Triplett 9065 Instruction Manual

Page 1
Model 9065
True RMS Data Logging MultiMeter
Instruction Manual
84-910 12/15
Page 2
Contents
I. Introduction----------------------------------------------------------1 II.Included Items------------------------------------------------------1 Ill. Rules for Safe Operation---------------------------------------2 IV. Electrical Symbols-----------------------------------------------3 V. Meter Structure----------------------------------------------------4 VI. LCD Display-------------------------------------------------------5 VII. Keys, Rotary Switch and Input Terminals-----------------6 VIII. Technical lndex--------------------------------------------------8
1. General Specications---------------------------------------8
2. Electrical Specications-------------------------------------9
IX. Measurement Operation--------------------------------------17
1. Meter Power Control----------------------------------------17
2. Meter Settings------------------------------------------------18
3. AC Voltage----------------------------------------------------20
4. DC Voltage----------------------------------------------------20
5. AC and DC Current-----------------------------------------21
6. Resistance----------------------------------------------------22
7. Conductance-------------------------------------------------24
8. Capacitance--------------------------------------------------24
9. Continuity------------------------------------------------------25
10. Diode Test----------------------------------------------------25
11. Frequency/Duty Cycle Measurement/Pulse Width-27
12. Temperature------------------------------------------------27
13. LPF-----------------------------------------------------------28
14. dBV-----------------------------------------------------------29
15. dBm----------------------------------------------------------29
16. Maximum Value and Minimum Value----------------29
17. Relative Value Mode-------------------------------------30
18. Peak Detection Mode------------------------------------30
19. Compare Mode--------------------------------------------30
20. Recording Measurement Data------------------------31
21. Communication-------------------------------------------33 X. Maintenance and Repair-------------------------------------34
1. General Maintenance and Repair---------------------34
2. Testing Fuses-----------------------------------------------34
3. Replacing Fuses-------------------------------------------35
4. Battery Charge---------------------------------------------35 X. Warranty-----------------------------------------------------------36 X. Notes---------------------------------------------------------------37
Page 3
I. Introduction
The Model 9065 is a 60000 count, or 4 5/6 digit,
handheld auto-ranging true RMS Digital Multimeter (hereinafter referred to as “the meter’’). The meter can be used to accurately measure AC and DC voltage, AC and DC current, resistance, conductance, capacitance, tem­perature, frequency and pulse width. It can also perform diode testing, continuity testing, and it provides useful test functions such as data hold, maximum/minimum/average measurements, comparison measurement, relative mea­surement, peak detection, trend capture and data record/ readback of as many as 20,000 measurements.
This Operating Manual provides information on safe and recommended uses of this product. Please read the rel­evant information carefully and observe all the Warnings and Notes.
Warning:
Please read the “Rules for Safe Operation” before using the meter.
II. Included Items
Open the package and the carrying case and inspect the contents. Please check that all of the following items are included:
1. One operating manual
2. A pair of test leads
3. Two K type temperature probes
4. One temperature connector
5. One charging adapter
6. One charging connector
7. One USB cable
8. One CD
9. One cloth bag
If you discover any missing or damaged pieces, please contact Triplett at 1-800 Triplett or support@triplett.com.
1
Page 4
III. Rules for Safe Operation
Please note the Warning symbol. Warnings indicate the conditions and actions which pose hazards to users or may damage the meter or equipment under test. This Meter is designed to comply with several safety standards, This meter conforms to UL STD. 61010-1,
61010-2-030, 61010-2-032, 61010-2-033, Certied to CSA
SID. C22.2 NO. 61010-1,61010-2-030, IEC STD 61010­2-032, 61010-2-033 in pollution degree 2, measurement category (CAT III 1000V, CAT IV 600V), double insulation as well as with IP65 standards for water resistance and dustproong. Use of the meter without following the oper­ating instructions can reduce or circumvent the protection provided by the meter.
Measurement Category Ill (CAT Ill) is intended for measurements performed in building installation. Examples may include measurements on distribution boards, circuit­breakers, wiring, including cables, bus-bars, junction boxes,
switches, socket-outlets in xed installations, and equipment
for industrial use and some other equipment, for example,
stationary motors with permanent connection to the xed
installation.
Measurement Category IV (CAT IV) is intended for measurements performed at the source of the low-voltage installation. Examples include electricity meters and mea­surements on primary overcurrent protection devices and ripple control units.
1. Never use a damaged meter. Before using the meter, check the meter case for visible cracks or any missing plastic parts. Pay special attention to the insulation areas around the connectors.
2. Before using the meter, ensure the battery cover is closed and secured. Before opening the battery cover, remove the test leads from the meter.
3. Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. If any test lead is damaged, replace it before using the meter.
4. Do not apply more than the rated voltage, as marked on the meter, between the terminals or between any terminal and ground.
5. Do not have the meter turned on when removing the cover or opening the case.
6. When working with a voltage >30V AC (RMS), >42VAC (peak), or over 60V DC, special care should be taken to avoid electric shock.
7. Replacement fuses must comply with the specications in
this Operating Manual.
8. Use the proper terminals, function and range for your mea­surement.
9. When measuring current, shut off the power to the circuit,
then connect the meter to the circuit. Remember: the meter and the circuit must be connected in series.
10. When making electrical connections, connect the com­mon test lead before connecting the main test lead; when disconnecting, disconnect the main test lead before discon­necting the common test lead.
2
Page 5
11. If the meter works improperly, do not use it. The protection measures of the meter may have failed. If in doubt, contact your distributor or Triplett Customer Service.
12. Do not store or use the meter in an environment with high temperature, humidity, strong magnetic elds, or with inam­mable or explosive materials.
13. When using the probes, keep your ngers behind the nger guards.
14. Do not use the low-pass lter to verify hazardous voltage,
as there may be an AC voltage over the indicated value. First,
measure the voltage without a lter to check whether it is hazardous, then select low-pass ltering.
15. Before testing resistance, continuity, conductance, diode or capacitance, shut off the power to the circuit and discharge all high-voltage capacitors.
