TSI PAN300 Service Manual

Page 1
POSITIVE AND NEGATIVE
DUCT ACCREDITATION
(PANDA) SYSTEM
AIRFLOW™ INSTRUMENTS

OPERATION AND SERVICE MANUAL

P/N 6005447, REV B
MARCH 2021
Page 2
Page 3
Copyright TSI Incorporated / 2011-2021 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824 LIMITATION OF WARRANTY AND LIABILITY (effective February 2015)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator's manual, to be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s
manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT,
SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE
PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE,
Page 4
SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI's Customer Service department at (800) 680-1220 (USA) or (001 651) 490-2860 (International) or visit www.tsi.com.
Trademarks
TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations.
Page 5
iii
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............. 1
CHAPTER 2 PREPARING PAN300 SYSTEM FOR
AIR DUCT LEAK TESTING ................................................................ 5
CHAPTER 3 PERFORMING A DUCT LEAKAGE TEST ................ 11
Measuring Duct Static Pressure ................................................... 11
Measuring Duct Leakage Flow ..................................................... 12
Turning on the PAN300 Duct Leakage Tester.............................. 12
Using Leakage Test Application in the Model TA465-P ............... 13
Troubleshooting Guide ................................................................. 16
APPENDIX A SPECIFICATIONS ..................................................... 17
APPENDIX B LEAKAGE TESTING STANDARDS HIGHLIGHTS ... 21
Standards Supported .................................................................... 21
EU Standards ............................................................................... 22
US Standards ............................................................................... 25
APPENDIX C TYPICAL SETUP....................................................... 29
APPENDIX D PROCEDURE FOR USING SMOKE PELLETS
IN LEAKAGE TESTERS ................................................................... 31
Page 6
iv
(This page intentionally left blank)
Page 7
1
Chapter 1
Unpacking and Parts Identification
Carefully unpack the PANDA system and instrument cases from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI® Incorporated (TSI®) immediately.
The PANDA system consists of the following:
Qty
Description
Part Number
Reference Picture
1
Low flow nozzle
6002598
1
Primary duct adapter spigot plus rubber bung (to fit to test duct)
6002638
1
Cam lock primary spigot (to connect flexi-duct to PANDA)
6002607
2
Ø4-in. (100-mm) adjustable over lock straps
6002683
1
13-ft (4-m) long Ø4-in. (100-mm) plastic flexible duct
6002667 2
20-in. (500-mm) silicone tubes (red)
AFL9020004
Page 8
2 Chapter 1
Qty
Description
Part Number
Reference Picture
2
20-in. (500-mm) silicone tubes (blue)
AFL9020005
1
16-ft (5-m) silicone tube (blue)
AFL9020005 1
Smoke cap holder assembly
AFL71549801
1
Smoke pellets
AFL9004167 1
K-type thermocouple probe
AFL82859201
3
Instrument adapter
AFL82859401
1
Operation and User’s
manual
6005447
Page 9
Unpacking and Parts Identification
3
The following two instruments should be used in conjunction with the PANDA unit:
TA465-P Multi­function Instrument
Refer to TA465 Operation and Service Manual supplied with the instrument for additional parts supplied as standard.
PVM620 Micromanometer
Refer to PVM620 Operation and Service Manual supplied with the instrument for additional parts supplied as standard.
IMPORTANT
Read Before Using the PANDA for the First Time
It is IMPORTANT that the PANDA be connected to the power supply using a 30 mA residual-current device (RCD) or Ground Fault Circuit Interrupter (GFCI).
It is IMPORTANT that the 110V PANDA (yellow power socket) be connected only to 110V to 120V supplies. Connecting it to a higher voltage supply will permanently damage the inverter.
The 220V/240V PANDA (blue power socket) should only be connected to 200V to 240V power supplies.
When storing the PANDA in a vertical position, please ensure that the straps holding the instrument box are in place and tightened first.
The PANDA is designed so that the instrument box and duct carry tube are removable to lighten the load when lifting.
DANGER
Turn the isolation switch to the OFF position and wait 10 minutes prior to disconnecting power.
Page 10
4 Chapter 1
(This page intentionally left blank)
Page 11
5
Chapter 2
Preparing PAN300 System for Air Duct Leak Testing
The following procedure should be followed carefully so that safe and accurate leakage testing can be achieved:
Successfully completing a duct leakage test requires certain information be compiled prior to starting the test. Refer to
Appendix B for a discussion of standards relating to duct
leakage testing. The list below indicates the information required:
Type of leakage test to be performed (Positive or negative).
Leakage standard to be followed.
Air tightness/leakage class to be achieved
Amount of ductwork to be tested, such as the complete
system or a statistical sample.
1. Select the section of the ductwork to be tested.
2. Calculate the surface area of the ductwork of the section to be tested.
3. Temporarily seal the all openings of the ductwork except one, which will be connected to the PAN300 duct leakage tester.
