Triplett IRT225 User Manual

Page 1
User Manual
IRT225
10:1 Mini IR Thermometer
Introduction
Congratulations on your purchase of the IRT225 10:1 infrared thermometer The IRT225 can quickly and accurately determine the surface temperature by measuring the infrared energy radiated from the target surface. It is suitable for non-contact surface temperature measurements.
Page 2
Safety
Warning:
LASER RADIATION DO NOT STARE INTO BEAM
OUTPUT < 1mW COMPLIES WITH EN60825-1:2014 To avoid eyes or personal injury, please read the following safety instructions
before using the thermometer:
Please do not irradiate people or animals with laser directly or
Please do not look at the laser directly or through other optical tools
If battery symbol on LCD display is flashing, replace batteries
Inspect the case before using the thermometer. Do not use the
Please refer to the emissivity information for the actual temperature.
Do not use the thermometer in an explosive gas, steam, or humid
To ensure measurement accuracy, place the thermometer in the
Avoid keeping the thermometer near high temperature objects for a
indirectly.
(telescope, microscope, etc).
immediately to avoid inaccurate measurements before using the thermometer.
thermometer if it appears damaged. Look for cracks or missing plastic.
Reflective objects can cause the actual temperature value to be higher than the measured temperature value. When measuring these objects, please be aware not to touch high temperature surfaces.
environment.
current environment for more than 30 minutes before using it.
long time
Page 3
Description
1
Laser
2
Infrared sensor
3
Trigger
4
Battery cover
5
LCD
6
°C/°F switching button
7
Battery compartment
METER DESCRIPTION
LCD DESCRIPTION
1. Function Indicator
Page 4
Operation
Viewing the Last Measured Value
In the off state, short press the trigger to turn on the thermometer, it will display the last measurement data that was held before the last shutdown.
Temperature Measurement
Aim the thermometer at the target, then pull and hold the trigger, the laser indicator on the LCD display shows that the laser is turned on and points to the target, the measured temperature value will be displayed on the LCD, and the maximum or minimum temperature of the target area can be measured by scanning up, down, left and right according to the laser spot indication. Note: The measurement distance must be determined by the size of the measured target according to the “distance to spot size ratio (D:S)” and “field of view”.
Hold
After the trigger is released, the last measured value and the HOLD icon will remain on the LCD display. If there is no operation for 8 seconds, the thermometer will automatically shut down and save the last measured value.
Unit Conversion
In the power-on state, open the battery cover and short press the °C/°F button to switch the temperature unit to Celsius (°C) or Fahrenheit (°F).
Distance to Spot Size Ratio (D:S)
As the distance (D) from the target being measured increases, the spot size (S) on the measured area becomes larger. The relationship between the distance and the spot diameter is as shown in the figure below. The spot diameter represents 90% of the energy in the circle. The maximum
D:S (10:1) will be obtained when the thermometer is 400mm (16’’) from the
target resulting in a spot diameter of 40mm (1.6’’).
Page 5
Field of View
Make sure that the measured target is larger than the spot diameter. The smaller the target is, the closer the test distance should be (please refer to D:S for the spot diameter at different distances). To obtain the best measurement result, it is recommended that the target being measured is
2.5 times larger than the spot diameter.
Page 6
Symptom
Problem
Action
OL appears when measuring
Measured value is greater than the maximum range
Stop measuring
-OL appears when measuring
Measured value is less than the minimum range
Stop measuring
Err appears when booting
Exceeding the minimum or maximum operating ambient temperature
Place the thermometer in a 0°C~50°C (32°F~122°F) environment, and it can be recovered after 30 minutes
Battery indicator flashes
Low battery
Replace batteries
Laser fails to work / weak laser
Low battery
Replace batteries
Troubleshooting
Page 7
The measurement is inaccurate
Low battery, emissivity mismatching, measured distance is too far, measured target diameter is less than 20mm, etc.
Please refer to the instructions for field of view, D:S, etc.
Maintenance
Cleaning and storage
1. The meter should be wiped with a damp cloth and mild detergent when necessary. Do not use solvents or abrasives.
2. Store the meter in an area with moderate temperature and humidity.
Battery Replacement
Please install or replace the batteries (AAA (LR03) 2*1.5V) as shown on the right.
Page 8
Specifications
Temperature range
-20°C~400°C (-4°F~752°F)
Accuracy Ambient temperature 21°C~25°C (70°F~77°F)
>0°C: ±2.0°C or ±2% of reading, whichever is greater ≤0°C: ±(2.0°C+0.1°C/°C) >32°F: ±4.0°F or ±2% of reading, whichever is greater ≤32°F: ±(4.0°F+0.1°F/°F)
Temperature coefficient
±0.1°C/°C or ±0.1%/°C, whichever is greater (±0.1°F/°F or ±0.1%/°F, whichever is greater)
D:S ratio
10:1 (calculated at 90% energy)
Emissivity
0.95
Response time
500ms
Spectral response
5.5um~14um Resolution
0.1°C (0.1°F)
Repeatability
1°C or 1.0%, whichever is greater (2°F or 1.0%, whichever is greater)
Laser
Single laser, wavelength 630nm~670nm Output power < 1mW, class 2 laser
Battery type
AAA * 2 | 1.5V
Battery life
≥20H (continuous operation when laser backlight is
turned on)
Product size
135mm*94mm*36mm
Product weight
About 148g (including batteries)
Operating temperature
0°C~50°C (32°F~104°F)
Storage temperature
-20°C~60°C (-4°F~140°F)
Operating humidity
<RH90% (non-condensing) Drop test
1m
Page 9
Safety Standards
CE Certification: EN61326-2-3: 2013 Laser safety standard: EN60825-1:2014
Reference Standard
JJG 856-2015
Warranty
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase. This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries are not covered by this warranty
Copyright © 2021 Triplett
www.triplett.com