TSI Alnor PAN231 Operating Manual

Page 1
POSITIVE AND NEGATIVE DUCT
ACCREDITATION (PANDA) SYSTEM
MODEL PAN231 SERIES
(PANDA LIGHT)

OPERATION AND SERVICE MANUAL

MAY 2013
Page 2
Page 3
Copyright©
TSI Incorporated / 2013 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824 LIMITATION OF WARRANTY AND LIABILITY (effective June 2011)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.) Seller warrants the goods sold hereunder, under normal use and service as described in the operator's manual, shall be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions: a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when
indicated in specifications, are warranted for 90 days from the date of shipment; b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals; c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in
workmanship and material, under normal use, for 90 days from the date of shipment; d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses,
batteries or other consumable materials. Only the original manufacturer's warranty applies; e. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect
to, and shall have no liability in connection with, goods which are incorporated into other products
or equipment, or which are modified by any person other than Seller. The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO
OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION
REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS OR ANY SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI's Customer Service department at (800) 874-2811 (USA) or (001 651) 490-2811 (International) or visit www.tsi.com.
Trademarks
TSI and TSI logo are registered trademarks of TSI Incorporated. Airflow is a trademark of TSI Incorporated.
Page 4
Page 5

CONTENTS

CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............. 1
CHAPTER 2 PREPARING PAN231 SYSTEM FOR AIR DUCT
LEAK TESTING ................................................................................... 5
CHAPTER 3 PERFORMING A DUCT LEAKAGE TEST ................ 11
Measuring Duct Static Pressure ................................................... 11
Measuring Duct Leakage Flow ..................................................... 12
Turning on the PAN231 Duct Leakage Tester.............................. 12
Using Leakage Test Application in the Model 9565-P .................. 12
Troubleshooting Guide ................................................................. 16
APPENDIX A SPECIFICATIONS ..................................................... 17
APPENDIX B LEAKAGE TESTING STANDARDS HIGHLIGHTS .. 21
Standards Supported .................................................................... 21
EU Standards ............................................................................... 22
US Standards ............................................................................... 25
APPENDIX C TYPICAL SETUP....................................................... 29
iii
Page 6
(This page intentionally left blank)
iv
Page 7
Qty
Description
Part Number
Reference Picture
1
Low flow nozzle
6002598
1
Primary duct adapter spigot plus rubber bung (to fit to test duct)
6002638
1
Cam lock primary spigot (to connect flexi-duct to PANDA)
6002607
2
Ø4-in. (100-mm) adjustable over lock straps
6002683
1
13-ft (4-m) long Ø4-in. (100-mm) plastic flexible duct
6002667 2
20-in. (500-mm) silicone tubes (red)
AFL9020004
Chapter 1
Unpacking and Parts Identification
Carefully unpack the PANDA system and instrument cases from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI immediately.
The PANDA system consists of the following:
1
Page 8
Qty
Description
Part Number
Reference Picture
2
20-in. (500-mm) silicone tubes (blue)
AFL9020005
1
16-ft (5-m) silicone tube (blue)
AFL9020005
1
K-type thermocouple probe
AFL82859201
3
Instrument adapter
AFL82859401
1
Instrument Box
6006490
1
Flex Duct Carry Tube
6006491
1
Operation and Service manual
6006886
2 Chapter 1
Page 9
The following two instruments should be used in conjunction with the
9565-P Multi-function Instrument
Refer to 9565 Operation and Service Manual supplied with the instrument for additional parts supplied as standard.
5815 Micromanometer
Refer to 5815 Operation and Service Manual supplied with the instrument for additional parts supplied as standard.
PANDA unit:
(continued on next page)
Page 10
IMPORTANT
Read Before Using the PANDA for the First Time
It is IMPORTANT that the 110V PANDA (yellow power socket) be connected only to 110V to 120V supplies. Connecting it to a higher voltage supply will permanently damage the inverter.
The 220V/240V PANDA (blue power socket) should only be connected to 200V to 240V power supplies.
When storing the PANDA in a vertical position, please ensure that the straps holding the instrument box are in place and tightened first.
The PANDA is designed so that the Instrument box and duct carry tube are removable to lighten the load when lifting.
4 Chapter 1
Page 11
Successfully completing a duct leakage test requires compiling certain information prior to starting the test. Refer to
Appendix B for a discussion of standards relating to duct
leakage testing. The list below indicates the information required:
Type of leakage test to be performed (Positive or
negative).
Leakage standard to be followed.  Air tightness/leakage class to be achieved  Amount of ductwork to be tested, such as the complete
system or a statistical sample.
C A U T I O N
Remove the power cord from the PANDA duct leakage tester before tilting it to the vertical position to avoid damaging the cord.
Chapter 2
Preparing PAN231 System for Air Duct Leak Testing
Carefully follow the procedures below to achieve safe and accurate leakage testing:
