Tripp Lite OMNISMART350HG, SMART1200XLHG User Manual

Owner’s Manual
1111 W. 35th Street Chicago, IL 60609 USA
Customer Support: (773) 869-1234 • www.tripplite.com
Important Safety Instructions
2
Quick Installation
3
Basic Operation
5
Storage & Service
Specifications
Copyright © 2003 Tripp Lite. All rights reserved.
SmartPro
®
is a registered trademark of Tripp Lite. OmniSmart™is a trademark of Tripp Lite.
Español
Français
SmartPro®& OmniSmart
Medical Grade UPS Systems
• Full Isolation (Less than 300µA Leakage)
• Hospital-Grade Plug & Receptacles
• Line-Interactive Operation
• 120V Input
• 330 - 1000VA Output Capacities
OmniSmart
UPS Systems with Hospital-Grade Plugs & Receptacles
• Hospital-Grade Plug & Receptacles
• Line-Interactive Operation
• 120V Input
• 450 - 1400VA Output Capacities
UL1778
UL1778
2601-1
Register on-line today for a chance to win a FREE Tripp Lite product!
www.tripplite.com/warranty
2
UPS Location Warnings
• Your UPS incorporates fusing only in the ungrounded phase conductor. UL2601-1 certified UPS systems (see Specifications) must not be used in countries other than the United States and Canada and must be used only in health care facilities on grounded systems where conditions of maintenance and supervision ensure that only qualified persons will service the electrical distribution system.
• Install your UPS indoors, away from excess moisture or heat, dust or direct sunlight and conductive contaminants.
• Your UPS is not suitable for use in the presence of a flammable anaesthetic mixture with air, oxygen or nitrous oxide.
• For best performance, keep the indoor temperature between between 32º F and 104º F (0º C and 40º C).
• Leave adequate space around all sides of the UPS for proper ventilation.
UPS Connection Warnings
• Connect your UPS directly to a properly grounded AC power outlet. Do not plug the UPS into itself; this will damage the UPS.
• Do not modify the UPS's plug, and do not use an adapter that would eliminate the UPS’s ground connection.
• Do not use extension cords to connect the UPS to an AC outlet. Your warranty will be voided if anything other than Tripp Lite surge suppressors are used to connect your UPS to an outlet.
• If the UPS receives power from a motor-powered AC generator, the generator must provide clean, filtered, computer-grade output.
Equipment Connection Warnings
• Do not use Tripp Lite UPS Systems for life support devices in which a malfunction or failure of a Tripp Lite UPS System could cause failure or significantly alter the performance of a life-support device.
• Do not connect surge suppressors or extension cords to the output of your UPS. This may damage the UPS and will void the surge suppressor and UPS warranties.
Battery Warnings
• Your UPS does not require routine maintenance. Do not open your UPS for any reason. There are no user-serviceable parts inside.
• Because the batteries present a risk of electrical shock and burn from high short-circuit current, observe proper precautions. Unplug and turn off the UPS before performing battery replacement. Use tools with insulated handles, and replace the existing batteries with the same number and type of new batteries (Sealed Lead-Acid). Do not open the batteries. Do not short or bridge the battery terminals with any object.
• Tripp Lite offers a complete line of UPS System Replacement Battery Cartridges (R.B.C.). Visit Tripp Lite on the Web at www.tripplite.com/support/battery/index.cfm to locate the specific replacement battery for your UPS.
• The UPS batteries are recyclable. Refer to local codes for disposal requirements, or in the USA only call 1-800-SAV-LEAD or 1-800-8-BATTERY (1-800-8-228-8379) or visit www.rbrc.com for recycling information. Do not dispose of the batteries in a fire.
• If your UPS is equipped with an external battery connector, only connect Tripp Lite external battery packs of the appropriate voltage.
• If your UPS is not equipped with an external battery connector, do not attempt to add external
Important Safety Instructions
SAVE THESE INSTRUCTIONS
This manual contains instructions and warnings that should be followed during the installation, operation and storage of all Tripp Lite UPS Systems. Failure to heed these warnings will void your warranty.
3
Quick Installation
Note! Although some installation procedures are common to all models, some procedures vary depending on whether you are installing a SmartPro model or an OmniSmart model.
For All Models—Plug your UPS into a 3-wire grounded, 120V AC, 60 Hz utility outlet.
