Tribier AOR500-P User Manual

TriBrer
AOR500-P
PON Optical Time Domain Reflectometer
USER'S GUIDE
WARNINGWARNING
You are cautioned that changes or modifications not espressly approved in this document could void yout authority to operate this equipment. To reduce the risk of fire or electric shock,do not expose this apparatus to rain or moisture. To avoid electrical shock, do not open the cabinet. Refer servicing to qualified personnel only.
NOTENOTE
As the laser is harmful to the eyes, do not attempt to disassemble the cabinet.
Precautions for Use
Use batteries
At the same time, can not use different style or different capacitance batteries. And only charge the rechargeable batteries.
Avoiding condensation problems
As much as possible, avoid sudden temperature changes. Do not attempt to use the drive immediately after moving it
from a cold to a warm location, to raising the room temperature suddenly, as condensation may form with in the drive. If the temperature changes suddenly while using the drive, Stop using it and take out batteries for at least an hour.
Storage
When long time no use,must take out the batteries to avoid destroying the device.
1
RR
TriBrer
RR
Standard
Host(with TF card)
Lithium battery
Carrying bag
2
Power adapter
USB cable
Description
24
5 6
25
23
21 22
8
7
9
16
17
19
20
18
13
14
15
1
2
3
4
10
11
12
USB interface
1
TF(micro SD) card slot
2
Power adapter socket
3
Charge indicator
4
Menu button
5 6
Cursor select button File operation button
7 8
OPM fast enter button VFL control button
9
Zoom control button
10
Full screen button
11
Power button
12
Cancel button
13
Real-time measurement button
14
Average measurement button
15
Up button
16
Right button
17
Down button
18
Left button
19
Confirming button
20
OTDR optic fiber connector(1310/1550nm)
21
OTDR optic fiber connector(1625nm)
22
VFL optic fiber connector
23
Anti-dust cover
24 25
LCD
RR
3
TriBrer
RR
On/off
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
File
ENTER
Button is used to turn on/off the device. Keep pressing it for 2 seconds
to turn on device. Short press it again to turn off the device.
Press button to
start measuring. But user should
SCAN
2012-01-01 WED 21:25
*
modify measurement setting by
OPM
VFL
ESC
REAL
SCAN
real requirement before test start.
A-B: ---.---Km --.--dB --.--dB/Km Km/divA-B: ---.---Km --.--dB --.--dB/Km Km/div
No. Distance Loss
Reflection
dB/Km
C.Loss
File Info:
4
2012-01-01 WED 21:25
Date and time
There are 10x10 grids in waveform region.
Horizontal grids mean distance, vertical grids
mean dB value. As shown in figure, each
horizontal grid represents 0.2km, each vertical
grid represents 4dB.
Measurement interface
2012-01-01 WED 21:25
40.00
36.00
A:0.000Km
32.00
B:0.000Km
28.00
24.00
20.00
16.00
12.00
8.00
4.00
0.00
A-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/divA-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/div
No. Distance Loss
1/3 0.000 0.00 -28.00 --.-- 0.00
Measure Info: 1550nm 2.0Km 10ns
Reflection
174622_0001.sor
C.Loss
dB/Km
Power
174622_0001.sor
Events list
File name
Waveform region
RR
A-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/div
The distance and Attenuation between A and B cursor
Every horizontal grid is 0.2Km
5
Measure info: 1550nm 2.0Km 10ns
Measurement setting info
TriBrer
RR
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
File
ENTER
Real-time and average measurement
Real-time measurement can quickly judge basic faults of optical fiber. Press
REAL
button to start real-time measuring. During measuring, you can change
range, zoom in or zoom out . Press button again or button to
REAL
stop. The device will not analyses event after real-time measurement in default.
Unless you turn on RT analyse in System settings, the device will analyse events
OPM
VFL
ESC
REAL
SCAN
according to the last refreshed waveform.
Average measurement can judge the line condition more accurate. It can get a
better SNR and fits high requirement circuit. Press button to start. User can
SCAN
set measurement time from 5 second to 180 second. The device analyses events and
generates event list automatically. Press button during measuring, device
escesc
will stop measuring, analyses events and generates event list automatically.
6
escesc
Event list
RR
No. Dis. Loss
1/3 0.000 0.00 -28.00 --.-- 0.00
Ref.
dB/Km
C.Loss
Event list on main interface
Event List
No.
1/6 2/6
3/6
4/6 5/6 6/6
Press key ENTER to locate the corresponding event.
Dis.
0.000
0.940
4.301
5.589
18.712
39.809
-0.05
Loss
0.00
0.08
0.12
-.--
-.--
Ref.
