Trotec BB20 User Manual

BB20
EN
OPERATING MANUAL
LAYER THICKNESS MEASURING DEVICE
TRT-BA-BB20-TC2019-05-001-EN
Notes regarding the operating manual................................. 2
Safety .....................................................................................2
You can download the current version of the operating manual and the EU declaration of conformity via the following link:
Information about the device................................................3
Transport and storage...........................................................5
Operation ...............................................................................5
Menu structure ....................................................................10
PC software..........................................................................11
Errors and faults..................................................................11
Maintenance and repair ......................................................11
Disposal ...............................................................................12

Notes regarding the operating manual

Symbols

Warning of electrical voltage
This symbol indicates dangers to the life and health of persons due to electrical voltage.
Warning
This signal word indicates a hazard with an average risk level which, if not avoided, can result in serious injury or death.
Caution
This signal word indicates a hazard with a low risk level which, if not avoided, can result in minor or moderate injury.
Note
This signal word indicates important information (e.g. material damage), but does not indicate hazards.
Info
Information marked with this symbol helps you to carry out your tasks quickly and safely.
Follow the manual
Information marked with this symbol indicates that the operating manual must be observed.
BB20
https://hub.trotec.com/?id=43284

Safety

Read this manual carefully before starting or using the device. Always store the manual in the immediate vicinity of the device or its site of use!
Warning Read all safety warnings and all instructions.
Failure to follow the warnings and instructions may result in electric shock, fire and/ or serious injury.
Save all warnings and instructions for future reference.
This appliance can be used by children aged from 8years and above and persons with reduced physical, sensory or mental capabilities or lack of experience and knowledge if they have been given supervision or instruction concerning use of the appliance in a safe way and understand the hazards involved. Children shall not play with the appliance. Cleaning and user maintenance shall not be made by children without supervision.
Do not use the device in potentially explosive rooms.
Do not use the device in aggressive atmosphere.
Protect the device from permanent direct sunlight.
Do not remove any safety signs, stickers or labels from the device. Keep all safety signs, stickers and labels in legible condition.
Do not open the device.
Observe the storage and operating conditions as given in the Technical data chapter.

Intended use

Only use the device for coating thickness measurements. Observe and comply with the technical data.
To use the device for its intended use, only use accessories and spare parts which have been approved by Trotec.
2 EN
layer thickness measuring device BB20

Improper use

1
2
8
7
9
5
4
3
10
6
Do not use the device in potentially explosive atmospheres, for measurements in liquids or at live parts. Trotec accepts no liability for damages resulting from improper use. In such a case, any warranty claims will be voided. Any unauthorised modifications, alterations or structural changes to the device are forbidden.

Personnel qualifications

People who use this device must:
have read and understood the operating manual, especially the Safety chapter.

Residual risks

Warning
Risk of suffocation! Do not leave the packaging lying around. Children may use it as a dangerous toy.
Warning
The device is not a toy and does not belong in the hands of children.

Information about the device

Device description

The deviceBB20 is used to determine the coating thickness on ferromagnetic and non-ferromagnetic metal surfaces. The measuring device functions according to the magnetic induction principle (for coatings on ferromagnetic surfaces) or the turbulent flow principle (for coatings on non-ferromagnetic surfaces).
The BB20 is a compact and versatile device suitable for quick, precise and non-destructive measurements of coating thicknesses.

Device depiction

Warning
Dangers can occur at the device when it is used by untrained people in an unprofessional or improper way! Observe the personnel qualifications!
Caution
Keep a sufficient distance from heat sources.
Note
To prevent damages to the device, do not expose it to extreme temperatures, extreme humidity or moisture.
Note
Do not use abrasive cleaners or solvents to clean the device.
No. Designation
1 USB interface
2 Blue button (back/ cancel)
3
button
4 ZERO button
5 Sensor
6 Battery compartment with cover
7 Power button
8
button
9 Red button (select/ confirm)
10 Display
EN 3
layer thickness measuring device BB20
Display
DIR GRO
NO=3 MAX=63
NFe
58
18 17
16
15
14
13
11
12
No. Designation
11 Working mode display
12 Battery indication
13 PC access display
14 NFe/Fe display
15 Indication of the unit
16 Layer thickness display
17 Statistical data display: AVG, MAX, MIN, SDEV
18 Number of measured values

