Trividia TRUEresult Quick Reference Manual

Page 1
INTRODUCTION Welcome to the TRUEresult Blood Glucose
Monitoring System Our Commitment to You
TRUEresult is a simple, accurate way to test whole blood glucose (sugar) level, anytime, anywhere.
Our goal is to provide quality healthcare products and dedicated customer service. For questions about TRUEresult products, please see cover for phone number.
is booklet contains all the information needed to get the most from the TRUEresult System. To start testing quickly, please see the Quick Reference Guide found inside the cover of this Owner’s Booklet.
Please read complete Owner’s Booklet and all product Instructions for Use before using the System.
1
Use By Date
For in vitro Diagnostic Testing Only
Lot
Number
Authorised
Representative
SYMBOLS:
Keep Dry
Caution
Single Use Only
Attention! Read Instructions for Use.
Sterile
Control
Storage Temperature Range
Manufactured By
Serial Number
Control Level
Biological Risk
Do Not Resterilise
For Assistance Call
Date of Manufacture
Page 2
2
IMPORTANCE OF BLOOD GLUCOSE MONITORING
e more you know about diabetes, the better you will be able to take care of yourself. A Doctor or Healthcare Professional will determine target ranges for blood glucose results and how oen to test. Having most results within the target ranges shows how well a treatment plan is working to control glucose levels. Keeping most results within the target ranges helps slow or stop complications of diabetes.
NEVER change a treatment plan without consulting with a Doctor or Healthcare Professional.
Use of the TRUEresult Blood Glucose Monitoring System in a manner not specied in this Owner’s Booklet is not recommended and may aect your ability to determine true blood glucose levels.
e TRUEresult System is an in vitro (outside body) quantitative system that is used for self-testing and point-of-care testing of only human whole blood. e most accurate results come from using fresh whole blood taken from the ngertip or forearm (capillary) or from the vein (venous).
Page 3
3
What you need to know when using the TRUEresult System:
• Read all product instructions for use before testing.
• Use only TRUEresult Test Strips and TRUEresult Control Solution with TRUEresult Meter.
• To help prevent false high results, wash hands before using the System to test blood, especially aer handling fruit or other foods containing sugar.
• Perform Control Tests before performing a blood glucose test for the rst time (see Control Test).
• Remove only one Test Strip at a time from vial when testing. Recap vial immediately aer re­moving Test Strip.
• NEVER reuse Test Strips. NEVER wipe Test Strips with water, alcohol or any cleaner. DO NOT attempt to clean and re-use Test Strips. Reuse of Test Strips will cause inaccurate results.
• NEVER add a second drop of sample to Test Strip. Adding more sample to the Test Strip gives an error message.
• Venous whole blood collected into sodium or lithium heparin blood collection tubes may be used for testing by Healthcare Professionals. Use of EDTA blood collection tubes is not recommended and may cause low results. Mix tube contents gently before using.
Page 4
4
IMPORTANT HEALTH and SAFETY INFORMATION:
e TRUEresult Blood Glucose Monitoring System is for one person use ONLY. DO NOT share your Meter or your Lancing Device with anyone, including family members. DO NOT use on more than one person. ALL parts of the TRUEresult Blood Glucose Monitoring System could carry blood­borne pathogens aer use, even aer cleaning and disinfection.
2,3
For cleaning and disinfecting the Meter, see Care, Cleaning/Disinfecting. For cleaning and disinfecting the lancing device, see the lancing device’s Instructions for Use.
We suggest cleaning the Meter when visibly dirty or if blood is on the Meter. Wash your hands thoroughly with soap and warm water aer handling the Meter, lancing device, or Test Strips as contact with blood presents an infection risk.
Reuse of devices labeled for single-use may result in product contamination and patient infection.
Page 5
5
•
DO NOT perform capillary blood glucose testing on critically ill patients. Capillary blood
glucose levels in critically ill patients with reduced peripheral blood ow may not reect the true physiological state. Reduced peripheral blood ow may result from the following conditions (for example):
4
~ shock ~ severe hypotension ~ severe dehydration ~ hyperglycaemia with hyperosmolarity,
with or without ketosis.
• Do not use TRUEresult for the diagnosis of diabetes or for testing blood glucose in neonates (newborns).
•
Do not use TRUEresult System during a xylose absorption test. is may falsely raise glucose results.
5
Please check with your Doctor before
using TRUEresult.
Page 6
6
TABLE OF CONTENTS
Phone Number, Fast Test Guide,
Expected Results ........................................... see covers
Introduction and Important Information ...........1
Know the System ..................................................8
Meter .......................................................................8
Test Strip ...............................................................11
Control Solution ..................................................13
Getting Started ....................................................14
Quality Control Testing ......................................15
Automatic Self-Test .............................................15
Control Test .........................................................16
How to Test Control Solution ...........................17
Testing Blood ......................................................20
Obtaining a Blood Sample .................................20
How to Test Blood Glucose ...............................22
Unusual Blood Glucose Test Results ................25
Ketone Test Alert .................................................26
TRUEresult System and Laboratory Testing ...27
Memory................................................................28
View Averages (7-, 14-, and 30-day) ................28
View Results .........................................................29
Page 7
7
Meter Set Up ........................................................30
Set Ketone Test Alert ..........................................32
Set Testing Reminders ........................................32
Care, Cleaning/Disinfecting ...............................34
Meter Care ............................................................34
Control Solution Care .........................................35
Test Strip Care .....................................................35
Changing Battery .................................................36
Troubleshooting ..................................................37
Messages................................................................38
Technical Information ........................................40
Performance Characteristics ..............................40
System Specications ..........................................44
Operating Range ..................................................44
Chemical Composition .......................................44
EMC Safety Information ....................................45
References ............................................................46
Notes ....................................................................47
Page 8
8
KNOW THE SYSTEM
Meter
“ + ” Button
Increase numbers in Time/Date Set Up for Meter
and Testing Reminders; turn on Testing Reminders and Ketone Test Alert; add ALT Symbol; move forward by date/time when viewing results in Memory.