1. Select the section of the ductwork to be tested.
2. Calculate the surface area of the ductwork of the section to
be tested.
3. Temporarily seal all the openings of the ductwork except one,
which will be connected to the PAN231 duct leakage tester.
4. Position the PAN231 unit as close to the remaining opening in the
ductwork as possible to minimize the flexible tubing needed. Minimize bends in the flexible tubing to reduce the pressure loss, giving the best performance.
5. Make sure the Fan Run/Stop Switch on the Fan Speed Controller
is in the Stop position and the multi-turn Fan Speed Control potentiometer is fully turned counter-clockwise using the picture of the VFD in Figure 1 as a reference. Plug the cord into the PANDA unit as shown in Figure 2 and Figure 3. Then connect the other end of the cord to a suitable electrical supply.
5
Page 12
Fan Speed Control
Fan Run/Stop Switch
Figure 2. Receptacle for
Power Cord
Figure 3. Connected Power Cord
Figure 1. Fan Speed Controller
Note: The settings for the inverter have been locked and cannot be
changed using the key pad.
Note: The 110V unit has a yellow receptacle and the 230V unit has
a blue receptacle.
6 Chapter 2
Page 13
7
6. Fit the primary duct adapter spigot (black sheet metal with rubber bung) to one end of the 4-in. (100-mm) diameter flexi­tube. Make an air-tight seal using one of the over lock straps and lever-locking cam provided as shown in Figure 4. Adjust the fit of the over lock strap with a screwdriver.
7. Securely attach the black primary duct adapter spigot/flexi-tube assembly to the opening on the ductwork to be pressure tested.
Figure 4. Flex Ductwork
Connected to Primary Adapter
Spigot. Bung not shown.
8. If the static pressure tap on the black Primary Duct Adapter is open to the duct, connect the 16-ft (5-m) long blue silicone tube to it as shown in Figure 5.
If the static pressure tap on the black Primary Duct Adapter is not open to the duct, drill a 4-mm hole in the duct and insert about 6 inches (10 mm) of the silicone tube into the duct. Seal around the hole with putty.
9. Connect the other end of the 4-in. (100-mm) flexi-tube to the cam lock connector (grey cast
Figure 5. Connecting Pressure
Tubing to Tap on Primary Duct
Adapter. Bung not shown.
aluminum without nozzle). Make an airtight seal using the other over lock strap (not shown) and lever-locking cam provided. Adjust the fit of the over lock strap with a screwdriver.
Preparing PAN341 System for Air Duct Leak Testing
Page 14
10. Determine if you are going to perform a high- or low-flow testing and positive or negative testing. Set-up the duct leakage tester by:
a. For positive pressure, high-
flow testing, remove the low flow nozzle if it is installed. Then, connect the grey cast-aluminum cam lock connector to the outlet side of the blower per Figure 6. Close both cam lock arms at the same time to ensure proper fit.
Connect the free end of the static pressure 16-ft (5-m) silicone tube to the positive pressure connector on the
5815.
Cam Lock Connector
Fan Outlet
Figure 6. Positive Pressure,
High-Flow Setup
Finally, connect the FLOW GRID pressure taps marked P1(+) and P2(-) to the appropriate connectors on the 9565-P
using the red and blue tubing.
b. For positive pressure, low-
flow testing, add the low­flow nozzle to the blower inlet if it is not installed per Figure 7. Then, connect the grey cast-aluminum cam lock connector to the outlet side of the blower per Figure 6. Close both cam lock arms at the same time to ensure proper fit.
Connect the free end of the static pressure 16-ft (5-m) silicone tube to the positive pressure connector on the
5815.
Low Flow
Nozzle
Fan Inlet
Figure 7. Positive Pressure,
Low-Flow Setup
Finally connect the pressure taps marked P1(+) and P2(-) on the nozzle to the appropriate connectors on the 9565-P using the red and blue tubing.
8 Chapter 2
Page 15
9
c. For negative pressure,
high-flow testing, remove the low flow nozzle if it is installed. Then, connect the grey cast aluminum cam lock connector to the inlet side of the blower per Figure 8. Close both cam lock arms at the same time to ensure proper fit.
Connect the free end of the static pressure 16-ft (5-m) silicone tube to pressure connector on the 5815.
Fan Inlet
Cam Lock Connector
Figure 8. Negative Pressure,
High-Flow Setup
Finally connect the pressure taps marked P1(+) and P2(-) to the appropriate connector on the 9565-P using the red and blue tubing.
d. For negative pressure, low-
flow testing, add the low­flow nozzle to the blower inlet if it is not installed. Then, connect the grey cast aluminum cam lock connector to the low-flow nozzle per Figure 9. Close both cam lock arms at the same time to ensure proper fit.
Connect the free end of the static pressure 16-ft (5-m) silicone tube to the positive pressure connector to the
5815.
Cam Lock Connector
Fan Inlet
Low-Flow Nozzle
Figure 9. Negative Pressure,
Low-Flow Set-Up
Finally connect the pressure taps marked P1(+) and P2(-) on the nozzle to the appropriate connectors on the 9565-P using the red and blue tubing.
Preparing PAN341 System for Air Duct Leak Testing
Page 16
(This page intentionally left blank)
10 Chapter 2
Page 17
Chapter 3
Performing a Duct Leakage Test
The PAN231 duct leakage test system includes a Model 5815 Micromanometer and a Model 9565-P Ventilation Meter. During duct leakage testing, the Model 5815 Micromanometer measures the duct static pressure while the Model 9565-P Ventilation Meter measures the airflow rate.
Refer to the Operation and Service Manuals for the Model 5815 Micromanometer and the Model 9565-P Ventilation Meter to use these instruments in other applications. If you do not have the
manuals, download them from TSI’s website www.tsi.com.

