Triplett 1101a User Manual

TRIPLETT MODEL 1101-A
1
COMPACT, DIGITAL MULTIMETER
WITH BACKLIT DISPLAY &
TEMPERATURE TEST
CONTENTS
1. Introduction
3. International Symbols
4. Product Features
5. Specifications
6. Front Panel
7. Measurement Procedures
8. Backlight and Hold
9. Maintenance
10. Warranty Information
1: INTRODUCTION
The Triplett Model 1101-A is a 3 1/2 digit, handheld Digital Multimeter with a backlit LCD display. At its core is an LSI (Large Scale Integr ation) integrated circuit which uses dual slope A/D conversion for stability and accuracy. The meter can measure AC and DC Voltage and Current, Resistance, Temperature, Transistor gain, and perform a Diode and Continuity Tes t. Overload protection makes the meter resistant to damage from accidental overloads.
2: SAFETY RULES & WARNINGS
2
84-848
This meter is NOT to be used to measure High Energy circuits (power
circuitry fused at greater than 4KW, such as distribution circuits,
power entrance circuits, etc.) or circuits classified by CE as
2.1 Read all instructions in this manual before using this meter.
2.2 Prior to using the meter in any situation wh ich could result in injury
2.3 Do not use this meter with its battery cover open, its rear case open,
2.4 When using this meter in schools and workshops, responsible teachers
2.5 Follow the recommendations of any Trade Organizations or
2.6 Do not open this meter for maintenance without first disconnecting
2.7 Repairs and maintenance must only be carried out by qualified service
2.8 Always set the meter to the appropriate range or mode before
2.9 Check the condition of the test leads before making a measurement.
2.10 Make sure test leads are properly inserted and seated in the meter's
2.11 Do not touch the tips of the test leads when making a measurement.
2.12 Before using the meter, examine bot h the meter and the test leads fo r
to the user, in order to verify that the meter is functional and producing a valid reading, test the meter on a ci rcui t(s ) kno wn to h ave potentials equivalent to the potential that is to be measured. For example, before using the meter to determine if an AC power line is energized with 120VAC, test the meter on a line known to be energized with 120VAC.
or with parts removed. Doing so may damage the meter and/or injure the user.
or skilled personnel must control the usage of this meter.
Regulatory Agencies whose scope encompasses the use of this meter.
the test leads from all external circuitry.
personnel or qualified electricians / te chn ici an s who know the dangers of, and the safety rules applicab le to this type of equipment.
connecting it to the circuitry to be tested.
Do not use the test leads if there is damaged insulation or exposed metal.
input jacks. A loose test lead may cause the user to believe that no hazard exists, when in fact, dangerous voltages or currents may be present.
Do not touch live circuitry when making a measurement.
damage. Do not use the meter if damage (damaged insulation, exposed metal, cracked case, burnt smell, etc.) is evident.
WARNING!!!
CATEGORY III (CAT III).
3
2.13 Insert the test leads in the jacks specified in the instructions for
2.14 Do not exceed the maximum voltage or current limitations of the
2.15 Do not appl y volt age or current to the inp ut of th e meter when it is set
2.16 Do not appl y volt age or current to the inp ut of th e meter when it is set
2.17 Do not attempt to measure a voltage source with the test leads
2.18 Do not rotate the Function switch with the test leads connected to the
2.19 Replace fuses only with exact or equivalent fuses. Do not "bridge"
2.20 Do not apply voltages to the input of the meter which are elevated
2.21 Do not continue to use meter wh en the "low power" symbol
2.22 Use caution when working with voltages above 25 volts AC or 35
2.23 This meter is not for use by children.
2.24 Do not use th is meter t o make measurements in adv erse en viron ments
2.25 Do not use meter in condensing atmospheres. That is, do not use
2.26 Do not use the meter if either the meter or the test leads are wet, either
2.27 Do not attempt immediate use of the meter when bringing it from a
performing particular tests. Inse rting the test leads in incorrect jack s can damage the meter and/or injure the user.
meter (see product specifications). Doing so may damage the meter and/or injure the user.
to any of the Ohms Ω ranges. Doing so may damage the meter and/or injure the user.
to the Diode Test / Continuity Beeper mode may damage the meter and/or injure the user.
