Triplett 8105 User Manual

Model 9200-A Mini AC Clamp-On
Instruction Manual
1. Triplett Model 9200-A Features
• Economical, Compact AC Current Clamp-On
• Easily Fits in Shirt Pocket!
• Measures from 0 to 300 Amps AC in Two Ranges
• Simple, Two Button Operation
• Large, 3 3/4 Digit, 4000 Count Display
• Autoranging
• Data Hold
• Overrange Indication
• Auto Power Off
• Low Battery Indication
• Powered by one 3V Battery
• Carrying Case, Wrist Strap and Battery included
• 1 Year Limited Warranty
2. Model 9200-A Specifications
Display: 3 3/4 Digit, 4000 Count LCD, 12mm high Operating Principle: Dual Slope Integration Range Selection: Autoranging Overload Indication: “OL” is displayed when input
exceeds 405 A
Low Battery Warning: “BAT” symbol is displayed Sampling Rate: 2.5 times per second Data Hold: Holds reading when DH Button
Overload Protection: AC Current, 400A AC for 60
Seconds Max
Operating Temperature & Humidity: 32° F to 104° F,
Storage Temperature & Humidity: 14° F to 122° F,
Safety / Agency Approvals: CE Mark,
Auto Power Off: Power turns off after 12 min. of
Battery: 1 - 3 Volt CR2032 Button Type (Included)
if product was turned on in the package. If so, replace battery with a fresh one.
Battery Life: 100 Hours Continuous Operation Conductor Size: 27mm Max (1.14 “) Dimensions: 5.8” (L) x 1.8” (W) x 0.8” (H) Weight: 3.4 oz.
is pressed
less than 80% non-condensing
less than 70% non-condensing atmosphere
IEC 1010-1 Overvoltage CAT III 300VAC, Class II and EMC Test Passed
non-use Note: Power consumption is reduced in Auto Power Off. For maximum battery life, turn the power off manually.
Note: The installed battery may be depleted
egnaRnoituloseRycaruccAtnerruCtupnI.xaM
A00.04Am01
A0.003Am001
:noitcetorPdaolrevO:noitcetorPdaolrevO
:noitcetorPdaolrevO:noitcetorPdaolrevO xaMsdnoceS06rofCAA004
:noitcetorPdaolrevO
:esnopseRycneuqerF:esnopseRycneuqerF
:esnopseRycneuqerF:esnopseRycneuqerF tgd5ddaA4.0~0*,zH06~zH05
:esnopseRycneuqerF
tnerruCCA
*)tgd5±gdr%8.1(±
ycaruccaevobaot%5.1dda,zH005~zH04
A003
)eniLV003xaM(
3. Front Panel, Display Details
Figure 1
3.1 Model 9200-A LCD
~ : Alternating Current (AC) BAT : Low Battery Warning AUTO : Autoranging DH : Data Hold A : Amps APO : Auto Power Off
3.2 On/Off and Data Hold Buttons
Momentarily press the ON/OFF Power Button to turn the unit on or off.
To save a reading in the display, momentarily press the DH (Data Hold) Button. The DH symbol will appear in the LCD display. To clear the saved reading, momen­tarily press the DH button again.
3.3 Clamp Head
When measuring AC Current, open Clamp Head by pressing the Lever and clamp around a single conduc­tor to be tested.
NOTE: You cannot measure the current to an appli­ance by clamping around its power cord. You must clamp around only the HOT or NEUTRAL conductor, not the entire cord.
3.4 Safety Guard
For safety, fingertips must NOT go past the Safety Guard.
4. SAFETY PRECAUTIONS
Electrical measurements can be very dangerous with­out proper knowledge of the risks involved. To reduce the possibility of injury, the following precautions must be observed. However, no written warnings or cautions can prevent the misuse, abuse, or careless operation of the product. Safe measurements are only possible if the user follows safety guidelines.
