Triplett Test Equipment & Tools TMP60, TMP60-NIST User manual

Page 1
User Manual
TMP60
Dual Input Type J/K Thermometer
www.GlobalTestSupply.com
Page 2
Introduction
Congratulations on your purchase of the Triplett TMP60 Digital Thermometer with J/K - type dual input. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Features
1. Large backlit display shows any combination of T1, T2, T1-
2. Relative time clock on MAX MIN and AVG provides a time
3. Electronic Offset function allows compensation of thermocouple
4. Readout in °C, °F, or Kelvin (K).
5. Auto Power Off mode (Sleep mode) increases battery life.
Safety
1. To prevent personal injury or meter damage, use the meter only
2. If the low battery indicator appears, replace the batteries
T2, plus MAX, MIN, AVG.
reference for major events.
errors to maximize overall accuracy.
as specified in this guide
immediately. Inaccurate temperature reading could result from low battery voltage.
3. Do not operate the meter in an explosive atmosphere
To avoid electrical shock or personal injury, do not apply more than 20Vrms between the thermocouple inputs or between the thermocouple and earth ground
WARNING
Do not leave batteries and packing material lying around unattended; they can be dangerous for children if they use them as toys
www.GlobalTestSupply.com
Page 3
Meter Description
Meter
1. T1 Thermocouple input connector
2. T2 Thermocouple input connector
3. Function buttons (Hold, MAX/MIN, Set)
4. Function buttons (Power, Enter,T1 T2
5. Battery compartment (rear)
T1-T2, C-F-K)
www.GlobalTestSupply.com
Page 4
Operation
Connecting thermocouples
1. This meter accepts one or two K or J-type thermocouples with spade lugs (sub-miniature type with one spade wider than the
2. Plug the thermocouple(s) into the meter's thermocouple input
other).
jack(s).
Measuring
1. Press the button to turn on the thermometer on. After 1 second the thermometer will display the T1 reading in the upper display and the T2 reading in the lower display. The display will indicate dashes (----) if no thermocouple is connected to the
2. Press the C-F-K button to select the desired temperature
3. Contact the object to be measured with the probe. Read the
input.
unit of measure. A display icon will reflect the selection.
measured temperature in the display.
measurement range. The display shows "- - - -" when a
thermocouple is not connected.
MAX/MIN/AVG mode
Press the MAX/MIN button to begin recording the Maximum, Minimum and Averag appear on the display and the elapsed time display will appear.
1. Press MAX/MIN to enter this mode.
2. Press MAX/MIN to step through maximum, minimum and average
readings.
3. Press and hold MAX/MIN for 3 seconds to return to normal mode.
REC MAX
www.GlobalTestSupply.com
Page 5
Data Hold
Press the HOLD button to freeze the reading in the display. HOLD will appear at the top right of the display. Press the HOLD button again to return to normal operation mode.
T1 T2 T1-T2
Press the T1 T2 T1-T2 button to cycle through the four display combinations.
No. Upper Display Lower display
1 T1 T2
2 T2 T1
3 T1-T2 T1
4 T1-T2 T2
Press the button to select the displayed temperature units
Backlight
Press the SETUP button to turn the LCD backlight ON/OFF.
Setup Mode Options
Use the Setup mode to select the thermocouple type, program a display offset, or disable/enable the sleep mode. The thermometer stores the settings in its memory.
Press and hold the SET button for 3 seconds to enter or exit the Setup mode. When the thermometer is in Setup mode, the display shows the SETUP icon. Use the ENTER buttons or the arrow buttons to scroll
parameters.
Set Type J or K thermocouple
o From within the Setup mode (see the Setup Mode Options
paragraph above for accessing Setup mode) use the
p Press ENTER to access the parameter.
or
www.GlobalTestSupply.com
Page 6
q Use the
r Press ENTER to save the setting.
s Press and hold the SET button for 3 seconds to return to the
normal operating mode or press ENTER to move to the
button to toggle between J and K.
Set Sleep/Non-sleep mode
The thermometer enables the sleep mode option by default where the meter will automatically shut off after 20 minutes of inactivity.
o From within the Setup mode (see the Setup Options paragraph
above for accessing Setup mode) use the
p Press ENTER to access the parameter.
q Use the button to select ON (sleep mode enable) or OFF (sleep
mode disabled).
r Press ENTER to store the new setting in memory.
s Press and hold the SET button for 3 seconds to return to the
normal operating mode or press ENTER offset setting.
or button to
Set T1 or T2 Offset
Use the offset feature (T1 or T2 Setup menu item) to adjust the
at a specific temperature. The allowable adjustment range is ± 5.0°C or ± 9.0 °F.
1. Plug the thermocouple into the input connector.
2. Place the thermocouple into a known, stable temperature environment (such as an ice bath or dry well calibrator). Allow the
3. From within the Setup mode (see the Setup Mode Options
reading to stabilize.
paragraph above for accessing Setup mode) press
ENTER.
or to
www.GlobalTestSupply.com
Page 7
4. Use the arrow or buttons to adjust the offset. The lower digits show the offset amount while the upper digits show the actual
5. Press the ENTER button to store the value and move to the next
6. Press and hold the SET button for 3 seconds to return to normal
temperature with the offset applied.
parameter.
mode.
Replacing the Batteries
If the low battery icon appears, replace the batteries immediately to avoid inaccurate readings.
1. Turn off the thermometer.
2. Loosen the screw and remove the battery door.
3. Replace the AAA batteries.
4. Replace the battery door and tighten the screw.
You, as the end user, are legally bound (EU Battery
ordinance) to return all used batteries, disposal in the household garbage is prohibited! You can hand over
your used batteries / accumulators at collection points in your community or wherever batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle
www.GlobalTestSupply.com
Page 8
Specifications
Dimensions:
Temperature Range:
K type thermocouples J type thermocouples
Temperature accuracy:
>-148°F (
T1, T2
100°C) ± [0.15%rdg+1.8°F (1°C)] <-148°F (
T1, T2
100°C) ± [0.5%rdg+3.6°F (2°C)]
T1-T2 ± [0.5%rdg+1°C (1.8 °F)]
Display Resolution: 0.1°<1000°, 1°>1000°
Battery:
Operating Temperature: 32°F to 122°F (0°C to 50°C)
-328°F to 2501°F (-200°C to 1372°C) (73°K to 1645°K)
-346°F to 2012°F (-210°C to 1100°C) (63°K to 1373°K)
Storage Temperature: 14°F to 122°F (10°C to 50°C)
Weight: 7 oz. (200g)
Warranty
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase. This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries are not covered by this warranty