“ S ” Button
Turn Meter on to view Average values, to view
results in Memory, and to access Meter Set Up.
“ - ” Button Decrease numbers in Time/Date Set Up for Meter
and Testing Reminders; turn o Testing Reminders and Ketone Test Alert; remove ALT Symbol; move backward by date/time when viewing results in Memory.
Top of Meter
1 32
Page 9
4
5
Back of Meter
6
Display Screen
Shows test results, messages, user prompts, other
information.
Test Port Insert TRUEresult Test Strip here. Strip Release Button Releases Test Strip aer testing. Battery Door
Use one non-rechargeable 3V lithium battery
(#CR2032), positive (“+”) side up (see Changing Battery).
Meter Label
Contains serial number of Meter. Identies Meter
when calling for assistance.
Docking Station Contacts
Contacts used for uploading Meter results to a
computer using a Docking Station (Please call for availability).
9
2
1
Front of Meter
3
Page 10
10
Meter Display
Full Display Screen
1. Memory Symbol
2. Time, Date
3. Time is AM/PM
(Note: Not seen if factory set to 24 hour clock.)
4.
Result is from 7-, 14-, and 30-day Average
5. Day of Week
6. Test Result
7. Unit of Measure
(Note: Factory set, cannot be changed by user.)
8. Ketone Test Alert Symbol
9.
Testing Reminder Symbol
10. Temperature Symbol
11.
Drop Symbol - Apply blood or Control Solution
12. Alternate Site (ALT) Symbol
13. Battery Symbol
14. Control Symbol
3
4
2
1
7
6
5
8
9
10
11
12
13
14
Page 11
11
Contact End -
Insert into Test Port with blocks
(contacts) facing up
.
Sample Tip -
Touch to top of sample (fresh, capillary or venous blood or Control Solution) aer inserting Contact End into Meter.
Sample Placement
Correct Incorrect
•
Do not touch Sample Tip to drop of sample unless Contact End is inserted into Meter.
• Do not apply blood or Control Solution to top of Test Strip.
• Do not smear or scrape drop with Test Strip.
• Do not apply more sample to the Test Strip aer testing begins.
• Do not insert Sample Tip with sample into Meter for testing. May damage Meter.
Test Strip
12
Top of Test Strip
Page 12
Lot Number ( ) - Used for identication
when calling for assistance.
Use By Dates ( ) - Write date rst opened on
vial label. Discard vial and unused Test Strips if either 4 months aer rst opening or date printed next to on vial label has passed, whichever comes rst.
Use of Test Strips or Control Solution past the Use By Dates may give incorrect test results. Discard out-of-date products and test with new products.
Control Test Range - Range of numbers in
which Control Test result must fall to assure the System is working properly.
12
Test Strip Vial Label
1
2
2
3
May 30, 2014
1
2
2
3
May 30, 2014
or
Page 13
13
Control Solution Bottle Label
Lot Number ( ) - Used for identication
when calling for assistance.
Use By Dates ( ) - Write date rst opened on
bottle label. Discard bottle if either 3 months aer rst opening or date printed next to on bottle label has passed, whichever comes rst.
Control Solution Level (1, 2 or 3) - Testing
at least 2 levels of Control Solution is recommended. Call the number on the cover of this Booklet for information on how to obtain dierent levels of Control Solution.
1
2
3
May 30, 2014
Control Solution
Page 14
GETTING STARTED
e TRUEresult Meter comes with pre-set time, date, Ketone Test Alert and all Testing Reminders o. Before using Meter for the rst time or aer bat tery is changed, check to make sure all of the Meter pre-set functions are correct. Update any setting as needed (see Meter Set Up).
e Meter turns on when a Test Strip is inserted into the Test Port or when “ S ” Button is pressed and released (see Memory and Meter Set Up). Meter turns o when the Test Strip is released from the Meter, “ S ” Button is pressed and held for 20 seconds, or aer 2 minutes of non-use.
Always check your supplies before using.
• Check Meter for damage (cracked Display, missing buttons, etc.). If damage is seen, do not use Meter. Call for assistance.
• Check Test Strip vial for damage (cracked vial, broken vial, etc.). Discard damaged vial and all contents (Test Strips) and use a new vial of Test Strips for testing.
•
Write date rst opened on Test Strip vial. Check Use By Dates (written and printed) before using any Test Strips from the vial. Do not use if 4months aer rst opening (written date) or if printed Use By Date has passed.
• Check Control Solution bottle for any leaks or broken cap. Discard bottle and open a new one for testing.
• Write date rst opened on Control Solution bottle label. Check Use By Dates (written and printed) before using. Do not use if 3 months aer rst opening (written date) or if printed Use By Date has passed.
14
Page 15
15
Quality Control Testing
To assure you are getting accurate and reliable results, TRUEresult oers two kinds of quality control tests. ese tests let you know that your TRUEresult System is working properly and your testing technique is good.
Automatic Self-Test
An Automatic Self-Test ensures that the Meter is working correctly.
1. Wash hands and dry thoroughly.
2. Remove one Test Strip from the
Test Strip vial and insert Test Strip into the Meter.