17. Do not measure a voltage or current higher than the allowed input values. When the range of measured values cannot be determined, set the functional range switch at the maximum-range position.
18. When the symbol shows on the LCD display, the battery should be promptly charged to ensure measurement accuracy.
19. Do not change the wiring within the meter to avoid injury
and damaging the meter.
20. A soft cloth and mild detergent should be used to clean the surface of the meter when servicing. No abrasive clean­ers or solvents should be used on the product.
21.Test a known voltage before use to conrm the meter is
functioning correctly.
22. If you want to replace the test lead, you need to replace it with a test lead of the same or higher grade of CAT Ill 1000V/CAT IV 600V.
Dangerous Voltage
When the meter detects voltage ≥ 30V or overload (OL),
the symbol will display as a warning for potential hazard­ous voltage.
Double Insulated
Grounding
Warning
AC (Alternating Current)
DC (Direct Current)
Continuity Buzzer
Diode
Capacitance
AC or DC (Alternating Current or Direct Current)
Danger High Voltage
Conforms to Standards of European Union.
This symbol signies the product complies with both
USA and Canadian electrical requirements.
IV. Electrical Symbols
3
Page 6
V. Meter Structure (see Figure 1)
1
Case
2
LCD
3
Function Keys
4
Rotary Switch
5
Input Terminals
4
Page 7
V. LCD Display (see Figure 2)
No. Function Description
1
Label of Function Keys
Indicates auxiliary functions under current measuring interface
2
Simulation Bar Analog display of input signals
3
Minus Sign Indicates negative reading
4
Lightning Symbol Danger- High Voltage
5
Time and Date Indicates the time and date set in internal
clock
6
Small Measurements
Real-time input values are displayed when the primary and secondary displays are covered by menus or pop-up messages.
7
Communications Indicates the USB communication is
enabled.
8
Sound Setting Indicates the button sound is enabled
(beeps when a button is pressed)
9
Battery Level Indicates the charge level of rechargeable
battery
10
Range Indicator Indicates the current range and range
mode for the meter (auto or manual)
11
Unit Indicates the measurement unit
12
Assisted Function Mode
Indicates the active assisted measuring function, such as Low Pass Filter
13
Assisted Function Display
Shows additional measuring information from the assisted function mode that is active.
5
Page 8
VII. Keys, Rotary Switch and Input Terminals
(1) Keys
The 14 keys on the meter are used to activate the alternate functions of the rotary switch, browse menus or control the me­ter power. The keys shown in Figure 3 are described in the following table.
Key Function
Turn on or turn off the power to the meter
F1 F2 F3 F4 Select the sub-functions and modes indicated by the labels on the LCD screen.
Cursor keys are used to select menu items, scroll through text and input data
HOLD Keep the current reading held on the display
RANGE Switch the range mode of the meter to Manual mode, then switch between all the available ranges. Long
press the button to return to Automatic range mode.
MAX MIN Starts to record Minimum & Maximum values.
SELECT Press to select the alternate function. Long press to enter Help Menu.
Press the key to switch the backlight brightness. Long press it to turn off backlight.
Knob Function
Measurement of AC Voltage
Measurement of mV in AC and mV in AC+DC
Measurement of voltage in DC and AC+DC
Measurement of mV in DC and temperature
Measurement of resistance, continuity and specic conductance
Diode test and capacitance measurement
(2) Rotary Switch (see Figure 4)
6
Page 9
Knob Function
Hz% mS-Pulse Measurement of frequency, duty cycle and pulse width
µA Measurement of A in AC, DC and AC+DC
mA Measurement of mA in AC, DC and AC+DC
A Measurement of ampere in AC, DC and AC+DC
Battery is charging
Terminal Description
A Measurement of frequency, duty cycle and pulse width
mA µA Measurement of A in AC, DC and AC+DC
COM Measurement of mA in AC, DC and AC+DC
VΩ °C Hz Measurement of ampere in AC, DC and AC+DC
(3) Input Terminals
In addition to the above, all four terminals are used for charging and temperature measurement through corresponding adapters .”Lead Error!” will show on the display to warn if the probes are inserted improperly.
7
Figure 3 & 4
Page 10
VIII. Technical Index
1. General Specications
The maximum voltage between any terminal and ground: 1000 V The fuse protection of mA or µA input terminals: 0. 8A H 1000V Fuse Type 6X32mm The fuse protection of the A input terminal: 10A H1000V Fuse Type 10X38mm Max. display: 60000 Range: Auto/Manual Polarity: Auto Operating temperature: -20°~50°C Storage temperature: -30°C~60°C
Relative humidity: ≤80% (0°C~30°C), ≤75% (30°C~40°C), ≤45% (40°C~50°C) Electromagnetic compatibility: In an RF E-eld of 1V/m: Overall accuracy= specied accuracy+ 5% of range. No specied ac­curacy guarantee for RF E-eld strength > 1V/m.
Operating altitude: 0~2000m
Temperature coefcient: 0.1X (specied accuracy)/ °C (<18°C or >28°C) Internal battery: Lithium battery of 7.4V 2200mAh
Power adapter: Input of 100V~240V,50/60Hz 0.2A max, Output of DC10V 500mA(short-circuit protection for output). External diameter of 5.5 mm and internal diameter of 2.5mm. Low battery: The symbol shows on the LCD. Dimension: About (225 X100 X 60) mm Weight: About 608g (including battery) Safety standards: IEC/EN61010-1, EN61010-2-030, EN 61010-2-033 in pollution degree 2, CAT III 1000V,CAT IV 600V IP65 standards for waterproof and dustrproof rating
8
Page 11
2. Electrical Specications
Accuracy: ±(% Reading + Digits), one-year calibration period.
