TriPollar STOP V User Manual

Page 1
FACIAL
RENEWAL AND REJUVENATION
USER MANUAL
ENGLISH
Page 2
1
ENGLISH
Congratulations on purchasing your TriPollar STOP V device!
Page 3
TABLE OF CONTENTS
INTRO DUCI NG TRI POLL AR STOP V 8
• OVERVIEW
• THE TRIPOLLAR STO P V KIT
USING T RIPO LLA R STOP V 11
• THE TRIPOLLAR STO P V DEVI CE
• BEFORE YOU BEGIN
• TREATI NG YOUR SK IN
• RECOMMENDED TREATMENT GUIDELINES
TROUBLESHOOTING 22
CUSTOM ER SUPPORT 22
TECHN ICAL SPECIFICATI ONS 23
APPENDIX 24
2
Page 4
3
IMPORTANT NOTICE
© Pollogen Ltd. All rights reserved. Design and technical specifications are subject to change without prior notice. Manufactured by Pollogen Ltd. Patented technology. The names Pollogen, TriPollar and STOP are registered trademarks of Pollogen Ltd.
All intellectual property rights in this publication are owned by Pollogen Ltd. and protec ted by applicable copyright laws and international treaty provisions. Pollogen Ltd. retains all rights not expressly granted. No part of this publication may be reproduced in any form whatsoever or used to make any derivative work without prior written approval by Pollogen Ltd.
All other trademarks are the property of their respective owners. Other company and brand products and service names are trademarks or registered trademarks of their respective holders.
Page 5
SAFETY
Important: Be sure to read the entire user manual prior to using the product. Always adhere to all warnings and cautions. Be sure to follow all safet y precautions below when using the TriPollar STOP V product. To learn more about TriPollar technology, visit our website at w ww.stop -age.com.
WARNINGS AND CAUTIONS
TriPollar STOP V is an electrical device that must be used properly. To reduce the risk of injury, adhere to all warnings and cautions below:
• Do not touch the device should it fall into water. Unplug the device immediately with dry hands.
• Do not place or store the device near any source of water, such as a sink or bathtub.
• Do not submerge the device under water.
• W hen plugged in, do not leave the device unattended.
• Before use, ensure that the system and its power supply do not show any visible damage. Do not operate if the electric cord is damaged or if the system is disassembled.
• Only use the device with the electric cord and power supply provided.
• Keep the device out of the reach of children.
• Be sure that your hands are completely dry when operating the device.
• Turn the device OFF and then unplug the power supply from the elec trical outlet after use or before cleaning and maintenance.
4
Page 6
5
• Use STOP Preparation Gel during treatment according to its instructions for use.
SAFETY INSTRUCTIONS
• Only use this device for its intended purpose, as described in this manual.
• Do not use the device over eyelids, closed eyes, thyroid area, lips, ears, breasts or genital areas.
• The sensation while using the device should be warm and pleasant. If you experience discomfort, discontinue use immediately.
Failure to follow the instructions in this manual may cause damage or burns to your skin.
DO NOT USE THE DEV ICE IF A NY OF THE FOLLOW ING APPLY TO YOU:
• Are under 18 years of age.
• Have a pacemaker or internal defibrillator.
• Have metal implants in the treatment area, such as gold threads. This does not apply to dental fillings or implants, which present no problem.
• Currently have cancer or a history of cancer, especially skin cancer or premalignant moles.
• Have severe concurrent conditions, such as cardiac disorders.
• Are pregnant or nursing.
• Have an impaired immune system due to immunosuppressive diseases such as AIDS, or are using immunosuppressive medications.
• Have any disease that is stimulated by heat, such as recurrent Herpes Simplex (within the treatment area).
Page 7
• Have any active skin condition within the treatment area, such as sores, psoriasis, eczema or a rash.
• Have a history of keloid scarring, abnormal wound healing or fragile skin.
• Have had any surgical procedure, laser resurfacing or deep chemical peeling in the treatment area within the past three months or are still undergoing a healing process.
• Have permanent fillers within the treatment area. If you have temporary dermal fillers within the treatment area, consult your physician before using the device.