Note: UL2601-1 listed UPS systems only meet UL standard 2601-1 for medical equipment when plugged into a hospital-grade outlet.
For SmartPro Models Only:
After you plug the UPS into a live AC outlet, the UPS will automatically charge its batteries, but will not supply power to its outlets until it is turned ON (see Step 3 below). The BATTERY CHARGE LED will be the only LED illuminated.
For OmniSmart Models Only:
Your UPS will run a self-test after it is plugged in. See Basic Operation to understand the results of its self-test.
For All Models—Plug your equipment into the UPS.*
* Your UPS is designed to only support computer equipment. You will overload the UPS if the total VA ratings for all the equipment you connect exceeds the UPS’s Output Capacity (see Specifications). To find your equipment’s VA ratings, look on their nameplates. If the equipment is listed in amps, multiply the number of amps by 120 to determine VA. (Example: 1 amp × 120 = 120 VA). If you are unsure if you have overloaded the UPS’s outlets, see “OUTPUT LOAD LEVEL” LED description.
For SmartPro Models Only—Turn the UPS ON.
Press and hold the “POWER” button for one second. The alarm will beep once briefly after one second has passed. Release the button.
For OmniSmart Models Only— Select UPS Operating Mode.*
Press the UPS/STANDBY button to toggle your UPS between the UPS (“POWER” LED lit) and the STANDBY (“POWER” LED flashing) modes. Choose the operating mode based on your location:
USA, Canada & Western Europe:
• Leave the UPS in the UPS mode at all times. All Other Countries:
• Put the UPS in the STANDBY mode when you are not using connected equipment.
(WARNING! When set to “CHARGE ONLY,” the UPS will not
provide battery backup during a blackout or brownout.)
• Put the UPS in the UPS mode when you are using connected equipment.
*See Basic Operation section for a complete explanation of
each mode.
3
2
1
1
2
3
4
These connections are optional. Your UPS will function properly without these connections.
Serial Communications
(Select models only)
Using the cable provided, connect the DB9 port of your computer to the DB9 port of your UPS. Download the PowerAlert UPS monitoring software program appropriate for your operating system from www.tripplite.com and install it on your computer.
USB Communications
(Select Models Only)
Using the cable provided, connect the USB port of your computer to the USB port of your UPS. Download the PowerAlert UPS monitoring software program appropriate for your operating system from www.tripplite.com and install it on your computer.
Tel/DSL Protection
(Select Models Only)
Using telephone cords, connect your wall jack to the UPS jack marked “LINE” (or “IN”) and your equipment to the UPS jack marked “EQUIP” (or “OUT”). This will protect your equipment from surges over the phone line, but you should make sure that your equipment is also protected against surges on the AC line.
UPS models with tel/DSL protection can protect two lines on one cable simultaneously. If you want to protect two lines that travel through separate cables, you must use “Y” splitters appropriate to your application to adapt them to your UPS's phone jacks.
External Batteries
(Select Models Only)
External batteries are only needed to extend run­time. Adding external batteries will increase recharge time as well as runtime.
The illustration shows the location of your UPS’s External Battery Connector, where you will insert the battery pack cable. Complete installation and mounting instructions for your battery pack appear in the battery pack’s owner’s manual. Make sure that cables are fully inserted into their connectors. Small sparks may result during battery connection; this is normal.
4
3
2
1
Quick Installation
optional
1
2
3
4
IN OUT
SEE MANUAL FOR
PROPER CONNECTION
NORM
55
Note! Although some operational features and functions are common to all models, some operational features vary depending on whether you are operating a SmartPro model or an OmniSmart model.
For SmartPro Models Only—“POWER” Button:
• To turn the UPS ON: with the UPS plugged into a live AC wall outlet,* press and hold the POWER button for about two seconds.** Release the button. If utility power is absent, you can “cold-start” the UPS (i.e.: turn it ON and supply power for a limited time from its batteries***) by pressing and holding the POWER button for about two seconds.**
• To turn the UPS OFF: with the UPS ON and receiving utility power, press and hold the POWER button for one second.** Then unplug the UPS from the wall outlet. The UPS will be completely OFF.
*After you plug the UPS into a live AC outlet, the UPS will automatically charge its batteries, but will not supply power to its outlets until it is turned ON. ** The alarm will beep once briefly after the indicated interval has passed. *** If fully charged.
