You are cautioned that changes or modifications not espressly
approved in this document could void yout authority to operate this
equipment.
To reduce the risk of fire or electric shock,do not expose this
apparatus to rain or moisture.
To avoid electrical shock, do not open the cabinet. Refer servicing to qualified personnel only.
NOTENOTE
As the laser is harmful to the eyes, do not attempt to disassemble the cabinet.
Precautions for Use
Use batteries
At the same time, can not use different style or different capacitance batteries. And only charge the rechargeable batteries.
Avoiding condensation problems
As much as possible, avoid sudden temperature changes. Do not attempt to use the drive immediately after moving it
from a cold to a warm location, to raising the room temperature suddenly, as condensation may form with in the drive. If the
temperature changes suddenly while using the drive,
Stop using it and take out batteries for at least an hour.
Storage
When long time no use,must take out the batteries to avoid destroying the device.
1
RR
TriBrer
RR
Standard
Host(with TF card)
Lithium battery
Carrying bag
2
Power adapter
USB cable
TriBrer
Description
24
5 6
25
23
21 22
8
7
9
16
17
19
20
18
13
14
15
1
2
3
4
10
11
12
USB interface
1
TF(micro SD) card slot
2
Power adapter socket
3
Charge indicator
4
Menu button
5
6
Cursor select button
File operation button
7
8
OPM fast enter button
VFL control button
9
Zoom control button
10
Full screen button
11
Power button
12
Cancel button
13
Real-time measurement button
14
Average measurement button
15
Up button
16
Right button
17
Down button
18
Left button
19
Confirming button
20
OTDR optic fiber connector(1310/1550nm)
21
OTDR optic fiber connector(1625nm)
22
VFL optic fiber connector
23
Anti-dust cover
24
25
LCD
RR
3
TriBrer
RR
On/off
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
File
ENTER
Button is used to turn on/off the device. Keep pressing it for 2 seconds
to turn on device. Short press it again to turn off the device.
waveform vertically. Keep pressing , then press Left/Right button to zoom in
or out horizontally. Press button to return to full screen display. Notice: the
focus of zooming is the location of the activated cursor.
9
TriBrer
RR
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
File
ENTER
Menu
Measure Settings
File
Open...
Format SD Card...
Press key MENU to switch the menu tag.
OPM
VFL
ESC
REAL
SCAN
System Settings
About
System Settings
Measure Settings
FileAbout
Wavelength
Scan Mode
Range
File
Pulse Width
Average Time
Splice Loss
Reflection Threshold
End Threshold
Refractive Rate
Scatter Coefficient
Press key ENTER to edit settings.
1550
Manual
80
Measure Settings
10000
Auto Power Off
120
Auto Name
0.05
Language/
65.0
brightness
2.0
Firmware Version:
Date
1.46832
Time
52.1
Length Unit
Firmware Version:
Factory Settings
Firmware Version:
Firmware Version: A4B75978
Language:
File
System Settings
nm
Km
System Settings
ns
10min
s
On
dB
English
Measure Settings
dB
60
dB
2013
2157
dB
Metric
No
10
About
System Settings
TEN1A20V35_MAIN_V1.2
1102
TEN1A20V35_ANALOG_V0.3
:
TEN1A20V35_TFT_V1.1
SEN1A20V35_MAIN_V1.2.0
FEN1A20V35_MAIN_V1.x.0
English
There are four pages of menu
which used to configure parameters.
Under waveform interface, press
MENU
button to switch the four
About
menus cyclically.
The four menus are:
1. File menu
2. Measure settings menu
3. System settings menu
4. About menu
Measure Settings System Settings
File
Open...
Format SD Card...
About
Menu-File menu
File
20140615
20140617
20140618
201406192014-06-19
20140622
Measure Settings
System Settings
TriBrer
About
2014-06-15
2014-06-17
2014-06-18
2014-06-22
RR
Press key MENU to switch to the menu tag.
Warnning
Enter File menu, there are two options,
Open and Formatting SD card. Select one
option and press button to enter
ENTER
the corresponding operation.
Attention: formatting SD card operation will delete all files and cannot recover, please be cautious.
Are you sure to format the SD card?
OK
11
Cancel
TriBrer
RR
Menu-File menu
File
20140615
20140617
20140618
20140619
20140620
Measure settings
File
20140615
20140617
20140618
20140619
20140620
System settings
Measure Settings
About
2014-06-15
2014-06-17
2014-06-18
2014-06-19
2014-06-20
System Settings
About
2014-06-15
2014-06-17
2014-06-18
2014-06-19
2014-06-20
12
Select Open option, there are folders. The folder
name is the date of the file saved, which is generated
automatically by system. Files measured in the same
day will be stored in the same folder. Folder can only
be delete, but not rename. And must delete all files
before delete the folder.
Press Up or Down button to select a folder and
press button to open delete tip. Then press
ENTER
button to finish deleting, or press
ESC
or button to cancel.
