TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective June 2011)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods sold hereunder, under normal use and service as described in the operator's
manual, shall be free from defects in workmanship and material for 24 months, or if less, the length of
time specified in the operator's manual, from the date of shipment to the customer. This warranty period
is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and
exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when
indicated in specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in
workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses,
batteries or other consumable materials. Only the original manufacturer's warranty applies;
e. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect
to, and shall have no liability in connection with, goods which are incorporated into other products
or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO
OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR
MERCHANTABILITY IS MADE. WITH RESPECT TO SELLER’S BREACH OF THE IMPLIED
WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT
INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE
DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION
REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER,
AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES
CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT,
STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE
REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR
REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR
REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE
PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS OR
ANY SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE
RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR
CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a
cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's
risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY,
which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF
WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing
signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our
customers, our service policy is designed to give prompt attention to any problems. If any
malfunction is discovered, please contact your nearest sales office or representative, or call TSI's
Customer Service department at (800) 874-2811 (USA) or (001 651) 490-2811 (International) or
visit www.tsi.com.
Trademarks
TSI and TSI logo are registered trademarks of TSI Incorporated.
Airflow is a trademark of TSI Incorporated.
Page 4
Page 5
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............. 1
EU Standards ............................................................................... 22
US Standards ............................................................................... 25
APPENDIX C TYPICAL SETUP....................................................... 29
iii
Page 6
(This page intentionally left blank)
iv
Page 7
Qty
Description
Part Number
Reference Picture
1
Low flow nozzle
6002598
1
Primary duct adapter
spigot plus rubber bung
(to fit to test duct)
6002638
1
Cam lock primary spigot
(to connect flexi-duct to
PANDA)
6002607
2
Ø4-in. (100-mm)
adjustable over lock
straps
6002683
1
13-ft (4-m) long
Ø4-in. (100-mm) plastic
flexible duct
6002667
2
20-in. (500-mm) silicone
tubes (red)
AFL9020004
Chapter 1
Unpacking and Parts Identification
Carefully unpack the PANDA system and instrument cases from the
shipping container. Check the individual parts against the list of
components below. If anything is missing or damaged, notify TSI
immediately.
The PANDA system consists of the following:
1
Page 8
Qty
Description
Part Number
Reference Picture
2
20-in. (500-mm) silicone
tubes (blue)
AFL9020005
1
16-ft (5-m) silicone tube
(blue)
AFL9020005
1
K-type thermocouple
probe
AFL82859201
3
Instrument adapter
AFL82859401
1
Instrument Box
6006490
1
Flex Duct Carry Tube
6006491
1
Operation and Service
manual
6006886
2 Chapter 1
Page 9
The following two instruments should be used in conjunction with the
9565-P Multi-function
Instrument
Refer to 9565 Operation and Service
Manual supplied with the instrument
for additional parts supplied as
standard.
5815 Micromanometer
Refer to 5815 Operation and Service
Manual supplied with the instrument
for additional parts supplied as
standard.
PANDA unit:
(continued on next page)
Page 10
IMPORTANT—
Read Before Using the PANDA for the First Time
It is IMPORTANT that the 110V PANDA (yellow power socket) be
connected only to 110V to 120V supplies. Connecting it to a higher
voltage supply will permanently damage the inverter.
The 220V/240V PANDA (blue power socket) should only be
connected to 200V to 240V power supplies.
When storing the PANDA in a vertical position, please ensure that the
straps holding the instrument box are in place and tightened first.
The PANDA is designed so that the Instrument box and duct carry
tube are removable to lighten the load when lifting.
4 Chapter 1
Page 11
Successfully completing a duct leakage test requires compiling
certain information prior to starting the test. Refer to
Appendix B for a discussion of standards relating to duct
leakage testing. The list below indicates the information
required:
Type of leakage test to be performed (Positive or
negative).
Leakage standard to be followed.
Air tightness/leakage class to be achieved
Amount of ductwork to be tested, such as the complete
system or a statistical sample.
C A U T I O N
Remove the power cord from the PANDA duct leakage tester before
tilting it to the vertical position to avoid damaging the cord.
