TSI Incorporated / 2008-2010 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective July 2000)
Seller warrants the goods sold hereunder, under normal use and service as described in the
operator's manual, shall be free from defects in workmanship and material for twenty-four (24)
months, or the length of time specified in the operator's manual, from the date of shipment to the
customer. This warranty period is inclusive of any statutory warranty. This limited warranty is
subject to the following exclusions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of
shipment.
b. Carbon monoxide (CO) Electro-chemical sensors are warranted for 12 months from the
date of shipment.
c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment.
d. Seller does not provide any warranty on finished goods manufa ctured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty
applies.
e. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty
with respect to, and shall have no liability in connection with, goods which are
incorporated into other products or equi pment, or which are modified by any person other
than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated
herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR
BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED
ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL
BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE
PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE
GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL,
CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE
RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS
OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12
months after a cause of action has accrued. The goods returned under warranty to Seller's
factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This
LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms
waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our
customers, our service policy is designed to give prompt attention to any problems. If any
malfunction is discovered, please contact your nearest sales office or representative, or call
Customer Service department at (800) 874-2811 (USA) or (1) 651-490-2811 (International).
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............ 1
APPENDIX A SPECIFICATIONS ..................................................... 13
iii
(This page intentionally left blank)
iv
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components below. If
anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
3. USB cable
4. CD-ROM with downloading software
5. Calibration collar
6. AC adapter
1
2
(This page intentionally left blank)
Chapter 1
Chapter 2
Setting-up
Supplying Power to the Model 7545 IAQ-CALC Indoor Air Quality
Meter
The Model 7545
or the optional AC adapter.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on the
inside of the battery compartment. The Model 7545 is designed to
operate with either alkaline or NiMH rechargeable batteries, although it
will not recharge NiMH batteries. Battery life will be s horter if NiMH
batteries are used. Carbon-zinc batteries are not recommended because
of the danger of battery acid leakage.
Using the Optional AC Adapter
When using the AC adapter, the batteries (if installed) will be bypassed.
Be sure to provide the correct voltage and frequency, which is marked
on the back of the AC adapter. The AC adapter is not a battery charger.
can be powered in one of two ways: four size AA batteries
Using The Probe
The sensing probe relies on the diffusion of air. For best results, try to keep
the sensing probe surrounded by moving air. Do not breathe on the probe.
Humans exhale CO
the probe to re-stabilize. Use the probe holder to support the probe when in
continuous data logging mode.
Connecting to a Computer
Use the USB Cable provided with the Model 7545 to connect the instrument
to a computer. Connect the end labeled “COMPUTER” to the computer
USB port and the other end to the data port of the Model 7545.
For more information on how to download stored data see Chapter 3 section
titled LogDat2™ Downloading Software
Caution: This symbol is used to indicate that the data port
of the Model 7545 is not intended for connection to a
public telecommunications network. Connect the USB
data port only to another USB port.
levels exceeding 10,000 ppm and it may take time for
2
.
3
(This page intentionally left blank)
4 Chapter 2
Chapter 3
Operation
Keypad Functions
ON/OFF Key
Arrow () Keys
(Enter) Key
Arrow ( or ) and
Menu Soft Keys
Common Terms
In this manual there are several terms that are used in different places. The
following is a brief explanation of the meanings of those terms.
Sample
Press to turn the Model 7545 on and off. During
the power up sequence the display will show the
following: Model Number, Serial Number,
Software Revision and Last Date Calibrated .
Press to scroll through choices while setting a
parameter. Pressing the
will lock the keypad to prevent unauthorized
adjustments to the instruments. To unlock the
keypad, press the
Press to accept a value or condition.
Press arrow keys to change choices while setting a
parameter. Press the Menu soft key to select the
Menu selections, which are Display Setup,
Settings, Barometric Pressure, Data Logging,
Applications, Calibration.
Consists of all of the measurement parameters
stored at the same time.
keys simultaneously
keys simultaneously.
Test ID
A group of samples. The statistics (average,
minimum, maximum, and count) are calculated for
each test ID. The maximum number of test IDs is
100.