Menu-File menu
TriBrer
R R
Measure settings
File
.
..
133210_0001.sor
133210_0003.sor
133210_0004.sor
File Info: 1550nm 370m 10ns
File
.
..
133210_0001.sor
133210_0003.sor
133210_0004.sor
File Info 1550nm 370m 10ns
System settings
Measure Settings
About
2014-06-18
2014-06-18133210_0002.sor
2014-06-18
System Settings
2014-06-18
Delete...
Rename...
About
2014-06-18
2014-06-18133210_0002.sor
2014-06-18
2014-06-18
13
Select one folder to display all files inside. Press
Up or Down button to select one file and press
ENTER
button to display it in main interface.
The file information at the bottom is the file's main
measurement setting.
Press button can delete or rename
the file. Filename can make up of 23 digits, alphabets
and special symbols at most. The last four
13
numbers_xxxx is generated by Automatic naming
function. Shut down this function will not generate.
TriBrer
RR
Menu-File menu
Save
File Name:
140950
CapsbackSave
abcde
hijklmn
opqrs
vwxyz
File Name:
140950
abcde
hijklmn
opqrs
vwxyz
f
t
CapsbackSave
.
g
u
7
Rename
4
12
_
f
t
.
89
5
0
g
u
6
3
#
7
4
12
_
89
5
0
14
When finish measuring an optic fiber, press
button to enter storage interface. Default filename is
hour/minute/second of first time storage. Press
Up/Down/Left/Right button to select alphabets, digits
and symbols on soft keyboard. You can input 23
alphabets at most. If Automatic naming is enabled, the
filename will automatically generated with four digits.
Without shutting down the device, the following stored
file's name will automatically plus 1. The file format is
6
.SOR.
3
On file recalling interface, press Right button can
#
modify filename, as the same operations above.
File
Menu-Measure settings
TriBrer
RR
FileMeasure Settings System SettingsAbout
Wavelength
Scan Mode
Range
Pulse Width
Average Time
Splice Loss
Reflection Threshold
End Threshold
Refractive Rate
Scatter Coefficient
Splice loss---treat as an event when the loss is higher than setting value.
Reflection threshold---treat as an event when the reflection is higher than setting value.
End threshold---treat as the end of optic fiber when the loss is higher than setting value.
Refractive rate---represent the average refractive index of entire optic fiber.
Scatter coefficient--- the intrinsic value of Rayleigh Scattering.
1550
Manual
40
1000
120
0.05
65.0
2.0
1.46832
52.1
nm
Km
ns
s
dB
dB
dB
dB
Measure settings menu is used to set relative measurement data, which
the judgment of events list is based on. Wrong setting might leads to wrong or
missing events.
Wavelength---wavelength of laser.
Scan mode---manual and auto mode. Under auto mode, it will match the
distance, range and pulse width.
Range---match with the length of measured optic fiber, usually over one level.
Pulse width---set the pulse width of output laser. Usually, small pulse width can
measure close event, large pulse width can measure remote distance, but
enlarge event's blind area.
Average time--- can set between 5 second and 180 second as average
measurement time.
15
TriBrer
RR
Menu-System settings
FileMeasure Settings
Auto Power Off
Auto Name
Optical Dectector
RT AnalyseOff
Brightness
Date
Time
Length Unit
Factory Settings
Press key ENTER to edit settings.
System Settings
10min
On
On
60
2013
2157
:
Metric
No
1102
About
System settings menu is used to set the device's basic functions.
Auto Power Off--- set time of automatic off or cancel this function.
Auto Name---name automatically when file is saving, can cancel this
function.
Optical Detector --- detect whether there is signal in optic fiber or
not before measuring.
RT Analyse --- set whether analyses events after real-time measuring
or not.
Brightness--- adjust brightness of LCD.
Date--- set year, month, day.
Time ---set hour, minute.
Length unit --- default is metric units.
Factory settings --- is used in resetting to default , and has no effect
on The set time and stored data.
16
Menu- About
TriBrer
RR
Measure Settings
File
Firmware Version:
Firmware Version:
Firmware Version:
Firmware Version: A4B75978
Language:
System Settings
TEN1A20V35_MAIN_V1.2
TEN1A20V35_ANALOG_V0.3
TEN1A20V35_TFT_V1.1
SEN1A20V35_MAIN_V1.2.0
FEN1A20V35_MAIN_V1.x.0
English
About
This menu is version details about device's configuration of
hardware and software. CPU number is used to generate upgrading
code, which should inform distributor or factory before software
upgrading.
17
TriBrer
RR
OPM function
Optical Power Meter
Power
Press key Down to change
the wavelength.
Press key Enter to start or
to stop measure.
+00.00
000.00
1310Onnm
dBm
mw
Press button on device with OPM function to enter
OPM measurement interface. Press button to start
OPM
ENTER
measuring. Press button to change the wavelength. The
meter's six calibration wavelengths are: 850nm, 1300nm, 1310nm,