3.
Meter turns on. e full Display appears and is replaced by the time and the blinking Drop Symbol. e Meter is working correctly and is ready to perform a Control or blood test.
Do not use Meter if:
•
e full Display does not appear (segments are missing),
•
e blinking Drop Symbol does not appear, or
•
An error message appears in the Display.
See Troubleshooting or call for assistance (see cover of this Booklet for phone number).
Full Display
Contacts
Facing Up
Page 16
16
Control Test
e Control Test checks that the System is working correctly and testing technique is good. Use ONLY TRUEresult Control Solution to perform Control Tests. Perform Control Tests:
•
For practice before using the System for the rst time,
•
When opening a new vial of Test Strips,
•
Occasionally as a vial of Test Strips is used,
•
If a Test Strip vial has been le opened or le in extreme heat, cold, or humidity,
•
Whenever a check on performance of the System is needed,
•
If results seem unusually high or low,
•
If Meter damage is suspected (Meter was dropped, crushed, wet, etc.).
Performing Control Tests with more than one level of Control Solution is recommended to ensure that your System is working properly. ree levels of TRUEresult Control Solution are available. Contact place of purchase or use the number on the cover of this Booklet for more information on how to obtain dierent levels of Control Solution.
Ranges printed on Test Strip vial label are for Control Test results only and are not suggested levels for your blood glucose. Do not drink Control Solution.
Page 17
17
How to Test Control Solution
Use ONLY TRUEresult Control Solution with the TRUEresult Meter.
1. Check supplies. See Getting Started.
2. Allow Control Solution, vial of Test
Strips and Meter to adjust to room temperature (20°C-25°C).
Note: Running a Control Test at
temperatures outside the range listed above may cause Control Solution to read as a blood test.
3. Wash hands. Dry thoroughly.
4. Gently swirl or invert Control Solution bottle to mix. DO NOT SHAKE!
5. Remove one Test Strip from vial. Close Test
Strip vial immediately. Use Test Strip quickly aer removal from vial.
Note: If Test Strip has been out of the
vial too long before testing, an error message appears upon insertion of the Test Strip into the Meter. Release and discard old Test Strip. Use new Test Strip for testing.
6. Insert Test Strip into Meter. Meter turns on and shows blinking Drop Symbol and time. Do not remove Test Strip.
Contacts
Facing Up
May 30, 2014
May 30, 2014
Page 18
18
7. With cap removed, turn Control Solution bottle upside down. Gently squeeze one drop of Control Solution onto a clean tissue. Wipe o bottle tip with the tissue.
Note:
If Test Strip is removed before testing is nished, an error message appears. Release and discard old Test Strip. Use new Test Strip for testing.
8.
Gently squeeze a drop of Control Solution onto a small piece of unused aluminum foil or clear plastic wrap. Discard foil or plastic wrap aer use.
Note:
Do not put drop on top of the Test Strip.
9.
With Test Strip still in Meter, touch Sample Tip to top of the drop of Control Solution. Allow drop to be drawn into Test Strip. Remove Test Strip from drop when Meter beeps.
Dashes appear across the Display to show Meter is testing.
Note: If Meter does not beep or begin testing soon
aer drawing up sample, release and discard Test Strip. Repeat test with new Test Strip. If problem persists, see Troubleshooting.
10. Aer testing is nished, the result appears in the display with the Control Symbol.
Drop Symbol
Meter Testing
Page 19
11.
Compare result to Control Test Range printed on Test Strip vial label for Control Solution Level you are using.
If result is in range,
System can be used for testing blood. If result does not fall within range, repeat test using a new Test Strip.
If result is still outside range aer a second Control Test, System should not be used for testing blood. Call for assistance (see cover for phone number).
12.
Aer result is shown, Strip Release Button ashes. Hold Meter with Test Strip pointing down. Press Strip Release Button to release and discard Test Strip in appropriate container. Meter turns o. Recap Control Solution bottle tightly.
Note:
Removing Test Strip before result displays cancels the test. An error message appears and the result is not stored in Memory. Retest with a new Test Strip and do not remove before result is displayed.
19
Control Symbol
Test Strip
Vial Label
Control Solution
Bottle Label
Page 20
20
TESTING BLOOD Obtaining a Blood Sample
Refer to lancing device’s Instructions for Use for detailed instructions.
e lancing device is for single patient use ONLY. For cleaning/disinfecting your lancing device see Lancing Device Care in the lancing device’s Instructions for Use. Wash your hands thoroughly with soap and warm water aer handling the Meter, lancing device, or Test Strips. Contact with blood presents an infection risk.
• NEVER share lancets or lancing device.
• Lancets are for single use only. Do not reuse lancets.
• To help prevent false high results, wash hands before using the System to test blood, especially aer handling fruit or other foods containing sugar.
From Fingertip
1. Prepare ngertip by washing hands in warm, soapy
water. Rinse well. Dry thoroughly.
2. Place end of lancing device equipped with lancet against tip of nger. Lance ngertip.
3. Set lancing device aside. To help blood drop form, lower hand to waist level, gently massage nger from palm to ngertip. Allow blood drop to form before attempting to apply to Test Strip.
4. Recap and remove used lancet from lancing device. Discard used lancet into appropriate container.
Note: Treat used Test Strips and lancets as a biological
risk. Dispose used Test Strips and lancets in approved container.
Page 21
21
From Forearm
Note: Some lancing devices include a special end cap
for alternate site testing. Check lancing device’s Instructions for Use.
1. Select area to be lanced. Wash with soap and warm water,
rinse and dry thoroughly.