Chapter 2
Preparing PAN231 System for Air Duct Leak
Testing
Carefully follow the procedures below to achieve safe and accurate
leakage testing:
1. Select the section of the ductwork to be tested.
2. Calculate the surface area of the ductwork of the section to
be tested.
3. Temporarily seal all the openings of the ductwork except one,
which will be connected to the PAN231 duct leakage tester.
4. Position the PAN231 unit as close to the remaining opening in the
ductwork as possible to minimize the flexible tubing needed.
Minimize bends in the flexible tubing to reduce the pressure loss,
giving the best performance.
5. Make sure the Fan Run/Stop Switch on the Fan Speed Controller
is in the Stop position and the multi-turn Fan Speed Control
potentiometer is fully turned counter-clockwise using the picture
of the VFD in Figure 1 as a reference. Plug the cord into the
PANDA unit as shown in Figure 2 and Figure 3. Then connect the
other end of the cord to a suitable electrical supply.
5
Page 12
Fan Speed Control
Fan Run/Stop Switch
Figure 2. Receptacle for
Power Cord
Figure 3. Connected Power Cord
Figure 1. Fan Speed Controller
Note: The settings for the inverter have been locked and cannot be
changed using the key pad.
Note: The 110V unit has a yellow receptacle and the 230V unit has
a blue receptacle.
6 Chapter 2
Page 13
7
6. Fit the primary duct adapter
spigot (black sheet metal with
rubber bung) to one end of the
4-in. (100-mm) diameter flexitube. Make an air-tight seal
using one of the over lock
straps and lever-locking cam
provided as shown in Figure 4.
Adjust the fit of the over lock
strap with a screwdriver.
7. Securely attach the black
primary duct adapter
spigot/flexi-tube assembly to
the opening on the ductwork
to be pressure tested.
Figure 4. Flex Ductwork
Connected to Primary Adapter
Spigot. Bung not shown.
8. If the static pressure tap on the
black Primary Duct Adapter is
open to the duct, connect the
16-ft (5-m) long blue silicone
tube to it as shown in Figure 5.
If the static pressure tap on the
black Primary Duct Adapter is
not open to the duct, drill a
4-mm hole in the duct and
insert about 6 inches (10 mm)
of the silicone tube into the
duct. Seal around the hole
with putty.
9. Connect the other end of the
4-in. (100-mm) flexi-tube to the
cam lock connector (grey cast
Figure 5. Connecting Pressure
Tubing to Tap on Primary Duct
Adapter. Bung not shown.
aluminum without nozzle). Make an airtight seal using the other
over lock strap (not shown) and lever-locking cam provided.
Adjust the fit of the over lock strap with a screwdriver.
Preparing PAN341 System for Air Duct Leak Testing
Page 14
10. Determine if you are going to perform a high- or low-flow testing
and positive or negative testing. Set-up the duct leakage
tester by:
a. For positive pressure, high-
flow testing, remove the
low flow nozzle if it is
installed. Then, connect
the grey cast-aluminum
cam lock connector to the
outlet side of the blower
per Figure 6. Close both
cam lock arms at the same
time to ensure proper fit.
Connect the free end of the
static pressure 16-ft (5-m)
silicone tube to the positive
pressure connector on the
5815.
Cam Lock
Connector
Fan Outlet
Figure 6. Positive Pressure,
High-Flow Setup
Finally, connect the FLOW GRID pressure taps marked
P1(+) and P2(-) to the appropriate connectors on the 9565-P
using the red and blue tubing.
b. For positive pressure, low-
flow testing, add the lowflow nozzle to the blower
inlet if it is not installed per
Figure 7. Then, connect
the grey cast-aluminum
cam lock connector to the
outlet side of the blower
per Figure 6. Close both
cam lock arms at the same
time to ensure proper fit.
Connect the free end of the
static pressure 16-ft (5-m)
silicone tube to the positive
pressure connector on the
5815.
Low Flow
Nozzle
Fan Inlet
Figure 7. Positive Pressure,
Low-Flow Setup
Finally connect the pressure taps marked P1(+) and P2(-) on
the nozzle to the appropriate connectors on the 9565-P
using the red and blue tubing.