4. Position the PAN300 unit as close to the remaining opening in the ductwork as possible to minimize the flexible tubing needed. Minimize bends in the flexible tubing to reduce the pressure loss, giving the best performance.
5. Make sure the Fan Control Switch on the Fan Speed Controller is in the OFF position and the multi-turn Fan Speed Control potentiometer is fully turned counter-clockwise using the pictures of the VFD in Figure 1 as a reference. Plug the cord into the PANDA unit as shown in Figure 2 and Figure 3. Then connect the other end of the cord to a suitable electrical supply.
Page 12
6 Chapter 2
CAUTION
Remove the power cord from the PANDA duct leakage tester before tilting it to the vertical position to avoid damaging the cord.
110 V 230 V
Figure 1. 110V and 230V Fan Speed Controllers
NOTE
The control pad on the Fan Speed Controller has been disabled.
Figure 2. Receptacle for
Power Cord
Figure 3. Connected Power Cord
Fan Speed
Control
Fan Control
Switch
On/Off Switch
Page 13
Preparing PAN300 System for Air Duct Leak Testing
7
6. Fit the primary duct adapter spigot (black sheet metal with rubber bung) to one end of the 4-in. (100-mm) diameter flexi-tube. Make an air-tight seal using one of the over lock straps and lever­locking cam provided as shown in Figure 4. Adjust the fit of the over lock strap with a screwdriver.
7. Securely attach the black primary duct adapter spigot/flexi-tube assembly to the opening on the ductwork to be pressure tested.
Figure 4. Flex Ductwork
Connected to Primary Adapter
Spigot. Bung not shown.
8. If the static pressure tap on the black Primary Duct Adapter is open to the duct, connect the 16-ft (5-m) long blue silicone tube to it as shown in Figure 5.
If the static pressure tap on the black Primary Duct Adapter is not open to the duct, drill a 4-mm hole in the duct and insert about 6 inches (10 mm) of the silicone tube into the duct. Seal around the hole with putty.
9. Connect the other end of the 4-in. (100-mm) flexi-tube to
Figure 5. Connecting Pressure
Tubing to Tap on Primary Duct
Adapter. Bung not shown.
the cam lock connector (grey cast aluminum without nozzle). Make an airtight seal using the other over lock strap and lever­locking cam provided. Adjust the fit of the over lock strap with a screwdriver.
10. Determine if you are going to perform a high- or low-flow testing and positive or negative testing. Set-up the duct leakage tester by:
Page 14
8 Chapter 2
a. For positive pressure,
high-flow testing, remove the low flow nozzle if it is installed. Then, connect the grey cast-aluminum cam lock connector to the outlet side of the blower per Figure 6. Close both cam lock arms at the same time to ensure proper fit.
Connect the free end of the 16-ft (5-m) silicone
Figure 6. Positive Pressure,
High-Flow Set-Up
tube to the pressure tap marked DUCT PRESSURE P3 (+) on the black square box.
Finally, verify that the FLOW GRID MODE pressure taps on the inlet tubes are connected to the FLOW GRID pressure taps, i.e., P1 (+) to P1 (+) and P2 (-) to P2 (-).
b. For positive pressure,
low-flow testing, add the low-flow nozzle to the blower inlet if it is not installed per Figure 7. Then, connect the grey cast­aluminum cam lock connector to the outlet side of the blower per Figure 6. Close both cam lock arms at the same time to ensure proper fit.
Figure 7. Positive Pressure,
Low-Flow Set-Up
Connect the free end of the 16-ft (5-m) silicone tube to the pressure tap marked DUCT PRESSURE P3 (+) on the black square box.
Fan
Outlet
Cam Lock Connector
Low Flow
Nozzle
Fan Inlet
Page 15
Preparing PAN300 System for Air Duct Leak Testing
9
c. For negative pressure,
high-flow testing, remove the low flow nozzle if it is installed. Then, connect the grey cast aluminum cam lock connector to the inlet side of the blower per Figure 8. Close both cam lock arms at the same time to ensure proper fit.
Connect the free end of
Figure 8. Negative Pressure,
High-Flow Set-Up
the 16-ft (5-m) silicone tube to pressure tap marked DUCT PRESSURE P4 (-) on the black square box.
Finally, verify that the FLOW GRID MODE pressure taps on the inlet tubes are connected to the FLOW GRID pressure taps, i.e., P1 (+) to P1 (+) and P2 (-) to P2 (-).
d. For negative pressure,
low-flow testing, add the low-flow nozzle to the blower inlet if it is not installed. Then, connect the grey cast aluminum cam lock connector to the low­flow nozzle per Figure 9. Close both cam lock arms at the same time to ensure proper fit.
Connect the free end of
Figure 9. Negative Pressure,
Low-Flow Set-Up
the 16-ft (5-m) silicone tube to pressure tap marked DUCT PRESSURE P4 (-) on the black square box.
Fan Inlet
Cam Lock Connector
Fan Inlet
Low Flow
Nozzle
Cam Lock Connector
Page 16
10 Chapter 2
(This page intentionally left blank)
Page 17
11
Chapter 3
Performing a Duct Leakage Test
The PAN300 duct leakage test system includes a Model PVM620 Micromanometer and a Model TA465-P Ventilation Meter. During duct leakage testing, the Model PVM620 Micromanometer measures the duct static pressure while the Model TA465-P Ventilation Meter measures the airflow rate.
Refer to the Operation and Service Manuals for the Model PVM620 Micromanometer and the Model TA465-P Ventilation Meter to use these instruments in other applications. If you do not have the manuals, download them from TSI® Incorporated’s website
www.tsi.com.

