ENERGY AND INFORMATION DEVICE
TRIOMED CGI-3
(the manual)
Version 1
Introduced on
10 December 2015
TALLINN
2016

EC DECLARATION OF CONFORMITY
MANUFACTURER:
Name of the company: Triomed EU OU (Ltd)
ID-code: No. 12146269
Address: Maealuse 4, EE-12618 Tallinn, Estonia
Telephone: +(372)56888978
E-mail: triomed-eu.eu
Triomed EU OU (Ltd) does hereby declare that we are solely responsible for
the following product:
Energy and information device “TRIOMED CGI-3“
(with integrated emitters, options: 40-43 GHz)
It is in conformity with the provisions of the following EC Directives, including any amendments, and with national legislation implementing these Directives:
Directive 2004/108/EC - Electromagnetic compatibility directive (EMC)
Directive 2001/95/EC - General Product Safety Directive (GPSD)
The following harmonized standards have been applied:
-
EN 55011:2009+A1:2010 “Industrial, scientic and medical equipment – Radiofrequency
disturbance characteristics - Limits and methods of measurement”.
- EN 61000-6-3:2007 + EN 61000-6-3:2007/A1:2011+ EN 61000-6-3:2007/AC:2012
“Electromagnetic compatibility (EMC) - Part 6-3: Generic standards – Emission standard
for residential, commercial and light-industrial environments”.
- EN60065:2014 “Audio, video and similar electronic apparatus – Safety requirements”
On the basis of Directive 2004/108/EC the device and its packaging will bear the CE- mark.
Sergey Pichchev
COO Triomed EU OU (Ltd)
Date and place of issue:
12 January 2016
Tallinn, Estonia

Dear Consumer!
This manual explains how to use the TRIOMED CGI-3 device at home.
The device is ready for use and can be utilised only for its intended
purpose in strict compliance with the operation instructions and
safety precautions.
Please read this manual carefully before using the device and follow all the instruc
tions. This will help you achieve better eciency and extend the service life of the
product.
In a case of improper use of the device, the right to make claims will be forfeited and
the risk of possible consequences should be taken exclusively by the owner.
This manual is an integral part of the device. To have the information at hand,
always keep the manual together with the device.
If you have any questions concerning the use of the device, please read the informa
tion provided on www.triomed-eu.eu
The manufactured device is carefully examined to ensure its normal operation and
is cleaned and disinfected using Aerodesin® 2000 disinfection agent.
The device is not a medical product.
The document cannot be amended without prior negotiation with
TRIOMED-EU Ltd. Last updated on 10 December 2015.

TRIOMED CGI-3 (the manual)
4
CE labelling
Non-ionising electromagnetic radiation
Device serial number
Not for general
(household) waste
Important information
about the device or its
operation
Consult instructions
for use
Fragile,
handle with care!
Caution,
consult accompanying
documents
On/o (press/press)
Point of contact
with the body
of the patient
The following signs have been used in the manual
and for labelling the device:

TABLE OF CONTENTS
INTRODUCTION 6
1. PURPOSE 7
2. MECHANISMS OF EFFECTS 8
3. TECHNICAL AND OPERATIONAL CHARACTERISTICS 11
3.1. General technical specications 11
3.2. Structure and functioning 13
3.3. Use of the device 15
4. PACKAGE CONTENTS 16
5. LABELLING 16
6. PACKAGING 17
7. DISPOSAL 17
8. WARRANTY 17
9. PREPARING THE DEVICE FOR USE 18
9.1. Operating restrictions 18
9.2. Safety precautions 19
9.3. Preparing the device for use 19
9.4. List of possible faults and suggested remedies 20
9.5. Technical maintenance 21
10. PROCEDURE FOR USING THE DEVICE 22
Programme 1. Harmony, 30 minutes 23
Programme 2. Regulatory, 10 minutes. 24
Programme 3. Recovery, 15 minutes. 27
10.1. Description of the process of using the device 28

TRIOMED CGI-3 (the manual)
6
INTRODUCTION
“Any treatment that is not based on the energy concept is pointless”
Albert Szent-Györgyi, doctor of medical sciences
Nobel Prize winner in 1937
Hungary
The health of a person depends a lot on his/her ability to adapt to
the environment. People and the environment interact with each other
all the time by exchanging of energy, information and matter. The bal
ance of this exchanging process determines, to a large extent, by our
health condition. To provide this energy and information exchange, the
human body has developed the energy and information system (EIS). It
is known not as a healing method but as a very powerful way for health
improvement.
A modern and eective way to live in harmony within the environ
ment is to strengthen one’s own energy and information system. This
is not yet very known as a healing method but is a very powerful way
for health improvement. A naturally balanced and harmonious body
enjoys much higher vitality, adaptability and survival capacity in our
fast-changing world.
TRIOMED CGI-3 will help you to maintain this crucial balance!