8 Chapter 2
Page 15
9
c. For negative pressure,
high-flow testing, remove
the low flow nozzle if it is
installed. Then, connect
the grey cast aluminum
cam lock connector to the
inlet side of the blower per
Figure 8. Close both cam
lock arms at the same time
to ensure proper fit.
Connect the free end of the
static pressure 16-ft (5-m)
silicone tube to pressure
connector on the 5815.
Fan Inlet
Cam Lock
Connector
Figure 8. Negative Pressure,
High-Flow Setup
Finally connect the pressure taps marked P1(+) and P2(-) to
the appropriate connector on the 9565-P using the red and
blue tubing.
d. For negative pressure, low-
flow testing, add the lowflow nozzle to the blower
inlet if it is not installed.
Then, connect the grey
cast aluminum cam lock
connector to the low-flow
nozzle per Figure 9. Close
both cam lock arms at the
same time to ensure
proper fit.
Connect the free end of the
static pressure 16-ft (5-m)
silicone tube to the positive
pressure connector to the
5815.
Cam Lock
Connector
Fan Inlet
Low-Flow Nozzle
Figure 9. Negative Pressure,
Low-Flow Set-Up
Finally connect the pressure taps marked P1(+) and P2(-) on
the nozzle to the appropriate connectors on the 9565-P
using the red and blue tubing.
Preparing PAN341 System for Air Duct Leak Testing
Page 16
(This page intentionally left blank)
10 Chapter 2
Page 17
Chapter 3
Performing a Duct Leakage Test
The PAN231 duct leakage test system includes a Model 5815
Micromanometer and a Model 9565-P Ventilation Meter. During duct
leakage testing, the Model 5815 Micromanometer measures the duct
static pressure while the Model 9565-P Ventilation Meter measures
the airflow rate.
Refer to the Operation and Service Manuals for the Model 5815
Micromanometer and the Model 9565-P Ventilation Meter to use
these instruments in other applications. If you do not have the
manuals, download them from TSI’s website www.tsi.com.
Measuring Duct Static Pressure
1. Turn ON the Model 5815.
2. Zero the Model 5815 pressure sensor with both ports open to
the atmosphere.
3. Connect the (+) port on the Model 5815 to measure the duct
static pressure (see Figure 5).
4. Leave the (-) port on the Model 5815 open to the atmosphere.
Note: Refer to the Model 5815 Operation and Service Manual
for instruction on use.
11
Page 18
C A U T I O N
The Model 5815 and Model 9565-P meters must be zeroed before
entering the Leakage Test Application.
1. Turn ON Model 9565-P.
2. Zero the Model 9565-P
pressure sensor with both
ports open to the atmosphere.
3. Connect the Model 9565-P to
the PAN231 by connecting the
(+) and (-) ports on the Model
9565-P to the P1 (+) and P2 (+) ports located on the inside
edge of the box shelf facing
the fan. See Figure 10.
4. Connect the thermocouple to
the Model 9565-P.
5. Insert the thermocouple probe
into the blower inlet through
the hole marked TC1.
Figure 10. Connecting
Instruments to PANDA Tester in
High Flow Mode
Measuring Duct Leakage Flow
Turning on the PAN231 Duct Leakage Tester
1. Power the PANDA unit on by plugging in the power cord.
2. Position the Fan Control switch to the RUN position to energize
the fan.
3. Increase the fan to the desired speed by turning the Fan Speed
Controller clockwise. To decrease the fan speed, turn the Fan
Speed controller counter-clockwise.
Using Leakage Test Application in the Model 9565-P
12 Chapter 4
Page 19
13
1. Press the MENU key to access
the menu system on the
Model 9565-P.
2. Use the keys to highlight
the Applications item.
3. Press the (ENTER) key to
access the Applications menu.
MENU
Zero Press
Display Setup
Settings
Flow Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Discover Printer
4. Select Leakage Test and
press key.
APPLICATIONS
Draft Rate
Heatflow
Turbulence
% Outside Air
Leakage Test
5. Select either the EN Standard
or SMACNA leakage test.
LEAKAGE TEST
EN Standard
SMACNA
1. Enter key parameters:
a. Surface Area of ductwork
section to be tested.
b. Static Pressure of test, as
measured by Model 5815
micromanometer.
c. Flow Device as Nozzle or
Flow Grid.