Measuring Duct Static Pressure

1. Turn ON the Model 5815.
2. Zero the Model 5815 pressure sensor with both ports open to the atmosphere.
3. Connect the (+) port on the Model 5815 to measure the duct static pressure (see Figure 5).
4. Leave the (-) port on the Model 5815 open to the atmosphere.
Note: Refer to the Model 5815 Operation and Service Manual
for instruction on use.
11
Page 18
C A U T I O N
The Model 5815 and Model 9565-P meters must be zeroed before entering the Leakage Test Application.
1. Turn ON Model 9565-P.
2. Zero the Model 9565-P pressure sensor with both ports open to the atmosphere.
3. Connect the Model 9565-P to the PAN231 by connecting the (+) and (-) ports on the Model 9565-P to the P1 (+) and P2 (+) ports located on the inside edge of the box shelf facing the fan. See Figure 10.
4. Connect the thermocouple to the Model 9565-P.
5. Insert the thermocouple probe into the blower inlet through the hole marked TC1.
Figure 10. Connecting
Instruments to PANDA Tester in
High Flow Mode

Measuring Duct Leakage Flow

Turning on the PAN231 Duct Leakage Tester

1. Power the PANDA unit on by plugging in the power cord.
2. Position the Fan Control switch to the RUN position to energize the fan.
3. Increase the fan to the desired speed by turning the Fan Speed Controller clockwise. To decrease the fan speed, turn the Fan Speed controller counter-clockwise.

Using Leakage Test Application in the Model 9565-P

12 Chapter 4
Page 19
13
1. Press the MENU key to access the menu system on the Model 9565-P.
2. Use the  keys to highlight the Applications item.
3. Press the (ENTER) key to access the Applications menu.
MENU Zero Press Display Setup
Settings
Flow Setup Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration Discover Printer
4. Select Leakage Test and press key.
APPLICATIONS Draft Rate Heatflow Turbulence % Outside Air
Leakage Test
5. Select either the EN Standard or SMACNA leakage test.
LEAKAGE TEST EN Standard SMACNA
1. Enter key parameters: a. Surface Area of ductwork
section to be tested.
b. Static Pressure of test, as
measured by Model 5815 micromanometer.
c. Flow Device as Nozzle or
Flow Grid.
LEAKAGE TEST Surface Area Static Pressure Flow Device Tightness Class Test Length Run Test