For OmniSmart Models Only—“UPS/STANDBY” Button:
Use the UPS/STANDBY button to do three things:
Switch your UPS’s Operating Mode: While your UPS is plugged into a
live AC outlet, press the UPS/STANDBY button and hold it until you hear a beep (about 2 seconds) to toggle between the following operating modes. Choose your UPS’s operating mode based on the regional guidelines in Step 3 of the Quick Installation section.
• UPS Mode: ENABLES battery backup. UPS Conditions: While receiv­ing adequate utility line power, the UPS supplies power to its AC recep­tacles and charges its batteries; its “POWER” indicator light will be lit. If the utility line power fails or becomes inadequate, the UPS will supply inverted AC power to its receptacles if its batteries are sufficiently charged. Setting Advantages: Provides battery backup during blackouts or brownouts.
• STANDBY Mode: DISABLES battery backup. UPS Conditions: While receiving adequate utility line power, the UPS supplies power to its AC receptacles and charges its batteries; the “POWER” indicator light will be flashing. Setting Advantages: Continues to charge the battery when power is present while turning OFF the inverter to prevent battery deple­tion during power outages when equipment is not in use.
Cold-Start Your UPS: You may “cold start” your UPS and use it as a stand-alone power source when utility power is not present, providing that the UPS battery is charged. To “cold start” your UPS, press and hold the UPS/STANDBY button until you hear a beep (about 2 seconds), then release it. The “BATTERY CHARGE/BATTERY POWER” indicator light will illuminate and AC power inverted from stored battery power will be provided at the UPS receptacles.
Shut Down Your UPS: Press and hold the UPS/STANDBY button when AC line power is absent (i.e. during a blackout, or when the UPS is unplugged) to deactivate your UPS.
Buttons
Basic Operation
6
For All Models—“MUTE/TEST” Button:
To Silence (or “Mute”) UPS Alarms: briefly press and release the MUTE/TEST button.*
To Run a Self-Test: with your UPS plugged in and turned ON, press and hold the MUTE/TEST button for two seconds.* Continue holding the button until the alarm beeps several times and the UPS performs a self test. See “Results of a Self-Test” below. Note: you can leave connected equipment on during a self-test. Your UPS, however, will not perform a self-test if it is not turned ON (see "POWER" Button description).
CAUTION! Do not unplug your UPS to test its batteries. This will remove safe electrical grounding and may introduce a damaging surge into your network connections.
Results of a Self-Test: The test will last approximately 10 seconds as the
UPS switches to battery to test its load capacity and battery charge.** If the “OUTPUT LOAD LEVEL” LED remains lit red and the alarm
continues to sound after the test, the UPS’s outlets are overloaded. To clear the overload, unplug some of your equipment and run the self-test repeatedly until the “OUTPUT LOAD LEVEL” LED is no longer lit red and the alarm is no longer sounding.
CAUTION! Any overload that is not corrected by the user imme­diately following a self-test may cause the UPS to shutdown and cease supplying output power in the event of a blackout or severe brownout.
If the “BATTERY WARNING” LED remains lit and the alarm continues to sound after the test, the UPS batteries need to be recharged or replaced. Allow the UPS to recharge continuously for 12 hours, and repeat the self-test. If the LED remains lit, contact Tripp Lite for service. If your UPS requires battery replacement, visit www.tripplite.com/sup­port/battery/index.cfm to locate the specific Tripp Lite replacement battery for your UPS.
* The alarm will beep once briefly after the indicated interval has passed. ** SmartPro models’ operation during the self-test: the “POWER” LED will be flashing and the “OUTPUT LOAD LEVEL” and “BATTERY CHARGE” LEDs will be lit and the UPS alarm will sound. OmniSmart models’ operation during the self-test: all LEDs will be lit and the UPS alarm will sound.
Basic Operation
continued
All Indicator Light descriptions apply when the UPS is plugged into an AC outlet and turned on.
“POWER” LED
For SmartPro Models Only: this green LED lights continuously
when the UPS is ON and supplying connected equipment with AC power from a utility source. The LED flashes and an alarm sounds (4 short beeps followed by a pause) to indicate the UPS is operating from its internal batteries during a blackout or severe brownout. If the blackout or severe brownout is prolonged, you should save files and shut down your equipment since internal battery power will eventually be depleted. See “BATTERY CHARGE” LED description below.
