Triner Scale TS-700 User Manual V2.0

OG Version 1.0_11-15-13
Compatible with DataLog
TM
Scale-to-PC Data Logging Software
TS 700-SS
Series 2
TS 700-MS
Series 2
TS 700-WB
TS-700 Series Digital Indicators
Includes Appendix for Scale Technicians
Set Up • Calibration • Connectivity
Operating Guide
Triner Scale & Mfg.Co., Inc.
8411 Hacks Cross Road Olive Branch, MS 38654
(800) 238-0152
www.trinerscale.com
© 2013 Triner Scale. All rights reserved.
!!!! CALIBRATION WARNING !!!!
Calibration AND inspection of calibration properties
is prohibited unless done so by
a qualified scale technician.
If digital indicator was purchased as part of a
complete scale, it is
FACTORY CALIBRATED AND READY FOR USE.
WARNING
The indicator has static sensitive components.
Do not make any connections without powering off the indicator.
DANGER
Electric shock risk.
Take necessary precautions to avoid risk of shock.
Operating Guide | TS-700 Series
1
Table of Contents
Specifications
TS-700 MS, SS and WB.................................................. 2
Quick Start: Connecting and Powering On
TS-700 MS ....................................................................... 3
TS-700 SS ........................................................................ 5
TS-700 WB ....................................................................... 7
Operation
Keys and Functions: TS-700 MS, SS and WB .............. 9
Parameters and Settings
How to Access and Adjust Parameters........................ 12
Example: How to Set Time & Date ............................... 12
Parameters Charts......................................................... 14
Using a Receipt Printer (optional)
TS-700 MS, SS & WB ................................................... 19
APPENDIX FOR SCALE TECHNICIANS
Calibration Procedure ........................................................A-1
Load Cell Connections
TS-700 MS & WB...................................................A-9
TS-700 SS ..............................................................A-9
Serial Port/RS-232 Connections
TS-700 MS ...........................................................A-10
TS-700 SS ............................................................A-11
TS-700 WB .......................................................... A-11
Data Output Formats, TS-700 MS, SS & WB
Remote Display Format.......................................A-12
PC Computer Continuous Sending Format........A-13
Serial Interface Reception Command.................A-13
Error Codes
TS-700 MS, SS and WB ......................................A-15
Factory Default Parameter Settings
TS-700 MS, SS and WB ......................................A-16
Operating Guide | TS-700 Series
2
Specifications
Resolution ......................... Display: 30,000
ADC................................... 2,000,000
Zero stability error.............. TK
0
< 0.1µV//K
Span stability error............. TK
spn
< ± 6 ppm//K
Sensitivity (internal) ........... 0. 3 µV /d
Input voltage...................... -30~30mV DC
Excitation circuit................. 5 VDC
Load Cell Max.................... 8, 350Ω
AC power........................... AC100~250V
Operation temp .................. - 10 °C ~ + 40 °C
Operation humidity............. ≤90% RH
Storage temp ..................... - 40 °C ~ + 70 °C
KEYPAD KEYS
*requires adjustment of parameter settings
KEY
FUNCTION
On/Off
Powers on, powers of the indicator
Hold
Hold weight on scale, or Hold peak weight*,
or Hold unstable/animal weight*
Total
Total Weight: total of multiple weighings
Kg/lb
Display weight in pounds or kilograms
Gross/Tare
Tare container weight, display gross or net
weight
Zero
Returns the indicated weight to “0”
Print
Prints receipt (requires optional printer)
PC connectivity requires Triner Scale’s DataLog PC™ software.
Operating Guide | TS-700 Series
3
Quick Start: Connecting & Powering On
TS-700 MS
Connect Weighing Platform Cable
Attach the home run cable from the weighing platform to the loadcell port located on the underside of the digital indicator.
Connect Any Optional Devices Referring to the manual provided with optional device (such as a receipt printer), connect the
Operating Guide | TS-700 Series
4
Quick Start: Connecting & Powering On
TS-700 MS (CONT’D)
device to the RS-232 port located on the underside of the digital indicator. Power on the device.
Connect AC Adapter Plug the AC adapter included with the digital indicator into the 9VDC 1200ma port located on the underside of the digital indicator.