2. Rub area vigorously or apply a warm dry compress to
increase blood ow.
3. Place end of lancing device equipped with lancet
rmly against forearm. Press trigger button. Apply rm pressure on lancing device for 10 seconds. Allow blood drop to form before attempting to apply to the Test Strip.
4. Recap and remove lancet from lancing device. Discard
used lancet into appropriate container.
Important Notes Regarding Forearm Testing
6
• Check with your Doctor or Healthcare Professional to see
if forearm testing is right for you.
• Results from forearm are not always the same as results
from ngertip.
• Use ngertip instead of forearm for more accurate results:
- Within 2 hours of eating, exercise, or taking insulin,
- If your blood sugar may be rising or falling rapidly
or your routine results are oen uctuating,
- If you are ill or under stress,
- If your forearm test results do not match how you feel,
- If your blood sugar may be low or high,
- If you do not notice symptoms when blood sugar
is low or high.
Page 22
22
How to Test Blood Glucose
1. Check supplies. See Getting Started.
2. Wash hands (and forearm for alternate site
testing). Rinse well and dry thoroughly.
3. Remove one Test Strip from vial. Close vial immediately. Use Test Strips quickly aer removal from vial.
4. With Meter o, insert Test Strip Contact End (blocks facing up) into Meter. Meter turns on and shows the Drop Symbol in the Display. Keep Test Strip in Meter until testing is nished.
---Alternate Site Testing---
To mark test as alternate site (forearm) result,
press “+” Button. ALT Symbol appears in the Display. Press “-” Button to remove ALT Symbol.
Note: If Test Strip has been out of the vial too long
before testing, an error message appears upon insertion of the Test Strip into the Meter. Release and discard old Test Strip. Use new Test Strip for testing.
Page 23
5. Lance ngertip or forearm. Allow drop to form
(see Obtaining a Blood Sample).
6. With Test Strip still in Meter, touch Sample Tip to top of blood drop and allow blood to be drawn into Test Strip. Remove from blood drop immediately aer the Meter beeps and dashes appear across Meter Display.
Note: If Meter does not beep or begin testing soon
aer touching blood drop to Sample Tip, discard Test Strip. Repeat test with new Test Strip and new blood drop. If problem persists, see Troubleshooting.
7. Aer the test is nished, result is displayed with date and time. e Strip Release Button ashes. Record result in log book.
23
Page 24
24
8. Hold Meter with Test Strip pointing down. Press Strip Release Button to discard Test Strip in the appropriate container.
Note: Removing Test Strip before result displays
cancels the test. An error message appears and result is not stored in Memory. Retest with a new Test Strip and do not remove before result is displayed.
9. Meter turns o. Result is stored in Memory with day, date and time.
Note: Treat used Test Strips and lancets as a
biological risk. Dispose used Test Strips and lancets in approved container.
Page 25
25
Unusual Blood Glucose Test Results
If you have symptoms of low or high glucose, check your blood glucose immediately. If your result does not match the way you feel, repeat test. If your results still do not match the way you feel, call your Doctor or Healthcare Professional.
• Low blood glucose (hypoglycaemia) symptoms may be trembling, sweating, intense hunger, nervousness, weakness, and trouble speaking.
•
High blood glucose (hyperglycaemia) symptoms may be intense thirst, a need to urinate oen, a dry mouth, vomiting and headache.
Meter reads blood glucose levels from
1.1-33.3 mmol/L (20-600 mg/dL).
If blood test result is less than 1.1 mmol/L (20 mg/dL), “Lo” appears in Meter Display. If blood test result is greater than 33.3 mmol/L (600 mg/dL), “Hi” appears in Meter Display. ALWAYS repeat test to conrm Low (“Lo”) and High (“Hi”) results. If results still display “Lo” or “Hi”, call your Doctor or Healthcare Professional
immediately. Note: “Lo” results are included in the Average as
1.1 mmol/L (20 mg/dL). “Hi” results are included as 33.3 mmol/L (600 mg/dL).
Page 26
26
Ketone Test Alert
If blood glucose test result is greater than 13.3 mmol/L (240 mg/dL), and Ketone Test Alert is turned on, “Ketone” appears in Display with glucose result (see Set Ketone Test Alert).
When the Ketone Alert Symbol is shown,
it does not mean that ketones have been detected in blood. Perform a ketone test per the health plan as prescribed by the Doctor or Healthcare Professional.
Ketone Test
Alert
Page 27
TRUEresult System and Laboratory Testing
e most accurate results come from using fresh, capillary whole blood from the ngertip or forearm. Venous whole blood collected into sodium or lithium heparin blood collection tubes may be used for testing by Healthcare Professionals. Use of EDTA blood collection tubes is not recommended and may cause low results. Mix tube contents gently before using.
When comparing results between TRUEresult and a laboratory system, perform a TRUEresult blood test within 30 minutes of laboratory test. Diabetes experts have suggested that 95% of glucose meter results should agree within 0.83 mmol/L (15 mg/dL) of a laboratory method when the glucose concentration is less than 5.55 mmol/L (100 mg/dL), and within 15% of a laboratory method when the glucose concentration is 5.55 mmol/L (100 mg/dL) or higher.
7
If the patient has recently eaten, ngerstick results from the TRUEresult System can be up to 3.9 mmol/L (70 mg/dL) higher than venous laboratory results.
8
27
Page 28
28
MEMORY View Averages (7-, 14-, and 30-Day)
e Averages feature allows you to view the average of all blood glucose results within a 7-, 14-, and 30-day period. Results marked as Control Test results are not included in the Averages.
