Safety

Notes regarding the operating manual................................. 1
Safety .....................................................................................1
Information about the device................................................2
Transport and storage...........................................................3
Operation ...............................................................................4
Measuring principle ..............................................................5
Maintenance and repair ........................................................6
Disposal .................................................................................6

Notes regarding the operating manual

Symbols

Warning of electrical voltage
This symbol indicates dangers to the life and health of persons due to electrical voltage.
Warning
This signal word indicates a hazard with an average risk level which, if not avoided, can result in serious injury or death.
Caution
This signal word indicates a hazard with a low risk level which, if not avoided, can result in minor or moderate injury.
Note
This signal word indicates important information (e.g. material damage), but does not indicate hazards.
Info
Information marked with this symbol helps you to carry out your tasks quickly and safely.
Follow the manual
Information marked with this symbol indicates that the operating manual must be observed.
You can download the current version of the operating manual and the EU declaration of conformity via the following link:
Read this manual carefully before starting or using the device. Always store the manual in the immediate vicinity of the device or its site of use!
Warning Read all safety warnings and all instructions.
Failure to follow the warnings and instructions may result in electric shock, fire and/ or serious injury.
Save all warnings and instructions for future reference.
This appliance can be used by children aged from 8years and above and persons with reduced physical, sensory or mental capabilities or lack of experience and knowledge if they have been given supervision or instruction concerning use of the appliance in a safe way and understand the hazards involved. Children shall not play with the appliance. Cleaning and user maintenance shall not be made by children without supervision.
Do not use the device in potentially explosive rooms.
Do not use the device in aggressive atmosphere.
Protect the device from permanent direct sunlight.
Do not remove any safety signs, stickers or labels from the device. Keep all safety signs, stickers and labels in legible condition.
Do not open the device with a tool.
Observe the storage and operating conditions as given in the Technical data chapter.

Intended use

Only use the device for moisture measurements of wood and building materials within the measuring range specified in the technical data.
To use the device for its intended use, only use accessories and spare parts which have been approved by Trotec.

Improper use

Do not use the device in potentially explosive atmospheres, for measurements in liquids or at live parts. Trotec accepts no liability for damages resulting from improper use. In such a case, any warranty claims will be voided. Any unauthorised modifications, alterations or structural changes to the device are forbidden.

Personnel qualifications

People who use this device must:
BM40
https://hub.trotec.com/?id=41251
1 EN
moisture measuring device BM40
have read and understood the operating manual, especially the Safety chapter.
Page 3

Residual risks

1
2
3
4
BM40
5
6
7
8
9
Warning
Risk of suffocation! Do not leave the packaging lying around. Children may use it as a dangerous toy.
Warning
The device is not a toy and does not belong in the hands of children.
Warning
Dangers can occur at the device when it is used by untrained people in an unprofessional or improper way! Observe the personnel qualifications!
Caution
Keep a sufficient distance from heat sources.
Note
To prevent damages to the device, do not expose it to extreme temperatures, extreme humidity or moisture.
Note
Do not use abrasive cleaners or solvents to clean the device.

Device depiction

No. Designation
1 Measuring probe
2 Display
3 Power/ MODE button
4 Battery compartment with cover
Display

Information about the device

Device description

The device BM40 is suited for the non-destructive material moisture measurement in mineral building materials and wood.
The measuring probe is pressed against the material to be tested. The moisture level is then measured by capacitive sensors using the dielectric measurement method, which is absolutely non-destructive since no electrodes or probes have to be inserted into the material.
For an increased accuracy of the measured values, which are directly indicated in % by weight, there are 10selectable material curves available for the most important types of wood and building materials.
The device further features the following functions:
Real-time measurement value display
Temperature measurement
Automatic switch-off
No. Designation
5 Measuring mode indication
6 °C / °F indication
7 % indication
8 Measurement value indication
9 Battery indication
EN 2
moisture measuring device BM40
Page 4

Technical data

Transport and storage

Parameter Value
Model BM40
Measuring principle Dielectric/capacitive
Electrode length Approx. 40 mm
Penetration depth, non-destructive 0 mm to 50 mm
Measuring range Solid brick 0 % to 20 %
Anhydrite screed 0 % to 20 %
Cement screed 0 % to 20 %
Sand-lime brick 0 % to 20 %
Gas concrete 0 % to 20 %
High-insulating
0 % to 20 %
brick
Concrete 0 % to 20 %
Hardwood 0 % to 50 %
Softwood 0 % to 50 %
Chipboard 0 % to 50 %
Temperature -10°C to 55°C
Resolution Wood 0.1 %
Mineral building
0.1 %
materials
Temperature 0.1°C
Ambient temperature -5 to 50°C with
0 to 85%RH
Power supply 1 x 9V battery
Automatic switch-off After approx. 10min
Dimensions (length x width x height) 168 x 35 x 75 mm
Weight 220 g
Note
If you store or transport the device improperly, the device may be damaged. Note the information regarding transport and storage of the device.