LEAKAGE TEST
Surface Area
Static Pressure
Flow Device
Tightness Class
Test Length
Run Test
Instrument Operation if EN Standard Test Protocol is Selected
Performing a Duct Leakage Test
Page 20
d. Leakage class as A, B, C, or D. Note that tests with negative
pressures must be selected as negative tests, as indicated
by -.
e. Test Length, or duration of leakage test, usually 5 minutes.
2. Increase the blower speed until
the desired static pressure is
achieved.
3. When the static pressure has
stabilized, select Run Test and
press .
4. The display will show the
readings on the right. Leakage
Factor and Leak Rate will
update in real time, while other
parameters will remain constant.
If the Leakage Factor and Leak
Rate are sufficiently stable,
press the START soft key or the
key to begin the leak test.
Pressing the ESC key will exit
back to the previous screen.
LEAKAGE TEST
Leakage Factor x.xx
Leak Limitx.xx
Leak Ratex.xx
5. After the leak test is complete, the Model 9565-P will prompt you
to press the SAVE or PRINT soft key. You can also press the ESC
key here to back out to the previous screen without saving the
data.
1. Enter key parameters:
a. Surface Area of ductwork
section to be tested.
b. Static Pressure of test, as
measured by Model 5815
micromanometer.
c. Flow Device as Nozzle or
Flow Grid.
LEAKAGE TEST
Surface Area
Static Pressure
Flow Device
Leakage Class
Test Length
Run Test
After completing leakage testing for a section of duct, you can move
onto the next section.
Instrument Operation if SMACNA Test Protocol is Selected
14 Chapter 4
Page 21
15
d. Leakage class as a number
from 1 to 48. Typical values
are 3, 6, 12, 24, or 48.
e. Test Length, or duration of
leakage test.
2. Increase the blower speed
until the desired static pressure
is achieved.
3. When the static pressure has
stabilized, select RunTest and
press .
4. The display will show the
readings on the right. Leakage
Factor and Leak Rate will
update in real time, while other
parameters will remain constant.
LEAKAGE TEST
Leakage Factor x.xx
Leak Limitx.xx
Leak Ratex.xx
StatusOK
If the Leak Factor and Leak Rate are sufficiently stable, press the
START soft key or the key to begin the leak test. Pressing the
ESC key will exit back to the previous screen.
5. After the leak test is complete, the Model 9565-P will prompt you
to press the SAVE or PRINT soft key. You can also press the
ESC key here to back out to the previous screen without saving
the data.
After completing leakage testing for a section of duct, you can move
onto the next section.
Performing a Duct Leakage Test
Page 22
Symptom
Recommended Action
Fan motor will not run.
Check the power connection.
Circuit Breaker may have
tripped.
Static pressure reading (on 5815)
is zero.
Check the connections.
Static pressure reading (on 5815)
is too low.
Required static pressure cannot
be achieved with motor speed
control settings at the maximum.
Leakage rate is too high.
Check for leaks using soap
bubbles or smoke pallets.
Alternatively, test a smaller
section of the ductwork.
Leak Flow (on 9565-P) shows
flashing XXX.XX.
Check the pressure tube
connections to the 9565-P
meter.
Leak flow is too low. Use low
flow nozzle adapter.
