Copyright
TSI Incorporated / 2007-2016 / All rights reserved.
Limitation Of Warranty And Liability (effective February 2015)
Seller warrants the goods, excluding software, sold hereunder, under normal use and
service as described in the operator's manual, to be free from defects in workmanship
and material for 24 months, or if less, the length of time specified in the operator's
manual, from the date of shipment to the customer. This warranty period is inclusive of
any statutory warranty. This limited warranty is subject to the following exclusions and
exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of
shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s
manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's
warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture.
Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the
manual allow an operator to replace consumables or perform recommended
cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned
according to the requirements of the manual.Unless specifically authorized in a
separate writing by Seller, Seller makes no warranty with respect to, and shall have no
liability in connection with, goods which are incorporated into other products or
equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS
stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT,
SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND
EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE
PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S
OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER
OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS
BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR
OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND
OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR
REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL
REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL
REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL
SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY
SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER
SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR
REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be
brought against Seller more than 12 months after a cause of action has accrued. The
goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and
will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY
AND LIABILITY, which contains the complete and exclusive limited warranty of Seller.
This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or
its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are
to our customers, our service policy is designed to give prompt attention to any
problems. If any malfunction is discovered, please contact your nearest sales office or
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components
below. If anything is missing or damaged, notify TSI immediately.
Supplying Power to the Alnor AXD610/Airflow Instruments
PVM610
The Alnor AXD610/Airflow Instruments PVM610 is powered with four
size AA batteries.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on
the inside of the battery compartment. The Alnor AXD610/Airflow
Instruments PVM610 is designed to operate with either alkaline
or NiMH rechargeable batteries, although it will not recharge
NiMH batteries. Battery life will be shorter if NiMH batteries are
used. Carbon-zinc batteries are not recommended because of
the danger of battery acid leakage.
Press to turn the Alnor AXD610/Airflow
Instruments PVM610 on and off. During the
power up sequence the display will show the
following: Model Number, Serial Number,
Software Revision and Last Date Calibrated.
ft/min / m/s Key
Pressing this key changes the display to read
air velocity, converted from velocity pressure
when connected to a pitot probe.
in H2O / Pa / hPa /
kPa / mm HG Key
Pressing this key changes the display to read
differential pressure.
0 Key
Pressing the 0 key zeroes the pressure
measurement.
Changing Units
To change units, first put the desired
measurement (air velocity or pressure) on the
display. Then press and hold the left,
unlabelled key for five seconds. Finally, use
the and ENTER key to select the units.
The Alnor AXD610/Airflow Instruments PVM610 requires very little
maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Alnor AXD610/Airflow Instruments
PVM610 to TSI for annual recalibration. Please contact one of TSI’s
offices or your local distributor to make service arrangements and to
receive a Return Material Authorization (RMA) number. To fill out an
Cases
If the instrument case or storage case needs cleaning, wipe it off with
a soft cloth and isopropyl alcohol or a mild detergent. Never immerse
the Alnor AXD610/Airflow Instruments PVM610. If the enclosure of
the Alnor AXD610/Airflow Instruments PVM610 becomes broken, it
must be replaced immediately to prevent access to hazardous
voltage.
Storage
Remove the batteries when storing the unit for more than one month
to prevent damage due to battery leakage.
Table 5-1 lists the symptoms, possible causes, and recommended
solutions for common problems encountered with the Alnor
AXD610/Airflow Instruments PVM610. If your symptom is not listed,
or if none of the solutions solves your problem, please contact TSI.
Table 5-1: Troubleshooting the Alnor AXD610/Airflow
Instruments PVM610
WARNING!
The pressure sensor is protected from damage up to 7 psi (48 kPa or
360 mmHg). At higher pressure it can burst!
Specifications are subject to change without notice.
Static/Differential Pressure:
Range1: -15 to +15 in. H2O (-28.0 to +28.0 mm Hg, -3735 to
+3735 Pa)
Accuracy: ±1% of reading ±0.005 in. H2O (±0.01 mm Hg, ±1 Pa)
Resolution: 0.001 in. H2O (0.1 Pa, 0.01 mm Hg)
Velocity From a Pitot Tube:
Range2: 250 to 15500 ft/min (1.27 to 78.7 m/s)
Accuracy3: ±1.5% at 2000 ft/min (10.16 m/s)
Resolution: 1 ft/min (0.1 m/s)
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C)
Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Pollution degree 1 in accordance with IEC 664
Transient over voltage category II
External Meter Dimensions:
3.3 in. 7.0 in. 1.8 in. (8.4 cm 17.8 cm 4.4 cm)
Meter Weight:
Weight with batteries: 0.6 lbs (0.27 kg)
Power Requirements:
Four AA-size batteries (included)
1
Overpressure range = 7 psi (190 in. H2O, 360 mmHg, 48 kPa).
2
Pressure velocity measurements are not recommended below 1,000 ft/min (5 m/s)
and are best suited to velocities over 2000 ft/min. Range can vary depending on
barometric pressure.
3
Accuracy is a function of converting pressure to velocity. Conversion accuracy