Instrument Operation if EN Standard Test Protocol is Selected

Performing a Duct Leakage Test
Page 20
d. Leakage class as A, B, C, or D. Note that tests with negative
pressures must be selected as negative tests, as indicated by -.
e. Test Length, or duration of leakage test, usually 5 minutes.
2. Increase the blower speed until the desired static pressure is achieved.
3. When the static pressure has stabilized, select Run Test and press .
4. The display will show the readings on the right. Leakage Factor and Leak Rate will update in real time, while other parameters will remain constant.
If the Leakage Factor and Leak Rate are sufficiently stable, press the START soft key or the
key to begin the leak test. Pressing the ESC key will exit back to the previous screen.
LEAKAGE TEST Leakage Factor x.xx Leak Limit x.xx Leak Rate x.xx
Status OK
Flow Device Flow Grid Baro Pressure 20.20 Temperature 20C Time 9:55
Standard TestXXX
Sample 0
(Sample Saved 1)
5. After the leak test is complete, the Model 9565-P will prompt you to press the SAVE or PRINT soft key. You can also press the ESC key here to back out to the previous screen without saving the data.
1. Enter key parameters: a. Surface Area of ductwork
section to be tested.
b. Static Pressure of test, as
measured by Model 5815 micromanometer.
c. Flow Device as Nozzle or
Flow Grid.
LEAKAGE TEST Surface Area Static Pressure Flow Device Leakage Class Test Length Run Test
After completing leakage testing for a section of duct, you can move onto the next section.

Instrument Operation if SMACNA Test Protocol is Selected

14 Chapter 4
Page 21
15
d. Leakage class as a number
from 1 to 48. Typical values are 3, 6, 12, 24, or 48.
e. Test Length, or duration of
leakage test.
2. Increase the blower speed until the desired static pressure is achieved.
3. When the static pressure has stabilized, select Run Test and press .
4. The display will show the readings on the right. Leakage Factor and Leak Rate will update in real time, while other parameters will remain constant.
LEAKAGE TEST Leakage Factor x.xx Leak Limit x.xx Leak Rate x.xx Status OK
Flow Device Flow Grid Baro Pressure 20.20 Temperature 20C Time 9:55 Standard TestXXX
Sample
0 (Sample Saved 1) Stop(Save) Print
If the Leak Factor and Leak Rate are sufficiently stable, press the
START soft key or the key to begin the leak test. Pressing the ESC key will exit back to the previous screen.
5. After the leak test is complete, the Model 9565-P will prompt you to press the SAVE or PRINT soft key. You can also press the ESC key here to back out to the previous screen without saving the data.
After completing leakage testing for a section of duct, you can move onto the next section.
Performing a Duct Leakage Test
Page 22
Symptom
Recommended Action
Fan motor will not run.
Check the power connection. Circuit Breaker may have
tripped.
Static pressure reading (on 5815) is zero.
Check the connections.
Static pressure reading (on 5815) is too low.
Required static pressure cannot be achieved with motor speed control settings at the maximum.
Leakage rate is too high.
Check for leaks using soap bubbles or smoke pallets. Alternatively, test a smaller section of the ductwork.
Leak Flow (on 9565-P) shows flashing XXX.XX.
Check the pressure tube
connections to the 9565-P meter.
Leak flow is too low. Use low
flow nozzle adapter.