Power On The Digital Indicator Press and hold the ON/OFF key for 2+ seconds to power on the indicator. A self-test routine will run, followed by “0” indicated weight. If a small amount of weight displays, press the Zero (word) key. Scale is ready to operate.
IMPORTANT NOTE: if equipped with optional built in rechargeable battery, fully charge for 18 hours before operating on battery power.
Operation
Refer to page 9 for operating instructions and functions details.
Certain settings in the TS-700 MS can be adjusted according to user preferences. Refer to page 14 for a listing of adjustable parameters and available settings options.
Operating Guide | TS-700 Series
5
Quick Start: Connecting & Powering On
TS-700 SS
Connect Weighing Platform Cable
Attach the home run cable from the loadcell port to the weighing platform junction box. Refer to installation instructions provided with the Triner Scale weighing platform.
Operating Guide | TS-700 Series
6
Quick Start: Connecting & Powering On
TS-700 SS (CONT’D)
Connect Any Optional Devices
Referring to the manual provided with optional device (such as a receipt printer), connect the device to the RS-232 port located on the underside of the digital indicator. Power on the device.
Connect AC Adapter Plug the AC adapter included with the digital indicator into the 9VDC 1200ma port located on the underside of the digital indicator.
Power On The Digital Indicator Press and hold the ON/OFF key for 2+ seconds to power on the indicator. A self-test routine will run, followed by “0” indicated weight. If a small amount of weight displays, press the ZERO (word) key. Scale is ready to operate.
IMPORTANT NOTE: Fully charge for 18 hours before powering the indicator from the built in rechargeable battery.
Operation Refer to page 9 for operating instructions and functions details.
Certain settings in the TS-700 MS can be adjusted according to user preferences. Refer to page 14 for a listing of adjustable parameters and available settings options.
Operating Guide | TS-700 Series
7
Quick Start: Connecting & Powering On
TS-700 WB
Connect the Weigh Bar Cables
Attach the home run cables to the loadcell ports and to the weigh bars. For weigh bar set up, refer to the User’s Guide included with livestock the scale.
Operating Guide | TS-700 Series
8
Quick Start: Connecting & Powering On
TS-700 WB (CONT’D)
Connect Any Optional Devices
Referring to the manual provided with optional device (such as a receipt printer), connect the device to the RS-232 port located on the underside of the digital indicator. Power on the device.
Connect AC Adapter Plug the AC adapter included with the digital indicator into the 9VDC 1200ma port located on the underside of the digital indicator.
Power On The Digital Indicator Press and hold the ON/OFF key for 2+ seconds to power on the indicator. A self-test routine will run, followed by “0” indicated weight. If a small amount of weight displays, press the ZERO (word) key. Scale is ready to operate.
IMPORTANT NOTE: Fully charge for 18 hours before powering the indicator from the built in rechargeable battery.
Operation Refer to page 9 for operating instructions and functions details.
Certain settings in the TS-700 MS can be adjusted according to user preferences. Refer to page 14 for a listing of adjustable parameters and available settings options.
Operating Guide | TS-700 Series
9
Operation
KEYS AND FUNCTIONS: TS-700 MS, SS & WB
KEY
FUNCTION/OPERATION
On/Off
Press and hold for 2+ seconds to operate.
To Hold Weight on Scale (manual hold): With weight on scale, press Hold key.
To Hold Peak Weight of Multiple Weighings (auto hold)*: To enable, press Hold key before loading scale.
To Hold Weight of an Unstable Load (auto hold)*: To enable, press hold key before loading scale.
To Release the Hold Function: Press hold key again to return to normal operation.
*Requires adjustment of parameter #11
Hold
In Settings Mode: Return Key
- Previous
- Pressed with other key(s) for functions.
Operating Guide | TS-700 Series
10
KEY
FUNCTION/OPERATION
To Add Weights:
With weight on scale, press the Total key to place weight and count in memory. Remove weight, place next item on scale, press Total to add weight to total weight and total count.
To View Totals: Press Total and Print key at the same time. The total weight will flash continuously.
To Print Totals:
Press and hold the Print key for 3 seconds while in View Totals mode (requires optional printer).
To Go Out of View Totals Mode: Press and hold the Total key. “Clr n” will display.
If you wish to continue adding to the totals, press the Print to accept the “Clear-No” prompt. Continue weighing and adding items as required.