• Have had Botox administered in the treatment area within the past three months.
• Have suspicious lesions within the treatment area.
• Currently use medications, herbal preparations, vitamins or food supplements that may cause fragile or dr y skin.
6
Page 8
7
POSSIBLE SIDE EFFECTS
The following side effects are rare. However, should they occur, they are usually temporary. If any of the following conditions appear and last more than 24 hours, consult your physician:
• Pain
• Damag e to natural skin texture (crusting, blistering, burns, fragile skin, bruising)
• Mild headache
For other aesthetic procedures or should you have any uncertainty regarding potential side effects, consult your physician before using the device.
Page 9
INTRODUCING TRIPOLLAR STOP V
This chapter introduces the TriPollar STOP V product and describes its components.
OVERVIE W
TriPollar STOP V is a safe, effec tive and easy-to-use skin renewal device that delivers professional level aesthetic results in the comfort of your own home. The TriPollar STOP V device produces genuine, noticeable, immediate and long-lasting results. The revolutionary TriPollar STOP V combines t wo effec tive technologies – TriPollar Radio Frequency (R F) and Dynamic Muscle Activation (DMA) that work simultaneously to deliver non-invasive collagen remodeling and muscle activity, as well as skin tightening and treatment of wrinkles.
Powered by Pollogen’s revolutionar y, patented TriPollar technology, TriPollar STOP V delivers RF energy into the dermis skin layer. TriPollar energy stimulates dermal activity to increase and accelerate the natural production of collagen, the protein responsible for the skin’s supple texture. Increased collagen levels in the skin result in tighter, more-supple skin.
DMA targets the SMAS (muscle) layer, toning and lifting it by stimulating gentle muscle contractions inducing a face lift effect. The combination of RF and DMA has a synergistic effect on the skin, improving the appearance of the skin.
The TriPollar STOP V system is intended for non-invasive skin renewal at home, specifically indicated for skin tightening and treatment of wrinkles. When the TriPollar STOP V device is used properly in accordance with the recommended treatment protocol, it effectively postpones the signs of aging.
8
Page 10
9
THE TRIPOLLAR STOP V KIT
The TriPollar STOP V kit contains the following components:
1. TRIPO LLAR S TOP V KIT ( FIG.1)
FIG.1
Page 11
2. TRIPOLL AR STOP V D EVICE ( FIG.2)
3. TRIPOLL AR STOP V P OWER SUP PLY AND ELEC TRIC CORD ( FIG.3)
4. TRIPOLL AR STOP V ELECTRI CAL OU TLET A DAPTER S (FIG.4)
5. STOP PREPARATION GEL (FI G.5)
After the co ntainer ha s been ope ned, the gel c an be used fo r up to 12 months.
Important: It is recommended to p erform a patch test 24 hour s prior to ini tial
use. See page 15 for details.
6. TRIPOLL AR STOP V USER MANUAL AND WARRAN TY PO LICY (F IG.6)
Preparation
FIG.5
FIG.2
FIG.3 FIG.6
FIG.4
Gel
User Manual/ International
Warranty
10
Page 12
11
USING TRIPOLLAR STOP V
This chapte r describ es the TriPoll ar STOP V dev ice and how to us e it.
THE TRIPOLLAR STOP V DEV ICE
The TriPollar STOP V device is used to increase and accelerate collagen production for skin renewal. TriPollar STOP V focuses heat deep within the dermis. The TriPollar STOP V’s DMA technology targets the SMA S layer, as shown below.
1. TRIPO LLAR S TOP V TRE ATMENT HEAD
2. EPIDERMIS
3. DERMIS
4. TRIPOLL AR STOP V P OWER IS FOCUSED
DEEP WIT HIN TH E DERMIS
5. HYPODERMIS
6. SMAS (MUSCLE)
FIG.7
Page 13
The components comprising the TriPollar STOP V device:
1. DEVICE BODY
2. TREATMENT HEAD
3. ELECTR ODE
4. TEMPERATURE SENSOR
5. POWER SUPPLY CONNECTION
6. GREEN/BLU E INDI CATOR LIGH T
7. ORANGE INDICATOR LIGHT
8. ON/OFF BUTTON
9. LOW RF
10. HIG H RF
11. LOW R F AND LOW D MA
12. HIGH RF A ND HIG H DMA
13. PROTECTIV E COVER
3.