Troubleshooting Guide

16 Chapter 4
Page 23
Pressure Measurement (5815)
Range
± 3,735 Pa
±15 inwg
Resolution
0.1 Pa
0.001 inwg
Accuracy
±1% of reading ±1 Pa
±1% of reading ±0.005 inwg
Actual duct static range
2500 Pa at 0 Flow
10 inwg at 0 Flow
Volume Flow Measurement (9565-P)
High Leakage Range (Flow Grid):
10 to 200 l/s 36 to 720 m3 /hr
21 to 424 cfm
Low Leakage Range (15 mm Low Flow Nozzle Adapter)
1 to 13 l/s
3.6 to 46.9 m3 /hr
2 to 27.5 cfm
Accuracy
± 2.5% of reading ± 0.01 l/s
± 2.5% of reading ± 0.04 m3 /hr
± 2.5% of reading ± 0.02 cfm
Resolution
0.01 l/s
0.01 m3 /hr
0.01 cfm
Temperature Measurement (9565-P)
K Type Thermo Couple Probe To EN60584 (IEC 584)
Barometric Pressure Measurement (9565-P)
Range
690 to 1,241 hPa
517.5 to
930.87 mm Hg
20.36 to
36.648 in Hg Accuracy
±2% of reading
±2% of reading
Weight
Carry Weight Total Weight
45 Kg 55 Kg
99 lbs 121 lbs
Dimensions (LxWxH)
1,130 mm x 660 mm x 600 mm
44.5 in x 26 in x
23.5 in
Power Requirements
230V Version
220 to 240 V, 1 Phase, 50/60 Hz 10A
110V Version
110 to 120 V, 1 Phase, 50/60 Hz 16A
9565-P and 5815
See specification sheets for details on individual instruments

Appendix A Specifications

(Specifications are subject to change without notice.)
17
Page 24
0
500
1000
1500
2000
2500
3000
0 50 100 150 200 250
Static Pressure Pa.
Volume Flow l/s
Panda Fan Performance Graph (metric units)
18 Appendix A
Page 25
19
0
1
2
3
4
5
6
7
8
9
10
11
0 50 100 150 200 250 300 350 400 450 500
Static Pressure ins.w g.
Volume Flow CFM
Panda Fan Performance Graph (imperial units)
Specifications
Page 26
(This page intentionally left blank)
20 Appendix A
Page 27
Standard
US- or
EU-
Based
Description
BS EN 12237:2003
EU
Ventilation for buildings DuctworkStrength and leakage of circular sheet metal ducts.
BS EN 1507:2006
EU
Ventilation for buildingsSheet metal air ducts with rectangular sectionRequirements for strength and leakage.
DW/143
EU
HVACA practical guide to Ductwork leakage testing.
Eurovent 2/2
EU
Air leakage rate in sheet metal air distribution systems.
SMACNA HVAC Air Duct Leakage Test manual, Second edition, 2012
US
Duct construction leakage classification, expected leakage rates for sealed and unsealed ductwork, duct leakage test procedures, recommendations on use of leakage testing, types of test apparatus and test setup and sample leakage analysis.
Appendix B
Leakage Testing Standards Highlights
Different standards are used throughout the world to specify duct air tightness and leakage requirements. The PAN231 duct leakage test system has a duct leakage application to automatically compare the actual leakage flow with the maximum allowed leakage flow for EN and SMACNA standards. Field technicians can also use the duct leakage application to determine actual leakage flow and manually compare it to maximum leakage from another standard. The PAN231 duct leakage test system cannot determine the appropriate leakage classification for a given duct.

Standards Supported

TSI has made every effort to accurately reflect the standards referenced. Please refer to the actual standards for more detailed information and to make the best interpretation of each statement.
21
Page 28
Air
Tightness
Class
Air Leakage
Limit (f
max
)
m3/s/m2
Static Pressure Limit (ps) Pa
Negative
Positive at
pressure class
1 2 3
A
200
400
B
500
400
1000
2000
C
750
400
1000
2000
D*
750
400
1000
2000 * Class D ductwork is only for special apparatus
Air
Tightness
Class
Air leakage
limit (f
max
)
m3*/s/m2
Static Pressure Limit (ps)
Pa
Negative
Positive
A
500
500
B
750
1000
C
750
2000
D*
750
2000
* Class D ductwork is only for special apparatus
The scope of the standards listed above includes many items other than duct leakage. This summary, however, is limited to duct leakage testing.