Measuring Duct Static Pressure

1. Turn ON the Model PVM620.
2. Zero the Model PVM620 pressure sensor with both ports open to
the atmosphere.
3. Connect the (+) port on the Model PVM620 to P3 (+) to measure
the duct static pressure (see Figure 6).
4. Leave the (-) port on the Model PVM620 open to the atmosphere.
NOTE
Refer to the Model PVM620 Operation and Service Manual for instructions on using data logging to automatically record the duct static pressure.
Page 18
12 Chapter 3

Measuring Duct Leakage Flow

Turning on the PAN300 Duct Leakage Tester

1. Switch the PANDA unit on. a. For 230V models position the Mains Power switch of the
inverter to the ON position to power the Inverter speed controller. The Fan motor is not energized.
b. 110V models DO NOT include a separate power switch. The
PANDA unit is turned on when the power cord is plugged in.
2. Position the Fan Control switch to the RUN position to energize the fan.
3. Increase the fan to the desired speed by turning the Fan Speed Controller clockwise. To decrease the fan speed, turn the Fan Speed controller counter-clockwise.
1. Turn ON Model TA465-P.
2. Zero the Model TA465-P pressure sensor with both ports open to the atmosphere.
3. Connect the Model TA465-P to the PAN300 by connecting the (+) and (-) ports on the Model TA465-P to the P1 (+) and P2 (+) ports located inside the black box of the PAN300 as shown in Figure 10.
Figure 10. Connecting Instruments
to PANDA Tester in
High Flow Mode
4. Connect the thermocouple to the Model TA465-P.
5. Insert the thermocouple probe into the blower inlet through the hole marked TC1 as shown in Figure 11.
Figure 11. Thermocouple
Insertion Hole
Thermocouple Hole
Page 19
Performing a Duct Leakage Test
13

