Triplett GSM400 User Manual

Page 1
User Manual
GSM400
CO2 Meter with RH% and Temperature
Introduction
Congratulations on your purchase of the Triplett GSM400 CO2 Meter with RH% and Temperature. This meter measures CO2 (Carbon Dioxide) levels, air temp., dew point, wet bulb temperature and humidity and is an ideal instrument for indoor air quality (IAQ) diagnosis.
Description
METER DESCRIPTION
1. Temperature and Humidity Sensor
2. CO2 Sensor (Rear)
3. LCD Display
4. AC Adaptor Connector
5. RS-232 Port
6. Keypad
7. Battery Compartment (Rear)
Page 2
LCD DISPLAY
O
%
1. CO2 concentration in ppm
2. Relative Humidity in %
3. Air Temperature, Dew Point or Wet Bulb Temperature
SYMBOLS
TWA Time weighted average (8 hours) STEL Short-term exposure limit (15 minutes weighted average) HOLD Freezes current reading on display MIN/MAX Minimum/Maximum readings
DP Dew point temperature AIR Air temperature WBT Wet bulb temperature % Unit of relative humidity C or F Celsius/Fahrenheit
Low battery indicator
BATTERY INSTALLATION
The meter is powered by 4 AA batteries or a DC adaptor. Install the batteries into the rear battery compartment observing correct polarity. When an adaptor is used, the batteries will be disconnected from the meter. The adaptor cannot be used as a battery charger. When the battery voltage drops below the required level, and “Lob” will appear on the display, a beeper sounds and readings
are no longer displayed. (Press any key but the SET to stop the beeps). Replace the batteries to resume normal operation.
CO2
You, as the end user, are legally bound (EU Battery ordinance) to return all used batteries, disposal in the household garbage is prohibited! You can hand over your used batteries / accumulators at collection points in your community or wherever batteries / accumulators are sold!
Page 3
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle
Operation
POWER ON/OFF
Press SET to turn the meter on and off. At power up, the meter emits a short beep and performs a 30 second countdown for meter warm up. It then enters the normal operation mode with current CO2, temperature and humidity readings displayed.
CO2
%
O
TAKING MEASUREMENT
The meter starts measurements when powered on and updates readings every second. If the operating environment changes (ex. from high to low temp.), it takes 30 seconds for CO2 sensor to respond and 30 minutes for RH.
NOTE: Do not hold the meter close to your mouth or any other source of CO2.
AIR, DP and WBT Measurements
Press the DP/WBT button to switch temperatures display. The lower left display will cycle from
“AIR” air temperature, “DP” dew point temperature and “WBT” wet bulb temperature.
DATA HOLD
Press the HOLD button to freeze the readings, the "HOLD” icon is displayed on the left top of the
display. All current readings are held unchanged, except STEL and TWA. Press “HOLD” again to
cancel the hold function.
BACKLIGHT
Press the MODE/ button for more than 1 second to activate or to cancel backlight function.
MIN,MAX,STEL,TWA
In the normal mode, press the Max/AV button to see the minimum, maximum, and weighted average readings. With each press of the Max/AV button, the meter displays MIN, MAX, STEL, TWA in sequence and then returns to the normal mode.
In MIN and MAX modes, the meter shows the minimum and maximum readings of CO2 (main display), AIR, DP or WB temperatures (lower left display) and Humidity (lower right display).
In STEL and TWA modes, the main display shows the weighted average of CO2 readings for the past 15 minutes (STEL) or 8 hours (TWA). The lower displays are the current measurements
STELCO2
%
O
NOTE:
8. If the meter has been powered on for less than 15 minutes, the STEL value will be the weighted average of readings taken since power on. As well, the TWA mode will display a weighted average of readings prior to 8 hours of operation.
9. The GSM400 takes at least 5 minutes to calculate STEL and TWA. The display shows “---- during the first 5 minutes from power on.
Page 4
3. The STEL and TWA values will keep updating every 5 minutes.
ALARM
The meter features an audible alarm to give warnings when CO2 concentration exceeds the set limit. It emits beeps (Abt.80dB) when CO2 level goes over the set value and stops when any key (except SET) is pressed or the readings fall below the set value. It beeps again if the value exceeds the limit. Restart the meter if the beeper will not stop.
AUTO POWER OFF
The meter turns off automatically after 20 minutes of inactivity. To override the function, press and hold down the SET and HOLD buttons until “n” appears in the display (approx. 2 seconds) while turning on the meter on. NOTE: Auto sleep function will be disabled during calibration mode.
SETUP (alarm limit and temperature scale)
In the normal mode press and hold the SET button for more than 1 sec to enter the setup mode. To exit the setup mode, press the CAL/Esc button when either P1.0 or P3.0 is displayed.
P1.0 CO2 ALARM Limit
When entering setup mode, P1.0 and “AL” are displayed. Press the MN/AV button to scroll to P1.1 for setting CO2 alarm threshold. The current CO2
set value will be blinking.
Press the MODE/▲ button to increase or the DP/WBT/▼ button to decrease the value. Each press adjusts 100 ppm. The alarm range is from 100 to 9900ppm. When the preferred alarm value is set, press the MN/AV button to exit and save the setting or the CAL/Esc button to exit without saving and return to P1.0.
P3.0 TEMPERATURE SCALE
Press the MODE/▲ button or the DP/WBT/▼ button in P1.0 to access P3.0 for setting the temperature scale. Press the MN/AV button to go into P3.1 for setting the temperature units. The currently selected units (°C or °F) will be blinking in the display. To switch units, press the MODE/▲ button. Press the MN/AV button to save the setting or press the CAL/Esc button exit without saving and return to P3.0.
