Available SDI sensors ...................................... B - 48
Available BNC electrodes................................. B - 49
Further accessories (optional).......................... B - 51
Disposal .......................................................... B - 52
Declaration of conformity................................. B - 52
Symbols
Hazardous electric current!
Warns about hazards from electric current which can lead to injuries or even
death.
Hazard from hot surfaces!
Warns of a burn hazard from hot surfaces.
Explosion hazard!
Warns of hazards from explosive substances.
Danger!
Warns of a hazard which can lead to personal injury.
Caution!
Warns of a hazard which can lead to
damage to property.
Notes regarding the operating manual
Preface
The operating manual only describes the multifunction measuring meter and its intended use.
Detailed descriptions of the sensors and optional
accessories as well as tips for proper and practical
use of the multifunction measuring meter are not
included in this operating manual.
For that information, see the separate Practice Handbook!
The current version of the operating manual and the
Definitions
ConceptMeaning
SDI connection
BNC connection
NTC
MultiMeasure Studio
Standard,
MultiMeasure Studio
Professional
Serial Digital Interface;
digital serial interface for connecting sensors
interface for connecting passive electrodes after a resistance process
process for measuring temperature with
the electric resistance of a temperature
sensor
PC software for complete analysis and
visualisation of measured values
Practice Handbook can be found at: www.trotec.de
B-1Operating manual – Multifunction measuring meter T3000GB
Page 3
Legal notice
This release replaces all previous versions. No part of
this publication may be reproduced without written
permission from TROTEC
electronically processing, duplicating or spreading the
publication. Subject to technical changes. All rights
reserved. Trademarks are used without guarantee
that they may be used freely and primarily following
the spelling of the manufacturer. Product names are
registered.
Changes to construction in the interests of constant
improvements to the product, as well as changes to
the shape and colour are reserved.
The scope of delivery may vary from product images.
This document was created with all due care.
The warranty is for 12 months. Damages caused by
incorrect use by untrained people or startup by unauthorised people are excluded from the warranty.
The device complies with the fundamental health and
safety requirements of the applicable EU regulations
and was tested at the factory for perfect functionality
multiple times. However, if faults in the functionality
occur and cannot be remedied with the measures in
the chapter "Errors and faults", please get in touch
with your dealer or distributor. When making a warranty claim, supply the device number (see the rear of
the device). The invoice acts as warranty certificate.
When manufacturer's instructions or legal regulations
have not been followed, or after unauthorised changes
to the device are made, the manufacturer is not
responsible for the resulting damages. Changes to the
device or unauthorised replacement of individual
parts can drastically impact the electrical safety of this
product and leads to the forfeit of the warranty. Liability does not extend to damages to people or property
caused by the device being used other than as
described in the instructions in this operating manual.
Subject to changes to technical design and model
changes as part of constant development and product
improvement without prior notice.
No liability is accepted for damages resulting from improper use. In such cases, entitlements to a warranty
are then also forfeited.
GBOperating manual – Multifunction measuring meter T3000B-2
Page 4
Information about the device
Description of the device
The T3000 Multifunction measuring meter is used for
carrying out detailed analysis in the fields of material
and construction diagnostics. To do so, the device has
a 24 bit analogue/digital converter which delivers precise results and is permanently stable compared with
analogue measuring instruments.
The operating elements are found on the front and
sides of the robust housing. A scratch proof colour
display with touch function (1) is allows entering and
selecting values and functions and also displaying detected results. You can also navigate the device software for selecting measuring functions by using the
cross control (3) and the "OK" key (5).
Pressing the "Back" key (2) returns you to the previous screen of the device software. Pressing the "Main
menu" key (6) opens the main menu which is also
shown when you start the device.
Pressing the "Illumination on/off" key (4) either
switches the background illumination for the colour
display and the keys on or dims them.
The sensor connections (one SDI connection (7) and
one BNC connection (8)) are found above the colour
display (1). At the SDI connection (7), connect the appropriate digital SDI sensor for the corresponding situation. The universal interface of the SDI sensors
allows the multifunction measuring meter to automatically detect the sensor construction and/or indicates
when the wrong type of sensor is connected. Depending on the sensor construction, it is calibrated to the
prevailing surrounding conditions after connection.
Any sensor settings are saved directly to the corresponding SDI sensor. The various measured values,
such as relative humidity, dew point or flow rate, are
automatically calculated by the SDI sensors and
transmitted to the multifunction measuring meter.
At the BNC connection (8), you connect passive wood
or building moisture sensors.
The measured results of the sensor which is connected to the SDI or BNC connection are shown in the device's colour display (1).
The "On/Off" key (10) and a USB connection (11) are
found on the sides of the device.
You can connect the device to a computer by using the
supplied USB connection cable. Then you can extract
and analyse your measured results with the optional
MultiMeasure Studio Standard or MultiMeasure Studio
Professional software. Contact TROTEC
®
customer
service if you want this alternative version.
B-3Operating manual – Multifunction measuring meter T3000GB
Page 5
Device depiction
1
2
3
4
5
6
8
7
9
1011
No. Operating element
1scratch proof colour display with touch function
2"Back" key
3cross control with "Up", "Down", "Left", "Right" keys
4"Illumination on/off" key
5"OK" key
6"Main menu" key
7SDI connection (5-pin)
8BNC connection
9battery compartment with battery cover
10 "On/Off" key
11 USB connection
GBOperating manual – Multifunction measuring meter T3000B-4
Page 6
Technical data
ParametersValues
Multifunction measuring meter
ModelT3000
Dimensions (L x W x H)approx. 170 x 62 x 34 mm
Operating range
Operating temperature0 °C to +50 °C
Relative air humidity
<90% or <20g/m
Height above sea level (NN)max. 3000 m
Storage
Relative air humidity< 95 % (non condensing)
Ambient temperature-20 °C to +60 °C
Energy
Battery4 x Alkaline LR6 AA batteries, 1.5 V or NiMH 1.2 V to 1,5 V (rechargeable batteries)
Optional power supply5 V USB
Power input, activeapprox. 400 mW
Battery lifespan, passiveapprox. 1 year
Battery lifespan, activeat least 24 h
Sensor supply5.5 V ±10 % DC, max. 200 mA
Compatibility
Sensors
*
SDI sensors (digital)
BNC electrodes
Measured value storage
for approx. 200 measuring projects consisting of up to 3 x 3,600 (=10,800) meas-
ured values, that is a total of 200 x 10,800 = 2,160,000 measured values
Amount of tiles in the gridmax. 50 x 40 (max. amount of grid fields in one measurement)
Real time clock with calender programmed to 2099
3
(non-condensing)
Air temperature and humidity
Material temperature
Material moisture
Air current
trace gas detection
H
2
Material moisture (construction materials
and wood) after resistance process
Trace gas sensor TS 800 SDI (optional)
Response sensitivity
Measuring range
1 ppm H
2
0 to 1000 ppm H
2
Reaction time< 1 s
Pore size at stainless steel sinter filter > 50 μm
Power supply11–14 V
Battery ("BatteryPack")12 V / 2 Ah lead battery
Current consumption
Ambient temperature
activeapprox. 400 mA
passiveapprox. 20 mA (sensor inactive)
Operation0 °C to +50 °C
Storage-20 °C to +50 °C
Relative air humidity< 95 % (non condensing)
*
Data changes reserved.