4.
2.
6.
1.
9. 10.
12.11. 13.
12
FIG.8
5.
8.
7.
Page 14
13
The indicator lights on the TriPollar STOP V device works as follows:
GREEN LIGHT S INDI CATE RF MODE
• A single green light means that the device is ON and is op erating in low­level RF mo de.
• Two green lights mean that the device is op erating in high-level RF mode.
BLUE LIGHTS I NDIC ATE RF + DMA MODE
• A single blue light means that the device is op erating in low-level RF mode and low-level DMA mode.
• Two blue lights mean that the device is operating in high-level RF mode and high-level DMA mode.
Page 15
Each press of the Power button changes the mode and power level.
3. 4.
1.2.
5. 6.
FIG. 9
1. ON/OFF BUTTON
2. GREEN/BLU E INDI CATOR LIG HT
3. LOW RF
4. HIGH RF
5. LOW RF AND LOW DMA
6. HIGH RF AND HI GH DMA
14
Page 16
15
BEFORE YOU BEGI N
Before using the TriPollar STOP V device, do the following:
1.
Open the box containing the device, and remove the d evice and it s power supply.
2. Inser t the adapter onto the electrical posts (Fig.10).
3. Twist the adapter clockwise to secure it (Fig.11).
4. Conne ct the power supply to the device (Fig.12).
5. Plug the power supply into an electrical outlet (Fig.13).
6. Remove the STOP Preparation Gel from its box.
Important: Before using the device for the first time, it is recommended to perform a patch test 24 hours prior to initial use. To do so, apply a small amount of STOP Preparation Gel to the inner arm and treat the area with the TriPollar STOP V device. Ensure that the skin does not have a negative reaction to the gel or treatment.
7. Remove the protective cover from the device.
FIG.11FIG.10
Page 17
TREATING YOUR SKIN
To treat your skin using the TriPollar S TOP V device:
1. Cleanse and dr y the treatment area.
Apply a ver y thin, even layer of STOP Preparation Gel over the entire
2. treatment area (face and/or neck). This gel must be used during the entire treatment session. Your skin may feel a warm sensation when the gel is applied.
3. Turn the device ON by pressing the Power button (Fig.14). Make sure that the green indicator light turns ON. Tip: A long press of the Power button turns the device OFF.
Select the RF power setting (Low/High) that is most comfortable for you.
4.
Important: During your first treatment session, we recommend that you use the Low power setting and then adjust the power level according to your skin’s tolerance for subsequent treatments.
FIG.12 FIG.13 FIG.14
16
Page 18
17
5. Divide the entire treatment area into two main sections - one for each side of your face and/or neck (right side and left side). Typically, you treat one side of your face/neck entirely before treating the other side. Divide each side into several working sections that are approximately twice the size of the device head. In general, you first treat one working section and then move to an adjacent section. You repeat this general technique and movement until you have treated one entire side of your face /neck. Then, you use the same technique to treat the other side of your face and neck (Fig.15).
6.
Place the device on your skin, ensuring that all four elec trodes are touching the skin. For best results, all four electrodes must be in full contact with the skin at all times during treatment.
7. Start moving the device in very slow circular motions or small figure eights, slowly heating your skin (Fig.16). The orange indicator light will turn ON when the skin is heated to the optimal temperature. In addition, a beep will sound to indicate that the optimal treatment temperature has been reached. Important: Be sure to keep the device constantly moving on your skin. Do not leave the device on one place for long without movement, as skin overheating may occur. Be sure to constantly move the device slowly over the skin to ensure proper heating of the treated area. Moving the device too fast will not achieve the intended effect.
Page 19
8.