Page 5
Calibration
CO2 CALIBRATION
The meter is calibrated to a standard 400ppm CO2 concentration at the factory
CAUTION: Do not calibrate the meter in an atmosphere of unknown CO2 concentration.
1. Place the meter in the 400ppm calibration chamber. Turn the
RH CALIBRATION
meter on and hold down the CAL/Esc and MODE/ buttons simultaneously to enter CO2
calibration mode. 400ppm and “CAL” will
blink on the LCD while performing calibration.
2. Wait about 5 minutes until the blinking stops. The calibration is then completed and the meter automatically returns to the normal mode.
NOTE: Ensure the batteries are fresh prior to calibration to
prevent an interruption or failed calibration.
The meter is calibrated to a standard 33% and 75% salt solution bottle. CAUTION: Do not calibrate the humidity without the default calibration
salt. Otherwise, permanent damage could occur. Contact the Triplett for calibration salts or services.
33% calibration
1. Plug the sensor probe into 33% salt bottle.
2. In the normal mode, Press and Hold the CAL/Esc and DP/WBT/▼ buttons to enter the
33% calibration. “CAL” and calibrating value
(32.7% if at 25°C) will blink on the LCD with current temperature at the left.
3. The meter is now calibrating, and will finish
in about 60 minutes when “CAL” and
humidity stop blinking.
Page 6
75% calibration
1. After the 33% calibration, plug the sensor probe into a 75% salt
bottle.
2. Press the MN/AV button to enter 75% calibration.
3. “CAL” and calibrating value (75.2% if at 25°C) will blink on the LCD with current temperature at the left.
4. The meter is now calibrating. Wait about 60 minutes until the blinking stops, then calibration is completed and the meter will return to normal mode.
NOTE: Single point calibrations are allowed. To calibrate 33% only, press
CAL/Esc and exit when 33% calibration is completed. To calibrate 75% only, press MODE/▲ within the 5 minutes while initializing 33% calibration.
Page 7
Specifications
Function
Range
Resolution
Accuracy
0 to 5000ppm
1ppm
±(%5rdg + 30ppm)
CO2
5000 to 9999ppm
1ppm
Not specified
Pressure dependence: +1.6% reading per kPa deviation from
normal pressure, 100kPa
Temperature
-10 to 60°C
0.1°
±0.6°C/0.9°F
14 to 140°F
Humidity
0.0 to 99.9%
0.1%
±3%(10 to 90%)
±5%(< or > 10 to 90%)
Wet Bulb
-5 to 60°C
0.1°
23 to 140°F
Calculated from RH and
Dew Point
-20 to 60°C
0.1°
Temperature
-4 to 140°F
Display
Triple LCD with backlight
Sensor Type
CO2: NDIR (non-dispersive infrared) technology
Humidity: Capacitance sensor;
Temperature (air): Thermistor
Operating Conditions
0 to 50oC (32 to 122oF); < 95% RH non­condensing
Storage Conditions
-20 to 60oC ( -4 to 140oF); <99% RH non­condensing
Power Supply
4 x 1.5V ‘AA’ batteries or AC adaptor (9V/1A)
Battery Life
approx. 24 hours (alkaline batteries)
Dimensions / Weight
200x70x57mm (7.9x2.7x2.3”)/190g (6.7 oz.)
Maintenance
CLEANING AND STORAGE
1. The meter should be cleaned with a damp cloth and mild detergent when necessary. Do not use solvents or abrasives.
Page 8
2. Store the meter in an area with moderate temperature and humidity (refer to the operating and storage range in the specifications chart earlier in this manual).
TROUBLESHOOTING
Can’t power on
Press SET for more than 0.3 seconds and try again. Check that the batteries have good contact and the correct polarity or that the AC adaptor is properly connected.
Slow response
Check whether the air flow channels on the rear of the meter are blocked.
Error messages
E01: CO2 sensor damaged. E02: The value is under range. E03: The value is over range. E04: The original data error results in this error (DP, WB) E07: Too low voltage to measure CO2, replace batteries or use an adaptor. E11: Retry humidity calibration. E17: Retry CO2 calibration. E31: Temperature sensor damaged. E34: Humidity sensor damaged.
Page 9
CO2 Levels and Guidelines
Non-Enforced Reference levels:
250 - 350 ppm – background (normal) outdoor air level
350- 1,000 ppm - typical level found in occupied spaces with good air exchange. 1,000 – 2,000 ppm - level associated with complaints of drowsiness and poor air. 2,000 – 5,000 ppm – level associated with headaches,
sleepiness, and stagnant, stale, stuffy air. Poor concentration, loss of attention, increased heart rate and slight nausea may also be present. >5,000 ppm – Exposure may lead to serious oxygen deprivation resulting in permanent brain damage, coma and even death.
Regulatory exposure limits:
ASHRAE Standard 62-1989: 1000ppm: CO2 concentration in occupied building should not exceed 1000ppm.
OSHA: 5000ppm: Time weighted average over five 8-hour work days should not exceed 5000ppm
Building bulletin 101 (Bb101): 1500ppm. UK standards for schools say that CO2 at averaged over the whole day(i.e. 9am to 3.30 pm) should not exceed 1500ppm.
Germany, Japan, Australia, UK...: 5000ppm, 8 hours weighted average in occupational exposure limit is 5000ppm
Warranty
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase. This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries are not covered by this warranty