Using Leakage Test Application in the Model TA465-P

CAUTION
The Model PVM620 and Model TA465-P meters must be zeroed before entering the Leakage Test Application.
1. Press the MENU key to access the menu system on the Model TA465-P.
2. Use the  keys to highlight the Applications item.
3. Press the (ENTER) key to access the Applications menu.
MENU Zero Press Display Setup
Settings
Flow Setup Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration Discover Printer
4. Select Leakage Test and press key.
APPLICATIONS Draft Rate Heatflow Turbulence % Outside Air
Leakage Test
5. Select either the EN Standard or SMACNA leakage test.
LEAKAGE TEST EN Standard SMACNA
Page 20
14 Chapter 3

Instrument Operation if EN Standard Test Protocol is Selected

1. Enter key parameters: a. Surface Area of ductwork
section to be tested.
b. Static Pressure of test, as
measured by Model PVM620 micromanometer.
c. Flow Device as Nozzle or
Flow Grid.
LEAKAGE TEST Surface Area Static Pressure Flow Device Tightness Class Test Length Run Test
d. Leakage class as A, B, C, or D. Note that tests with negative
pressures must be selected as negative tests, as indicated by -.
e. Test Length, or duration of leakage test, usually 5 minutes.
2. Increase the blower speed until the desired static pressure is achieved.
3. When the static pressure has stabilized, select Run Test and press .
4. The display will show the readings on the right. Leakage Factor and Leak Rate will update in real time, while other parameters will remain constant.
If the Leakage Factor and Leak Rate are sufficiently stable,
LEAKAGE TEST Leakage Factor x.xx Leak Limit x.xx Leak Rate x.xx
Status OK
Flow Device Flow Grid Baro Pressure 20.20 Temperature 20C Time 9:55
Standard TestXXX
Sample 0
(Sample Saved 1)
press the START soft key or the key to begin the leak test. Pressing the ESC key will exit back to the previous screen.
5. After the leak test is complete, the Model TA465-P will prompt you to press the SAVE or PRINT soft key. You can also press the ESC key here to back out to the previous screen without saving the data.
After completing leakage testing for a section of duct, you can move on to the next section.
Page 21
Performing a Duct Leakage Test
15

Instrument Operation if SMACNA Test Protocol is Selected

1. Enter key parameters: a. Surface Area of ductwork
section to be tested.
b. Static Pressure of test, as
measured by Model PVM620 micromanometer.
c. Flow Device as Nozzle or
Flow Grid.
d. Leakage class as a number
from 1 to 48. Typical values are 3, 6, 12, 24, or 48.
e. Test Length, or duration of
leakage test.
2. Increase the blower speed until the desired static pressure is achieved.
3. When the static pressure has stabilized, select Run Test and press .
4. The display will show the readings on the right. Leakage
LEAKAGE TEST Surface Area Static Pressure Flow Device Leakage Class Test Length Run Test
LEAKAGE TEST Leakage Factor x.xx Leak Limit x.xx Leak Rate x.xx Status OK
Flow Device Flow Grid Baro Pressure 20.20 Temperature 20C Time 9:55 Standard TestXXX
Sample
0 (Sample Saved 1) Stop(Save) Print
Factor and Leak Rate will update in real time, while other parameters will remain constant.
If the Leak Factor and Leak Rate are sufficiently stable, press the
START soft key or the key to begin the leak test. Pressing the ESC key will exit back to the previous screen.
5. After the leak test is complete, the Model TA465-P will prompt you to press the SAVE or PRINT soft key. You can also press the ESC key here to back out to the previous screen without saving the data.
After completing leakage testing for a section of duct, you can move on to the next section.
Page 22
16 Chapter 3
DANGER
Turn the isolation switch to the OFF position and wait 10 minutes prior to disconnecting power.

