TSI Alnor CF910 Operating Manual

Indoor Air Quality

Indoor Air Quality Meter

Alnor® Model CF910
AIRFLOWTM Model IAQ910
Operation and Service Manual
Copyright©
TSI Incorporated / 2007-2008 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective July 2000)
Seller warrants the goods sold hereunder, under normal use and service as described in the operator's manual, shall be free from defects in workmanship and material for twenty-four (24) months, or the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions: a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of shipment.
b. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of shipment.
c. Seller does not provide any warranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies.
d. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty
with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call Customer Service department at +44 (0) 149 4 459200 (UK), (800) 874-2811 (USA), or (1) 651­490-2811 (International).
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............
CHAPTER 2 SETTING-UP................................................................. 3
Supplying Power to the Alnor CF910/AIRFLOW IAQ910 .................3
Installing the Batteries ............................................................... 3
Using the Integral Probe............................................................ 3
CHAPTER 3 OPERATION .................................................................5
Keypad Functions ........................................................................... 5
CHAPTER 4 MAINTENANCE............................................................7
Recalibration...................................................................................7
Cases..............................................................................................7
Storage ...........................................................................................7
CHAPTER 5 TROUBLESHOOTING.................................................. 9
APPENDIX A SPECIFICATIONS..................................................... 11
1
iii
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
3. Calibration collar
1
Chapter 2
Setting-up

Supplying Power to the Alnor CF910/AIRFLOW IAQ910

The Alnor CF910/A

Installing the Batteries

Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The Alnor CF910/AIRFLOW IAQ910 is designed to operate with either alkaline or NiMH rechargeable batteries, although it will not recharge NiMH batteries. Battery life will be shorter if NiMH batteries are used. Carbon-zinc batteries are not recommended because of the danger of battery acid leakage.

Using the Integral Probe

The sensing probe relies on the diffusion of air. For best results, try to keep the sensing probe surrounded by moving air. Do not breathe on the probe. Humans exhale CO take time for the probe to re-stabilize.
IRFLOW IAQ910 is powered with four size AA batteries.
levels exceeding 10,000 ppm and it may
2
3
Chapter 3
Operation

Keypad Functions

ON/OFF Key
Sample Key
Stats Key
CAL Key
Press to turn the Alnor CF910/A on and off. During the power up sequence the display will show the following: Model Number, Serial Number, Software Revision, and Last Date Calibrated.
Press and release to start the sample, then press the END key to stop the sample.
Press to view the minimum, maximum and average of the last sample taken.`
Press to begin the field-calibration process. To re­calibrate in the field, first connect zero gas to the instrument and open the regulator. Then use the
ST and ENTER keys to select ZERO. The
instrument will automatically make zero itself after one minute.
After zeroing the sensor, connect span gas to the instrument and open the regulator. Then use the
ST and ENTER keys to select SPAN. The
instrument will take one minute to make a reading. Then use the concentration displayed on the instrument to match the concentration on the gas cylinder.
ST and ENTER keys to adjust the
IRFLOW IAQ910
5
Chapter 4
Maintenance
The Alnor CF910/AIRFLOW IAQ910 requires very little maintenance to keep it performing well.

Recalibration

To maintain a high degree of accuracy in your measurements, we recommend that you return your Alnor CF910/A annual recalibration. Please contact one of TSI’s offices or your local distributor to make service arrangements and to receive a Return Material Authorization (RMA) number. To fill out an online RMA form, visit TSI’s website at
U.S. & International UK
TSI Incorporated Alnor Products Stirling Road 500 Cardigan Road Cressex Business Park Shoreview MN 55126-3996 High Wycombe Tel: (800) 874-2811 Bucks (651) 490-2811 HP12 3RT United Kingdom Fax: (651) 490-3824 Tel: +44 (0) 149 4 459200 Fax: +44 (0) 149 4 459700
The Alnor CF910/A using the CALIBRATION menu. These field adjustments are intended to make minor changes in calibration to match a user's calibration standards. The field adjustment is NOT intended as a complete calibration capability. For complete, multiple-point calibration and certification, the instrument must be returned to the factory.
http://service.tsi.com.
IRFLOW IAQ910 can also be recalibrated in the field
IRFLOW IAQ910 to TSI for
TSI Instruments Ltd.

Cases

If the instrument case or storage case needs cleaning, wipe it off with a soft cloth and isopropyl alcohol or a mild detergent. Never immerse the Alnor CF910/A IAQ910, it must be replaced immediately to prevent access to hazardous voltage.

Storage

Remove the batteries when storing the unit for more than one month to prevent damage due to battery leakage.
IRFLOW IAQ910. If the enclosure of the Alnor CF910/AIRFLOW
7
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the Alnor CF910/A IAQ910. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI.
Table 5-1: Troubleshooting the Alnor CF910/A
Symptom Possible Causes Corrective Action
No Display Unit not turned on Switch unit on. Low or dead batteries Replace batteries or plug
Dirty battery contacts Clean the battery
High CO2 levels Breathing on probe Shield probe from breath
Instrument Error message appears
Fault in instrument Factory service required
IRFLOW IAQ910
in AC adapter.
contacts.
and allow instrument to re-stabilize.
on instrument.
WARNING!
Remove the probe from excessive temperature immediately: excessive heat can damage the sensor. Operating temperature limits can be found in Appendix A, Specifications.
IRFLOW
9
Appendix A
Specifications
Specifications are subject to change without notice.
CO
:
2
Range: 0 to 5000 ppm Accuracy Resolution: 1 ppm Sensor type: Non-Dispersive Infrared (NDIR)
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C) Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters Relative humidity up to 80% RH, non-condensing Pollution degree 1 in accordance with IEC 664 Transient over voltage category II
External Meter Dimensions:
3.3 in. × 9.6 in. × 1.8 in. (8.4 cm × 24.4 cm × 4.4 cm)
Meter Weight:
Weight with batteries: 0.6 lbs (0.27 kg)
1
: ±3% of reading or ±50 ppm, whichever is greater
Meter Display Dimensions:
Primary display: 4-digit LCD, 0.6 in. (15 mm) digit height Secondary display: 3.5-digit LCD, 0.3 in. (8 mm) digit height
Power Requirements:
Four AA-size batteries (included)
1
At 77°F (25°C). Add uncertainty of ±0.2%/°F (±0.2%/°C) away from calibrated temperature
11
TSI Incorporated 500 Cardigan Road Shoreview, MN 55126 USA Toll Free (800) 874-2811 Telephone (651) 490-2811 Fax: (651) 490-3824
E-mail: customerservice@alnor.com Web: www.alnor.com
France Tel: +33 491 95 21 90 E-mail: tsifrance@tsi.com Germany Tel: +49 241 523030 E-mail: tsigmbh@tsi.com
Part No. 1980591 Rev. C Copyright © 2008 by TSI Incorporated
TSI Instruments Ltd. Stirling Road Cressex Business Park High Wycombe Bucks HP12 3RT United Kingdom Tel: +44 (0) 149 4 459200 Fax: +44 (0) 149 4 459700
E-mail: info@airflowinstruments.co.uk Web: www.airflowinstruments.co.uk
Quality Assured to ISO 9001:2000
Website: www.tsiinc.fr
Website: www.tsiinc.de