To Clear the Totals Memory: Press the Zero/Up Arrow key to change the “Clr n” display to “Clr y”. Press the Print key to accept the “Clear-Yes” prompt.
Total
In Settings Mode: Previous/Home Key
First press = previous Second press = home
Press to change pounds/kilograms
kg/lb
In Settings Mode: Left Arrow Key
Press to move active digit one unit over to the left
Operating Guide | TS-700 Series
11
KEY
FUNCTION/OPERATION
For use after a TARE weight has been stored in the indicator. Press key to toggle view between gross and net weight.
PLEASE NOTE that there may be a momentary pause in the display when going from gross to tare weight.
Gross
In Settings Mode: Right Arrow Key
Press to move active digit one unit over to the right
With weight on the scale, press to Tare the weight.
Press again to remove Tare weight.
Tare
In Settings Mode: Down arrow key
Press to reduce value of active (flashing) digit.
Press key to zero the scale.
Zero
In Settings Mode: Up arrow key
Press to increase value of active (flashing) digit.
Print
Press key to print a ticket (requires optional printer).
In Settings Mode: Enter Key Press to accept entry. Press to proceed to next item.
IF EQUIPPED WITH INTERNAL RECHARGEABLE BATTERY: Fully charge the battery for 18 hours before operating on battery power.
Operating Guide | TS-700 Series
12
Parameters and Settings
CAUTION: Use care when accessing and
adjusting your indicator’s parameters. Several parameters can be accessed that, if altered, will change your indicator’s operations.
HOW TO ACCESS AND ADJUST THE TS-700 MS PARAMETERS
Press and hold down the PRINT and HOLD keys. “C” will display. Release the keys. “C08” or “C01” will display*, with last digit flashing. The digit flashing is the active digit, and can be adjusted up or down by using the Zero/up arrow key or Tare/down arrow key. Use the kg-lb/left arrow key and the Gross/right arrow key to change which digit is active (flashing).
*TIP: Parameters C01 through C07 are for typically locked out and are only accessible by scale technician during calibration.
Navigate to the setting parameter you wish to adjust and use the arrow keys to change the settings values. Press the Print/enter key to accept your settings, then press the Total/home key to exit.
EXAMPLE: To adjust the date setting in parameter C16:
With the 8 flashing in “C08”, press the Tare/down arrow key to reduce the 8 to 6.
Operating Guide | TS-700 Series
13
Press the kg-lb/left arrow key to move the active digit from the 6 to the 0 (the 0 will flash).
Use the Zero/up arrow key to increase the 0 to 1 (C16 will display).
Press the Print/enter key.
The display will show digits for month, day and year. Use the arrow keys as discussed above to navigate to and change the digits.
Press the Print/enter key to accept your settings.
Press the Total/home ( ) key to exit.
Operating Guide | TS-700 Series
14
PARAMETERS CHART
(For calibration parameters, see Appendix for Scale Technicians, page A-4)
PARAMETER
FUNCTION
OPTIONS
C08 Warning tone
0 = Disable warning tone 1 = warning tone on
C09 Automatic
Power-off
0 = Disable auto off 10 = Power off in 10
minutes.
30 = Power off in 30
minutes.
60 = Power off in 60
minutes.
C10 Power saving
setting (display backlight on/off)
0 = Disable power saving
setting 3 = Display off in 3 min. 5 = Display off in 5 min.
C11
Hold function
0 = Disable hold function 1 = Hold peak weight 2 = Hold weight on scale 3 = Hold animal weight 4 = Reserved
Note: “hold peak weight”
will display the max. load of multiple weighings.
C12 Kg/lb
conversion
0 = kg/lb conversion
disabled 1 = kg/lb conversion
enabled
C13 Upper/lower
limit alarm
C14 Lower limit
alarm value
Enter numeric values for
upper and lower checkweighing limits.
Operating Guide | TS-700 Series
15
PARAMETER
FUNCTION
OPTIONS
C15 Inner Code
display
Enter C15 to check the
inner code.
C16
Date
Set the date, from left to
right: year/month/day
C17
Time
Set the time from left to
right: hour/min./sec.
C18 Communication
setting
(Serial interface data output method)
0 = Serial interface data
output disabled
1 = Continuous sending,
remote display
2 = Print key sends data to
printer or PC.
3 = PC command request
(DataLogPC software)
4 = PC continuous
sending.