B-5Operating manual – Multifunction measuring meter T3000GB
Page 7
Scope of delivery
Scope of delivery includes:
• 1 x T3000 Multifunction measuring meter
• 1 x USB connection cable
• 4 x Alkaline LR6 AA, 1.5 V batteries
• 1 x Display protection film
• 1 x Getting started guide
• 1 x Factory test certificate
• MultiMeasure Studio Standard PC software (to be
downloaded from www.trotec.de under the
"Downloads" menu)
Intended use encapsulates e.g.:
• the analysis of:
– supply and exhaust air flows
– fluctuations in relative humidity
– formation of condensation
– poor machine cooling
– porous seals
– heat build up and temperature fluctuations
– excessively dry or damp materials
• leak location with trace gas
Safety
Carefully read the operating manual before using
the device and keep it within reach!
• Do not use the device in atmospheres containing
oil, sulphur, chlorine or salt.
• Ensure that all connection cables are protected
from damages (e.g. from kinks or crushing).
• Protect the device from permanent direct sunlight.
• Observe the storage and operating conditions (see
chapter "Technical data").
Intended use
Only use the T3000 multifunction measuring meter for
preventive maintenance and risk analysis in the fields
of material, construction and climate diagnostics.
Here, read and observe the technical data.
To use the device for its intended use, only connect
and use accessories and spare parts which have been
approved by TROTEC
®
.
Improper use
Do not use the device in potentially explosive atmospheres, or for measurements in liquids. TROTEC® accepts no liability for damages resulting from improper
use. In such a case, entitlements to a warranty are forfeited. Any unauthorised modifications, alterations or
structural changes to the device are forbidden.
Personnel qualifications
People who use this device must:
• know and understand the dangers of working near
live parts.
• take measures to protect themselves from direct
contact with live parts.
• have read and understood the operating manual,
especially the "Safety" chapter.
For maintenance or repair work which requires
the housing to be opened, contact TROTEC® customer service. Devices which have been opened
unlawfully are void of any warranty and warranty
claims.
GBOperating manual – Multifunction measuring meter T3000B-6
Page 8
Residual risks
12
13
Hazardous electric current!
Work on the electrical components must
only be carried out by an authorised specialist company.
Hazardous electric current!
Never measure live parts.
Danger!
Do not leave the packaging lying around.
Children may use it as a dangerous toy.
Danger!
The device is not a toy and does not belong in the hands of children.
Danger!
Dangers can occur at the device when it
is used by untrained people in an unprofessional or improper way. Observe the
personnel qualifications.
Caution!
To prevent damages to the device, do not
expose it to extreme temperatures,
extreme humidity or moisture.
Caution!
To prevent damages to the device or to a
connected sensor due to overheating, observe the permissible operating and
measuring ranges of the device and the
connected sensor. The corresponding
specifications are provided in the product
catalogue for measuring devices or at
www.trotec.de under the "Product Catalog – Measuring instruments – Multifunction" menu.
Transport and storage
Transport
To safely transport the multifunction measuring meter
and accessories, use the optional MultiMeasure carry
case (12) recommended by TROTEC
protect the device during use and transport with an
optional device bag (13). Contact TROTEC
service if you want this alternative version.
Storage
When the device is not being used, observe the following storage conditions:
•dry,
• protected from dust and direct sunlight,
• with a plastic cover to protect it from invasive dust,
if necessary.
• The storage temperature is the same as the range
given in the chapter "Technical data".
• When storing the device for a long time, remove
the batteries.
• To store the device, use the optional carry case
(see chapter "Transport and storage") wherever
possible.
®
. Otherwise,
®
customer
B-7Operating manual – Multifunction measuring meter T3000GB
Page 9
Operation
20
15
6
2
191816 17
Inserting the batteries
• The text "SDI", "BNC" or "NTC" (16) indicates the
connected sensor type.
• Display of the current time (17)
• The "padlock" symbol (18) appears when you
press and hold the "On/Off" key for approx.
1 second during a measurement. The touch function of the colour display is locked. To release the
lock, press and hold the "On/Off" key for approx.
1 second.
• The "lightbulb" symbol (19) appears when you
have used the "Illumination on/off" key to switch
on background illumination (see chapter "Information about the device").
• Battery loading status indicator (20); a plug symbol
is shown when power is supplied via a USB connection.
Switching on
1. Press and hold the "On/Off" key for approx.
3 seconds until the device beeps.
2. Release the "On/Off" key.
– The colour display is switched on. The device is
ready to go as soon as the main menu is
shown.
Switching off
1. Press and hold the "On/Off" key for approx.
3 seconds until the device beeps.
2. Release the "On/Off" key.
– The colour display is switched off.
Description of screen elements
When using the device, take special note of the following important operating elements and displays:
• The "Back" key (2) opens the previous menu.
• The "Main menu" key (6) opens the main menu.
• Name of the current screen (15)
All options which can be selected via touch function
can also be selected by using the cross control and
the "OK" key.
For safety reasons, some options can only be selected
and configured by using the cross control and the
"OK" key (e.g. date and time in the "Settings" screen).
GBOperating manual – Multifunction measuring meter T3000B-8
Page 10
Set language
21
6
1. Press the "Settings" key in the main menu.
2. Press the language selection key (21) in the "Settings" screen.
3. Press the key with the desired language from the
following screen.
– The selected screen language is activated im-
mediately.
Main menu
You can open the following menus from the main
menu:
Measuring mode: Carrying out measurements.
Archive:Open archived measured values.
Settings:Make any device settings.
"Measuring mode" screen
The following measuring methods are available in the
"Measuring mode" screen:
4. Press the "Main menu" key (6) to open the main
menu.
B-9Operating manual – Multifunction measuring meter T3000GB
Wood moisture:Measure moisture in wood (with BNC elec-
Building moist:Measure moisture in walls and floors (with SDI
1. Touch the displayed type of wood (22 in image A.).
2. Press the "OK" key to activate the "Wood type" input box
(see image B.).
3. Press the "Up" or "Down" keys to select the initial letter of
the desired type of wood.
4. Press the "Right" key to select any other letters of the
desired type of wood.
5. Press the "OK" key.
6. Use the "Down" key to jump to the list of wood types and
select the desired type of wood. Press the "Left" or "Right"
keys to scan forward or backwards through the "Wood
type" list page by page.
7. Press the "OK" key to confirm the selected type of wood.
23M%:Display the measured value as a mass percent-
24Select the temperature sensor (INT) which is built into the
T3000 or select the manually specified temperature value (FIX;
see image C.). The measured or manually specified tempera-
ture value is shown to the right of the "INT" or "FIX" key.
25Resets the measured values shown by 27 to zero.
26Specifies how measured values by 27 are shown (see
image D.):
Minimum: Always shows the smallest detected measured
Maximum: Always shows the largest detected measured
Average:Shows the average value of all detected
Hold:Pauses the current detected measured value and
Off:Switches off the measured values by 27.
27Shows the measured values according to the specifica-
tions under 26:
Minute:Duration of the measured interval
M%:Shows the measured value (e.g. as average
°C:Shows the measured (INT) or manually specified
28Carries out a grid measurement (see image E., page B - 11).
29Specifies the measuring range in height x width (edge lengths)
and their activation/deactivation as well as the specified tem-
perature value (FIX) (see image F. , page B - 12).
30Carries out an automatic measurement for the duration of an
already specified recording interval. The recording interval can
be specified in the following screen (see image G. and
image H., page B - 13).
31Saves the measured values in the archive.
age (relative mass proportion of the humidity to
the total mass).
value from a measuring period.
value from a measuring period.
measured values from a measuring period.
shows it continually (when this option is
selected).
value) as a mass percentage (relative mass proportion of the humidity to the total mass).
temperature values (FIX).