Technical data

Parameter Value
Accuracy 0 to 50μm
(0.1μm) 50 to 850μm (1μm) 850 to 1250μm (0.01mm) 0 to 1.968mils (0.001mils)
1.968 to 33.46mils (0.01mils)
33.46 to 49.21mils (0.1mils)
Minimum bending radius
Diameter of the smallest measuring surface
Critical primary layer thickness
Ambient temperat­ure
Power supply 2 batteries 1.5 V AAA
Dimensions (length x width x height)
Weight 152 g
1.5 mm 3 mm
7 mm 5 mm
0.5μm 0.3μm
0 to 40°C (32 to 104°F) at 20 to 90%RH
114x 27x 54 mm
0 to 50μm (0.1μm) 50 to 850μm (1μm) 850 to 1250μm (0.01mm) 0 to 1.968mils (0.001mils)
1.968 to 33.46mils (0.01mils)
33.46 to 49.21mils (0.1mils)
Parameter Value
Model BB20
Sensor
Functional principle Magnetic induction Turbulent flow
Measuring range 0 to 1250 μm
Guaranteed tolerance (of the measured value)
F N
0 to 49.21 mils
0 to 850μm (±3 % ±1μm) 850 to 1250μm (±5 %) 0 to 33.46mils (±3 % ±0.039mils)
33.46 to 49.21mils (±5 %)
0 to 1250 μm 0 to 49.21 mils
0 to 1000μm (±3 % ±1.5μm) 850 to 1250μm (±5 %) 0 to 33.46mils (±3 % ±0.059mils)
33.46 to 78.7mils (±5 %)
Scope of delivery
1 x Layer thickness measuring device BB20
2 x 1.5 V batteries AAA
1 x Transport case
1xSteel and aluminium base plate with different film thicknesses for calibration
1 x USB cable
1x Operating manual
1x PCsoftware
4 EN
layer thickness measuring device BB20

Transport and storage

Operation

Note
If you store or transport the device improperly, the device may be damaged. Note the information regarding transport and storage of the device.

Transport

For transporting the device, use the transport case included in the scope of delivery in order to protect the device from external influences.