Once the orange light turns ON and a beep sounds, continue moving the device within one working section using the same slow circular motions for a few moments and then move to an adjacent working sec tion. The new working section now includes the heated section and the new sec tion, thus enlarging the area. You may want to start your treatment by treating the section at the jaw line, and then moving to the nasal area and so on, moving from the bottom to the top of your face (Fig.17).
The device’s automatic temperature-optimization mechanism protects the skin from overheating. After the skin has reached the maximal treatment temperature, the green indicator light goes OFF and the orange indicator light goes O N to indicate that the energy source has been automatically shut down. The device automatically turns itself back ON when the skin temperature drops below the optimal temperature or when you move the device to an adjacent treatment area where the skin temperature is lower. As you move from one section to another, the orange indicator light turns OFF if the temperature of the skin in an area is not warm enough. This is normal. Therefore, as you work within one working section and move back to the previous section, you maintain the heat in that area too. In this way, as you move back and forth between working sections, you are able maintain the heat in both sections for as long as possible.
WORKING SECTIONS
TREATMENT
AREA
FIG.16 FIG.17FIG.15
18
Page 20
19
For effective treatment, the goal is to maintain the optimal temperature for several seconds before moving to the next treatment section. This means that the orange indicator light should be illuminated as much as possible when treating an area. Important: The orange indicator light turns ON and OFF, depending on the temperature of the skin in the current area of treatment. The orange light is ON when the temperature of the skin in that area is suf ficiently heated and is O FF when it is not. In addition, a beep sounds whenever the orange indicator light turns ON.
For optimal treatment, it is recommended to begin the treatment using the Low power setting in RF mo de and to follow the guidelines below:
• If the orange indicator light and sound indicators do not appear within 5 minutes, or if you do not feel any heating, tr y moving the device as slowly as possible without stopping the motion. Slow, circular, constant movements are more effective than rapid movements. The slower the movement, the faster the treatment area heats internally.
If the orange indicator light and sound indicators still do not appear after 3-5 minutes, select a higher power level. If the indicators still do not appear after increasing to the High power setting, try moving the device more slowly, without stopping the motion.
• If the orange indicator light and sound indicators turn ON immediately or if you feel excessive heat, selec t a lower power level.
9. Repeat this technique one working sec tion at a time until the entire treatment area has been heated.
10. For best results, after treating all working sections on one side of your
Page 21
20
face/neck, go over the entire treatment area in large sweeping circles for few minutes to rewarm the entire area, without waiting for the orange indicator light to illuminate. This action maintains the optimal temperature and achieves uniform heat across the entire treatment area. This completes the treatment session for one side of your face and neck.
11. Repeat steps 4-10 for the other side of your face and neck.
12. Switch to RF + DMA mode. We recommend starting at the Low power level. Important: We recommend that you only use RF + DMA mode on the jaw line. This mode is not recommended on other facial areas.
13. Start moving the device using a ver y slow linear motions along the jaw line.
14.
To ensure optimal results, use the device for 5 minutes on each treatment side of the jaw for a total of 10 minutes. After the entire treatment area has been treated, the treatment session is complete. As a natural result of the treatment, the skin may show some redness (erythema) and minor skin swelling (edema) in the treatment area. Both should disappear af ter a short time.
15. Press the Power but ton for two seconds to turn the device O FF.
16. Unplug the power supply from the electrical outlet.
17. Gently remove the STOP Preparation Gel from your skin and wash your face. You can apply a moisturizer if you like.
18.
Clean the electrodes of the device with a damp cloth or paper towel and cover the device tip with its protective cover.
Warning: Do not submerge the TriPollar STOP V device under water.
Page 22
21
RECOMMENDED TREATMENT GUIDELINES
The following treatment guidelines are recommended for using the TriPollar STOP V device:
Treating the entire face with RF takes approximately 20 minutes. For shor ter treatment times, focus on selected areas of concern. Treatment times differ for each person.
• Treating the face with DMA takes approximately 10 minutes.
• For optimal results, use the TriPollar STOP V device 2-3 times a week for a period of 6 -8 weeks.
• Make sure that you allow at least one day in bet ween treatments on the same treatment area.
• To maintain long-term result s after finishing the full course of treatments, we recommend that you use your TriPollar STOP V device 2- 4 times a month or as needed.