GBOperating manual – Multifunction measuring meter T3000B-10
32digits:Dimensionless display of the measured value;
33Use the grid measurement function to carry out a grid
measurement (e.g. of a wall). The grid measurement can be
carried out either with or without height and length details, but
the grid is always used.
Proceed as follows:
1. Press the "OK" key to select the grid.
2. Use the cross control to select the first grid measuring spot
(e.g. "2-1" in image E.).
3. Press the "OK" key to confirm the selected grid measuring
spot. => The dimensionless measured value is recorded.
4. Use the keys of the cross control to select the next grid
measuring spot and repeat the process.
34Enters the current measured value to the marked grid measur-
ing spot (e.g. "2-1" in image E.). Then, the grid measuring
spot below it is selected (e.g. "2-2" in image E.).
Use this function if no measurement can be made at this position because of a power cabinet or a window etc.
35Skips the grid measuring spot below the currently selected
grid measuring spot (e.g. "2-1" in image E.) and selects the
grid measuring spot after it (e.g. "2-3" in image E.).
36Skips to the start of the next column.
37Saves the measured values in the archive.
the lower the measured electric resistance, the
higher the moisture in the wood and therefore,
the higher the digit value.
B-11Operating manual – Multifunction measuring meter T3000GB
43M%:Display the measured value as a mass percent-
44Shows the measured (INT) or the specified (FIX) temperature
45Opens a screen for selecting the recording duration for meas-
46Starts recording.
47Stops the current recording. The detected values are auto-
48Pauses the current recording.
age (relative mass proportion of the humidity to
the total mass).
value for temperature compensation.
uring over a long period (see image H.).
The key turns red once recording has started.
matically saved to the current measuring project.
Key 46 flashes.
Press this key again to continue recording.
B-13Operating manual – Multifunction measuring meter T3000GB
Page 15
"Measuring mode – Building moist" screen
Note!
The following applies for the screen descriptions below:
• When using BNC electrodes, the measured value is
shown as the dimensionless unit "digits".
• When using dielectric measuring SDI sensors, the
measured value is shown as "digits", "M%" or
"CM%".
GBOperating manual – Multifunction measuring meter T3000B-14
Page 16
57565554
49
53
A.
50
52
B.
C.
51
49 Selection for connected SDI sensor (see image B.):
Without material:
Anhydrite floor: A building moisture measurement is carried out for
Cement floor: A building moisture measurement is carried out for
Selection for connected BNC electrode (see image A.):
Without material:
50 For selection "Without material" (with BNC electrode):
digits:Dimensionless display of the measured value; the
For selection "Anhydrite floor" or "Cement floor" (with SDI
sensor TS 660 SDI):
M%:Display the measured value as a mass percentage
CM%:Shows the measured value according to the calcium
51 Resets the measured values shown by 53 to zero.
52 Specifies how measured values by 53 are shown (see
image C.):
Minimum:Always shows the smallest detected measured
Maximum:Always shows the largest detected measured value
Average:Shows the average value of all detected measured
Hold:Pauses the current detected measured value and
Off:Switches 53 display off.
53 Shows the measured values according to the specifications
under 52:
Minute:Duration of the measured interval
M%:Shows the measured value (e.g. as average value)
CM%:Shows the measured value (e.g. as average value)
54 Carries out a grid measurement (see image D., page B - 16).
55 Specifies the measuring range in height x width as well as the limit
value for the alarm signal (see image E. and image F., page B - 17).
56 Carries out an automatic measurement for the duration of an
already specified recording interval. The recording interval can be
specified in the following screen (see image G. and image H.,
page B - 18).
57 Saves the measured values in the archive.
A non-destructive, indicative (digits) measurement
is carried out.
anhydrite floor (with SDI sensor TS 660 SDI). The
measurement results (indicative) in M% and CM%
are detected with an equation which is suitable for
this material.
cement floor (with SDI sensor TS 660 SDI). The
measurement results (indicative) in M% and CM%
are detected with an equation which is suitable for
this material.
A resistance measurement is carried out.
lower the measured electric resistance, the higher
the moisture in the material and therefore, the
higher the digit value.
(relative mass of the humidity to the total mass).
carbide method.
value from a measuring period.
from a measuring period.
values from a measuring period.
shows it continually (when this option is selected).
as a mass percentage (relative mass proportion of
the humidity to the total mass).
according to the calcium carbide method.
B-15Operating manual – Multifunction measuring meter T3000GB
Page 17
"Measuring mode – Building moisture – Grid measurement" screen
57
D.
58
626160
59
57digits:Dimensionless display of the measured value;
58Use the grid measurement function to carry out a grid meas-
urement (e.g. of a wall). The grid measurement can be carried
out either with or without height and length details, but the
grid is always used.
Proceed as follows:
1. Press the "OK" key to select the grid.
2. Use the cross control to select the first grid measuring spot
(e.g. "2-1" in image D.).
3. Press the "OK" key to confirm the selected grid measuring
spot. => The dimensionless measured value is recorded.
4. Use the keys of the cross control to select the next grid
measuring spot and repeat the process.
59Enters the current measured value to the marked grid measur-
ing spot (e.g. "2-1" in image D.). Then, the grid measuring
spot below it is selected (e.g. "2-2" in image D.).
Use this function if no measurement can be made at this position because of a power cabinet or a window etc.
60Skips the grid measuring spot below the currently selected
grid measuring spot (e.g. "2-1" in image D.) and selects the
grid measuring spot after it (e.g. "2-3" in image D.).
61Skips to the start of the next column.
62Saves the measured values in the archive.
the lower the measured electric resistance, the
higher the moisture in the material and therefore, the higher the digit value.
GBOperating manual – Multifunction measuring meter T3000B-16
Page 18
"Measuring mode – Building moisture – Measuring range/limit value alarm signal" screen
67
63
E.
64
65
66
F.
63Specifies the total height of the grid area to be measured. The
value is used during grid measurement (see image D.).
64Specifies the total width of the grid area to be measured. The
value is used during grid measurement (see image D.).
65Activates ("On") or deactivates ("Off") the use of specified
edge lengths (63, 64) during grid measurement. For deactiva-
tion, you must specify the corresponding edge lengths of the
grid area in the PC software after grid measurement.
66Specifies the limit value for the alarm signal of the dimension-
less measuring unit "digits".
Once the limit value is reached, an alarm signal is emitted.
Note!
This option is only enabled when an SDI sensor with alarm
function is connected (see image F.). This option is greyed out
when connecting a BNC electrode (see image E.).
67Saves the settings.
Note!
The settings saved here are also used for all following
measurements, unless they are deactivated (in the
corresponding menu item)!
B-17Operating manual – Multifunction measuring meter T3000GB
Page 19
"Measuring mode – Building moisture – Automatic measurement" screen
G.
68
69
737271
70
H.
68For selection "Without material" (see image A.,
page B - 14):
digits:Dimensionless display of the measured value;
For selection "Anhydrite floor" or "Cement floor" (see
image A., page B - 14):
M%:Display the measured value as a mass percent-
CM%:Shows the measured value according to the cal-
69Shows the remaining time until the automatic measurement
finishes.
70Opens a screen for selecting the recording duration for meas-
uring over a long period (see image H.).
71Starts recording.
The key turns red once recording has started.
72Stops the current recording. The detected values are automat-
ically saved to the current measuring project.
73Pauses the current recording.
Key 71 flashes.
Press this key again to continue recording.
the lower the measured electric resistance, the
higher the moisture in the material and therefore, the higher the digit value.
age (relative mass proportion of the humidity to
the total mass).
cium carbide method.