Troubleshooting Guide

Symptom
Recommended Action
Fan motor will not run.
Check the power connection.
Circuit Breaker may have
tripped.
Static pressure reading (on PVM620) is zero.
Check the connections.
Static pressure reading (on PVM620) is too low.
Required static pressure cannot be achieved with motor speed control settings at the maximum.
Leakage rate is too high. Check for leaks using soap bubbles or smoke pallets. Alternatively, test a smaller section of the ductwork.
Leak Flow (on TA465-P) shows flashing XXX.XX.
Check the pressure tube connections to the TA465-P meter.
Leak flow is too low. Use low flow nozzle adapter.
CAUTION
If recommended actions are used and the problem persists DO NOT service further. Return to TSI® Incorporated for servicing.
Page 23
17
Appendix A
Specifications
Pressure Measurement (PVM620)
Range ..................................
± 3,735 Pa (± 15 in. W.G.)
Resolution ...........................
0.1 Pa (0.001 in. W.G.)
Accuracy .............................
1% of reading ± 1 Pa (± 0.005 in. W.G.)
Actual duct static range .......
± 2,500 Pa (± 10 in. W.G.) at Zero Flow
Volume Flow Measurement (TA465-P)
Wilson radial flow grid .........
High leakage range: 10 to 200 l/s (36 to 720 m3/hr, 21 to 424 cfm)
15 mm conical inlet
nozzle adapter ....................
Low leakage range: 1 to 13 l/s (3.6 to 46.9 m3/hr, 2 to 27.5 cfm)
Resolution ...........................
0.01 l/s (0.01 m3/hr, 0.01 cfm)
Accuracy .............................
± 2.5% of reading ± 0.01 l/s (± 0.04 m3/hr, ± 0.02 cfm)
Temperature Measurement (TA465-P)
K Type thermocouple probe
To EN60584 (IEC 584)
Barometric Pressure Measurement (TA465-P)
Range ..................................
690 to 1,241 hPa (517.5 to 930.87 mm Hg,
20.36 to 36.648 in. Hg)
Accuracy .............................
± 2% of reading
Power Requirements
Model PAN331* ..................
220 to 240 V, 1 Phase, 50/60 Hz, 10A
Model PAN331-110* ...........
110 to 120 V, 1 Phase, 50/60 Hz, 16A
Model PAN311** .................
220 to 240 V, 1 Phase, 50/60 Hz, 10A
Model PAN311-110** ..........
110 to 120 V, 1 Phase, 50/60 Hz, 16A
Weight ................................
71 kg (157 lbs.)
Dimensions (L x W x H) ....
1,130 mm x 660 mm x 510 mm (44.5 in. x 26 in. x 20 in.)
TA465-P and PVM620 .......
See specification sheets for details on individual instruments
* Model: instruments included ** Model: instruments NOT included
(Specifications subject to change without notice.)
Page 24
18 Appendix A
0
500
1000
1500
2000
2500
3000
0 50 100 150 200 250
Static Pressure Pa.
Volume Flow l/s
Panda Fan Performance Graph (metric units)
Page 25
Specifications
19
0
1
2
3
4
5
6
7
8
9
10
11
0 50 100 150 200 250 300 350 400 450 500
Static Pressure ins.w g.
Volume Flow CFM
Panda Fan Performance Graph (imperial units)
Page 26
20 Appendix A
Environmental Requirements
Ambient Air Temperature ....
-10 to +40°C
Relative Humidity ................
0~95%, non-condensing
Altitude ................................
2000m or less
Transportation and Storage
Storage Temperature ..........
-10 to +40°C
IMPORTANT
Read Before Using the PANDA for the First Time
It is IMPORTANT that the PANDA be connected to the power supply using a 30 mA residual-current device (RCD) or Ground Fault Circuit Interrupter (GFCI).
It is IMPORTANT that the 110V PANDA (yellow power socket) be connected only to 110V to 120V supplies. Connecting it to a higher voltage supply will permanently damage the inverter.
The 220V/240V PANDA (blue power socket) should only be connected to 200V to 240V power supplies.
Page 27
21
Appendix B
Leakage Testing Standards Highlights
Different standards are used throughout the world to specify duct air tightness and leakage requirements. The PAN300 duct leakage test system has a duct leakage application to automatically compare the actual leakage flow with the maximum allowed leakage flow for EN and SMACNA standards. Field technicians can also use the duct leakage application to determine actual leakage flow and manually compare it to maximum leakage from another standard. The PAN300 duct leakage test system cannot determine the appropriate leakage classification for a given duct.