Storage

When the device is not being used, observe the following storage conditions:
dry and protected from frost and heat
protected from dust and direct sunlight
For storing the device, use the transport case included in the scope of delivery in order to protect the device from external influences.
The storage temperature is the same as the range given in the Technical data chapter.
Remove the batteries from the device.
Inserting the batteries
Insert the supplied batteries before first use.
1. Loosen the screw and open the battery compartment(6).
2. Insert the new batteries in the battery compartment(6) with correct polarity.
3. Place the cover back on the battery compartment(6) and reinsert the screw.
Switching the device on
Info
Note that moving from a cold area to a warm area can lead to condensation forming on the device's circuit board. This physical and unavoidable effect can falsify the measurement. In this case, the display shows either no measured values or they are incorrect. Wait a few minutes until the device has become adjusted to the changed conditions before carrying out a measurement.
1. Press the Power button(7).
ð The display will be switched on and the device ready for
operation.
Menu
Press the Red button(9) to open the main menu.
Press the (3) or button(8) to select the desired menu item.
In order to confirm the selected menu item, press the Red button(9) again.
Press the Blue button(2) to return to the previous menu item.
EN 5
layer thickness measuring device BB20
Setting the measuring mode
To adjust the measuring mode, please proceed as follows:
1. Press the Red button(9) to open the main menu.
2. In the menu, select the menu item Options and confirm by pressing the Red(9) button.
3. Select the Measure mode menu item, then, select the desired mode. Confirm with the Redbutton(9).
Single mode:
After every measurement, an acoustic signal will be emitted. All measurements are automatically saved.
Continuous mode:
The sensor must not be lifted between the individual measurements. No acoustic signal will be emitted after the measurement. All measurements are automatically saved.
Setting the working mode
To set the working mode, please proceed as follows:
1. Press the Red button(9) to open the main menu.
2. In the menu, select the menu item Options and confirm by pressing the Red(9) button.
3. Select the Working mode menu item, then, select the desired mode. Confirm with the Redbutton(9).
The device offers two different working modes:
DIRECT:
This mode is suitable for quick and easy measurements. Measurements are cached during a measurement series.
Once the device is switched off or changes into another working mode, the measurement results will be deleted. The statistical analysis programme is able to assess80measurements. Once the memory is full, the oldest measurements will be overwritten.
GROUP:
The GROUP mode comprises group1to4. In each group, 80individual and 5statistical measurements can be saved.
Calibration and limit values can be adjusted and saved individually.
Once the memory is full, current measurements will no longer be saved. Measurements can be carried out as usual.
Setting the sensor mode
To adjust the sensor mode, please proceed as follows:
1. Press the Red button(9) to open the main menu.
2. In the menu, select the menu item Options and confirm by pressing the Red(9) button.
3. Select the Used probe menu item, then, select the desired mode. Confirm with the Redbutton(9).
AUTO:
The sensor automatically selects a measuring mode.
Fe:
The sensor operates according to the magnetic induction principle if it is placed on a magnetic surface.
Not Fe:
The sensor operates according to the turbulent flow principle if it is placed on non-magnetic metals.
Units
You can choose between metric (μm and mm) and imperial(mils) units.
If the measured value exceeds850μm, the device switches to mm.
To change the unit, proceed as follows:
1. Press the Red button(9) to open the main menu.
2. In the menu, select the menu item Options and confirm by pressing the Red(9) button.
3. Select the Unit setting menu item, then, select the desired unit. Confirm with the Redbutton(9).
Reset
Resetting deletes the entire data in all memories. In order to carry out a reset, follow these steps:
1. Switch the device off.
2. Simultaneously press the Zero button(4) and the Power button(7).
ð sure to reset will be displayed.
3. Press the Red button(9) to confirm or the Blue button(2) to cancel the process.
The device automatically switches back on.
Background illumination
In the menu, you can control the background illumination using Options and Backlight, during a measurement, use the Bluebutton(2).
6 EN
layer thickness measuring device BB20
Indication of statistics
In the menu item LCD Statistic, you can choose between the following indications:
Average
Maximum
Minimum
Standard deviation (Sdev.)
The selected indication appears in the bottom right corner of the display in the statistical data indication(17).
Statistics of the measured values appear on the bottom left corner in the display in the measured values indication(18).
To view all statistical values of the respective groups, go to the menu and select Statisticview.
Measured value indication
To view all measured values of the respective groups, go to the menu and select Measureview.
Automatic switch-off
The device switches off after 3 minutes of inactivity. To deactivate automatic switch-off, go to the Options section in the menu and select the Auto poweroff function. Select Disable and confirm by pressing the Red(9) button.
Device calibration
Carry out a calibration before each measuring operation.
You can perform the calibration e.g. at an untreated or uncoated spot of the object to be tested or use the set included in the scope of delivery.
Calibration mode
1. Press the Red button(9) to open the main menu.
2. Select the Calibration menu item, then select Enable.
ð You return automatically back to Calibration.
3. Press the Blue button(2) to return back to the start screen. Calibration is carried out here.
ð The following messages appear on the display:
cal zero: There is no one-point or two-point calibration.
cal1to2: There is a one-point or two-point calibration.
zeroy: There is a zero-point calibration.
Info