GBOperating manual – Multifunction measuring meter T3000B-18
Page 20
"Measuring mode – Flowrate" screen
828180
74
79
A.
75
76
B.
77
78
74Shows the measured flow rate as a numerical value in the
selected unit (e.g. m/s).
This unit is selected in the "Settings" screen, page B - 30.
75Shows the measured flow rate as visual bars.
76Shows the measured temperature value.
77Resets the measured values shown by 79 to zero.
78Specifies how measured values by 79 are shown (see
image B.):
Minimum: Always shows the smallest detected measured
Maximum: Always shows the largest detected measured
Average:Shows the average value of all measured values
Hold:Pauses the current detected measured value and
Off:Switches 79 display off.
79Shows the measured values according to the specifica-
tions under 78:
Minutes:Duration of the measured interval
m/s:Shows the measured flow rate as a numerical
°C:Shows the measured temperature value.
80Carries out an automatic measurement for the duration of an
already specified recording interval. The recording interval can
be specified in the following screen (see image C. and
image D., page B - 20).
81Specifies the measuring range in height x width as well as the
form of the object to be measured (e.g. pipe or shaft) for the
volumetric flow measurement (see image E., page B - 21).
82Saves the measured values in the archive.
value from a measuring period.
value from a measuring period.
which have been detected so far from a measuring period.
shows it continually (when this option is
selected).
value in the selected unit (e.g. m/s).
This unit is selected in the "Settings" screen,
page B - 30.
B-19Operating manual – Multifunction measuring meter T3000GB
83Shows the measured flow rate as a numerical value in the
selected unit (e.g. m/s).
This unit is selected in the "Settings" screen, page B - 30.
84Shows the measured flow rate as visual bars.
85Shows the measured temperature value.
86Shows the remaining time until the automatic measurement
finishes.
87Opens a screen for selecting the recording duration (see
image D.).
88Starts recording.
The key turns red once recording has started.
89Stops the current recording. The detected values are automat-
ically saved to the current measuring project.
90Pauses the current recording.
Key 88 flashes.
Press this key again to continue recording.
GBOperating manual – Multifunction measuring meter T3000B-20
91Selects the form of the object to be measured. The following
options are available (see image F.):
1. Square
2. Round
3. Off
Depending on the selected form, a different equation is used
to calculate the measured values.
92Specifies the diameter of the object to be measured.
93Specifies the height of the object to be measured.
94Specifies the width of the object to be measured.
95Saves the settings.
Note!
The settings saved here are also used for all following
measurements, unless they are deactivated (in the
corresponding menu item)!
B-21Operating manual – Multifunction measuring meter T3000GB
Page 23
"Measuring mode – Humidity" screen
B.
100
96
98
A.
97
99
101
96Shows the following measured values:
Temperature:Temperature value in °C
Humidity:humidity in %RH (relative humidity)
Dew point: Dew point in °C
97Specifies how measured values by 99 are shown (see
image B.):
Minimum: Always shows the smallest detected measured
Maximum: Always shows the largest detected measured
Average:Shows the average value of all measured values
Hold:Pauses the current detected measured value and
Off:Switches 99 display off.
98Resets the measured values shown by 99 to zero.
99Shows the measured values according to the specifica-
tions under 97:
Temperature:Temperature value in °C
Humidity:humidity in %RH (relative humidity)
Dew point: Dew point in °C
100 Carries out an automatic measurement for the duration of an
already specified recording interval. The recording interval can
be specified in the following screen (see image C. and
image D., page B - 23).
101 Saves the measured values in the archive.
value from a measuring period.
value from a measuring period.
which have been detected so far from a measuring period.
shows it continually (when this option is
selected).
GBOperating manual – Multifunction measuring meter T3000B-22
Dew point: Dew point in °C
103 Shows the remaining time until the automatic measurement
finishes.
104 Opens a screen for selecting the recording duration for
measuring over a long period (see image D.).
105 Starts recording.
The key turns red once recording has started.
106 Stops the current recording. The detected values are auto-
matically saved to the current measuring project.
107 Pauses the current recording.
Key 105 flashes.
Press this key again to continue recording.
B-23Operating manual – Multifunction measuring meter T3000GB
Page 25
"Measuring mode – Temperature" screen
B.
112
108
110
A.
109
113
111
108 Shows the measured temperature in °C.
109 Specifies how measured values by 111 are shown (see
image B.):
Minimum: Always shows the smallest detected measured
Maximum: Always shows the largest detected measured
Average:Shows the average value of all measured values
Hold:Pauses the current detected measured value and
Off:Switches 111 display off.
110 Resets the measured values shown by 111 to zero.
111 Shows the measured value according to the specifications
under 109:
Minutes:Shows the measurement duration in minutes.
Temperature:Shows the measured temperature in °C.
112 Carries out an automatic measurement for the duration of an
already specified recording interval. The recording interval can
be specified in the following screen (see image C. and
image D., page B - 25).
113 Saves the measured values in the archive.
value from a measuring period.
value from a measuring period.
which have been detected so far from a measuring period.
shows it continually (when this option is
selected).
GBOperating manual – Multifunction measuring meter T3000B-24
Page 26
"Measuring mode – Temperature – Automatic measurement" screen
C.
114
115
119118117
116
D.
114 Shows the measured temperature in °C.
115 Shows the remaining time until the automatic measurement
116 Opens a screen for selecting the recording duration for
117 Starts recording.
118 Stops the current recording. The detected values are auto-
119 Pauses the current recording.
finishes.
measuring over a long period (see image D.).
The key turns red once recording has started.
matically saved to the current measuring project.
Key 117 flashes.
Press this key again to continue recording.
B-25Operating manual – Multifunction measuring meter T3000GB
Page 27
"Measuring mode – Hydrogen" screen
120
121
122
120 Shows the measured hydrogen concentration in digits (dimen-
sionless measured value). This measured value is shown rela-
tive to the set base value (see chapter "Preparing for
measurement with trace gas sensor (heating phase)"). Nega-
tive measured values cannot be shown.
121 Shows the measured hydrogen concentration in digits (dimen-
sionless measured value). This measured value is shown rela-
tive to the set reference value (see chapter "Differentiated
indication measurement"). Negative measured values can also
be shown.
122 Exits the screen.
Note!
When using the differentiated indication measurement, new
basic concentrations always occur which the displayed
H
2
measured values are based on. An evaluation and comparison
of the measured dimensionless measured values is then not
worthwhile, because the measurements may relate to con-
stantly changing H
ues are not saved.
So, only use the Measuring mode – Hydrogen for locating
leaks.
concentrations. Thus, the measured val-
2
GBOperating manual – Multifunction measuring meter T3000B-26
Page 28
"Archive" screen
125
123
124
A.
In the "Archive" screen, you can view archived measuring projects or open them for further processing:
Archive, image A.:
123 Shows the measuring projects which are saved in the archive.
The currently selected archive entry is highlighted.
Navigate in the archive as follows:
1. Press the "Down" key on the cross control until the desired
archive entry is selected.
2. You may need to turn the page by pressing the "Left" or
"Right" keys on the cross control.
3. Press the "OK" key on the cross control to confirm the
selected archive entry. You can also press "OK" (124) on
the screen key. => Image B. opens.
124 Selects the currently selected archive entry.
=> Image B. opens.
125 Deletes all entries from the archive. => A safety prompt
opens. Confirm it by pressing the "OK" key if you want to
delete all entries. Otherwise, touch the "Cancel" key.
