Congratulations on your purchase of the Triplett RHT32 Digital
Psychrometer. This unique meter designed as pocket size, battery operated
for Humidity, Dry Bulb, Dew Point, Wet Bulb, External Temperature &
Temperature Differential measurement. The sensor is also specially
protected by protective turn cap.
The psychrometer is a micro processor-based design. A must device for
HVAC engineers use.
Turn the protective cap (top of
meter), in the direction of the OPEN
arrow to allow air into the sensor
cavity. Turn the cap in the direction
of the CLOSE arrow to protect the
sensors. Always open the cap to
take measurements and close when
storing.
Meter Power
Press the power button for one second to switch the meter ON/OFF.
The meter will perform a short self-test when switched ON.
Auto Power Off (Sleep Function)
The meter will turn itself off after 5 minutes.
To override Auto Power Off function, press + HOLD while the meter
is off. When "n" appears (see figure A), release the HOLD button. The
meter is now in Non-Sleep Mode.
Figure A
Mode Options
1. Turning on the protective sensor cap in counterclockwise direction.
2. Turn meter on by pressing power button (see figure B).
3. Press C/F key for more than 1 second to convert reading to
desired unit. Both the temperature and relative humidity
measurement will display simultaneously (see figure C).
into socket before turning the meter on, the error code will also
appear when you check the MIN/MAX value.
NOTE: Low battery power tends to give inaccurate readings,
therefore make sure you have enough power.
Calibration
WARNINGS
the memory by pressing the "ON/OFF" button before step 4. During
step 4, press the "ON/OFF" button to exit the calibration mode.
23°C.
this indicates the
calibration has failed. See troubleshooting section below.
1. Turn the meter off and plug the sensor probe into 33% salt bottle.
Press the "ON/OFF + C/F " buttons for more than one second to
enter calibration mode.
2. "32.8%" will flash on the display, when "---.-" is showing on the
display, that means the values are invalid in the calibration process.
After 30 min, the flashing will stop to indicate the procedure is
finished.
3. Move the sensor to 75% salt bottle and press "MN/MX" for more
than one second to enter 75.3% calibration. "75.3%" will flash
on the display.
4. The flashing will stop after 30 minutes, at this point the
entire calibration is completed and the calibration data
has been saved in memory.
5. Clean the meter housing with the sensor cap closed, using a
lightly damp cloth and mild detergent when necessary. Do not
use solvents or abrasives.
6. Store the meter -- with the protective cap closed -- in an area with
moderate temperature and humidity (refer to the operating/storage
conditions specifications).
Battery Replacement
When the batteries need replacing, the symbol will appear on the
compartment, observing correct polarity.
Never dispose of used batteries or rechargeable batteries
in household waste. As consumers, users are legally
required to take used batteries to appropriate collection
sites, the retail store where the batteries were purchased,
or wherever batteries are sold.
Disposal: Do not dispose of this instrument in household waste.
The user is obligated to take end-of-life devices to a designated
collection point for the disposal of electrical and electronic
equipment.
Other Battery Safety Reminders
Never dispose of batteries in a fire. Batteries may explode or
leak.
Never mix battery types. Always install new batteries of the
same type.
Triplett / Jewell Instruments extends the following warranty to the original purchaser of
these goods for use. Triplett warrants to the original purchaser for use that the products
sold by it will be free from defects in workmanship and material for a period of (1) one
year from the date of purchase. This warranty does not apply to any of our products
which have been repaired or altered by unauthorized persons in any way or purchased
from unauthorized distributors so as, in our sole judgment, to injure their stability or
reliability, or which have been subject to misuse, abuse, misapplication, negligence,
accident or which have had the serial numbers altered, defaced, or removed.
Accessories, including batteries are not covered by this warranty