plugged into the meter's 10A jack. Doing so may damage the meter and/or injure the user.
circuitry to be tested. Doing so may damage the meter or the circuitry, and/or injure the user.
fuses out of circuit. Doing so may damage the meter and/or injure the user.
above the earth ground potential by more than 600VDC / AC peak. Doing so may damage the meter and/or injure the user.
displayed. The displayed reading may be in error and lead the user to believe that no hazard exists, when in fact, dangerous voltages or currents may be present.
volts DC. Such voltages may cause a life threatening electrical shock.
such as rain, snow, fog, or locations with steam, explosive gases or dusts. Doing so may damage the meter and/or injure the user.
meter in conditions where ambient temperature and hu midity could cause condensation of water inside of meter. Doing so may cause injury to the user.
from exposure to the weather, or after cleaning the case of the meter. Doing so may cause injury to the user.
cold environment to a warm environment. Condensation of water, inside and outside of the meter, may produce dangerous measuring
. Doing so
is
4
conditions. Allow the meter to warm to room temperature before
2.28 Do not modify the meter. Changing the design may make the meter
2.29 Use cautio n when attempting to evaluate if a dangerous voltage is
2.30 Do not touch the metallic portion of one test lead if the other test lead
2.31 Remove the battery when the meter may be left unused for longer
2.32 Do not use the meter if there is evidence of chemical leakage from the
2.33 Do not attempt to use meter when no display is present on LCD.
2.34 Do not attempt to measure voltages on capacitors holding dangerous
2.35 Do not use this meter to measure voltages or currents in industri al
2.36 Do not use this meter to measure current in circuits whose open
2.37 Always replace the meter's internal fuse with an exact replacement.
2.38 When you use the meter to check a high-voltage circuit, do not try to
2.39 If there is any doubt abou t the condition of the meter (i.e. safe vs
using.
unsafe and may result in injur y to the user.
present. The meter will not read AC voltage if it is set to DC, nor will it read DC if it is set to AC. For example, if the meter is set to 200VDC, it will not measure a dangerous AC vo ltage, even if the probes are inserted into a household AC wall outlet.
is connected to a live circuit. The current from the live circuit may pass through the meter and appear on the unconnected test lead. Failure to observe this warning may result in user injury.
than 1 month. Chemical leakage from the battery could damage the meter, leading to user injury.
battery. Leakage could damage meter and lead to injury of user.
Doing so may damage the meter and/or injure the user.
charges. These charges may be of lethal potential, and may have enough energy to cause damage to the meter, or injury to the user. These capacitors can store electricity, even when the circuitry which they are in has been disconnected from its power source.
applications where High Energy circuitry is used. High Energy circuitry is usually regarded as any circuitry capable of supplying over 4000 watts to a load before a safety device (like a fuse) interrupts the circuit. See product specifications for Category rating of this meter.
circuit voltage exceeds 500V AC/DC. The meter's fuses are rated at 500V max.
Do not substitute a 250V fuse for a 500V fuse.
connect both test leads at once. Instead, clip one probe to the neutral or ground lead of the circuit (usually a bare, green, or white wire in AC wiring circuits) using the insulated slip-on Alligator Clips. Then probe for voltages with the other test lead. This helps prevent you from accidentally touching a hot wire, since you need to concentrate on only one test lead.
unsafe), remove the meter from service and secure it in a location that will prevent its unintentional use.
5
2.40 Do not use the meter if it does not appear to work correctly on all
2.41 Do not use the meter if it has undergone long-term storage under
2.42 Do not use the meter if it may have been damaged in transport.
2.43 Al ways connect one of the meter’s al ligator clips to the low sid e of a
2.44 To avoid damage to the meter and possible user injury, disconnect test
2.45 Avoid usage near strong magnetic fields (magnets, loudspeakers,
2.46 Av oi d us age n ear s tro ng el ectrost atic fiel ds (high v oltage p ower lin es,