Ambient temperature: 23°C± 5°C;
Ambient humidity: 75%RH; Temperature coefcient: 0.1x (Accuracy)/ °C (<18 °C or >28°C)
2. Electrical Specications
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
60mV 0.001mV
45~1kHz 1k~10kHz 10k~20kHz 20k~100kHz
±(0.6%+60) ±(1.2%+60) ±(3%+60) ±(4%+60)
600mV 0.01V
45~1kHz 1k~10kHz 10k~20kHz 20k~100kHz
±(0.3%+30) ±(1.2%+40) ±(3%+40) ±(4%+40)
6V 0.0001V
45~1kHz 1k~10kHz 10k~20kHz 20k~100kHz
±(0.3%+30) ±(1.2%+40) ±(3%+40) ±(4%+40)
60V 0.001V
45~1kHz 1k~10kHz 10k~20kHz 20k~100kHz
±(0.3%+30) ±(1.2%+40) ±(3%+40) ±(4%+40)
600V 0.01V
45~1kHz 1k~10kHz 10k~20kHz 20k~100kHz
±(0.3%+30) ±(1.2%+40) ±(3%+40) Only for reference
1000V 0.1V
45~1kHz 1k~5kHz 5k~10kHz 10k~100kHz
±(0.6%+30) ±(3%+40) ±(6%+40) Only for reference
• Input impedance: About 10MΩ
• Overload protection: 1000V
• Display: True RMS value for 10% to 100% of the range.
9
Page 12
(2) DC Voltage
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
60mV 0.001mV ± (0.025%+20)
600mV 0.01mV
± (0.025%+5)6V 0.0001V
60V 0.001V
600V 0.01V
± (0.03%+5)
1000V 0.1V
(3) AC Voltage + DC Voltage
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
60mV 0.001mV
50~1kHz 1k~10kHz 10k~35kHz
+(1%+80) +(3%+40) +(6%+40)
600mV 0.01V
50~1kHz 1k~10kHz 10k~35kHz
+(1%+80) +(3%+40) +(6%+40)
6V 0.0001V
50~1kHz 1k~10kHz 10k~35kHz
+(1%+80) +(3%+40) +(6%+40)
60V 0.001V
50~1kHz 1k~10kHz 10k~35kHz
+(1%+80) +(3%+40) +(6%+40)
600V 0.01V
50~1kHz 1k~10kHz 10k~35kHz
+(1%+80) Only for reference Only for reference
1000V 0.1V
50~1kHz 1k~10kHz 10k~35kHz
+(1.2%+80) Only for reference Only for reference
• Input impedance: About 10MΩ
• Overload protection: 1000V
• Display: True RMS value for 10% to 100% of the range.
10
Page 13
• Input impedance: About 10MΩ
• Overload protection: 1000V
• Display: True RMS value for 10% to 100% of the range.
(4) AC Current
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
600µA 0.01µA
45~1kHz 1k~10kHz
+(0.6%+40) +(1.2%+40)
6000µA 0.1µA
45~1kHz 1k~10kHz
+(0.6%+20) +(1.2%+40)
60mA 0.001mA
45~1kHz 1k~10kHz
+(0.6%+40) +(1.2%+40)
600mA 0.01mA
45~1kHz 1k~10kHz
+(0.6%+20) +(1.2%+40)
10A 0.001A
45~1kHz 1k~10kHz
+(1%+20) +(3%+40)
• Display: True RMS value for 10% to 100% of the range.
• Overload protection: µAmA range: 0.8A H 1000V Fuse Type ɸ 6x32 mm 10 A range: 10A H 1000V Fuse Type ɸ 10x38mm
• Switch on for 30 seconds and suspend measurement for 10 minutes for 20A. Not specied for over 10A.
11
Page 14
(5) DC Current
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
600µA 0.01 µA +(0.08%+20)
6000µA 0.1 µA + (0.08%+10)
60mA 0.001 mA + (0.08%+20)
600mA 0.01 mA +(0.15%+10)
10A 0.001A +(0.5%+10)
• Overload protection: µAmA range: 0.8A H 1000V Fuse Type ɸ 6x32 mm 10 A range: 10A H 1000V Fuse Type ɸ 10x38mm
• Switch on for 30 seconds and suspend measurement for 10 minutes for 20A. Not specied for over 10A.
(6) AC Current + DC Current
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
600µA 0.01µA
50~1kHz 1k~10kHz
+(0.8%+40) +(2.0%+40)
6000µA 0.1µA
50~1kHz 1k~10kHz
+(0.8%+20) +(2.0%+40)
60mA 0.001mA
50~1kHz 1k~10kHz
+(0.8%+40) +(2.0%+40)
600mA 0.01mA
50~1kHz 1k~10kHz
+(0.8%+20) +(2.0%+40)
10A 0.001A
50~1kHz 1k~10kHz
+(1.2%+20) +(3%+40)
12
Page 15
• Display: True RMS value for 10% to 100% of the range.
• Overload protection: µAmA range: 0.8A H 1000V Fuse Type ɸ 6x32 mm 10 A range: 10A H 1000V Fuse Type ɸ 10x38mm
• Switch on for 30 seconds and suspend measurement for 10 minutes for 20A. Not specied for over 10A.
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
600Ω 0.01Ω In REL state: ±(0.05%+10)
6kΩ 0.0001kΩ
+(0.05%+2)60kΩ 0.001kΩ
600kΩ 0.01kΩ
6MΩ 0.0001MΩ +(0.3%+10)
60MΩ 0.001MΩ +(0.3%+10)
(7) Resistance
• Overload protection: 1000V
• Humidity for 60 MΩ: <50%
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
60ns 0.01nS +(2%+10)
(8) Conductance
• Overload protection: 1000V
• Humidity: <50%
13
Page 16
(9) Capacitance
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
6nF 0.001nF +(3%+10)
60nF 0.01nF +(2.5%+5)
600nF 0.1nF
+(2%+5)
6µF 0.001µF
60µF 0.01µF
600µF 0.1µF
6mF 1µF +(5%+5)
60mF 10µF Not specied
• Overload protection: 1000V
• Display digits: 6000
(10) Temperature
Range Resolution Accuracy
-40°C~40°C
0.1°C
+(2.0%+30)
40°C~400°C +(1.0%+30)
400°C ~1000°C +2.5%
-40°F~104°F
0.2°F
±(2.5%+50)
104°F~752°F +(1.5%+50)
752°F~1832°F +2.5%
Overload protection: 1000V
• Two-channel temperature measurement can be performed via temperature connectors.