Transport

For transporting the device use a suitable bag to protect it from external influences.

Storage

When the device is not being used, observe the following storage conditions:
dry and protected from frost and heat
protected from dust and direct sunlight
with a cover to protect it from invasive dust if necessary
The storage temperature is the same as the range given in the Technical data chapter.
Remove the batteries from the device.
Scope of delivery
1 x Moisture measuring device
1 x 9V battery
1 x Quick guide
3 EN
moisture measuring device BM40
Page 5

Operation

+ -
Inserting the battery
Insert the supplied battery into the device before first use.
Note
Make sure that the surface of the device is dry and the device is switched off.
Selecting the measuring mode
For an increased accuracy of the measured values, which are directly indicated in % by weight, 10selectable material curves are available for the following types of wood and building materials:
Measuring
Designation Measuring range
mode
B1 Solid brick 0 % to 20 %
B2 Anhydrite screed 0 % to 20 %
B3 Cement screed 0 % to 20 %
B4 Sand-lime brick 0 % to 20 %
B5 Gas concrete 0 % to 20 %
B6 High-insulating brick 0 % to 20 %
B7 Concrete 0 % to 20 %
H1 Hardwood (e.g. maple, birch,
0 % to 50 % beech, Douglas fir, oak, ash, walnut)
H2 Softwood (e.g. spruce, pine,
0 % to 50 % cherry, larch, Meranti, poplar)
H3 Chipboard 0 % to 50 %
°C Temperature -10°C to 55°C
1. Open the battery compartment at the rear of the device by sliding down the cover.
2. Use the battery clip to connect the 9V battery with correct polarity.
3. Place the battery with the battery clip into the battery compartment.
4. Slide the cover back onto the battery compartment until it locks in place.
Switching the device on
Info
Note that moving from a cold area to a warm area can lead to condensation forming on the device's circuit board. This physical and unavoidable effect can falsify the measurement. In this case, the display shows either no measured values or they are incorrect. Wait a few minutes until the device has become adjusted to the changed conditions before carrying out a measurement.
1. Press the power button(3) for approx. 1s.
ð The device starts in the measuring mode°C. ð The measured value is displayed in real time on the
measurement value display(8).
1. Repeatedly press the MODE button(3) until the desired measuring mode is set.
Moisture measurement
Info
When pressing the contact surface of the measuring probe against the object, apply a steady contact pressure (approx. 10N which corresponds to approx. 1kg of weight force) but do not bend it to the point of overlapping. In order to become a better judge of what the right level of contact pressure might be, we recommend pressing the device a few times against a pair of scales.
1. Use the MODE button(3) to select the desired measuring mode.
2. Hold the device against the object at an angle of 90° and make sure that the entire width of the measuring probe rests on the contact surface.
ð The measured value is displayed in real time on the
measurement value display(8).
Switching the device off
The device switches off automatically after approx. 10min of non-use.
1. Press the power button(3) for approx. 3s.
ð The device switches off.
EN 4
moisture measuring device BM40
Page 6