5
Time Constant
The time constant is an averaging period. It is used
to dampen the display. If you are experiencing
fluctuating flows, a longer time constant will slow
down those fluctuations. The display will update
every second, but the displayed reading will be the
average over the last time constant period. For
example, if the time constant is 10 seconds, the
display will update every second, but the displayed
reading will be the average from the last 10
seconds. This is also referred to as a “moving
average”.
Logging Interval
The logging interval is a frequency period that the
instrument will log readings. For example, if the
logging interval is set to 30 minutes, each sample
will be the average of the last 30 minutes.
Menus
DISPLAY SETUP
Display setup menu is where you will setup the desired parameters to be
displayed on the running screen. With a parameter highlighted you can then
use the ON soft key to have it show up on the running screen or select the
OFF soft key to turn off the parameter. Use PRIMARY soft key to have a
parameter show up on the running screen in a larger display. Only one
parameter can be selected as a primary, and up to 2 secondary parameters
can be selected at one time.
SETTINGS
Settings menu is where you can set the general settings. These include
Language, Beeper, Select Units, Time Constant, Contrast, Set Time, Set
Date, Time Format, Date Format, Number Format, Backlight and Auto Off.
Use the or soft keys to scroll through the settings for each option and
use the
key to accept settings.
BAROMETRIC PRESSURE
Barometric Pressure menu is where you set the barometric pressure, which
can affect the CO
measurement.
2
ZERO CO
Zero CO menu item sets the baseline (zero) reading for the CO sensor. Only
zero the CO sensor in fresh air, where no CO is present, or with the probe
connected to zero gas.
6 Chapter 3
Mode
Log Settings
Auto-save
Log Interval
Cont-key
Log Interval
DATA LOGGING
Measurements
Measurements to be logged are independent of measurements on the
display, and must therefore be selected under DATA LOGGING
Measurements.
Log Mode/Log Settings
You can set Log Mode to Manual, Auto-save or Cont-key.
•Manual mode does not automatically save data, but instead prompts
the user to save a sample.
•In Auto-save mode, the user manually takes samples that are
automatically logged.
•In Cont-key mode, the user starts taking readings and logging by
pressing the
measurements until the
key. The instrument will continue taking
key is pressed again.
•Auto-save and Cont-Key modes have the following additional Log
Settings:
APPLICATIONS
You can choose % Outside Air in the Applications menu. After choosing
this application, take measurements or enter data for each line.
LogDat2™ Downloading Software
The Model 7545 comes with special software called LogDat2
Downloading Software, which is designed to provide you with maximum
flexibility and power. To install this software on your computer, follow the
instructions on the label of the LogDat2
Operation7
•Pressing the keys simultaneously will lock the keypad to
prevent unauthorized adjustments to the instruments. To unlock the
keypad, press the
keys simultaneously.
Delete Data
Use this to delete all data, delete test or delete sample.
% Memory
This option displays the memory available. Delete All, unde r Delete
Data, will clear memory and reset the memory available.
™
™ CD-ROM.
To download data from the Model 7545, connect the supplied computer
interface USB cable to the Model 7545 and to a computer USB port. Then
run the LogDat2
™ downloading software. Within the LogDat2™ software,
either select the tests to be downloaded or double-click on a test to open it.
8 Chapter 3
U.S. & International
TSI Incorporated
500 Cardigan Road
Shoreview MN 55126-3996
Tel: (800) 874-2811
(651) 490-2811
Fax: (651) 490-3824
Chapter 4
Maintenance
The Model 7545 requires very little maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model 7545 to TSI for annual recalibration.
Please contact one of TSI’s offices or your local distributor to make service
arrangements and to receive a Return Material Authorization (RMA)
number. To fill out an online RMA form, visit TSI’ s website at
http://service.tsi.com
The Model 7545 can also be recalibrated in the field using the
CALIBRATION menu. These field adjustments are intended to make minor
changes in calibration to match a user's calibration standards. The field
adjustment is NOT intended as a complete calibration capability. For
complete, multiple-point calibration and certification, the instrument must be
returned to the factory.
.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft
cloth and isopropyl alcohol or a mild detergent. Never immerse the
Model 7545. If the enclosure of the Model 7545 or the AC adapter becomes
broken, it must be replaced immediately to prevent access to hazardous
voltage.
Storage
Remove the batteries when storing the unit for more than one month to
prevent damage due to battery leakage.
9