5 = Remote display
continuous sending format.
C19
Baud rate
0 = 1200 1 = 2400 2 = 4800 3 = 9600
C20
CAUTION: Changing these parameters will affect the indicator’s performance.
Zero range (amount that can be zeroed out with the Zero key)
0 = Disable manual zero
setting 1 = ±1% max capacity 2 = ±2% max capacity 4 = ±4% max capacity 10 = ±10% max capacity 20 = ±20% max capacity 100 = ±100% max capacity
C21
Zero range at power-up
0 = no initial zero setting 1 = ±1% max capacity 2 = ±2% max capacity 5 = ±5% max capacity 10 = ±10% max capacity 20 = ±20% max capacity
Operating Guide | TS-700 Series
16
PARAMETER
FUNCTION
OPTIONS
C22
CAUTION: Changing these parameters will effect the indicator’s performance.
Automatic Zero: tracking range
0 = Disable zero
tracking
0.5 = ±0.5d
1.0 = ±1.0d
2.0 = ±2.0d
3.0 = ±3.0d
4.0 = ±4.0d
5.0 = ±5.0d Note:
d = division
The zero tracking range
cannot be more than manual zero range.
C23
Automatic Zero; tracking time
0 = Disable zero tracking
time 1 = 1 second 2 = 2 seconds 3 = 3 seconds
C24
CAUTION: Changing these parameters will effect the indicator’s performance.
Overload range
00 = close overload range
01d -- 99d
(d = division)
C25
Negative display
0 = -9d 10 = 10% max. capacity 20 = 20% max. capacity 50 = 50% max. capacity 100 = 100% max. capacity
Operating Guide | TS-700 Series
17
PARAMETER
FUNCTION
OPTIONS
C26
CAUTION: Changing these parameters will affect the indicator’s performance.
Standstill time
0 = quick 1 = medium 2 = slow
C27
Standstill range
1 = 1d 2 = 2d 5 = 5d 10 = 10d
(d = division)
C28
CAUTION: Changing these parameters will affect the indicator’s performance.
Digital filter To compensate for unstable loads on the weighing platform (for example, animal weighing).
0 = close dynamic filter 1 = 1 digital filter strength 2 = 2 digital filter strength 3 = 3 digital filter strength 4 = 4 digital filter strength 5 = 5 digital filter strength 6 = 6 digital filter strength
C29
CAUTION: Changing these parameters will affect the indicator’s performance.
Noise filter
0 = disable noise filter 1 = 1 digital filter strength 2 = 2 digital filter strength 3 = 3 digital filter strength
C30
Format of date print out
0 = yy.mm.dd 1 = mm.dd.yyyy 2 = dd.mm.yyyy 3 = yyyy.mm.dd
C31
Analog output setting
0 = 0~5Vouput 1 = 4~20mA output
C32
RESERVED
-------
Operating Guide | TS-700 Series
18
PARAMETER
FUNCTION
OPTIONS
C33
Relay output setting (optional)
0 = Close relay output 1 = Open relay output
function 1
2 = Open relay output
function 2
3 = Reserved menu
C34
RESERVED
-------
C35
RESERVED
--------
C36
CAUTION: Changing these parameters will affect the indicator’s performance.
Gravity of calibration location
(Only available to technician during calibration)
C36 = 9.7000 -- 9.9999
C37
CAUTION: Changing these parameters will affect the indicator’s performance.
Gravity of destination
(Only available to technician during calibration)
C37 = 9.7000 -- 9.9999
C38
Reserved
-------
TIP: When in parameter settings mode, you can move forward through all parameters and view the settings by pressing the PRINT (
) key
repeatedly.
Operating Guide | TS-700 Series
19
Using A Receipt Printer
To print receipts with an optional printer:
- With power off, attach the printer cable to the connector socket on the underside of the indicator.
- Power on the printer & indicator.
- Place weight on the scale and press the
Print key.
- If receipt does not print, settings may need to be adjusted in parameters C18 & C19.
- To adjust the format of how the date prints, see options in parameter C30.
Typical receipt
Operating Guide | TS-700 Series
A-1
Appendix For Scale Technicians
Calibrating a TS-700 Series Indicator
NOTICE: The TS-700 SS/MS/WB are NTEP
Class III Legal for Trade approved and certified. Unlocking the calibration parameters is not allowed by persons other than qualified scale technicians.