The icons beside the saved measuring projects show
each measuring mode (see image A.). The meaning is
as follows:
Building moisture measurement
Wood moisture measurement
Flowrate measurement
Temperature measurement
Relative humidity measurement
Grid measurement
Spot measurement
Time measurement
B-27Operating manual – Multifunction measuring meter T3000GB
Page 29
Archive, image B.:
B.
126
127
128
C.
129
130
Archive, image C.:
126 Opens the selected archive entry (see image C., page B - 28).
127 Deletes the selected archive entry. => A safety prompt opens.
Confirm it by pressing the "OK" key if you want to delete the
entry. Otherwise, touch the "Cancel" key.
128 Closes image B. and returns to image A. ("Archive" screen).
129 Closes the selected archive entry and returns to image A.
("Archive" screen, page B - 27).
130 Opens the editing mode for the selected grid measurement.
This editing mode is only available for grid fields. Single
measurements or measurements over a long period cannot be
edited in the device.
GBOperating manual – Multifunction measuring meter T3000B-28
Page 30
"Settings" screen
131
132
133
134
135
136
137
138
139
140
You can configure the device as follows in the "Settings" screen:
Settings 1/3:
Settings 2/3:
131 Selects the screen language (see chapter "Set language").
132 Selects the unit system:
Metric:Activates the metric unit system (e.g. for use in
Imperial:Activates the imperial unit system (e.g. for use in
133 Sets the screen brightness.
This option can only be selected by using the cross control
below the colour display.
1. Press the "Down" key on the cross control until the scale is
2. Press the "Left" or "Right" keys on the cross control to
134 Sets the date and time.
This option can only be selected by using the cross control
below the colour display.
1. Press the "Down" key on the cross control until the date is
2. Press the "OK" key on the cross control.
3. Press the "Left" or "Right" keys on the cross control to
4. Press the "Up" or "Down" keys on the cross control to
5. Press the "OK" key on the cross control.
135 Opens the next screen.
136 Either specifies the period for automatic dimming of the colour
display or deactivates the function:
30 sec, 1 min, 5 min, off
137 Either specifies the period for automatic switch-off of the
colour display or deactivates the function:
10 min, 30 min, 1 hour, off
138 Switches signal tones/key tones on or off.
139 Selects the base value to be shown in measuring mode:
Dew point: Displays the value as dew point.
Absolute
humidity:
Mixing ratio: Displays the value as mixing ratio (g/kg) of the
140 Opens the next screen.
continental Europe).
the USA).
selected.
reduce (left) or increase (right) the screen brightness.
selected.
=> The entire row is selected.
select the value to be configured.
increase or reduce the value to be configured.
=> The row is deselected.
Displays the value as absolute humidity in g/m
mass of steam to the mass of the dry gas.
3
.
B-29Operating manual – Multifunction measuring meter T3000GB
Page 31
Settings 3/3:
141
141 Specifies the unit for volumetric flow:
3
/s, m3/min, m3/h, l/min
m
GBOperating manual – Multifunction measuring meter T3000B-30
Page 32
Carrying out a measurement
A.
B.
C.
D.
E.
Note:
Note that moving from a cold area to a warm area can
lead to condensation forming on the device's circuit
board. This physical and unavoidable effect can falsify
the measurement. In this case, the colour display
shows either no measured values or they are incorrect. Wait a few minutes until the device has become
adjusted to the changed conditions before carrying
out a measurement.
After connecting a sensor and selecting the measuring mode, the measurement starts automatically.
Carrying out a measurement with SDI sensor (e.g.
with TS 300 SDI humidity sensor)
B-31Operating manual – Multifunction measuring meter T3000GB
Page 33
Carrying out a measurement with BNC electrode
F.
90°
A.
B.
C.
D.
(e.g. with TS 60 hand electrode)
GBOperating manual – Multifunction measuring meter T3000B-32
Page 34
E.
F.
G.
H.
Shut down procedure
1. Switch off the device by pressing the "On/Off" key
(see chapter "Switching off").
2. Detach connecting cables and sensors.
3. Clean the device according to the chapter
"Maintenance".
4. Store the device according to the chapter
"Storage".
B-33Operating manual – Multifunction measuring meter T3000GB
Page 35
PC software
Use the MultiMeasure Studio Standard PC software
(free standard version) or MultiMeasure Studio
Professional (paid professional version) to carry out a
detailed analysis and visualisation of your measured
results. You can only use all configuration,
visualisation and functional options of the
multifunction measuring meter when using this PC
software and a TROTEC
You can open a basic display of your measured values
at any time in the device (see chapter "Description of
screen elements").
®
USB dongle (professional).
Installation conditions
Ensure that the following minimum requirements for
installing the MultiMeasure Studio Standard or MultiMeasure Studio Professional PC software are fulfilled:
• Supported operating systems (32 or 64 bit version):
– Windows XP from service pack 3
–Windows Vista
–Windows 7
–Windows 8
• Software requirements:
– Microsoft Excel
– Microsoft .NET Framework 3.5 SP1 (is otherwise
automatically installed during the software installation)
• Hardware requirements:
– Processor speed: 1.0 GHz, minimum
– USB connection
– Internet connection
– 512 MB RAM, minimum
– 1 GB hard disk space, minimum
– TROTEC
®
USB dongle (Professional) for using
the professional version of the PC software
GBOperating manual – Multifunction measuring meter T3000B-34
Page 36
Installing the PC software
1. Download the current PC software from the Internet. To do so, visit the website www.trotec.de.
Click "Downloads" and then "Software Download". In the "Software" search box, select "MultiMeasure Studio Standard" and then follow the
other instructions. If you want to use the optionally available professional version of the PC software "MultiMeasure Studio Professional", then
get in touch with your TROTEC
vice.
2. Unpack the downloaded file and double click the
"Setup.exe" file.
3. Follow the instructions of the installation wizard.
®
customer ser-
Starting the PC software
1. Connect the multifunction measuring meter to
your PC via the USB connection cable provided in
the scope of delivery.
®
2. If necessary, connect the TROTEC
to a free USB port on your PC.
– You need the TROTEC® USB dongle (Profes-
sional) to use the professional version of the PC
software.
– The TROTEC® USB dongle (Professional) is
automatically detected by the operating system.
– If you only connect the TROTEC® USB dongle
(Professional) to your PC after starting the PC
software, click the "Parameters" menu item in
the PC software. Afterwards, click the USB
symbol ("Dongle check") to read the connected
®
TROTEC
USB dongle (Professional).
3. Switch on the multifunction measuring meter (see
chapter "Switching on") if necessary.
4. Start the MultiMeasure Studio Standard or MultiMeasure Studio Professional software.
USB dongle
Information about using the PC software is provided in
the online help.
B-35Operating manual – Multifunction measuring meter T3000GB
Page 37
Errors and faults
The accurate functionality of the device was tested
during production a number of times. However, if
functionality faults do occur, then check the device
according to the following list.
The device does not switch on:
• Check the loading status of the batteries. Replace
the batteries when the battery symbol in the colour
display only shows one bar. If the battery symbol
is red, then the battery voltage is insufficient.
• Check that the batteries are properly positioned.
Check the polarity is correct.
• Never carry out an electrical check yourself;
instead, contact your TROTEC
The device runs but no measured values are
shown:
• Check whether the multifunction measuring meter
is in the correct sensor mode.
• Check the connection cable to the SDI or BNC
sockets is connected properly.
• Check the used connection cable and its connections as well as the connections to the multifunction measuring meter for damages (e.g. broken
cable, damaged pins etc.). Use a different connection cable of the same type to rule out possible
faults.