TRIOMED CGI-3 (the manual)
7
1. PURPOSE
TRIOMED CGI-3 device is designed to protect and restore your health
by inuencing your energy and information system (EIS) and the surround
-
ing environment with a specially selected wave signal of the Si-Ge crystal.
TRIOMED CGI-3 is multifunctional
and can be used:
• to harmonise the human EIS;
• to maintain the energy balance of the body;
• restoration of the normal functioning of cells;
• to eliminate consequences of stress;
• to prevent stressful situations;
• general rejuvenation of the body.
TRIOMED CGI-3 device will benet all mem
bers of the family regardless of age and gender.
The device is convenient to use and easy to oper
ate, economical, safe, secure and lightweight.
It is recommended for everyday use at home to
raise vitality, boost health and improve psycho
emotional condition.
In terms of its utilisation, the device is classi
ed as a product of multiple cyclic use.

TRIOMED CGI-3 (the manual)
8
2. MECHANISMS OF EFFECTS
According to Oriental teachings, we are all one unit consisting of two
inseparable parts; the physical body and the energy and information sys
tem (EIS). Every process that takes place in the body, in an organ or in a
cell, has a double nature: An energy change will lead to changes in the
biochemical process as well. And every biochemical reaction will lead to
a change in the energy system.
The energy and information system of a human being consists of
the central energy-informational system of channels, the peripheral
system of channels, the system of energy centers and points of energy
distribution.
The human body works harmoniously if its energy channels are clear
and therefore the ow of the energy is optimal.
Changes in the biochemical metabolism occur due to the changed
energy waves which have it dicult to ow through the energy centres.
The changed energy waves disturb the normal functions of the centres
and therefore prevent a healthy energy ow to the peripheral channels
and internal organs. This results in a lack of energy for providing the neces
-
sary processes in the body and its proper function.

TRIOMED CGI-3 (the manual)
9
The disruption of normal EIS function can be also caused by the
external factors: adverse ecological situations, electromagnetic smog,
poor-quality water, unhealthy food, stress, emotional and psychological
tension etc.
The treatment with the electromagnetic waves of one or another part
of the body will improve the whole EIS (energy and information) system.
The research has revealed that the stimulation with the device pro
ducing electromagnetic signals which are similar to those of the human
EIS system has a benecial eect on the body.
The SI-GE crystal used in the TRIOMED CGI-3 device can work in two
modes: In an active mode (when the crystal is adapted and activated by
being applied on the body) or in the background mode when there is
no direct contact with the body anymore. When working in the active
mode, the crystal uses the BioTrEM technology to tune in to the fre
quency parameters of the EIS. And, after that the device is switched o,
the crystal continues transmitting the same frequency waves, but in the
background mode.
The research of human energy with the help of GDV (gas discharge
visualization) equipment has convincingly showed that the radiation of
the Si-Ge crystal used in TRIOMED CGI-3 aects the human EIS positively.

TRIOMED CGI-3 (the manual)
10
The BioTrEM technology
makes it possible to use the
eect of electronic home
opathy and adjust the distortions in the functioning of
the energy and information
system (EIS) returning it to
its harmonious state and
restoring the normal opera
tion of cells thus rejuvenating the body. The rejuvenation eect is also achieved
on account of increased sen
sitivity of receptors of cellular
membranes.
It is well known that people start aging when membrane receptors
become less sensitive to the signals sent by hypothalamus, the main reg
ulatory centre of the body. Human energy research by means of GDV has
been kindly carried out by Manniskans Resurser AB, Stockholm, Sweden.
The ancient oriental medicine arms that there is a lot of diseases but
just one reason behind them – destroyed harmony.