-51.74
-51.74
-54.67
-.--
-41.03
-46.02
dB/Km
-.--
0.24
0.28
-.--
0.22
0.21
C.Loss
0.00
0.26
1.18
4.07
9.41
Event List
-.--
After measurement or open a waveform in memory, there is
event list on the bottom of waveform interface. Press
ENTER
button to show the whole event list. Six types of events as
followed:
Optic fiber start
Reflection event
Attenuation event
Mix event
Gain event
optic fiber ending
Press Up or Down button to select an event which needs to be
located by cursor on the waveform. Then press button
ENTER
to return to waveform interface. The cursor will stay on the
position of the selected event. Press button to return
ESC
to waveform interface.
7
TriBrer
RR
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
File
ENTER
Real-time measurement tips
The device supports user to change measurement range during real-time
measurement. Start Real-time measuring function and press Up or Down button
to increase or decrease the range. And it also supports to change the cursor position
and zoom in/out waveform in realtime, which means you can zoom in partial
waveform while measuring to judge the network fault.
OPM
VFL
ESC
REAL
SCAN
8
Cursor selection and waveform zoom in/out
RR
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
File
ENTER
This device has two cursors, A and B.
Default distance is 0m, activated cursor
turned to bright red. Press
Marker
button switch to another cursor. Press Left
or Right button to move cursor. Press Up or
OPM
VFL
ESC
REAL
SCAN
Down button to previous or next event. You
can calculate the distance and attenuation
between two cursors.
Keep pressing , then press and Up/Down button to zoom in or out
2012-01-01 WED 21:25
40.00
36.00
A:0.000Km
32.00
28.00
24.00
20.00
16.00
12.00
8.00
4.00
0.00
A-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/divA-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/div
No. Distance Loss
1/3 0.000 0.00 -28.00 --.-- 0.00
Measure Info: 1550nm 2.0Km 10ns
B:0.583Km
Reflection
dB/Km
C.Loss
*
waveform vertically. Keep pressing , then press Left/Right button to zoom in
or out horizontally. Press button to return to full screen display. Notice: the
focus of zooming is the location of the activated cursor.
9
TriBrer
RR
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
File
ENTER
Menu
Measure Settings
File
Open... Format SD Card...
Press key MENU to switch the menu tag.
OPM
VFL
ESC
REAL
SCAN
System Settings
About
System Settings
Measure Settings
File About
Wavelength
Scan Mode
Range
File
Pulse Width
Average Time
Splice Loss
Reflection Threshold
End Threshold
Refractive Rate
Scatter Coefficient
Press key ENTER to edit settings.
1550
Manual
80
Measure Settings
10000
Auto Power Off
120
Auto Name
0.05
Language/
65.0
brightness
2.0
Firmware Version:
Date
1.46832
Time
52.1
Length Unit
Firmware Version:
Factory Settings
Firmware Version:
Firmware Version: A4B75978
Language:
File
System Settings
nm
Km
System Settings
ns
10min
s
On
dB
English
Measure Settings
dB
60
dB
2013
21 57
dB
Metric
No
10
About
System Settings
TEN1A20V35_MAIN_V1.2
11 02
TEN1A20V35_ANALOG_V0.3
:
TEN1A20V35_TFT_V1.1
SEN1A20V35_MAIN_V1.2.0
FEN1A20V35_MAIN_V1.x.0
English
There are four pages of menu
which used to configure parameters.
Under waveform interface, press
MENU
button to switch the four
About
menus cyclically.
The four menus are:
1. File menu
2. Measure settings menu
3. System settings menu
4. About menu
Measure Settings System Settings
File
Open...
Format SD Card...
About
Menu-File menu
File
20140615 20140617 20140618
20140619 2014-06-19
20140622
Measure Settings
System Settings
About
2014-06-15 2014-06-17 2014-06-18
2014-06-22
RR
Press key MENU to switch to the menu tag.
Warnning
Enter File menu, there are two options,
Open and Formatting SD card. Select one
option and press button to enter
ENTER
the corresponding operation.
Attention: formatting SD card operation will delete all files and cannot recover, please be cautious.
Are you sure to format the SD card?
OK
11
Cancel
TriBrer
RR
Menu-File menu
File
20140615 20140617 20140618
20140619
20140620
Measure settings
File
20140615 20140617 20140618
20140619
20140620
System settings
Measure Settings
About
2014-06-15 2014-06-17 2014-06-18
2014-06-19
2014-06-20
System Settings
About
2014-06-15 2014-06-17 2014-06-18
2014-06-19
2014-06-20
12
Select Open option, there are folders. The folder
name is the date of the file saved, which is generated
automatically by system. Files measured in the same
day will be stored in the same folder. Folder can only
be delete, but not rename. And must delete all files
before delete the folder.