SAFETY CONSIDERATIONS
Read all WARNINGS and CAUTIONS in this manual,
especially those identified with the symbol,
prior to using the product. Failure to do so may
result in product damage or injury to the user.
WARNING!!!
AC Power Lines are very dangerous and without proper knowledge, can be lethal to the user. Power Lines that are considered High Energy Circuits are
especially dangerous. High Energy Circuitry is
usually regarded as any circuitry capable of
supplying over 4000 watts to a load before a safety
device (like a fuse) interrupts the circuit. Arcs of an
explosive nature can develop in this type of
circuitry, causing serious injury or even death.
The Triplett Model 9200-A is not designed for use
in High Energy Circuitry!
4.1 Warnings
Condition of Meter
Make sure that the meter is not damaged in any way (cracks, missing parts, etc.) before every measurement. Make sure that the body is free of dust, grease, and moisture.
Figure 2
High Energy Circuits
This product is not intended for use on High Energy Circuits such as Distribution Transformers, Bus Bars, Large Motors. etc., which are very dangerous.
Maximum Voltage and Current
The maximum current is 300A AC. The maximum volt­age is 300V AC. Do not attempt to measure current that might exceed the specified value or current on lines whose voltage exceeds 300 VAC.
Safety Guard
Do not let fingers go over the Safety Guard when grip­ping the Model 9200-A.
4.2 General Warnings and Cautions
Warning 1 - Do not let children use this instrument
or those who are unable to recognize the dangers of electrical measurements.
Warning 2 - Do not make measurements while bare-
footed. This greatly increases the possibility of elec­trical shock.
Warning 3 - Do not attempt to clean the case with
any harsh solvents. If necessary, use mild detergent and water.
Warning 4 - Avoid severe mechanical shock or vi-
bration, extreme temperature or very strong magnetic field.
Warning 5 - Remove the battery when not in use for
an extended period of time. A low battery might leak electrolyte and corrode the internal components.
5. Measurement Procedures
5.1 Preparation For Use
1. Instruction Manual
Prior to use, read this Instruction Manual carefully and acquaint yourself with the specifications and functions of the instrument. Especially read and observe care­fully the Safety Precautions in Section 4.
2. Battery
One 3V CR2032 battery is installed in this instrument. When the battery is low and “BAT” is displayed on the LCD, open the Battery Cover and replace the battery with a fresh one. Refer to section 6.1 Battery Replace­ment for details.
3. Overrange Indication
If the Input Value exceeds approximately 405 Amps AC, “OL” is displayed on the LCD.
WARNING!!!
DO NOT attempt to make any measurements that
might exceed 300A AC (max 300V Line)
4. Auto Power Off
Approximately 12 minutes after the operation of either button on the product, the unit will go into a ‘sleep’ mode to reduce its power consumption. In the sleep mode, the LCD will turn off, but the unit is still operat­ing at low power. To turn the unit off completely (which maximizes the battery life), press the ON/OFF button before the unit goes to sleep, or if the unit is sleeping, press the ON/OFF button to turn the unit on, then press it again to turn the unit off.
5. Symbols
The following International Symbols shown on the in­strument and in this instruction manual aid in the safe operation of the instrument.
: Caution (Refer to Instruction Manual) ~ : Alternating Current (AC) : Earth Ground : Double Insulation
5.2 AC Current (~A) Measurements
WARNING!!!
Maximum Voltage of AC Current Line is 300V AC.
DO NOT attempt to make any current
measurements that might exceed
300A AC, the Maximum Input Current.
Prior to use, read this Instruction manual
carefully and acquaint yourself with the
specifications and functions of the instrument.
Especially, read and observe carefully the
Safety Precautions in Section 4.
1. Momentarily press the ON/OFF button to turn the
power on.
WARNING!!!
For safe measurements, fingertips must be
placed behind the Safety Guard of this instrument.
DO NOT touch any part of the Power Line
or the Electrical Circuit.
2. Open Clamp Head by pressing the Lever and clamp
around a single conductor. See Figure 3.