TRIOMED CGI-3 (the manual)
11
3. TECHNICAL AND OPERATIONAL
CHARACTERISTICS
3.1. General technical specications
The device is produced without using any harmful chemical substances in compliance with the TRIOMED LLC technical documents.
As far as potential risks of use are concerned, the device has been
designed as a product with internal safe power supply.
The TRIOMED CGI-3 device is manufactured with in-built generator
producing electromagnetic waves in the millimeter band (wave length
between 7,5 and 6,98 mm). The software allows to generate MM EMR
with varying modulation during the session. The device features are:
3 programmers of low-frequency modulation of the MM.
The main technical characteristics of the TRIOMED CGI-3 device are
given in Table 1 below.
Tabl e 1
No
Characteristics Description
1.
Start–up time no more than 5 sec
2.
Type of work with specied characteristics continuous, uninterrupted

TRIOMED CGI-3 (the manual)
12
No
Characteristics Description
3.
Automatic shutdown function after the end of the programme
4.
Overall dimensions no more than 75×45×13 mm
5.
Weight no more than 0.1 kg
6.
Mean time between failures no less than 1,500 hours
7.
Life cycle of the device minimum 8 years
8.
Body material polycarbonate plastic
9.
Ambient temperature during use between + 10 and + 35oС
10.
Rated air humidity (combination of relative humidity and temperature)
80 % at 25 °С
11.
Rated direct voltage 3.0 V
12.
Current Consumption no more than 30 mA
13.
Power consumption no more than 100 mVA
The device is supplied with a CR2032 battery.
The external surfaces of the device are disinfected using a 3% solu
tion of hydrogen peroxide or a 1% water solution of chlorhexidine. Soak
a cloth in a disinfecting solution, wring it out and wipe the device.
No special safety measures are required.

TRIOMED CGI-3 (the manual)
13
3.2. Structure and functioning
The device is a monoblock unit. The front panel of the body of the
device (Fig. 1) houses the control button and 2 LED indicators. The back
panel (Fig. 2) houses two fastening screws and the working surface with
the millimetre EMR generator underneath. The appearance of the device
and the elements of its structure are shown in Figures 1 and 2. There is MM
EMR generator, LED indicators, a control button, fastening screws, fasten
-
ing strap location.
location of
MM EMR generator
LED indicators
control button
fastening screws
strap fastening
Figure 1. Front panel Figure 2. Back panel

TRIOMED CGI-3 (the manual)
14
The device is switched on by pressing the control button. While the
button is kept pressed, the programmes no.1 and 2 are selected in turn
as indicated by the switching on of one or two LEDs for approx. 2 seconds
(Fig. 3 - 5). If the control button is not released, the programme switching
cycle is repeated. The release of the button during the indication of a
particular programme activates this programme. MM EMR will switch on.
To conrm that the device is working in accordance with the selected
programme, the corresponding LEDs on the front panel are switched
on (Fig. 3 - 5) and a sound signal is produced briey at certain intervals
(3-4 sec). To use the device in the mute mode, keep the control button
pressed after completion of the rst programme selection cycle. Wait
for the LED combination selected earlier and release the control button.
The device is programmed to switch on the sound in every other pro
gramme selection cycle. Having nished its operation according to the
selected programme the device switches o automatically. The device
can be switched oat any time by pressing the control button.

TRIOMED CGI-3 (the manual)
15
3.3. Use of the device
Choose a suitable program (Section 10.1). Find out the LED combination by which it is indicated (Fig. 3 - 5). Switch on the device by pressing the control button and hold it up to the moment the chosen LED
combination will come up. The device will start working according to the
selected program after the control button is released.
The LED indication of the programmes in the selection and operation
modes is shown in Fig. 3 - 5.
Figure 3. Programme 1
Figure 4. Programme 2
Figure 5. Programme 3

TRIOMED CGI-3 (the manual)
16
4. PACKAGE CONTENTS
Package contents:
• TRIOMED CGI-3 device;
• CR2032 battery;
• user manual;
• consumer packaging;
• warranty.
5. LABELLING
The marking is shown on the label at the bottom of the device.
The label species:
EV-0001/15
TRIOMED CGI-3
Figure
5. Label
• name of the device;
• factory number (device identification num
ber - serial number/two last digits of the year
of manufacture):
• CE marking;
The package species:
• manufacturer, handling;
• battery type;

TRIOMED CGI-3 (the manual)
17
• power consumption;
• operational symbols;
• CE marking;
Other important information is provided in the Declaration
of Conformity and User manual.
6. PACKAGING
The packaging protects the device from weather and mechanical
damages. The packaging (box) provides all the required information
in English and the language of the seller’s country about the product,
manufacturer and authorized representative in the EU. It also contains
handling symbols.
7. DISPOSAL
The device is produced in accordance with the EU requirements for the content of harmful chemical substances.
The device should be disposed of into a special container
for radioelectronic equipment.
8. WARRANTY
The warranty is provided on a separate sheet which can be found
in the box.