Troubleshooting Guide
16 Chapter 4
Page 23
Pressure Measurement (5815)
Range
± 3,735 Pa
±15 inwg
Resolution
0.1 Pa
0.001 inwg
Accuracy
±1% of reading
±1 Pa
±1% of reading
±0.005 inwg
Actual duct static range
2500 Pa at 0 Flow
10 inwg at 0 Flow
Volume Flow Measurement (9565-P)
High Leakage Range
(Flow Grid):
10 to 200 l/s
36 to 720 m3 /hr
21 to 424 cfm
Low Leakage Range
(15 mm Low Flow Nozzle
Adapter)
1 to 13 l/s
3.6 to 46.9 m3 /hr
2 to 27.5 cfm
Accuracy
± 2.5% of reading
± 0.01 l/s
± 2.5% of reading
± 0.04 m3 /hr
± 2.5% of reading
± 0.02 cfm
Resolution
0.01 l/s
0.01 m3 /hr
0.01 cfm
Temperature Measurement (9565-P)
K Type Thermo Couple Probe To EN60584 (IEC 584)
Barometric Pressure Measurement (9565-P)
Range
690 to 1,241 hPa
517.5 to
930.87 mm Hg
20.36 to
36.648 in Hg
Accuracy
±2% of reading
±2% of reading
Weight
Carry Weight
Total Weight
45 Kg
55 Kg
99 lbs
121 lbs
Dimensions (LxWxH)
1,130 mm x 660
mm x 600 mm
44.5 in x 26 in x
23.5 in
Power Requirements
230V Version
220 to 240 V, 1 Phase, 50/60 Hz 10A
110V Version
110 to 120 V, 1 Phase, 50/60 Hz 16A
9565-P and 5815
See specification sheets for details on
individual instruments
Appendix A
Specifications
(Specifications are subject to change without notice.)
17
Page 24
0
500
1000
1500
2000
2500
3000
050100150200250
Static Pressure Pa.
Volume Flow l/s
Panda Fan Performance Graph (metric units)
18 Appendix A
Page 25
19
0
1
2
3
4
5
6
7
8
9
10
11
050100150200250300350400450500
Static Pressure ins.w g.
Volume Flow CFM
Panda Fan Performance Graph (imperial units)
Specifications
Page 26
(This page intentionally left blank)
20 Appendix A
Page 27
Standard
US- or
EU-
Based
Description
BS EN 12237:2003
EU
Ventilation for buildings—
Ductwork—Strength and leakage of
circular sheet metal ducts.
BS EN 1507:2006
EU
Ventilation for buildings—Sheet
metal air ducts with rectangular
section—Requirements for strength
and leakage.
DW/143
EU
HVAC—A practical guide to
Ductwork leakage testing.
Eurovent 2/2
EU
Air leakage rate in sheet metal air
distribution systems.
SMACNA HVAC Air
Duct Leakage Test
manual, Second
edition, 2012
US
Duct construction leakage
classification, expected leakage
rates for sealed and unsealed
ductwork, duct leakage test
procedures, recommendations on
use of leakage testing, types of test
apparatus and test setup and
sample leakage analysis.
Appendix B
Leakage Testing Standards Highlights
Different standards are used throughout the world to specify duct air
tightness and leakage requirements. The PAN231 duct leakage test
system has a duct leakage application to automatically compare the
actual leakage flow with the maximum allowed leakage flow for EN
and SMACNA standards. Field technicians can also use the duct
leakage application to determine actual leakage flow and manually
compare it to maximum leakage from another standard. The PAN231
duct leakage test system cannot determine the appropriate leakage
classification for a given duct.
Standards Supported
TSI has made every effort to accurately reflect the standards
referenced. Please refer to the actual standards for more detailed
information and to make the best interpretation of each statement.
21
Page 28
Air
Tightness
Class
Air Leakage
Limit (f
max
)
m3/s/m2
Static Pressure Limit (ps) Pa
Negative
Positive at
pressure class
1 2 3
A
200
400
B
500
400
1000
2000
C
750
400
1000
2000
D*
750
400
1000
2000
* Class D ductwork is only for special apparatus
Air
Tightness
Class
Air leakage
limit (f
max
)
m3*/s/m2
Static Pressure Limit (ps)
Pa
Negative
Positive
A
500
500
B
750
1000
C
750
2000
D*
750
2000
* Class D ductwork is only for special apparatus
The scope of the standards listed above includes many items other
than duct leakage. This summary, however, is limited to duct
leakage testing.
EU Standards
Ductwork classification and maximum air leakage. Note that EN1507,
EN12237 Eurovent 2/2 and DW/143 all have the same formula to
determine f
l/s/m2 whereas others use m3/s/m2.