• Ensure that the appropriate sensor for the measurement is being used. Here, also observe the
Practice Handbook or the product catalogue for
measuring devices.
• Ensure that the colour display is switched on. Possibly use the "Illumination on/off" key (see chapter
"Device depiction").
®
customer service.
• Check the room temperature and the relative
humidity. Check the device's permissible
operating range complies with the technical data.
• Check whether the multifunction measuring meter
responds to touching the colour display. If it shows
no reaction despite an enabled colour display and
sufficient battery power, then the device firmware
may have crashed. Restart the multifunction
measuring meter. To do so, remove a battery and
reinsert it.
Does your device still not operate correctly after
these checks?
®
Contact TROTEC
customer service.
GBOperating manual – Multifunction measuring meter T3000B-36
Page 38
Maintenance
Maintenance and care intervals
Maintenance and care intervalbefore every start
check SDI, BNC, USB connections for dirt and foreign
objects and clean if necessary
clean housingXX
visually check whether the device is dirtyXX
replace batteriesX
check for damagesX
carry out a test runX
X
when
necessary
at least every
4 weeks
at least annually
B-37Operating manual – Multifunction measuring meter T3000GB
Page 39
Activities for before the start of maintenance
1. Switch off the device (see chapter "Switching
off").
2. Detach connecting cables and sensors.
For maintenance or repair work which
requires the housing to be opened,
contact TROTEC
®
customer service.
Devices which have been opened unlawfully are void of any warranty and
warranty claims.
Visual inspection of the device
1. Check the device for dirt and damages.
2. Check the SDI, the BNC and the USB connection
for dirt and damages.
3. Check the colour display for dirt and damages.
4. Check that the batteries and battery cover sit
properly.
Damaged connections can falsify measurements and
measurement results. A damaged colour display can
influence how measured results are shown. In this
case, contact your TROTEC® customer service or replace the device.
Cleaning the device
1. Use a soft, lint-free cloth for cleaning.
2. Dampen the cloth with clean water. Do not use
sprays, solvents, alcohol-based or abrasive
cleaners to dampen the cloth.
3. Clear dirt from the housing, the connections and
the colour display.
GBOperating manual – Multifunction measuring meter T3000B-38
Page 40
Using the trace gas sensor TS 800 SDI
(optional)
Hazard from hot surfaces!
The silver sensor protection cap (sinter
filter) of the trace gas sensor heats up
during operation to a temperature approximately 40 °C above ambient temperature.
Do not touch the sensor protection cap.
Explosion hazard!
Do not use pure hydrogen as a trace gas.
There is a risk of explosion!
Only use the gas mixture recommended
in chapter "Trace gas and calibration
gas".
Caution!
To prevent damages to the trace gas sensor, never hold it in stagnant water, other
liquids, mud or fine powdered or powdery
substances.
Trace gas and calibration gas
Use a mixture of 95 % nitrogen and 5 % hydrogen as
trace gas. This gas mixture is not flammable, not explosive, not poisonous and not corrosive.
Do not use this trace gas to calibrate the trace gas
sensor. The calibration gas must contain oxygen.
®
TROTEC
thetic air as calibration gas.
recommends 50 ppm of hydrogen in syn-
Functioning principle
When using the trace gas sensor TS 800 SDI, it is possible to detect and locate leakages in systems purged
with trace gas without causing any damage. To do so,
an indicative measurement of the hydrogen concentration is carried out. The measurement of easily volatile and flammable gases is also possible to an extent.
Hydrogen is known for its small molecule size and
high rate of diffusion. After being filled into the system
to be checked, this allows the hydrogen contained in
the trace gas to exit through small openings (e.g. tears
in the sleeves of pressure tanks, pipes, tanks) and
back into the surrounding air. Hydrogen can also diffuse through surrounding and covering materials (e.g.
flooring, plaster, mineral joining material). Thus, covered leakages in the home technology can be limited
by detection of the hydrogen or even precisely located. The surrounding material does not need to be
damaged to do so.
Due to its high resolution accuracy, the trace gas sensor TS 800 SDI already detects low concentrations of
hydrogen from 1 ppm. Here, the detection is carried
out at a heated sensor, which permanently detects the
hydrogen concentration in the surroundings within the
measuring range from 0 ppm to 1000 ppm.
During measurement, rising and falling hydrogen concentrations are signalled acoustically at the carry handle or the trace gas sensor TS 800 SDI as well as
displayed numerically on the colour display of the
T3000 multifunction measuring meter.
The trace gas sensor TS 800 SDI is not designed for
absolutely determining the hydrogen content in surrounding air!
B-39Operating manual – Multifunction measuring meter T3000GB
Page 41
Scope of delivery
177
13
142
T3000
148
The scope of delivery of the TS 800 SDI trace gas sensor set contains:
• 1 x Trace gas sensor TS 800 SDI (177)
• 1 x Battery pack (142)
• 2 x TC 30 SDI connecting cable (148)
• 1 x Device bag (13)
• 1 x Charger (not shown)
For a battery voltage between 11.5 V and 11 V, the
lead battery is run in the reserve range. Two short
note signals are sounded. The trace gas sensor can
still remain in operation.
If the lead battery is discharged, a three-beep signal
sounds every 10 seconds. The colour display of the
T3000 multifunction measuring meter displays the
flashing value "-1000.0" (battery voltage < 11 V).
If the lead battery capacity is lower, then the operation
is automatically interrupted to prevent damages to the
lead battery.
Switch off the trace gas sensor (see chapter "Switching the trace gas sensor on and off") and use the
charger provided to load the battery pack.
If the lead battery is not completely discharged and
another minimum voltage is available, you can also
run the trace gas sensor directly via the charger (connected to the battery pack). In this case, the lead battery is loaded simultaneously.
However, first ensure that the lead battery is definitely not completely discharged!
Caution!
Information about the battery pack
The battery pack contains a special lead battery which
allows continuous operation for approx. 4 hours.
Always store and load the lead battery at
temperatures of 15 °C to 35 °C. Otherwise, there is a risk of decreased capacity
and long-term damages to the lead battery!
Always use the charger provided to load the battery
pack completely and without interruption. Then use
the battery pack until it is nearly completely empty.
A 'memory effect' does not occur if the lead battery is
only occasionally not completely discharged and loaded. But try to avoid interim charging wherever possible, as this can influence the lead battery lifetime.
GBOperating manual – Multifunction measuring meter T3000B-40
Page 42
Connecting the trace gas sensor to the T3000
A.
B.
C.
D.
E.
F.
G.
multifunction measuring meter
B-41Operating manual – Multifunction measuring meter T3000GB
Page 43
Switching the trace gas sensor on and off
H.
143144145146 147
The trace gas sensor is also switched on and off when
the multifunction measuring meter is switched on and
off.
Note:
Ensure that the trace gas sensor is correctly connected to the T3000 multifunction measuring meter. The
trace gas sensor remains connected if it is only connected to the battery pack.
Operating elements of the trace gas sensor
Use the operating elements at the battery pack for
operating the trace gas sensor.
No. Operating element
Connection cable socket for the T3000 multifunction
143
measuring meter
144 Connection cable socket for the trace gas sensor
"Sensitivity" selector switch:
Position 1:
Maximum interval for measured value difference 500
(Device only reacts to larger amounts of hydrogen.)
Position 2:
Maximum interval for measured value difference 200
Position 3:
145
Maximum interval for measured value difference 100
Position 4:
Maximum interval for measured value difference 50
Position 5:
Maximum interval for measured value difference 10
(Device even reacts to small amounts of hydrogen.)