• Temperature sensor: Applicable to K type(chromel-silicel) thermocouple. Spare parts are point contact Ktype (chromel-silicel) thermocouple (only applicable to the measurement when temperature is below 800°C)
14
Page 17
(11) Frequency
Range Resolution Accuracy
60Hz 0.001 Hz +(0.02%+8)
600Hz 0.01 Hz
+(0.01%+5)
6kHz 0.0001kHz
60kHz 0.001kHz
600kHz 0.01kHz
6MHz 0.0001MHz
60MHz 0.001MHz
• Overload protection: 1000V
• Display digits: 6000
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
10%~90%(10Hz~2kHz) 0.01% +(1.2%+30)
(12) Duty Cycle
• Overload protection: 1000V
• Humidity: <50%
Range Resolution Accuracy Tolerance: ± (% Reading + Digits)
250mS 0.001mS~0.01mS +(1.2%+30)
(13) Pulse Width
• Overload protection: 1000V
• When the rise time is less than 1 µs, the signals center on trigger level.
• The pulse width is greater than 2 µs for 10Hz to 200kHz. The pulse width depends on signal frequency.
15
Page 18
Range Resolution Remark
0.01Ω
Open circuit voltage is around 3V.
When the buzzer is set for Short Circuit warning; If the impedance tested is
less than 10Ω, the buzzer continuously sounds. If the impedance tested is
greater than 50Ω, the buzzer does not sound.
When the buzzer is set for Open Circuit warning; If the impedance tested is
greater than 50Ω, the buzzer continuously sounds. If the impedance tested is
less than 10Ω, the buzzer does not sound.
(14) Continuity Test
• Overload protection: 1000V
Range Resolution Remark
0.0001V
Open circuit voltage is around 3V. The forward voltage drop value of the mea-
sured PN junction is approximately ≤ 3V.
When the buzzer is selected in Diode Test Mode; it will beep briefly for the
normal semiconductor junction. If the semiconductor junction shorts out (im-
pedance <10Ω), the alarm will continuously sound. Typical silicon PN junction
voltage drops vary between 0.5~0.8V.
(15) Diode Test
• Overload protection: 1000V
16
Page 19
IX. Measurement Operation
1. Meter Power Control
1) To manually power up and power down the meter:
When the meter is off, long press to start the meter. When the meter is on, long press to shut it off. The meter cannot be powered off when charging.
2) Indicators for Battery Level:
The meter is powered by a rechargeable lithium battery. The indicators for battery level are in the upper right corner of the
When the battery capacity is lower than 3% of full capacity, the meter will automatically shut down.
3) Backlight Control
If the display is not visible in low-light situations, press the backlight button to switch the backlight brightness. Long press the button to turn off the backlight and enter power saving mode. When the backlight is off, the green light
ashes to indicate that the meter is still collecting data. Press any key or turn the rotary switch to turn on the backlight again.
4) Auto Power-off
When the Auto Power Off is set, the meter will turn off if no button is pressed or the dial selector is not turned within the time set. To disable Auto Power Off, set the Power Off mode to “OFF” under the SETUP – Auto Power Save Menu.
5) Power Saving Mode
Set the control time for dimming the brightness and turning off the display, found in the SETUP – Auto Power Save Menu
17
Page 20
2. Meter Settings
Press the function key labelled, SETUP to access the general settings for the meter. Press the cursor keys to navigate through the settings menu.
1) Keypad Tone
This function enables or disables sound when a button is pressed. The beeper symbol in the upper right corner indicates that sound is enabled.
2) Lead Alarm Buzzer
This function enables or disables the alarm sound for improper probe lead connection.
3) Communication
Set as ON to enable USB communication. A symbol will appear in the upper right corner when active. When Communication Mode is OFF, the symbol will disappear.
4) Date & Time
With “Date & Time” highlighted in the SETUP Menu, press the function key labelled, “SET” to set the date and time. Press (LEFT) or (RIGHT) to select the eld to edit, and press (UP) or (DOWN) to enter different numbers, then press the
function key “OK” to conrm. To cancel the settings, press the function key labelled, “CANCEL”.
5) Auto Power Save Options
With “Auto Power Save” highlighted in the SETUP Menu, press the function key labelled, “SET” to edit the time of inactivity to dim the brightness, turn off the display, and power down the meter. Press (UP) or (DOWN) to move the cursors to select different items. Press (LEFT) or (RIGHT) to enter the time, in minutes, to activate the power saving feature.
Menu Item Description Set Value
Brightness Down
Dim the display brightness after the elapsed
time of inactivity.
ON: 1-60 Min OFF: This function is disabled
Display Off Turn off the display after the elapsed time. ON: 1-60 Min OFF: This function is disabled
Power Off Turn off the power after the elapsed time. ON: 1-60 Min OFF: This function is disabled
Press the function key “OK” to conrm the above settings. To cancel the settings, press the function key “CANCEL”.
18
Page 21
6) More Settings
With “More Settings” highlighted in the SETUP Menu, press the function key “ENTER” to: Set the language of the Help Menu, Format Memory, Reset the meter settings, or check the model number, serial number and available memory space.
Press the cursor keys to select between the following menu items.
- Help Menu Language Press the function key labelled, “SET” to set the Help Menu language. Press (UP) or (DOWN) to select a different lan
guage. Then press the function key “OK” to conrm. To cancel the settings, press the function key “CANCEL”.
- Memory Format
With “Memory Format” highlighted, press the function key labelled, “FORMAT” to enter the memory format menu, then
press the function key labelled “YES” to conrm. To cancel the format, press the function key labelled, “NO”.