2.47 Avoid usage near strong RF fields (radio or televisio n transmitters,
2.48 The LCD can be damaged by prolong ed exposure to intense sunlight
2.49 When replacing the battery, dispose of the depleted battery in
ranges and in all modes.
unfavorable conditions.
power circuit first. Never clamp onto a hot wire first, (usually red, black, or blue in AC wiring circuits.) If you clamp onto a hot wire first, and touch the other probe, you could receive a shock.
leads from test points before changing the function/range.
transformers, motors, coils, relays, contactors, electromagnets, etc.). The meter may display readings that are in error, causing the user to misinterpret the hazards present. For example, the meter may indicate a low voltage when high voltage s are actually present.
televisions, computer monitors, etc.). The meter may display readings that are in error, causing the user to misinterpret the hazards present. For example, the meter may indicate a low voltage when high voltages are actually present.
walkie talkies, cellular phones, etc.). The meter may display readings that are in error, causing the user to misinterpret the hazards present. For example, the meter may indicate a low voltage when high voltages are actually present.
or to high temperatures. Avoid using or storing the meter under such conditions.
accordance with any prevailing safety or environmental regulations.
3: INTERNATIONAL SYMBOLS
6
The following International Symbols are used throughout this manual and on the case of the meter to identify, caution, or warn the user of important product limitations or important operational procedures that must be followed to ensure safe usage of the product.
Read Instruction Manual for Cautions and Warnings
HIGH VOLTAGE: A dangerous voltage may be present. Use
caution and follow instructions and warnings to avoid electric shock. Failure to do so may result in user injury.
GROUND: The voltage rating associated with this symbol identifies
the maximum allowable voltage that may be applied between the identified terminal(s) an d earth ground.
WARNING!!!
~ AC: Alternating Current
DC: Direct Current
FUSE: For continued safety, replace only with exact or equivalent
fuses.
DOUBLE INSULATION: This meter uses Double Ins ulated
construction. Do not alter the construction and use care not to degrade the insulation when replacin g fuses and battery.
7
4: PRODUCT FEATURES
4.1 20 Measurement Ranges
4.2 0.5" (12mm) high LCD display
4.3 3 1/2 digit resolution (2000 counts)
4.4 Backlit LCD display
4.5 Protective shock absorbing "boot"
4.6 Diode Test
4.7 Continuity Beeper
4.8 Transistor Gain Test
4.9 Temperature Measurement
4.10 Fused µA / mA ranges
4.11 Overload protection
4.12 CAT I (600VDC, 750VAC), CAT II (300V AC/DC) on Voltage Ranges
8
5: SPECIFICATIONS
5.1 Display: .............................. 0.5" (12mm) high LCD
5.2 Display Resolution: ........... 2000 counts, 0000 to 1999
5.3 Overrange Indication: ........ First digit displays "1",
remaining digits are blank
5.4 Measurement Rate: ............ 2 to 3 measurements per second
5.5 Low Power Annunciator: ...
5.6 Case Dimensions: .............. 5.4 x 3.1 x 1.6 inches (L x W x H)
(including boot)
5.7 Weight with battery: ......… 0.55 lbs.
(including boot)
5.8 Power Source: .................. 9 volt battery
5.9a Overvoltage Category: ..... Category II (CAT II) IEC 1010, 300V
(Installation Category) AC/DC. For measuring local level, appliances, portable equipment, etc. For applications where transient voltages are not present, or are of low amplitude. Not for use in high energy industrial installation measurement. Not for use on circuitry where source capability exceeds 4KW.
5.9b Overvoltage Category : .. Category I (CAT I) IEC 1010, 600VDC,
600VAC. For measuring in low power consumer electronics where transient voltages are not present. Not for use in high energy industrial installation measurement. Not for use on circuitry where source capability exceeds 4KW.
5.10 Fuses: ............ mA Jack: 300mA / 500V, 5 x 20mm: 3207-130
10A Jack: UNFUSED
5.11 Supplied Accessories: ... Boot: 10-4271
Test Leads: 79-760 Temperature Probe: 79-795 Instruction Manual: 84-848
Applies to
Voltage
Measurement
Only
Applies to
Voltage
Measurement
Only
9
Note:
a) The following accuracy specifications are valid at 23 degrees C, +/- 5 degrees C, Relative Humidity less than 75% b) The specifications are in the form " +/- (x % of reading + dgt)" where dgt is "Least Significant DIGIT".

