Triplett Test Equipment & Tools MM300-NIST, MM300 User manual

Page 1
User Manual
MM300 True RMS Manual Ranging Digital Multimeter
Introduction
Congratulations on your purchase of the Triplett MM300 True RMS Manual Ranging Digital Multimeter This instrument performs AC/DC voltage, AC/DC Current, Resistance, Audible Continuity, Diode, Temperature measurements and Frequency measurements, it is a 3 3/4 digits, 3999 counts
Safety
Circuits under test must be de-energized and isolated before connections are made (except for
International Safety Symbols
voltage measurements).
Circuit connections must not be touched during a test. Use extreme caution when working near bare conductors and bus bars. Accidental contact with conductors could result in electrical shock.
Use caution when working near voltages above 60VDC or 30VACrms.
After insulation tests, capacitors must be discharged.
Test leads (including alligator clips) must be in good working order, clean and without broken
or cracked insulation.
When servicing, use only specified replacement parts.
Caution, refer to this manual before using this meter
Find Quality Products Online at: sales@GlobalTestSupply.com
Dangerous Voltages
Meter is protected throughout by double or reinforced insulation
www.GlobalTestSupply.com
Page 2
DC (Direct Current).
AC (Alternating Current).
DC or AC
Important safety information. Refer to the manual.
Dangerous voltage maybe present.
Earth ground.
Low battery
Fuse.
Diode
Continuity test
°C
Centigrade
°F
Fahrenheit
AUTO
Auto range
Conforms to European Union directive.
Double insulated.
APO
APO auto power off
Battery capacity
Relative test
Data hold
ELECTRICAL SYMBOLS
Meter Description
1) Protective Holster
2) LCD Display
3) Select.” Button When you measure the current, pressing this button will switch the meter between DC function and AC function. When you measure the diode or the continuity, pressing this button will switch the meter between the diode and the continuity functions.
4) ” Button After pressing the button, the present reading is held on
the display, meanwhile “ ” is displayed on the LCD
as an indicator. To exit the Hold Mode, press the button again and the indicator “ ” will disappear.
Press the button for 2 seconds to turn the Backlight ON. Press again for 2 seconds to turn OFF.
5) Function/Range Rotary Switch This switch can be used to select desired function and range.
6) “A” Jack Plug-in connector for the red test lead for Current (200mA ~ 10A) measurements.
7) “COM” Jack Plug-in connector for black (negative) test lead.
8) “INPUT” Jack
Plug-in connector for the red test lead for all measurements except current (≥200mA) measurements
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Operation
DC & AC VOLTAGE MEASUREMENT
1. Connect red test lead to INPUT” jack, Black lead to “COM” jack.
2. Set RANGE switch to desired VOLTAGE range, if the voltage to be measured is not known beforehand, set switch to the highest range and reduce it until satisfactory reading is obtained.
3. Connect test leads to device or circuit being measured.
4. Turn on power of the device or circuit being measured voltage value will appear on Digital Display along with the voltage polarity.
DC/AC CURRENT MEASUREMENT
1. Red lead to “INPUT”. Black lead to “COM” (for measurements between 200mA and 10A connect red lead to “10A” jack with fully depressed.)
2. Range switch to desired range. The default function of the meter is the measurement of DCA.
3. Press “select” button to switch to ACA measurement mode.
4. Open the circuit to be measured, and connect test leads in series with the load in with current is to measure.
5. Read current value on Digital Display.
6. Additionally, “A” function is designed for intermittent use only. Maximum contact time of the test leads with the circuit is 15 seconds, with a minimum intermission time of seconds between tests.
RESISTANCE MEASUREMENT
1. Red lead to “INPUT”. Black lead to “COM”.
2. Range switch to desired Ω range.
3. If the resistance being measured is connected to a circuit, turn off power and discharge all capacitors before measurement.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
4. Connect test leads to circuit being measured.
5. Read resistance value on Digital Display.
DIODE MEASUREMENT
1. Red lead to “INPUT”, Black lead to “COM”.
2. RANGE switch to “ ” range.
3. Connect the red test lead to the anode of the diode to be measured and black test lead to cathode.
4. The forward voltage drop in mV will be displayed. If the diode is reversed, figure “OL” will be shown.
TEMPERATURE MEASUREMENT
1. Range switch to ºC or ºF range, it will display room temperature in ºC or ºF value.
2. Connect the K-type thermoelectric couple to “INPUT” and “COM” jacks.
3. The display will read Temperature value ºC or ºF.
4. Press “select” button, can switch the unit of measurement of temperature. The default is temperature measurement in Celsius.
NOTE: The K-type thermocouple Max. Operating temperature of Probe: 250˚C/482˚F (300˚C/572˚F short-term). The sensor supplied with the instrument is an ultra- fast response naked bead thermocouple suitable for many general purpose applications.