For OmniSmart Models Only: this green light will turn ON whenev­er your UPS is receiving normal AC line power. It will flash while the UPS is in CHARGE ONLY mode to indicate that the UPS will not provide battery backup during a blackout or brownout.
“VOLTAGE CORRECTION” LED (For Select SmartPro and OmniSmart Models): this green LED lights continuously whenever
the UPS is automatically correcting high or low AC voltage on the utility line without the assistance of battery power. The UPS will also emit a slight clicking noise. These are normal, automatic operations of the UPS. No action is required on your part.
“OUTPUT LOAD LEVEL” LED (For Select SmartPro and OmniSmart Models):
For SmartPro Models Only: this multicolored LED indicates the
approximate electrical load of equipment connected to the UPS's AC outlets. It will turn from green (light load) to yellow (medium load) to red (overload). If the LED is red (either illuminated continuously or flashing), clear the overload immediately by unplugging some of your equipment from the outlets until the LED changes from red to yellow (or green) and the alarm is no longer sounding. CAUTION! Any overload that is not corrected by the user immediately may cause the UPS to shut down and cease supplying output power in the event of a blackout or severe brownout.
For OmniSmart Models Only: this red light will turn ON continuously when the UPS is providing power from battery or after the UPS runs a self-test to indicate that the UPS's inverter is overloaded. If it lights up, immediately remove some of the equipment connected to the UPS and run a self-test. Large overloads may cause your UPS to shut down.
7
Indicator Lights
Basic Operation
continued
“BATTERY CHARGE / BATTERY POWER” LED:
For SmartPro Models Only: when the UPS is operating from utility
power, this LED indicates the approximate charge state of the UPS’s internal batteries: red indicates the batteries are beginning to charge; yellow indicates the batteries are roughly midway through charging; and green indicates the batteries are fully charged. When the UPS is operating from battery power during a blackout or severe brownout, this LED indicates the approximate amount of energy (ultimately affecting runtime) that the UPS’s batteries will provide: red indicates a low level of energy; yellow indicates a medium level of energy; and green indicates a high level of energy. Since the runtime performance of all UPS batteries will gradually deplete over time, it is recommended that you periodically perform a self-test (see “MUTE/TEST” Button description) to determine the energy level of your UPS batteries BEFORE a blackout or severe brownout occurs. During a prolonged blackout or severe brownout, you should save files and shut down your equipment since battery power will eventually be depleted. When the LED turns red and an alarm sounds continuously, it indicates the UPS’s batteries are nearly out of power and UPS shut down is imminent.
For OmniSmart Models Only: this LED will illuminate when your UPS is providing your equipment with battery backup power.
“BATTERY WARNING” LED: this LED lights red and an alarm sounds intermittently after you initiate a self test (See “MUTE/TEST” Button description) to indicate the UPS batteries need to be recharged or replaced. Allow the UPS to recharge continuously for 12 hours, and repeat the self-test. If the LED continues to light, contact Tripp Lite for service. If your UPS requires battery replacement, visit www.trip­plite.com/support/battery/index.cfm to locate the specific Tripp Lite replacement battery for your UPS.
8
Basic Operation
continued
9
Hospital Grade AC Receptacles: The receptacles provide your con­nected equipment with AC line power during normal operation and battery power during blackouts and brownouts. They also protect your equipment against damaging surges and line noise.
Tel/DSL Protection Jacks (Select Models Only): These jacks protect connected equipment against surges travelling over the modem/fax or telephone line. Connecting your equipment to these jacks is optional. Your UPS will still work properly without this connection.
DB9 Port (Select Models Only): The DB9 port connects your UPS to any workstation or server. Use with Tripp Lite’s PowerAlert UPS monitoring software and a DB9 cable to enable your computer to automatically save open files and shut down equipment during a blackout. The UPS can also report power availability and UPS status, and select UPS models have other capabilities. Consult your PowerAlert manual for more information.
USB Port (Select Models Only): The USB port connects your UPS to any workstation or server. Use with Tripp Lite’s PowerAlert UPS monitoring software and an USB cable to enable your computer to automatically save open files and shut down equipment during a blackout. The UPS can also report power availability and UPS status, and select UPS models have other capabilities. Consult your PowerAlert manual for more information.
Battery Replacement Door: Under normal conditions, the original battery in your UPS will last several years. Refer to “Battery Warnings” in the Safety section on page 2.