DC Voltage

Range Resolution Accuracy
200mV 0.1mV
2000mV 1mV
20V 10mV 200V 100mV 500V 1V ± (0.8% rdg + 2 digits)

Input Impedance: All ranges are 10M

Overload Protection: 500Vrms
± (0.5% rdg + 2 digits)

AC Voltage

Range Resolution Accuracy
200V 100mV 500V 1V

Input Impedance: All ranges are approx. 5M

Frequency: 40Hz to 400Hz Overload Protection: 200mV is 230V (AC or DC Current). All other ranges are 500Vrms. Display: Average Value (RMS of Sine Wave)

DC Current

Range Resolution Accuracy
2000µA 1µA
20mA 10µA
200mA 100µA ± (1.2% rdg + 2 digits)
10A 10mA ± (2.0% rdg + 5 digits)

Overload Protection: 0.3A / 500V fuse, 10A range is unfused.

Max. Current Input: 10A (10 seconds max, 15 minute cool down)
Burden Voltage: 300mV max
± (1.2% rdg + 10 digits)
± (1.0% rdg + 2 digits)
10

Resistance

Range Resolution Accuracy
200 0.1 ± (0.8% rdg + 5 digits)
2000 1
20K 10
200K 100
20M 10K ± (1.0% rdg + 5 digits)
Open Circuit Voltag e: Less than 3V on all ranges
Overload Protection: All ranges 500Vrms (AC or DC Current).

Temperature

Range Resolution Accuracy
-40°F to 302°F ± (1.0% rdg + 5 digits) 303°F to 1832°F
Overload Protection: 500Vrms Thermocouple: K-Type Note: The provided thermocouple is only intended to measure
Temperatures up to 300 degrees F. Exceeding this temperature may damage the probe.
hFE
1 degree Fahrenheit

Diode Test and Continuity Beeper

Displays relative
reading in mV,
not actual diode voltage
Beeper sounds if
meter displays <70.
(Approx. 30)