Please bear in mind that the device at hand is a precision measuring device that can determine coating thicknesses of no more than a few micrometres (1µm is equivalent to one thousandth of a millimetre). The surface condition of most measuring objects is hardly ever perfectly even and homogeneous, even though it might appear differently to the naked eye. Observed under the microscope, even the smoothest surface looks much like a mountain and valley landscape. The teensiest of scratches, cavities or contaminations can therefore already have a negative effect on the expected measurement result, seeing as they will also be a part of the measurement to a greater or lesser extent. It does, however, not affect the accuracy of the device. Even after the calibration unexpected measurement deviations of a few micrometres always have to be regarded in this context. Hence it is important to handle the supplied calibration accessories with care to avoid scratches and dirt on their respective surface as far as possible.
EN 7
layer thickness measuring device BB20
Zero point calibration
ü An uncoated sample is required.
1. Place the sensor onto an uncoated sample. The calibration sample is to be identical with the actual sample in terms of material composition, shape and surface condition.
ð The following message will be displayed:
<x.x μm>
2. Lift the sensor again and keep it away from the sample (at least 10cm).
3. Press and hold the Zero button(4) for approx. 2seconds.
ð The following result appears on the display:
<0.0μm> CAL1 ZeroY
ð The calibration is terminated.
One-point calibration
This calibration is suited for high-precision measurements, e.g. for thin coatings.
ü A calibration film, a coated and an uncoated sample are
required.
1. Activate the calibration mode as described before.
2. Carry out a zero-point calibration.
3. Place the calibration film on an uncoated sample.
4. Place the sensor on the sample.
ð The measurement is effected.
5. Lift the sensor again.
6. Adjust the required film thickness by using the Down/ Right(3) or the Up/Left(8) button. The film thickness should be at least the size of the estimated coating thickness.
7. Repeat this step several times.
8. Now carry out the measurement on a coated sample.
9. Press the Zero button(4) to adopt the current calibration.
10. To exit the current calibration, press the Blue button(2).
Two-point calibration
This calibration is especially suited for measurements on rough surfaces or for high-precision measurements.
ü You need two films of different film thickness. The thicker
film should be approx.1.5 times thicker than the thinner film.
1. At first, carry out a zero-point calibration.
2. Using the first film, carry out a one-point calibration.
3. Repeat this step with the second film.
4. Place the sensor onto the coating that is to be measured. After a brief acoustic signal is emitted, you can lift the sensor again.
ð The measured value appears on the display.
Calibration on radiated surfaces
Values of layer thickness measurements carried out on radiated surfaces are usually too high. The average thickness can be determined as follows:
Method A
1. Carry out the steps for one-point and two-point calibration as described above. Use a smooth sample with the same surface curvature and material type as the measuring object that is to be used later on.
2. In order to determine the average value Xo, carry out approx.10 measurements on the uncoated sample.
3. Afterwards, carry out approx.10 measurements on the coated sample to determine the average value Xm.
ð The difference between both average values is the
average coating thickness Xeff. Take the standard deviation of both values Xm and Xo into account.
Xeff= (Xm - Xo)±S
Method B
1. Carry out a zero calibration with approx.10measurements on an uncoated sample.
2. Afterwards, carry out a film calibration on the uncoated sample.
3. Repeat this calibration with films of different thicknesses (max. thickness50μm). Together, these films should be as thick as the assumed layer thickness.
ð The layer thickness can be read from the display.
Method C
1. Carry out the steps for two-point calibration with two films as described above.
2. Use several 50μm films to approach the actual surface as precisely as possible.
ð The layer thickness can be read from the display.
8 EN
layer thickness measuring device BB20
General information on measurements
If the calibration has been carried out correctly, the measured value lies within the guaranteed measuring tolerance. Incorrect measured values can be deleted in the menu. The last value results from the statistical calculation and the guaranteed degrees of tolerance of the measured value.
Strong electrical or electromagnetic fields can influence the measured values.
The measuring tip is to be cleared of contamination prior to each calibration.
Wait for at least 4seconds after each measurement before proceeding to the next measurement as the device does not respond to measurement sequences carried out too quickly.
Deleting functions
In order to delete measured values, select Delete in the menu, then select the measured values you would like to delete.
Current Data
Deletes the last measured value.
All Data
All data can be deleted during the respective working mode.
Group Data
Comprises Delete all data. In addition, the limit values and one-point and two-point calibration values will be deleted.
Statistical measurements
The device is able to calculate statistics based on 80measurements. A total of 400measured values can be saved.
No measured values are saved in direct mode, however, calculating statistics is possible.
If you change between the different working modes or if the device is switched off, statistics are deleted from the direct mode.
The following statistical values are calculated:
NO.: Number of measured values in working mode
AVG: Average
Sdev.: Standard deviation
MAX: Maximum value
MIN: Minimum value
Memory space full
If the memory space is full in group mode, statistics will no longer be updated. However, measurements can still be carried out. These measurements will not be part of the statistics.
In single measurement mode, FULL appears on the display.
If the memory space is full in direct mode, old measurements are replaced by new ones and statistics will be updated.
Limit value functions
Limit values can be entered using the Limit function. It is possible to enter limit values before, during or after carrying out measurement series.
1. Press the Red button(9) to open the main menu.
2. Select the Limit menu item, then select Limit setting.
ð Using the (3) and (8) buttons, you can
determine the upper limit value in the High limit section and the lower limit value in the Low limit section.
Each measured value beyond the fixed tolerance limit appears on the display as follows:
H: Measured value is above upper limit value.
L: Measured value is below lower limit value.
EN 9
layer thickness measuring device BB20