External Battery Connector (Select Models Only): Use to connect one or more Tripp Lite battery packs for additional runtime. Refer to Specifications and/or the label next to the connector to determine the appropriate variety of battery pack to use. Refer to the battery pack instruction manual for complete installation information and important safety warnings.
Input Breaker (Select Models Only): Protects your electrical circuit from overcurrent draw from the UPS load. If this breaker trips, remove some of the load, then reset it by pressing the breaker in.
Equipotential Connection (Select Models Only): Use this to connect any equipment that requires a chassis ground.
Other UPS Features
Basic Operation
continued
10
Storage
For SmartPro Models Only: CAUTION! Your UPS has an internal power source. Its outlets may still deliver current, even after the UPS is unplugged, until the UPS is completely turned OFF (deactivated). Before storing your UPS, turn it completely OFF: with the UPS ON and receiving utility power, press and hold the POWER button for one second (an alarm will beep once briefly after the interval has passed); then, unplug the UPS from the wall outlet. If you store your UPS for an extended period of time, recharge the UPS batteries once every three months: plug the UPS into a wall outlet; allow it to charge for 12 hours; and then unplug it and place it back in storage. Note: after you plug the UPS in, it will automatically begin charging its batteries; however, it will not supply power to its outlets (see Quick Installation section). If you leave your UPS batteries discharged for an extended period of time, they will suffer a permanent loss of capacity.
For OmniSmart Models Only: All connected equipment should be turned off, then disconnected from the UPS to avoid battery drain. Unplug your UPS from its AC receptacle. CAUTION:
Your UPS has an internal power source. Its outlets may still deliver current after it is unplugged, until the UPS is deactivated. To deactivate the UPS, press and hold its UPS/STANDBY button. Your UPS is now ready for storage. If you plan on storing your UPS
for an extended period of time, fully recharge the UPS batteries once every three months by plugging the UPS into a live AC outlet and letting the UPS charge for 12 hours. If you leave your UPS batteries discharged for an extended period of time, they will suffer a permanent loss of capacity.
Storage & Service
Power Sensitivity Adjustment (Select Models Only): This dial is normally set fully counterclockwise, which enables the UPS to pro­tect against waveform distortions in its AC input. When such distor­tion occurs, the UPS will normally switch to providing PWM sine wave power from its battery reserves for as long as the distortion is present. In areas with poor utility power or where the UPS’s input power comes from a backup generator, chronic waveform distortion could cause the UPS to switch to battery too frequently, draining its battery reserves. You may be able to reduce how often your UPS switches to battery due to waveform distortion by experiment­ing with different settings for this dial. As the dial is turned clock­wise, the UPS becomes more tolerant of variations in its input power’s AC waveform. NOTE: The further the dial is adjusted clock­wise, the greater the degree of waveform distortion the UPS will allow to pass to connected equipment. When experimenting with dif­ferent settings for this dial, operate connected equipment in a safe test mode so that the effect on the equipment of any waveform distortions in the UPS’s output can be evaluated without disrupting critical oper-
Basic Operation
continued
NORM DELAY
11
Storage & Service
continued
Service
Before returning your UPS for service, follow these steps:
1. Review the installation and operation instructions in this manual to ensure that the service problem does not originate from a misreading of the instructions. Also, check that the UPS System’s circuit breaker(s) are not tripped. This is the most common cause of service inquiries which can be easily remedied by following the resetting instructions in this manu­al.
2. If the problem continues, do not contact or return the UPS to the dealer. Instead, call Tripp Lite at (773) 869-1233. A service technician will ask for the UPS's model number, serial number and purchase date and will attempt to correct the problem over the phone.
3. If the problem requires service, the technician will issue you a Returned Material Authorization (RMA) number, which is required for service. If you require packaging, the technician can arrange to send you proper packaging. Securely pack the UPS to avoid damage during shipping. Do not use Styrofoam beads for packaging. Any damages (direct, indirect, special, incidental or consequential) to the UPS incurred during shipment to Tripp Lite or an authorized Tripp Lite service center is not covered under warranty. UPS Systems shipped to Tripp Lite or an authorized Tripp Lite service center must have transportation charges prepaid. Mark the RMA number on the outside of the package. If the UPS System is within the 2-year warranty period, enclose a copy of your sales receipt. Return the UPS for service using an insured carrier to the address given to you by the Tripp Lite service technician.