EU Standards

Ductwork classification and maximum air leakage. Note that EN1507, EN12237 Eurovent 2/2 and DW/143 all have the same formula to determine f l/s/m2 whereas others use m3/s/m2.
EN 1507 (rectangular ductwork)
, the Air Leakage Limit, although DW/143 uses units of
max
EN12237 (circular ductwork)
22 Appendix B
Page 29
Eurovent 2/2 Air Tightness For Installed Duct Testing:
Air Tightness
Class
Air leakage limit (f
max
)
m3/s/m2
A B C
Duct
Pressure
Class
Static Pressure
Limit
Maximum
Air
Velocity
m/s
Air leakage
limits l/s/m2
Positive
Pa
Negative
Pa
Low-pressure
– Class A
500
500
10
0.027*p
t
0.65
Medium-
pressure –
Class B
1000
750
20
0.009*p
t
0.65
High pressure
– Class C
2000
750
40
0.003*p
t
0.65
DW/143: A Practical Guide to Ductwork Leakage Testing
Leakage Testing Standards Highlights 23
Figure 11. Allowable Air Leakage Rates from DW/143
Page 30
The measured leakage flow rates shall be corrected if the
temperature and/or barometric pressure are different from standard conditions (+20ºC and 101 325 Pa) as follows:
where:
qv=corrected flow leakage rate q
= measured flow leakage rate
measured
t = measured temperature (ºC) p = measured barometric pressure (Pa)
The test report shall give the following general information of the
test performed:
o Date and place o Test personnel and witness o Test equipment, including pressuring means and measuring
instruments
o Air temperature and barometric pressure during the test o Building and project reference o Design of installed ductwork including dimensions, thickness
of materials, types of stiffening, length, type of duct/tubes and fittings, assembly method and distance of hangers/supports
o Required air tightness class and design operating pressure of
the installed ductwork
o Installer of ductwork o Manufacturer of the ductwork o Measured values of:
1. Ductwork surface area (A)
2. Total joint length (L)
3. Test pressure (p
test
)
4. Leakage flow rate (qv) corrected for temperature and barometric pressure
5. Pressurizing time
o Calculated values of
1. Leakage factor (f)
2. Air leakage limit (f table above at the measured test pressure (p
) according to the formulas given in
max
)
test
o Air tightness class achieved
24 Appendix B
Page 31
For tests including several test pressures it is recommended to
Duct Class
½-, 1-, 2-inwg
3-inwg
4-, 6-, 10-inwg
Seal Class
C B A
Sealing
Applicable
Transverse Joints Only
Transverse
Joints and
Seams
Joints, Seams
and All Wall
Penetrations
Leakage Class
Rectangular
Metal
24
12
6
Round
Metal
12 6 3
plot the leakage factors as a function of test pressure in a diagram together with the air leakage limit curve.

US Standards

Ductwork classification and maximum air leakage
Maximum air leakage is then defined as
0.65
F=CLP where: F = Maximum air leakage (cfm/100 ft2)
CL = Leakage class P = Pressure (inwg)
Leakage Testing Standards Highlights 25
Figure 12. Allowable Air Duct Leakage from Round Ducts,
per SMACNA Standard
Page 32
Figure 13. Allowable Air Duct Leakage from Rectangular Ducts,
per SMACNA Standard
The SMACNA standard does not generally require correcting
leakage flow rates to standard conditions, unless:
1. Air temperature <40ºF or >100ºF
2. Elevation <1500 ft above sea level
3. Duct static pressure <-20 inwg or >+20 inwg Should one of these conditions not be satisfied, then correcting the leakage to standard conditions may be done using one of these formulas:
1. ACFM=SCFM * (460+T)/530 where T = actual dry bulb air temperature (ºF) moisture is negligible pressure between -20 and +20 inwg
2. ACFM = SCFM * 0.075/d where d = air density from psychrometric chart
3. ACFM=lb dry air/minute * humid volume (ft3/lb dry air)
26 Appendix B
Page 33
The SMACNA standard does not specify the information to be
reported, but instead defers to project documents. However, the SMACNA standard does include a sample test report that includes:
o Test date and place o Test personnel and witness o Building and project reference o Duct section tested o Specified leakage class, test pressure and duct construction
pressure class
o Measurements of:
Ductwork surface area  Leakage flow and calculations required to determine
leakage flow
Leakage Testing Standards Highlights 27
Page 34
(This page intentionally left blank)
28 Appendix B
Page 35
Appendix C
Typical Setup
29
Page 36
(This page intentionally left blank)
30 Appendix C
Page 37
USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 4 91 11 87 64 Germany Tel: +49 241 523030
India Tel: +91 80 67877200 China Tel: +86 10 8219 7688 Singapore Tel: +65 6595 6388
TSI Incorporated – Visit our website www.tsi.com for more information.
P/N 6006886 Rev A ©2013 TSI Incorporated Printed in U.S.A.
Loading...