Menu structure

Main menu Submenu 1 Submenu 2
Statistic view (Statistic view)
Options Measuring mode Single mode
Limit values (Limit) Limit value settings
Average view Minimum view Maximum view Number view Standard deviation view (Sdev.view)
Working mode (Working mode)
Used probe (Used probe)
Units (Unit setting)
Background illu­mination (Back- light)
LCD Statistic
LCDStatistic
Automatic switch­off (Autopoweroff)
(Limitsettings)
Delete limit value (Delete limit)
-
Continuous mode
Direct Group 1 Group 2 Group 3 Group 4
AUTO Fe Not Fe
μm mils mm
On Off
Average Maximum Minimum Standard deviation
Activate Deactivate
Upper limit value Lower limit value
-
Main menu Submenu 1 Submenu 2
Delete Current data -
All data
Group data
Measured value
- ­indication (Measure view)
Calibration Activate (Enable) -
Deactivate (Disable)
Delete zeroN
Delete zeroF
10 EN
layer thickness measuring device BB20

PC software

Errors and faults

Using the software CoatingThicknessTester stored measurement data can be called up and saved via USB cable.
The software is available for download at www.trotec.de.
Info
The supplied free software is designed for useful basic functionalities. Trotec assumes no liability with regard to this free software and also provides no support on that score. Trotec accepts no liability concerning the use of this free software and is under no obligation to make adjustments or to further develop updates or upgrades.
Installation requirements
Ensure that the following minimum requirements for installing the PC software are fulfilled:
Supported operating systems:
– Windows 7 – Windows 8 – Windows 8.1 – Windows 10
Hardware requirements:
– Standard USB interface – Min. 7MB of free hard disk space – Recommended resolution: 1280*1024 with16bit – .NET Framework 2.0 or higher
The device has been checked for proper functioning several times during production. If malfunctions occur nonetheless, check the device according to the following list.
The device does not switch on:
Check the charging status of the batteries. Change the batteries, if required.
Check that the batteries are properly positioned. Check the polarity is correct.
Never carry out an electrical check yourself; instead, contact your Trotec customer service.
Table of faults
The following error codes can be displayed:
Error code Cause of error
Err1, Err2, Err3 Faulty contact between sensor and
mainboard
Err1 Faulty signal of eddy-current sensor
Err2 Faulty signal of magnetic induction sensor
Err3 Faulty signals of both sensors
Err4, Err5, Err6 No meaning, reserved for later versions of
the device
Err7 Faulty measurement. There might be a
hardware problem.
Installing the PC software
1. Insert the data medium with the software into the drive or download the current software from the Service area of Trotec download centre.
2. Double-click the installation file Setup.exe.
3. Follow the instructions of the installation wizard.
Starting the PC software
1. Connect the device to the PC using the supplied USB cable.
2. Start the PC software.
ð The software connects to the device. ð Data stored on the device can now simply be visualised
in form of a chart or graphic.
Saving measured values (export)
You can export a selected group as csv or txt file and save it on your PC. There the table looks as it does in the software.
1. Open the desired file.
2. You select either Save *.TXT or Save *.CSV.

Maintenance and repair

Battery change

A battery change is required, when the battery indication(12) lights up or the device can no longer be switched on. See chapter Operation

Cleaning

Clean the device with a soft, damp and lint-free cloth. Make sure that no moisture enters the housing. Do not use any sprays, solvents, alcohol-based cleaning agents or abrasive cleaners, but only clean water to moisten the cloth.

Repair

Do not modify the device or install any spare parts. For repairs or device testing, contact the manufacturer.
EN 11
layer thickness measuring device BB20

Disposal

The icon with the crossed-out waste bin on waste electrical or electronic equipment stipulates that this equipment must not be disposed of with the household waste at the end of its life. You will find collection points for free return of waste electrical and electronic equipment in your vicinity. The addresses can be obtained from your municipality or local administration. For further return options provided by us please refer to our website https://de.trotec.com/shop/.
The separate collection of waste electrical and electronic equipment aims to enable the re-use, recycling and other forms of recovery of waste equipment as well as to prevent negative effects for the environment and human health caused by the disposal of hazardous substances potentially contained in the equipment.
In the European Union, batteries and accumulators must not be treated as domestic waste, but must be disposed of professionally in accordance with Directive 2006/66/EC of the European Parliament and of the Council of 6September 2006 on batteries and accumulators. Please dispose of batteries and accumulators according to the relevant legal requirements.
12 EN
layer thickness measuring device BB20
Trotec GmbH
Grebbener Str. 7 D-52525 Heinsberg
+49 2452 962-400 +49 2452 962-200
info@trotec.com www.trotec.com
Loading...