ALL MODELS: Place the Indicator in Calibration Mode and Determine if the Calibration Parameters are Accessible
1) Power on the indicator.
2) Press the PRINT and HOLD keys. “C” will display. Release the keys.
3) If the calibration parameters are accessible, “C01” will display, with the 1 flashing. Proceed to page 8, “CALIBRATION PROCEDURE”.
NOTE: If C08 Displays instead of C01, the calibration switch must be adjusted. Disconnect the AC adapter and power off the indicator.
DEFAULT SETTINGS CAUTION Placing the indicator in calibration mode enables access to the Default Settings Parameter, CO7.
NEVER activate Parameter C07 unless you are a qualified scale technician, and:
A) You understand that all custom settings will be deleted permanently, and B) You understand how to restore all custom parameters to the required settings based on the scale’s specifications and usage.
Operating Guide | TS-700 Series
A-2
Appendix For Scale Technicians
Calibration Switch, TS-700 MS: Remove the screws and the back plate of the enclosure. A white-capped calibration switch is located at the bottom left corner of the circuit board. Press the switch fully in and release. This will place the switch in the up position, allowing access to the calibration parameters. Loosely reattach the rear plate.
Calibration Switch, TS-700 SS: Remove the screws and the back plate of the enclosure. A calibration switch is located at the bottom left corner of the circuit board. Move the switch to the “On” position, allowing access to the calibration parameters. Loosely reattach the rear plate.
Calibration Switch, TS-700 WB: Disconnect all cables and cords. To remove the back plate of the enclosure, place the indicator upside down (not face down) on a work surface. Locate the two slotted holes near the connection sockets. Insert a flat blade screwdriver into each slot (one at a time), pushing the blade inward while angled towards the back of the indicator. This will release the retaining prong and enable removal of the back plate. Carefully lift the rear plate.
A white-capped calibration switch is located at the bottom left corner of the circuit board. Press the switch fully in and release. This will place the switch in the up position, allowing access to the calibration parameters. Loosely reattach the rear plate.
Operating Guide | TS-700 Series
A-3
Appendix For Scale Technicians
ALL MODELS: After enabling access to calibration parameters, proceed as follows:
CALIBRATION PROCEDURE
1) Power on the indicator.
2) Press the PRINT and HOLD keys. “C” will display. Release the keys.
3) “C01” will display, with the 1 flashing. The indicator is now is Parameter Settings mode.
Navigating in Settings Mode When the indicator is in settings mode, refer to the graphic printed below the key to determine the keypad key’s function.
The and arrow keys are used to increase/decrease the value of the flashing digit, or to navigate up and down a list of preset options.
The and arrow keys are used to move the active digit over to the left or to the right.
The key and key can be used to move forward or backward through the parameters. Also,
the key is used to enter/accept changes after they have been made to the parameter settings.
The key can be used to exit settings mode.
With the indicator in settings mode and the calibration weight nearby, proceed with the calibration process while referring to the following chart:
NOTE: After completing and verifying the calibration, set the calibration switch to the “Off” position and securely fasten the rear plate.
Operating Guide | TS-700 Series
A-4
Appendix For Scale Technicians
PARAMETER
FUNCTION
OPTIONS
C01 Default weighing
unit
1 = kg 2 = lb
Description:
With C01 flashing, press the
key, then use the up or down arrow keys to change the parameter setting as needed.
Press the key to accept the setting and continue on to the next parameter.
C02 Decimal location
0 = none 1 = one decimal place 2 = two decimal places 3 = three decimal places
Description:
With C02 flashing, press the
key, then use the up or down arrow keys to change the parameter setting as needed.
Press the key to accept the setting and continue on to the next parameter.
Operating Guide | TS-700 Series
A-5
Appendix For Scale Technicians
PARAMETER
FUNCTION
OPTIONS
C03 Graduations
1 = one unit 2 = two units
5 = five units 10 = ten units 20 = twenty units 50 = fifty units
Description:
With C03 flashing, press the
key, then use the up or down arrow keys to change the parameter setting as needed.
Press the key to accept the setting and continue on to the next parameter.
C04
Capacity
Description:
With C04 flashing, press the
key, then use the left and right arrow keys to select the active digit, then use the up and down keys to adjust the digits to the desired capacity, i.e., 5,000. Press the key to accept
the setting and continue on to the next parameter.