Note:
If a Control Test is performed outside the recommended testing temperature (see How to Test Control Solution), the Control Solution may read as a blood test and be included in the Averages.
Start with Meter o. Press and release “ S ” Button. Display scrolls through 7-, 14-, and 30-Day Average values.
Note: If there are no Average
values, three dashes are displayed for 7-, 14-, and 30-day Averages.
No Average
7-Day Average
14-Day Average
30-Day Average
Page 29
29
Ketone Test
Alert
Control Symbol
Alternate Site
Symbol
View Results
Memory stores 500 total blood and Control Test results, which are displayed from most recent to oldest. When the Memory is full, the oldest result is replaced with the newest result.
1.
Press and release “ S ” Button. Meter displays 7-, 14-, and 30-day Averages.
2. Press and release “ S ” Button again to view most
recent result in Memory. If there are no results in Memory, dashes appear with the Memory Symbol.
3.
Press “ + ” Button and release to advance to the next result. Press “ + ” Button to scroll forward through blood results or “ - ” Button to scroll backwards through blood results.
• Test results marked as
alternate site display ALT Symbol.
• Control Test results display the
Control Symbol. If no Control Test has been done Display shows dashes and the Control Symbol.
• Test results above 13.3 mmol/L
(240 mg/dL) display Ketone Test Alert Symbol, when Ketone Test Alert is turned on during Meter Set Up.
Page 30
METER SET UP
Meter Set Up is to be used if changes need to be made to the pre-set time and date, to turn on the Ketone Test Alert and Testing Reminders, or if settings need to be reset because of battery change.
Note: Meter turns o aer 2 minutes of non-use.
If Meter turns o at any time during Set Up, all settings are saved. e Meter takes you back to Step #1.
1. Press and hold “ S ” Button until
the full Display is shown and a tone sounds (around 5 seconds). Release “ S ” Button.
2. e hour ashes. Press the “ + ”
Button to scroll up the hour or the “ - ” Button to scroll down the hour. Release “ + ” or “ - ” Button when the correct hour displays.
Like many alarm clocks, to set the
hours for “AM” or “PM”, the hours must scroll until the correct hour with “AM” or “PM” displays. Press “ S ” Button and release to save the hour and go to the next step.
Note: Clock is factory set for either 12 hour or 24
hour clock. is feature cannot be changed.
30
Page 31
3.
e minutes ash. Slowly press the “ + ” Button to scroll minutes up or the “ - ” Button to scroll minutes down. Release “ + ” or “ - ” Button when the correct minutes are displayed. Press “ S “ Button and release to save the minutes and go
to the next step.
4.
e month (number) ashes. Slowly press and release the “ + ” Button to increase the month and the “ - ” Button to de crease the month. Press and release “ S ” Button to save the month and go to the next step.
5. Slowly press and release the “ + ”
Button to increase the day and the “ - ” Button to de crease the day. Press and release ” S ” Button to save the day and go to the next step.
6.
Slowly press and release the “ + ” Button to increase the year or the “ - ” Button to decrease the year. Press and release the “ S ” Button to save the year.
Note: Day of the week (Mon, Tue,
Wed, u, Fri, Sat, Sun) self­adjusts when month, day or year are changed.
Note:
Meter beeps every time a setting is conrmed (“ S ” Button is pressed).
31
Page 32
Set Ketone Test Alert
7. Aer setting year, press “ + ” or
“ - ” Button to turn Alert on or o. Press “ S ” Button to set and go to the Testing Reminders.
Set Testing Reminders
Up to four Testing Reminders per day may be set. Reminder sounds at set time for 10 seconds. To set the Testing Reminders:
8.
Aer pressing “ S ” Button to set Ketone Test Alert, Display shows rst Reminder setting (A-1). Press “ + ” Button (on) or “ - ” Button (o) to turn Reminder on or o.
9. If “on” is chosen, press “ S ” Button.
e hour ashes. Press the “ + ” Button to scroll up the hour or the “ - ” Button to scroll down the hour. Release “ + ” or “ - ” Button when the correct hour displays. Like many alarm clocks, to set the hours for “AM” or “PM”, the hours must scroll until the correct hour with “AM” or “PM” displays. Press “ S ” Button and release to save the hour and go to the minutes set up.
If “o” is chosen and the “ S ” Button is
pressed, the Meter goes to the next Testing Reminder.
32
Ketone Test
Alert
Page 33
10. e minutes ash. Press “ + ”
Button or “ - ” Button to set the minutes. Press “ S ” Button to set.
11. Turn Reminders on and repeat setting the time
for next 3 Reminders (if needed).
12. Press and hold “ S ” Button when
nished to turn o Meter. Meter turns o aer 2 minutes of non-use.
Note: If Testing Reminders
are set, the Testing Reminder Symbol appears in all Displays.
33
Testing Reminder Symbol
Page 34
34
CARE, CLEANING/DISINFECTING Caring for TRUEresult
• Store System (Meter, Control Solution, Test
Strips) in Carrying Case to protect from liquids, dust and dirt.
• Store in a dry place at
2°C-30°C (room temperature).
• DO NOT REFRIGERATE
OR FREEZE.
Meter Care
• Do not keep meter in an area where it may be
crushed (i.e. back pocket, drawer, bottom of bag, etc.).
Cleaning removes blood and soil, disinfecting
removes infectious agents.
To Clean the Meter:
• Clean the Meter when visibly dirty or if blood is
on the Meter.
• Never put Meter in liquids or allow any liquids to
enter the Test Port.
