Congratulations on your purchase of the Triplett TMP50 Type K/J
Thermometer This meter. This single input contact thermometer offers fast
response and laboratory accuracy and works with K-type or J-type
thermocouples as temperature sensors and offers a 0.1 0C/0.1 0 F resolution.
2. Press the C-F-K button to select the desired temperature unit of
MAX/MIN/AVG mode
measure. A display icon will reflect the selection.
3. Touch the object to be measured with the probe. Read the
measured temperature on the display.
Note: If the display indicates “OL” the temperature is outside the
measurement range. The display shows "- - - -" when a
thermocouple is not connected.
Press MAX/MIN to capture and step through the maximum (MAX),
minimum (MIN) and average (AVG) temperature values. These values will
be shown in the lower display with an elapsed time reference below
(hrs:mins, mins:secs).
1. Press MAX/MIN to enter this mode.
2. Press MAX/MIN to step through maximum, minimum and average
readings.
3. Press and hold MAX/MIN for 3 seconds to return to normal mode.
Data Hold
Press the HOLD button to freeze the reading in the display. HOLD will
appear at the top right of the display. Press the HOLD button again to
return to normal operation mode.
Backlight
Press the SETUP button to switch the backlight OFF/ON.
Setup option
The setup option is used to select the thermocouple type that matches the type
that will be connected to the meter, to switch the sleep mode OFF/ON, and to
adjust the thermocouple offset.
Note: In the setup mode, the upper display shows the menu items and
the lower display shows the selection for a given menu item.
Press and hold the SET button for 3 seconds to enter or exit the
Setup mode. When the thermometer is in Setup mode, the
display shows the SETUP display icon.
Set Type J or K thermocouple
1. Press ▲ or ▼to scroll to the “TYPE” setting (shown on the bottom left
of the display).
2. Press ENTER to choose the option.
3. Press ▼ to toggle between J and K.
4. Press and hold the SET button for 3 seconds to save the setting and
return to normal mode.
Set Sleep/Non-sleep mode
The thermometer enters sleep mode by default. The meter will
automatically shut off after 20 minutes of inactivity (no buttons pressed).
When the thermometer is in Setup mode, the display shows the SETUP
display icon.
1. Press ▲ or ▼to scroll to the SLP setting.
2. Press ENTER to choose the option; On or OFF will appear in the
display.
3. Use the arrow key ▼ to select ON or OFF.
4. Press ENTER to store the new setting.
5. Press and hold the SET button for 3 seconds to return to normal
mode.
Set Offset
Use the offset option in Setup (T1 menu item) to adjust the
thermometer’s readings to compensate for known temperature offset at
a specific temperature. The adjustment range is ± 5.0°C or ± 9.0 °F.
1. Plug the thermocouple into the input connector.
2. Place the thermocouple into the known, stable temperature
environment (such as an ice bath or dry well calibrator). Allow the
reading to stabilize.
3. Use the arrow keys ▲ ▼to scroll to the “T1” option and then press
ENTER.
4. Change the offset temperature reading (upper display) by
pressing ▼until the desired temperature is reached.
5. Press and hold the SET button for 3 seconds to return to normal
mode.
Replacing the Batteries
1. Switch OFF the thermometer.
2. Loosen the battery compartment screw and remove the compartment
door.
3. Replace the AAA batteries.
4. Replace the battery door and tighten the screw.
Never dispose of used batteries or rechargeable batteries
in household waste. As consumers, users are legally
required to take used batteries to appropriate collection
sites, the retail store where the batteries were purchased,
Disposal: Do not dispose of this instrument in household waste. The user is
obligated to take end-of-life devices to a designated collection point for the
disposal of electrical and electronic equipment.
or wherever batteries are sold.
Other Battery Safety Reminders
o Never dispose of batteries in a fire. Batteries may explode or leak.
• Never mix battery types. Always install new batteries of the same type.
Temperature Range:
K type thermocouples -200°C to 1372°C (-328°F to 2501°F)
J type thermocouples -210°C to 1100°C (-346°F to 2012°F)
Note that thermocouple wire may be insulated with materials that cannot
withstand temperatures in the ranges listed above. Check that the
thermocouples used are rated for the temperature measurements to be
made.
Warranty Information
Triplett / Jewell Instruments extends the following warranty to the original purchaser of
these goods for use. Triplett warrants to the original purchaser for use that the products sold
by it will be free from defects in workmanship and material for a period of (1) one year from
the date of purchase. This warranty does not apply to any of our products which have been
repaired or altered by unauthorized persons in any way or purchased from unauthorized
distributors so as, in our sole judgment, to injure their stability or reliability, or which have
been subject to misuse, abuse, misapplication, negligence, accident or which have had the
serial numbers altered, defaced, or removed. Accessories, including batteries are not
covered by this warranty