EN 1507 (rectangular ductwork)
, the Air Leakage Limit, although DW/143 uses units of
max
EN12237 (circular ductwork)
22 Appendix B
Page 29
Eurovent 2/2 Air Tightness For Installed Duct Testing:
Air Tightness
Class
Air leakage limit (f
max
)
m3/s/m2
A B C
Duct
Pressure
Class
Static Pressure
Limit
Maximum
Air
Velocity
m/s
Air leakage
limits
l/s/m2
Positive
Pa
Negative
Pa
Low-pressure
– Class A
500
500
10
0.027*p
t
0.65
Medium-
pressure –
Class B
1000
750
20
0.009*p
t
0.65
High pressure
– Class C
2000
750
40
0.003*p
t
0.65
DW/143: A Practical Guide to Ductwork Leakage Testing
Leakage Testing Standards Highlights 23
Figure 11. Allowable Air Leakage Rates from DW/143
Page 30
The measured leakage flow rates shall be corrected if the
temperature and/or barometric pressure are different from
standard conditions (+20ºC and 101 325 Pa) as follows:
where:
qv=corrected flow leakage rate
q
= measured flow leakage rate
measured
t = measured temperature (ºC)
p = measured barometric pressure (Pa)
The test report shall give the following general information of the
test performed:
o Date and place
o Test personnel and witness
o Test equipment, including pressuring means and measuring
instruments
o Air temperature and barometric pressure during the test
o Building and project reference
o Design of installed ductwork including dimensions, thickness
of materials, types of stiffening, length, type of duct/tubes and
fittings, assembly method and distance of hangers/supports
o Required air tightness class and design operating pressure of
the installed ductwork
o Installer of ductwork
o Manufacturer of the ductwork
o Measured values of:
1. Ductwork surface area (A)
2. Total joint length (L)
3. Test pressure (p
test
)
4. Leakage flow rate (qv) corrected for temperature and
barometric pressure
5. Pressurizing time
o Calculated values of
1. Leakage factor (f)
2. Air leakage limit (f
table above at the measured test pressure (p
) according to the formulas given in
max
)
test
o Air tightness class achieved
24 Appendix B
Page 31
For tests including several test pressures it is recommended to
Duct Class
½-, 1-, 2-inwg
3-inwg
4-, 6-, 10-inwg
Seal Class
C B A
Sealing
Applicable
Transverse
Joints Only
Transverse
Joints and
Seams
Joints, Seams
and All Wall
Penetrations
Leakage Class
Rectangular
Metal
24
12
6
Round
Metal
12 6 3
plot the leakage factors as a function of test pressure in a
diagram together with the air leakage limit curve.
US Standards
Ductwork classification and maximum air leakage
Maximum air leakage is then defined as
0.65
F=CLP
where: F = Maximum air leakage (cfm/100 ft2)
CL = Leakage class
P = Pressure (inwg)
Leakage Testing Standards Highlights 25
Figure 12. Allowable Air Duct Leakage from Round Ducts,
per SMACNA Standard
Page 32
Figure 13. Allowable Air Duct Leakage from Rectangular Ducts,
per SMACNA Standard
The SMACNA standard does not generally require correcting
leakage flow rates to standard conditions, unless:
1. Air temperature <40ºF or >100ºF
2. Elevation <1500 ft above sea level
3. Duct static pressure <-20 inwg or >+20 inwg
Should one of these conditions not be satisfied, then correcting
the leakage to standard conditions may be done using one of
these formulas:
1. ACFM=SCFM * (460+T)/530
where T = actual dry bulb air temperature (ºF)
moisture is negligible
pressure between -20 and +20 inwg
2. ACFM = SCFM * 0.075/d
where d = air density from psychrometric chart
The SMACNA standard does not specify the information to be
reported, but instead defers to project documents. However, the
SMACNA standard does include a sample test report that
includes:
o Test date and place
o Test personnel and witness
o Building and project reference
o Duct section tested
o Specified leakage class, test pressure and duct construction
pressure class
o Measurements of:
Ductwork surface area
Leakage flow and calculations required to determine
leakage flow
Leakage Testing Standards Highlights 27
Page 34
(This page intentionally left blank)
28 Appendix B
Page 35
Appendix C
Typical Setup
29
Page 36
(This page intentionally left blank)
30 Appendix C
Page 37
USA Tel: +1 800 874 2811
UK Tel: +44 149 4 459200
France Tel: +33 4 91 11 87 64
Germany Tel: +49 241 523030
India Tel: +91 80 67877200
China Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
TSI Incorporated – Visit our website www.tsi.com for more information.