• Wipe Meter with a clean, lint-free cloth
dampened with 70% Isopropyl alcohol.
• Let Meter air dry thoroughly before using to test.
• Do not use bleach to clean the Meter. For assistance call the phone number on the cover of
the Owner’s Booklet.
2°C
30°C
Page 35
35
Control Solution Care
• Write date opened on Control Solution label.
Discard if either 3 months aer rst opening or aer date printed next to on bottle label has passed, whichever comes rst.
• Store at 2°C-30°C (room temperature). DO NOT REFRIGERATE
OR FREEZE.
• Aer use, wipe bottle tip clean and recap tightly.
Test Strip Care
• Store Test Strips in original vial only. Do not
transfer old Test Strips to new vial or store Test Strips outside of vial.
•
Write date opened on Test Strip vial. Discard unused Test Strips from vial if either 4 months aer rst opening or aer date printed next to on vial label has passed, whichever comes rst. Use of Test Strips past either date may give incorrect results.
• Close vial immediately aer removing Test Strip.
• Store in a dry place below
30°C (room temperature).
DO NOT REFRIGERATE OR FREEZE.
• Do not reuse Test Strip. NEVER wipe Test Strips
with water, alcohol or any cleaner. DO NOT attempt to clean and re-use Test Strips. Reuse of Test Strips will cause inaccurate results.
• Do not bend, cut or alter Test Strips in any way.
2°C
30°C
1°C
30°C
Page 36
Battery
Door
Tab
36
Changing Battery
A low battery displays Battery Symbol while continuing to function. A dead battery displays Battery Symbol, beeps, and then turns o.
To replace battery:
1. Li tab on Battery Door.
2. Turn Meter over, tap gently on
the palm of your other hand to loosen and remove battery.
3. Discard old battery into
appropriate container.
4. Insert new battery, positive (“+”)
facing up. Close Battery Door.
Note: Use non-rechargeable 3V lithium battery
(#CR2032).
5.
Press “ S ” Button to turn Meter on and check time, date, and Testing Alerts and Reminders (see Meter Set Up). If Meter does not turn on, check that battery was installed properly. If not, remove and reinsert battery and turn Meter on by pressing “ S ” Button. Call for assistance if problem persists.
Low
Batteries might explode if mishandled or incorrectly replaced. Do not dispose of battery in re. Do not take apart or attempt to recharge battery. Dispose according to local/country specic regulations.
Page 37
37
Action
Remove Test Strip. Re-insert correctly. Remove Test Strip. Re-insert Test Strip fully into Meter. Repeat with new Test Strip. Replace battery. Battery positive (“+”) side must face up. Call for assistance.
Action
Repeat test with new Test Strip and larger drop. Repeat test with new Test Strip. Apply sample within 2 minutes of inserting Test Strip. Repeat with new Test Strip. Call for assistance.
Reason
Test Strip inserted upside down or backwards Test Strip not fully inserted Test Strip Error Dead or no battery Battery in backwards
Meter Error
Reason
Sample drop too small
Sample applied aer two minute shut-o
Problem with Test Strip Problem with Meter
TROUBLESHOOTING
1) Aer inserting Test Strip, Meter does not turn on.
2) Aer applying sample, test does not start/ Meter does not beep or begin testing.
For assistance, see cover for phone number.
Page 38
38
Messages
Display
Reason
Temperature
Error
•
Too Cold/
Too Hot
Sample Not Detected or
Using Wrong
Test Strip
Used Test
Strip
or Test Strip
Outside of
Vial Too Long
Meter
Error
Test Strip Error,
Very high blood
glucose result
(higher than
33.3 mmol/L (600 mg/dL)
Action
Move Meter and Test
Strips to area between
10°C-40°C; wait 10
minutes for System to
reach room temperature
before testing.
Retest with new
Test Strip and larger sample.
Repeat with new Test
Strip. Make sure sample
is touched to edge of Test
Strip (not top). If error
persists, call for assistance.
Call for
assistance.
Retest with new Test Strip. If error
persists, call for assistance. If you
have symptoms such as fatigue, excess
urination, thirst, or blurry vision
follow your healthcare professional’s
advice for high blood glucose.
Page 39
39
If error message still appears, any other error message appears, or troubleshooting does not solve the problem, call for assistance.
Display Action
Retest with new Test
Strip. Make sure result
is displayed before
removing Test Strip.
Result was not recorded
in Memory. Retest with
a new Test Strip. If error
persists, call for assistance.
Call for assistance.
Low: About 50 tests can be
done before battery dies.
Dead: Battery Symbol appears and
beep sounds before Meter turns o.
Change the battery.
Retest with new Test Strip.
If result is still “Hi”
(High) or “Lo” (Low)
contact Doctor
immediately.
Messages (continued)
Reason
Test Strip
Removed
During Test
Memory
Error
Communication
Error
Low or
Dead
Battery
Out of Range
High Results
> 33.3 mmol/L
(600 mg/dL)
Low Results
< 1.1 mmol/L
(20 mg/dL)
Page 40
40
Performance Characteristics
5
Precision: Precision describes the variation between results. ere are
two types of precision results measured, repeatability (using blood) and intermediate precision (using control solution). Repeatability: N=100 Mean (mmol/L) 3.4 5.1 8.5 12.1 19.2 Mean (mg/dL) 62 91 153 218 345 SD (mmol/L) 0.11 0.14 0.27 0.37 0.66 SD (mg/dL) 2.0 2.6 4.8 6.6 11.9 %CV 3.2 2.9 3.1 3.0 3.4 Intermediate Precision: N=100 Mean (mmol/L) 3.1 7.6 19.5 Mean (mg/dL) 56 136 351 SD (mmol/L) 0.09 0.24 0.69 SD (mg/dL) 1.7 4.3 12.4 %CV 3.1 3.2 3.5 System Accuracy: Diabetes experts have suggested that 95% of glucose meter results should agree within + 0.83 mmol/L (15 mg/dL) of the medical laboratory values at glucose concentrations below 5.55 mmol/L (100 mg/dL) and within + 15% of the medical laboratory values at glucose concentrations at or above 5.55 mmol/L (100 mg/dL). e tables below show how oen healthcare professionals (HCP) and users achieve these goals using capillary fingertip and forearm blood samples when glucose results are not fluctuating. e laboratory reference instrument is the Yellow Springs Instrument (YSI) Model 2300.