Standards Supported

Standard
US- or EU-
Based
Description
BS EN 12237:2003
EU
Ventilation for buildingsDuctwork Strength and leakage of circular sheet metal ducts.
BS EN 1507:2006
EU
Ventilation for buildingsSheet metal air ducts with rectangular section Requirements for strength and leakage.
DW/143
EU
HVACA practical guide to Ductwork leakage testing.
Eurovent 2/2
EU
Air leakage rate in sheet metal air distribution systems.
SMACNA HVAC Air Duct Leakage Test manual, First edition, 1985
US
Duct construction leakage classification, expected leakage rates for sealed and unsealed ductwork, duct leakage test procedures, recommendations on use of leakage testing, types of test apparatus and test setup and sample leakage analysis.
TSI® has made every effort to accurately reflect the standards referenced. Please refer to the actual standards for more detailed information and to make the best interpretation of each statement.
Page 28
22 Appendix B
The scope of the standards listed above includes many items other than duct leakage. This summary, however, is limited to duct leakage testing.

EU Standards

Ductwork classification and maximum air leakage. Note that EN1507, EN12237 Eurovent 2/2 and DW/143 all have the same formula to determine f
max
, the Air Leakage Limit, although DW/143 uses units of
l/s/m2 whereas others use m3/s/m2.
EN 1507 (rectangular ductwork)
Air
Tightness
Class
Air Leakage
Limit (f
max
)
m3/s/m2
Static Pressure Limit (ps) Pa
Negative
Positive at
pressure class
1 2 3
A 200
400
B 500
400
1000
2000
C 750
400
1000
2000
D* 750
400
1000
2000
* Class D ductwork is only for special apparatus
EN12237 (circular ductwork)
Air
Tightness
Class
Air leakage
limit (f
max
)
m3*/s/m2
Static Pressure Limit (ps)
Pa
Negative
Positive
A 500
500
B 750
1000
C 750
2000
D* 750
2000
* Class D ductwork is only for special apparatus
Page 29
Leakage Testing Standards Highlights 23
Eurovent 2/2 Air Tightness For Installed Duct Testing:
Air Tightness
Class
Air leakage limit (f
max
)
m3/s/m2
A B C
DW/143: A Practical Guide to Ductwork Leakage Testing
Duct
Pressure
Class
Static Pressure
Limit
Maximum
Air Velocity
m/s
Air leakage
limits l/s/m2
Positive
Pa
Negative
Pa
Low-pressure
– Class A
500
500
10
0.027*p
t
0.65
Medium-
pressure –
Class B
1000
750
20
0.009*p
t
0.65
High pressure
– Class C
2000
750
40
0.003*p
t
0.65
Figure 12. Allowable Air Leakage Rates from DW/143
Page 30
24 Appendix B
The measured leakage flow rates shall be corrected if the temperature and/or barometric pressure are different from standard conditions (+20ºC and 101 325 Pa) as follows:
where:
qv=corrected flow leakage rate q
measured
= measured flow leakage rate t = measured temperature (ºC) p = measured barometric pressure (Pa)
The test report shall give the following general information of the test performed:
o Date and place o Test personnel and witness o Test equipment, including pressuring means and measuring
instruments
o Air temperature and barometric pressure during the test o Building and project reference o Design of installed ductwork including dimensions, thickness
of materials, types of stiffening, length, type of duct/tubes and fittings, assembly method and distance of hangers/supports
o Required air tightness class and design operating pressure of
the installed ductwork
o Installer of ductwork o Manufacturer of the ductwork o Measured values of:
1. Ductwork surface area (A)
2. Total joint length (L)
3. Test pressure (p
test
)
4. Leakage flow rate (qv) corrected for temperature and barometric pressure
5. Pressurizing time
o Calculated values of
1. Leakage factor (f)
2. Air leakage limit (f
max
) according to the formulas given in
table above at the measured test pressure (p
test
)
o Air tightness class achieved
Page 31
Leakage Testing Standards Highlights 25
For tests including several test pressures it is recommended to plot the leakage factors as a function of test pressure in a diagram together with the air leakage limit curve.