- Reset All Setting
With “Reset All Setting highlighted, Press the function key “RESET” to reset the meter to default settings, then press the
function key “YES” to conrm. To cancel the reset, press the function key “NO”.
- About
Press the function key “ABOUT” to check product model, serial number and available memory space.
19
Page 22
3. AC Voltage
1) Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2) Set the rotary switch to V or mV as shown in Figure 5. Connect the test leads to the power or load under test in parallel.
3) Directly read the measured voltage values on the display. True virtual values are displayed for AC measurement.
4) Press the function key labelled, “MENU” to access additional features for measuring AC voltage. Press
the cursor keys to select menu items. The red highlighting indicates the selected item. Press F1 to enter the corresponding measuring mode, press F2 to enter Relative value mode when available, and press F3 to set dBm impedance (AC V only), and press F4 to close the measurement features menu.
Attention:
After completing all the measuring operations, discon­nect the test leads and the circuit under test Do not input a voltage higher than 1000V. Higher voltage may be measured but it may damage the meter.
• When measuring high voltage, special care should be taken to avoid electric shock.
• After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
• The meter’s True RMS response is calibrated to an AC si­nusoidal input signal. Measuring non-sinusoidal AC signals
will have decreased accuracy.
4. DC Voltage
1) Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2) Set the rotary switch to DC V or DC mV as shown in Figure 6. Connect the test leads to the power or load under test in parallel.
3) Directly read the measured voltage values on the display.
4) Press the function key labelled, “MENU” to access
additional features for measuring DC voltage. Press the cursor keys to select menu items. The red highlighting indicates the selected item. Press F1 to enter the corresponding measuring mode, press F2 to enter Relative value mode when available, and press F4 to close the measurement features menu.
20
Page 23
Attention:
• Do not input a voltage higher than 1000V. Higher voltage may be measured but it may damage the meter.
• When measuring high voltage, special care should be taken to avoid electric shock.
• After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
5. AC and DC Current
1) Insert the red test lead into the µAmA or A terminal and the black test lead into the COM terminal.
2) Set the rotary switch to the position as shown in Figure 7. Press the key “SELECT” to measure either AC or DC. Connect the test leads to the test circuit in series.
3) Directly read the measured current values on the display.
4) Press the function key labelled, “MENU” to access
additional features for measuring AC or DC current. Press the cursor keys to select menu items.
The red highlighting indicates the selected item. Press F1 to enter the corresponding measuring mode, press F2 to enter Relative value mode when available, and press F4 to close the measurement features menu.
Warning:
• Before connecting to the test circuit, turn off the power to
the circuit rst and discharge all the high-voltage
capacitors.
• Use proper input terminals and functions for measurement.
If the current amperage cannot be estimated, rst measure
using the largest range.
• When the test lead is inserted in a current terminal, do not connect its test probe to any circuit in parallel, it could blow the fuses within the meter and damage the meter.
• After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
21
Page 24
22
Page 25
6. Resistance
1) Insert the red test lead into the Ω terminal and the black
test lead into the COM terminal.
2) Set the rotary switch to the Ω nS position. The
meter will default to resistance measurement. Connect the test leads to both ends of the measured resistance as shown in Figure 8.
3) Directly read the measured resistance values on the display.
Attention:
• If measuring an open circuit or resistance value exceeds the maximum range of the meter, “OL” will show on the display.
• When measuring the in-circuit resistance, all the power within the measured circuit must be shut off before measurement and all the capacitors must be discharged to ensure a correct measurement, and to avoid damaging the meter.
• When measuring low resistance, the test leads may
add additional resistance between 0.10 Ω to 0.20Ω. To
obtain accurate readings, Relative measurement can be used. First short the input test leads, then press the function key labelled, “MENU” and press F2 to enter Relative measurement mode. Perform the low resistance measurement and the meter will automatically subtract the resistance of the shorted test leads.
• If the resistance value is greater than 0. 50Ω with shorted
test leads, the test leads should be checked for loose connections or other factors.
• When measuring a resistance >1MΩ, the readings require
a few seconds to stabilize. This is normal for high resis tance measurements. In order to obtain stable readings, short test lines can be used for the measurement.
• Do not input a voltage in resistance measurement mode to avoid possible product damage or personal injury.
• After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test
23
Page 26
7. Conductance
1) Insert the red test lead into the Ω terminal and the black
test lead into the COM terminal.
2) Set the rotary switch to the measurement Ω nS .
Press the SELECT key twice to select Conductance measurement. Connect the test leads to both ends of the measured resistance as shown in Figure 8.
3) Directly read the measured conductance value on the display.
Attention:
• When measuring the in-circuit resistance, all the power within the measured circuit must be shut off before measurement and all the capacitors must be discharged to ensure a correct measurement, and to avoid damaging the meter.
• Do not input a voltage when in conductance measurement mode to avoid possible product damage or personal injury.
• After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
8. Capacitance
1) Insert the red test lead into the terminal and the black test lead into the COM terminal.
2) Set the rotary switch to the measurement press the SELECT key to select Capacitance measurement. Connect the test leads to both ends of the
measured capacitance as shown in Figure 9.
3) Directly read the measured capacitance value on the display.
Attention:
• If the measured capacitance shorts or capacitance value exceeds the maximum range of the meter, “OL” will show on the display.
• For the measurement of small capacitance within range, relative measurement REL should be used to avoid the
inuence of test lead capacitance for correct readings.
• For capacitance measurements > 600 µF, the readings require a few seconds to be stable.
• To ensure the measuring accuracy, the capacitor should be discharged completely then put onto the meter to measure, especially for a capacitor with high voltage to avoid damage to the meter and personal injury.
• Do not input a voltage when in capacitance measurement mode to avoid possible product damage or personal injury.
• After completing all the measuring operations, disconnect the connection between the test leads and the capacitor under test.
24
Page 27
9. Continuity Test
1) Insert the red test lead into the Ω terminal and the black
test lead into the COM terminal.