Page 41
41
For Healthcare Professionals
99.3% of TRUEresult fingertip values performed by healthcare professionals (HCP) fell within
0.83 mmol/L (15 mg/dL) of the YSI results at glucose levels < 5.55 mmol/L (100 mg/dL) and within 15% at glucose levels > 5.55 mmol/L (100 mg/dL).
Fingertip Samples (HCP vs. YSI) for glucose concentrations < 5.55 mmol/L (100 mg/dL)
Within
+ 0.28 mmol/L (5 mg/dL)
Within
+ 0.56 mmol/L (10 mg/dL)
Within
+ 0.83 mmol/L (15 mg/dL)
74/180 (41%) 155/180 (86%) 178/180 (98.8%)
Fingertip Samples (HCP vs. YSI) for glucose concentrations > 5.55 mmol/L (100 mg/dL)
Within + 5% Within + 10% Within + 15%
272/420 (65%) 395/420 (94%) 418/420 (99.5%)
Fingertip Samples for glucose concentrations between 1.1-33.3 mmol/L (20-600 mg/dL)
Within + 0.83 mmol/L (15 mg/dL) or + 15%
596/600 (99.3%)
Forearm Samples (HCP vs. YSI) for glucose concentrations < 5.55 mmol/L (100 mg/dL)
Within
+ 0.28 mmol/L (5 mg/dL)
Within
+ 0.56 mmol/L (10 mg/dL)
Within
+ 0.83 mmol/L (15 mg/dL)
18/28 (64%) 26/28 (93%) 28/28 (100%)
Forearm Samples (HCP vs. YSI) for glucose concentrations > 5.55 mmol/L (100 mg/dL)
Within + 5% Within + 10% Within + 15%
50/72 (69%) 64/72 (89%) 72/72 (100%)
Parkes Error Grid: 100% of individual fingertip glucose measured values performed by healthcare professionals fell within Zone A of the Parkes Error Grid (PEG).
100% of TRUEresult forearm values performed by healthcare professionals (HCP) fell within
0.83 mmol/L (15 mg/dL) of the YSI results at glucose levels < 5.55 mmol/L (100 mg/dL) and within 15% at glucose levels > 5.55 mmol/L (100 mg/dL).
Forearm Samples for glucose concentrations between 1.1-33.3 mmol/L (20-600 mg/dL)
Within + 0.83 mmol/L (15 mg/dL) or + 15%
100/100 (100%)
Parkes Error Grid: 100% of individual forearm glucose measured values performed by healthcare professionals fell within Zone A of the Parkes Error Grid (PEG).
Page 42
42
Venous Samples (HCP vs. YSI) for glucose concentrations < 5.55 mmol/L (100 mg/dL)
Within
+ 0.28 mmol/L (5 mg/dL)
Within
+ 0.56 mmol/L (10 mg/dL)
Within
+ 0.83 mmol/L (15 mg/dL)
10/28 (35.7%) 18/28 (64.3%) 28/28 (100%)
Venous Samples (HCP vs. YSI) for glucose concentrations > 5.55 mmol/L (100 mg/dL)
Within + 5% Within + 10% Within + 15%
126/214 (58.9%) 191/214 (89.3%) 213/214 (99.5%)
Venous Blood
99.6% of TRUEresult venous values performed by healthcare professionals (HCP) fell within
0.83 mmol/L (15 mg/dL) of the YSI results at glucose levels < 5.55 mmol/L (100 mg/dL) and within 15% at glucose levels > 5.55 mmol/L (100 mg/dL).
Venous Samples for glucose concentrations between 1.1-33.3 mmol/L (20-600 mg/dL)
Within + 0.83 mmol/L (15 mg/dL) or + 15%
241/242 (99.6%)
Parkes Error Grid: 99.6% of individual venous glucose measured values performed by healthcare professionals fell within Zone A and 0.4% fell within Zone B of the Parkes Error Grid (PEG).
For Consumers
99% of TRUEresult fingertip values performed by users fell within 0.83 mmol/L (15 mg/dL) of the YSI results at glucose levels < 5.55 mmol/L (100 mg/dL) and within 15% at glucose levels > 5.55 mmol/L (100 mg/dL).
Fingertip Samples (User vs. YSI) for glucose concentrations < 5.55 mmol/L (100 mg/dL)
Within
+ 0.28 mmol/L (5 mg/dL)
Within
+ 0.56 mmol/L (10 mg/dL)
Within
+ 0.83 mmol/L (15 mg/dL)
11/23 (48%) 21/23 (91%) 23/23 (100%)
Fingertip Samples (User vs. YSI) for glucose concentrations > 5.55 mmol/L (100 mg/dL)
Within + 5% Within + 10% Within + 15%
38/77 (49%) 66/77 (86%) 76/77 (99%)
Fingertip Samples for glucose concentrations between 1.1-33.3 mmol/L (20-600 mg/dL)
Within + 0.83 mmol/L (15 mg/dL) or + 15%
99/100 (99%)
Parkes Error Grid: 100% of individual fingertip glucose measured values performed by users fell within Zone A of the Parkes Error Grid (PEG).