Measuring principle

The measuring device at hand serves for the rough determination of the material or wood moisture content according to the dielectric measuring principle. Fields of application include the non-destructive moisture measurement of sawn timber, lumber and firewood as well as of mineral building materials.
The dielectric measuring method is an indirect measurement method, since the moisture content of the measured material is deduced from the changed dielectric permittivity.
Influencing variables suitable to change this parameter, e.g. electrically conducting materials or dissolved salts, hence also have a direct influence on the determined measured values. Since the exact composition of building materials is generally unknown before a moisture measurement, the displayed measured values can only be regarded as an indicator of the moisture content.
General notes on use
Calibration curves for measuring the moisture level in different mineral building materials, hardwood, softwood and chipboard are stored in the device. If precise material moisture values are required, an additional control measurement applying the Darr procedure is recommended. Especially for precise wood moisture measurements for other wood temperatures or with due regard to the type and bulk density of a certain timber it is recommended to use a wood moisture measuring device with temperature comparison function and selection option for the specific wood type calibration.
When selecting the measuring positions observe the following:
– Always measure the material moisture at three different
positions in order to achieve a sufficient accuracy by means of the arithmetic average.
– Do not measure the face, because there are mainly dry
areas.
– Preferably do not measure at cracks, branches or resin
pockets either.
Notes on use regarding the material moisture measurement:
No further settings are required for the quick determination of the material moisture at the device. Nevertheless, when assessing the measurement results one has to bear in mind that the results can be influenced by various factors.
The measurement's accuracy depends on the contact pressure of the measuring electrodes. Ideally, they should be applied with a steady contact pressure of approx. 10N (this corresponds to approx. 1kg of weight force), however not bent to the point of overlapping.
Measuring depth
– Depending on the material's bulk density and degree of
moisture penetration, the device's penetration depth ranges up to 50mm. Conclusions regarding zones located at a lower level cannot be drawn.
Material characteristics
– Increased occurrences of soluble salts in the measured
material can distort the result of the measurement. The more salts present, the higher the conductivity of the material and the higher will be the displayed measurement value.
– If possible, do not measure wood with a temperature
below -5°C. Too low wood temperatures distort the result of the measurement.
– The material should be preferably homogenous and not
have any air pockets.
Electrically conducting materials or electrical fields
– If a construction material contains electrically
conducting substances, the dielectric constant is also increased, which then simulates high moisture values. Consequently, a too high measured value will be displayed.
– From visual inspection it is not always apparent,
whether there are any electrically conducting substances inside the construction material. Some of the main error sources here are in particular reinforcements, metal laminations and conducting insulation such as slag in timber beam ceiling constructions. Especially in case of insulation materials with metal lamination, measured values are often misinterpreted.
– Avoid static charge of the material to be measured due
to friction. Static charge distorts the measurement result.
– With a wood moisture of less than 10%RH,
electrostatic forces may occur at the good to be measured. This can falsify the result of the measurement. Experience has shown that this occurs at the outlet of veneer drying plants. Remove the static charge by use of suitable grounding measures.
– For optimum conditions neither electrical fields nor
conducting materials should be present.
5 EN
moisture measuring device BM40
Page 7
Surface condition
Li-Ion
– Oily and/or aqueous timber preservatives affect the
measurement result.
– The surface of the material to be measured should be
as level as possible to permit thorough contact of the measuring probe.
Moisture distribution
– With material thicknesses of less than 2 cm there is a
danger of humidity values from adjacent material layers affecting the measured value.
– Ideally, the moisture should be evenly distributed within
the material to be measured.
Quantitative conclusions about the moisture content of the mineral material to be measured can only be drawn by applying the Darrprocedure or the CMmethod.

Maintenance and repair

Battery change

A battery change is required when the battery indication(9) is displayed or the device can no longer be switched on (see chapter Inserting the battery).

Cleaning

Clean the device with a soft, damp and lint-free cloth. Ensure that no moisture enters the housing. Do not use any sprays, solvents, alcohol-based cleaning agents or abrasive cleaners, but only clean water to moisten the cloth.

Disposal

The icon with the crossed-out waste bin on waste electrical or electronic equipment stipulates that this equipment must not be disposed of with the household waste at the end of its life. You will find collection points for free return of waste electrical and electronic equipment in your vicinity. The addresses can be obtained from your municipality or local administration. For further return options provided by us please refer to our website www.trotec24.com.
The separate collection of waste electrical and electronic equipment aims to enable the re-use, recycling and other forms of recovery of waste equipment as well as to prevent negative effects for the environment and human health caused by the disposal of hazardous substances potentially contained in the equipment.
In the European Union, batteries and accumulators must not be treated as domestic waste, but must be disposed of professionally in accordance with Directive 2006/66/EC of the European Parliament and of the Council of 6September 2006 on batteries and accumulators. Please dispose of batteries and accumulators according to the relevant legal requirements.

Repair

Do not modify the device or install any spare parts. For repairs or device testing, contact the manufacturer.
EN 6
moisture measuring device BM40
Page 8
Page 9
Trotec GmbH & Co. KG
Grebbener Str. 7 D-52525 Heinsberg
+49 2452 962-400 +49 2452 962-200
info@trotec.com www.trotec.com
Loading...