2) Set the rotary switch to the measurement Ω nS ,
press the SELECT key to select the Continuity test . Connect the test leads to both ends of the measured resistance as shown in Figure 8. Press the function key labelled, “MENU” to select measurement options. Pressing the function key labelled, “SHORT” will put the meter in Short Circuit Alarm mode. Pressing the function key labelled, “OPEN” will put the meter in Open Circuit Alarm mode.
When the buzzer is set for Short Circuit warning; If the
impedance tested is less than 10Ω, the buzzer continuously sounds. If the impedance tested is greater than 50Ω, the
buzzer does not sound.
When the buzzer is set for Open Circuit warning; If the
impedance tested is greater than 50Ω, the buzzer continu­ously sounds. If the impedance tested is less than 10Ω, the
buzzer does not sound.
3) Directly read the measured resistance value on the display.
Warning
• When measuring the in-circuit resistance, all the power within the measured circuit must be shut off before measurement and all the capacitors must be discharged to ensure a correct measurement.
• Do not input a voltage when in continuity test mode to avoid possible product damage or personal injury.
• After completing all the measurements, disconnect the connection between the test leads and the circuit under test.
10. Diode
1) Insert the red test lead into the terminal and the
black test lead into the COM terminal. The polarity of red test lead is “+” and “-” for black test lead.
25
Page 28
2) Set the rotary switch to the position. The meter will default to Diode measurement. Connect the test leads to both ends of the diode as shown in Figure 10. Directly read the approximate forward PN junction voltage of the measured diode on the display.
3) Press the function key labelled, “MENU” to choose measurement options. Pressing the function key labelled,“ALARM” will enable the beeper. It will beep
briey when connected to a normal semiconductor
junction; If the measurement shorts out, it will beep continuously. Typical Silicon PN junction drops vary between 0.5V ~ 0.8V. Pressing the function key labelled, “NORMAL” will run the diode test without the beeper function.
Attention:
• If the measurement is an open circuit, or the diode polarity is reversed, “OL” will be displayed.
• When measuring an in-circuit diode, all the power within the measured circuit must be shut off before measurement and all the capacitors must be discharged.
• Open-circuit voltage of diode test is around 3V.
• Do not input a voltage when in diode test mode to avoid possible product damage or personal injury.
• After completing all measurements, disconnect the connection between the test leads and the circuit under test.
26
Page 29
11. Frequency/Duty Cycle Measurement /Pulse Width
1) Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2) Set the rotary switch to the measurement Hz%
ms-Pulse position. Press the SELECT key to select Hz,
Duty Cycle (%), or ms-Pulse. Connect the test leads to
the signal source under test in parallel as shown in Figure 11.
3) Directly read the measured values of frequency, duty cycle or pulse width on the display.
Attention:
• Simulation bar displays the frequency of the measured signal for duty cycle and pulse width.
• After completing all measurements, disconnect the connection between the test leads and the circuit under test.
12. Temperature
1) Set the rotary switch to the measurement mV °C °F then press the SELECT key to choose Celsius °C or Fahrenheit °F. Insert the temperature connector into the four input terminals and connect two temperature probes to the temperature connector as shown in Figure 12. The temperature probes detect the surface temperature of the object under test.
2) Directly read the temperature values of the two sensors on the display.
3) Press the function key labelled, “MENU” to choose measurement options. Scroll through the menu with
keys and press F1 to choose the highlighted option. Press F2 while in the Menu to enter Relative Mode. Press F4 to close the Menu.
27
Page 30
Attention:
• If the ambient temperature for the meter is outside the range of 18 °C to 28 °C, it may cause measurement errors. The measurement effects are more obvious at low temperatures.
• Remove the temperature probes after completing all the measurements.
• Point contact K type (chromel-silicel) thermocouple
accuracy specication is for temperature measurements
below 230 °C.
13. LPF Measurement
1) Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2) Set the rotary switch to v. Connect the test leads to the power or load to be tested in parallel as shown in Figure 5.
3) Press the function key labelled, “MENU” to choose measurement options. Press the cursor keys
to highlight “Low Pass”, then press the function key labelled, “Low Pass” to enter LPF measuring mode.
4) The meter performs the measurement in AC mode.
The AC signals go through a lter which lters out
voltage signals with frequencies higher than 1KHz. As
shown in the following gure, the low-pass lter can
measure the composite signals of sinusoidal waves generated by inverter and variable-frequency motor.
Attention:
• To avoid electric shock or personal injury, do not use the
low-pass lter to verify hazardous voltage, there may be
a voltage over the indicated value. First, measure the
voltage without a lter to check whether it is hazardous, then select low-pass ltering.
28
Page 31
• In the LPF measuring mode, the meter will turn to manual mode. Press the RANGE key to select a range. When the
low-pass lter is enabled, automatic range is unavailable.
• Do not input a voltage above 1000V. Higher voltage may be measured but it poses the risk of damaging the meter.
• After completing all measurements, disconnect the connection between the test leads and the circuit under test.
14. dBV
1) Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2) Set the rotary switch to v. Connect the test leads to the power or load to be tested in parallel as shown in Figure 5.
3) Press the function key labelled, “MENU” to choose measuring options. Press the cursor keys to
highlight “dBV”, then press the function key labelled, “dBV” to enter dBV measurement mode.
4) dBV is the primary measurement shown and AC voltage value is the secondary value shown on the display. AC voltage of the measured signal is also displayed on the
15. dBm
1) Insert the red test lead into the V terminal and the black test lead into the COM terminal.
2) Set the rotary switch to V. Connect the test leads to the power or load to be tested in parallel as shown in Figure 5.
3) Press the function key labelled, “MENU” to choose measurement options. Press the cursor keys
to highlight “dBm”. A reference impedance must be used to calculate the dBm measurement. Press the function key labelled, “RES” to select a reference impedance value. Press (UP) or (DOWN) to scroll
between ten dened reference values: 4, 8, 16, 25, 32, 50, 75, 600, 1000Ω, and User Dened. While selecting a User dened reference impedance, press the function
key labelled, “EDIT” and use the cursor keys to input a reference impedance value between
4-1,200Ω. Press the function key labelled, “OK” to set
reference values. Press the function key labelled, “dBm” to enter dBm measurement.