Operating Guide | TS-700 Series
A-6
Appendix For Scale Technicians
PARAMETER
FUNCTION
OPTIONS
C05 Zero Calibration
0 = skip zero calibration 1 = proceed with zero
calibration
Description:
With C05 flashing, press the
key, then use the up arrow key to change the 0 parameter setting to 1.
Press the key to proceed. The zero cal will count down from 10.
Press the key to proceed to the next parameter.
Operating Guide | TS-700 Series
A-7
Appendix For Scale Technicians
PARAMETER
FUNCTION
OPTIONS
C06
Calibration
0 = skip calibration
1 = proceed with calibration
Description:
With C06 flashing, press the
key, then use the up arrow key to change the parameter setting to 1.
Press the key to proceed. “Span” will briefly display, then all digits will display.
Load the weighing platform, then adjust the displayed digits to match the weight on the platform, i.e, 4,000. Press the key to
proceed. The indicator will
calibrate, then display
“CalEnd”.
PRESS THE KEY TO
ACCEPT THE
CALIBRATION AND EXIT
SETTINGS MODE.
CAUTION -- IMPORTANT NOTICE!!
After calibrating, be sure to exit settings mode without activating parameter C07!
Parameter C07 is the parameter that can DELETE all custom settings and reset the indicator to factory settings.
Operating Guide | TS-700 Series
A-8
Appendix For Scale Technicians
PARAMETER
FUNCTION
OPTIONS
C07 CAUTION!!
FACTORY RESET… Erases all custom settings!
0 = skip reset to factory
defaults
1 = proceed with reset to
defaults
Description: To reset to factory defaults, change the 0 setting to 1 and press the key.
Operating Guide | TS-700 Series
A-9
Appendix For Scale Technicians
Load Cell Connections
The indicator can connect with a maximum of eight (8) 350Ω load cells, 4 wire or 6 wire configurations.
Excitation voltage for the load cell is 5VDC, the maximum output current is 120mA.
TS-700MS/WB Load Quick-Connect Pins
TS-700 SS Load Cell Terminal Block
Operating Guide | TS-700 Series
A-10
Appendix For Scale Technicians
Serial Port Connection
RS232 CONNECTION
TS-700 MS
The RS232 DB9 serial port can be utilized to connect the indicator to a peripheral device such as a printer or remote display, or to interface the indicator with a PC.
NOTE: PC interface requires Triner DataLog PC™ software to be installed on the PC.
Indicator to PC Connection
Operating Guide | TS-700 Series
A-11
Appendix For Scale Technicians
TS-700 SS
The 3-wire terminal block located on the main board can be utilized to connect the indicator to a peripheral device such as a printer or remote display, or to interface the indicator with a PC.
NOTE: PC interface requires Triner DataLog PC™ software to be installed on the PC.
TS-700 SS RS-232 Terminal Block
TS-700 WB
The RS232 3-pin quick-connect serial port can be utilized to connect the indicator to a peripheral device such as a printer or remote display, or to interface the indicator with a PC.
NOTE: PC interface requires Triner DataLog PC™ software to be installed on the PC.
Operating Guide | TS-700 Series
A-12
Appendix For Scale Technicians
Data Output Formats
REMOTE DISPLAY FORMAT
State A
Bits: 0,1,2
0 1 2
Decimal point
position
1 0 0
XXXXXX0
0 1 0
XXXXXXX
1 1 0
XXXXX-X
0 0 1
XXXX-XX
1 0 1
XXX-XXX
Bits: 3,4
Division
0
1
X1
1
0
X2
State B
BITS
FUNCTION
Bits0
Gross = 0, net=1
Bits1
Symbol: positive = 0,negative =1
Bits2
Overload (or under zero)=1
Bits3
Dynamic = 1
Bits4
Unit: lb=0, kg=1
Bits5
Constant 1
Bits6
Constant 0
State C
Bit2
Bit1
Bit0
unit
0 0 0
Kg or lb
0 0 1
g
0 1 0
t
Bit 3
printing=1
Bit 4
Extend
display=1
Bit 5
Constant 1
Bit 6
Constant 0
Operating Guide | TS-700 Series
A-13
Appendix For Scale Technicians
PC COMPUTER CONTINUOUS SENDING FORMAT
, ,
CR
L F
S 1 S 2 S 3 Data S 4
S1: Weight status:
ST = standstill US = not standstill OL = overload
S2: Weight mode:
GS = gross mode NT = net mode
S3: Weight of positive and negative, “+” or ” –“
S4: “kg” or “lb”
Data: Weight value, including decimal point
CR: Carriage return
LF: Line feed
Operating Guide | TS-700 Series
A-14
Appendix For Scale Technicians
SERIAL INTERFACE RECEPTION COMMAND
RS232COM serial interface can receive simple ASCII command.