TRIOMED CGI-3 (the manual)
18
9. PREPARING THE DEVICE FOR USE
9.1. Operating restrictions
The device can be used only after reading the User Manual.
It is forbidden
• to use the device without reading the User Manual;
•
to use the faulty or damaged device;
• to use the device in rooms with high humidity;
• to put the device into water;
• to let water and chemical substances get inside the device,
• to handle the device roughly, expose it to excessive
mechanical vibrations or shocks, crush or drop the device,
• to use power supply devices not recommended by the
manufacturer;
• to keep the device in places accessible to children and
animals;
• to use the device after it has been stored at a temperature
below 0°С without leaving it rst for at least 4 (four) hours
to lie unpacked at the room temperature.

TRIOMED CGI-3 (the manual)
19
9.2. Safety precautions
No special safety precautions are required
9.3. Preparing the device for use
Before switching on the device, inspect the outside of
the device and make sure that the body is not damaged.
IT IS FORBIDDEN to use the device with the damaged body!
Fig. 7 shows how to insert (replace)
the battery.
• use a cross-point screwdriver to
unfasten the screws;
• take o the top panel;
• insert a CR2032 battery into the
battery holder observing the polar
-
ity (in the case of replacement
remove the old battery rst);
• reinstall the top panel and fas
ten it to the bottom panel with
the screws.
Figure 7. Battery replacement

TRIOMED CGI-3 (the manual)
20
The level of the battery charge is indicated by the brightness of the
LEDs.
The serviceability of the device should be checked before every use.
Normal functioning of the device is described in Section 3.2 “Structure
and functioning”.
9.4. List of possible faults and suggested remedies
Possible faults and suggested remedies are listed in Table 2.
Tabl e 2
No
Signs of a fault Likely reason Remedy
1
The LEDs do not switch
on when the control
button is pressed
The battery is defective
or has discharged.
Replace the battery.
If following the insertion of a non-defective battery the device
cannot be switched
on, send it to be
repaired.
2
Lack of sound in the
speakers following
the activation of the
programme
Electric circuit failure Send the device to be
repaired

TRIOMED CGI-3 (the manual)
21
3
The device does not
exit the selection mode
and does not switch o
Electric circuit failure Send the device to be
repaired
4
The mute mode cannot
be selected
Electric circuit failure Send the device to be
repaired
5
The battery in the
device discharges
quickly (within less
than a month) even
if the device is used
rarely.
Electric circuit failure Send the device to be
repaired
In the case of other faults please contact the Seller. The addresses and
contact numbers can be found on the package.
9.5. Technical maintenance
No technical maintenance is provided during the life cycle of the
product. The serviceability of the device and the characteristics of
the radiation it generates are checked once a year at the technical
maintenance centres of the seller.

TRIOMED CGI-3 (the manual)
22
10. PROCEDURE FOR USING THE DEVICE
This Manual regulates the use of the device. Exposure of the human
biological energy to the low intensity radiation of the Si-Ge crystal
maintains and strengthens a person`s energy and information system,
restores the integrity of the energy ow, provides restoration of the har
monious functioning of cells and, as a consequence, general rejuvenation of the body.
Programmes 1-2 launches the low-intensity wave radiation gener
ated by the Si-Ge crystal with varied distribution of the low-frequency
modulation and integral capacity during their work. The combination of
Programmes ensures multifunctionality and enhances the eect. Select
programme according to treatment of preference.
The device may be used to stimulate:
•
the central energy-informational system of the channels;
• the peripheral system of the channels;
• the system of energy centers;
• problem areas.
The eects of the treatment on all these systems have a long-term
eect. In the beginning, shortly after a treatment, an initial positive eect
is apparent. This eect will increase during the next 2-3 weeks. That is
why intervals of up to three weeks are required between the cures.

TRIOMED CGI-3 (the manual)
23
Programme 1. Harmony, 30 minutes
(directly on the body or for distant stimulation)
The maximum duration of stimulation per each 24-hour
period is 2 hours.
The programme is designed to stimulate the whole EIS
of a person. For that purpose, the device should be worn
on a strap or string on the neck or placed at a distance of 50
cm from the body. The radiation of the Si-Ge crystal struc
tures water particles present in the air which, having been
restructured, start reradiating and create the all-around
radiation eect which is maintained by water particles
even after the device is switched o. The surrounding
space is thus now cleaned from electromagnetic smog
and that is why MM radiation is so vitally important for all
biological objects (people, animals, plaints).
The interaction between the Si-Ge crystal radiation and
all the nodes and energy ows in the EIS of all the people
present in the room are harmonized. The benet from
the harmonized conditions improves the body’s overall
resistance towards unfavourable factors, strengthening its
recovery properties. People become more tolerant towards
each other. The result is an oasis of harmonious existence.