AUDIBLE CONTINUITY TEST
1. Red lead to “INPUT”, Black lead to “COM”.
2. RANGE switch to “ ” range.
3. Connect test leads to two points of circuit to be tested. If the resistance is lower than 30Ω±20Ω, the buzzer will sound.
BATTERY TEST
1. Connect the black test lead to the "COM" jack and the red test lead to the "INPUT" jack (Note: The polarity of the red test lead is positive "+").
2. According to the different type of the battery (1.5V, 9V,) to be tested, set the range switch to the desired BATT range.
3. Connect the test leads to the battery to be tested.
4. Read the reading on the display. The polarity of the red test lead connection will be indicated.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
RANGE
RESOLUTION
ACCURACY
2V
1mV
±0.8%of rdg+5D
20V
10mV
200V
100mV
600V
1V
±1.0% of rdg+5D
RANGE
RESOLUTION
ACCURACY
200V
100mV
±1.0% of rdg+5D 600V
1V
” FUNCTION
5. After pressing the button, the present reading is held on the display, meanwhile
” is displayed on the LCD as an indicator. To exit the Hold Mode, press the button again and the indicator “ ” will disappear.
6. Pressing the button more than 2 seconds will turn the Backlight ON and OFF.
Auto Power Off
If you don’t operate the meter for about 15 minutes, it will range switch or
press a button. In the power off state, press and “Select” button to Rotary range switch, you can cancel the auto power off function, “APO” symbol disappear from the LCD.
Specifications
Accuracies are guaranteed for 1 year, 23ºC±5ºC, less than 80%RH
DC VOLTAGE
OVERLOAD PROTECTION: 220V rms AC for 200mV range and 600V DC or
600V rms for all ranges.
AC VOLTAGE
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
RESPONSE: Average responding, calibrated in rms of a sine wave.
RANGE
RESOLUTION
ACCURACY
200mA
100µA
±1.5% of rdg+4D
10A
10mA
±2.0% of
rdg+10D
Range
Resolution
Accuracy
200mA
100µA
±(1.5% of rdg + 4D)
10A
10mA
±(2.0% of rdg + 4D)
RANGE DESCRIPTION
Built-in buzzer sounds if resistance is
less then 30±20Ω
FREQUENCY RANGE: 45Hz ~ 450Hz OVERLOAD PROTECTION: 600V DC or 600V rms for all ranges.
AUDIBLE CONTINUITY
OVERLOAD PROTECTION: 15 second maxi- mum 220 V rms.
DC CURRENT
OVERLOAD PROTECTION: 500mA/600V fuse or 10A/600V fuse MEASURING VOLTAGE DROP: 200mV
AC CURRENT
Overload Protection: µA and mA ranges: F0.5A/600V fuse 10A ranges: F10A/600V fuse Max. Input Current: “INPUT” jack: 200mA “A” jack: 10A (For measurements>5A: duration <10 seconds, interval >15 minutes) Voltage Drop: 200mA 10A ranges: 200mV Frequency Range: 40Hz ~ 400Hz Response: Average, calibrated in rms of sine wave
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
RESISTANCE
RANGE
RESOLUTION
ACCURACY
200Ω
0.1Ω
±1.0% of rdg+10D
2KΩ
20KΩ
10Ω
200KΩ
100Ω
2MΩ
1KΩ
RANGE
RESOLUTION
ACCURACY
-40℃~150℃
1℃
±
1.0%+4D
150℃~1370℃
±
1.5%+15D
-40℃~302℃
1℃
±
1.0%+4D
302℃~1999℃
±
1.5%+15D
Range
Resol.
Internal Resistance
9V
10mV
91
1.5V
1mV
150Ω
OVERLOAD PROTECTION: 15 seconds maxi- mum 220Vrms.
TEMPERATURE (with K-TYPE PROBE)
BATTERY TEST
BATTERY AND FUSE REPLACEMENT
Fuse rarely need replacement and blow almost always as a result of operator error. If “ ” appears in display, it indicates that the battery should be replaced. To replace battery & Fuse (500mA/600V or 10A/600V) remove the 1 screws in the bottom of the battery case, simply remove the old, and replace with a new one. Be careful to observe polarity .Remove the hostel and remove the bottom cover of the meter to replace the fuse. Please replace the appropriate fuse according to operation manual.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
ACCESSORIES
Operator’s instruction manual  Set of test leads  Gift box  K-type thermoelectric couple  1.5V 3, AAA s
General specifications
Display : LCD, 1999 counts updates 2/sec LCD Size : 55 x 31mm Polarity Indication : “-” displayed automatically Over-range Indication : “OL” displayed Low Battery Indication : “ ” displayed Range select : auto or manual Operation Temperature : 0°C to 40°C, less than 80%RH Storage Temperature : -10°C to 50°C, less than 85%RH Battery Type : 1.5V x 3, AAA size Dimension (H×W×D) : 145×70×35mm Weight : Approx 157g
Approvals
Warranty Information
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase. This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries are not covered by this warranty
Copyright © 2019 Triplett
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...