US Standards

Ductwork classification and maximum air leakage
Duct Class
½-, 1-, 2-inwg
3-inwg
4-, 6-, 10-inwg
Seal Class
C B A
Sealing
Applicable
Transverse Joints Only
Transverse
Joints and
Seams
Joints, Seams and
All Wall
Penetrations
Leakage Class
Rectangular
Metal
24
12
6
Round
Metal
12 6 3
Maximum air leakage is then defined as F=CLP
0.65
where: F = Maximum air leakage (cfm/100 ft2)
CL = Leakage class P = Pressure (inwg)
Page 32
26 Appendix B
Figure 13. Allowable Air Duct Leakage from Round Ducts,
per SMACNA Standard
Figure 14. Allowable Air Duct Leakage from Rectangular Ducts,
per SMACNA Standard
The SMACNA standard does not generally require correcting leakage flow rates to standard conditions, unless:
1. Air temperature <40ºF or >100 ºF
2. Elevation <1500 ft above sea level
3. Duct static pressure <-20 inwg or >+20 inwg
Page 33
Leakage Testing Standards Highlights 27
Should one of these conditions not be satisfied, then correcting the leakage to standard conditions may be done using one of these formulas:
1. ACFM=SCFM * (460+T)/530 where T = actual dry bulb air temperature (ºF) moisture is negligible pressure between -20 and +20 inwg
2. ACFM = SCFM * 0.075/d where d = air density from psychrometric chart
3. ACFM=lb dry air/minute * humid volume (ft3/lb dry air)
The SMACNA standard does not specify the information to be reported, but instead defers to project documents. However, the SMACNA standard does include a sample test report that includes:
o Test date and place o Test personnel and witness o Building and project reference o Duct section tested o Specified leakage class, test pressure and duct construction
pressure class
o Measurements of:
Ductwork surface area ▪ Leakage flow and calculations required to determine
leakage flow
Page 34
28 Appendix B
(This page intentionally left blank)
Page 35
29
Appendix C
Typical Setup
Page 36
30 Appendix C
(This page intentionally left blank)
Page 37
31
Appendix D
Procedure for Using Smoke Pellets in Leakage Testers
As shown in the sketch below, a rubber bung is fitted into the hole in the duct adapter which holds the wire coil.
When a smoke pellet is required to be used, remove the bung and fit a pellet into the wire coil as shown in the sketch.
Light the pellet and immediately plug into the hole in the duct adapter and proceed with the test.
The pellet should emit dense white smoke for up to a minute.
Duct
Pellet Holder
Smoke Pellet
Flexible Ducting
Duct Adapter
Page 38
32 Appendix D
(This page intentionally left blank)
Page 39
Page 40
TSI Instruments Ltd.
Visit our website at www.tsi.com for more information.
UK Tel: +44 149 4 459200 France Tel: +33 1 41 19 21 99
Germany Tel: +49 241 523030
P/N 6005447 Rev B ©2021 TSI Incorporated Printed in U.S.A.
Loading...