in. × 7.0 in. × 1.8 in. (8.4 cm × 17.8 cm × 4.4 cm)
Meter Probe Dimensions:
Probe length (including handle): 11.15 in. (28.3 cm) Probe diameter of tip: 0.75 in. (19.0 mm)
Meter Weight:
Weight with batteries: 0.6 lbs (0.27 kg)
Power Requirements:
Four AA-size batteries (included) or AC adapter (optional) 9 VDC, 300 mA, 4-18 watts (input voltage and frequency vary depending on which adapter is used)
1
Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F (0.03°C/°C) for
change in instrument temperature.
2
Accuracy with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F (0.2% RH/°C) for
change in probe temperature. Includes 1% hysteresis.
14 Appendix A
TSI Incorporated 500 Cardigan Road Shoreview, MN 55126 USA Toll Free (800) 874-2811 Telephone (651) 490-2811 Fax: (651) 490-3824
E-mail: customerservice@alnor.com Web: www.alnor.com
France Tel: +33 491 95 21 90 E-mail: tsifrance@tsi.com Germany Tel: +49 241 523030 E-mail: tsigmbh@tsi.com
Part No. 1980592 Rev. C Copyright © 2008 by TSI Incorporated
TSI Instruments Ltd. Stirling Road Cressex Business Park High Wycombe Bucks HP12 3RT United Kingdom Tel: +44 (0) 149 4 459200 Fax: +44 (0) 149 4 459700
E-mail: info@airflowinstruments.co.uk Web: www.airflowinstruments.co.uk
Quality Assured to ISO 9001:2000
Website: www.tsiinc.fr
Website: www.tsiinc.de
Loading...