Page 43
43
Forearm Samples (User vs. YSI) for glucose concentrations < 5.55 mmol/L (100 mg/dL)
Within
+ 0.28 mmol/L (5 mg/dL)
Within
+ 0.56 mmol/L (10 mg/dL)
Within
+ 0.83 mmol/L (15 mg/dL)
16/28 (57%) 23/28 (82%) 28/28 (100%)
Forearm Samples (User vs. YSI) for glucose concentrations > 5.55 mmol/L (100 mg/dL)
Within + 5% Within + 10% Within + 15%
42/72 (58%) 61/72 (85%) 72/72 (100%)
100% of TRUEresult forearm values performed by users fell within 0.83 mmol/L (15 mg/dL) of the YSI results at glucose levels < 5.55 mmol/L (100 mg/dL) and within 15% at glucose levels > 5.55 mmol/L (100 mg/dL).
Forearm Samples for glucose concentrations between 1.1-33.3 mmol/L (20-600 mg/dL)
Within + 0.83 mmol/L (15 mg/dL) or + 15%
100/100 (100%)
Parkes Error Grid: 100% of individual forearm glucose measured values performed by users fell within Zone A of the Parkes Error Grid (PEG).
User Performance Evaluation: A study evaluating glucose values from fingertip capillary blood samples obtained by 100 lay persons showed the following results: 100% within + 0.83 mmol/L (15 mg/dL) of the medical laboratory values at glucose concentrations below 5.55 mmol/L (100 mg/dL) and 99% within + 15% of the medical laboratory values at glucose concentrations at or above 5.55 mmol/L (100 mg/dL).
Page 44
44
SYSTEM SPECIFICATIONS
Result Range: 1.1-33.3 mmol/L (20-600 mg/dL) Sample Size: Minimum 0.5 microlitre (0.5 μL) Sample:
Fresh capillary whole blood, venous blood drawn in sodium or lithium heparin blood collection tubes, or Control Solution.
Test Time: Results in as little as 4 seconds Result Value: Plasma values Assay Method: Electrochemical Power Supply: One 3V lithium battery
#CR2032 (non-rechargeable)
Battery Life: Approximately 2146 tests or 1.5 years Automatic shut-o: Aer two minutes of non-use Weight: 47 grams Size: 89 x 55 x 17 mm Memory Size: 500 glucose results
Operating Range (Meter & Test Strips):
Relative Humidity: 10-90% (Non-condensing) Temperature: 10°C-40°C Haematocrit: 20-55% Altitude: 3094 metres
Note: Use within specied environmental conditions only.
Chemical Composition
TRUEresult Test Strips: Glucose dehydrogenase-FAD
(Aspergillus sp.), mediators, buers and stabilisers.
TRUEresult Control Solution: Water, d-glucose, buers, viscosity enhancing agent, salts, dye and preservatives.
Page 45
45
EMC Safety Information
is meter meets the electromagnetic immunity requirements as per ISO 15197:2013 Annex A. It meets the electromagnetic emissions requirements as per EN 61326 series. Interference from the meter to other electronically driven equipment is not anticipated. e electromagnetic environment should be evaluated prior to operation of the device.
Do not use the meter in a very dry environment, especially one in which synthetic materials are present. Do not use the meter close to sources of strong electromagnetic radiation, as these may interfere with the proper operation.
Page 46
REFERENCES
1. Joslin Diabetes Center. Goals for Blood Glucose Control
[Electronic Version]. Retrieved February 16, 2012 from http://www.joslin.org/info/Goals-for-Blood-Glucose­Control.html.
2. FDA Public Health Notication: Use of Fingerstick Devices
on More than One Person Poses Risk for Transmitting Blood Borne Pathogens: Initial Communication Update
11/29/2010 [Electronic Version]. Retrieved February 22, 2012 from http://www.fda.gov/MedicalDevices/Safety/ AlertsandNotices/ucm224025.htm.
3. CDC Clinical Reminder: Use of Fingerstick Devices
on More than one Person Poses Risk for Transmitting Bloodborne Pathogens [Electronic Version]. Retrieved
February 22, 2012 from http://www.cdc.gov/injectionsafety/Fingerstick­DevicesBGM.html.
4. Atkin, S.H., et. al. Fingerstick Glucose Determination in
Shock. Annals of Internal Medicine, 114:1020-1024 (1991).
5. Data on le.
6. U.S. Food and Drug Administration. Blood Glucose
Meters, Getting the Most Out of Your Meter. [Electronic Version]. Retrieved July 6, 2009: www. fda.gov/MedicalDevices/Safety/ AlertsandNotices/ TipsandArticlesonDeviceSafety/ucm109371.htm.
7. International Organization for Standardization. In vitro
diagnostic test systems. Requirements for blood-glucose monitoring system for self-testing in managing diabetes mellitus. Reference number ISO 15197:2013 (E). Geneva:
International Organization for Standardization; 2013.
8. Larsson-Cohn U: Dierence between capillary and venous
blood glucose during oral glucose tolerance tests. Scand J Clin Lab Invest 36:805-808, 1976.
46
Page 47
47
NOTES:
Page 48
48
E4INP03FDA Rev. 30
NOTES:
Loading...