TSI 8380 Product information

ACCUBALANCE®
AIR CAPTURE HOOD
MODEL 8380
The 8380 AccuBalance® Air Capture Hood is a multipurpose electronic air balancing instrument primarily used for eciently taking direct air volume readings at diusers and grilles. It features a detachable micromanometer which can be used with optional probes for increased exibility in multiple measurement applications. Oering durable, trouble-free operation, this lightweight, ergonomically designed capture hood kit saves time and money by combining multiple measurement tools into one
®
package. The 8380 AccuBalance create healthy and energy ecient environments while meeting local codes, guidelines and regulations for ventilation systems.
Air Capture Hood helps you
Features and Benets
+ Ergonomic design and ultra light weight for easy,
one-person operation
+ Automatically senses and displays supply or
return ows, saving time on the job
+ Back pressure compensation ensures accurate readings
+ Multiple hood sizes available for easy, cost eective
use across multiple jobs
+ Detachable digital micromanometer oers exibility
to use in multiple applications
+ Includes Swirl X Flow Conditioner for use with twist
or swirl type supply air diusers
+ Compatible LogDat™ Mobile Remote Reader and Data Logger
Software option simplies documenting of results and
emailing of reports
+ Capture hood stand eliminates the need for ladders
(reaching diusers up to 4,5m (15 ft.)
Applications
+ Test and balance contractors
+ Commissioning agents
+ Facilities managers
+ Health and safety specialists
+ Ventilation system installers
UNDERSTANDING, ACCELERATED
DETACHABLE
MICROMANOMETER
MODEL 8715
The 8380 AccuBalance® Air Capture Hood includes a detachable 8715 micromanometer— one of the most advanced, versatile, and easy to use micromanometers on the market today. The 8715 features an auto-zeroing pressure sensor that increases measurement resolution and accuracy along with an intuitive menu structure for ease of operation.
Model 8715 (Micromanometer shown with standard and optional accessories)
Features and Benets
+ Accurately measures pressure, velocity and ow to help you
meet industry standards
+ Auto-zeroing pressure sensor reduces user-steps and saves time
+ Automatic density correction increases reading accuracy
+ Intuitive menu structure allows for ease of use and setup
+ Large graphic display with backlight oers easy-to-use interface
– Displays up to ve measurements simultaneously
– On-screen messages and instructions
– Programmed for multiple languages
+ Integrated Log-Tchebyche duct traverse application
simplies calculations
+ Bluetooth communications for transferring data
or remote polling
+ Optional LogDat™ Mobile Android™ App connects to the
instrument via bluetooth to remotely take readings and datalog
measurements for review or export
+ Includes downloading software with USB cable
+ Accommodates optional pitot, air ow (straight pitot),
temperature/relative humidity, velocity matrix, or
thermoanemometer probes for use in multiple applications
Plug and play thermoanemometer probes enables use in multiple applications.
SPECIFICATIONS
ACCUBALANCE® AIR CAPTURE HOOD MODEL 8380 DETACHABLE MICROMANOMETER MODEL 8715
Velocity Range
Pitot probes
0.125 to 78 m/s (25 to 15,500 ft/min)
Air ow probe 0.125 to 12.5 m/s (25 to 2,500 ft/min)
Velocity matrix
Accuracy
Units
0.125 to 12.5 m/s (25 to 2,500 ft/min) ±3% of reading ±0.04 m/s (±7 ft/min)
at velocities >0.25 m/s (50 ft/min) m/s, ft/min
Resolution 0.01 m/s (1 ft/min)
Pressure
Dierential pressure
Absolute pressure
Accuracy
Units
Resolution
±3735 Pa (±15 in. H maximum safe operating pressure
356 to 1016 mm Hg (15 to 40 in. Hg) ±2% of reading ±0.025 Pa H
dierential; ±2% of reading absolute in. H
O, in. Hg, Pa, hPa, kPa, mm Hg, cm Hg,
2
O, cm H2O
mm H
2
0.001 Pa H Hg (0.01 in. Hg) absolute
2
O); 37.5 kPa (150 in. H2O),
2
O (±0.0001 in.) static and
2
O (0.00001 in.) static and dierential; 1 mm
Volume
Range
Accuracy
Units m
Resolution
42 to 4250 m capture hood, supply and return
±3% of reading ±12 m >85 m
3
/h, ft3/min, l/s, m3/min
3
1 m
3
/h (25 to 2,500 ft3/min)
3
/h (>50 ft3/min)
/h (1 ft3/min)
3
/h (±7 ft3/min) at ows
Temperature
Sensor in base 4.4 to 60°C (40 to 140°F)
Temperature/RH probe -10 to 60°C (14 to 140°F)
Accuracy ±0.3°C (±0.5°F)
Units
Resolution
°C, °F
0.1°C (0.1°F)
RH
Range 5 to 95% RH (temperature/RH probe)
Accuracy ±3% RH
Resolution 0.1% RH
Instrument Temperature Range
Operating
Storage
4.4 to 60°C (40 to 140°F)
-20 to 71°C (-4 to 160°F)
Statistics
min, max, average and sum
Data Storage
26,500 samples, time and date stamped
Logging Interval
User selectable
Response Time
2 to 8 seconds, dierential pressure sensor
Power Requirements
Four AA-size cells or AC adapter
Physical Characteristics
Dimensions (micromanometer only)
Weight with Batteries
18.8 cm x 11.4 cm x 5.8 cm (7.4 in. x 4.5 in. x 2.3 in.)
8715 0.5 kg (17 oz.) 8380 3.4 kg (7.4 lb.)