Press Up or Down button to select a folder and
press button to open delete tip. Then press
ENTER
button to finish deleting, or press
ESC
or button to cancel.
Menu-File menu
R R
Measure settings
File
. .. 133210_0001.sor
133210_0003.sor 133210_0004.sor
File Info: 1550nm 370m 10ns
File
. .. 133210_0001.sor
133210_0003.sor 133210_0004.sor
File Info 1550nm 370m 10ns
System settings
Measure Settings
About
2014-06-18
2014-06-18133210_0002.sor
2014-06-18
System Settings
2014-06-18
Delete...
Rename...
About
2014-06-18
2014-06-18133210_0002.sor
2014-06-18 2014-06-18
13
Select one folder to display all files inside. Press
Up or Down button to select one file and press
ENTER
button to display it in main interface.
The file information at the bottom is the file's main
measurement setting.
Press button can delete or rename
the file. Filename can make up of 23 digits, alphabets
and special symbols at most. The last four
13
numbers_xxxx is generated by Automatic naming
function. Shut down this function will not generate.
TriBrer
RR
Menu-File menu
Save
File Name:
140950
Caps back Save
a b c d e
h i j k l m n
o p q r s
v w x y z
File Name:
140950
a b c d e
h i j k l m n
o p q r s
v w x y z
f
t
Caps back Save
.
g
u
7
Rename
4
1 2
_
f
t
.
8 9
5
0
g
u
6
3
#
7
4
1 2
_
8 9
5
0
14
When finish measuring an optic fiber, press
button to enter storage interface. Default filename is
hour/minute/second of first time storage. Press
Up/Down/Left/Right button to select alphabets, digits
and symbols on soft keyboard. You can input 23
alphabets at most. If Automatic naming is enabled, the
filename will automatically generated with four digits.
Without shutting down the device, the following stored
file's name will automatically plus 1. The file format is
6
.SOR.
3
On file recalling interface, press Right button can
#
modify filename, as the same operations above.
File
Menu-Measure settings
RR
File Measure Settings System Settings About
Wavelength
Scan Mode
Range
Pulse Width
Average Time
Splice Loss
Reflection Threshold
End Threshold
Refractive Rate
Scatter Coefficient
Splice loss---treat as an event when the loss is higher than setting value.
Reflection threshold---treat as an event when the reflection is higher than setting value.
End threshold---treat as the end of optic fiber when the loss is higher than setting value.
Refractive rate---represent the average refractive index of entire optic fiber.
Scatter coefficient--- the intrinsic value of Rayleigh Scattering.
1550
Manual
40
1000
120
0.05
65.0
2.0
1.46832
52.1
nm
Km
ns
s
dB
dB
dB
dB
Measure settings menu is used to set relative measurement data, which
the judgment of events list is based on. Wrong setting might leads to wrong or
missing events.
Wavelength---wavelength of laser.
Scan mode---manual and auto mode. Under auto mode, it will match the distance, range and pulse width.
Range---match with the length of measured optic fiber, usually over one level.
Pulse width---set the pulse width of output laser. Usually, small pulse width can measure close event, large pulse width can measure remote distance, but enlarge event's blind area.
Average time--- can set between 5 second and 180 second as average measurement time.
15
TriBrer
RR
Menu-System settings
File Measure Settings
Auto Power Off
Auto Name
Optical Dectector
RT Analyse Off
Brightness
Date
Time
Length Unit
Factory Settings
Press key ENTER to edit settings.
System Settings
10min
On
On
60
2013
21 57
:
Metric
No
11 02
About
System settings menu is used to set the device's basic functions.
Auto Power Off--- set time of automatic off or cancel this function.
Auto Name---name automatically when file is saving, can cancel this function.
Optical Detector --- detect whether there is signal in optic fiber or not before measuring.
RT Analyse --- set whether analyses events after real-time measuring or not.
Brightness--- adjust brightness of LCD.
Date--- set year, month, day.
Time ---set hour, minute.
Length unit --- default is metric units.
Factory settings --- is used in resetting to default , and has no effect on The set time and stored data.
16
Menu- About
RR
Measure Settings
File
Firmware Version:
Firmware Version:
Firmware Version:
Firmware Version: A4B75978
Language:
System Settings
TEN1A20V35_MAIN_V1.2
TEN1A20V35_ANALOG_V0.3
TEN1A20V35_TFT_V1.1
SEN1A20V35_MAIN_V1.2.0
FEN1A20V35_MAIN_V1.x.0
English
About
This menu is version details about device's configuration of
hardware and software. CPU number is used to generate upgrading
code, which should inform distributor or factory before software
upgrading.