Page 23
22
TROUBLESHOOTING
This chapter describ es how to troubleshoot problems that may arise when using the TriPollar STOP V device.
If your TriPollar STOP V d evice is not working, take the following steps before contacting the service center:
• Ensure that you have removed the device’s protective cover.
• Make sure that the electric cord is securely connected to the device and the Power supply adapter is firmly connected.
• Press the Power button and verif y that the green indicator light illuminates.
• Ensure that you applied the S TOP Preparation Gel according to instructions described in this manual. The TriPollar STOP V device does not work properly when the preparation gel is not applied.
CUSTOMER SUPPORT
This chapter provides customer support contact information for the TriPollar STOP V device.
For ser vice, refer to the store where you purchased the TriPollar S TOP V device or visit our website at www.stop-age.com. When contacting a lo cal ser vice center, have the device serial number ready. The serial number is located on the top panel of the device. Customer satisfaction is our top priority. To help us provide you with the best support possible, please send us your comments and suggestions. Contact information is provided on the back cover of this manual, or you can contact us via our website (see above).
Page 24
23
TECHNICAL SPECIFICATIONS
This chapter presents the specifications for the TriPollar STOP V system.
TABLE 1
Technical Specifications
Weight: 85gr Physical Dimensions: H=146mm L=50mm W=40mm TriPollar RF Frequency: 1MHz
TABLE 2
Operating conditions
Temperature: +5°C to +30°C Humidity: 20RH to 80RH Rating: AC 100Vac to 240Vac 0.6 A 50/60Hz DC 8V 1.5A
TABLE 3
Transport and Storage
Temperature: 0°C to 40 °C Humidity: 10RH to 90RH
TABLE 4
Electrical Safety
The power supply is classified as Class ll double insulation
Page 25
24
HAZARDOUS MATERIALS DECLARATION
This appendix provides the hazardous materials declaration for the TriPollar STOP V device.
Restriction of Hazardous Substance Material Declaration
TABLE 5 - IS IN ACCORDANCE WITH THE PROVISIONS OF SJ/T 11364.
Part Name
DIODE
MOSFET
UK AC PIN
EU AC PIN
OTHER
LEGEND
X
Indicates that the named hazardous substance contained in at least one of
PB HG CD CR
X
X
X
X
the homogeneous materials used for this component may be above the limit requirement of GB/ T 26572. The restricted materials are used in accordance with manufacturing needs. Use of such materials has no impact on the device’s safety and no harmful effect on device users. This declaration is included
to remind device users that the device must be disposed of reasonably and in accordance with the environmental policy of your locale.
Indicates that the named hazardous substance contained in all of the homogeneous materials for this component is below the limit requirement of GB/ T 26572.
Hazardous Substances
6+
PBB PBDE
Page 26
POLLOGEN LTD. 6 KAUFMAN ST. GIBOR H OUSE P.O.BOX 50320 TEL AVIV, ISR AEL 6801298 TEL : +972( 0)3 5104110 FAX: +972(0)3 5174112 EMAIL: INFO@POLLOGEN.COM
© POLLOGEN LTD., ALL RI GHTS RESERVED. 14700200 EU / REV 3 DESIGN AND TECHNICAL SPECIFICATION ARE SUBJECT TO CHANGE WITHO UT PRI OR NOTI CE. MANU FACTURED AND DIS TRIBUTED BY PO LLOGEN LTD. PATENTED ISR AELI TECHNOLOGY, MADE IN VI ETNA M. THE NAME STOP IS T RADEMARK OF P OLLOGEN LTD. THE NAMES POL LOGEN AN D TRIP OLLAR ARE REG ISTERED TRADEMARK S OF POLLOGEN LTD. FOR MORE INFORMATION, PLEASE VISIT: WWW.STOP-AGE.COM
O BELIS S. A. 53 BLVD. GÉNÉR AL WAHIS B-1030 BRUSSELS, BELGIUM TEL: +32(0)2 732595 4 FAX: +32(0)2 7326 003 EMAIL: MAIL@OBELIS.NET