Transistor hFE test

Can measure NPN or PNP
transistor hFE. Range: 0-1000 ß
± (0.8% rdg + 2 digits)
± (1.5% rdg + 15 digits)
Test current about 1mA.
Open circuit voltage about 2.8V
Test current about 1mA.
Open circuit voltage about 2.8V
Base current about
10µA,Vce about 2.8V
11
6: FRONT PANEL
1) Backlit LCD Display
2) Backlight and Hold button
3) Rotary Function / Range Selector Switch
4) Transistor Test Jack
5) COM Jack: Common (black) jack for all measurements
6) 10A Jack: Input for currents from 200mA to 10A
7) V/ /mA / F Jack: Input for Voltage, Ohms, µA, mA, and Temperature
12
7: MEASUREMENT PROCEDURES
7.1 DC Voltage Measurement:
If the magnitude of the voltage to be measured is unknown, always start
by setting the meter to the highest range, and then to lower ranges,
Do not rotate the RANGE switch with the input applied to the meter.
If the input voltage is higher than 500VDC, or exceeds the limitations
of a CAT I 600V measurement, do not attempt to measure!
until a satisfactory reading is obtained.
WARNING!
This meter will display voltages that exceed the maximum voltage that can safely be applied to the meter. The fact that the meter displays the
voltages does not mean that it is safe to measure these voltages.
7.1.1 Connect the black test lead to the COM jack and the red test lead to
7.1.2 Set the RANGE switch to a V position. If the magnitude of the
7.1.3 With power disconnected from the circuit to be measured, connect the
7.1.4 If the display indicates overrange, i.e. "1- - -", remove power from the
7.1.5 If the displayed value is less than "200" (decimal po int not shown), a
the V/ /mA / F jack.
voltage is unknown, set the RANGE switch to the 500 position. If the input voltage is higher than 500VDC, or exceeds the limitations of a CAT I 600V measurement, do not attempt to measure!
test probes to the measurement points. Apply power to the circuit. The LCD will display the DC volt age.
circuit, disconnect the test probes, and rotate the RANGE switch to the next higher position. Reconnect the test probes, reapply power, and read the voltage on the LCD display. If th e RANGE switch is already at the highest position (i.e. 500 VDC), the input voltage exceeds the measurement capability of the meter and should not be measured.
more accurate reading may be obtained by setting the RANGE switch to a lower range. Remove power from the circuit, disconnect the test
WARNING!
13
probes, and rotate the RANGE switch to the next lower position. Reconnect the test probes, reapply power, and read the voltage on the LCD display. If the RANGE switch is already on t he lowest position (i.e. 200mV DC), no greater measurement resolution can be obta ined.
7.2 AC Voltage Measurement:
If the magnitude of the voltage to be measured is unknown, always start
by setting the meter to the highest range, and then to lower ranges,
Do not rotate the RANGE switch with the input applied to the meter.
If the input voltage is higher than 500VAC, or exceeds the limitations
of a CAT I 600V measurement, do not attempt to measure!
until a satisfactory reading is obtained.
WARNING!
This meter will display voltages that exceed the maximum voltage that can safely be applied to the meter. The fact that the meter displays the
voltages does not mean that it is safe to measure these voltages.
WARNING!
7.2.1 Connect the black test lead to the COM jack and the red test lead to
7.2.2 Set the RANGE switch to a V
7.2.3 With power disconnected from the circuit to be measured, connect the
7.2.4 If the display indicates overrange, i.e. "1- - -", remove power from the
the V/ /mA / F jack.
~ position. If the magnitude of
the voltage is unknown, set the RANGE switch to the 500 position. If the input voltage is higher than 500VAC, do not attempt to measure!
test probes to the measurement points. Apply power to the circuit. The LCD will display the AC volt age.
circuit, disconnect the test probes, and rotate the RANGE switch to the next higher position. Reconnect the test probes, reapply power, and read the voltage on the LCD display. If th e RANGE switch is already at the highest position (i.e. 500VAC), the input voltage exceeds the measurement capability of the meter and should not be measured.
14
p
7.2.5 If the displayed value is less than "200" (decimal po int not shown), a
7.3 DC Current Measurement:
7.3.1 Connect the black test lead to the COM jack. Connect the red test
7.3.2 If the red test lead is inserted into the 10A jack, set the RANGE
7.3.3 If the red test lead is inserted into the V/ /mA / F jack, set the
7.3.4 If the display indicates overrange, i.e. "1- - -", remove power from the
more accurate reading may be obtained by setting the RANGE switch to a lower range. Remove power from the circuit, disconnect the test probes, and rotate the RANGE switch to the next lower position. Reconnect the test probes, reapply power, and read the voltage on the LCD display. If the RANGE switch is already on t he lowest position (i.e. 200V AC), no greater measurement resolution can be obtained.
If the magnitude of the current to be measured is unknown, always start
by setting the meter to the highest range, and then to lower ranges,
Do not rotate the RANGE switch with the input applied to the meter.
If the input current is higher than 10A, do not attempt to measure! If
the input current is between 5A and 10A, limit the measuring time to 10
seconds, and allow 15 minutes for the meter to cool down between