17
TriBrer
RR
OPM function
Optical Power Meter
Power
Press key Down to change the wavelength.
Press key Enter to start or to stop measure.
+00.00
000.00
1310Onnm
dBm
mw
Press button on device with OPM function to enter
OPM measurement interface. Press button to start
OPM
ENTER
measuring. Press button to change the wavelength. The
meter's six calibration wavelengths are: 850nm, 1300nm, 1310nm,
1490nm, 1550nm, 1625nm.
Press button to quit OPM function.
ESC
18
VFL
RR
2012-01-01 WED 21:25
40.00
36.00
A:0.000Km
32.00
B:0.000Km
28.00
24.00
20.00
16.00
12.00
8.00
4.00
0.00
A-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/divA-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/div
No.
Distance
Loss
Reflection
dB/Km
C.Loss
1/3 0.000 0.00 -28.00 --.-- 0.00
Measure Info; 1550nm 2.0Km 10ns
This device has VFL function, the output power is about 1nW.
*
Press button to control red laser on-flashing-off. When
VFL
Laser is on, the prompt icon will occur at upper right corner.
VFL laser output connector
19
TriBrer
RR
TF card and USB communication
All measured waveforms are stored in TF card. Using
mini USB cable can store data to PC. PC will show a new
disk icon, data inside. According to filename, user can
store or recall files.
2012-01-01 WED 21:25
40.00
36.00
A:0.000Km
32.00
B:0.000Km
28.00
24.00
20.00
16.00
12.00
8.00
4.00
0.00
A-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/divA-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/div
No.
1/3 0.000 0.00 -28.00 --.-- 0.00
Measure Info: 1550nm 2.0Km 10ns
Distance
Loss
Reflection
dB/Km
*
C.Loss
20
User upgrading
RR
Measure Settings
File
Firmware Version:
Firmware Version:
Firmware Version:
Firmware Version: A4B75978
Language:
TEN1A20V35_MAIN_V1.2
TEN1A20V35_ANALOG_V0.3
TEN1A20V35_TFT_V1.1
SEN1A20V35_MAIN_V1.2.0
FEN1A20V35_MAIN_V1.x.0
English
System Settings
About
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
File
ENTER
When device needs upgrading, user does not need to send it
back to factory. Prepare materials before upgrading as followed:
1. CPU ID
2. TF(micro SD) card, capacity is less than 16GB
Inform the CPU ID to distributor or factory, you can get
upgrading firmware only for your device. Store the firmware to TF
ard through computer. Insert TF card to device when it is power off.
MENU
OPM
VFL
ESC
REAL
SCAN
Press and button and hold them, then press
button to turn on the device, upgrading is starting. You
can delete upgrading file in TF card when finish upgrading.
SCAN
21
TriBrer
RR
Charging
2012-01-01 WED 21:25
40.00
36.00
A:0.000Km
32.00
B:0.000Km
28.00
24.00
20.00
16.00
12.00
8.00
4.00
0.00
A-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/divA-B: 000.000Km 00.00dB 00.00dB/Km 0.200Km/div
No.
Distance
Reflection
dB/Km
C.Loss
Loss
1/3 0.000 0.00 -28.00 --.-- 0.00
Measure Info: 1550nm 2.0Km 10ns
This device has lithium batteries inside, and can only use the
*
power adapter from factory to charge it. Insert the adapter to
device, and finish charging in 8-10 hours. Red charge indicator
means charging, while green
means finish.
Optical Time Domain ReflectometerOptical Time Domain Reflectometer
CHARGE
MENU
Marker
ENTER
22
File
OPM
VFL
ESC
REAL
SCAN
Dynamic range
Optic fiber type Wavelength
Optic fiber connector
LCD Test range
Pulse width
Measurement time
Attenuation blind area Event blind area Distance accuracy
Data storage
Communication connector VFL power power
Battery lifetime Working temperature
Storage temperature Relative humidity
Weight
Appearance size
Detail parameters
PB-3SN25: 35/34/34dB PC-3CN25: 38/37/37dB SM optical fiber
1550nm 20nm/1310nm 20nm/1625nm 20nm FC/PC(1310/1550nm),FC/APC(1625nm)interchange able adapter(Optional SC/ST/LC)
3.5 inch color LCD
PB-3SN25: 120Km PC-3CN25: 140Km
10ns~10us
5s~180s 8m@Range 2Km,Pulse width=10ns
1.5m@Range 2Km,Pulse width=10ns
1.25m Micro SD card, less than 16GB
USB
1mW Special lithium battery packet
Standby>15 hour, measurement>8 hour
0 ~-50
-20 ~+70 <90%
0.75Kg
197mmx107mmx67mm
23
RR
TriBrer
RR
TriBrer
RR