TRIOMED CGI-3 (the manual)
24
Programme 2. Regulatory, 10 minutes.
This programme is designed to work with the main
energy centres, channels and problem areas.
The programme is indicated to be used in cases of
fatigue, increased irritability, overexertion, psycho-emo
tional stress, sleep disorders and mood decline/depression.
It promotes an adequate reaction of the organism to any
stimulus. It optimizes all stages of reaction to the aggres
sive factor causing stress. To activate the stress-busting
eects, it is recommended to use the device to stimulate
the projection of the chakras on the arms in the area of the
wrist, the centre of the palm of the left hand (left handed
should use their right hand palm) and on the head - in the
tragus of the ear, with the slow circular motion.
The duration of stimulation for one zone is 12 minutes, twice a day. In
case of chronic stress, exposure rate is 10 - 15 days.
The energy centres are stimulated to get the energy balance of the
body. In order to restore the ow of energy through the channels of the
hands and feet, use the device to stimulate the ngertips and toe tips
respectively, for 1 minute each, starting with the left little nger and end
-
ing with the right little nger, successively.

TRIOMED CGI-3 (the manual)
25
At the same time due to the regulation of energy ows and nodes,
human EIS is recovered, the general resistance of the organism to the
unfavourable factors increases and a restoration of functional reserves
occurs.
The regulatory programme allows diagnosing assessments of the
human EIS, identifying areas that work with the distortions and by stim
-
ulating them, bringing the interaction of cells in the body to harmony.
The diagnostic assessment should be carried out in a calm, quiet and
relaxed atmosphere to pay attention to the body response.
In the case of slow-owing (sluggish) processes the following is
recommended:
1.
Take a comfortable position, calm down and start stimulating the left
palm with the device making slow, circular movements during 5 minutes
(2 circles/minute), left handed should use their right hand palm.
2. Pay attention to your body. Gradually, not right away, you can feel
your body response. It may be in the form of tingling, itching or slight
ly painful sensations in dierent places. You should identify the most
pronounced response. This will be the location of the body`s reaction
to the impact (trigger zone). You should switch on the device again,
bringing it on that spot and start the healing session. Treat just one
trigger zone per session.

TRIOMED CGI-3 (the manual)
26
A woman suffered from a fatigue
syndrome.
The rst photo shows
a lack of energy in the
area of the left kidney
and the lungs, as well as
reduced adaptation. The
diagnostic assessment
of the palm identied
the problem areas. The
stimulation of the both
areas with Triomed
CGI-3 was carried out
during 10 minutes for
each one. The photo
taken after the stimula
tion shows an increased
energy in the both areas
and the restored adap
tation as well.

TRIOMED CGI-3 (the manual)
27
The following days continue the stimulation with the same sequence.
The body may respond in the same spot or elsewhere. The body will determine the response zone in accordance with its own needs. The pain
area can be the same or a new one. Make the treatment session where
it is needed. The duration of stimulation of one zone is 10 minutes.
Programme 3. Recovery, 15 minutes.
With the program “Recovery”, the impact of background
emission can be performed. To do this, you have to bring
the little window of the device to the selected area, and
only then start the programme. The device can be left in
the chosen area for some more time even after the end of it.
The information signal of the Si-Ge crystal helps EIS rec
ognize the problem automatically and starts the appropriate reactions of the body aimed at solving it.
The impact on the body`s problem area with the “Recov
ery” programme leads to rapid recovery and normal operation of the area.
The worse the body`s condition (i.e. abrasions, burns bleedings), the
more preferable it is to use the programme “Recovery” since it acceler
-
ates the regeneration of damaged tissues.

TRIOMED CGI-3 (the manual)
28
10.1. Description of the process of using the device
• Take a comfortable position.
• To begin the procedure, select the suitable programme of stimulation
and switch on the device by pressing the control button.
• When using Programme 1 place the device on the body part with the
top panel facing up (the small round dimple on the panel, where the
square crystal is visible).
• To switch on the device and to begin the session, press the control
button. When continuing pressing the control button, the device suc
cessively passes on to the next programmes in line (which is visible by
the programme lights ca. 2 seconds/programme). When the control
button is released at a selected programme, it starts working and the
speaker produces a sound.
• The duration of each treatment session is determined by the pro
gramme in use. It is not recommended to break off the session. At the
end of the treatment session the device will switch off automatically.
• You can switch off the device earlier at any time if needed, by pressing
the control button.