measurements. Do not attempt to measure current in a circuit whose
lead to the 10A jack unless it is known that the input current is less than 200mA. If the current is less than 200mA, con nect the red test lead to the V/ /mA / F jack.
switch to the 10 A circuit to be tested, connect the test leads of the meter in series with the circuit. Apply power to the circuit. The LCD will display the DC current.
RANGE switch to the 200m A from the circuit to be tested, connect the test leads of the meter in series with the circuit. Apply power to the circuit. The LCD will display the DC current.
circuit, disconnect the test probes, and rotate the RANGE switch to the next higher position. Reconnect the test probes, reapply power, and read the current on the LCD display. If the meter is already set to
until a satisfactory reading is obtained.
o
WARNING!
en circuit voltage exceeds 500V.
position. With power removed from the
position. With power removed
15
measure 10A, the current being tested exceeds the measurement capability of the meter and cannot be measured.
7.3.5 If the displayed value is less than "200" (decimal po int not shown), a more accurate reading may be obtained by setting the RANGE switch to a lower range. Remove power from the circuit, disconnect the test probes, and rotate the RANGE switch to the next lower posit ion (and, if necessary, move the red test lead from the 10A jack to the V/ /mA / F jack). Reconnect the test probes, reapply power, and r ead the voltage on the LCD display. If the RANGE switch is already on the lowest position (i.e. 2000 uA), no greater measurement resolution can be obtained.
7.4 Resistance Measurement:
WARNING!
Do not apply voltage or current to the meter when it is set
to any of the Ohms ranges!
7.4.1 Connect the black test lead to the COM jack, and the red test lead to the V/ /mA / F jack.
7.4.2 Set the RANGE switch to the posit ion that is appropriate for the device or circuit to be measured. The LCD display will indicate overrange (i.e. "1 - - -").
7.4.3 Connect the test leads to the device or circuit being measured. Observe correct polarity if appropriate.
7.4.4 If the display indicates overrange, i.e. "1- - -", rotate the RANGE switch to the next higher position and observe the reading on the LCD display. If the RANGE switch is already at the highest po sition (i.e. 20M Ohms), the resistance exceeds the value measurable by the meter.
7.4.5 If the displayed value is less than "200" (decimal po int not shown), a more accurate reading may be obtained by setting the RANGE switch to a lower range. Rotate the RANGE switch to the next lower position and observe the reading on the LCD display. If the RANGE switch is already on the lowest position (i.e. 200 Ohms), no greater measurement resolution can be obtained.
Note: On the 20 range. The displayed value must be less than "20" to obtain a reading on a lower range.
16
7.5 Continuity Beeper:
WARNING!
Do not apply voltage or current to the meter when it is set
7.5.1 Connect the black test lead to the COM jack and the red test lead to
7.5.2 Set the RANGE switch to the
7.5.3 Connect the test prob es to the device or circuit to be tested.
7.5.4 If the resistance of the device or circuit is less than approximately 30
Ohms, the continuity beeper will sound.
Note: The reading displayed on the LCD is not an accurate indication of the resistance of the device or circuit being measured.
7.6 Diode Test:
7.6.1 Connect the black test lead to the COM jack and the red test lead to
7.6.2 Set the RANGE switch to the
7.6.3 Connect the test probes to the device or circuit to be tested. To test a
to the Diode Test / Continuity Beeper mode .
the V/ /mA / F jack.
Do not apply voltage or current to the meter when it is set to the Diode Test / Continuity Beeper mode .
the V/ /mA / F jack.
simple diode, connect the red test probe to th e Anod e o f the d iode an d the black test lead to the Cathode ("banded" end) of the diode. The LCD will indicate the relative voltage drop of the diode. Reverse the connections of the test probes to the diode. The LCD sho uld indicate overrange ("1 - - -"). If the Continuity Beeper sounds when the leads
range.
WARNING!
range.
17
are connected in either direction to the diode, the diode is probab ly
Note: The reading displayed on the LCD is not an accurate indication of the voltage drop of the device or circuit being measured.
7.7 Temperature Test:
shorted.
Do not apply voltage or current to the meter when it is set to the
Do not attempt to measure the temperature of an electrically "live" surface.
The provided thermocouple is only intended to measure temperatures
up to 300 degrees F. Exceeding t his temperature may damage the probe.
7.7.1 Connect the black lead of thermocouple probe to the COM jack and
7.7.2 Set the RANGE switch to the degree F range.
7.7.3 Press the tip of the thermocouple probe against the surface to be
7.8 Transistor Gain Test:
7.8.1 Set the RANGE switch to the hFE position.
7.8.2 Insert the transistor whose gain is to be measur ed into the test socket.
Notes:
1) The Transistor Gain Test does not indicate if a transistor is "good" or "bad".
It only tests for gain assuming that the transistor is good.
2) For safety purposes, the contacts in the test socket are recessed down into the
socket. This may cause difficulty when attempting to measure the gain of transistors removed from a circuit board. To test transistors with short leads, solder a small length of solid wire to each lead before inserting into test socket.