6.35 mm (1/4 in.) OD straight ports with
Pressure Connection
barbed ends for use with 4.76 mm (3/16 in.) ID exible tubing
Model 8380-B 8380 8380-STA 8715
Description
Basic 610 mm x 610 mm (2 ft x 2ft) AccuBalance Capture Hood Kit
Standard 610 mm x 610 mm (2 ft x 2ft) AccuBalance Capture Hood Kit
Bundled 610 mm x 610 mm (2 ft x 2ft) AccuBalance Capture Hood Kit
Micromanometer
Capture hood base, poles, frame and fabric + + +
Micromanometer (4) support poles (6) support poles
(4) AA alkaline batteries
+ + + + +
+ +
+
(4) AA rechargeable NiMH batteries + + +
(2) battery holders + + + +
Multi-country AC power adaptor + + +
46 cm (18 in.) pitot probe
+ + +
5.0 m (16 ft.) tubing + + +
(2) static pressure probes
+ + +
Neck strap + + +
Capture hood stand +
Android Tablet loaded with LogDat Mobile +
Wheeled carrying case + + +
Handheld carrying case
LogDat CH downloading software with cable + + + +
User manual Calibration certicate, pressure: 5-points (dierential),
3-points (barometric), 3-points (temperature) Calibration certicate, ow:
7-points (supply), 7-points (return)
+ + + +
+ + + +
+ + +
Kit
Use the stand and tablet app to do single-person balancing of a system
+
SPECIFICATIONS
ACCUBALANCE® AIR CAPTURE HOOD MODEL 8380
DETACHABLE MICROMANOMETER MODEL 8715
Recommended Optional Accessories Optional Probes
Hood Kits
801180 (standard) 610 mm x 610 mm (2 ft x 2 ft)
801200 305 mm x 1220 mm (1 ft x 4 ft)
801216 610 mm x 915 mm (2 ft x 3 ft)
801201 610 mm x 1220 mm (2 ft x 4 ft)
801202 305 mm x 1525 mm (1 ft x 5 ft)
801203 915 mm x 915 mm (3 ft x 3 ft)
801206
801207
305 mm x 1,220 mm (1 ft x 4 ft) and 610 mm x 1,220 mm (2 ft x 4 ft)
305 mm x 1,525 mm (1 ft x 5 ft) and 915 mm x 915 mm (3 ft x 3 ft)
801209 406 mm x 406 mm (16 in. x 16 in.)
801210
801211
133 mm x 1220 mm (5.25 in. x 4 ft)
710 mm x 710 mm (28 in. x 28 in.)
801212 710 mm x 1270 mm (28 in. x 50 in.)
801215
801204 (BSC*)
801205 (BSC*)
* The BSC hood kits are used to certify Class II bio-safety cabinets by taking direct
in-ow measurements for NSF compliance.
305 mm x 915 mm (1 ft x 3 ft)
205 mm x 535 mm (8 in. x 21 in.)
255 mm x 535 mm (10 in. x 21 in.)
Duct Plugs
634650002 9.5 mm (3/8 in.) diameter - 1000 pieces
634650003 9.5 mm (3/8 in.) diameter - 5000 pieces
Printer
8934
Wireless Bluetooth printer
LogDat™ Mobile Software
LogDat Mobile*
Remote reader and data logger Android Software App available via Google Play
Capture Hood Stand
CH-Stand*
Extends up to 4.5 m (15 ft) with 8380 attached to take readings from ceiling diuser without the use of a ladder. Capture hood is secured onto quad bracket and two extension pole sections can be raised to desired height and locked in place. Hood stand uses wheels for ease of movement and portability.
Airow Probe 800187
Straight air ow probe, 46 cm (18 in.). Used to perform a duct traverse and to measure face velocity measurements. Ideal for small diameter ductwork.
Temperature and Humidity Probe 800220
Telescopic temperature and humidity probe, extends 230-990 mm (9-39 in.). Used for measuring inside of duct work. Can be inserted into a standard 8 mm (5/16 in.) diameter hole typically use for pitot traverses with the ability to calculate wet bulb and dewpoint temperatures.
Thermoanemometer Air Velocity Probes Models 960, 962, 964, and 966
Available in straight or articulating construction, and with or without a relative humidity sensor. Models with a relative humidity sensor can also calculate wet bulb and dewpoint temperature.
Velocity Matrix 801090
16 point Telescopic Velocity Matrix. Used for measuring face velocities of HEPA lters, chemical fume hood, laminar ow benches, lter banks, kitchen exhausts and other applications where a large surface area needs to be measured. Grid covers 0.09 m velocity while minimizing the eects of
2
(1 ft2) and averages the air
turbulence to produce a stable reading.
Pitot Probes
634634000
634634001*
634634002
634634003
634634005
8 mm–30 cm (5/16-12 in.) diameter
8 mm–46 cm (5/16-18 in.) diameter
8 mm–61 cm (5/16-24 in.) diameter
8 mm–91 cm (5/16-36 in.) diameter
8 mm – 152 cm (5/16-60 in.) diameter
*included in specic bundles. Please refer to model matrix on page 3.
Specications subject to change without notice.
TSI and the TSI logo are registered trademarks, and Airow, the Airow logo, LogDat and ProHood are trademarks of TSI Incorporated.
Android and Google Play are trademarks of Google Inc.
UNDERSTANDING, ACCELERATED
TSI Incorporated
USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 1 41 19 21 99
- Visit our website www.tsi.com for more information.
India Tel: +91 80 67877200 China Tel: +86 10 8219 7688 Singapore Tel: +65 6595 6388
Germany Tel: +49 241 523030
P/N 5 00143 2 Rev G (A4) ©2020 TSI Incorporated
Printed in U.S.A.
Loading...