TSI PAN300 Service Manual

POSITIVE AND NEGATIVE
DUCT ACCREDITATION
(PANDA) SYSTEM
AIRFLOW™ INSTRUMENTS

OPERATION AND SERVICE MANUAL

P/N 6005447, REV B
MARCH 2021
Copyright TSI Incorporated / 2011-2021 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824 LIMITATION OF WARRANTY AND LIABILITY (effective February 2015)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator's manual, to be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s
manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT,
SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE
PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE,
SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI's Customer Service department at (800) 680-1220 (USA) or (001 651) 490-2860 (International) or visit www.tsi.com.
Trademarks
TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations.
iii
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............. 1
CHAPTER 2 PREPARING PAN300 SYSTEM FOR
AIR DUCT LEAK TESTING ................................................................ 5
CHAPTER 3 PERFORMING A DUCT LEAKAGE TEST ................ 11
Measuring Duct Static Pressure ................................................... 11
Measuring Duct Leakage Flow ..................................................... 12
Turning on the PAN300 Duct Leakage Tester.............................. 12
Using Leakage Test Application in the Model TA465-P ............... 13
Troubleshooting Guide ................................................................. 16
APPENDIX A SPECIFICATIONS ..................................................... 17
APPENDIX B LEAKAGE TESTING STANDARDS HIGHLIGHTS ... 21
Standards Supported .................................................................... 21
EU Standards ............................................................................... 22
US Standards ............................................................................... 25
APPENDIX C TYPICAL SETUP....................................................... 29
APPENDIX D PROCEDURE FOR USING SMOKE PELLETS
IN LEAKAGE TESTERS ................................................................... 31
iv
(This page intentionally left blank)
1
Chapter 1
Unpacking and Parts Identification
Carefully unpack the PANDA system and instrument cases from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI® Incorporated (TSI®) immediately.
The PANDA system consists of the following:
Qty
Description
Part Number
Reference Picture
1
Low flow nozzle
6002598
1
Primary duct adapter spigot plus rubber bung (to fit to test duct)
6002638
1
Cam lock primary spigot (to connect flexi-duct to PANDA)
6002607
2
Ø4-in. (100-mm) adjustable over lock straps
6002683
1
13-ft (4-m) long Ø4-in. (100-mm) plastic flexible duct
6002667 2
20-in. (500-mm) silicone tubes (red)
AFL9020004
2 Chapter 1
Qty
Description
Part Number
Reference Picture
2
20-in. (500-mm) silicone tubes (blue)
AFL9020005
1
16-ft (5-m) silicone tube (blue)
AFL9020005 1
Smoke cap holder assembly
AFL71549801
1
Smoke pellets
AFL9004167 1
K-type thermocouple probe
AFL82859201
3
Instrument adapter
AFL82859401
1
Operation and User’s
manual
6005447
Unpacking and Parts Identification
3
The following two instruments should be used in conjunction with the PANDA unit:
TA465-P Multi­function Instrument
Refer to TA465 Operation and Service Manual supplied with the instrument for additional parts supplied as standard.
PVM620 Micromanometer
Refer to PVM620 Operation and Service Manual supplied with the instrument for additional parts supplied as standard.
IMPORTANT
Read Before Using the PANDA for the First Time
It is IMPORTANT that the PANDA be connected to the power supply using a 30 mA residual-current device (RCD) or Ground Fault Circuit Interrupter (GFCI).
It is IMPORTANT that the 110V PANDA (yellow power socket) be connected only to 110V to 120V supplies. Connecting it to a higher voltage supply will permanently damage the inverter.
The 220V/240V PANDA (blue power socket) should only be connected to 200V to 240V power supplies.
When storing the PANDA in a vertical position, please ensure that the straps holding the instrument box are in place and tightened first.
The PANDA is designed so that the instrument box and duct carry tube are removable to lighten the load when lifting.
DANGER
Turn the isolation switch to the OFF position and wait 10 minutes prior to disconnecting power.
4 Chapter 1
(This page intentionally left blank)
5
Chapter 2
Preparing PAN300 System for Air Duct Leak Testing
The following procedure should be followed carefully so that safe and accurate leakage testing can be achieved:
Successfully completing a duct leakage test requires certain information be compiled prior to starting the test. Refer to
Appendix B for a discussion of standards relating to duct
leakage testing. The list below indicates the information required:
Type of leakage test to be performed (Positive or negative).
Leakage standard to be followed.
Air tightness/leakage class to be achieved
Amount of ductwork to be tested, such as the complete
system or a statistical sample.
1. Select the section of the ductwork to be tested.
2. Calculate the surface area of the ductwork of the section to be tested.
3. Temporarily seal the all openings of the ductwork except one, which will be connected to the PAN300 duct leakage tester.
4. Position the PAN300 unit as close to the remaining opening in the ductwork as possible to minimize the flexible tubing needed. Minimize bends in the flexible tubing to reduce the pressure loss, giving the best performance.
5. Make sure the Fan Control Switch on the Fan Speed Controller is in the OFF position and the multi-turn Fan Speed Control potentiometer is fully turned counter-clockwise using the pictures of the VFD in Figure 1 as a reference. Plug the cord into the PANDA unit as shown in Figure 2 and Figure 3. Then connect the other end of the cord to a suitable electrical supply.
6 Chapter 2
CAUTION
Remove the power cord from the PANDA duct leakage tester before tilting it to the vertical position to avoid damaging the cord.
110 V 230 V
Figure 1. 110V and 230V Fan Speed Controllers
NOTE
The control pad on the Fan Speed Controller has been disabled.
Figure 2. Receptacle for
Power Cord
Figure 3. Connected Power Cord
Fan Speed
Control
Fan Control
Switch
On/Off Switch
Loading...
+ 28 hidden pages