NOTE: You cannot measure the current to an appli­ance by clamping around its power cord. You must clamp around only the HOT or NEUTRAL conductor, not the entire cord.
3. Read the current on the LCD.
4. To capture the reading, activate the Data Hold by
momentarily pressing the DH button.
6. Maintenance
6.1 Battery Replacement
WARNING!!!
Remove Clamp Head from the circuit being
measured before attempting to replace the battery.
1. If the battery is low and “BAT” is displayed on the
LCD, replace the battery.
2. Turn the Model 9200 off.
3. Remove the screw on the Battery Cover on the back
side of the instrument.
4. Replace the low battery with a fresh one - 3V CR2032 type.
NOTE: A small screwdriver or pair of tweezers may be helpful in removing the depleted battery. See Figure 4"
5. Replace the Battery Cover and screw. Refer to Figure 4.
Figure 4
6.2 Periodic Test and Calibration
Periodic tests and calibration are necessary to ensure safe and accurate measurements. It is recommended that the instrument be checked and calibrated once each year and/or after it is repaired.
TRIPLETT PRODUCT
RETURN INSTRUCTIONS
In the unlikely event that you must return your Triplett equipment for repair,the following steps must be taken.
1) Call 1-800-TRIPLETT to obtain a Return Material Authorization (RMA) number from Customer Service.
2) Enclose a copy of the original sales receipt showing date of purchase.
3) Clearly print the RMA number on the outside of the shipping container.
4) Return to: Triplett / Jewell Instruments 850 Perimeter Road Manchester, NH 03103 ATTN: Repair Dept.
Be sure to include a full description of the problem, and a telephone number, street address, or email address, where you can be contacted, and a return address where the meter can be shipped to upon repair.
Triplett One Year Limited Warranty
Triplett / Jewell Instruments warrants instruments and test equipment manufactured by it to be free from defective material or workmanship and agrees to repair or replace such products which, under normal use and service, disclose the defect to be the fault of our manufacturing, with no charge within one year of the date of original purchase for parts and labor. If we are unable to repair or replace the product, we will make a refund of the purchase price. Consult the Instruction Manual for instructions regarding the proper use and servicing of instruments and test equipment. Our obligation under this warranty is limited to repairing, replacing, or making refund on any instrument or test equipment which proves to be defective within three years from the date of original purchase.
This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries and fuses, not of our manufacture used with this product are not covered by this warranty.
To register a claim under the provisions of this warranty, contact Triplett / Jewell Instruments’ Customer Service Department for a Return Authorization Number (RMA) and return instructions. No returned product will be accepted without an RMA number. Upon our inspection of the product, we will advise you as to the disposition of your claim.
ALL WARRANTIES IMPLIED BY LAW ARE HEREBY LIMITED TO A PERIOD OF ONE YEAR FROM DATE OF PURCHASE, AND THE PROVISIONS OF THE W ARRANTY ARE EXPRESSLY IN LIEU OF ANY OTHER WARRANTIES EXPRESSED OR IMPLIED.
The purchaser agrees to assume all liability for any damages and bodily injury which may result from the use or misuse of the product by the purchaser, his employees, or others, and the remedies provided for in this warranty are expressly in lieu of any other liability Triplett / Jewell Instruments may have, including incidental or consequential damages.
Some states (USA ONLY) do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you. No representative of Triplett / Jewell Instruments or any other person is authorized to extend the liability of Triplett / Jewell Instruments in connection with the sale of its products beyond the terms hereof.
Triplett / Jewell Instruments reserves the right to discontinue models at any time, or change specifications, price or design, without notice and without incurring any obligation.
This warranty gives you specific legal rights, and you may have other rights which vary from state to state.
Figure 3
5. When measurements are finished, remove Clamp Head from the conductor.
Triplett Model 9200-A Mini AC Clamp-On Meter Cat. No.: 9200
Made in China
84-878