Position 6: Beep deactivated
"Reset" key:
146
Defines the current measured value as the new reference
value.
147 Connection for charger
GBOperating manual – Multifunction measuring meter T3000B-42
Page 44
Preparing for measurement with trace gas
sensor (heating phase)
1. Ensure that the trace gas sensor and the T3000
multifunction measuring meter are properly connected to each other.
2. Ensure that the T3000 multifunction measuring
meter is switched on.
Hazard from hot surfaces!
The silver sensor protection cap (sinter
filter) of the trace gas sensor heats up
during operation to a temperature
approximately 40 °C above ambient temperature.
Do not touch the sensor protection cap.
After being switched on, the trace gas sensor is heated for approx. 1 minute at approx. 600 °C – regardless on the position of the "Sensitivity" selector
switch.
During the heating phase, the trace gas sensor carries
out a self-calibration to detect the base value for
hydrogen concentration in the surroundings.
3. During the heating phase, ensure that the
trace gas sensor is not near a relevant source
of hydrogen. Otherwise, an increased hydrogen concentration can influence the self-calibration.
If you have disabled the trace gas sensor beep ("Sensitivity" selector switch at position 6), then the end of
the heating phase is indicated with a short beep. The
trace gas sensor is ready for operation.
If you have enabled the trace gas sensor beep ("Sensitivity" selector switch at position 1–5), then this
beep is not emitted at the end of the heating phase.
The readiness to operate is automatically indicated at
least once per second with a beep.
During heating, the "Hydrogen 2" value on the colour
display or the T3000 multifunction measuring meter
drops from 50 to 0. Upon reaching zero, the trace gas
sensor is ready for operation.
B-43Operating manual – Multifunction measuring meter T3000GB
Page 45
Measuring with the trace gas sensor
1. Start by setting the "Sensitivity" selector switch
to position 5.
– The leakage rate of the exiting hydrogen is not
yet known at the start of a measurement. If the
"Sensitivity" selector switch is at position 5,
then even minimal amounts of hydrogen are
detected.
2. Inspect the surrounding area for raised hydrogen
concentration.
– The colour display of the T3000 multifunction
measuring meter shows an increasing
measured value when you approach an area
with raised hydrogen concentration. In
addition, the trace gas sensor emits a beep at
higher frequency and shorter intervals when
the measured values increase.
– The colour display of the T3000 multifunction
measuring meter shows a decreasing
measured value when you leave an area with
increased hydrogen concentration. The beep of
the trace gas sensor is emitted at longer
intervals and lower frequency.
3. Adjust the position of the "Sensitivity" selector
switch during the measurement according to conditions on site.
Note:
It is not possible to display measured values below the
base values (negative measured values) in standard
operation.
To inspect surrounding areas in which the hydrogen
concentration is below the base value, use the differentiated indication measurement by pressing the "Reset" key (see chapter "Differentiated indication
measurement").
Differentiated indication measurement
1. Press the "Reset" key on the battery pack to
specify the current measured value as the new
reference value.
– A short beep confirms the selection of the new
reference value.
– The T3000 multifunction measuring meter's
colour display shows two different measured
values in the "Measuring mode – Hydrogen"
screen at the same time.
For the differentiated indication measurement, the
trace gas sensor also reacts to minimal changes to
hydrogen concentration when the nearby hydrogen
concentration is raised.
If you reach areas with higher or lower hydrogen
concentration, the shown measured value in the lower
measured value display in the "Measuring mode –
Hydrogen" screen changes in relation to the reference
value. For increasing hydrogen concentration, it rises.
For decreasing hydrogen concentration, even negative measured values are shown in relation to the reference value.
The base value for the beep intervals is no longer the
benchmark, but instead the selected reference value
is.
From a combination of the basis value and reference
value, as well as a systematic procedure, you can also
locate hydrogen leaks in poor conditions.
GBOperating manual – Multifunction measuring meter T3000B-44
Page 46
Schematic example of measured value
Reactivation of the dif-
ferentiated indication
measurement
Schematic example of measured value displays
Standard operation:
Differentiated indication measurement:
Activation of the dif-
ferentiated indication
measurement
Environment 1
Base hydrogen concen-
tration is 200 digits
Startup and base value
calibration
Environment 2
Base hydrogen concen-
tration is 300 digits
Environment 3
Base hydrogen concen-
tration is 100 digits
Hydrogen1 digits
0.0
Hydrogen2 digits
0.0
Hydrogen1 digits
100.0
Hydrogen2 digits
100.0
Hydrogen1 digits
0.0
Hydrogen2 digits
0.0
Hydrogen1 digits
0.0
Hydrogen2 digits
-100.0
Hydrogen1 digits
0.0
Hydrogen2 digits
0.0
Hydrogen1 digits
100.0
Hydrogen2 digits
200.0
Hydrogen1 digits
0.0
Hydrogen2 digits
0.0
displays
In the following example, three bordering, fictional
measuring environments are assumed:
• In environment 1, a base hydrogen concentration
of 200 digits applies.
• In environment 2, a base hydrogen concentration
of 300 digits applies.
• In environment 3, a base hydrogen concentration
of 100 digits applies.
The base hydrogen concentrations are not known to
the example user, but the reader does know them for
understandability.
The trace gas sensor is commissioned in
environment 1. After the start up phase and automatic
calibration, the base value 0 for the prevailing base
hydrogen concentration (200 digits) is specified.
The measurements start in standard operation.
In environment 1, the relative measured value 0 is
shown in the upper and lower measured value display.
The user proceeds to environment 2. Both measured
value displays show a relative measured value of 100,
relating to the base value.
Finally, the user proceeds to environment 3. Here,
both displays show 0 again, but the base hydrogen
concentration is 100 digits lower than the concentration below the base value.
This is due to the fact that no measured values below
the base value (negative measured values) can be
shown in standard operation.
B-45Operating manual – Multifunction measuring meter T3000GB
The user then returns to environment 1 to use the
differentiated indication measurement by pressing the "Reset" key (see chapter "Differentiated
indication measurement").
In environment 1, the user activates the differentiated
indication measurement and defines an additional reference value for the current environmental concentration.
The upper display now shows the measured value relating to the base value. The lower display shows the
measured value relating to the reference value.
Because both the base value and the reference value
were specified in this environment, both measured
value displays show the value 0 in this environment.
Now the user returns to environment 3. The upper display shows the measured value 0 here relating to the
base value; while the lower displays shows the measured value -100 relating to the reference value.
Page 47
While still in environment 3, the user carries out a differentiated indication measurement. The measured
value display of the reference value now changes
from -100 to 0 because this environment was specified as the new reference value. Afterwards, the user
proceeds to environment 2. Here, the upper display
shows the measured value 100 as at the start because the base value was not changed. Due to the renewed differentiated indication measurement in
environment 3 and thus a renewed reference value
specification, the lower measured value display now
shows a measured value of 200 in environment 2.
Note:
The correlation of measured value display and H
concentration is not linear, but logarithmic. The
displayed digit value does not automatically equal the
ppm value!
Practical example
The measuring position is located in the basement of
a house. Earlier pressure testing determined that a
pressure water line was to be checked. This pressure
water line proceeds through three basement rooms
without windows.
After purging the line with air from a compressor and
then feeding the trace gas, the gas is emitted from the
leakage point. This point should now be found. Because no damp areas on the floor are visible, every
room must be individually checked.
Because the room cannot be aired, the leakage must
be located with the prevailing, high H
concentration.