4) dBm is the primary measurement shown and AC voltage value is the secondary value shown on the display. AC voltage of the measured signal is also displayed on the simulation bar.
16. Maximum and Minimum Value
Press the “MAX MIN” key to activate Max Min Mode. Real-time measurements are the primary value displayed. Maximum, minimum, and average values, as well as elapsed time, start date and times corresponding to the three values are secondarily displayed.
29
Page 32
Press the function key labelled, “EXIT” to exit Max Min Mode.
17. Relative Value Mode
Press the function key labelled, “MENU” to enter the measurement option menu. While in the menu, press the function key labelled, “REL” to enter Relative Value Mode*. The meter will store the value shown at the time that “REL” is pressed as the “Reference” measurement. Press the function key labelled, “REL” to store the present value as the “Reference”. The real-time measurement will be displayed as “Measurement”. The primary value displayed will be the Rela­tive value (Measurement – Reference). Press the “MENU” function key to select other modes and exit Relative Value Mode. *NOTE: When applicable. Relative Value Mode is not available for all measurement types.
18. Peak Detection Mode
While in a Voltage or Current measuring mode, press the “MENU” function key to choose measurement options. While in the menu, press the function key labelled, “PEAK” to acti­vate Peak Detection Mode. Response time is 1ms. Transient values can be more accurately measured by using Peak Mode while using the Record function.
19. Compare Mode
Press the F3 function key labelled, “COMP” to begin Com­pare Mode. Press (UP) or (DOWN) to scroll between the settings of Compare Mode. Press the “EDIT” function key to edit the highlighted Compare Mode setting.
1) Pass Mode
• INNER(Low Value ≤ Input Value ≤ High Value)
• OUTER (Input Value < Low Value OR Input Value >High
Value)
• <Value
• >Value
Press the “OK” function key to conrm the above settings.
To cancel the settings, press the “CANCEL” function key.
2) Beep Mode
While highlighting the “Beep Mode” setting, press the “EDIT” function key to edit the Beep settings. Press the keys
to select one of the three following options:
• PASS ON
The meter will beep when the Pass Mode criteria are met, and the meter displays, “PASS”.
• FAIL ON
The meter will beep when the Pass Mode criteria are NOT met, and the meter displays, “FAIL”.
• OFF
The meter will not beep in Compare Mode.
Press the function key OK to conrm the above
settings. To cancel the settings, press the “CANCEL” function key.
3) Low Value or High Value or Value
This setting controls the value(s) for the Pass Mode criteria. After completing the settings, press the “START” function key to start the measurement in Compare Mode. Press the “EXIT” function key to exit Compare Mode.
30
Page 33
20. Recording Measurement Data
Press the function key labelled, “SAVE” to enter the menu for saving, recording or viewing data. Press the cursor keys
to select the following options.
1) Save
While in the Save menu, press the “SAVE” function key to store the measurement currently displayed. Up to 20,000 measurements can be saved on the meter.
2) View Save
In the Save menu with “View Save” highlighted, press the
“VIEW” function key to view stored save les. Press or
hold the function keys labelled, “PREV” or “NEXT” to scroll through the saved data. Press the “DELETE” function key to delete the selected saved data. Press the “RETURN” func­tion key to return to the last menu. In addition to displaying
the saved data, the le save # and total number of saved les are displayed in the lower left corner. The date and time
of the saved data are displayed in the lower right corner.
3) Delete All Save
In the Save menu with “Delete All Save” highlighted, press the function key labelled, “DELETE” to delete all saved data
les. Press the function key labelled, “YES” to conrm. To
cancel the operation, press the function key labelled, “NO”.
4) Record
In the Save menu with “Record” highlighted, press the func­tion key labelled, “ENTER” to enter the continuous recording mode with a maximum of 10000 data points.
Press (UP) or (DOWN) to edit the three following settings.
• Edit Name
In the Record Menu with “Edit Name” highlighted, press the “EDIT” function key to edit the name of the recording. Press (LEFT) or (RIGHT) to move the cursor and select where to edit. Press the F1 key to switch the input to Uppercase mode, then press (UP) or (DOWN) to input uppercase letters. Press the F2 key to switch the input to Lowercase mode, then press (UP) or (DOWN) to input lowercase let­ters. Press the F3 key to switch the input to Digit mode, then press (UP) or (DOWN) to enter numbers or symbols. Press the F4 key to save the Record name and exit edit mode.
• Set Interval
With “Set Interval” highlighted, press the “EDIT” function key to set the time interval between recorded data points. Press or hold (LEFT) or (RIGHT) to move the cursors to select where to edit. Press or hold (UP) or (DOWN) to enter differ­ent numbers. The interval time can be set from 1 sec to 60 min between data points.
• Set Duration
With “Set Duration” highlighted, press the “EDIT” func­tion key to set the duration of the recording. Press or hold (LEFT) or (RIGHT) to move the cursors to select where to edit. Press or hold (UP) or (DOWN) to enter different numbers. The duration time can be set in days, hours and minutes.
• Max Duration
Displays the maximum time for continuous recording with the current settings. When ready to Record, press the “START” function key to start a recording with the current settings.
31
Page 34
When Recording, the characters “REC” show on the display
with a ashing red dot, as shown in Figure 13. The relevant
display information is shown in the following table.
To manually stop recording, press the “STOP” function key. After stopping the recording, the record data will be displayed. See the viewing options in the following “view Record” section.
5) View Record
In the Save Menu with “View Record” highlighted, press the “View” function key to view the information on recordings. The data will be displayed as shown in Figure 14. The basic display information is in the following table.