Command word and role as follows:
Command
Name
Function
T
TARE
Save and clear tare
Z
ZERO
Zero gross weight
P
PRINT
Print the weight
R
G.W/N.W
Read gross weight or net weight
C
Kg/lb
Kg/lb conversion
G
G.W
Check gross weight at net weight mode
Data Format
Print Format
ID.NO. ....................004 (Serial No.)
Date:.......................XX.XX. XX (yy.mm.dd)
Time: ......................XX.XX.XX (hh.mm.ss)
GROSS ..................NNNNNN lb/kg
TARE......................NNNNNN lb/kg
NET ........................NNNNNN lb/kg
Operating Guide | TS-700 Series
A-15
Appendix For Scale Technicians
Error Codes
ERROR
CAUSE
SOLUTION
uuuuuu
1. Overloaded
condition.
2. Wrong
connection at load cell.
3. Load cell has
quality problem.
1. Reduce the weight.
2. Check load cell
connection
3. Inspect load cell. Check
the input and output.
nnnnnn
1. Calibration is
incorrect.
2. Bad connections.
3. Load cell has
quality problem.
1. Check scale is resisted
or not, foot is kept level or not.
2. Check load cell
connection.
3. Check input and output
resistance of load cell.
ERR1
Overweight calibration error. Incorrect weight input or incorrect weight on weighing platform.
1. Input weight correctly
during calibration.
2. Place correct weight on
weighing platform.
ERR2
Underweight calibration error. Incorrect weight input or incorrect weight on weighing platform.
The calibration weights Minimum is 10% of Max. cap. Recommend 60%­80% of Max. Cap.
ERR3
During calibration, the input signal is negative.
Check all connections. Check load cell. If connections and load cell are okay, PCB needs replacing.
Operating Guide | TS-700 Series
A-16
Appendix For Scale Technicians
ERROR
CAUSE
SOLUTION
ERR4
During calibration, the signal is unstable.
After the platform is stable, start calibration.
ERR5
Circuitry error
Replace the PCB.
Factory Default Settings
PARAMETER
INSTRUCTION
DEFAULT
C01
Calibration
1
C02
Decimal digits
0
C03
Resolution
1
C04
Max. capacity
10000
C05
Empty calibration
0
C06
Capacity calibration
0
C07
Restore default
0
C08
Warning tone
1
C09
Power-off automatically
0
C10
Power saving mode
0
C11
Hold function
0
C12
Disable kg/lb conversion
1
Operating Guide | TS-700 Series
A-17
Appendix For Scale Technicians
PARAMETER
INSTRUCTION
DEFAULT
C13
Upper limit alarm
000000
C14
Under limit alarm
000000
C15
Inner code
----
C16
Date setting
----
C17
Time setting
----
C18
Serial interface data output
0
C19
Serial interface Baud rate
3(9600)
C20
Zero manually
10
C21
Initial zero
10
C22
Zero tracking range
0
C23
Zero tracking time
1
C24
Overload range
9
C25
Negative range
10
C26
Standstill time
1
C27
Standstill range
2
C28
Dynamic filter
0
C29
Noise filter
2
Operating Guide | TS-700 Series
A-18
Appendix For Scale Technicians
PARAMETER
INSTRUCTION
DEFAULT
C30
Print format
0
C31
Analog signal options
1
C32
4~20mA testing
4
C33
Relay output setting
1
C34
Reserved
0
C35
Reserved
6
C36
Calibration location gravity
9.7936
C37
Destination gravity
9.7936
C38
Version No.
----
C39
Reserved menu
----
Triner Scale & Manufacturing Co. Inc.
8411 Hacks Cross Road Olive Branch, MS 38654
(800) 238-0152
www.trinerscale.com