the red lead of thermocouple probe to the V/ /mA / F jack.
measured. The tip may also be immersed in non-corrosive liquids.
Observe the correct polarity (NPN vs PNP) and pinout (E (Emitter), B (Base), C (Collector).
WARNING!
Temperature measurement mode.
User injury may result.
18
8.1 Backlight:
8.2 Hold:
8: BACKLIGHT AND HOLD
The 1101-A incorporates a light green LED backlight into the LCD assembly. Since the backlight uses more power than anything else in the meter, an anti-tease function on the button prevents accidental turn-on, and a 20 second tim e -out prevents extended battery drain.
To turn on backlight, press and hold the Backlight / Hold button for about 3 seconds, until the backlight comes on. Release the button. The backlight will stay on for about 20 seconds and then turn itself off.
A Hold feature is incorporated i nto the 1101-A. It is a momentary Hold feature that only holds a reading as long as the associated button is kept depressed. Since the button that perform s Hold is al so the Backlight button, if the Hold feature is engaged for longer than 3 seconds, the Backlight will turn on.
To activate the Hold feature, press and hold the Backlight / Hold button. The Hold annunciator will appear in the LCD display for as long as the button is held depressed.
19
9.1 Battery and Fuse Replacement:
9: MAINTENANCE
9.1.1 On the back of the case, remove the rubber bumpers that cover the
9.1.2 Gently open case, star ting at the bottom where the screws were
removed.
9.1.3 When the back of the case is removed, remove the depleted batt ery or
damaged fuse, replace, and reassem ble t he meter.
20
case screws. Remove the screws.
Note: When replacing the fuse, use the exact type specified. Do not substitute a 250V fuse for the 500V fuse.
9.2 Cleaning:
9.3 Calibration:
9.4 Service:
To clean the outside of the meter, use a cloth dampened with a mild detergent solution. Do not use any abrasive cleansers, or chemical solvents that may damage the case of the meter.
Periodic calibration of the meter will insure that it is accurate and performing in accordance with its design specifications. A one year interval is suggested. Annual calibration is not included in the warranty.
TRIPLETT PRODUCT RETURN INSTRUCTIONS
In the unlikely event that you must return your Triplett equipment for repair, the following steps must be taken.
1) Call 1-800-TRIPLETT to obtain a Return Material Authorization
(RMA) number from Customer Service.
2) Enclose a copy of the original sales receipt showing date of purchase.
3) Clearly print the RMA number on the outside of the shipping container.
4) Return to: Triplett Corporation
One Triplett Drive Bluffton, OH 45817
ATTN: Repair Dept.
Be sure to include a full description of the problem, and a telephone number, street address, or email address, where you can be contacted, and a return address where the meter can be shipped to upon repair.
21
The Triplett Corporation warrants instruments and test equipment manufactured by it to be free from defective material or workmanship and agrees to repair or replace such products which, under normal use and service, disclose the defect to be the fault of our manufacturing, with no charge within three years (one year on calibration) of the date of original purchase for parts and labor. If we are unable to repair or replace the product, we will make a refund of the purchase price. Consult the Instruction Manual for instructions regarding the proper use and servicing of instruments and test equipment. Our obligation under this warranty is limited to repairing, replacing, or making refund on any instrument or test equipment which proves to be defective within three years from the date of original purchase.
This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries and fuses, not of our manufacture used with this product are not covered by this warranty.
To register a claim under the provisions of this warranty, return the instrument or test equipment to Triplett Corporation, Service Department, One Triplett Drive, Bluffton, Ohio 45817, transportation prepaid. Upon our inspection of the product, we will advise you as to the disposition of your claim.
ALL WARRANTIES IMPLIED BY LAW ARE HEREBY LIMITED TO A PERIOD OF THREE YEARS (ONE YEAR ON CALIBRATION) FROM DATE OF PURCHASE, AND THE PROVISIONS OF THE WARRANTY ARE EXPRESSLY IN LIEU OF ANY OTHER WARRANTIES EXPRESSED OR IMPLIED.
The purchaser agrees to assume all liability for any damages and bodily injury which may result from the use or misuse of the product by the purchaser, his employees, or others, and the remedies provided for in this warranty are expressly in lieu of any other liability Triplett Corporation may have, including incidental or consequential damages.
10: TRIPLETT THREE YEAR
LIMITED WARRANTY
22
Some states (USA ONLY) do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you. No representative of Triplett Corporation or any other person is authorized to extend the liability of Triplett Corporation in connection with the sale of its products beyond the terms hereof.
Triplett Corporation reserves the right to discontinue models at any time, or change specifications, price or design, without notice and without incurring any obligation.
This warranty gives you specific legal rights, and you may have other rights which vary from state to state.
23
24
Loading...