Information Description
Name Name of recording
Interval Time between data points
Duration Duration of recording from start to stop
Samples Total number of data points recorded
Maximum Maximum recorded value
Average Average value of the recorded data
Minimum Minimum recorded value
REC Record le number and total number of
recordings
Start Time and date at beginning of recording
Information Description
Elapsed Time Run time displayed in hours: minutes
Remaining Time
Time left displayed in hours: minutes: seconds
Samples Total number of recorded data points for
the active recording
Start Time and date of the start of the
recording.
32
Page 35
While viewing a Record, press the “PREV” function key to display the previous record. Press the “NEXT” function key to display the next record. Press the “RETURN” function key to return to the previous menu. Press the function key la­belled, “TREND” to view the recording graphically, as shown in Figure 15. The display information of trend data is in the following table:
No. Description
1 Measured value corresponding to cursor
2 Date and time of measurement at the cursor
3 Cursor
4 Trend line
5 Elapsed time labels on X-axis
6 Name of recording
7 Time and date at the start of the recording.
Viewing the trend chart: Press or hold the F2 key to move the cursor left. Press or hold the F3 key to move the cur­sor right. The cursor moves one data point for each press. Holding, or long pressing the keys will scroll through data points more quickly. Press (UP) or (DOWN) to scale the graph vertically. Press (LEFT) or (RIGHT) to scale the graph horizontally.
Press the F1 key to delete this record. Press the function
key labelled, “YES” to conrm the deletion. To cancel the
deletion, press the “NO” function key.
6) Delete All Record
In the Save menu with “Delete All Record” highlighted, press the function key labelled, “DELETE” to delete all record-
ings. Press the “YES” function key to conrm. To cancel the
operation, press the “NO” function key.
21. Communication
33
Page 36
USB Communication
Turn on communication via settings (see detailed operations in the section 2. Meter Settings). The symbol will appear in the upper right corner on the display as shown in Figure 15. The meter performs USB communication with the supplied USB cable (standard accessory) connecting to the PC.
X. Maintenance and Repair
1. General Maintenance and Repair
Regularly clean the meter case with damp cloth and mild de­tergent. Do not use abrasives, isopropyl alcohol or solvents. Dirt or moisture on the terminals can affect readings but can also enable the “Lead Error” warning. Clean the terminals according to the following steps:
1) Turn off the meter and remove all test leads.
2) Clean up the dirt on the terminals.
3) Soak a clean cotton swab in mild detergent and water.
Clean each terminal with the cotton swab. Dry each terminal with canned compressed air to force water and detergent to
ow out of the terminals.
4) In the case of abnormal meter function, stop using it and send it for repair.
5) When the meter needs to be veried or repaired, qualied
service personnel or a designated maintenance department are required to repair it.
2. Testing Fuses
As shown in Figure 17, put the meter in resistance mode. Insert a test lead into the terminal as shown in Figure 17 and contact the probe tip of the test lead with the metal in the of a current input terminal. If the message, “Lead Error!” (connection error for test leads) appears, it shows that the probe tip is inserted too deeply in the input current terminal. Draw out the test lead a little until the error message disap­pears and OL (overload) or resistance reading appears on the display. If the resistance reading for A terminal is less
than 0.50Ω, it shows that the fuse F2 is normal. If the
reading is “OL”, you need to replace F2; If the resistance
reading of the µAmA terminal is less than 1.2MΩ, it shows
that the fuse F1 is normal. If the reading is OL, F1 needs to be replaced;
34
Page 37
3. Replacing Fuses
Inspect or replace the meter fuse as shown in Figure 18 according to the following steps:
1) Turn off the meter and remove the test leads from the terminals.
2) Use a at-head screwdriver to turn the screw on the bat- tery cover counterclockwise, then remove the battery cover.
3) Gently pry one end of the fuse and then remove the fuse from the clip.
4) The required fuse installed for mA or µA input terminal:
0.8A H 1000V Fuse Type 6X32mm (F1) The required fuse installed for A input terminal: 10A H 1000V Fuse Type 10X38 mm (F2)
5) Reinstall the battery cover and then turn the screw clockwise to tighten the battery cover.
4. Battery Charge
When the indicator of battery level in upper right corner is less than 5% of full capacity, the meter should be immedi­ately charged, otherwise it will affect the measurement
accuracy. As shown in Figure 19, set the rotary switch to
. The message, “Please plug in AC adapter!” will prompt you to insert a power adapter for charging. Connect the power connector to the four terminals in the meter, then insert the power adapter into the power connector, as shown in Figure 19. The message, “Charging” on the display indi­cates it is in charging with a progress bar displaying battery charge to the nearest 5%. The power button will light red. When the battery is fully charged, the red light will turn off and the charging process stops. A “Charge complete” message will display when fully charged.
Attention:
You must use the power adapter specied by Triplett.
35
Page 38
XI. Warranty
Triplett extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (3) three years from the date of purchase.
This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batter­ies and fuses, are not covered by this warranty.
To register a claim under the provisions of this warranty, please contact the distributor from which you purchased the product from for warranty consideration.
ALL WARRANTIES IMPLIED BY LAW ARE HEREBY LIMITED TO A PERIOD OF THREE YEARS FROM DATE OF PURCHASE, AND THE PROVISIONS OF THE WAR­RANTY ARE EXPRESSLY IN LIEU OF ANY OTHER WAR­RANTIES EXPRESSED OR IMPLIED.
The purchaser agrees to assume all liability for any damag­es and bodily injury which may result from the use or misuse of the product by the purchaser, his employees, or others, and the remedies provided for in this warranty are expressly
in lieu of any other liability Triplett may have, including inci­dental or consequential damages.
Some states (USA ONLY) do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you No rep­resentative of Triplett or any other person is authorized to extend the liability of Triplett in connection with the sale of its products beyond the terms hereof.
Triplett reserves the right to discontinue models at any time,
or change specications, price or design, without notice and
without incurring any obligation.
This warranty gives you specic legal rights, and you may
have other rights which vary from state to state.
36
Page 39
XII. Notes
37
Page 40
Model 9065 Part Number: 9065
Made in China
84-910 10/15
Loading...