Optional Probes for VELOCICALC 9555 Series
Multi-Functional Anemometers
Model Probe Description
960 Air Velocity and Temperature, straight probe
962 Air Velocity and Temperature, articulating probe
964 Air Velocity, Temperature, and Humidity, straight probe
966 Air Velocity, Temperature, and Humidity, articulating probe
995 4 in. (100 mm) Rotating Vane probe
496 1.5 in. (35mm) Rotating Vane probe
792 Surface Temperature probe
794 Air Temperature probe
980 Indoor Air Quality probe
982 Indoor Air Quality probe, with CO
Specifications
VELOCICALC
Models 9555, 9555-A, 9555-P, 9555-X and Optional Probes
Velocity (Pitot Tube for Meter Models 9555, 9555-A, 9555-P)
Range
1
250 to 15,500 ft/min (1.27 to 78.7 m/s)
Accuracy
2
±1.5% at 2,000 ft/min (10.16 m/s)
Resolution 1 ft/min (0.01 m/s)
Duct Size
Dimensions 1 to 250 inches in increments of 0.1 in.
(1 to 635 cm in increments of 0.1 cm)
Volumetric Flow Rate
Range Actual range is a function of velocity, pressure, duct size,
and K factor
Static/Differential Pressure (Meter Models 9555, 9555-A, 9555-P)
Range
3
-15 to +15 in. H2O (-28.0 to +28.0 mm Hg,
-3735 to +3735 Pa)
Accuracy ±1% of reading ±0.005 in. H
2
O (±0.01 mm Hg, ±1 Pa)
Resolution 0.001 in. H
2
O (0.1 Pa, 0.01 mm Hg)
Barometric Pressure
Range 20.36 to 36.648 in. Hg (517.15 to 930.87 mm Hg)
Accuracy ±2% of reading
Instrument Temperature Range
Operating (Electronics)
40 to 113°F (5 to 45°C)
Storage -4 to 140°F (-20 to 60°C)
Data Storage Capabilities
Range 26,500+ samples and 100 test IDs
Logging Interval
1 second to 1 hour
Time Constant
User selectable
External Meter Dimensions
3.8 in. x 8.3 in. x 2.1 in. (9.7 cm x 21.1 cm x 5.3 cm)
Meter Weight with Batteries
0.8 lbs. (0.36 kg)
Power Requirements
Four AA-size batteries or AC adapter
9555 9555-A 9555-P 9555-X
Probe that measures velocity,
temperature, and humidity
includes
964 probe
includes
966 probe
optional optional
Pressure measurement
•••
Calculates flow, wet bulb, dew
point, standard/actual
••
optional optional
Optional velocity and
temperature probe
••••
Optional rotating vane probe
••••
Optional IAQ probes (CO2,
temperature, humidity, CO)
••••
Data Logging (manual, auto save
continuous)
••••
Data logging software
••••
Optional Bluetooth printer
••••
Certificate of Calibration
••••
1
Pressure velocity measurements are not recommended below 1000 ft/min (5 m/s) and are
best suited to velocities over 2,000 ft/min (10.00 m/s). Range can vary depending on
barometric pressure.
2
Accuracy is a function of converting pressure to velocity. Conversion accuracy improves when
actual pressure values increase.
3
Overpressure range = 190 in. H2O (360 mmHg, 48 kPa).