2
To do so, a reference value is enabled with the differentiated indication measurement (see chapter "Differentiated indication measurement").
Thus, the user carries out an active compensation relating to the H
surrounding concentration. This is now
2
adopted as reference value 0.
The device emits the corresponding tone and/or
shows the according value for rising, deviating concentrations, e.g. directly above or near the leakage.
The sensitivity control with its six settings also re-
2
mains functional.
It is thus possible to locate very high H
concentra-
2
tions in surroundings where trace gas is already enriched.
The first measurement occurs in basement room 1 –
unsuccessful. After approx. 5 minutes, the next unsuccessful measurement occurs in basement room 2.
Thus, the leakage must be located in basement
room 3. After 10 minutes, the user enters the third
basement room with the enriched trace gas. The device shows a high base level and emits beeps at a high
frequency.
GBOperating manual – Multifunction measuring meter T3000B-46
Page 48
Errors and faults at the trace gas sensor
The accurate functionality of the trace gas sensor was
tested during production a number of times. However,
if functionality faults do occur, then check the trace
gas sensor according to the following list.
Fault at basic calibration:
• Never heat the trace gas sensor near a source of
hydrogen.
The displayed measured value rises over a long
period continually by approx. 20 to 30 digits:
• This so-called basic noise occurs when the "Sensitivity" selector switch is set to position 5 or 6.
This effect is normal and negligible over the entire
measuring range of 1000 digits.
The trace gas sensor reacts sensitively to sources
of interference such as flammable gases:
• Due to the physical principle of the detection process, the trace gas sensor does not only react to
hydrogen, but also to other flammable gases (e.g.
alcohol-air mixtures, methane or hydrocarbons).
• Use the differentiated indication measurement to
bypass existing, permanent sources of interference on a case by case basis (see chapter "Differentiated indication measurement"). This lowers
the sensitivity to sources of interference and raises
the sensitivity for detecting hydrogen.
Cleaning the trace gas sensor
Caution!
The silver sensor protection cap is sealed
with a special sticker.
Do not remove the silver sensor protection cap. Otherwise, the sensor may be
damaged.
Caution!
Never use water or harsh cleaners for
cleaning. These can damage the sensor
head if they come into contact with it.
1. After measuring with the trace gas sensor, wait
approx. 5 minutes before cleaning it. After approx. 5 minutes, the silver sensor protection cap
has cooled down sufficiently.
2. Only clean the silver sensor protection cap externally by using a short bristled brush (e.g. file
brush) while applying only slight pressure.
3. Ensure that the silver sensor protection cap is
screwed on tightly. Particulate must not enter below the silver sensor protection cap or directly
onto the sensor element.
4. Remove any loose dust with compressed air. Ensure that the compressed air flow is only used
near the silver sensor protection cap.
B-47Operating manual – Multifunction measuring meter T3000GB
Page 49
Available SDI sensors
170 171 172 173
174175 176 177 178
An overview of the currently available SDI sensors is
provided below.
Further details about SDI sensors and their use is provided in the separately Practice Handbook.
No.SDI sensor
170TS 210 SDI
Climate sensor
171TS 230 SDI
Climate sensor
172TS 250 SDI
Climate sensor
173TS 660 SDI
Dielectric humidity sensor
No.SDI sensor
174TS 610 SDI
Microwave humidity sensor
175TS 410 SDI
Anemometer sensor
176TS 470 SDI
Anemometer sensor
177TS 800 SDI
Trace gas sensor system (also see chapter "Using the
trace gas sensor TS 800 SDI (optional)")
178TS 131/150 SDI
Surface temperature sensor, 5 mm
GBOperating manual – Multifunction measuring meter T3000B-48
Page 50
Available BNC electrodes
179180181
182183184
An overview of the currently available BNC electrodes
is provided below.
Further details about BNC electrodes and their use is
provided in the separately available Practice Handbook.
No.BNC electrode
179TS 16/300
Flat electrodes, 1 mm, insulated
180TS 16/200
Flat electrodes, 1 mm, insulated
181TS 12/300
Round electrodes, 4 mm, insulated
No.BNC electrode
182TS 12/200
Round electrodes, 4 mm, insulated
183TS 4/300
Round electrodes, 2 mm
184TS 4/200
Round electrodes, 2 mm
B-49Operating manual – Multifunction measuring meter T3000GB
Page 51
No.BNC electrode
185186187188
189190
185TS 70
Ram electrode
186TS 8/300
Round electrodes, 4 mm
187TS 8/200
Round electrodes, 4 mm
188TS 60
Hand electrode
No.BNC electrode
189TS 24/250
Layer depth electrodes, 8 mm
190TS 20/110
Brush electrodes, 7 mm, insulated
GBOperating manual – Multifunction measuring meter T3000B-50
Page 52
Further accessories (optional)
The following accessories are optionally available.
Contact TROTEC® customer service in this case:
AccessoriesArticle number
MultiMeasure case 2ZB9119017
MultiMeasure case 53510200921
Holster 2 MM series-
Holster 5 MM series-
Battery pack for TS 800 SDI3510200209
Charger for battery pack3510200208
Telescopic rod for T3000 sensors3510200221
Stainless steel sinter filter for T 200, T 250,
TS 200 SDI
Teflon coated electrode tips, 45 mm long3510200212
Teflon coated electrode tips, 60 mm long3510200213
Replacement electrode tips, not insulated 3510200214
Contact mass3510200217
TC 20 BNC connection cable3510200024
TC 30 SDI connection cable3510200025
Calibration block for SDI sensors3510200216
Calibration ampoules for calibration block
(supplied for humidity values 35 %, 50 %
and 80 %)
Test block for BNC electrodes3510200226
Display protection film3510200220
MultiMeasure Studio Professional PC software for analysing the measured results
3510200211
3510200215
3510204010
B-51Operating manual – Multifunction measuring meter T3000GB
Page 53
DisposalDeclaration of conformity
In the European Union, electronic equipment must not be treated as domestic
waste, but must be disposed of profes-
sionally in accordance with Directive
2002/96/EC of the European Parliament and Council
of 27th January 2003 concerning old electrical and
electronic equipment. After the end of its use, please
dispose of this device in a manner appropriate to the
relevant legal requirements.
In the European Union, batteries must not be treated
as domestic waste, but must be disposed of professionally in accordance with Directive 2006/66/EC of
the European Parliament and Council of 6th September 2006 concerning batteries and accumulators.
Please dispose of batteries in a manner appropriate to
the relevant legal requirements.
in accordance with the EC Low Voltage Directive
2006/95/EC and the EC Directive 2004/108/EC about
electromagnetic compatibility.
Herewith, we declare that the T3000 multifunction
measuring meter was developed, constructed and
produced in compliance with the named EC directives.
Applied harmonised standards:
EN 61326-1:2006,
EN 61326-2-1:2006,
IEC 61326-1:2005,
IEC 61326-2-1:2005
The marking is found on the rear of the device.
Manufacturer:
Trotec GmbH & Co. KG
Grebbener Straße 7
D-52525 Heinsberg
Phone:+49 2452 962-400
Fax:+49 2452 962-200
E-mail:info@trotec.de
Heinsberg, 05/04/2013
Managing Director: Detlef von der Lieck
GBOperating manual – Multifunction measuring meter T3000B-52
Page 54
TROTEC® GmbH & Co. KG • Grebbener Straße 7 • D-52525 Heinsberg
Tel.: +49 2452 962-400 • Fax: +49 2452 962-200
www.trotec.com • E-mail: info@trotec.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.